Clinical samples were used to perform a method comparison study between the INNOVANCE Anti-Xa assay and the predicate device (HemosIL Liquid Anti-Xa) to demonstrate substantial equivalence.
Patients undergoing anticoagulant therapy with apixaban; Sample Size: 301; Number of Sites: 4
HemosIL Liquid Anti-Xa on the ACL TOP
Passing-Bablok regression analysis
Indications for Use
INNOVANCE Anti-Xa assay in combination with INNOVANCE Heparin Calibrator is an In-vitro diagnostic automated chromogenic assay for the quantitative determination of unfractionated heparin (UFH) and low molecular weight heparin (LMWH) activity in human plasma collected from venous blood samples in 3.2 % sodium citrated tubes in the clinical laboratory. The quantitative determination of UFH and LMWH can be performed on the BCS XP System, CS-2500 System, CS-5100 System and the CA-660 System. For use with plasma from patients undergoing anticoagulant therapy with either UFH or LMWH. INNOVANCE Anti-Xa assay in combination with INNOVANCE Apixaban Standards provides quantitative determination of the concentration of apixaban in human plasma collected from venous blood samples in 3.2 % sodium citrated tubes in the clinical laboratory. The quantitative determination of apixaban can be performed on CS-2500 System. For use with plasma from patients undergoing anticoagulant therapy with apixaban in situations where quantification of apixaban levels may be indicated: · Patient with bleeding, · Patient with risk for bleeding (e.g. during perioperative management), · Patient with conditions affecting pharmacokinetics (e.g. deteriorating renal function, extremes of body weight, treatment with other drugs known to affect pharmacokinetics of apixaban).
Device Story
INNOVANCE Anti-Xa is a one-stage chromogenic assay for quantitative measurement of apixaban, UFH, and LMWH in 3.2% citrated human plasma. Input: patient plasma sample mixed with reagent containing Factor Xa and chromogenic substrate. Principle: Factor Xa converts substrate to paranitroaniline; apixaban or heparin/antithrombin complex inhibits Factor Xa, reducing paranitroaniline formation. Output: light absorption at 405 nm measured by automated coagulation analyzers (CS-2500, CS-5100, BCS XP, CA-660). Used in clinical laboratories by technicians/pathologists. Results assist clinicians in managing patients on anticoagulant therapy, particularly during bleeding events or perioperative care. Benefits include precise drug level monitoring to guide clinical decision-making.
Clinical Evidence
Bench testing only. Precision studies (n=240 per sample) showed CVs ranging from 1.25% to 2.57%. Linearity verified from 10.5 to 378 ng/mL. LoD is 3.3 ng/mL; LoQ is 15.8 ng/mL. Method comparison study (n=301) against the predicate device yielded a correlation coefficient of 0.989 and a Passing-Bablock regression of y = 1.026x - 8.590 ng/mL. Interference testing confirmed no significant impact from common endogenous substances or drugs.
Technological Characteristics
One-stage chromogenic assay. Reagents: Factor Xa (bovine) and chromogenic substrate (Suc-Ile-Glu(piperidin-1-yl)-Gly-Arg-pNA.HCl). Energy source: Photometric measurement at 405 nm. Connectivity: Integrated with automated coagulation analyzers (BCS XP, CS-2500, CS-5100, CA-660). Stability: 72 hours on-board, 8 weeks open-vial. Standards: CLSI EP05-A3, EP07-A2/ED3, EP09-A3, EP34-ED1.
Indications for Use
Indicated for patients undergoing anticoagulant therapy with UFH, LMWH, or apixaban. Apixaban testing is indicated for patients with bleeding, risk for bleeding (e.g., perioperative), or conditions affecting pharmacokinetics (e.g., renal impairment, weight extremes, drug interactions). Not established for neonate or pediatric populations.
