K240333 · Guangdong Genial Technology Co., Ltd. · FLL · Oct 30, 2024 · General Hospital
Device Facts
Record ID
K240333
Device Name
Wearable Digital Thermometer (T31)
Applicant
Guangdong Genial Technology Co., Ltd.
Product Code
FLL · General Hospital
Decision Date
Oct 30, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Wearable Digital Thermometer is a battery-operated electronic device with intended use of measuring and monitoring human armpit temperature continuously via wireless signal transmission of the measuring result. Wearable Digital Thermometer is reusable and intended for armpit temperature monitoring for persons of all age. The temperature data of device is not intended to replace the advice, diagnosis, nor treatment recommendations of doctor. Wearable Digital Thermometer can be used at home and healthcare center.
Device Story
Wearable Digital Thermometer (T31) is a reusable, battery-operated electronic device for continuous, remote armpit temperature monitoring. The device consists of a host unit, mobile App (Android/iOS), and medical-grade double-sided adhesive stickers. The host, featuring a 304 stainless steel probe and NTC resistor, is affixed to the axilla. It measures temperature in direct mode and transmits data via Bluetooth (2.4G BLE) to a paired smartphone. The App displays real-time temperature readings in Celsius or Fahrenheit. The device includes an automatic low-battery detection circuit that transmits alerts to the receiver. Used in home or clinical settings by patients or caregivers; the output assists in monitoring temperature trends. It does not provide medical diagnosis or treatment recommendations.
Clinical Evidence
Bench testing only. Performance validated against ASTM E1112-00 (2018) and ISO 80601-2-56:2017. Electrical safety and EMC testing performed per IEC 60601-1, IEC 60601-1-2, and IEC 60601-1-11. Biocompatibility testing conducted per ISO 10993-5 and ISO 10993-10. Usability testing performed per IEC 60601-1-6.
Technological Characteristics
Materials: 304 stainless steel probe, medical-grade double-sided adhesive. Sensing: NTC resistor. Energy: 3.0V/10mAh non-rechargeable battery. Connectivity: Bluetooth 2.4G BLE. Form factor: Wearable patch. Software: App-based display. Standards: ISO 80601-2-56, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, ISO 10993-1, ISO 10993-5, ISO 10993-10.
Indications for Use
Indicated for continuous armpit temperature monitoring in persons of all ages. Not intended to replace medical advice, diagnosis, or treatment. Suitable for home and healthcare center use.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 30, 2024
Guangdong Genial Technology CO., Ltd. % Klem Hou RA Manager Constant Medical Technology Co., Ltd. No. 2021-6, 2nd Floor, No. 228 Huadi Avenue Middle Liwan District, Guangzhou Zhaoqing, Guangdong 526437 China
Re: K240333
Trade/Device Name: Wearable Digital Thermometer (T31) Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical electronic thermometer Regulatory Class: Class II Product Code: FLL Dated: September 30, 2024 Received: September 30, 2024
Dear Klem Hou:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
David Walloschek
David Wolloscheck, Ph.D. Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K240333
Device Name Wearable Digital Thermometer (T31)
# Indications for Use (Describe)
Wearable Digital Thermometer is a battery-operated electronic device with intended use of measuring and monitoring human armpit temperature continuously via wireless signal transmission of the measuring result. Wearable Digital Thermometer is reusable and intended for armpit temperature monitoring for persons of all age. The temperature data of device is not intended to replace the advice, diagnosis, nor treatment recommendations of doctor. Wearable Digital Thermometer can be used at home and healthcare center.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
| ☐ Prescription Use (Part 21 CFR 801 Subpart D) | |
| ☒ Over-The-Counter Use (21 CFR 801 Subpart C) | |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# K240333 - 510(k) Summary
## 1. Submission sponsor
Name: GUANGDONG GENIAL TECHNOLOGY CO., LTD.
