K240369 · BunkerHill Health · JAK · Sep 27, 2024 · Radiology
Device Facts
Record ID
K240369
Device Name
CAC (gated) Algorithm
Applicant
BunkerHill Health
Product Code
JAK · Radiology
Decision Date
Sep 27, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K240369 · Sep 27, 2024
CAC (gated) Algorithm
BunkerHill Health
Retrospective clinical gated CT studies
The device performance was validated in a stand-alone retrospective study using clinical CT images from six geographically diverse sites to evaluate the agreement between the device output and established ground truth for coronary artery calcium quantification.
Pivotal testing study; Stand-alone retrospective study
Patients aged 30 years and above (gated CT studies); Number of Sites: 6
Established ground truth
Cohen's weighted kappa for 4-category score group assessment
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Coronary Artery Calcium Detection
Deep-learning algorithms
Cohen's weighted kappa > 0.90
Cohen's weighted kappa of 0.972 (95%CI 0.958, 0.987)
—
—
Pivotal testing study: gated CT studies from six geographically diverse sites.
—
Indications for Use
CAC (gated) is a software device intended for use in estimating presence and quantity of coronary artery calcium for patients aged 30 years and above. The device automatically analyzes non-contrast electrocardiogram (ECG) gated cardiac computed tomography (CT) images collected and outputs the segmentation (intended for informational purposes only) and quantification of detected calcium. The output of the subiect device is made available to the physician on-demand as part of his or her standard workflow. The device-generated quantification can be viewed in the patient report at the discretion of the physician, and the physician also has the option of viewing the device-generated calcium segmentation in a diagnostic image viewer. The subject device output in no way replaces the original patient report or the original cardiac CT scan; both are still available to be viewed and used at the discretion of the physician. The device is intended to provide information to the physician to provide assistance during review of the patient's case. Results of the subject device are not intended to be used on a stand-alone basis and are solely intended to aid and provide information to the physician. In all cases, further action taken on a patient should only come at the recommendation of the physician after further reviewing the patient's results.
Device Story
Software as a Medical Device (SaMD) for coronary artery calcium (CAC) assessment; inputs non-contrast ECG-gated cardiac CT images in DICOM format. Uses deep-learning algorithms to automatically localize, segment, and quantify calcium burden. Outputs include exact Agatston scores for the whole heart and individual coronary arteries (LCA, LAD, LCX, RCA), plus categorization into four Agatston ranges (0, 1-99, 100-399, 400+). Used in medical facilities by interpreting physicians; output provided on-demand within standard clinical workflow. Physician reviews device-generated segmentation and quantification alongside original CT scans to assist in clinical case review. Does not replace standard of care or diagnostic reports; intended to aid clinical decision-making.
Clinical Evidence
Retrospective study using gated CT studies from six geographically diverse sites. Primary endpoint: agreement of cardiovascular disease risk category between device and ground truth. Metric: Cohen's weighted kappa for 4-category assessment. Result: 0.972 (95% CI 0.958, 0.987), exceeding the 0.90 acceptance criterion.
Technological Characteristics
SaMD; deep-learning algorithm for image analysis. Inputs: DICOM format non-contrast ECG-gated cardiac CT images (slice thickness up to 3mm). Threshold: 130 HU. Outputs: Agatston scores and calcium segmentation. Connectivity: interfaces with standard CT systems/diagnostic viewers. No hardware components.