Regulatory Classification
Identification
A heparin and direct oral factor Xa inhibitor drug test system is intended for the detection of heparin and direct oral factor Xa inhibitors in human specimens collected from patients taking heparin or direct oral factor Xa inhibitors. This device is intended to aid in the management of therapy in conjunction with other clinical and laboratory findings.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include the following:
(i) Detailed documentation of analytical device performance studies and results demonstrating acceptable analytical performance with a sufficient number of specimens tested in order to obtain unbiased estimates of analytical performance. This documentation shall include the following as appropriate to the technology, specimen types tested, and intended use of the device:
(A) Studies and results for that demonstrate device precision including repeatability and reproducibility, using quality controls and clinical samples, when appropriate. Precision studies must assess specimens for each indicated drug at concentrations throughout the measuring range of the device including near clinically relevant levels, as appropriate. The study must evaluate different sources of variability including, as appropriate, between-run, between-operator, between-lot, between-instrument, between-day, and between-site;
(B) Studies and results that demonstrate that the device is free from clinically significant interference, from endogenous and exogenous interferents associated with the target population(s), and interferents that are specific for, or related to, the technology or methodology of the device;
(C) Data to demonstrate appropriate specimen stability for the intended sample matrices under the intended conditions for specimen collection, handling, and storage described in the device labeling;
(D) Studies and results that demonstrate the linear range, limit of blank (LoB), limit of detection (LoD), and limit of quantitation (LoQ), as applicable to the technology of the device; and
(E) For any devices intended for use for near patient testing, studies and results that demonstrate the robustness of the device in the hands of the intended user, including the entire testing procedure, pre-analytical specimen processing steps, and results interpretation.
(ii) Detailed documentation of clinical performance testing in which the performance is analyzed relative to a comparator that FDA has determined is appropriate. Specimens must be representative of the intended use population(s) and must cover the full range of the device output and any clinically relevant decision points as appropriate.
(2) The labeling required under § 809.10(b) of this chapter must include:
(i) Identification of any known interferents, including all endogenous, exogenous, technology-specific, and patient population-specific interferents, specific to the test outputs. The information must include the concentration(s) or level(s) of the interferent at which clinically significant interference was found to occur, and the concentration range or levels at which interference was not found to occur;
(ii) A prominent statement that the device is not intended for use in monitoring patients taking heparin or direct oral factor Xa inhibitors; and
(iii) Limiting statements indicating, as applicable:
(A) That the device should only be used in conjunction with information available from clinical evaluations and other diagnostic procedures; and
(B) That the device is not specific to the direct oral factor Xa inhibitor that has been evaluated and may detect the presence of other direct factor Xa inhibitors that have not been evaluated.
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY ONLY
## I Background Information:
A 510(k) Number
K240315
B Applicant
Siemens Healthcare Diagnostics Products GmbH
C Proprietary and Established Names
INNOVANCE Anti-Xa
INNOVANCE Heparin Calibrator; INNOVANCE Heparin UF and LMW Controls
INNOVANCE Apixaban Standards; INNOVANCE Apixaban Controls
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| QLU | Class II | 21 CFR 864.7295 - Heparin and Direct Oral Factor Xa Inhibitor Drug Test System | HE - Hematology |
## II Submission/Device Overview:
A Purpose for Submission:
Clearance of INNOVANCE Anti-Xa assay for apixaban measurement with INNOVANCE Apixaban Standards and Apixaban Controls
B Measurand:
Apixaban, unfractionated heparin (UFH) and low molecular weight heparin (LMWH)
C Type of Test:
Chromogenic assay
## III Intended Use/Indications for Use:
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
K240315 - Page 2 of 12
A Intended Use(s):
See Indications for Use below.
B Indication(s) for Use:
INNOVANCE Anti-Xa assay in combination with INNOVANCE Heparin Calibrator is an In-vitro diagnostic automated chromogenic assay for the quantitative determination of unfractionated heparin (UFH) and low molecular weight heparin (LMWH) activity in human plasma collected from venous blood samples in 3.2 % sodium citrated tubes in the clinical laboratory. The quantitative determination of UFH and LMWH can be performed on the BCS XP System, CS-2500 System, CS-5100 System and the CA-660 System. For use with plasma from patients undergoing anticoagulant therapy with either UFH or LMWH.
INNOVANCE Anti-Xa assay in combination with INNOVANCE Apixaban Standards provides quantitative determination of the concentration of apixaban in human plasma collected from venous blood samples in 3.2 % sodium citrated tubes in the clinical laboratory. The quantitative determination of apixaban can be performed on CS-2500 System. For use with plasma from patients undergoing anticoagulant therapy with apixaban in situations where quantification of apixaban levels may be indicated:
- Patient with bleeding
- Patient with risk for bleeding (e.g. during perioperative management)
- Patient with conditions affecting pharmacokinetics (e.g. deteriorating renal function, extremes of body weight, treatment with other drugs known to affect pharmacokinetics of apixaban).
