HCG Home Use Pregnancy Test Strip (Colloidal Gold), HCG Home Use Pregnancy Test Cassette (Colloidal Gold), HCG Home Use Pregnancy Test Midstream (Colloidal Gold)
K240242 · Anhui Deepblue Medical Technology Co., Ltd. · LCX · Oct 11, 2024 · Clinical Chemistry
Device Facts
Record ID
K240242
Device Name
HCG Home Use Pregnancy Test Strip (Colloidal Gold), HCG Home Use Pregnancy Test Cassette (Colloidal Gold), HCG Home Use Pregnancy Test Midstream (Colloidal Gold)
Applicant
Anhui Deepblue Medical Technology Co., Ltd.
Product Code
LCX · Clinical Chemistry
Decision Date
Oct 11, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1155
Device Class
Class 2
Indications for Use
The HCG Home Use Pregnancy Test Strip (Colloidal Gold) is a rapid chromatographic immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in urine to aid in the early detection of pregnancy. The test strip is designed for over-the-counter use. Important note regarding positive results: this test may give positive results even if you are not pregnant. All results should be confirmed by your healthcare providers, especially when making decisions about future medical care. This device is intended for home use only. The HCG Home Use Pregnancy Test Cassette (Colloidal Gold) is a rapid chromatographic immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in urine to aid in the early detection of pregnancy. The test cassette is designed for over-the-counter use. Important note regarding positive results: this test may give positive results even if you are not pregnant. All results should be confirmed by your healthcare providers, especially when making decisions about future medical care. This device is intended for home use only. The HCG Home Use Pregnancy Test Midstream (Colloidal Gold) is a rapid chromatographic immunoassay for the quantitative detection of human chorionic gonadotropin (hCG) in urine to aid in the early detection of pregnancy. The test midstream is designed for the over-the-counter use. Important note regarding positive results: this test may give positive results even if you are not pregnant. All results should be confirmed by your healthcare providers, especially when making decisions about future medical care. This device is intended for home use only.
Device Story
Rapid chromatographic immunoassay for qualitative hCG detection in urine; utilizes monoclonal hCG antibodies and colloidal gold tracer. Specimen migrates via capillary action; positive results indicated by red line at test region; control line confirms procedural validity. Available in strip, cassette, and midstream formats. Intended for over-the-counter home use by lay users. Results provide early pregnancy indication; positive results require clinical confirmation by healthcare providers for medical decision-making.
Clinical Evidence
Bench testing only. Precision studies (single-site/single-day, single-site/multi-day, multi-site) performed with 60-900 replicates per concentration (0-100 mIU/mL). Interference testing (endogenous/exogenous substances, pH, specific gravity, β-core fragment) showed no interference. Method comparison study (n=300) and lay-user study (n=300) demonstrated 100% agreement with professional testing using the predicate device.
Technological Characteristics
Lateral flow sandwich immunochromatographic assay. Components: nitrocellulose membrane with anti-rat-α-hCG monoclonal antibody test line; glass fiber pad with gold nanoparticle-conjugated anti-rat-β-hCG monoclonal antibody; goat anti-mouse polyclonal control line. Formats: strip, cassette (with dropper/cup), or midstream (plastic housing). No electronic components; visual readout.
Indications for Use
Indicated for qualitative detection of hCG in urine to aid early pregnancy detection. Intended for over-the-counter, home use by women of childbearing age.
Regulatory Classification
Identification
A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.
Predicate Devices
hCG Urine Test Strip, hCG Urine Test Cassette, hCG Urine Test Midstream (K200133)
Submission Summary (Full Text)
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY ONLY
## I Background Information:
A 510(k) Number
K240242
B Applicant
Anhui DEEPBLUE Medical Technology Co. LTD.
C Proprietary and Established Names
HCG Home Use Pregnancy Test Strip (Colloidal Gold), HCG Home Use Pregnancy Test Cassette (Colloidal Gold), HCG Home Use Pregnancy Test Midstream (Colloidal Gold)
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| LCX | Class II | 21 CFR 862.1155 - Human Chorionic Gonadotropin | CH - Clinical Chemistry |
## II Submission/Device Overview:
A Purpose for Submission:
New device
B Measurand:
Human Chorionic Gonadotropin (hCG)
C Type of Test:
Qualitative chromatographic immunoassay
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
K240242 - Page 2 of 13
# III Intended Use/Indications for Use:
## A Intended Use(s):
See Indications for Use below.
## B Indication(s) for Use:
The HCG Home Use Pregnancy Test Strip (Colloidal Gold) is a rapid chromatographic immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in urine to aid in the early detection of pregnancy. The test strip is designed for over-the-counter use.
Important note regarding positive results: this test may give positive results even if you are not pregnant. All results should be confirmed by your healthcare providers, especially when making decisions about future medical care.
This device is intended for home use only.
The HCG Home Use Pregnancy Test Cassette (Colloidal Gold) is a rapid chromatographic immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in urine to aid in the early detection of pregnancy. The test cassette is designed for over-the-counter use.