Indications for Use
Indicated for estimating presence and quantity of coronary artery calcium in patients aged 30 years and above using non-contrast ECG-gated cardiac CT images. Not for stand-alone diagnostic use; intended as an adjunctive tool for physicians.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
September 27, 2024
BunkerHill Health % Nitva Naravanan Director of Regulatory Affairs 436 Bryant Street SAN FRANCISCO, CA 94107
Re: K240369
Trade/Device Name: CAC (gated) Algorithm Regulation Number: 21 CFR 892.1750 Regulation Name: Computed Tomography X-Ray System Regulatory Class: Class II Product Code: JAK Dated: August 23, 2024 Received: August 26, 2024
Dear Nitya Narayanan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang
Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Submission Number (if known)
K240369
Device Name
CAC (gated) Algorithm
### Indications for Use (Describe)
CAC (gated) is a software device intended for use in estimating presence and quantity of coronary artery calcium for patients aged 30 years and above. The device automatically analyzes noncontrast electrocardiogram (ECG) gated cardiac computed tomography (CT) images collected and outputs the segmentation (intended for informational purposes only) and quantification of detected calcium.
The output of the subiect device is made available to the physician on-demand as part of his or her standard workflow. The device-generated quantification can be viewed in the patient report at the discretion of the physician, and the physician also has the option of viewing the device-generated calcium segmentation in a diagnostic image viewer. The subject device output in no way replaces the original patient report or the original cardiac CT scan; both are still available to be viewed and used at the discretion of the physician.
The device is intended to provide information to the physician to provide assistance during review of the patient's case. Results of the subject device are not intended to be used on a stand-alone basis and are solely intended to aid and provide information to the physician. In all cases, further action taken on a patient should only come at the recommendation of the physician after further reviewing the patient's results.
Type of Use (Select one or both. as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(K) SUMMARY
## BunkerHill Health Inc.'s CAC (gated) Device
BunkerHill Health Inc.
436 Bryant Street
San Francisco CA 94107
Phone: (650) 842-0198 Contact Person: Nishith Khandwala
Date Prepared: September 18, 2024
### Proposed Device
| Proprietary Name | CAC (gated) Device |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification Name | Computed tomography x-ray system |
| Regulation Number | 21 CFR 892.1750 |
| Product Code | JAK |
| Requlatory Class | ============================================================================================================================================================================== |
# Predicate Device(s)
| Proprietary Name | Imbio's CAC Software |
|------------------------|----------------------------------|
| Premarket Notification | K230112 |
| Classification Name | Computed tomography x-ray system |
| Regulation Number | 21 CFR 892.1750 |
| Product Code | JAK |
| Regulatory Class | II |
| Proprietary Name | iCAC Device |
|------------------------|----------------------------------|
| Premarket Notification | K230223 |
| Classification Name | Computed tomography x-ray system |
| Regulation Number | 21 CFR 892.1750 |
| Product Code | JAK |
| Regulatory Class | II |
### Device Description
Bunkerhill CAC (gated) is a software as a medical device (SaMD) product that interfaces with compatible and commercially available CT systems. Bunkerhill CAC (gated) localizes, quantifies,
{5}------------------------------------------------
and categorizes coronary artery calcification in non-contrast, electrocardiogram (ECG) gated, chest CT studies. The core features of the product are:
- Categorization of the coronary artery calcium burden in the form of a range of Agatston ● scores. Calcium score groupings are defined as one of the four following ranges of Agatston units:
- a. Group 1: 0 Agatston units
- b. Group 2: 1-99 Agatston units
- Group 3: 100-399 Agatston units C.
- d. Group 4: 400+ Agatston units
- Quantification of the overall coronary artery calcium burden in the form of an exact Agatston Score.
- o Quantification of each coronary artery's (left main (LCA), left anterior descending (LAD), left circumflex (LCX), and right coronary artery (RCA)) calcium burden in the form of an exact Agatston score.
- Localization of estimated calcium burden in the form of a CAC segmentation applied to a ● copy of the original CT scan (intended for informational purposes only).
### Intended Use / Indications for Use
CAC (gated) is a software device intended for use in estimating presence and quantity of coronary artery calcium for patients aged 30 years and above. The device automatically analyzes non-contrast electrocardiogram (ECG) gated chest computed tomography (CT) images collected and outputs the seqmentation (intended for informational purposes only) and quantification of detected calcium.
