K240327 · Unik Orthopedics, Inc. · JWH · Jun 27, 2024 · Orthopedic
Device Facts
Record ID
K240327
Device Name
UNIKO PointCloud™ Knee Instruments
Applicant
Unik Orthopedics, Inc.
Product Code
JWH · Orthopedic
Decision Date
Jun 27, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K240327 · Jun 27, 2024
UNIKO PointCloud™ Knee Instruments
Unik Orthopedics, Inc.
Published clinical literature (Matziolis - JBJS 2007; Ensini - CORR 2007) reporting outcomes of conventional total knee arthroplasty
The sponsor compared the accuracy of their device (tested in a cadaver study) against performance data for conventional total knee arthroplasty reported in peer-reviewed clinical literature to demonstrate substantial equivalence.
Literature review; Clinical performance benchmark; Total knee arthroplasty
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Literature comparison of conventional TKA accuracy; Literature review / Comparative analysis
Patients undergoing conventional total knee arthroplasty
Conventional total knee arthroplasty (as reported in literature)
The UNIKO PointCloud™ Knee Instruments are intended to be used as a surgical instrument to assist in the positioning of Total Knee Replacement components intraoperatively and in guiding the marking of bone before cutting provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient MRI scans. The UNIKO PointCloud™ Knee Instruments are compatible with the femoral and tibial components of the DJO Surgical EMPWR 3D Knee System and the Maxx Orthopedics Freedom Total Knee System. The Indications for Use of the DJO Surgical EMPWR 3D Knee System and the Maxx Orthopedics Freedom Total Knee System remain the same as those cleared in the manufacturer's clearances for the implant system.
Device Story
Patient-specific cutting guides (jigs) for total knee arthroplasty (TKA); machined from polyoxymethylene. Input: patient MRI scans processed via Vault® System Surgical Planning Software (K124051) to create individualized surgical plan. Output: physical cutting guides used intraoperatively by orthopedic surgeons to guide initial distal femoral and proximal tibial bone cuts. Guides establish references for femoral orientation; surgeon completes procedure using standard implant-specific instrumentation. Benefits: improved alignment accuracy compared to conventional instrumentation; consistent bone resection based on pre-operative plan. Used in clinical/surgical settings.
Clinical Evidence
Cadaver study with 3 experienced orthopedic surgeons performing 3 TKAs each (9 total). Primary endpoints: angular measurements (varus/valgus, flexion/extension, internal/external rotation) and resection thickness accuracy compared to pre-operative plan, predicate (K193312), and literature (Matziolis 2007, Ensini 2007). Results: accuracy comparable to predicate and superior to conventional surgery literature; pinning accuracy rated excellent; bone resections consistent.
Indicated for patients undergoing total knee replacement requiring assistance in positioning components and marking bone, provided anatomic landmarks are identifiable on MRI. Compatible with DJO Surgical EMPWR 3D and Maxx Orthopedics Freedom Total Knee systems.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 27, 2024
Unik Orthopedics, Inc. Charlie Chi Founder & CTO 701 Fortune Drive. Unit E San Jose, California 95131
Re: K240327
Trade/Device Name: UNIKO PointCloudTM Knee Instruments Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Codes: JWH. OOG. MBH Dated: April 5, 2024 Received: April 8, 2024
Dear Charlie Chi:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
| Peter G. | Digitally signed by Peter G. Allen -S |
|----------|---------------------------------------|
| Allen -S | Date: 2024.06.27<br>12:06:43 -04'00' |
For: Lixin Liu, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K240327
Device Name UNIKO PointCloud™ Knee Instruments
#### Indications for Use (Describe)
The UNIKO PointCloud™ Knee Instruments are intended to be used as a surgical instrument to assist in the positioning of Total Knee Replacement components intraoperatively and in guiding the marking of bone before cutting provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient MRI scans.
The UNIKO PointCloud™ Knee Instruments are compatible with the femoral and tibial components of the DJO Surgical EMPWR 3D Knee System and the Maxx Orthopedics Freedom Total Knee System. The Indications for Use of the DJO Surgical EMPWR 3D Knee System and the Maxx Orthopedics Freedom Total Knee System remain the same as those cleared in the manufacturer's clearances for the implant system.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for UNIKO Orthopedics. The word "UNIKO" is written in large, blue, sans-serif font. Below the word "UNIKO" is the phrase "UNIK ORTHOPEDICS" in a smaller, black, sans-serif font. To the right of the word "UNIKO" is a large, gray circle.
