K240320 · Edan Instruments, Inc. · MHX · Jun 28, 2024 · Cardiovascular
Device Facts
Record ID
K240320
Device Name
Patient Monitor (RespArray)
Applicant
Edan Instruments, Inc.
Product Code
MHX · Cardiovascular
Decision Date
Jun 28, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1025
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The RespArray™ patient monitor is intended to be used for monitoring, storing, reviewing of, and to generate alarms for, multiple physiological parameters of adults, pediatrics and neonates. The monitors are intended for use by trained healthcare professionals in hospital environments. The monitor is for prescription use only. The monitored physiological parameters include: ECG, respiration (RESP), temperature (TEMP), oxygen saturation of arterial blood (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), and carbon dioxide (CO2). The arrhythmia detection and ST Segment analysis are intended for adult patients. The SpO2 (Nellcor™) module is intended to be used for spot-check or continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR), in motion and no motion conditions, and in patients who are well or poorly perfused. The Microstream™ capnography module is intended for continuous non-invasive monitoring of carbon dioxide concentration of the expired and inspired breath (etCO2) and respiration rate (RR). The monitor also provides the clinician with integrated pulmonary index (IPI), apnea per hour (A/hr) and oxygen desaturation index (ODI) values. IPI is not intended for patients up to the age of one year. A/hr and ODI are intended for ages 22 and up. The monitors are not intended for MRI environments.
Device Story
RespArray patient monitor performs continuous monitoring of physiological parameters (ECG, RESP, TEMP, SpO2, PR, NIBP, CO2). Device processes input signals via integrated modules (e.g., Nellcor SpO2, Microstream capnography) to provide real-time data, trend storage, and alarm generation for abnormal conditions. Used in hospital environments by trained healthcare professionals. Output displayed on monitor screen for clinical review; assists clinicians in patient assessment and timely intervention. Benefits include continuous patient surveillance and automated detection of physiological abnormalities.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), alarm systems (IEC 60601-1-8), and performance standards for ECG, NIBP, respiratory gas, temperature, and pulse oximetry modules.
Technological Characteristics
Multi-parameter patient monitor; AC power (100-240V) and rechargeable battery (6800 mAh). Modules: ECG (3/5 lead), NIBP (oscillation), SpO2 (Nellcor), CO2 (Microstream), TEMP. Connectivity: Wireless coexistence (IEEE ANSI C63.27). Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-2-27, IEC 80601-2-30, IEC 60601-2-49, ISO 80601-2-55, ISO 80601-2-56, ISO 80601-2-61.
Indications for Use
Indicated for monitoring, storing, reviewing, and alarming physiological parameters (ECG, RESP, TEMP, SpO2, PR, NIBP, CO2) in adults, pediatrics, and neonates. Arrhythmia/ST analysis for adults only. IPI for patients >1 year; A/hr and ODI for patients ≥22 years. For use by trained healthcare professionals in hospital environments.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
June 28, 2024
Edan Instruments, Inc Tracy Yue Regulatory Affairs Engineer 15 Jinhui Road, Jinsha Community, Kengzi Sub-district, Pingshan District, Shenzhen, Guangdong 518122 China
Re: K240320
Trade/Device Name: Patient Monitor (RespArray) Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector And Alarm (Including ST-Segment Measurement And Alarm) Regulatory Class: Class II Product Code: MHX, DSI, MLD, DRT, DXN, DSK, FLL, DQA, CCK, DSB Dated: February 2, 2024 Received: May 24, 2024
Dear Tracy Yue:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Kimberly N. Crowley -S
For: Jennifer Shih Kozen Assistant Director DHT2A: Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices
{2}------------------------------------------------
OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
K240320
Device Name
Patient Monitor (RespArray)
#### Indications for Use (Describe)
The RespArray™ patient monitor is intended to be used for monitoring, storing, reviewing of, and to generate alarms for, multiple physiological parameters of adults, pediatrics and neonates. The monitors are intended for use by trained healthcare professionals in hospital environments. The monitor is for prescription use only.
The monitored physiological parameters include: ECG, respiration (RESP), temperature (TEMP), oxygen saturation of arterial blood (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), and carbon dioxide (CO2).
The arrhythmia detection and ST Segment analysis are intended for adult patients.