The output of the subject device is made available to the physician on-demand as part of his or her standard workflow. The device-generated quantification can be viewed in the patient report at the discretion of the physician, and the physician also has the option of viewing the device-generated calcium segmentation in a diagnostic image viewer. The subject device output in no way replaces the original patient report or the original ECG-gated cardiac CT scan; both are still available to be viewed and used at the discretion of the physician.
The device is intended to provide information to the physician to provide assistance during review of the patient's case. Results of the subject device are not intended to be used on a stand-alone basis and are solely intended to aid and provide information to the physician. In all cases, further action taken on a patient should only come at the recommendation of the physician after further reviewing the patient's results.
### Summary of Technological Characteristics
At a high level, the subject and predicate devices are based on the following same technological elements:
- . Both the predicate and the subject device use deep-learning algorithms to identify the presence of coronary artery calcium deposits and quantify calcium burden in adult patients.
- Both devices analyze computed tomography (CT) imaqes that are sent to the software in o DICOM format.
- Both the predicate and the subject device quantify the calcium burden of the coronary arteries.
{6}------------------------------------------------
- . Both devices serve as support tools to provide information to the physician. However, they do not replace clinical evaluation and do not alter the standard of care.
- . Both devices segment the calcium area on an image and generate a report.
A table comparing the key features of the subject and predicate devices is provided below.
| Characteristic | Subject Device<br>(gated) algorithm | Primary Predicate Device:<br>CAC SW<br>Imbio Inc's<br>(K230112) | Secondary Predicate<br>Device: iCAC (K230223) | Summary |
|-----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use /<br>Indications for Use | CAC (gated) is a software<br>device intended for use in<br>estimating presence and<br>quantity of coronary artery<br>calcium for patients aged<br>30 years and above. The<br>device automatically<br>analyzes non-contrast<br>electrocardiogram (ECG)<br>gated chest computed<br>tomography (CT) images<br>collected and outputs the<br>segmentation (intended for<br>informational purposes<br>only) and quantification of<br>detected calcium. | Imbio CAC Software is<br>intended for use as a non-<br>invasive post-processing<br>software to evaluate<br>calcified plaques in the<br>coronary arteries, which<br>present a risk for coronary<br>artery disease. Imbio CAC<br>Software uses machine<br>learning to analyze thoracic<br>CT images and outputs a<br>summary report containing<br>Agatston score, arterial<br>age, and calcified lesion<br>mass and volume metrics<br>of the calcification burden<br>for the whole heart and<br>individual coronary artery<br>level. Additionally, Imbio<br>CAC Software outputs<br>annotated images<br>previewing the<br>segmentation of<br>calcifications for<br>informational purposes<br>only. Imbio CAC Software<br>is limited to the<br>quantification of detected<br>possible calcifications in<br>adult patients ≥ 29 years of<br>age. It does not diagnose<br>coronary artery disease.<br>The device output will be<br>available to the users as<br>part of the standard DICOM<br>viewing workflow. The<br>Imbio CAC Software results<br>are not intended to be used<br>on a stand-alone basis for | iCAC is a software device<br>intended for use in<br>estimating presence and<br>quantity of coronary artery<br>calcium for patients aged 30<br>years and above during<br>routine care. The device<br>automatically analyzes non-<br>gated, chest computed<br>tomography (CT) images<br>collected during routine<br>care and outputs a visual<br>representation of estimated<br>coronary artery calcium<br>segmentation (intended for<br>informational purposes<br>only) and both exact and<br>four-category quantitative<br>estimates of the patient's<br>coronary artery calcium<br>burden in Agatston units.