K240279 · bioMerieux, Inc. · QAT · May 1, 2024 · Immunology
Device Facts
Record ID
K240279
Device Name
VIDAS TBI (GFAP, UCH-L1)
Applicant
bioMerieux, Inc.
Product Code
QAT · Immunology
Decision Date
May 1, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5830
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K240279 · May 1, 2024
VIDAS TBI (GFAP, UCH-L1)
bioMerieux, Inc.
Frozen serum specimens from the ALERT-TBI prospective multicenter study; Frozen serum specimens from the BRAINI multicenter study
The sponsor used retrospective clinical samples from two established cohorts to evaluate the clinical sensitivity and specificity of the VIDAS TBI (GFAP, UCH-L1) test as an aid in determining the need for a head CT scan.
BRAINI study; Retrospective analysis of stored serum samples from a multicenter study; Follow-up/Duration: Not applicable; Study Period: Samples collected within 4 years prior to analysis
Subjects 18+ years old presenting with suspected mild TBI (GCS 13-15) within 12 hours of injury; Sample Size: 562; Number of Sites: 3
The VIDAS TBI (GFAP, UCH-L1) test is composed of two automated assays - VIDAS TBI (GFAP) and VIDAS TBI (UCH-L1) - to be used on the VIDAS 3 instrument for the quantitative measurement of Glial Fibrillary Acidic Protein (GFAP) and Ubiquitin C-terminal Hydrolase (UCH-L1) in human serum using the ELFA (Enzyme Linked Fluorescent Assay) technique. The results of both assays are required to obtain an overall qualitative test interpretation. The overall qualitative VIDAS TBI (GFAP, UCH-L1) test result is used, in conjunction with clinical information, to aid in the evaluation of patients (18 years of age or older), presenting within 12 hours of suspected mild traumatic brain injury (Glasgow Coma Scale score 13-15), to assist in determining the need for a Computed Tomography (CT) scan of the head. A negative interpretation of VIDAS TBI (GFAP, UCH-L1) test is associated with the absence of acute intracranial lesions visualized on a head CT scan.
Device Story
VIDAS TBI (GFAP, UCH-L1) is an automated in vitro diagnostic test for human serum; utilizes ELFA (Enzyme Linked Fluorescent Assay) on the VIDAS 3 instrument. Input: serum sample; process: three-step sandwich immunoassay using SPRs (Solid Phase Receptacles) coated with anti-GFAP or anti-UCH-L1 antibodies; output: quantitative biomarker concentrations and qualitative interpretation. Used in clinical settings by laboratory professionals. Instrument automatically calculates concentrations via calibration curves and MLE (Master Lot Entry) data. Clinicians use results alongside patient history to assess mild TBI; negative results correlate with absence of acute intracranial lesions on head CT, potentially reducing unnecessary imaging.
Clinical Evidence
Clinical validation used 1,911 samples from the ALERT-TBI study and 562 from the BRAINI study. ALERT-TBI cohort: 120 CT-positive, 1791 CT-negative. Sensitivity 96.7% (91.7-99.1%), Specificity 41.2% (38.9-43.5%), NPV 99.5% (98.6-99.9%). BRAINI cohort: 74 CT-positive, 488 CT-negative. Sensitivity 97.3% (90.6-99.7%), Specificity 34.2% (30.1-38.5%), NPV 98.8% (95.8-99.9%).
Technological Characteristics
ELFA (Enzyme Linked Fluorescent Assay) sandwich immunoassay. Materials: SPRs coated with specific antibodies. Energy: Automated instrument-based. Connectivity: Master Lot Entry (MLE) via barcode. Standards: ISO 17511:2020 for traceability; CLSI EP05-A3, EP06-Ed2, EP07-Ed3, EP17-A2, EP28-A3 for analytical validation.
Indications for Use
Indicated for patients 18+ years old presenting within 12 hours of suspected mild traumatic brain injury (GCS 13-15) to assist in determining the need for a head CT scan. Negative results are associated with the absence of acute intracranial lesions.
Regulatory Classification
Identification
A brain trauma assessment test is a device that consists of reagents used to detect and measure brain injury biomarkers in human specimens. The measurements aid in the evaluation of patients with suspected mild traumatic brain injury in conjunction with other clinical information to assist in determining the need for head imaging per current standard of care.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) The 21 CFR 809.10(b) compliant labeling must include detailed descriptions of and results from performance testing conducted to evaluate precision, accuracy, linearity, analytical sensitivity, interference, and cross-reactivity. This information must include the following:
(i) Performance testing of device precision must, at minimum, use one unmodified clinical specimen from the intended use population with concentration of the brain injury biomarker(s) near the medical decision point. Contrived specimens that have been generated from pooling of multiple samples or spiking of purified analyte to cover the measuring range may be used, but the contrived samples must be prepared to mimic clinical specimens as closely as possible. This testing must evaluate repeatability and reproducibility using a protocol from an FDA-recognized standard.
(ii) Device performance data must be demonstrated through a clinical study and must include the following:
(A) Data demonstrating clinical validity including the clinical sensitivity and specificity, and positive and negative predictive value of the test in the intended use population of patients with suspected mild traumatic brain injury (
*i.e.,* Glasgow Coma Score (GCS) of 13-15), or equivalent standard of care for determination of severity of traumatic brain injury (TBI).(B) Study must be performed using the operators and in settings that are representative of the types of operators and settings for which the device is intended to be used.
(C) All eligible subjects must meet the well-defined study inclusion and exclusion criteria that define the intended use population. The prevalence of diseased or injured subjects in the study population must reflect the prevalence of the device's intended use population, or alternatively, statistical measures must be used to account for any bias due to enrichment of subpopulations of the intended use population.
(D) All eligible subjects must have undergone a head computerized tomography (CT) scan or other appropriate clinical diagnostic standard used to determine the presence of an intracranial lesion as part of standard of care and must also be evaluated by the subject device. All clinical diagnostic standards used in the clinical study must follow standard clinical practice in the United States.
(E) Relevant demographic variables and baseline characteristics including medical history and neurological history. In addition, head injury characteristics, neurological assessments, and physical evidence of trauma must be provided for each subject. This information includes but is not limited to the following: Time since head injury, time from head injury to CT scan, time from head injury to blood draw, GCS score or equivalent, experience of loss of consciousness, presence of confusion, episodes of vomiting, post-traumatic amnesia characteristics, presence of post-traumatic seizures, drug or alcohol intoxication, mechanism of injury, acute intracranial lesion type, neurosurgical lesion, and cranial fracture.
(F) Each CT scan or other imaging result must be independently evaluated in a blinded manner by at least two board-certified radiologists to determine whether it is positive or negative as defined by the presence or absence of acute intracranial lesions. This independent review must be conducted without access to test results of the device. Prior to conducting the review, the criteria and procedures to be followed for scoring the images must be established, including the mechanism for determining consensus.
(G) All the clinical samples must be tested with the subject device blinded to the TBI status and the neurological-lesion-status of the subject.
(H) Details on how missing values in data are handled must be provided.
(I) For banked clinical samples, details on storage conditions and storage period must be provided. In addition, a specimen stability study must be conducted for the duration of storage to demonstrate integrity of archived clinical samples. The samples evaluated in the assay test development must not be used to establish the clinical validity of the assays.
(iii) Performance testing of device analytical specificity must include the most commonly reported concomitant medications present in specimens from the intended use population. Additionally, potential cross-reacting endogenous analytes must be evaluated at the highest concentration reported in specimens from the intended use population.
(iv) Expected/reference values generated by testing a statistically appropriate number of samples from apparently healthy normal individuals.
(2) The 21 CFR 809.10(a) and (b) compliant labeling must include the following limitations:
(i) A limiting statement that this device is not intended to be used a stand-alone device but as an adjunct to other clinical information to aid in the evaluation of patients who are being considered for standard of care neuroimaging.
(ii) A limiting statement that reads “A negative result is generally associated with the absence of acute intracranial lesions. An appropriate neuroimaging method is required for diagnosis of acute intracranial lesions.”
(iii) As applicable, a limiting statement that reads “This device is for use by laboratory professionals in a clinical laboratory setting.”
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY ONLY
## I Background Information:
A 510(k) Number
K240279
B Applicant
bioMérieux, Inc.
C Proprietary and Established Names
VIDAS TBI (GFAP, UCH-L1)
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| QAT | Class II | 21 CFR 866.5830 - Brain Trauma Assessment Test | IM - Immunology |
## II Submission/Device Overview:
A Purpose for Submission:
New device
B Measurand:
Glial Fibrillary Acidic Protein (GFAP) and Ubiquitin C-terminal Hydrolase (UCH-L1)
C Type of Test:
Automated Enzyme Linked Fluorescent Assay (ELFA), Quantitative
## III Intended Use/Indications for Use:
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
K240279 - Page 2 of 24
A Intended Use(s):
See Indications for Use below.
B Indication(s) for Use:
The VIDAS TBI (GFAP, UCH-L1) test is composed of two automated assays - VIDAS TBI (GFAP) and VIDAS TBI (UCH-L1) - to be used on the VIDAS 3 instrument for the quantitative measurement of Glial Fibrillary Acidic Protein (GFAP) and Ubiquitin C-terminal Hydrolase (UCH-L1) in human serum using the ELFA (Enzyme Linked Fluorescent Assay) technique. The results of both assays are required to obtain an overall qualitative test interpretation.