The performance of this device has not been established in neonate and pediatric patient populations.
C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
For in vitro diagnostic use only
D Special Instrument Requirements:
When used with INNOVANCE Heparin Calibrator:
- BCS XP System
- Automated Blood Coagulation Analyzer CA-600 series
- Automated Blood Coagulation Analyzer CS-2500
- Automated Blood Coagulation Analyzer CS-5100
When used with INNOVANCE Apixaban Standards:
- Automated Blood Coagulation Analyzer CS-2500 (K172286)
IV Device/System Characteristics:
A Device Description:
{2}
INNOVANCE Anti-Xa assay is a one stage chromogenic assay based on a synthetic chromogenic substrate and on Factor Xa inactivation. The reagent kit consists of two components. One component (Reagent) contains Xa, the other (Substrate) a chromogenic substrate specific for Xa. The assay provides quantitative UFH/LMWH and apixaban results on 3.2% citrated human plasma when used with INNOVANCE Heparin Calibrator and INNOVANCE Apixaban Standards.
For measuring apixaban:
INNOVANCE Apixaban Standards are lyophilized reagents of two levels (Standard 0 and 1). They consist of human plasmas with either no apixaban (Standard 0) or a defined concentration of apixaban calibrated against an apixaban reference preparation (Standard 1).
INNOVANCE Apixaban Controls are lyophilized reagents of two levels (Control 0 and 1). They consist of human plasmas with defined concentrations of apixaban.
For measuring UFH/LMWH:
The INNOVANCE Heparin Calibrator consists of five calibrator levels. INNOVANCE Heparin Calibrator 1 represents plasma containing no heparin. INNOVANCE Heparin Calibrators 2, 3, 4 and 5 are plasmas with defined activities of LMWH and are calibrated against the WHO (World Health Organization) International Standards for UFH and LMWH.
The INNOVANCE Heparin Control consists of plasmas with defined activities of either UFH (control 1 and 2) or LMWH (control 1 and 2).
## B Principle of Operation
INNOVANCE Anti-Xa assay is one stage chromogenic assay. Upon mixing of INNOVANCE Anti-Xa REAGENT and INNOVANCE Anti-Xa SUBSTRATE, Xa converts the chromogenic substrate into two products, one of them is paranitroaniline. The formation of paranitroaniline can be quantified by the coagulation analyzer employing light absorption at a specific wavelength (405 nm).
In the presence of a heparin containing sample, the formation of paranitroaniline will be reduced in a time dependent manner. This is due to inhibition of Xa by the heparin/AT complex. This complex is formed in the patient's plasma and competes with the substrate conversion by Xa. The concentration of the complex is not only dependent on the concentration of heparin but also on the availability of the patient's endogenous antithrombin. By comparison to a reference curve the heparin activity of the sample can be quantified.
In the presence of an apixaban containing sample factor Xa is inhibited directly by this inhibitor. Comparison to an inhibitor specific reference curve allows quantification of the inhibitor concentration in the sample.
## V Substantial Equivalence Information:
### A Predicate Device Name(s):
HemosIL Liquid Anti-Xa
K240315 - Page 3 of 12
{3}
B Predicate 510(k) Number(s):
K223187
# C Comparison with Predicate(s):
| Device & Predicate Device(s): | K240315 | K223187 |
| --- | --- | --- |
| Device Trade Name | INNOVANCE Anti-Xa | HemosIL Liquid Anti-Xa |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | INNOVANCE Anti-Xa assay in combination with INNOVANCE Heparin Calibrator is an In-vitro diagnostic automated chromogenic assay for the quantitative determination of unfractionated heparin (UFH) and low molecular weight heparin (LMWH) activity in human plasma collected from venous blood samples in 3.2 % sodium citrated tubes in the clinical laboratory. The quantitative determination of UFH and LMWH can be performed on the BCS XP System, CS-2500 System, CS-5100 System and the CA-660 System. For use with plasma from patients undergoing anticoagulant therapy with either UFH or LMWH. INNOVANCE Anti-Xa assay in combination with INNOVANCE Apixaban Standards provides quantitative determination of the concentration of apixaban in human plasma collected from venous blood samples in 3.2 % sodium citrated