K240258 · Makrite Industries, Inc. · FXX · Apr 18, 2024 · General, Plastic Surgery
Device Facts
Record ID
K240258
Device Name
Disposable Medical Face Mask (M643BE)
Applicant
Makrite Industries, Inc.
Product Code
FXX · General, Plastic Surgery
Decision Date
Apr 18, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4040
Device Class
Class 2
Indications for Use
The Disposable Medical Face Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. The Disposable Medical Face Mask is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This a single use, disposable device(s), provided non-sterile.
Device Story
Disposable, non-sterile, flat-folded medical face mask; three-layer construction (spun-bond polypropylene inner/outer layers, melt-blown polypropylene filter layer); features iron-core/polypropylene nose piece and nylon/spandex ear loops. Intended for single-use by healthcare personnel and patients to provide barrier protection against microorganisms, body fluids, and particulates. Used in clinical infection control settings. Device is worn over the nose and mouth, secured by ear loops and adjusted via the nose piece. Benefits include reduced potential exposure to blood and body fluids during medical procedures.
Clinical Evidence
No clinical data. Bench testing only. Performance verified against ASTM F2100-23, ASTM F2101 (BFE ≥95%), EN 14683 (Delta P <5.0 mmH2O/cm²), ASTM F1862 (synthetic blood penetration at 80 mmHg), 16 CFR 1610 (Flammability Class 1), and ASTM F3502 (PFE ≥80%). Biocompatibility confirmed per ISO 10993-5 and ISO 10993-10.
Technological Characteristics
Three-layer mask: spun-bond polypropylene (outer/inner), melt-blown polypropylene (filter). Nose piece: iron core coated with polypropylene. Ear loops: nylon and spandex. Dimensions: 175x95mm. Non-sterile, single-use. Complies with ASTM F2100-23. Biocompatible per ISO 10993.
Indications for Use
Indicated for use by patients and healthcare personnel to protect against transfer of microorganisms, body fluids, and particulate material, and to reduce exposure to blood and body fluids in infection control practices.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 18, 2024
Makrite Industries, Inc. % Ivy Wang Technical Manager Shanghai Sungo Management Consulting Company Limited 14th Floor, 1500# Century Avenue Shanghai. 200122 China
Re: K240258
Trade/Device Name: Disposable Medical Face Mask (M643BE) Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical apparel Regulatory Class: Class II Product Code: FXX Dated: January 31, 2024 Received: January 31, 2024
Dear Ivy Wang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices
{2}------------------------------------------------
OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
K240258
Device Name
Disposable Medical Face Mask (M643BE)
Indications for Use (Describe)
The Disposable Medical Face Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. The Disposable Medical Face Mask is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This a single use, disposable device(s), provided non-sterile.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
> Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(K) Summary
# K240258
(As required by 21 CFR 807.92)
Date prepared: April 9, 2024
#### A. Applicant:
Name: MAKRITE INDUSTRIES INC. Address: 11F-5., NO.79, Sec.1, Xintai 5th Rd., XiZhi Dist., New Taipei City 221, Taiwan Contact Person: Bob Wen Tel: +886-2-26982419 Fax: +886-2-26982423 Email: bobwen@makrite.com.tw
Submission Correspondent: Primary contact: Ms. Ivy Wang Shanghai SUNGO Management Consulting Co., Ltd. Room 1401, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-58817802 Email: haiyu.wang@sungoglobal.com Secondary contact: Mr. Raymond Luo Room 1401, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-68828050 Email: zxfda(@sungoglobal.com
#### B. Device:
Trade Name: Disposable Medical Face Mask Common Name: Disposable Medical Face Mask Model: M643BE
Regulatory Information Classification Name: Surgical Mask Classification: Class II Product code: FXX Regulation Number: 21 CFR 878.4040 Review Panel: Surgical Apparel
## C. Identification of Primary Predicate device: K22809
Anqing Mayfield Medical Limited. Trade/Device Name: Medical Disposable Face Masks
{5}------------------------------------------------
Regulatory Information Classification Name: Surgical Mask Classification: Class II Product code: FXX Regulation Regulation Number: 21 CFR 878.4040 Review Panel: Surgical Apparel
#### D. Indications for use:
The Disposable Medical Face Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. The Disposable Medical Face Mask is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This a single use, disposable device(s), provided non-sterile.
#### E. Device Description:
The Disposable Medical Face Mask is blue color, single use, flat-folded masks with nose piece and ear loops. The blue colorant is polypropylene (PP) master batch.
The body of the mask is composed of three layers: the inner (3rd layer) and outer (1st layers are made of spun-bond polypropylene, the middle (2nd layer) is made of melt blown polypropylene. The nose piece is made of Iron core coated with polypropylene, ear loop is made of Nylon and Spandex.