The SpO2 (Nellcor™) module is intended to be used for spot-check or continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR), in motion and no motion conditions, and in patients who are well or poorly perfused.
The Microstream™ capnography module is intended for continuous non-invasive monitoring of carbon dioxide concentration of the expired and inspired breath (etCO2) and respiration rate (RR). The monitor also provides the clinician with integrated pulmonary index (IPI), apnea per hour (A/hr) and oxygen desaturation index (ODI) values. IPI is not intended for patients up to the age of one year. A/hr and ODI are intended for ages 22 and up.
The monitors are not intended for MRI environments.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(K) Summary
# Prepared in accordance with the content and format regulatory requirements of 21 CFR Part 807.92
# 1. Submitter:
| Applicant: | Edan Instruments, Inc.<br>#15 Jinhui Road, Jinsha Community,<br>Kengzi Sub-District, Pingshan District,<br>Shenzhen, 518122 P.R.China.<br>Tel: +86(0755) 26858736<br>Fax: +86(0755) 26882223 |
|-----------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact person: | Tracy Yue |
| Contact person: | Tracy Yue |
|-----------------|-------------------|
| Preparing date: | February 02, 2024 |
# 2. Device name and classification:
| Trade name: | Patient Monitor, Model: RespArray | | | |
|-----------------------------|---------------------------------------------------------------------------|--------------|--|--|
| Common/Usual Name: | Patient Monitor | | | |
| Classification Name: | 21 CFR 870.1025 | | | |
| | Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms) | | | |
| Regulatory class: | Class II | | | |
| Primary product code: | MHX-Monitor, Physiological, Patient (With Arrhythmia Detection Or Alarms) | | | |
| Subsequent product<br>code: | Regulation number/Device | Product Code | | |
| | | | | |
| Regulation number/Device | Product Code |
|---------------------------------------------------------------------------|--------------|
| 21 CFR 870.2340<br>Detector and Alarm, Arrhythmia | DSI |
| 21 CFR 870.1025<br>Monitor, ST Segment with Alarm | MLD |
| 21 CFR 870.2300<br>Monitor, Cardiac (Incl. Cardiotachometer & Rate Alarm) | DRT |
| 21 CFR 870.1130<br>System, Measurement, Blood-Pressure, Non-Invasive | DXN |
| 21 CFR 870.1110<br>Computer, Blood-Pressure | DSK |
| 21 CFR 880.2910<br>Thermometer, Electronic, Clinical | FLL |
| 21 CFR 870.2700<br>Oximeter | DQA |
| 21 CFR 870.1400<br>Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase | CCK |
{5}------------------------------------------------
Traditional 510(k) of RespArray Patient Monitor
| 21 CFR 870.2770 | DSB |
|--------------------------|-----|
| Impedance plethysmograph | |
3. Predicate Device(s): 1) Edan Instruments, Inc., Patient Monitor, Model: RespArray, K220308 (Primary predicate device)
- 2) Edan Instruments, Inc., Vital Signs Monitor, Model: iM3s, iM3As, iM3Bs, iHM3s K202892 (Reference device)
- 3) Covidien LLC., MicrostreamTM CO2 NanoPod, K213911 (Reference device)
#### The RespArray patient monitor (hereinafter called RespArray) can perform long-time 4. Device Description: continuous monitoring of multiple physiological parameters. Also, it is capable of storing, displaying, analyzing and controlling measurements, and it will indicate alarms in case of abnormity so that doctors and nurses can deal with them in time.
- The RespArray™ patient monitor is intended to be used for monitoring, storing, reviewing 5. Indication for Use of, and to generate alarms for, multiple physiological parameters of adults, pediatrics and neonates. The monitors are intended for use by trained healthcare professionals in hospital environments.
The monitor is for prescription use only.
The monitored physiological parameters include: ECG, respiration (RESP), temperature (TEMP), oxygen saturation of arterial blood (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), and carbon dioxide (CO2).
The arrhythmia detection and ST Segment analysis are intended for adult patients.
The SpO2 (Nellcor™) module is intended to be used for spot-check or continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR), in motion and no motion conditions, and in patients who are well or poorly perfused.
