TENS (Mode 1~6) To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, back of the neck, arm, leg, and foot due to strain from exercise or normal household work activities by applying current to stimulate nerve. EMS (Mode 1~6) It is intended to stimulate healthy muscles in order to improve and facilitate muscle performance.
Device Story
Transcutaneous Electrical Nerve Stimulator (9029SCM) is a portable, battery-powered device for home use. It delivers electrical pulses via dual-channel output to electrode patches attached to the patient's skin. The device features 12 total modes (6 TENS, 6 EMS) and 18 intensity levels, controlled via an LCD interface. It operates by transmitting biphasic symmetric square wave pulses to peripheral nerves for pain relief or to muscles for performance improvement. The device is intended for over-the-counter use. Healthcare providers or patients use the device to manage muscle pain or facilitate muscle conditioning. Benefits include non-invasive pain management and muscle stimulation.
Clinical Evidence
No clinical data. Bench testing only. Compliance verified with IEC 60601-1 (general safety), IEC 60601-1-2 (EMC), IEC 60601-1-11 (home healthcare), and IEC 60601-2-10 (nerve/muscle stimulators). Biocompatibility of electrode patches verified per ISO 10993-5 and ISO 10993-10.
Technological Characteristics
ABS plastic housing; Type BF applied part; 4.5V DC power (3x AAA batteries); dual-channel output; biphasic symmetric square waveform. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-10, ISO 10993-5, ISO 10993-10.
Indications for Use
Indicated for temporary relief of pain in shoulder, waist, back, neck, arm, leg, and foot due to exercise or household strain (TENS mode); indicated for stimulation of healthy muscles to improve/facilitate muscle performance (EMS mode).
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 22, 2024
Hong Kong Etech Groups Limited % Rilev Chen RA Specialist Feiying Drug & Medical Consulting Technical Service Group Rm 2401 Zhenye International Business Center, No. 3101-90 Oianhai Road Shenzhen, Guangdong 518052 China
Re: K240292
Trade/Device Name: Transcutaneous Electrical Nerve Stimulator (9029SCM) Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: Class II Product Code: NUH, NGX Dated: January 31, 2024 Received: February 1, 2024
Dear Riley Chen:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Tushar Bansal -S
for Heather Dean, PhD
Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation
{2}------------------------------------------------
Enclosure
{3}------------------------------------------------
### Indications for Use
Submission Number (if known)
K240292
Device Name
Transcutaneous Electrical Nerve Stimulator (9029SCM)
Indications for Use (Describe)
TENS (Mode 1~6)
To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, back of the neck, arm, leg, and foot due to strain from exercise or normal household work activities by applying current to stimulate nerve.
EMS (Mode 1~6)
It is intended to stimulate healthy muscles in order to improve and facilitate muscle performance.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
| 510(k) #: K240292 | | 510(k) Summary | Prepared on: 2024-03-22 |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|----------------|-------------------------|
| Contact Details<br>21 CFR 807.92(a)(1) | | | |
| Applicant Name | HONG KONG ETECH GROUPS LIMITED | | |
| Applicant Address | RM 747,7/F STAR HSE 3 SALISBURY RD TST Hong Kong Hong Kong<br>999077 China | | |
| Applicant Contact Telephone | +86159197487099 | | |
| Applicant Contact | Ms. Haixin Chen | | |
| Applicant Contact Email | chinaetech65@etechgroups.com | | |
| Correspondent Name | Feiying Drug & Medical Consulting Technical Service Group | | |
| Correspondent Address | Rm 2401 Zhenye International Business Center, No. 3101-90,<br>Qianhai Road Shenzhen Guangdong 518052 China | | |
| Correspondent Contact Telephone | +86 13660660449 | | |
| Correspondent Contact | Ms. Riley Chen | | |
| Correspondent Contact Email | c3714930@gmail.com | | |
| Device Name<br>21 CFR 807.92(a)(2) | | | |
| Device Trade Name | Transcutaneous Electrical Nerve Stimulator (9029SCM) | | |
| Common Name | Transcutaneous electrical nerve stimulator for pain relief | | |
| Classification Name | Stimulator, Nerve, Transcutaneous, Over-The-Counter | | |
| Regulation Number | 882.5890 | | |
| Product Code(s) | NUH, NGX | | |
| Legally Marketed Predicate Devices<br>21 CFR 807.92(a)(3) | | | |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | | Product Code |
| K191982 | Low-frequency Multi-function physiotherapy instrument | | NUH |
| K133929 | Health Expert Electronic Stimulator, Model: AST-300C and AST+ | | NUH |
| Device Description Summary<br>21 CFR 807.92(a)(4) | | | |
| 1)Stylish appearance, compact size, humanized design for easy portability.<br>2)TENS and EMS modes with 6 different modes each, catering to various treatment needs and a wider range of users.<br>3)Dual-channel output, allowing control of intensity for each channel.<br>4)Unique combination of electric pulses for a different sensory experience.<br>5)LCD display for clear and easy use.<br>6)Powered by three AAA batteries (size 7), providing long-lasting usage and convenient for home and travel use.<br>7) The electrode patches used with the device are previously cleared under K221589 for safety use. The main unit delivers the electrical | | | |
{5}------------------------------------------------
## Intended Use/Indications for Use
21 CFR 807.92(a)(5)
21 CFR 807.92(a)(5)
21 CFR 807.92(a)(6)
TENS (Mode 1~6)
To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back of the neck, arm, leg, and foot due to strain from exercise or normal household work activities by applying current to stimulate nerve. EMS (Mode 1~6)
It is intended to stimulate healthy muscles in order to improve and facilitate muscle performance.