<br><br>The output of the subject<br>device is made available to<br>the physician on-demand as<br>part of his or her standard<br>workflow. The device-<br>generated calcium score or<br>score group can be viewed<br>in the patient report at the<br>discretion of the physician,<br>and the physician also has<br>the option of viewing the<br>device-generated calcium<br>segmentation in a<br>diagnostic image viewer.<br>The subject device output in<br>no way replaces the original<br>patient report or the original | Same: Both primary and<br>secondary predicates are<br>non-invasive post-<br>processing SW to evaluate<br>calcified plaque in coronary<br>arteries. Both use CT<br>images. Similar to primary<br>predicate device, CAC<br>(gated) outputs a summary<br>report containing Agatston<br>score, i.e. the calcification<br>burden for the whole heart<br>and at individual coronary<br>artery level. Annotated<br>images previewing the<br>segmentation of<br>calcifications is provided for<br>information purposes. Both<br>subject and predicates are<br>not intended to be used on<br>a stand-alone basis. |
| | The output of the subject<br>device is made available to<br>the physician on-demand<br>as part of his or her<br>standard workflow. The<br>device-generated<br>quantification can be<br>viewed in the patient report<br>at the discretion of the<br>physician, and the<br>physician also has the<br>option of viewing the<br>device-generated calcium<br>segmentation in a<br>diagnostic image viewer.<br>The subject device output<br>in no way replaces the<br>original patient report or the<br>original ECG-gated cardiac<br>CT scan; both are still<br>available to be viewed and | | | |
| Characteristic | Subject Device (gated) algorithm | Primary Predicate Device:<br>Imbio Inc's CAC SW<br>(K230112) | Secondary Predicate Device: iCAC (K230223) | Summary |
| | used at the discretion of the physician.<br>The device is intended to provide information to the physician to provide<br>assistance during review of the patient's case. Results<br>of the subject device are<br>not intended to be used on<br>a stand-alone basis and are<br>solely intended to aid and<br>provide information to the<br>physician. In all cases,<br>further action taken on a<br>patient should only come at<br>the recommendation of the<br>physician after further<br>reviewing the patient's<br>results. | clinical decision-making or<br>otherwise preclude clinical<br>assessment of CT images. | ECG-gated cardiac CT<br>scan; both are still available<br>to be viewed and used at<br>the discretion of the<br>physician.<br>The device is intended to<br>provide information to the<br>physician to provide<br>assistance during review of<br>the patient's case. Results<br>of the subject device are<br>not intended to be used on<br>a stand-alone basis and are<br>solely intended to aid and<br>provide information to the<br>physician. In all cases,<br>further action taken on a<br>patient should only come at<br>the recommendation of the<br>physician after further<br>reviewing the patient's<br>results. | |
| Type of Interpretation | Adjunctive information | Adjunctive information | Adjunctive information | Same |
| Intended User | Interpreting physicians | Interpreting physicians | Interpreting physicians | Same |
| Patient population | Patients aged 30 years and above | Patients above the age of 29 | Patients aged 30 years and above | Same |
| Anatomical location | Cardiac | Thoracic, Chest, cardiac | Chest | Same |
| Intended location | Medical facility | Medical facility | Medical facility | Same |
| Rx or OTC | Rx | Rx | Rx | Same |
| Measurement scale | Agatston units | Agatston units | Agatston units | Same |
| Product code | JAK | JAK | JAK | Same |
| Regulation number | 21 CFR §892.1750 | 21 CFR §892.1750 | 21 CFR §892.1750 | Same |
| Modality | Computed tomography (CT) | Computed tomography (CT) | Computed tomography (CT) | Same |
| Image format | DICOM | DICOM | DICOM | Same |
| Supported CT scan | Cardiac gated CT scan | Non-cardiac-gated and<br>cardiac gated CT scan | Non-cardiac-gated CT scan | Similar. Primary predicate<br>is for both gated and non-<br>gated scans therefore |
| Characteristic | Subject Device -CAC (gated) algorithm | Primary Predicate Device: Imbio Inc's CAC SW (K230112) | Secondary Predicate Device: iCAC (K230223) | Summary |
| |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.