Each mask contains ear loops to secure the mask over the user's face and mouth with a bendable nose piece to firmly fit over the nose. This device is not made with natural rubber latex.
The Disposable Medical Face Masks are sold and are intended to be single use, disposable devices.
The mask is designed and manufactured in accordance with ASTM F2100-23 Standard Specification for Performance of Materials Used in Medical Face Masks.
#### F. Comparison of technological characteristics with the predicate device
Provided below table 1 is a comparison of the proposed device with the predicate device.
| Table 1 Comparison of Proposed and Predicate Devices | | | |
|---------------------------------------------------------|------------------------------|-------------------------------|---------|
| Comparison between proposed device and predicate device | | | |
| Comparison Items | proposed device | Predicate Device | Result |
| Device Name | Disposable Medical Face Mask | Medical Disposable Face Masks | Similar |
| 510k Number | K240258 | K222809 | --- |
| Product Code | FXX | FXX | Same |
| Regulation Number | 21 CFR 878.4040 | 21 CFR 878.4040 | Same |
| Regulatory Class | Class II | Class II | Same |
| | | able 1 Comparison of Proposed and Predicate Devices |
|--|--|-----------------------------------------------------|
|--|--|-----------------------------------------------------|
{6}------------------------------------------------
| | Indications for Use | The Disposable Medical Face<br>Mask is intended to be worn to<br>protect both the patient and<br>healthcare personnel from<br>transfer of microorganisms,<br>body fluids and particulate<br>material. The Disposable<br>Medical Face Mask is intended<br>for use in infection control<br>practices to reduce the<br>potential exposure to blood<br>and body fluids. This a single<br>use, disposable device(s),<br>provided non-sterile. | The Medical Disposable Face<br>Mask is intended to be worn to<br>protect both the patient and<br>healthcare personnel from<br>transfer of microorganisms,<br>body fluids and particulate<br>material. The face mask is<br>intended for use in infection<br>control practices to reduce the<br>potential exposure to blood and<br>body fluids. This is a single<br>use, disposable device,<br>provided nonsterile. | Similar | |
|--------------------------------------|--------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|--|
| Size<br>specification<br>(Dimension) | | 175×95mm | 17.5×9.5cm<br>14.5×9.5cm | Similar | |
| Color | | Blue | Blue | Same | |
| Design Feature | | Ear Loops, Flat-pleated, 3 layers | Ear Loops, Flat-pleated, 3 layers | Same | |
| Use | | Single Use, Disposable | Single Use, Disposable | Same | |
| Materials | Outer layer | Spun-bond polypropylene | Spunbond Polypropylene | Same | |
| | Inner layer | Spun-bond polypropylene | Spunbond Polypropylene | Same | |
| | Middle layer<br>(Filter layer) | Melt blown polypropylene filter | Melt-blown Polypropylene | Same | |
| | Nose piece | Iron core coated with polypropylene | Aluminum with Polypropylene<br>(PP) covering | Different.<br>The proposed device<br>has different material<br>of nose piece and ear<br>loop to the predicate<br>device. | |
| | Ear loop | Nylon, Spandex | Polyester, Spandex | | |
| OTC use | | Yes | Yes | Same | |
| Sterility | | Non-Sterile | Non-Sterile | Same | |
| | ASTM F2100 Level | ASTM F2100-23 Level 1 | ASTM F2100-19 Level 1 | Similar | |
| Fluid<br>Performance | Resistance | 32 out of 32 per lot pass at 80 mmHg 3 non-consecutive lots tested | 32 out of 32 per lot pass at 80 mmHg 3 non-consecutive lots tested | Same | |
| | ASTM F1862 | | tested | tested | |
{7}------------------------------------------------
| Particulate Filtration<br>Efficiency | ≥ 80%<br>ASTM F3502-2022a 8.1 | ≥ 95%<br>ASTM F2299 | Similar. Both masks<br>meet the particulate<br>filtration efficiency<br>requirements for their<br>respective versions of<br>ASTM F2100 |
|----------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|
| Bacterial Filtration<br>Efficiency<br>ASTM F2101 | ≥ 95% | ≥ 95% | Same |
| Differential<br>Pressure<br>(Delta P)<br>EN14683 Annex C | < 5.0mmH₂O/cm² | < 5.0mmH₂O/cm² | Same |
| Flammability<br>CFR 1610 | Class 1 | Class 1 | Same |
| Biocompatibility | ISO 10993<br>Under the conditions of the<br>studies, non-cytotoxic, not an<br>irritant, and not a sensitizer | ISO 10993<br>Under the conditions of the<br>studies, non-cytotoxic, not an<br>irritant, and not a sensitizer | Same. |
#### Difference Analysis:
The proposed device has same intended use and similar structure, parameters, and performance as the predicate device. The materials of inner layer and outer layer of the proposed device are also similar to those of the predicate device.