tubes in the clinical laboratory. The quantitative determination of apixaban can be performed on CS-2500 System. For use with plasma from patients undergoing anticoagulant therapy with apixaban in situations where quantification of apixaban levels may be indicated: | HemosIL Liquid Anti-Xa is an automated chromogenic assay for in vitro diagnostic use by laboratory professionals in clinical laboratories. The assay provides quantitative results on 3.2% citrated human plasma for the following analytes based on the calibrators used: ·When used with HemosIL Heparin Calibrators: Quantitative determination of unfractionated heparin (UFH) and low molecular weight heparin (LMWH) activity on the ACL TOP Family and ACL TOP Family 50 Series. ·When used with HemosIL Apixaban Calibrators: Quantitative determination of apixaban on the ACL TOP Family and ACL TOP Family 50 Series through measurement of Factor Xa activity, which is inversely proportional to the apixaban level. With HemosIL Apixaban Calibrators, the assay is intended to measure apixaban concentrations in patients on apixaban therapy in the following situations where measurement of apixaban levels could be useful to have as additional information: - Patients at risk for major bleeding - Patients experiencing a bleeding episode |
K240315 - Page 4 of 12
{4}
K240315 - Page 5 of 12
| | • Patient with bleeding
• Patient with risk for bleeding (e.g. during perioperative management)
• Patient with conditions affecting pharmacokinetics (e.g. deteriorating renal function, extremes of body weight, treatment with other drugs known to affect pharmacokinetics of apixaban).
The performance of this device has not been established in neonate and pediatric patient populations. | • When used with HemosIL Rivaroxaban Calibrators:
Quantitative determination of rivaroxaban on the ACL TOP Family and ACL TOP Family 50 Series through measurement of Factor Xa activity, which is inversely proportional to the rivaroxaban level. With HemosIL Rivaroxaban Calibrators, the assay is intended to measure rivaroxaban concentrations in patients on rivaroxaban therapy in the following situations where measurement of rivaroxaban levels could be useful to have as additional information:
- Patients at risk for major bleeding
- Patients experiencing a bleeding episode
The assay is not a stand-alone test and the results should be used in conjunction with other clinical and laboratory findings.
For use in adult population. For prescription use only. |
| --- | --- | --- |
| Sample Type | 3.2% citrated human plasma | Same |
| Measurement | Quantitative | Same |
| Test Principle | Chromogenic | Same |
| Reporting Unit | ng/mL | Same |
| **General Device Characteristic Differences** | | |
| Instrumentation | For measuring Apixaban: AUTOMATED BLOOD COAGULATION ANALYZER CS-2500 (K172286)
For measuring Heparin:
• BCS XP System
• Automated Blood Coagulation Analyzer CA-600 series
• Automated Blood Coagulation Analyzer CS-2500
• Automated Blood Coagulation Analyzer CS-5100 | ACL TOP Family (K160276)
ACL TOP Family 50 Series (K150877) |
{5}
K240315 - Page 6 of 12
| Open Reagent Stability | 8 weeks | 1 month |
| --- | --- | --- |
| Composition | The INNOVANCE Anti-Xa kit includes the following components:
- Ready to use liquid containing: FXa, bovine (~0.7 IU/mL) buffers/stabilizers, preservatives pH 8
- Ready to use liquid containing: Suc-Ile-Glu(piperidin-1-yl)-Gly-Arg-pNA.HCl (1.25 mg/mL) buffers/stabilizers, preservatives pH 5
INNOVANCE Anti-Xa Size:
- INNOVANCE Anti-Xa Reagent: 5 x 3.2 mL vial of a liquid preparation.
INNOVANCE Anti-Xa Substrate: 5 x 4.0 mL vial of liquid chromogenic substrate. | The HemosIL Anti-Xa kit includes the following components:
- Factor Xa reagent: Liquid preparation containing purified bovine Factor Xa (approximately 5.5 nkat/mL), Tris-Buffer, EDTA, dextran sulfate, sodium chloride and bovine serum albumin.
- Chromogenic substrate: liquid chromogenic substrate S-2732 (Approximately 1.2 mg/mL) and bulking agent.
The HemosIL Anti-Xa kit sizes: 4 mL Kit Vial Size (Size 1):
- Factor Xa reagent (Cat. No. 0020302612): 5 x 2.5 mL vial of a liquid preparation containing purified bovine Factor Xa (approximately 5.5 nkat/mL), Tris- Buffer, EDTA, dextran sulfate, sodium chloride and bovine serum albumin.