The overall qualitative VIDAS TBI (GFAP, UCH-L1) test result is used, in conjunction with clinical information, to aid in the evaluation of patients (18 years of age or older), presenting within 12 hours of suspected mild traumatic brain injury (Glasgow Coma Scale score 13-15), to assist in determining the need for a Computed Tomography (CT) scan of the head. A negative interpretation of VIDAS TBI (GFAP, UCH-L1) test is associated with the absence of acute intracranial lesions visualized on a head CT scan.
C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
For in vitro diagnostic use
D Special Instrument Requirements:
VIDAS 3 instrument (K141133)
IV Device/System Characteristics:
A Device Description:
The VIDAS TBI (GFAP, UCH-L1) test is composed of two assays: VIDAS TBI (GFAP) and VIDAS TBI (UCH-L1), each with sufficient reagents for 60 determinations.
VIDAS TBI (GFAP) kit contains the following:
- GFAP Strips (STR) (60): Ready-to-use.
- GFAP solid phase receptacles (SPR) (2x30): coated with mouse monoclonal IgG anti-GFAP antibodies. Ready-to-use.
- GFAP Calibrators (S1) (1x2.4 mL): Recombinant human GFAP diluted in buffer + stabilizer + preservative.
- GFAP Control (C1) (1x1.6 mL): Recombinant human GFAP diluted in buffer + stabilizer + preservative.
VIDAS TBS (UCH-L1) kit contains the following:
- UCH Strips (STR) (60): Ready-to-use.
{2}
- UCH Solid Phase Receptacles (SPR) (2x30): coated with mouse monoclonal IgG anti-UCH-L1 antibodies. Ready-to-use.
- UCH Calibrators (S1) (1x2.4 mL): Recombinant human UCH-L1 diluted in buffer + stabilizer + preservative.
- UCH-L1 Control (C1) (1x1.6 mL): Recombinant human UCH-L1 diluted in buffer + stabilizer + preservative.
The master lot entry (MLE) barcode printed on the outer label of the packaging contains specifications for assay calibration and concentration information of calibrator (S1) and control (C1) for each assay.
Both assays are to be used on VIDAS 3 instrument.
## B Principle of Operation:
The VIDAS TBI (GFAP, UCH-L1) test is composed of two assays, VIDAS TBI (GFAP) and VIDAS (UCH-L1). Each combines a three-step enzyme immunoassay sandwich method with a final fluorescent detection (ELFA). All the assay steps are performed automatically by the VIDAS 3 instrument. The sample is transferred into the well containing sample diluent. The sample/diluent mixture is cycled in and out of the Solid Phase Receptable (SPR) device several times to allow the antigen (GFAP or UCH-L1) to bind with the immunoglobulins that are fixed to the interior wall of the SPR device. The bounded complex is presented to antibodies labeled with biotin (conjugate) to form a sandwich immunocomplex. For the third immunological step, the complex is presented to anti-biotin antibodies labeled with alkaline phosphatase (tracer) using the SPR and form the final immunological complex "Antibody-Analyte-Conjugate-Tracer". Unbound tracer is eliminated during washing steps. During the final detection step, the substrate (4-Methylumbelliferyl phosphate) is cycled in and out of the SPR device. The conjugate enzyme catalyzes the hydrolysis of this substrate into a fluorescent product (4-Methylumbelliferone), which is measured at 450 nm. The intensity of the fluorescence is proportional to the concentration of antigen in the sample. At the end of each assay, separate results of VIDAS TBI (GFAP) and VIDAS TBI (UCH-L1) are automatically calculated by the instrument.
Final VIDAS TBI (GFAP, UCH-L1) test interpretation is performed manually by the user. The results interpretation is shown in the following table:
| GFAP Assay Result (relative to cut-off)^{A} | UCH-L1 Assay Result (relative to cut-off)^{B} | VIDAS TBI (GFAP, UCH-L1) Interpretation |
| --- | --- | --- |
| Negative | Negative | Negative |
| Negative, Invalid | Positive | Positive |
| Positive | Negative, Invalid | Positive |
| Positive | Positive | Positive |
| Invalid | Negative | Not Reportable^{C} |
| Negative | Invalid | Not Reportable^{C} |
| Invalid | Invalid | Not Reportable^{C} |
A "Positive" means the GFAP concentration is equal to or above the cut-off value of 22.0 pg/mL and "Negative" means the GFAP concentration is below the cut-off value of 22.0 pg/mL.
B "Positive" means the UCH-L1 concentration is equal to or above the cut-off value of 327.0 pg/mL and "Negative" means the UCH-L1 concentration is below the cut-off value of 327.0 pg/mL.
C Samples with "Invalid" results that yield a "Not Reportable" VIDAS TBI (GFAP, UCH-L1) result may be retested once to obtain a "Negative" or "Positive" result.
K240279 - Page 3 of 24
{3}
A positive clinical interpretation of VIDAS TBI (GFAP, UCH-L1), in conjunction with clinical assessment, can suggest and support the need for a head CT (Computed Tomography) scan. In contrast, a negative clinical interpretation is associated with absence of acute intracranial lesions visualized on a head CT scan.
# V Substantial Equivalence Information:
A Predicate Device Name(s):
Banyan BTI
B Predicate 510(k) Number(s):
DEN170045
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K240279 | DEN170045 (Predicate) |
| --- | --- | --- |
| Device Trade Name | VIDAS TBI (GFAP, UCH-L1) | Banyan BTI |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | The VIDAS TBI (GFAP, UCH-L1) test is composed of two automated assays - VIDAS TBI (GFAP) and VIDAS TBI (UCH-L1) - to be used on the VIDAS 3 instrument for the quantitative measurement of Glial Fibrillary Acidic Protein (GFAP) and Ubiquitin C-terminal Hydrolase (UCH-L1) in human serum using the ELFA (Enzyme Linked Fluorescent Assay) technique. The results of both assays are required to obtain an overall qualitative test interpretation.The overall qualitative VIDAS TBI (GFAP, UCH-L1) test result is used, in conjunction with clinical information, to aid in the evaluation of patients (18 years of age or older), presenting within 12 hours of suspected mild traumatic brain injury (Glasgow Coma Scale | The Banyan BTI is an in vitro diagnostic chemiluminescent enzyme-linked immunosorbent assay (ELISA). The assay provides a semi-quantitative measurement of the concentrations of ubiquitin C-terminal hydrolase-L1 (UCH-L1) and glial fibrillary acidic protein (GFAP) in human serum and is used with the Synergy 2 Multi-mode Reader.The assay results obtained from serum collected within 12 hours of suspected head injury are used, along with other available clinical information, to aid in the evaluation of patients 18 years of age and older with suspected traumatic brain injury (Glasgow Coma Scale score 13-15).A negative assay result is associated with the absence of |
K240279 - Page 4 of 24
{4}
| | score 13-15), to assist in determining the need for a Computed Tomography (CT) scan of the head. A negative interpretation of VIDAS TBI (GFAP, UCH-L1) test is associated with the absence of acute intracranial lesions visualized on a head CT scan. | acute intracranial lesions visualized on a head CT Computed Tomography) scan. The Banyan BTI is for prescription use only. |
| --- | --- | --- |
| Intended Use Setting | Clinical laboratory | Same |
| Measurands | GFAP and UCH-L1 | Same |
| Assay Technology | Enzyme-linked fluorescent assay | Chemiluminescent immunoassay |
| Reportable Result | Quantitative results for GFAP and UCH-L1 with a qualitative interpretation of test results | Same |
| Assay Format | Two separate test kits | Same |
| Specimen Type | Serum | Same |
| Analytical Measuring Interval | GFAP: 10.0 – 320.0 pg/mL
UCH-L1: 80.0 – 2560.0 pg/mL | Same |
| Assay Cut-off | GFAP: 22.0 pg/mL
UCH-L1: 327.0 pg/mL | Same |
| General Device Characteristic Differences | | |
| Assay Procedure | Automated immunoassay | Manual ELISA |
| Sample Volume | 200 μL for VIDAS TBI (GFAP)
200 μL for VIDAS TBI (UCH-L1) | 150 μL for Banyan GFAP
100 μL for Banyan UCH-L1 |
| Time to result | ~ 40 min | ~ 4 hours |
VI Standards/Guidance Documents Referenced:
The following Clinical and Laboratory Standards Institute (CLSI) guidelines were used:
- CLSI EP05-A3, Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline – Third Edition
- CLSI EP06-Ed2, Evaluation of the Linearity of Quantitative Measurement Procedures – Second Edition
- CLSI EP07-A3, Interference Testing in Clinical Chemistry – Third Edition
- CLSI EP09c 3rd Edition, Measurement Procedure Comparison and Bias Estimation Using Patient Samples
- CLSI EP17-A2, Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline – Second Edition
- CLSI EP25-A, Evaluation of Stability of In Vitro Diagnostic Reagents; Approved Guideline
K240279 - Page 5 of 24
{5}
- CLSI EP28-A3c, Defining Establishing and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline – Third Edition
- CLSI EP35, Assessment of Equivalence or Suitability of Specimen Types for Medical Laboratory Measurement Procedures – First Edition
- CLSI EP37, Supplemental Tables for Interference Testing in Clinical Chemistry – First Edition
## VII Performance Characteristics (if/when applicable):
### A Analytical Performance:
All results met the manufacturer’s pre-determined acceptance criteria.