# 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
## Applicant Information:
| Owner Name: | Unik Orthopedics, Inc. |
|-----------------|----------------------------|
| Address: | 1701 Fortune Drive, Unit E |
| | San Jose, CA 95131 |
| Phone: | (408) 883-5842 |
| Contact Person: | Charlie Chi, Ph.D. |
| Phone: | (408) 887-5842 |
| Date Prepared: | 21 June 2024 |
## Device Information:
| Classification: | ClassII |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name: | UNIKOPointCloud™ KneeInstruments |
| Common name: | CuttingGuide |
| Classification name: | Knee joint patellofemorotibial polymer/metal/<br>polymer semi-constrained cemented prosthesis<br>Knee joint patellofemorotibial metal/polymer porous-<br>coated uncemented prosthesis |
| Regulation number: | 21 CFR 888.3560 and 21 CFR 888.3565 |
| Classification Code: | JWH, OOG, MBH |
## Predicate Device:
The subject of this submission, the UNIKO PointCloud™ Knee Instruments, is substantially equivalent to the Primary Predicate, Stryker ShapeMatch Cutting Guides cleared by FDA through 510(k) K122053, and the Secondary Predicate, UNIKO PointCloud™ Knee Instruments which were cleared by FDA through 510(k) K193312. The subject device has the same intended use, indications for use, design, material, operational principles, and performance as the secondary predicate device, UNIKO PointCloud™ Knee Instruments. Additionally, the subject device has substantially equivalent intended use, indications for use, design, materials, and operational principles to the Primary Predicate, Stryker ShapeMatch Cutting Guides.
## Device Description:
The purpose of this submission is to add an additional cleared knee implant system to the previously cleared UNIKO PointCloud™ Knee Instruments cleared through 510(k) K193312. The predicate device was cleared for compatibility with the DJO Surgical EMPWR 3D Knee System and this submission provides for additional compatibility with the Maxx Orthopedics Freedom Total Knee System
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for UNIKO UNIK ORTHOPEDICS. The word UNIKO is in blue, with the O being a gray circle. Underneath UNIKO is the text UNIK ORTHOPEDICS.
The UNIKO PointCloud™ Knee Instruments use the TKA surgical plan that is output from the Vault® System Surgical Planning Software which was previously cleared by FDA through K124051. The surgeon plans a primary total knee replacement surgery using the Vault System and the output data file from the individualized surgical plan is utilized by the UNIKO PointCloud™ Knee Instruments to create the necessary files for the production of patient specific cutting guides for the initial cuts to the distal femur and proximal tibia.
The UNIKO PointCloud™ Knee Instruments are patient specific cutting quides which are machined from polyoxymethylene material by the by means of customized off-the-shelf software. These guides (also called jigs) aid the surgeon in making the initial distal femoral and the initial proximal tibial bone cuts along with establishing the references for femoral orientations used during total knee arthroplasty surgery. The surgeon then continues the surgical procedure with the conventional knee instrumentation provided by the implant manufacturer for the implant and size specific cuts required for implantation of the femoral and tibial total knee implants.
The UNIKO PointCloud™ Knee Instruments are patient specific and are intended as single use instruments compatible with the TKA femoral and tibial components of the DJO Surgical EMPWR 3D Knee System and the Maxx Orthopedics Freedom Total Knee System. The Indications for Use of the DJO Surgical EMPWR 3D Knee System and the Maxx Orthopedics Freedom Total Knee System remain the same as those cleared in the manufacturer's clearance for those implant systems.
#### Intended Use / Indication for Use:
The UNIKO PointCloud™ Knee Instruments are intended to be used as a surgical instrument to assist in the positioning of Total Knee Replacement components intraoperatively and in guiding the marking of bone before cutting provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient MRI scans.
The UNIKO PointCloud™ Knee Instruments are compatible with the femoral and tibial components of the DJO Surgical EMPWR 3D Knee System and the Maxx Orthopedics Freedom Total Knee System. The Indications for Use of the DJO Surqical EMPWR 3D Knee System and the Maxx Orthopedics Freedom Total Knee System remain the same as those cleared in the manufacturer's clearances for the implant system.