Address: I-6-05-02 11th Road, Area B, Guangfozhao Economic Cooperation Zone, Zhagang
Town, Huaiji County, Zhaoqing City, Guangdong Province, 526437, P.R. China Zhaoqing
Guangdong 526437 China
Contact person: Jun Yu
Title: Deputy general manager
Tel: +86- 136 0284 9551
Email: admin@genial.cn
# 2. Submission correspondent
Name: Constant Medical Technology Co., Ltd. Address: No. 2021-6, 2nd Floor, No. 228 Huadi Avenue Middle, Liwan District, Guangzhou Zhaoqing Guangdong 526437 China Contact person: Klem Hou Title: RA manager Tel: +86 178 7557 1727 Email: pharmaklem@outlook.com
#### 3. Device
Name of Device: Wearable Digital Thermometer
Model(s): T31
Common or Usual Name: Clinical Electronic Thermometer
Classification Name: Clinical Electronic Thermometer
Regulatory Class: II
Product Code: FLL
Regulation Number: 21 CFR 880.2910
{5}------------------------------------------------
#### 4. Predicate device(s)
| Manufacturer | Predicate Device | 510(k) Number |
|--------------------|----------------------------------------------------------------|---------------|
| iWEECARE Co., Ltd. | Temp Pal (Smart Thermometer Patch)<br>Model Number: STP-MB01-1 | K222588 |
#### Device description న్.
The Wearable Digital Thermometer consists of a host, App software and stickers. A comprehensive Android and iOS App are provided to access the host from a smart device. It is used for measuring and monitoring armpit temperature in real-time continuously and remotely via Bluetooth to smart device. The medical double-sided sticker are used to fix and stick the host in the user`s axilla.
The NTC temperature sensor is located closely to the stainless steel sheet that make sure the accuracy of temperature data measured.
The subject device could measure and monitor temperature in real-time continuously and remotely via Bluetooth to smart phone.
The Temp Pal is the combination device of thermometer and Bluetooth communication unit intended to be worn at axilla to monitor the armpit temperature continuously. The subject device is a direct mode clinical thermometer where the output temperature is not adjusted.
For the monitoring operation, switch the thermometer on and stick the thermometer in the user's axilla. The thermometer will make a Bluetooth connection between the thermometer and the receiver automatically (User should setup Bluetooth properly on receiver). Then the thermometer starts to measure the body temperature.
The wireless thermometer uses a rechargeable battery for operation.
When the battery is low, internal circuit will detect the low battery condition automatically and send "low battery" signal through Bluetooth communication unit to receiver.
Power source: Battery 3.0V/10mAh
Operation mode: Direct mode
Measuring results display method: The measuring results are transmitted to smart phone and display by APP.
Data communication method: Wireless 2.4G Bluetooth BLE.
#### Indications for Use 6.
Wearable Digital Thermometer is a battery-operated electronic device with intended use of
{6}------------------------------------------------
measuring and monitoring human armpit temperature continuously via wireless signal transmission of the measuring result. Wearable Digital Thermometer is reusable and intended for armpit temperature monitoring for persons of all age. The temperature data of device is not intended to replace the advice, diagnosis, nor treatment recommendations of doctor. Wearable Digital Thermometer can be used at home and healthcare center.
# 7. Comparison of Technological Characteristics with the Predicate Device(s)
Wearable Digital Thermometer has the same intended use, mode of action and operational characteristics as the predicate devices. Any minor differences between the subject device and the listed predicate devices do not raise any issues of safety or efficacy. Performance data supports that the device is safe and as effective as the predicate device for its intended use.
Therefore, The Wearable Digital Thermometer may be found substantially equivalent to its predicate devices.
Wearable Digital Thermometer is compared with the following Predicate Devices in terms of intended use, design, material, specifications, and performance:
1) K222588 (Predicate Device), " Temp Pal (Smart Thermometer Patch) ", manufactured by "
