UltraThineer Zirconia Slurry is indicated for use by dental technicians in the construction of custom-made ceramic restorations. These include copings, crowns, inlays and onlays, veneers, and bridges (3-unit anterior bridges).
Device Story
UltraThineer Zirconia Slurry is a ceramic material used by dental technicians to fabricate custom dental restorations via 3D printing. The slurry, a mixture of zirconia powder and binder, is poured into a 3D printer vat. The printer uses a photocuring lamination process, irradiating the slurry with 420nm UV light to cure the material into a shape defined by an STL file derived from patient dental data. Following printing, the restoration is sintered to achieve final properties. The device is provided in a sealed HDPE bottle. The output is a custom-made ceramic restoration (coping, crown, inlay, onlay, veneer, or bridge) used by dentists to restore patient dentition. The device benefits patients by enabling the production of precise, custom-fitted dental prosthetics.
Clinical Evidence
No clinical testing was performed. Substantial equivalence is supported by non-clinical bench testing and biocompatibility evaluations conducted in accordance with ISO 6872:2015 and ISO 10993-1. Bench testing confirmed compliance with requirements for radioactivity, flexural strength (>300 MPa), linear thermal expansion coefficient, chemical solubility (<100µg/cm²), and shelf-life. Biocompatibility testing included chemical characterization, cytotoxicity, sensitization, oral mucosal irritation, acute systemic toxicity, and material-mediated pyrogenicity.
Technological Characteristics
Ceramic slurry (Zirconia-5Y/BMF-5YZ-1.6B) for 3D printing. Photocuring lamination process using 420nm UV light. Complies with ISO 6872 Type I Class III. Biocompatible per ISO 10993-1. Non-sterile, single-use. Provided in 0.75L HDPE bottle.
Indications for Use
Indicated for dental technicians constructing custom ceramic restorations including copings, crowns, inlays, onlays, veneers, and 3-unit anterior bridges.
Regulatory Classification
Identification
Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 12, 2024
Boston Micro Fabrication % Eric Bannon Sr. VP, Regulatory and Clinical Affairs AlvaMed 935 Great Plain Ave Suite 166 Needham, Massachusetts 02942
Re: K240419
Trade/Device Name: UltraThineer Zirconia Slurry (5Y-B) Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain Powder For Clinical Use Regulatory Class: Class II Product Code: EIH Dated: February 12, 2024 Received: February 13, 2024
Dear Eric Bannon:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Bobak Shirmohammadi -S
For Michael E. Adjodha, M.ChE., RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices
{2}------------------------------------------------
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
K240419
Device Name
UltraThineer Zirconia Slurry (5Y-B)
Indications for Use (Describe)
UltraThineer Zirconia Slurry is indicated for use by dental technicians in the construction of custommade ceramic restorations. These include copings, crowns, inlays and onlays, veneers, and bridges (3-unit anterior bridges).
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### K240419
### 510(K) SUMMARY
#### Name and Address of Submitter
BMF Precision, Inc. 8 Mill & Main St., Suite 310 Maynard, MA 01754 USA Phone: +1 978-637-2050 John Kawola CEO jkawola@bmf3d.com Phone: +1 781-492-3548
#### Correspondent/Primary Contact Person
Eric Bannon Vice President of Requlatory and Clinical Affairs AlvaMed, Inc. ebannon@alvamed.com Phone: +1 (781) 710-8243 Fax: +1 (617) 249-0955
#### Submission Information
Date Summary Prepared: Name of Device: Device Classification Name: Common or Usual Name: Classification: Product Code:
27 January 2023 UltraThineer Zirconia Slurry (5Y-B) Porcelain Powder for Clinical Use Dental Zirconia Material Class II EIH (21 CFR 872.6660)
#### Device Description
The UltraThineer Zirconia Slurry is a ceramic slurry intended for use in fabricating custom-made ceramic restorations using 3D printing and sintering processes.
There is a single formulation:
- Zirconia-5Y/BMF-5YZ-1.6B .
The UltraThineer Zirconia Slurry is provided as a slurry contained in a sealed tamper-proof and opaque HDPE bottle, with product labeling. The bottle containing the slurry has dimensions of 3.5" in diameter and 7.75" in height and has a maximum volume of 0.75L.