The proposed device has different materials for the nose piece and ear loops than the predicate device. However, the proposed device meets the requirements of ISO 10993. The results do not show any adverse effect. Thus, this difference will not affect the safety and effectiveness between the proposed device and the predicate device.
The Proposed device has different Sub-micron particulate filtration efficiency because the testing of proposed device has different standards and testing methods than the predicate device. The proposed device has been tested and the test results shown that the proposed device meets of Particulate Filtration Efficiency in ASTM F2100-23.
#### G. Summary of Non-Clinical Testing
- > Performance testing
Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards and the requirements stated in the Guidance for Industry and FDA Staff: Surgical Mask - Premarket Notification [510(K)] Submission issued on March 5, 2004:
· ASTM F2100, Standard Specification for Performance of Materials Used in Medical Face Masks.
· ASTM F2101, Standard Test Method for Evaluating the Bacterial Filtration Efficiency (BFE) Of Medical Face Mask Materials, Using A Biological Aerosol of Staphylococcus Aureus;
· EN 14683, Medical Face Masks- Requirements and Test Methods Annex C;
{8}------------------------------------------------
• ASTM F1862/F 1862M, Standard Test Method for Resistance of Medical Face Masks to Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume at A Known Velocity)
- · 16 CFR 1610, Standard for the Flammability of clothing textiles;
· ASTM F3502, Standard Specification for Barrier Face Coverings.
Performance testing are summarized in below table 2.
| Test Methodology | Purpose | Acceptance Criteria<br>for Level 3 Barrier | Result |
|-------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|--------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Bacterial Filtration<br>Efficiency<br>ASTM F2101 | Measure bacterial<br>filtration efficiency | ≥95% | Passed<br>3 non-consecutive lots tested, using a<br>sample size of 32/lot<br>Lot 0923: ≥95%<br>Lot 1023: ≥95%<br>Lot 1123: ≥95% |
| Differential Pressure<br>(mmH2O/cm²)<br>EN 14683:2019<br>Annex C | Determine<br>breathability of the<br>mask | <5.0 mmH2O/cm² | Passed<br>3 non-consecutive lots tested, using a<br>sample size of 32/lot<br>Lot 0923: <5.0<br>Lot 1023: <5.0<br>Lot 1123: <5.0 |
| Sub-micron<br>Particulate Filtration<br>Efficiency<br>ASTM F3502-2022a<br>8.1 | Measure initial particle<br>filtration efficiency | ≥80% | Passed<br>3 non-consecutive lots tested, using a<br>sample size of 32/lot<br>Lot 0923: ≥80%<br>Lot 1023: ≥80%<br>Lot 1123: ≥80% |
| Resistance to<br>Penetration by<br>Synthetic Blood<br>ASTM F1862-17 | Evaluate the resistance<br>to penetration by<br>impact of small volume<br>of synthetic blood | 32 out of 32 pass at 80<br>mmHg | Passed<br>3 non-consecutive lots tested, using a<br>sample size of 32/lot<br>Lot 0923: 32 out of 32 pass at 80<br>mmHg<br>Lot 1023: 32 out of 32 pass at 80<br>mmHg<br>Lot 1123: 32 out of 32 pass at 80<br>mmHg |
| Flammability<br>16 CFR Part<br>1610-2008 | Response of materials<br>to heat and flame | Class I | Passed<br>3 non-consecutive lots tested, using a<br>sample size of 32/lot<br>Lot 0923: Class 1<br>Lot 1023: Class 1<br>Lot 1123: Class 1 |
| ISO 10993-5 | Evaluate potential for | Under the conditions | Passed |
| | Table 2: Summary of Performance Testing | |
|--|-----------------------------------------|--|
{9}------------------------------------------------
| Tests for in vitro<br>cytotoxicity | cytotoxicity | of the study,<br>non-cytotoxic | |
|-------------------------------------------------------------------|----------------------------------------------|-----------------------------------------------------------|--------|
| ISO 10993-10<br>Tests for irritation<br>and skin<br>sensitization | Evaluate potential for<br>irritation | Under the conditions<br>of the study, not an<br>irritant | Passed |
| | Evaluate potential for<br>skin sensitization | Under the conditions<br>of the study, not a<br>sensitizer | Passed |
>> Biocompatibility of patient-contacting material
The biocompatibility of the subject device was determined via the evaluation according to the ISO 10993-5: 2009 and ISO 10993-10:2010 standards.
## H. Summary of Clinical Testing
No clinical study is included in this submission.
#### I. Conclusion
The conclusions drawn from the nonclinical tests demonstrate that the subject device in this 510(K) submission, the Disposable Medical Face Mask, is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K222809.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.