## Substantial Equivalence:
Like the Primary Predicate, Stryker ShapeMatch Cutting Guides, and the Secondary Predicate, UNIKO PointCloud™ Knee Instruments, the subject UNIKO PointCloud™ Knee Instruments are intended to be used as a Knee Arthroplasty Implantation System. That is, a device accessory or set of device accessories that aids the surgeon in performing the implantation of the knee implant. Also, like both predicate devices, the UNIKO PointCloud™ Knee Instruments is indicated for use with femoral and tibial components of compatible FDA cleared total knee implant systems. See the table below for the specific implant systems for each of the predicate devices.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for UNIKO Orthopedics. The word "UNIKO" is written in large, blue, sans-serif font. Below the word "UNIKO" is the phrase "UNIK ORTHOPEDICS" in a smaller, sans-serif font. To the right of the word "UNIKO" is a gray circle.
## Technological Characteristics/Performance Data:
Substantial equivalence in materials and technological characteristics of the UNIKO PointCloud™ Knee Instruments and the predicate device is outlined in the table below:
| Product | UNIKO<br>PointCloud™ Knee<br>Instruments | Stryker<br>ShapeMatch<br>Cutting Guides | UNIKO<br>PointCloud™ Knee<br>Instruments | |
|---------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|
| 510(k) number | K240327 | K122053 – Primary<br>Predicate | K193312 –<br>Secondary Predicate | Conclusion<br>Substantially<br>Equivalent |
| Manufacturer | Unik Orthopedics Inc. | Stryker Corporation | Unik Orthopedics Inc. | |
| Product Code | MBH, JWH, OOG* | MBH, JWH, OOG* | MBH, OOG* | |
| Regulation | 21 CFR 888.3565<br>21 CFR 888.3560 | 21 CFR 888.3565**<br>21 CFR 888.3560** | 21 CFR 888.3565***<br>21 CFR 888.3560*** | SAME |
| Intended Use /<br>Indication | | | | |
| Intended Use | Intended to be used<br>as a Knee<br>Arthroplasty<br>Implantation System.<br>Intended to be used<br>to assist in the<br>implantation of a<br>specific knee<br>arthroplasty device or<br>a set of specific knee<br>arthroplasty devices. | Intended to be used<br>as a Knee<br>Arthroplasty<br>Implantation System.<br>Intended to be used<br>to assist in the<br>implantation of a<br>specific knee<br>arthroplasty device<br>or a set of specific<br>knee arthroplasty<br>devices. | Intended to be used<br>as a Knee<br>Arthroplasty<br>Implantation System.<br>Intended to be used<br>to assist in the<br>implantation of a<br>specific knee<br>arthroplasty device<br>or a set of specific<br>knee arthroplasty<br>devices. | SAME |
| Indications for<br>Use | The UNIKO<br>PointCloud™ Knee<br>Instruments are<br>intended to be used<br>as a surgical<br>instrument to assist<br>in the positioning of<br>Total Knee<br>Replacement<br>components intra-<br>operatively and in<br>guiding the marking<br>of bone before<br>cutting provided that<br>anatomic landmarks<br>necessary for<br>alignment and<br>positioning of the<br>Implant are<br>identifiable on patient<br>MRI scans.<br><br>The UNIKO<br>PointCloud™ Knee<br>Instruments are<br>compatible with the<br>femoral and tibial<br>components of the<br>DIO Surgical | The ShapeMatch<br>Cutting Guides are<br>intended to be used<br>as patient-specific<br>surgical<br>instrumentation to<br>assist in the<br>positioning of total<br>knee arthroplasty<br>components<br>intraoperatively and<br>in guiding the<br>marking of bone<br>before cutting<br>provided that<br>anatomic landmarks<br>necessary for<br>alignment and<br>positioning of the<br>implant are<br>identifiable on<br>patient imaging<br>scans.