{7}------------------------------------------------
# iWEECARE Co., Ltd."
| Comparison<br>Elements | Subject Device | Predicate<br>Device<br>(K222588) | Remark |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Device name | Wearable Digital Thermometer | Temp Pal (Smart Thermometer<br>Patch) | -- |
| Model | T31 | STP-MB01-1 | -- |
| Indications for<br>use | Wearable Digital Thermometer<br>is a battery-operated electronic<br>device with intended use of<br>measuring and monitoring<br>human armpit temperature<br>continuously via wireless signal<br>transmission of the measuring<br>result. Wearable Digital<br>Thermometer is reusable and<br>intended for armpit temperature<br>monitoring for persons of all<br>age. The temperature data of<br>device is not intended to replace<br>the advice, diagnosis, nor<br>treatment recommendations of<br>doctor. Wearable Digital<br>Thermometer can be used at<br>home and healthcare center. | Temp Pal is a battery-operated<br>electronic device with intended<br>use of measuring and<br>monitoring human armpit<br>temperature continuously via<br>wireless signal transmission of<br>the measuring result. This<br>system is reusable and intended<br>for armpit temperature<br>monitoring for persons of all<br>age. The temperature data of<br>device is not intended to<br>replace the advice, diagnosis,<br>nor treatment recommendations<br>of doctor. Temp Pal can be used<br>at home and healthcare center. | Same |
| Principle of<br>Operation | For the monitoring operation,<br>switch the thermometer on and<br>stick the thermometer in the<br>user's axilla. The thermometer<br>will make a Bluetooth<br>connection between the<br>thermometer and the receiver<br>automatically (User should setup<br>Bluetooth properly on receiver).<br>Then the thermometer starts to<br>measure the body temperature<br>by means of testing the NTC<br>resistor's resistance value and<br>calculates the body temperature<br>continuously and sends the<br>temperature data to the receiver<br>through Bluetooth connection. | For the monitoring operation,<br>switch the thermometer on and<br>stick the thermometer in the<br>user's axilla. The thermometer<br>will make a Bluetooth<br>connection between the<br>thermometer and the receiver<br>automatically (User should<br>setup Bluetooth properly on<br>receiver). Then the<br>thermometer starts to measure<br>the body temperature by<br>means of testing the NTC<br>resistor's resistance value and<br>calculates the body temperature<br>continuously and sends the<br>temperature data to the receiver<br>through Bluetooth connection. | Same |
| Thermometer<br>Type | Clinical Electronic Thermometer | Clinical Electronic Thermometer | Same |
| Anatomical<br>Application | Patients' armpit | Patients' armpit | Same |
| Sensor | NTC Resistor | NTC Resistor | Same |
{8}------------------------------------------------
| Display | iOS device display, Android<br>device display | iOS device display, Android<br>device display | Same |
|------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------|---------------------|
| Power<br>Requirements | Non-Rechargeable Battery<br>3.0V/10mAh | Rechargeable<br>Battery<br>3.7V/10mAh | Different<br>Note 1 |
| Temperature<br>range<br>(Measurement<br>Range) | 68-113°F (20-45°C) | 77-113°F (25-45°C) | Different<br>Note 2 |
| Accuracy | ±0.1°C (35.00°C or39.00°C) - +<br>0.18°F (95°F - 102.20°F)<br>±0.2°C (<35.00°C or>39.00°C)<br>or 0.36°F (<95°F or >102.20°F) | 0.09 °F (± 0.05 °C) | Different<br>Note 3 |
| Valid<br>transmission<br>distance | 10 meters | Up to 5 meters | Different<br>Note 4 |
| APP<br>Temperature<br>display | Display in either °C or°F | Display in either °C or°F | Same |
| Operation<br>mode | Direct mode | Direct mode | Same |
| Reusability | Reusable | Reusable | Same |
| shelf life | 5 years | 24 months | Different<br>Note 5 |
|----------------------------------------------------------------|----------------------------------------------------------------|----------------------------------------------------------------|---------------------|
| composition of<br>material that<br>contacts to the<br>patients | 304 Stainless Steel Probe and<br>double sided medical adhesive | 304 Stainless Steel Probe and<br>double sided medical adhesive | Same |
# Comparison in Detail(s):
# Note 1: Power Requirements
The subject device is powered by a non-rechargeable battery while the predicate device is powered by a rechargeable battery. If the low power signal displayed in APP, the subject device should change the battery and the predicate device should be recharged, and both of the subject device and the
{9}------------------------------------------------
predicate device should stop working.