- Chromogenic substrate (Cat. No. 0020302622): 5 x 3 mL vial of liquid chromogenic substrate S- 2732 (approximately 1.2 mg/mL) and bulking agent.
- Factor Xa reagent (Cat. No. 0020303610): 5 x 5 mL vial of a liquid preparation containing purified bovine Factor Xa (approximately 5.5 nkat/mL), Tris-Buffer, EDTA, dextran sulfate, sodium chloride and bovine serum albumin. |
| Linearity | 20 to 700 ng/mL for apixaban | 20 to 1000 ng/mL for apixaban |
| Calibrators | INNOVANCE Apixaban Standards
Target Levels: 0 and 420 ng/mL | HemosIL Apixaban Calibrators
Target Levels: 0 and 500 ng/mL |
| Controls | INNOVANCE Apixaban Controls
Target Levels: 70 and 260 ng/mL | HemosIL Apixaban Controls
Target Levels: 75 and 300 ng/mL |
{6}
VI Standards/Guidance Documents Referenced:
CLSI EP05-A3: Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline – Third Edition.
CLSI EP06: Evaluation of Linearity of Quantitative Measurement Procedures, 2nd Edition
CLSI EP07: Interference Testing in Clinical Chemistry, 3rd Edition.
CLSI EP17-A2: Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline – Second Edition.
CLSI EP34: Establishing and Verifying an Extended Measuring Interval Through Specimen Dilution and Spiking, 1st Edition.
CLSI EP37: Supplement Tables for Interference Testing in Clinical Chemistry, 1st Edition.
CLSI EP25: Evaluation of Stability of In Vitro Medical Laboratory Test Reagents, 2nd Edition.
ISO 14971:2019: Medical devices – Application of risk management to medical devices.
ISO 17511 Second edition 2020-04: In vitro diagnostic medical devices – Requirements for establishing metrological traceability of values assigned to calibrators, trueness control materials and human samples.
VII Performance Characteristics (if/when applicable):
A Analytical Performance:
1. Precision/Reproducibility:
The study was conducted over 20 days with two runs per day and two replicates per run for a total of 240 determinations (i.e., 80 determinations per reagent lot). This study was performed at three (3) internal sites with one SYSMEX CS-2500 instrument per site, by using five patient plasma pool samples and two QC materials with varying levels of apixaban. For investigating different assay and standard lots, each sample was analyzed by: 1) three reagent lots combined with one standard lot, 2) three standard lots combined with one reagent lot, 3) three control lots combined with one reagent and one standard lot. Precision estimates were calculated for each of the following variance components: repeatability (within-run), between-run, between-day, between-lot and total imprecision (within-laboratory). The results are provided in the summary table below.
Study 1 – Assay lots
| Sample | N | Mean | Repeatability | | Between-Run | | Between-Day | | Between-Lot | | Within-Laboratory | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | SD | CV% | SD | CV% | SD | CV% | SD | CV% | SD | CV% |
