ThermoChem HT-2000 System and Intraperitoneal Hyperthermia (IPH) Procedure Kit (HT-2000)
Applicant
Thermasolutions, LLC
Product Code
MLW · Gastroenterology, Urology
Decision Date
May 5, 2025
Decision
SESK
Submission Type
Traditional
Regulation
21 CFR 876.5630
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The intended use of the ThermoChem HT-2000 System is to raise the core temperature of the peritoneum to a desired target temperature, in patients who require warming, including oncology patients, by continuously lavaging with circulated and warmed Lactated Ringer's Solution, U.S.P., or other physician specified fluids.
Device Story
ThermoChem HT-2000 System consists of reusable ThermoChem unit with touchscreen monitor and single-use sterile Intraperitoneal Hyperthermia (IPH) Procedure Kit. Device functions by heating and circulating sterile fluids (e.g., Lactated Ringer's Solution) to raise peritoneal core temperature to physician-selected target. Operated by clinicians in clinical settings for therapies such as continuous hyperthermic peritoneal perfusion. Touchscreen interface allows user control of system functions. Output is warmed, circulating fluid lavage; clinical decision-making relies on physician selection of fluids and therapy parameters. Benefits include controlled temperature-targeted warming of peritoneum for oncology patients.
Clinical Evidence
No clinical performance testing was conducted. Substantial equivalence is supported by bench testing data from the predicate device (K131583).
Technological Characteristics
System comprises reusable unit with touchscreen monitor and single-use disposable IPH kit. Operates via fluid heating and circulation. No changes in materials, design, or energy source compared to predicate. Software interface provides user control.
Indications for Use
Indicated for patients requiring peritoneal warming, including oncology patients, via continuous lavage with warmed Lactated Ringer's Solution or other physician-specified fluids.
Regulatory Classification
Identification
(1) A peritoneal dialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions, and that consists of a peritoneal access device, an administration set for peritoneal dialysis, a source of dialysate, and, in some cases, a water purification mechanism. After the dialysate is instilled into the patient's peritoneal cavity, it is allowed to dwell there so that undesirable substances from the patient's blood pass through the lining membrane of the peritoneal cavity into this dialysate. These substances are then removed when the dialysate is drained from the patient. The peritoneal dialysis system may regulate and monitor the dialysate temperature, volume, and delivery rate together with the time course of each cycle of filling, dwell time, and draining of the peritoneal cavity or manual controls may be used. This generic device includes the semiautomatic and the automatic peritoneal delivery system.(2) The peritoneal access device is a flexible tube that is implanted through the abdominal wall into the peritoneal cavity and that may have attached cuffs to provide anchoring and a skin seal. The device is either a single use peritioneal catheter, intended to remain in the peritoneal cavity for less than 30 days, or a long term peritoneal catheter. Accessories include stylets and trocars to aid in the insertion of the catheter and an obturator to maintain the patency of the surgical fistula in the abdominal wall between treatments. (3) The disposable administration set for peritoneal dialysis consists of tubing, an optional reservoir bag, and appropriate connectors. It may include a peritoneal dialysate filter to trap and remove contaminating particles. (4) The source of dialysate may be sterile prepackaged dialysate (for semiautomatic peritoneal dialysate delivery systems or “cycler systems”) or dialysate prepared from dialysate concentrate and sterile purified water (for automatic peritoneal dialysate delivery systems or “reverse osmosis” systems). Prepackaged dialysate intended for use with either of the peritoneal dialysate delivery systems is regulated by FDA as a drug.
Special Controls
*Classification.* Class II (special controls). The following accessories are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9: A catheter finger grip that is non-patient contacting and intended for single use with a peritoneal catheter; a continuous ambulatory peritoneal dialysis (CAPD) belt; and a catheter stand that does not include weigh scales.
{0}
[LOGO]
FDA
U.S. FOOD & DRUG
ADMINISTRATION
May 5, 2025
ThermaSolutions, LLC
% Jessica Czamanski
Regulatory, Quality & Compliance Consultant
DuVal & Associates, P.A.
Suite 1820, Medical Arts Building
825 Nicollet Mall
Minneapolis, Minnesota 55402
Re: K240447
Trade/Device Name: ThermoChem HT-2000 System and Intraperitoneal Hyperthermia (IPH) Procedure Kit (HT-2000)
Regulation Number: 21 CFR 876.5630
Regulation Name: Peritoneal dialysis system and accessories
Regulatory Class: Class II
Product Code: MLW
Dated: March 5, 2025
Received: March 6, 2025
Dear Jessica Czamanski:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit/tray. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K240447 - Jessica Czamanski
Page 2
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-
{2}
K240447 - Jessica Czamanski
Page 3
devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Maura Rooney -S
Maura Rooney
Assistant Director
DHT3A: Division of Renal, Gastrointestinal, Obesity, and Transplant Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K240447
Device Name
ThermoChem HT-2000 System and Intraperitoneal Hyperthermia (IPH) Procedure Kit (HT-2000)
Indications for Use (Describe)
The intended use of the ThermoChem HT-2000 System is to raise the core temperature of the peritoneum to a desired target temperature, in patients who require warming, including oncology patients, by continuously lavaging with circulated and warmed Lactated Ringer's Solution, U.S.P., or other physician specified fluids.