For use, the slurry is poured into a vat inside the 3D printer for printing the dental veneer. After printing, the veneer is sintered to yield the final product. Each veneer will be custom fabricated on a named patient prescription basis on the order of a dentist.
#### Predicate Device
INNI-CERA by Aon Co., Ltd (K211308)
{5}------------------------------------------------
### Valid Predicate Discussion
Inni-Cera was selected as the valid predicate device used to support the 510(k) submission because it was cleared using well-established methods. Inni-Cera was the only valid predicate device that could be identified which meets the expected predicate performance. After conducting a search on the Manufacturer and User Facility Device Experience (MAUDE) Database, Medical Device Reporting (MDR), MedSun Reports Database, Medical Device Safety and CBER Safety & Availability (Biologics) and Recall Database websites it was found that there are no known unmitigated use-related or design safety issues and the predicate device has not been subject to a design-related recall.
### Table 1: Valid Predicate Device
| Valid Predicate<br>Device | A - Well<br>established<br>methods | B - Meets or<br>exceeds<br>expected<br>predicate<br>performance | C - Unmitigated<br>use-related or<br>design-related<br>safety issues | D - Associated<br>design-related<br>recall |
|---------------------------|-----------------------------------------------------------------------------|--------------------------------------------------------------------------------------|-------------------------------------------------------------------------------|--------------------------------------------|
| Inni-Cera | Used relevant<br>methods that<br>were published<br>in the public<br>domain. | History of Safe<br>use, established<br>due to duration<br>of device on the<br>market | No known<br>unmitigated use-<br>related or design<br>related safety<br>issues | No design-<br>related recall<br>identified |
### Indications for Use
UltraThineer Zirconia Slurry is indicated for use by dental technicians in the construction of custom-made ceramic restorations. These include copings, crowns, inlays and onlays, veneers, and bridges (3-unit anterior bridges).
### Table 2: Comparison of Predicate Device and Subject Device
| Contents | Subject Device | Predicate Device |
|--------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device: | UltraThineer Zirconia Slurry (5Y-B) | INNI-CERA |
| Manufacturer: | Boston Micro Fabrication | Aon Co., Ltd |
| 510(k) Number: | Pending | K211308 |
| Product Code(s): | EIH | EIH |
| Regulation Number<br>& Description: | 21 CFR 872.6660<br>Porcelain powder for clinical use | 21 CFR 872.6660<br>Porcelain powder for clinical use |
| Classification | II | II |
| Device<br>Description: | UltraThineer Zirconia Slurry device is<br>intended for use in fabricating custom-<br>made ceramic restorations using a 3D<br>printer additive manufacturing<br>process. There is one formulation, | INNI-CERA is intended for use in<br>fabricating custom-made ceramic<br>restorations using a 3D printer additive<br>manufacturing process. It is a product<br>that is used after fabricating and |
| Contents | Subject Device | Predicate Device |
| | BMF-5YZ-1.6B. It is a product that is<br>used after fabricating and sintering<br>with a mixture of zirconia powder and<br>binder. The 3D printer used to<br>fabricate the UltraThineer Zirconia<br>Slurry is a photocuring lamination<br>method using an STL file (dental<br>restoration shape), which is irradiated<br>with light in the strong UV region<br>(420nm) to cure the photocurable<br>mixture to form a model. This product<br>corresponds to ISO 6872 Type I Class<br>III. The material is used in a 3D printer,<br>which prints the shape determined by<br>an STL file converted from patients'<br>teeth data. The 3D printer is not<br>included with the product. The<br>UltraThineer Zirconia Slurry device<br>cannot be reused. | sintering with a mixture of zirconia<br>powder and binder. The DLP 3D<br>printer used to fabricate the INNI-<br>CERA is a photocuring lamination<br>method using an STL file (dental<br>restoration shape), which is irradiated<br>with light in the strong UV region<br>(420nm) to cure the photocurable<br>mixture to form a model. This product<br>corresponds to ISO 6872 Type I Class<br>III. The material is used in a 3D<br>printer, which prints the shape<br>determined by an STL file converted<br>from patients' teeth data. 3D printer is<br>not included with the product. INNI-<br>CERA cannot be reused. |