The Microstream™ capnography module is intended for continuous non-invasive monitoring of carbon dioxide concentration of the expired and inspired breath and carbon dioxide concentration of the exhaled breath and inspired (etCO2) and respiration rate (RR). The monitor also provides the clinician with integrated pulmonary index (IPI), apnea per hour (A/hr) and oxygen desaturation index (ODI) values. IPI is not intended for patients up to the age of one year. A/hr and ODI are intended for age 22 and up. The monitors are not intended for MRI environments.
#### 6. Predicate Device Comparison
The table below compares the indication for use and key technological feature of the subject devices to the predicate device (Patient Monitor Model RespArray, K220308).
{6}------------------------------------------------
| Item | <Subject Device><br>RespArray | <Predicate Device><br>RespArray | Comparison<br>Result |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| Manufacture<br>r/K# | Current Submission | Edan Instrument.Inc/K220308 | — |
| Intended Use | The RespArray™ patient monitor is<br>intended to be used for monitoring,<br>storing, reviewing of, and to generate<br>alarms for, multiple physiological<br>parameters of adults, pediatrics and<br>neonates. The monitors are intended<br>for use by trained healthcare<br>professionals in hospital environments.<br>The monitor is for prescription use<br>only.<br>The monitored physiological<br>parameters include: ECG, respiration<br>(RESP), temperature (TEMP), oxygen<br>saturation of arterial blood (SpO2),<br>pulse rate (PR), non-invasive blood<br>pressure (NIBP), and carbon dioxide<br>(CO2).<br>The arrhythmia detection and ST<br>Segment analysis are intended for<br>adult patients.<br>The SpO2 (Nellcor™) module is<br>intended to be used for spot-check or<br>continuous non-invasive monitoring of<br>functional oxygen saturation of arterial<br>hemoglobin (SpO2) and pulse rate | The RespArray™ patient monitor is<br>intended to be used for monitoring,<br>storing, reviewing of, and to generate<br>alarms for, multiple physiological<br>parameters of adults, pediatrics and<br>neonates. The monitors are intended<br>for use by trained healthcare<br>professionals in hospital environments.<br>The monitor is for prescription use<br>only.<br>The monitored physiological<br>parameters include: ECG, respiration<br>(RESP), temperature (TEMP), oxygen<br>saturation of arterial blood (SpO2),<br>pulse rate (PR), non-invasive blood<br>pressure (NIBP), and carbon dioxide<br>(CO2).<br>The arrhythmia detection and ST<br>Segment analysis are intended for<br>adult patients.<br>The SpO2 (Nellcor™) module is<br>intended to be used for spot-check or<br>continuous non-invasive monitoring of<br>functional oxygen saturation of arterial<br>hemoglobin (SpO2) and pulse rate | Same |
| | (PR), in motion and no motion<br>conditions, and in patients who are<br>well or poorly perfused.<br><br>The Microstream™ capnography<br>module is intended for continuous | (PR), in motion and no motion<br>conditions, and in patients who are<br>well or poorly perfused.<br><br>The Microstream™ capnography<br>module is intended for continuous | |
| | non-invasive monitoring of carbon dioxide concentration of the expired and inspired breath (etCO2) and respiration rate (RR).<br>The monitor also provides the clinician with integrated pulmonary index (IPI), apnea per hour (A/hr) and oxygen desaturation index (ODI) values. IPI is not intended for patients up to the age of one year. A/hr and ODI are intended for ages 22 and up.<br>The monitors are not intended for MRI environments. | non-invasive monitoring of carbon dioxide concentration of the expired and inspired breath (etCO2) and respiration rate (RR).<br>The monitor also provides the clinician with integrated pulmonary index (IPI), apnea per hour (A/hr) and oxygen desaturation index (ODI) values. IPI is not intended for patients up to the age of one year. A/hr and ODI are intended for ages 22 and up.<br>The monitors are not intended for MRI environments. | |
| | ECG monitor | | |
| Lead Mode | 3 Electrodes; 5 Electrodes; | 3 Electrodes; 5 Electrodes; | Same |
| Arrhythmia Analysis | ASYSTOLE, VFIB/VTAC, COUPLET, VT > 2, BIGEMINY, TRIGEMINY, VENT, R on T, PVC, TACHY, BRADY, MISSED BEATS, IRR, VBRADY, PNC, PNP | ASYSTOLE, VFIB/VTAC, COUPLET, VT > 2, BIGEMINY, TRIGEMINY, VENT, R on T, PVC, TACHY, BRADY, MISSED BEATS, IRR, VBRADY, PNC, PNP | Same |