| A-PP1 | 240 | 31.4 | 0.9 | 2.93 | 0.5 | 1.70 | 1.6 | 3.39 | 0 | 0.00 | 1.9 | 6.19 |
| A-PP50 | 240 | 50.1 | 0.7 | 1.45 | 0.1 | 0.22 | 0.9 | 1.70 | 0.4 | 0.86 | 1.2 | 2.40 |
| A-C1 | 240 | 74.9 | 1.2 | 1.58 | 0.7 | 0.99 | 1.2 | 1.63 | 0.0 | 0.00 | 1.9 | 2.48 |
| A-PP2 | 240 | 95.8 | 1.2 | 1.25 | 0.0 | 0.00 | 1.5 | 1.58 | 0.2 | 0.22 | 1.9 | 2.03 |
| A-C2 | 240 | 277.6 | 6.1 | 2.20 | 1.9 | 0.69 | 3.0 | 1.08 | 4.4 | 1.57 | 8.3 | 2.57 |
| A-PP3 | 240 | 320.5 | 4.0 | 1.24 | 3.6 | 1.11 | 4.0 | 1.23 | 11.1 | 3.47 | 13.0 | 2.09 |
| A-PP4 | 240 | 503.5 | 5.3 | 1.05 | 9.2 | 1.83 | 3.8 | 0.75 | 7.8 | 1.55 | 13.7 | 2.73 |
| A-PP5 | 240 | 629.9 | 12.9 | 2.04 | 19.7 | 3.13 | 11.5 | 1.82 | 18.0 | 2.86 | 31.8 | 5.05 |
K240315 - Page 7 of 12
{7}
Study 2 – Standard lots
| Sample | N | Mean | Repeatability | | Between-Run | | Between-Day | | Between-Lot | | Within-Laboratory | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | SD | CV% | SD | CV% | SD | CV% | SD | CV% | SD | CV% |
| A-PP1 | 240 | 31.0 | 1.1 | 3.43 | 0.7 | 2.32 | 1.4 | 4.66 | 0.3 | 0.92 | 2.0 | 6.30 |
| A-PP50 | 240 | 51.1 | 0.8 | 1.59 | 0.3 | 0.57 | 0.9 | 1.80 | 1.4 | 2.80 | 1.9 | 3.73 |
| A-C1 | 240 | 74.6 | 1.3 | 1.70 | 0.9 | 1.18 | 1.1 | 1.44 | 0.0 | 0.00 | 1.9 | 2.52 |
| A-PP2 | 240 | 95.9 | 1.3 | 1.35 | 0.0 | 0.00 | 1.2 | 1.24 | 0.0 | 0.00 | 1.8 | 1.83 |
| A-C2 | 240 | 280.1 | 7.3 | 2.61 | 1.2 | 0.42 | 4.4 | 1.57 | 0.0 | 0.00 | 8.6 | 3.08 |
| A-PP3 | 240 | 331.0 | 5.9 | 1.79 | 4.0 | 1.21 | 2.7 | 0.81 | 6.7 | 2.02 | 10.2 | 3.07 |
| A-PP4 | 240 | 508.3 | 5.9 | 1.16 | 8.5 | 1.66 | 5.2 | 1.02 | 0.0 | 0.00 | 11.5 | 2.27 |
| A-PP5 | 240 | 647.3 | 16.5 | 2.55 | 23.1 | 3.57 | 14.6 | 2.25 | 12.1 | 1.87 | 34.1 | 5.27 |
Study 3 – Control lots
| Sample | N | Mean | Repeatability | | Between-Run | | Between-Day | | Between-Lot | | Within-Laboratory | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | SD | CV% | SD | CV% | SD | CV% | SD | CV% | SD | CV% |
| A-C1 | 240 | 74.8 | 1.3 | 1.68 | 0.9 | 1.15 | 1.1 | 1.53 | 0.7 | 0.89 | 2.0 | 2.70 |
| A-C2 | 240 | 282.3 | 7.7 | 2.72 | 3.4 | 1.21 | 3.3 | 1.18 | 3.4 | 1.20 | 9.7 | 3.42 |
The reproducibility study was conducted at three sites over five days, with two runs per day and three replicates per run for a total of 90 determinations. The study design included three SYSMEX CS-2500 instruments (one instrument per site), one assay lot of INNOVANCE Anti-Xa, one standard lot and one control lot. Five pooled patient plasmas and two QC materials representing variable levels of apixaban were evaluated. Precision estimates were calculated for each of the following variance components: within-run, between-run, between-day, between-site and total imprecision (reproducibility). The results for reproducibility are provided in the summary table below.