The Power Requirements differences between the subject and predicate devices does not raise new concerns of safety and effectiveness for the clinical use.
# Note 2: Temperature range (Measurement Range)
The maximum measured temperature of the subject device and predicate device are same and the minimum measured temperature of the subject device is a little lower than that of predicate device which is not important for that the human body temperature is not nearly close to 20°C.
The Temperature range (Measurement Range) differences between the subject and predicate devices does not raise new concerns of safety and effectiveness for the clinical use.
#### Note 3: Accuracy
The accuracy of the subject device is higher than that of the predicate device.
The accuracy of the subject device has met the requirement of ISO 80601-2-56:2017 and the accuracy differences between the subject and predicate devices does not raise new concerns of safety and effectiveness for the clinical use.
#### Note 4: Valid transmission distance
The transmission distance of the subject device is further than that of the predicate device.
The Valid transmission distance differences between the subject and predicate devices does not raise new concerns of safety and effectiveness for the clinical use.
#### Note 5: Shelf life
The shelf life of the subject device is further than that of the predicate device.
The shelf life differences between the subject and predicate devices does not raise new concerns of safety and effectiveness for the clinical use.
## 8. Performance Data
{10}------------------------------------------------
The following performance data are provided in support of the substantial equivalence determination.
# 1) Performance test
The Continuous measurement, Intermittent Determination and Direct clinical thermometer measure meet the specification of the product and the relevant standards are listed below.
- ASTM, E1112-00 (2018), Standard Specification for Electronic Thermometer for Intermittent A
Determination of Patient Temperature
- > ISO 80601-2-56 Second edition 2017-03, Medical electrical equipment - Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement
### 2) Electromagnetic Compatibility and Electrical Safety
Electrical safety and EMC testing has been performed to, and passed, the following standards:
- > IEC 60601-1 Edition 3.2 2020-08, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- > IEC 60601-1-2 Edition 4.1 2020-09. Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- » IEC 60601-1-11 Edition 2.1 2020-07, Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- A ANSI C63.4-2014, FCC Part 15 Subpart B
- 》 ANSI C63.4-2014, FCC Part 15 Subpart C
#### 3) Software Verification and Validation
Software documentation consistent with Basic Documentation Level of concern is submitted in this 510(k). System validation testing presented in this 510(k) demonstrated that all software requirement specifications have been met and all software hazards have been mitigated to acceptable risk levels. The relevant standards and guidance are listed below.
- A IEC 62304 Edition 1.1 2015-06, Medical device software - Software life cycle processes
- A Content of Premarket Submissions for Device Software Functions-Guidance for Industry and Food and Drug Administration Staff
- > Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions-Guidance for Industry and Food and Drug Administration Staff
{11}------------------------------------------------
### 4) Biocompatibility
The biocompatibility evaluation for the body-contacting components of the Wearable Digital Thermometer has been conducted in accordance with the "Use of International Standard ISO 10993-
1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk
management process" : Guidance for Industry and Food and Drug Administration Staff", as
recognized by FDA. The testing has been performed to, and passed, including:
- » ISO 10993-5:2009, Biological Evaluation of Medical Devices -Part 5: Tests for In Vitro Cytotoxicity
- > ISO 10993-10:2010, Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitization
#### 5) Shelf life
The Reliability Test of the Wearable Digital Thermometer has been conducted in accordance with relevant guidance and standards. The testing has been performed to, and passed, including:
- > Guidance of Shelf Life of Medical Devices (1991)
## 6) Usability
Usability testing has been performed to, and passed, the following standards:
- > IEC 60601-1-6:2020 Medical electrical equipment –Part 1-6: General requirements for safety -Collateral standard: Usability
## 7) Summary
Based on the above performance as documented in this application, the Wearable Digital
Thermometer is found to have a safety and effectiveness profile that is similar to the predicate device.
#### 9. Conclusions
Based on the comparison of intended use, design, materials and performance, the Wearable Digital Thermometer is to be concluded substantial equivalent to its predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.