#### 1. Precision/Reproducibility:
##### a. Within-Laboratory Precision:
A study was conducted per CLSI guideline EP05-A3 to evaluate the within-laboratory precision of the VIDAS TBI (GFAP) and VIDAS TBI (UCH-L1) assays. Two separate panels, one for each assay and each consisting of five human serum samples, were tested over the course of 20 days at one site using one VIDAS 3 instrument and one assay lot. Each panel includes two serum samples close to the respective assay cut-off value, and three other samples to cover the analytical measuring interval of each assay. Samples were tested each day with two runs per day, two replicate measurements per run for a total of 80 replicates per sample.
The data were analyzed quantitatively and qualitatively. The results are summarized in the tables below.
**Quantitative analysis:**
| GFAP Assay | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Sample | N | Mean (pg/mL) | Within-Run | | Between-Run | | Between-Day | | Within-Laboratory | |
| | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| PP01^{A} | 80 | 16.39 | 0.85 | 5.2 | 0.00 | 0.0 | 0.26 | 1.6 | 0.89 | 5.4 |
| PP02^{A} | 80 | 32.45 | 1.49 | 4.6 | 0.00 | 0.0 | 0.48 | 1.5 | 1.56 | 4.8 |
| PP03^{A} | 80 | 82.63 | 2.06 | 2.5 | 2.45 | 3.0 | 0.00 | 0.0 | 3.21 | 3.9 |
| PP04^{B} | 80 | 209.22 | 5.83 | 2.8 | 0.00 | 0.0 | 2.69 | 1.3 | 6.42 | 3.1 |
| PP05^{B} | 80 | 298.17 | 9.52 | 3.2 | 4.51 | 1.5 | 2.14 | 0.7 | 10.74 | 3.6 |
$^{\mathrm{A}}$ Native human sera from a subject with traumatic brain injury
$^{\mathrm{B}}$ Native human sera from subject with meningeal hemorrhage
K240279 - Page 6 of 24
{6}
K240279 - Page 7 of 24
| UCH-L1 Assay | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Sample | N | Mean (pg/mL) | Within-Run | | Between-Run | | Between-Day | | Within-Laboratory | |
| | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| PP06A | 80 | 136.11 | 7.86 | 5.8 | 1.98 | 1.5 | 3.84 | 2.8 | 8.97 | 6.6 |
| PP06* | 79 | 135.24 | 6.41 | 4.7 | 0.00 | 0.0 | 3.18 | 2.4 | 7.15 | 5.3 |
| PP07A | 80 | 270.37 | 10.48 | 3.9 | 6.49 | 2.4 | 2.76 | 1.0 | 12.63 | 4.7 |
| PP08A | 80 | 380.79 | 14.88 | 3.9 | 0.00 | 0.0 | 5.29 | 1.4 | 15.79 | 4.1 |
| PP09C | 80 | 821.76 | 22.80 | 2.8 | 0.00 | 0.0 | 11.93 | 1.5 | 25.73 | 3.1 |
| PP10C | 80 | 2200.11 | 67.63 | 3.1 | 52.66 | 2.4 | 31.31 | 1.4 | 91.25 | 4.1 |
A Native human sera from a subject with traumatic brain injury
C Native human sera spiked with recombinant protein
* Analysis without the outliers
Qualitative analysis:
| GFAP Assay (Cut-Off: 22.0 pg/mL) | | | | |
| --- | --- | --- | --- | --- |
| Panel Member | Mean (pg/mL) | Total Number Replicates | Qualitative Agreement | |
| | | | n/N (Positives/Total Replicates) | %Correct Call |
| PP01A | 16.39 | 80 | 0/80 | 100 |
| PP02C | 32.45 | 80 | 80/80 | 100 |
| PP03C | 82.63 | 80 | 80/80 | 100 |
| PP04C | 209.22 | 80 | 80/80 | 100 |
| PP05C | 298.17 | 80 | 80/80 | 100 |
A Negative (GFAP concentration is below the cut-off value)
C Positive (GFAP concentration is equal to or above the cut-off value)
| UCH-L1 Assay (Cut-Off: 327.0 pg/mL) | | | | |
| --- | --- | --- | --- | --- |
| Panel Member | Mean (pg/mL) | Total Number Replicates | Qualitative Agreement | |
| | | | n/N (Positives/Total Replicates) | %Correct Call |
| PP06A | 136.11 | 80 | 0/80 | 100 |
| PP06* | 135.24 | 79 | 0/79 | 100 |
| PP07B | 270.37 | 80 | 0/80 | 100 |
| PP08B | 380.79 | 80 | 80/80 | 100 |
| PP09C | 821.76 | 80 | 80/80 | 100 |
| PP10C | 2200.11 | 80 | 80/80 | 100 |
A Negative (UCH-L1 concentration is below the cut-off value)
B Near cut-off (mean $\pm 25\%$ of the cut-off value)
C Positive (UCH-L1 concentration is equal to or above the cut-off value)
* Analysis without the outliers
b. Lot-to-Lot Precision
A study was conducted per the CLSI guideline EP05-A3 to evaluate the effects of between-lot variability. The same two sample panels evaluated in the within-laboratory
{7}
precision study were tested in five replicates, one run per day, during five days, on three different lots of the VIDAS TBI (GFAP) assay and three different lots of the VIDAS TBI (UCH-L1) assay using one VIDAS 3 instrument. A total of 75 replicates were obtained per panel member. The data were analyzed quantitively and qualitatively. The results are summarized in the tables below.
Quantitative analysis:
| GFAP Assay | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Sample | N | Mean (pg/mL) | Within-Run | | Between-Run/Day | | Between-Lot | | Total | |
| | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| PP01A | 75 | 15.63 | 0.93 | 5.9 | 0.00 | 0.0 | 0.62 | 4.0 | 1.11 | 7.1 |
| PP02A | 75 | 31.03 | 1.16 | 3.7 | 0.12 | 0.4 | 1.56 | 5.0 | 1.95 | 6.3 |
| PP03A | 75 | 75.29 | 4.69 | 6.2 | 0.00 | 0.0 | 4.62 | 6.1 | 6.58 | 8.7 |
| PP04B | 75 | 199.54 | 5.93 | 3.0 | 2.47 | 1.2 | 10.47 | 5.2 | 12.28 | 6.2 |
| PP05B | 75 | 282.23 | 8.23 | 2.9 | 2.36 | 0.8 | 16.08 | 5.7 | 18.22 | 6.5 |
A Native human sera from a subject with traumatic brain injury
B Native human sera from subject with meningeal hemorrhage
| UCH-L1 Assay | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Sample | N | Mean (pg/mL) | Within-Run | | Between-Run/Day | | Between-Lot | | Total | |
| | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| PP06A | 75 | 141.30 | 6.58 | 4.7 | 4.28 | 3.0 | 1.14 | 0.8 | 7.93 | 5.6 |
| PP07A | 75 | 274.52 | 10.72 | 3.9 | 2.61 | 1.0 | 3.42 | 1.2 | 11.55 | 4.2 |
| PP08A | 75 | 388.91 | 14.54 | 3.7 | 0.00 | 0.0 | 7.74 | 2.0 | 16.47 | 4.2 |
| PP09C | 75 | 837.10 | 30.20 | 3.6 | 0.00 | 0.0 | 16.75 | 2.0 | 34.54 | 4.1 |
| PP10C | 75 | 2253.47 | 104.87 | 4.7 | 0.00 | 0.0 | 86.28 | 3.8 | 135.80 | 6.0 |
A Native human sera from a subject with traumatic brain injury
C Native human sera spiked with recombinant protein
Qualitative analysis:
| GFAP Assay (Cut-Off: 22.0 pg/mL) | | | | |
| --- | --- | --- | --- | --- |
| Panel Member | Mean (pg/mL) | Total Number Replicates | Qualitative Agreement | |
| | | | n/N (Positives/Total Replicates) | %Correct Call |
| PP01A | 15.63 | 75 | 0/75 | 100 |
| PP02C | 31.03 | 75 | 75/75 | 100 |
| PP03C | 75.29 | 75 | 75/75 | 100 |
| PP04C | 199.54 | 75 | 75/75 | 100 |
| PP05C | 282.23 | 75 | 75/75 | 100 |
A Negative (GFAP concentration is below the cut-off value)
C Positive (GFAP concentration is equal to or above the cut-off value)
K240279 - Page 8 of 24
{8}
| UCH-L1 Assay (Cut-Off: 327.0 pg/mL) | | | | |
| --- | --- | --- | --- | --- |
| Panel Member | Mean (pg/mL) | Total Number Replicates | Qualitative Agreement | |
| | | | n/N (Positives/Total Replicates) | %Correct Call |
| PP06A | 141.30 | 75 | 0/75 | 100 |
| PP07B | 274.52 | 75 | 0/75 | 100 |
| PP08B | 388.91 | 75 | 75/75 | 100 |
| PP09C | 837.10 | 75 | 75/75 | 100 |
| PP10C | 2253.47 | 75 | 75/75 | 100 |
A Negative (UCH-L1 concentration is below the cut-off value)
B Near cut-off (mean ±25% of the cut-off value)
C Positive (UCH-L1 concentration is equal to or above the cut-off value)
c. Multi-Site Reproducibility
A study was conducted per CLSI guideline EP05-A3 to evaluate the effects of site-to-site variability. The same two sample panels evaluated in the within-laboratory precision study were tested in triplicate, two runs per day with at least two hours between runs, during five days in three different laboratories including two in the U.S. and one OUS. One lot of the VIDAS TBI (GFAP) assay, one lot of the VIDAS TBI (UCH-L1) assay and one VIDAS 3 instrument at each site was used. Combining all sites, a total of 90 replicates were obtained per panel member. The data were analyzed quantitively and qualitatively. The results are summarized for each assay in the table below.