| Sample | N | Mean | Repeatability | | Between-Run | | Between-Day | | Between-Site | | Reproducibility | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | SD | CV% | SD | CV% | SD | CV% | SD | CV% | SD | CV% |
| A-PP1 | 90 | 33.5 | 1.8 | 5.32 | 0.00 | 0.00 | 0.9 | 2.66 | 4.1 | 12.37 | 4.6 | 13.73 |
| A-PP50 | 90 | 48.6 | 0.8 | 1.68 | 0.6 | 1.31 | 0.0 | 0.00 | 1.5 | 3.05 | 1.8 | 3.72 |
| A-C1 | 90 | 75.7 | 1.4 | 1.87 | 0.0 | 0.00 | 0.9 | 1.17 | 3.1 | 4.05 | 3.5 | 4.62 |
| A-PP2 | 90 | 102.3 | 0.6 | 1.57 | 1.0 | 0.93 | 0.7 | 0.68 | 3.6 | 3.48 | 4.1 | 3.99 |
| A-C2 | 90 | 291.9 | 13.2 | 4.53 | 0.0 | 0.00 | 3.0 | 1.03 | 14.0 | 4.48 | 19.5 | 6.68 |
| A-PP3 | 90 | 320.0 | 7.0 | 2.19 | 1.2 | 0.39 | 2.9 | 0.90 | 10.1 | 3.17 | 12.7 | 3.97 |
| A-PP4 | 90 | 458.4 | 8.0 | 1.75 | 9.4 | 2.06 | 2.3 | 0.49 | 13.8 | 3.00 | 18.6 | 4.07 |
| A-PP5 | 90 | 670.4 | 19.7 | 2.94 | 0.0 | 0.00 | 6.6 | 0.98 | 8.3 | 1.24 | 22.4 | 3.34 |
2. Linearity:
Linearity studies were performed using three reagent lots of INNOVANCE Anti-Xa on one SYSMEX CS-2500 instrument. A dilution series of apixaban in plasma with assigned values that were prepared by mixing blank (normal citrated pooled plasma containing no apixaban) with a high pool (normal citrated plasma pool spiked with apixaban). A total of 11 samples covering the range of 10.5 to $378\mathrm{ng / mL}$ apixaban were included in this study, with each sample tested in four replicates. A dilution recovery study was conducted to establish extend measuring interval. Samples with five different concentrations were prepared by spiking
K240315 - Page 8 of 12
{8}
apixaban. Each sample was separated in two aliquots, and each aliquot was manually diluted two times using Standard Human Plasma. Each dilution was measured two times by two operators so the total number of measurements per sample is eight.
Based on the results of the Linearity studies and Detection Limit studies, the claimed assay reportable range is 20–700 ng/mL.
## 3. Analytical Specificity/Interference:
Interference studies were conducted based on the CLSI EP07, 3rd Edition. The study evaluated potentially interfering endogenous and exogenous substances with one reagent lot and one standard lot on one SYSMEX CS-2500 instrument using a dose-response method. Potentially interfering endogenous and exogenous substances were spiked into low plasma pool (containing 30–50 ng/mL apixaban) and high plasma pool (containing 280–350 ng/mL apixaban) in nine dilution levels and each sample level were run in four replicates.
Apixaban results are not affected by the interferences up to the respective concentrations listed in the table below:
| Interfering Substances | No interference up to |
| --- | --- |
| Hemoglobin | 191 mg/dL |
| Bilirubin, conjugated | 31 mg/dL |
| Bilirubin, unconjugated | 17 mg/dL |
| Lipids | 150 mg/dL |
| Rivaroxaban | 5.3 ng/mL |
| Fondaparinux | 60.0 ng/mL |
| Danaparoid sodium | 0.05 U/mL |
| LMWH | 0.07 IU/mL |
| UFH | 0.06 IU/mL |
| Edoxaban | 5.8 ng/mL |
| Warfarin | 5.340 mg/dL |
| Atorvastatin | 0.077 mg/dL |
| Isosorbide dinitrate | 0.600 mg/dL |
## 4. Assay Reportable Range:
The reportable range is 20–700 ng/mL.
## 5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
Traceability
K240315 - Page 9 of 12
{9}
A traceability and uncertainty of the analytical values for INNOVANCE Apixaban Standard 1 and Controls to purified apixaban reference material for the assay INNOVANCE Anti-Xa was conducted on the Automated Blood Coagulation Analyzer CS-2500, to evaluate if standards/controls can be related to stated references, usually national or international standards, through an unbroken chain of comparisons all having stated uncertainties. Since no international certified reference material for apixaban is available, the study was performed with 11 plasmas spiked with "In house" reference material (a stock solution of weighed apixaban) dissolved in dimethyl sulfoxide and with values assigned to defined concentrations of apixaban in plasma. Value assignments are traceable to purified apixaban reference material.
Value assignment of the commercial INNOVANCE Apixaban Standards is performed on the SYSMEX CS-2500 using a Master Standard lot which is traceable to the purified Reference Material.
Apixaban Standard 0: 0 ng/mL
Apixaban Standard 1: 420 ng/mL
Value assignment of the commercial INNOVANCE Apixaban Controls is performed on the SYSMEX CS-2500 using released INNOVANCE Apixaban Standards lots which are calibrated against a Master Standards lot that is traceable to the purified Reference Material.
Apixaban Control 1: 77 ng/mL
Apixaban Control 2: 284 ng/mL
## Reagent stability
The stability study results support the following claim:
The shelf-life for INNOVANCE Anti-Xa is 24 months when stored at 2–8°C. The shelf-life for INNOVANCE Apixaban Standards and Controls is 30 months when stored at 2–8°C.