### Indications for Use Comparison
The subject device and predicate devices have the same indications for use.
### Technological Comparison
The technical characteristics of Transcutaneous Electrical Nerve Stimulator (Model:9029SCM) are substantially equivalent to the predicate devices in the following aspects:
1. the same intended use: both use electrical stimulation to temporary relief of pain under TENS mode and facilitate musce performance by using EMS mode.
2. operate with the same mode of action: the device would generate electrical pulses and transmit it to the electrodes, which are attached to the patient's skin to the underlying peripheral nerves and muscle.
3. the same design: two output channel for voltage controly, which can be adjusted separately, and also design with display screen for treatment parameters display.
4. similar output parameters, such as Maximum Output Current, Pulse frequency, Maximum Phase Charge, Maximum Average Current, Maximum Current Density(t.m.s.), and Maximum Average Power Density. The product waveform parameters were tested according to FDA guidance documents and the requirements of IEC 60601-2-10, the tests are all passed.
5. the same waveform and shape: biphasic symmetric square wave, rectangular
6. the same safety features: Type BF applied part.
7. the same housing materials: ABS plastic for main unit.
8. similar enrgy source as the primary predicate device: both use three AAA batteries (4.5V DC), and they all pass electrical safety test.
The difference between subject device and the predicate devices mainly includes the following:
1. The numbers of output modes (12 modes) and the output intensity level (18 gears) are different from those of the predicate devices (50 modes and 16 steps for K191982, 25 modes and 99 steps for K133929), but the output parameters of each mode were tested and found to be similar with the predicate devices, and the subject device has passed IEC 60601-2-10 test, so these differences do not lead to safety and effectiveness problems.
2. The weight (main unit: 11.5g) and dimensions (108.5x33.8x67.5mm) of the subject device are different from the predicate devices, but these differences are insignificant and will not affect the safety and effectiveness of the device.
Thus, the subject device is determined to be substantially equivalent to the predicate devices.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
Nor-clinical testings have been conducted to verify that the Transcutaneous Electrical Nerve Stimulator meets all design specifications which supports the conclusion that it's Substantially Equivalent (SE) to the predicate devices. The testing results demonstrate that the subject device complies with the following standards:
网 IEC 60601-1, Medical electrical equipment -- Part 1: General requirements for basic safety and essential performance മ IEC 60601-1-2, Medical electrical equipments for basic safety and essential performance - Collateral standard: Electromagnetic disturbances - Requirements and tests
| EC 60601-1-11, Medical electrical equipment -- Part 1-11: General requirements for basic safety and essential performance -- Collateral standard: Requirements for medical electrical electrical systems used in the home healthcare environment മ IEC 60601-2-10, Medical electrical equipment -- Part 2-10: Particular requirements for the basic safety and essential performance of nerve and muscle stimulators
The body-contacting components of this device are electrode patches the electrode patches from qualified supplier which has obtained FDA clearance with a 510(k) number of K221589 and been marketed to US market. So we have reason to believe that the electrode patches are safe for the users. The electrode patches comply with the following standards. இ ISO 10993-5, Biological Evaluation of Medical Devices -- Part 5: Tests for InVitroCytotoxicity
図SO 10993-10, Biological Evaluation of Medical Devices - Part 10: Tests for Irritation and Skin Sensitization.
{6}------------------------------------------------
We have also conducted:
l Software verification and validation test according to the requirements of the FDA "Guidance for Pre Market Submissions and for Software Contained in Medical Devices" .
The waveform test report has also been conducted to verify the output specifications of the device according to "Guidance Document for Powered Muscle Stimulator 510(k)s"
The clinical test is not applicable, there is no clinical data.
The subject device and predicate devices have indications for use, technological characteristics. The subject device is as safe, as effective and performs as well as the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.