Quantitative analysis:
| GFAP Assay | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Sample | N | Mean (pg/mL) | Within-Run | | Between-Day | | Between-Site | | Total | |
| | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| PP01A | 90 | 15.41 | 0.92 | 6.0 | 0.67 | 4.3 | 1.44 | 9.4 | 1.97 | 12.8 |
| PP01* | 89 | 15.49 | 0.82 | 5.3 | 0.59 | 3.8 | 1.37 | 8.8 | 1.83 | 11.8 |
| PP02A | 90 | 31.38 | 1.22 | 3.9 | 0.57 | 1.8 | 1.29 | 4.1 | 2.07 | 6.6 |
| PP03A | 90 | 76.35 | 2.92 | 3.8 | 0.73 | 1.0 | 4.60 | 6.0 | 6.17 | 8.1 |
| PP04B | 90 | 200.60 | 5.24 | 2.6 | 3.08 | 1.5 | 8.98 | 4.5 | 11.20 | 5.6 |
| PP05B | 90 | 289.00 | 8.24 | 2.9 | 0.76 | 0.3 | 9.49 | 3.3 | 13.55 | 4.7 |
A Native human sera from a subject with traumatic brain injury
B Native human sera from subject with meningeal hemorrhage
* Analysis without the outliers
| UCH-L1 Assay | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Sample | N | Mean (pg/mL) | Within-Run | | Between-Day | | Between-Site | | Total | |
| | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| PP06A | 90 | 141.60 | 5.92 | 4.2 | 1.72 | 1.2 | 2.86 | 2.0 | 7.67 | 5.4 |
| PP06* | 89 | 141.25 | 4.68 | 3.3 | 2.09 | 1.5 | 3.13 | 2.2 | 6.99 | 5.0 |
| PP07A | 90 | 278.43 | 17.10 | 6.1 | 0.00 | 0.0 | 7.73 | 2.8 | 19.07 | 6.8 |
| PP07* | 89 | 279.86 | 9.73 | 3.5 | 0.00 | 0.0 | 7.19 | 2.6 | 13.43 | 4.8 |
K240279 - Page 9 of 24
{9}
K240279 - Page 10 of 24
| UCH-L1 Assay | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Sample | N | Mean (pg/mL) | Within-Run | | Between-Day | | Between-Site | | Total | |
| | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| PP08A | 90 | 391.28 | 10.88 | 2.8 | 0.00 | 0.0 | 6.95 | 1.8 | 14.12 | 3.6 |
| PP09C | 90 | 844.83 | 24.51 | 2.9 | 8.75 | 1.0 | 5.67 | 0.7 | 26.64 | 3.2 |
| PP10C | 90 | 2266.22 | 67.51 | 3.0 | 41.26 | 1.8 | 18.76 | 0.8 | 83.70 | 3.7 |
$^{\mathrm{A}}$ Native human sera from a subject with traumatic brain injury
$^{\mathrm{C}}$ Native human sera spiked with recombinant protein
*Analysis without the outliers
Qualitative precision:
| GFAP Assay (Cut-Off: 22.0 pg/mL) | | | | |
| --- | --- | --- | --- | --- |
| Panel Member | Mean (pg/mL) | Total Number Replicates | Qualitative Agreement | |
| | | | n/N (Positives/Total Replicates) | %Correct Call |
| PP01A | 15.41 | 90 | 0/90 | 100 |
| PP01* | 15.49 | 89 | 0/89 | 100 |
| PP02C | 31.38 | 90 | 90/90 | 100 |
| PP03C | 76.35 | 90 | 90/90 | 100 |
| PP04C | 200.60 | 90 | 90/90 | 100 |
| PP05C | 289.00 | 90 | 90/90 | 100 |
$^{\mathrm{A}}$ Negative (GFAP concentration is below the cut-off value)
$^{\mathrm{C}}$ Positive (GFAP concentration is equal to or above the cut-off value)
*Analysis without the outliers
| UCH-L1 Assay (Cut-Off: 327.0 pg/mL) | | | | |
| --- | --- | --- | --- | --- |
| Panel Member | Mean (pg/mL) | Total Number Replicates | Qualitative Agreement | |
| | | | n/N (Positives/Total Replicates) | %Correct Call |
| PP06A | 141.60 | 90 | 0/90 | 100 |
| PP06* | 141.25 | 89 | 0/89 | 100 |
| PP07B | 278.43 | 90 | 0/90 | 100 |
| PP07* | 279.86 | 89 | 0/89 | 100 |
| PP08B | 391.28 | 90 | 90/90 | 100 |
| PP09C | 844.83 | 90 | 90/90 | 100 |
| PP10C | 2266.22 | 90 | 90/90 | 100 |
$^{\mathrm{A}}$ Negative (UCH-L1 concentration is below the cut-off value)
$^{\mathrm{B}}$ Near cut-off (mean $\pm 25\%$ of the cut-off value)
$^{\mathrm{C}}$ Positive (UCH-L1 concentration is equal to or above the cut-off value)
*Analysis without the outliers
2. Linearity:
The linearity of the VIDAS TBI (GFAP) and VIDAS TBI (UCH-L1) assays on the VIDAS 3 instrument was evaluated in accordance with CLSI guideline EP06-Ed2. Two separate dilution series, one for each assay, and each composed of 11 sample pools, were prepared by mixing a High sample and a Low sample to cover the analytical measuring interval (AMI) of
{10}
each assay. The High sample for each assay was generated by spiking positive serum sample into a negative serum sample (for GFAP) or pools (for UCH-L1). The Low sample for each assay was analyte depleted pooled serum. With the exception of the low concentration sample pool, which was tested in eight replicates, all other sample pools in the dilution series were tested in four replicates using two lots of each assay. The observed values were evaluated against the expected values and weighted least squares linear regression was performed. The $\% \mathrm{CV}$ of the concentrations measured, the predicted values, deviations from linearity, and $\%$ deviations from the regression line were calculated for each sample. The results are summarized in the table below:
| Assay | Lot # | Test Range (pg/mL) | Slope (95% CI) | Intercept (95% CI) | R2 | % Deviation |
| --- | --- | --- | --- | --- | --- | --- |
| GFAP | 1 | 6.7 – 354.5 | 1.024 [0.999; 1.050] | -0.716 [-1.722; 0.289] | 0.9988 | -2.15 to 9.07% |
| UCH-L1 | 1 | 58.9 – 3034.0 | 1.023 [1.006; 1.040] | -1.857 [-5.165; 1.452] | 0.9995 | -6.23 to 1.72% |
| | 2 | 40.2 – 2769.1 | 0.980 [0.958; 1.002] | 6.818 [3.472; 10.164] | 0.9990 | -13.05 to 1.75% |
The study supports the linear range of $6.7 - 354.5\mathrm{pg / mL}$ for the GFAP assay and 58.9-2769.1 $\mathrm{pg / mL}$ for the UCH-L1 assay, with a deviation from linearity within $10\%$ (except the Low sample with the UCH-L1 value of $40.2~\mathrm{pg / mL}$ has a deviation of $-13.05\%$ from linearity). The results support the linearity throughout the reportable ranges of $10.0 - 320.0~\mathrm{pg / mL}$ for GFAP and $80.0 - 2560.0~\mathrm{pg / mL}$ for UCH-LI.
# Hook effect
The potential hook effect was evaluated. A serum sample spiked to a high antigen level was used to prepare a dilution series with 12 samples. Each dilution was tested in triplicate for each assay using one lot of VIDAS TBI (GFAP) and one lot of VIDAS TBI (UCH-L1) assay on one VIDAS 3 instrument. No hook effect was observed for each assay using serum samples with antigen concentrations exceeding $200,000~\mathrm{pg / mL}$ GFAP and $400,000~\mathrm{pg / mL}$ UCH-L1, when tested on the VIDAS 3 instrument.
# 3. Analytical Specificity/Interference:
Studies to evaluate the analytical specificity/interference were conducted using at least one lot of VIDAS TBI (GFAP) assay, one lot of VIDAS TBI (UCH-L1) assay and one VIDAS 3 instrument. Two levels of GFAP and UCH-L1 in pooled serum samples were evaluated in the studies: low $(17.0 - 27.0\mathrm{pg / mL})$ and high $(180.0 - 220.0\mathrm{pg / mL})$ for GFAP, and low $(300.0 - 350.0\mathrm{pg / mL})$ and high $(1300.0 - 1700.0\mathrm{pg / mL})$ for UCH-L1. The low GFAP and UCH-L1 level corresponds to the concentration close to the respective clinical decision threshold. The high UCH-L1 serum sample was spiked with recombinant UCH-L1 to achieve the targeted concentration.