| | RESP monitor from ECG | | |
| Principle of Operation | Thoracic impedance | Thoracic impedance | Same |
| Measurement Range | 0 rpm to 200 rpm | 0 rpm to 200 rpm | Same |
| | NIBP monitor | | |
| Principle of Operation | oscillation | oscillation | Same |
| Measurement Range | Measurement range:<br>Adult Pediatric Neonate<br>Systolic 25-290 25-240 25-140<br>Diastolic 10-250 10-200 10-115<br>Mean 15-260 15-215 15-125 | Measurement range:<br>Adult Pediatric Neonate<br>Systolic 25-290 25-240 25-140<br>Diastolic 10-250 10-200 10-115<br>Mean 15-260 15-215 15-125 | Same |
| PR from NIBP | | | |
| Measurement range | 40 bpm to 240 bpm | Same | |
| Continuous TEMP | | | |
| Measurement Range | 0 °C to 50 °C (32 °F to 122 °F) | Same | |
| SpO2 monitor | | | |
| Module | Nell-1 | Same | |
| Spot Temp | | | |
| Measuring Range | 16 °C ~ 43 °C (61 °F ~ 110 °F) | The added spot temp is substantial equivalent to the Spot Temp Module cleared in K202892 | |
| Power supply | | | |
| AC power | 100 V to 240 V~<br>50 Hz/60 Hz | Same | |
| Rechargeable Battery | Yes | Same | |
| Battery Capacity | 6800 mAh | The battery comply with IEC 62133-2. | |
| CO2 Monitor | | | |
| Module | Microstream™ NanoMediCO2 EtCO2 | The changed CO2 module is substantial equivalent to the CO2 module cleared in | |
{7}------------------------------------------------
{8}------------------------------------------------
As seen in the comparison tables, the subject and predicate devices have similar design features and performance specifications. The technological differences between the subject and predicate devices do not raise different questions of safety or effectiveness.
# 7. Performance Data:
## Non-clinical data:
## Electrical safety and electromagnetic compatibility (EMC)
The Resparray Patient Monitors were assessed for conformity with the relevant requirements of the following 510(k)_Summary 5 / 7
{9}------------------------------------------------
standards and found to comply:
- ANSI AAMI ES60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021] Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) [Including Amendment 2 (2021)]
- ANSI AAMI IEC 60601-1-2:2014 [Including AMD 1:2021] Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: electromagnetic disturbances - Requirements and tests.
#### Performance testing-Bench
Edan has conducted functional and system level testing to validate the performance of the results of the bench testing show that the subject device meets its accuracy specification and meet relevant consensus standards.
- IEC 60601-1-8:2020 Medical electrical equipment - part 1-8: general requirements for basic safety and essential performance - collateral standard: general requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems
- IEC 60601-2-27:2011 Medical electrical equipment--Part 2-27: Particular requirements for the basic safety and essential performance of electrocardiographic monitoring equipment
- IEC 80601-2-30:2018 Medical electrical equipment - part 2-30: particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers
- IEC 60601-2-49:2018 Medical electrical equipment –Part 2-49: Particular requirements for the basic safety and essential performance of multifunction patient monitoring equipment
- ISO 80601-2-55:2018 Medical electrical equipment - part 2-55: particular requirements for the basic safety and essential performance of respiratory gas monitors
- ISO 80601-2-56:2017+A1:2018 Medical electrical equipment - part 2-56: particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement
- ISO 80601-2-61:2017 Medical electrical equipment - part 2-61: particular requirements for basic safety and essential performance of pulse oximeter equipment
- IEEE ANSI C63.27:2017 American National Standard for Evaluation of Wireless Coexistence
#### Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and Food and Drug Administration Staff, "Content of Premarket Submissions for Device Software Functions".
#### Clinical data: Not applicable.
#### Summary
510(k)_Summary
{10}------------------------------------------------
The non-clinical performance testing showed that the subject devices are as safe and as effective as the predicate device.
### 8. Conclusion
The bench testing data and software verification and validation demonstrate that RespArray Patient Monitor is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.