Type of Use (Select one or both, as applicable)
☑
Prescription Use (Part 21 CFR 801 Subpart D)
☐
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
**DUVAL & ASSOCIATES**
Drug, Device and Food Law
1820 Medical Arts Building
825 Nicollet Mall
Minneapolis, MN 55402
612-338-7170 Main
# **K240447 – 510(k) Summary**
# **I. Submitter**
Submitter Name: ThermaSolutions LLC
Contact Person: Rhea Johny
Regulatory Specialist
Address: 1889 Buerkle Road
White Bear Lake MN 55110
United States
Telephone: (651) 374-9728
Email: rjohny@thermasolutions.com
Date of Preparation: May 5th, 2025
# **II. Application Correspondent**
Contact's Name: DuVal & Associates, P. A.
Contact Person: Jessica Czamanski
Address: Medical Arts Building
825 Nicollet Mall
Suite 1820
Minneapolis MN 55402
United States
Telephone: (754) 422-9101
Email: jczamanski@duvalfdalaw.com
# **III. Device**
Trade Name: ThermoChem HT-2000 System and Intraperitoneal Hyperthermia (IPH)
Procedure Kit (HT-2000)
Common Name: Peritoneal dialysis system and accessories
Classification Name: Warmer, Peritoneal Dialysate
Product Classification: Class II
Regulation Number: 21 CFR 876.5630
Product Code: MLW
# **IV. Predicate Device**
Manufacturer: ThermaSolutions LLC
Device Name: ThermoChem HT-2000
*Drug, Device and Food Law*
Page 1 of 3
{5}
DUVAL & ASSOCIATES
Drug, Device and Food Law
1820 Medical Arts Building
825 Nicollet Mall
Minneapolis, MN 55402
612-338-7170 Main
510(k) Number: K131583
Product Classification: Class II
Regulation Number: 21 CFR 876.5630
Product Code: MLW
# V. Device Description
The ThermoChem HT-2000 System is composed of the ThermoChem Unit and touchscreen monitor, and the Intraperitoneal Hyperthermia (IPH) Procedure Kit. The ThermoChem HT-2000 and IPH Procedure pack are used to raise the core temperature of the peritoneum to a desired target temperature by continuously lavaging the peritoneum with circulating and warmed sterile solution. The ThermoChem unit and monitor are reusable components of the system. The IPH Procedure kit consists of an array of sterile, single-use, disposable components. The touchscreen monitor on the ThermoChem unit provides for the user interface with the ThermoChem system. The touch screen can be used to control the main functions of the system. The ThermoChem HT-2000 System is comprised of the same hardware as the predicate K31583 device.
The ThermoChem HT-2000 is a tool for the heating and circulation of fluid. Physicians are responsible for selecting the fluids used and the therapy delivered. Oncology patients, undergoing a temperature-targeted warming of the peritoneum with continuous circulation and lavage, may be subject to physician specified fluids for therapies such as continuous hyperthermic peritoneal perfusion. There are no known drugs, other than Lactated Ringer's Solution, U.S.P., specifically approved for these therapies.
# VI. Intended Use/Indications for Use
The intended use of the ThermoChem HT-2000 System is to raise the core temperature of the peritoneum to a desired target temperature, in patients who require warming, including oncology patients, by continuously lavaging with circulated and warmed Lactated Ringer's Solution, U.S.P., or other physician specified fluids.
# VII. Indications for Use Comparison
The proposed indications for use clarify that the patient population includes oncology patients and that the clinical has the responsibility to specify which fluids other than Lactated Ringer's should be used with the device, adding specificity to the indications for use under the device's general intended use.
# VIII. Comparison of Technological Characteristics
The subject ThermoChem System has not undergone any significant changes in materials, design, energy source, or other features compared to the K131583 clearance. The only change that is the
Drug, Device and Food Law
Page 2 of 3
{6}
**DUVAL & ASSOCIATES**
Drug, Device and Food Law
1820 Medical Arts Building
825 Nicollet Mall
Minneapolis, MN 55402
612-338-7170 Main
subject of this 510(k) is the indication for use change which does not result in a change to the previously cleared intended use.
# **IX. Performance Testing – Bench**
All performance data included in K131583 for the predicate ThermoChem HT-2000, also supports the proposed device.
# **Performance Testing – Animal**
This submission does not include any animal performance testing. It was determined that no such testing is required to demonstrate substantial equivalence.
# **Performance Testing – Clinical**
This submission does not include any clinical performance testing. It was determined that no such testing is required to demonstrate substantial equivalence.
# **X. Conclusion**
The proposed ThermoChem HT-2000 has the same intended use, operating principle, fundamental technology, materials, and manufacturing as the predicate device. The difference in indications for use, clarifies the patient population for which the device is intended which is the same population that the predicate device has been used on since its clearance. These differences in indications for use do not introduce new questions for safety and effectiveness. The information provided in this submission demonstrated that the subject device, ThermoChem HT-2000, is substantially equivalent to its predicate, ThermoChem HT-2000 K131583.
---*Drug, Device and Food Law*
Page 3 of 3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.