# a. Endogenous and exogenous interference
K240279 - Page 11 of 24
{11}
Test samples were created by spiking the pooled serum samples with the potentially interfering substances listed in the table below at test concentrations recommended in CLSI EP37 Ed1. Control samples were spiked only with the appropriate solvent used to prepare the stock solution of the potentially interfering substance. Each test and control sample were evaluated with five to seven replicates in one assay run for each potentially interfering substance. The %recovery was calculated by comparing measurements of the test and control samples. A substance was identified as an interferent if the difference in the means between the test and control samples was not within 10% of the control sample mean for GFAP or UCH-L1. For any substances identified as an interferent at the initial concentration tested, a dose response analysis was performed.
| Endogenous Substances Not Found to Interfere | | |
| --- | --- | --- |
| | VIDAS TBI (GFAP) | VIDAS TBI (UCH-L1) |
| Potentially interfering substance | Concentration | |
| Bilirubin (conjugated) | 0.4 g/L | 0.4 g/L |
| Bilirubin (unconjugated) | 0.4 g/L | 0.4 g/L |
| Biotin | 3840 ng/mL | 3510 ng/mL |
| Hemoglobin | 10 g/L | 0.6 g/L |
| Human Albumin | 60 g/L | 60 g/L |
| Human anti-mouse antibody (HAMA) | 2000 ng/mL | 2000 ng/mL |
| Rheumatoid factor | 802 IU/mL | 175 IU/mL |
| Triglycerides | 30 g/L | 30 g/L |
| Total proteins | 120 g/L | 120 g/L |
K240279 - Page 12 of 24
{12}
| Exogenous Substances Not Found to Interfere | |
| --- | --- |
| Potentially interfering substance | Concentration |
| Acetaminophen | 15.6 mg/dL |
| Acetylsalicylic acid (Aspirin) | 3.0 mg/dL |
| Benzoylecgonine tetrahydrate | 37.5 ng/mL |
| Cardene (Nicardipine) | 0.047 mg/dL |
| Clopidogrel hydrogensulfate (Plavix) | 4.5 mg/dL |
| Coumadin (Warfarin) | 7.5 mg/dL |
| EDDPA | 125 ng/mL |
| Ethanol | 600.1 mg/dL |
| Ibuprofen | 21.9 mg/dL |
| Lopressor (Metroprolol + tartrate salt) | 18.7 μmol/L |
| Methadone hydrochloride | 0.318 mg/dL |
| d-Methamphetamine | 125 ng/mL |
| Methaqualone | 37.5 ng/mL |
| Morphine monohydrate | 0.78 mg/dL |
| Metoclopramide | 0.225 mg/dL |
| Ondansetron hydrochloride dihydrate | 0.96 mg/dL |
| Oxazepam | 0.435 mg/dL |
| Phencyclidine hydrochloride | 3.1 ng/mL |
| Propoxyphene | 0.321 mg/dL |
| Secobarbital | 1.59 mg/dL |
$^{A}$ 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine
## b. Cross-reactivity
For cross-reactivity, each potential cross-reacting substance was tested at a concentration that corresponds to its highest reported physiological level reported in circulation according to literature, as applicable, indicated in the table below. The effect of each potential cross-reactant was evaluated by comparing the test results from a control sample, spiked with the appropriate solvent used to create the stock solution of the potential cross-reactant (Control Pool), with the test results from a sample spiked with the potentially cross-reacting substance (Test Pool). Each test and control sample for each potentially cross-reacting substance was evaluated with five replicates. The %recovery was calculated by comparing measurements of the test and control samples. A substance was identified as an interferent if the difference in the means between the test and control samples was not within 10% of the control sample mean for GFAP or UCH-L1. None of the tested cross-reacting substances were found to interfere with the VIDAS TBI (GFAP) and VIDAS TBI (GFAP) assays.
K240279 - Page 13 of 24
{13}
| Concentrations Found Not to Interfere | |
| --- | --- |
| Potential Cross-Reactant | No Interference Observed Up To |
| GFAP Assay | |
| Desmin | 127 ng/mL |
| Internexin | 77 ng/mL |
| Keratin type II | 10 ng/mL |
| Neurofilmament light | 68 pg/mL |
| Neurofilmament medium | 8.6 ng/mL |
| Neurofilmament heavy | 77 ng/mL |
| Peripherin | 5 ng/mL |
| Vimentin | 354 ng/mL |
| UCH-L1 Assay | |
| UCH-L3 | 354 ng/mL |
# 4. Assay Reportable Range:
The reportable range for both VIDAS TBI (GFAP) and VIDAS TBI (UCH-L1) is the same as the analytical measuring interval for both assays.
Reportable range: VIDAS TBI (GFAP): 10.0 -320.0 pg/mL
VIDAS TBI (UCH-L1): 80.0 -2560.0 pg/mL
Assay results are preceded by the symbol for greater than $(>)$ or less than $(<)$ if the result is outside of the reportable range.
# 5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
# a. Traceability
There are no internationally recognized reference standards for GFAP and UCH-L1. Calibration of the VIDAS TBI (GFAP, UCH-L1) assay is traceable to in-house reference calibrators for each of the two proteins.
# b. Kit Stability
Shelf-life stability: Reagent shelf-life of the VIDAS TBI (GFAP, UCH-L1) was determined on three VIDAS 3 instruments using three reagent lots of each assay. One kit lot was exposed to a series of thermal shocks to simulate variations in temperature during shipping. A panel of four pooled serum samples covering the respective AMIs of each assay were tested. The mean value for each sample at T0, and sequent timepoints (3, 6, 9, 12 and 13 months) were determined in triplicate in three runs. The absolute and relative recovery of the test samples with respect to the initial measurement at T0 was evaluated. The results support that the VIDAS TBI (GFAP, UCH-L1) is stable up to 8 months when stored at $2 - 8^{\circ}\mathrm{C}$ .
In-use stability: Stability of the VIDAS TBI (GFAP, UCH-L1) after first opening was evaluated for kit components including SPR (storage pouch opened, then tightly closed),
K240279 - Page 14 of 24
{14}
Calibrator S1 and Control C1 (up to four cycles of successive opening and closing of the vials) and the reagent kit (12 cycles of 70 minutes at 30°C ± 2°C). A panel of four pooled serum samples covering the respective AMIs of each assay were tested in triplicate at test timepoints. The results support the following in-use stability claims:
- SPRs were stable up to 8 months after first opening and stored at 2–8°C.
- Reagent kit was stable up to 12 cycles of 70 minutes at 30°C and back to 2–8°C.
- Calibrator S1 and Control C1 were stable up to four cycles from 2–8°C to 25°C.
A study was also performed to determine the stability of each assay from storage at 2–8°C or after subjecting to different temperature conditions before testing. The results support that assay reagents could remain at 18–25°C for up to 4 hours before testing.
**On-board stability**: A study was performed to determine the stability of the reagents, i.e., SPR and strips, inside the VIDAS 3 instrument. A panel of four samples that includes Calibrator S1, Control C1 and two samples, one with analyte concentration close to the cut-off of the respective assay, was tested at the following time intervals using reagent on-board: 0 min, 6 min ± 1, 22 min ± 1, and 32 min ± 2. The results support the stability of the reagents in the VIDAS 3 instrument for up to 8 hours at 37°C and storage of samples in the strip in the VIDAS 3 instrument up to 20 minutes at 37°C before assaying.
**Sample Stability**
A study was performed to evaluate the stability of fresh serum specimens in primary tubes using one lot of the VIDAS TBI (GFAP) assay and one lot of the VIDAS TBI (UCH-L1) assay on one VIDAS 3 instrument. Forty specimens with concentrations of GFAP and UCH-L1 covering the respective assays were collected in serum tubes with separator gel and coagulation activator SST II Advance (SST). All specimens were centrifuged and then tested in duplicate with both assays at different timepoints under storage conditions to mimic customer use. Passing and Bablok regression was used to compare the result of the reference condition to the result of the other conditions. In addition, stability of the sample under freeze/thaw cycles was also evaluated by testing 40 specimens collected in SST tubes using one lot of the VIDAS TBI (GFAP) assay and two lots of the VIDAS TBI (UCH-L1) assay on two VIDAS 3 instruments.
The results support the sample is stable:
- up to three hours at 18–25°C after centrifugation
- up to 43 hours at 2–8°C in closed primary tubes
- for one free/thaw cycle (when stored at -19 to -31°C or -60°C for 5 months)
**Detection Limit**:
The Limit of Blank (LoB), Limit of Detection (LoD), and Limit of Quantitation (LoQ) studies were conducted in accordance with CLSI guideline EP17-A2. The studies evaluated two lots of VIDAS TBI (GFAP) assay and two lots of VIDAS TBI (UCH-L1) assay on one VIDAS 3 instrument.
K240279 - Page 15 of 24
{15}
For LoB, four blank samples corresponding to very low concentrations of native GFAP were tested in duplicate over 8 days on one VIDAS 3 instrument using two reagent lots of the VIDAS TBI (GFAP) assay. The five blank samples corresponding to very low concentrations of native UCH-L1 were tested in duplicate over 8 days on one VIDAS 3 instrument using two reagent lots of the VIDAS TBI (UCH-L1) assay. The LoB was evaluated for each lot of reagent and the final LoB was determined as the greater of the LoB values obtained for each reagent lot for each assay.
