K240253 · Aerin Medical, Inc. · BSP · Mar 21, 2024 · Anesthesiology
Device Facts
Record ID
K240253
Device Name
Reach Needle
Applicant
Aerin Medical, Inc.
Product Code
BSP · Anesthesiology
Decision Date
Mar 21, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5150
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Reach Needle is used for in injecting local anesthetics into a patient to provide regional anesthesia.
Device Story
Reach Needle is a sterile, single-use, 27-gauge stainless steel needle with a reinforced shaft and plastic luer-lock hub. Designed for regional anesthesia, it features a 4-inch length, a 30-degree distal bend, and a short-bevel Quincke-style tip with a 2mm distal stop for consistent injection depth. Used by clinicians to access difficult anatomical areas, such as nasal structures, for local anesthetic delivery. The device provides a physical conduit for fluid administration; it does not involve electronic processing, software, or automated decision-making. Benefits include reduced needle flex and controlled injection depth, facilitating precise anesthetic placement.
Clinical Evidence
No clinical data. Substantial equivalence established via non-clinical bench testing, including compliance with ISO 80369-7, ISO 9626, ISO 7864, and package integrity per ISO 11607-1/2. Biocompatibility testing performed per ISO 10993-1, including cytotoxicity, irritation, sensitivity, systemic toxicity, and pyrogenicity.
Technological Characteristics
27-gauge stainless steel cannula with reinforced plastic hub; 4-inch length; 30-degree distal bend; short-bevel Quincke tip; 2mm distal stop; luer lock connector. Standards: ISO 80369-7 (connectors), ISO 9626 (needle tubing), ISO 7864 (sterile needles), ISO 11607-1/2 (packaging), ISO 10993-1 (biocompatibility). Sterile, single-use.
Indications for Use
Indicated for adults (≥ 22 years) requiring regional anesthesia via injection of local anesthetics.
Regulatory Classification
Identification
An anesthesia conduction needle is a device used to inject local anesthetics into a patient to provide regional anesthesia.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 21, 2024
Aerin Medical Inc. Teri Feeley Regulatory Affairs Manager 2565 Leghorn St. Mountain View, California 94043
Re: K240253
Trade/Device Name: Reach Needle Regulation Number: 21 CFR 868.5150 Regulation Name: Anesthesia Conduction Needle Regulatory Class: Class II Product Code: BSP Dated: January 26, 2024 Received: January 30, 2024
Dear Teri Feeley:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.
{2}------------------------------------------------
See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/2/Picture/3 description: The image shows the text "Bradley Q. Quinn -S" in a large, sans-serif font. The text is black and is set against a white background. The letters are evenly spaced and the text is easy to read.
Bradley Quinn Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known)
K240253
Device Name
Reach Needle
Indications for Use (Describe)
The Reach Needle is used for in injecting local anesthetics into a patient to provide regional anesthesia.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Aerin Medical. The word "Aerin" is in a large, blue, sans-serif font. Below "Aerin" is the word "MEDICAL" in a smaller, sans-serif font. To the right of "Aerin" are several green circles of varying sizes.
# 510(k) Summary
## General Information
| Submitter Information | |
|------------------------------------|-----------------------------------------------------------------|
| Company: | Aerin Medical Inc. |
| Submitter address: | 2565 Leghorn St.<br>Mountain View, CA 94043 |
| Contact Person: | Teri Feeley<br>Regulatory Affairs Manger<br>Phone: 210-827-1618 |
| Establishment Registration Number: | 3011625895 |
| Date Prepared: | 15 Mar 2024 |
| Name of Device | |
| Brand Name: | Reach Needle |
| Common Name: | Needle, Conduction, Anesthetic (w/wo introducer) |
| Classification Name: | Anesthesia conduction needle |
| Device Class: | Class II |
| Product Code: | BSP |
| CFR Regulation: | 21 CFR 868.5150 |
| Predicate Device: | Entellus Reinforced Anesthesia Needle (K163435) |
| Device Description | |
The Reach Needle is a sterile, single use, 4-inch-long needle used for administering anesthetic solutions for regional anesthesia. The Reach Needle is 27-qauge needle that is reinforced to reduce needle flex during administration of anesthesia solutions. The needle tip is a short bevel quincke style tip that extends approximately 2mm past the reinforcing shaft, providing a distal stop for consistent injection depth. The distal end of the needle has a slight bend to help access difficult to reach areas such as nasal anatomy. The needle hub is compatible with luer lock connectors. The Reach Needle is intended for use in adults (≥ 22 years).
### Indications for Use
The Reach Needle is used for injecting local anesthetics into a patient to provide regional anesthesia.
### Summary of Technological characteristics compared to the predicate device
The subject device has the same technological characteristics as the predicate including the principle of operation, performance, and design features. The subject device and predicate device both have a stainless-steel cannula that is reinforced with a plastic hub to deliver local anesthetic solutions for regional anesthesia. Both devices are 27-gauge needles with a reinforced sleeve, a quincke style tip, 2mm distal stop, and a bend incorporated at the distal end of the reinforced cannula. The subject device cannula is 4 inches long with a 30° bend and a Short Bevel tip; whereas the predicate device cannula is 3.5 inches long with an 18° bend and long, beveled tip.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Aerin Medical. The word "Aerin" is in a large, blue, sans-serif font. Below "Aerin" is the word "MEDICAL" in a smaller, sans-serif font. To the right of "Aerin" are several green circles of varying sizes, arranged in a pattern.
# 510(k) Summary
| | Subject Device | Predicate Device |
|------------------------------|----------------|---------------------------------------------------------------------------------------------|
| Intended Use | Same | For use in injecting local<br>anesthetics into a patient to<br>provide regional anesthesia. |
| Use | Same | Single use |
| Materials | Same | Stainless steel needle and<br>plastic hub |
| Hub design | Same | Standard luer lock connector |
| Needle gauge | Same | 27G |
| Needle length | 4 inches | 3.5 inches |
| Device reinforcement | Same | Reinforced sleeve |
| Needle bend | 30 degree bend | 18 degree bend |
| Insertion stop (distal stop) | Same | 2mm insertion depth stop |
| Needle tip | Same | Quincke style |
| | Short bevel | Long bevel |
### Summary of non-clinical test
Device Performance was tested to comply to:
- ISO 80369-7 ●
- ISO 9626 ●
- ISO 7864 ●
Package Integrity per ISO 11607-1 and 11607-2.
Reach needle is an externally communicating device with direct tissue contact for limited duration; biocompatibility testing is in compliance with ISO 10993-1 and included:
- Cytotoxicity .
- . Irritation
- Sensitivity ●
- Systemic toxicity ●
- . Pyrogenicity
### Summary of clinical test
Clinical testing was not necessary for this device.
#### Conclusion
The subject device is substantially equivalent to the predicate device based on a comparison of the intended use, indications for use, and technological characteristics and the minor differences do not raise new questions of safety or effectiveness. Therefore, the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.