<br><br>The ShapeMatch<br>Cutting Guides are<br>intended for use with<br>the CR, PS, CS<br>components of the<br>Triathlon Knee | The UNIKO<br>PointCloud™ Knee<br>Instruments are<br>intended to be used<br>as a surgical<br>instrument to assist<br>in the positioning of<br>Total Knee<br>Replacement<br>components intra-<br>operatively and in<br>guiding the marking<br>of bone before<br>cutting provided that<br>anatomic landmarks<br>necessary for<br>alignment and<br>positioning of the<br>Implant are<br>identifiable on patient<br>imaging scans.<br><br>The UNIKO<br>PointCloud™ Knee<br>Instruments are<br>compatible with the<br>femoral and tibial<br>components of the<br>DIO Surgical | Substantially<br>Equivalent<br>(The only<br>differences<br>are in the<br>compatible<br>knee implant<br>systems) |
| Product | UNIKO<br>PointCloud™ Knee<br>Instruments | Stryker<br>ShapeMatch<br>Cutting Guides | UNIKO<br>PointCloud™ Knee<br>Instruments | Conclusion |
| 510(k) number | K240327 | K122053 – Primary<br>Predicate | K193312 –<br>Secondary Predicate | |
| Manufacturer | Unik Orthopedics Inc. | Stryker Corporation | Unik Orthopedics Inc. | |
| | EMPWR 3D Knee<br>System and the<br>Maxx Orthopedic<br>Freedom Total Knee<br>System. The<br>Indications for Use of<br>the DJO Surgical<br>EMPWR 3D Knee<br>System and the<br>Maxx Orthopedic<br>Freedom Total Knee<br>System remain the<br>same as those<br>cleared in the<br>manufacturer's<br>clearance for the<br>implant system. | System. The<br>indications for use of<br>the Triathlon Knee<br>System when used<br>with the ShapeMatch<br>Cutting Guides are:<br><i>General Total Knee<br/>Arthroplasty (TKA)</i><br>Indications:<br>-Painful, disabling<br>joint disease of the<br>knee resulting from;<br>degenerative<br>arthritis, rheumatoid<br>arthritis or post-<br>traumatic arthritis.<br>-Post-traumatic loss<br>of knee joint<br>configuration and<br>function.<br>-Moderate varus,<br>valgus, or flexion<br>deformity in which<br>the ligamentous<br>structures can be<br>returned to adequate<br>function and stability.<br>-Failed<br>reconstruction<br>procedures which did<br>not involve the<br>implantation of<br>hardware on the<br>condylar surfaces.<br><i>Additional<br/>Indications for<br/>Posterior<br/>Stabilized(PS)</i> :<br>-Ligamentous<br>instability requiring<br>implant bearing<br>surface geometries<br>with increased<br>constraint.<br>-Absent or non-<br>functioning posterior<br>cruciate ligament.<br>-Severe<br>anteroposterior<br>instability of the knee<br>joint. | EMPWR 3D Knee<br>System. The<br>Indications for Use of<br>the DJO Surgical<br>EMPWR 3D Knee<br>System remain the<br>same as those<br>cleared in the<br>manufacturer's<br>clearance for the<br>implant system. | |
| Product | UNIKO<br>PointCloud™ Knee<br>Instruments | Stryker<br>ShapeMatch<br>Cutting Guides | UNIKO<br>PointCloud™ Knee<br>Instruments | Conclusion |
| 510(k) number | K240327 | K122053 – Primary Predicate | K193312 – Secondary Predicate | |
| Manufacturer | Unik Orthopedics Inc. | Stryker Corporation | Unik Orthopedics Inc. | |
| Materials | | | | |
| Femur Jig Guide | polyoxymethylene | polyoxymethylene | polyoxymethylene | SAME |
| Femur Cutslot | polyoxymethylene | polyoxymethylene | polyoxymethylene | SAME |
| Tibia Jig Guide | polyoxymethylene | polyoxymethylene | polyoxymethylene | SAME |
| Tibia Cutslot | polyoxymethylene | N/A – one piece | polyoxymethylene | SAME |
| Design | | | | |
| Two part jig and cutslot | Yes | No, one piece jig and cut slot | Yes | Substantially Equivalent |
| Cutting Jigs are single use | Yes | Yes | Yes | SAME |
| Cutting Slots are single use | Yes | Yes | Yes | SAME |
| Cutting Guides are provided non-sterile | Yes | Yes | Yes | SAME |
| Technological Characteristics | | | | |
| Software creates pre-op surgical plan based on MRI data | Yes | Yes…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.