For LoD, seven low-level of GFAP serum samples were tested in five replicates per day for 8 days using two lots of the VIDAS TBI (GFAP) assay on one VIDAS 3 instrument to obtain 40 measurements for each sample, totaling 280 measurements per assay lot. The sample protocol was used to test six low-level of UCH-L1 serum samples using two lots of the VIDAS TBI (UCH-L1) assay, resulting a total of 240 measurements per assay lot. The final LoD was determined as the greater of the LoD values obtained for each reagent lot for each assay.
For LoQ, the data generated from the LoD study above was estimated based on the lowest concentration quantifiable with an intermediate precision of ≤15% CV and ≥ LoD for each assay.
The results are summarized in the table below:
| Assay | LoB (pg/mL) | LoD (pg/mL) | LoQ (pg/mL) | Lower Limit of AMI/ Reportable Range (pg/mL) |
| --- | --- | --- | --- | --- |
| GFAP | 4.4 | 5.4 | 5.4 | 10.0 |
| UCH-L1 | 41.8 | 48.1 | 48.1 | 80.0 |
7. Assay Cut-Off:
VIDAS TBI (GFAP): 22.0 pg/mL
VIDAS TBI (UCH-L1): 327.0 pg/mL
B Comparison Studies:
1. Method Comparison with Predicate Device:
Not applicable
2. Matrix Comparison:
Sample matrix equivalence was evaluated for the VIDAS TBI (GFAP, UCH-L1) in accordance with CLSI guideline EP35 1st Ed for serum samples collected with three different blood collection tube types: serum tube with coagulation activator (CAT), serum tube with separator gel and coagulation activator SST II Advance (SST) and Rapid Serum tube with thrombin-based clot activator and separator gel (RST). Blood from donors was drawn on each of the tested blood collection tubes.
K240279 - Page 16 of 24
{16}
K240279 - Page 17 of 24
# CAT vs SST
A set of samples was prepared to cover the AMI of the respective assay. All specimens were centrifuged and then tested in duplicate on one VIDAS 3 instrument using one assay lot for each assay. Passing-Bablok regression analysis was performed to compare the result of the SST to the result of the CAT used as reference. Results are summarized in the table below for each assay.
| Assay | N | SST (pg/mL) | CAT (pg/mL) | Slope (95% CI) | Intercept (95% CI) | R |
| --- | --- | --- | --- | --- | --- | --- |
| GFAP | 31 | 10.4 – 314.2 | 11.6 – 307.7 | 1.02 [0.99; 1.04] | -0.4 [-1.1; 1.2] | 0.997 |
| UCH-L1 | 35 | 95.2 – 2106.0 | 97.3 – 2395.2 | 1.00 [0.94; 1.05] | -5.9 [-18.9; 16.7] | 0.994 |
# SST vs RST
A separate study was performed using a set of samples that cover the entire AMI of the respective assay to validate the use of RST tube. All samples were tested in duplicate. Passing-Bablok regression analysis was performed to compare the result of the RST to the result of the SST used as reference. Results are summarized in the table below for each assay.
| Assay | N | SST (pg/mL) | RST (pg/mL) | Slope (95% CI) | Intercept (95% CI) | R |
| --- | --- | --- | --- | --- | --- | --- |
| GFAP | 41 | 10.0 – 326.4 | 12.1 – 279.9 | 1.02 [0.99; 1.04] | 0.8 [0.0; 1.8] | 0.992 |
| UCH-L1 | 46 | 60.0 – 1764.7 | 55.7 – 1348.5 | 0.98 [0.92; 1,04] | 9.8 [-7.4; 26.0] | 0.988 |
# C Clinical Studies:
1. Clinical Sensitivity and Clinical Specificity:
The performance of the VIDAS TBI (GFAP, UCH-L1) as an aid to assist physicians in determining the need for a computed tomography (CT) scan of the head was evaluated in two studies: the first study used samples from the ALERT-TBI study, the second study used samples from the BRAINI study.
Clinical study using samples from the ALERT-TBI study
The study was conducted by testing frozen serum specimens that were previously collected by Banyan Biomarkers (San Diego, CA) from subjects participating in their prospective multicenter study “A Prospective Clinical Evaluation of Biomarkers of Traumatic Brain Injury” (ALERT-TBI, protocol ATO-06). This study enrolled subjects over the age of 18 who presented to a health care facility (HCF) or emergency department (ED) with suspected head injuries and had blood withdrawn within 12 hours of head injury. A total of 1911 frozen and de-identified serum samples of the ALERT cohort were tested in singlicate at three clinical sites (that include two in the U.S.) on the VIDAS 3 instrument, using three lots and two calibrations per lot by operators representative of end-users of the VIDAS TBI (GFAP, UCH-L1) test. Each specimen was associated with a standard of care computed tomography (CT) scan of the head from the same study subject. Under the ALERT-TBI study protocol, the CT scans were independently reviewed by a panel of radiologists and a subject’s CT scan was
{17}
classified as positive if intracranial lesions were present. The standard of care head CT scans were previously classified as positive or negative for intracranial lesions as part of the ALERT-TBI study. Additional details regarding the inclusion and exclusion criteria, evaluation of CT scans and definition of intracranial lesion can be found in DEN170045¹ and published study². The demographic characteristics of the enrolled subjects evaluated with the VIDAS TBI (GFAP, UCH-L1) test are presented in the table below.
The demographic characteristics and CT scan results for all 1911 evaluable subjects (120 CT-positive patients and 1791 CT-negative patients) are presented in the table below.
| Demographic Characteristics (ALERT TBI Cohort) | Head CT-Scan Results | | Total |
| --- | --- | --- | --- |
| | Positive | Negative | |
| N | 120 | 1791 | 1911 |
| Age^{A} (Years) | | | |
| Mean | 58.8 | 48.5 | 49.1 |
| Median | 58.5 | 48.0 | 49.0 |
| Standard Deviation | 18.29 | 21.02 | 21.00 |
| Range | 20 – 95 | 18 – 98 | 18 – 98 |
| Gender, N (%) | | | |
| Male | 70 (58.3%) | 1010 (56.4%) | 1080 (56.5%) |
| Female | 50 (41.7%) | 781 (43.6%) | 831 (43.5%) |
| Race^{B}, N (%) | | | |
| White | 98 (81.7%) | 1250 (69.8%) | 1348 (70.5%) |
| Black or African American | 16 (13.3%) | 487 (27.2%) | 503 (26.3%) |
| Asian | 5 (4.2%) | 24 (1.3%) | 29 (1.5%) |
| Native Hawaiian/Pacific Islander | 1 (0.8%) | 2 (0.1%) | 3 (0.2%) |
| American Indian or Alaska Native | 1 (0.8%) | 10 (0.6%) | 11 (0.6%) |
| Unknown | 1 (0.8%) | 27 (1.5%) | 28 (1.5%) |
| Ethnicity, N (%) | | | |
| Hispanic or Latino | 1 (0.8%) | 89 (5.0%) | 90 (4.7%) |
| Not Hispanic or Latino | 118 (98.3%) | 1701 (95.0%) | 1819 (95.2%) |
| Not Reported | 1 (0.8%) | 1 (0.1%) | 2 (0.1%) |
| ¹ Age was calculated relative to the date of informed consent
² Subjects could have indicated more than one race | | | |
Other baseline characteristics collected for all evaluable subjects were similar in the CT scan-positive and CT scan-negative groups, except for alcohol use, which was slightly lower but not statistically significant in the CT scan-positive group (refer to DEN170045¹).
The following head injury characteristics were collected for all evaluable subjects and summarized in the table below. The mean time from head injury to blood draw was 3.5 hours. Most subjects had a GCS score of 15 (1798/1911 or 94.1% in the total cohort, 94/120 or 78.3% in CT scan-positive subjects and 1738/1827 or 95.1% in CT scan-negative
K240279 - Page 18 of 24
¹ FDA. De Novo request for evaluation of automatic class III designation of the Banyan Brain Trauma Indicators (BTI) 2018 [Available from: https://www.accessdata.fda.gov/cdrh_docs/reviews/DEN170045.pdf
² Bazarian JJ, Biberthaler P, Welch RD, Lewis LM, Barzo P, Bogner-Flatz V, et al. Serum GFAP and UCH-L1 for prediction of absence of intracranial injuries on head CT (ALERTTBI): a multicentre observational study. The Lancet Neurology. 2018;17(9):782-789.