The in-use stability for INNOVANCE Anti-Xa is 8 weeks at 2–8°C. The in-use stability for INNOVANCE Apixaban Standards is 40 hours at 2–8°C and up to 12 hours at 15–25°C. The in-use stability for INNOVANCE Apixaban Controls is 48 hours at 2–8°C, up to 20 hours at 15–25°C and 4 weeks at ≤ -18°C.
The on-board stability for INNOVANCE Anti-Xa is 24 hours. The on-board stability for INNOVANCE Apixaban Standards is 4 hours. The on-board stability for INNOVANCE Apixaban Controls is 5 hours when transferred in Sample Cup and placed on the sample rack, 24 hours when transferred to SLD mini cup and placed on the reagent table.
## Sample Stability
Sample stability studies were performed to support the recommended storage and handling instructions found in the device labeling. Twenty nine citrated plasma samples (4 samples with an apixaban concentration ≤ 50 ng/mL, 25 samples with an apixaban concentration ≥ 50 ng/mL) were tested after storage in the following temperature ranges: at 2 to 8°C in primary tube (plasma stored over cells) and in secondary tube (plasma siphoned from cells), at 15 to 25°C in primary tube and in secondary tube, at ≤ -18°C in secondary tube, at ≤ -70°C in secondary tube and at ≤ -70°C (one frozen thaw cycle). The study was performed using one reagent lot, one standard lot, and one control lot, with four replicate measurements at each time point for each sample. The study data demonstrate that apixaban in citrated plasma samples can be stored at 15–25°C in primary tube (plasma stored over cells) and in
K240315 - Page 10 of 12
{10}
secondary tube (plasma siphoned from cells) for 4 hours; stored in secondary tube at ≤ -18°C for one month; stored in secondary tube at ≤ -74°C for 5 months.
## 6. Detection Limit:
The limit of blank (LoB), limit of detection (LoD), and limit of quantitation (LoQ) for the test system was determined following the CLSI EP17-A2 guideline. Each study design included two reagent lots on one SYSMEX CS-2500 instrument.
The limit of blank (LoB) was determined using five independent analyte-free samples tested with four replicates per sample, two reagent lots and two standard lots over three days on one system, with a total of 60 determinations per reagent lot. The LoB was determined to be 1.0 ng/mL.
Limit of detection (LoD) was determined using five low analyte samples prepared by spiking platelet-poor plasma from five donors containing no apixaban, tested with four replicates per sample, with total 60 determinations per reagent lot. The LoD was determined to be 3.3 ng/mL.
The limit of quantitation (LoQ) was determined using five low analyte samples prepared by spiking platelet-poor plasma from five donors containing no apixaban, tested with four replicates per sample, with total 60 determinations per reagent lot. The LoQ was determined to be 15.8 ng/mL.
## 7. Assay Cut-Off:
Not applicable
## B Comparison Studies:
### 1. Method Comparison with Predicate Device:
The method comparison was performed by using fresh and frozen samples to compare the INNOVANCE Anti-Xa assay on the CS-2500 System to the HemosIL Liquid Anti-Xa on the ACL TOP for the apixaban measurement. The study was conducted by evaluating a total of 301 clinical samples (up to 10% contrived samples spiked with apixaban) collected fresh and/or after frozen storage at three different U.S. clinical sites and one site outside of the U.S. Passing-Bablok regression analyses were performed for the dataset collected for each site and all sites combined. The following table summarizes the regression analysis performed for the combined dataset.
| N | Range | Coefficient of correlation | Slope (95% CI) | Intercept (95%CI) |
| --- | --- | --- | --- | --- |
| 301 | 23–602 ng/mL | 0.989 | 1.026 (1.006, 1.047) | -8.590 (-11.593, -5.677) |
### 2. Matrix Comparison:
Not applicable
## C Clinical Studies:
K240315 - Page 11 of 12
{11}
K240315 - Page 12 of 12
1. Clinical Sensitivity:
Not applicable
2. Clinical Specificity:
Not applicable
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Not applicable
D. Clinical Cut-Off:
Not applicable
E. Expected Values/Reference Range:
Not applicable
VIII. Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX. Conclusion:
The submitted information in this premarket notification is complete and does support a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.