| Indications For<br>Use | UltraThineer Zirconia Slurry is<br>indicated for use by dental technicians<br>in the construction of custom-made<br>ceramic restorations. These include<br>copings, crowns, inlays and onlays,<br>veneers, and bridges (3-unit anterior<br>bridges). | INNI-CERA is indicated for use by<br>dental technicians in the construction<br>of custom-made ceramic restorations.<br>Coping<br>Crown<br>Inlay & Onlay<br>Veneer<br>Bridge (3-unit anterior bridges) |
| ISO 6872 Type/<br>Classification: | Type 1/Class 3 | Type 1/Class 3 |
| Material Type | Slurry | Slurry |
| Provided Sterile: | Non-Sterile | Non-Sterile |
| Single-Use or<br>Reusable: | Single Use | Single Use |
| Biocompatibility | According to EN ISO 10993-1 | According to EN ISO 10993-1 |
| Flexural Strength | >300 Mpa<br>(meeting ISO 6872 requirements) | >300 Mpa<br>(meeting ISO 6872 requirements) |
| Linear Thermal<br>Expansion<br>Coefficient | Mean linear thermal expansion<br>coefficient within 0.5 x 10-6 K-1 of the<br>value reported by the manufacturer | Unknown |
| Contents | Subject Device | Predicate Device |
| | (meeting ISO 6872 requirements) | (meeting ISO 6872 requirements) |
| Chemical Solubility | < 100µg/cm² | < 100µg/cm² |
| | (meeting ISO 6872 requirements) | (meeting ISO 6872 requirements) |
| Radioactivity | Activity concentration of uranium238 less than 1.0Bqg-1<br>(meeting ISO 6872 requirements) | Activity concentration of uranium238 less than 1.0Bqg-1<br>(meeting ISO 6872 requirements) |
| Production Type | 3D Printing | 3D Printing |
{6}------------------------------------------------
{7}------------------------------------------------
### Summary of Substantial Equivalence Comparison to the Predicate Device
The information provided for the subject device Indication for Use, Material, Biocompatibility, Flexural Strength, Linear Thermal Expansion Coefficient, Chemical Solubility and Radioactivity are equivalent to the predicate device, Inni-Cera cleared under K 211308. Both are zirconia slurry materials intended for 3D printing of dental implants followed by sintering. Both devices are composed of biocompatible material according to ISO 10993-1 and meet the requirements of ISO 6872. Both devices are intended for use by dental technicians in the construction of custom-made ceramic restorations.
### Biocompatibility Testing
UltraThineer Zirconia Slurry (5Y-B) was evaluated according to ISO 10993-1. The results of these tests confirmed that it met the biocompatibility requirements. Biocompatibility testing included:
- Chemical Characterization .
- Cytotoxicity
- Sensitization ●
- Oral Mucosal Irritation ●
- Acute Systemic Toxicity ●
- . Material Mediated Pyrogenicity
#### Non-Clinical Performance Data: Bench
The UltraThineer ceramic materials were evaluated according to ISO 6872:2015. The results of those test confirmed that the UltraThineer met all requirements of the standard. Testing included: Radioactivity; Flexural Strength; Linear Thermal Expansion Coefficient; Chemical Solubility; and; Shelf-Life Testing.
#### Performance Data: Clinical Testing
No clinical testing was applicable to this submission.
{8}------------------------------------------------
#### Conclusions
Like the predicate device originally cleared by the FDA. UltraThineer Zirconia Slurry has been shown to be safe through Bench Testing and Biocompatibility testing according to the above mentioned ISO standards. The data support substantial equivalence of the UltraThineer Zirconia Slurry subject device to the legally marketed predicate device.
The performance of the UltraThineer Zirconia Slurry meets the requirements of the non-clinical bench testing conducted to support substantial equivalence. Based on the available information, the subject device and the predicates are similar indication for use, material, performance data and biocompatibility.
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification that it can be concluded that the subject device is substantially equivalent with predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.