{18}
subjects). The percentage of subjects with GCS scores of 13 and 14 were higher in the CT scan-positive subjects (7/120 or 5.8% for GCS of 13 and 19/120 or 15.8% for GCS of 14) compared to the CT scan-negative subjects (15/1791 or 0.8% for GCS of 13 and 72/1791 or 4.0% for GCS of 14).
| Head Injury Characteristics (ALERT TBI Cohort) | Head CT-Scan Results | | Total |
| --- | --- | --- | --- |
| | Positive | Negative | |
| N | 120 | 1791 | 1911 |
| Glasgow Coma Score – N (%) | | | |
| 13 | 7 (5.8%) | 15 (0.8%) | 22 (1.2%) |
| 14 | 19 (15.8%) | 72 (4.0%) | 91 (4.8%) |
| 15 | 94 (78.3%) | 1704 (95.1%) | 1798 (94.1%) |
| Neurological Assessment - Number (%) of Subjects Experiencing | | | |
| Loss of Consciousness | 82 (68.3%) | 727 (40.6%) | 809 (42.3%) |
| Confusion | 44 (36.7%) | 314 (17.5%) | 358 (18.7%) |
| Vomiting | 14 (11.7%) | 131 (7.3%) | 145 (7.6%) |
| Vomiting two or more episodes | 10 (8.3%) | 62 (3.5%) | 72 (3.8%) |
| Post Traumatic Amnesia (PTA) | 81 (67.5%) | 552 (30.8%) | 633 (33.1%) |
| Retrograde PTA ≥ 30 min | 22 (18.3%) | 68 (3.8%) | 90 (4.7%) |
| Persistent Anterograde PTA | 42 (35.0%) | 172 (9.6%) | 214 (11.2%) |
| Seizures | 2 (1.7%) | 11 (0.6%) | 13 (0.7%) |
| Subjects with Drug or Alcohol Intoxication at Time of Presentation to Facility | 33 (27.5%) | 370 (20.7%) | 403 (21.1%) |
| Dangerous Mechanism of InjuryB | 27 (22.5%) | 370 (20.7%) | 397 (20.8%) |
| Physical EvidenceC | | | |
| Visible Trauma Above the Clavicle | 101 (84.2%) | 1109 (61.9%) | 1210 (63.3%) |
| Suspected Open or Depressed Skull Fracture | 14 (11.7%) | 46 (2.6%) | 60 (3.1%) |
| Signs of Basal Skull Fracture | 10 (8.3%) | 26 (1.5%) | 36 (1.9%) |
| Presence of Neurosurgical Lesion | 5 (4.2%) | 0 (0.0%) | 5 (0.3%) |
| Time from Head Injury to Examination, (hours)A | | | |
| N | 1789 | 120 | 1909D |
| Mean (SD) | 1.8 (1.73) | 1.7 (1.91) | 1.7 (1.90) |
| Median | 1.2 | 1.1 | 1.1 |
| Range (minimum, maximum) | (0.3, 7.8) | (0.1, 33.4) | (0.1, 33.4) |
| Time from Head Injury to CT-Scan, (hours)A | | | |
| N | 1789 | 120 | 1909D |
| Mean (SD) | 2.8 (2.08) | 2.8 (2.10) | 2.8 (2.10) |
| Median | 2.1 | 2.2 | 2.2 |
| Range (minimum, maximum) | (0.5, 10.9) | (0.2, 33.5) | (0.2, 33.5) |
| Time from Head Injury to Blood Draw, (hours)A | | | |
| N | 1788 | 120 | 1908E |
| Mean (SD) | 3.8 (1.92) | 3.5 (2.06) | 3.5 (2.06) |
| Median | 3.3 | 3.2 | 3.2 |
K240279 - Page 19 of 24
{19}
| Range (minimum, maximum) | (0.3, 9.3) | (0.3, 35.3) | (0.3, 35.3)F |
| --- | --- | --- | --- |
A Time since head injury calculated relative to time that subject was first examined by medical personnel at facility.
$^{\mathrm{B}}$ Dangerous mechanism of injury was pedestrian struck by a motor vehicle, an occupant ejected from a motor vehicle, or a fall from an elevation of 3 or more feet or 5 stairs.
Prior to head CT.
Two subjects did not have time from head injury to examination and CT scan recorded.
$^{\mathrm{E}}$ Three subjects did not have time from head injury to blood draw recorded.
F Two subjects had blood draws taken more than 12h (18h and 35.3h) from head injury. Please note that the "efficacy" ALERT population was used for this study to be aligned with Banyan testing strategy (worst case scenario). This population included all subjects for whom head CT scan results were obtained.
To estimate clinical performance characteristics, the VIDAS TBI (GFAP, UCH-L1) test result was compared to the consensus head CT scan result for each patient. The performance estimates are summarized in the table below.
| ALERT TBI Sample Cohort | Head CT-Scan Result | | Total | |
| --- | --- | --- | --- | --- |
| | | Positive | | Negative |
| VIDAS TBI (GFAP, UCH-L1) | Positive | 116 | 1053 | 1169 |
| | Negative | 4 | 738 | 742 |
| Total | | 120 | 1791 | 1911 |
| Clinical Sensitivity: 96.7% (116/120) (95% CI: 91.7-99.1%)Clinical Specificity: 41.2% (738/1791) (95% CI: 38.9-43.5%)Negative Predictive Value (NPV)A: 99.5% (738/742) (95% CI: 98.6-99.9%)Positive Predictive Value (PPV)B: 9.9% (116/1169) (95% CI: 8.3-11.8%)Likelihood Ratio Negative (LRN): 0.10 (95% CI: 0.04-0.23)Likelihood Ratio Positive (LRP): 1.6 (95% CI: 1.51-1.69)CT scan positive prevalence rate in study: 6.3% (120/1911) | | | | |
| A Adjusted NPV for 6% CT scan positive prevalence rate (DEN170045): 99.3% (95% CI: 98.5-99.7%)B Adjusted PPV for 6% CT scan positive prevalence rate (DEN170045): 9.3% (95% CI: 8.8-9.7%) | | | | |
Of the 120 subjects with positive CT scan results, four had negative results from the VIDAS TBI (GFAP UCH-L1) test. The rate of false negative (FN) results was $3.33\%$ (4/120). None of these four subjects were identified with a lesion requiring surgical intervention. Of the 1791 subjects with negative CT scan results, 738 had a negative VIDAS TBI (GFAP UCH-L1) result (clinical specificity $= 41.2\%$ ). The rate of False Positive (FP) results was $58.8\%$ (1053/1791). The Negative Predictive Value (NPV) of the assay was $99.5\%$ (738/742). The results showed that the clinical performance of the VIDAS TBI (GFAP UCH-L1) test is characterized by high clinical sensitivity and high NPV, comparable to that demonstrated by the Banyan BTI (DEN170045) with clinical sensitivity of $97.5\%$ and NPV of $99.6\%$ , supporting clinical validity as an aid in the evaluation of the need for a CT scan in subjects presenting with a GCS score of 13 to 15.
Analyses of assay performance by gender, age, and time from injury relative to blood draw are shown in the table below.
K240279 - Page 20 of 24
{20}
| | Sensitivity N (%) (95% CI) | Specificity N (%) (95% CI) | NPV N (%) (95% CI) | PPV N (%) (95%CI) |
| --- | --- | --- | --- | --- |
| All Subjects N=1911 | 116/120 (96.7%) (91.7-99.1%) | 738/1791 (41.2%) (38.9-43.5%) | 738/742 (99.5%) (98.6-99.9%) | 116/1169 (9.9%) (8.3-11.8%) |
| Gender | | | | |
| Male N=1080 (56.5%) | 67/70 (95.7%) (88.0-99.1%) | 413/1010 (40.9%) (37.9-44.0%) | 413/416 (99.3%) (97.9-99.9%) | 67/664 (10.1%) (8.0-12.6%) |
| Female N=831 (43.5%) | 49/50 (98.0%) (89.4-99.9%) | 325/781 (41.6%) (38.2-45.1%) | 325/326 (99.7%) (98.3-100.0%) | 49/505 (9.7%) (7.4-12.6%) |
| Age | | | | |
| <65 years N=1410 (73.8%) | 74/77 (96.1%) (89.0-99.2%) | 674/1333 (50.6%) (48.0-53.3%) | 675/678 (99.6%) (98.7-99.9%) | 74/732 (10.1%) (8.1-12.5%) |
| ≥65 years N=501 (26.2%) | 42/43 (97.7%) (87.7-99.9%) | 63/458 (13.8%) (10.9-17.2%) | 63/64 (98.4%) (91.6-100.0%) | 42/437 (9.6%) (7.2-12.7%) |
| Time from Injury to Blood Draw | | | | |
| 0-4 hours N=1443 (75.5%) | 83/86 (96.5%) (90.1-99.3%) | 545/1357 (40.2%) (37.6-42.8%) | 545/548 (99.5%) (98.4-99.9%) | 83/895 (9.3%) (7.5-11.4%) |
| 4-8 hours N=386 (20.2%) | 28/28 (100.0%) (87.7-100.0%) | 164/358 (45.8%) (40.7-51.0%) | 164/164 (100.0%) (97.8-100.0%) | 28/222 (12.6%) (8.9-17.6%) |
| 0-8 hours N=1829 (95.7%) | 111/114 (97.4%) (92.5-99.5%) | 709/1715 (41.3%) (39.0-43.7%) | 709/712 (99.6%) (98.8-99.9%) | 111/1117 (9.9%) (8.3-11.8%) |
| 8-12 hours N=80 (4.2%) | 5/6 (83.3%) (43.6-97.0%) | 29/74 (39.2%) (28.9-50.6%) | 29/30 (96.7%) (82.8-99.9%) | 5/50 (10.0%) (4.3-21.4%) |
There was little variation in NPV and PPV between males and females and with increasing time from injury. However, the differences did not translate into statistically significant differences in assay performance. The analysis by age shows that the clinical specificity is lower in patients $\geq 65$ years of age (13.8%) than in patients $< 65$ years of age (50.6%), and conversely, the clinical sensitivity is higher in the older population (97.7% versus 96.1%), respectively.
# Clinical study using samples from the BRAINI study
A supplemental clinical validation of the GFAP and UCH-L1 cut-offs was performed with serum samples from subjects participating in the multi-site study within 4 years after collection and stored frozen at $-60^{\circ}\mathrm{C}$ . The BRAINI study was designed following the main inclusion and non-inclusion criteria defined for the ALERT-TBI cohort. Only the serum
K240279 - Page 21 of 24
{21}
specimens collected and stored less than 3.5 years (42 months) in accordance with the proposed intended use of the VIDAS TBI (GFAP, UCH-L1) test were included in the analysis (n=562). The demographic and head injury characteristics of the evaluable subjects are presented in the tables below.
| Demographic CharacteristicsA
(BRAINI Sub-Cohort) | Head CT-Scan Results | | Total |
| --- | --- | --- | --- |
| | Positive | Negative | |
| N | 74 | 488 | 562 |
| Age (Years) | | | |
| Mean | 54.5 | 48.6 | 49.4 |
| Median | 55.5 | 47.0 | 49.0 |
| Standard Deviation | 20.14 | 21.51 | 21.41 |
| Range | 19 – 95 | 18 – 100 | 18 – 100 |
| Gender, N (%) | | | |
| Male | 52 (70.3%) | 309 (63.3%) | 361 (64.2%) |
| Female | 22 (29.7%) | 179 (36.7%) | 201 (35.8%) |
| AInformation relating to race and ethnic origin was not collected as part of the BRAINI study | | | |
| Head Injury Characteristics
(BRAINI Sub-Cohort) | Head CT-Scan Results | | Total |
| --- | --- | --- | --- |
| | Positive | Negative | |
| N | 74 | 488 | 562 |
| Time from Head Injury to Examination (hours) | | | |
| Mean (SD) | 1.5 (0.95) | 1.6 (1.41) | 1.6 (1.41) |
| Median | 1.2 | 1.2 | 1.2 |
| Range (min, max) | (0.3, 4.4) | (0.0, 10.0) | (0.0, 10.0) |
| Time from Head Injury to CT-Scan (hours) | | | |
| Mean (SD) | 4.2 (2.10) | 4.7 (2.61) | 4.7 (2.55) |
| Median | 3.7 | 4.2 | 4.2 |
| Range (min, max) | (1.0, 12.3) | (0.1, 22.8) | (0.1, 22.8) |
| Time from Head Injury to Blood Draw (hours) | | | |
| Mean (SD) | 5.2 (2.79) | 5.0 (2.73) | 5.0 (2.73) |
| Median | 4.5 | 4.3 | 4.3 |
| Range (min, max) | (1.4, 11.4) | (0.5, 11.9) | (0.5, 11.9) |
| Glasgow Coma Score | | | |
| 13 | 0 (0.0%) | 1 (0.2%) | 1 (0.2%) |
| 14 | 13 17.6%) | 23 (4.7%) | 36 (6.4%) |
| 15 | 61(82.4%) | 464 (95.1%) | 525 (93.4%) |
| Neurological Assessment: Number (%) of Subjects Experiencing | | | |
| Loss of ConsciousnessA | 46 (62.2%) | 189 (38.7%) | 235 (41.8%) |
| Vomiting | 5 (6.8%) | 19 (3.9%) | 24 (4.3%) |
| Vomiting Two or More Episodes | 9 (12.2%) | 27 (5.5%) | 36 (6.4%) |
| Post traumatic Amnesia (PTA) | 43 (58.1%) | 150 (30.7%) | 193 (34.3%) |
| Seizures | 4 (5.4%) | 21 (4.3%) | 25 (4.4%) |
| Subjects with Drug or Alcohol Intoxication at Time of Presentation to FacilityB | 9 (12.2%) | 62 (12.7%) | 71 (12.6%) |
K240279 - Page 22 of 24
{22}
| Dangerous Mechanism of InjuryC | 10 (13.5%) | 18 (3.7%) | 28 (5.0%) |
| --- | --- | --- | --- |
| Physical Evidence | | | |
| Suspected Open or Depressed Skull Fracture | 11 (14.9%) | 25 (5.1%) | 36 (6.4%) |
| Signs of Basal Skull Fracture | 3 (4.1%) | 3 (0.6%) | 6 (1.1%) |
A Loss of consciousness was assessed at patient admission.
B Subjects with other type of intoxication were also included, such as cannabis intoxication, cocaine intoxication etc.
C Dangerous mechanism of injury was pedestrian hit by a car.
The clinical performance of the VIDAS TBI (GFAP, UCH-L1) was evaluated by testing 562 serum specimens of this BRAINI sub-cohort at three clinical sites including two in the U.S., with operators of the test blinded to the CT results associated with the study subjects. The intracranial lesions were defined as any trauma-induced or related finding visualized upon head CT scan, and may have included acute epidural hematomas, acute subdural hematomas, indeterminate extra-axial lesions, cortical contusions, parenchymal hematomas, nonhemorrhagic contusions, ventricle compression, brain herniation, intraventricular hemorrhage, subarachnoid hemorrhage, petechial hemorrhage and global or focal brain edema.
| BRAINI Sub-Cohort | Head CT-Scan Result | | Total | |
| --- | --- | --- | --- | --- |
| | | Positive | | Negative |
| VIDAS TBI (GFAP, UCH-L1) | Positive | 72 | 321 | 393 |
| | Negative | 2 | 167 | 169 |
| Total | | 74 | 488 | 562 |
| Clinical Sensitivity: 97.3% (72/74) (95% CI: 90.6-99.7%)Clinical Specificity: 34.2% (167/488) (95% CI: 30.1-38.5%)Negative Predictive Value (NPV)A: 98.8% (167/169) (95% CI: 95.8-99.9%)Positive Predictive Value (PPV)B: 18.3% (72/393) (95% CI: 14.8-22.4%)Likelihood Ratio Negative (LRN): 0.10 (95% CI: 0.04-0.23)Likelihood Ratio Positive (LRP): 1.5 (95% CI: 1.51-1.69)CT scan positive prevalence rate in study: 13.2% (74/562) | | | | |
| AAdjusted NPV for 6% CT scan positive prevalence rate (DEN170045): 99.3% (95% CI: 98.5-99.7%)BAdjusted PPV for 6% CT scan positive prevalence rate (DEN170045): 9.3% (95% CI: 8.8-9.7%) | | | | |
The results show similar clinical performance of the VIDAS TBI (GFAP, UCH-L1) test to the predicate device, when evaluated with serum samples collected within 3.5 years in a subgroup of the BRANI study cohort.
2. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Not applicable
D Clinical Cut-Off:
Refer to assay cut-off
E Expected Values/Reference Range:
K240279 - Page 23 of 24
{23}
The expected values from 513 apparently healthy donors (53.8% male and 46.2% female) ranging in age from 18 to 85 were determined in accordance with CLSI guideline C28-A3c. The mean (SD) age was 47.3 (15.9) years. The races/ethnicity in the population is 25.2% Caucasian, 24.8% Hispanic, 25% Asian, and 25% African American. Data analysis was performed separately for the GFAP and UCH-L1 biomarkers. The 2.5th and 97.5th percentiles for both biomarkers were calculated non-parametrically. The results are shown below.
| Biomarker | N | Median (pg/mL) | 95% Reference Interval (2.5th – 97.5th percentile) |
| --- | --- | --- | --- |
| GFAP | 513 | 10.4 pg/mL | <10.00 – 31.63 pg/mL |
| UCH-L1 | 513 | 94.9 pg/mL | < 80.00 – 249.70 pg/mL |
Out of 513 samples, 470 (91.6%) samples were tested negative by the VIDAS TBI (GFAP UCH-L1) test, and 43 (8.4%) samples were tested positive by the VIDAS TBI (GFAP UCH-L1). Among 43 tested positive by the VIDAS TBI (GFAP, UCH-L1), one was positive for both GFAP and UCH-L1, 39 tested positive for GFAP only, and three tested positive for UCH-L1 only. Distribution of GFAP and UCH-L1 by age and gender in this study is summarized in the table below:
| Distribution of GFAP and UCH-L1 by Gender within Age | | | |
| --- | --- | --- | --- |
| Age (years) | Gender | GFAP (2.5th – 97.5th percentile) | UCH-L1 (2.5th – 97.5th percentile) |
| ≤ 30 (n=102) | Female (n=50) | <10 to 17.4 pg/mL | <80.0 – 156.8 pg/mL |
| | Male (n=52) | <10 to 18.4 pg/mL | <80.0 – 178.0 pg/mL |
| 31 – 45 (n=129) | Female (n=55) | <10 to 16.8 pg/mL | <80.0 – 207.2 pg/mL |
| | Male (n=74) | <10 to 15.8 pg/mL | <80.0 – 216.6 pg/mL |
| 46 – 60 (n=162) | Female (n=77) | <10 to 32.2 pg/mL | <80.0 – 206.2 pg/mL |
| | Male (n=85) | <10 to 30.4 pg/mL | <80.0 – 374.9 pg/mL |
| ≥61 (n=120) | Female (n=55) | <10 to 68.6 pg/mL | <80.0 – 282.0 pg/mL |
| | Male (n=65) | <10 to 36.8 pg/mL | <80.0 – 458.8 pg/mL |
It is the responsibility of each laboratory to establish its own reference ranges for the population of patients it serves, as expected values may be affected by different factors including age and gender.
## VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
## IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K240279 - Page 24 of 24
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.