Important note regarding positive results: this test may give positive results even if you are not pregnant. All results should be confirmed by your healthcare providers, especially when making decisions about future medical care.
This device is intended for home use only.
The HCG Home Use Pregnancy Test Midstream (Colloidal Gold) is a rapid chromatographic immunoassay for the quantitative detection of human chorionic gonadotropin (hCG) in urine to aid in the early detection of pregnancy. The test midstream is designed for the over-the-counter use.
Important note regarding positive results: this test may give positive results even if you are not pregnant. All results should be confirmed by your healthcare providers, especially when making decisions about future medical care.
This device is intended for home use only.
## C Special Conditions for Use Statement(s):
OTC - Over The Counter
The device is intended for home-use only and not intended for clinical-care settings.
## D Special Instrument Requirements:
None
{2}
K240242 - Page 3 of 13
## IV Device/System Characteristics:
### A Device Description:
The test comes in three test formats (strip, cassette, and midstream) named HCG Home Use Pregnancy Test Strip (Colloidal Gold), HCG Home Use Pregnancy Test Cassette (Colloidal Gold) and HCG Home Use Pregnancy Test Midstream (Colloidal Gold), respectively. The test strip format is packaged as a single test strip sealed in a desiccated aluminum pouch with instructions for use. The midstream format is a test strip assembled in a plastic housing with an absorbent tip and is packaged as a single test sealed in a desiccated aluminum pouch with instructions for use. The cassette format is the test strip assembled in a plastic cassette and is packaged as a single test with a dropper, a collection container, and instructions for use.
### B Principle of Operation:
HCG Home Use Pregnancy Test Strip (Colloidal Gold), HCG Home Use Pregnancy Test Cassette (Colloidal Gold) and HCG Home Use Pregnancy Test Midstream (Colloidal Gold) are qualitative, lateral flow sandwich immunochromatographic assays for the in vitro detection of human chorionic gonadotropin (hCG) in urine. Each test reagent strip consists of an anti-rat-α-hCG monoclonal antibody coated test line on the nitrocellulose membrane and a dried glass fiber membrane pad containing an anti-rat-β-hCG monoclonal antibody labeled with gold nanoparticle conjugate. The control antibodies are goat anti-mouse polyclonal antibody coated at the quality control region on the nitrocellulose membrane.
To use the strip format of the test, the user dips the sample pad into urine collected in a collection cup for at least 10 seconds. To use the cassette format, the user applies 3 full drops of urine specimen to the specimen well on the cassette. To use the midstream format, the user applies a specimen by keeping the absorbent tip in a urine midstream for at least five (5) seconds.
After application of the urine specimen, the test result is shown in the result window and read visually after 5 minutes of urine application. Two distinct colored lines, one in the test zone and another in the control zone indicate a positive test result (pregnant). Absence of a colored line in the test zone and only a colored line in the control zone indicates a negative test result (not pregnant). Absence of a colored line in the control zone, even in the presence of a colored line in the test zone, indicates an invalid test result.
### V Substantial Equivalence Information:
#### A Predicate Device Name(s):
hCG Urine Test Strip, hCG Urine Test Cassette, hCG Urine Test Midstream
#### B Predicate 510(k) Number(s):
K200133
{3}
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K240242 | K200133 |
| --- | --- | --- |
| Device Trade Name | HCG Home Use Pregnancy Test Strip (Colloidal Gold) | hCG Urine Test Strip |
| | HCG Home Use Pregnancy Test Cassette (Colloidal Gold) | hCG Urine Test Cassette |
| | HCG Home Use Pregnancy Test Midstream (Colloidal Gold) | hCG Urine Test Midstream |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | Qualitative detection of human chorionic gonadotropin (hCG) in urine to aid in the early detection of pregnancy. | Same |
| Specimen | Urine | Same |
| Methodology | Chromatographic immunoassay | Same |
| Analytical Sensitivity | 25 mIU/mL | Same |
| General Device Characteristic Differences | | |
| Specific Gravity Interference | No interference in urine with specific gravity 1.000 -1.035 | No interference in urine with specific gravity 1.000 -1.039 |
| High Dosage Hook Effect | No high dosage hook effect for hCG up to 1,000,000 mIU/mL | No high dosage hook effect for hCG up to 850,000 mIU/mL |
VI Standards/Guidance Documents Referenced:
Clinical Laboratory Standard Institute (CLSI) EP07: Interference Testing in Clinical Chemistry, 3rd edition.
CLSI EP25-A: Evaluation of Stability of In Vitro Diagnostic Reagents.
CLSI EP05-A3: Evaluation of Precision of Quantitative Measurement Procedures; 3rd edition
K240242 - Page 4 of 13
{4}
VII Performance Characteristics (if/when applicable):
A Analytical Performance:
1. Precision/Reproducibility:
Single-Site/Single-Day Precision Study
A precision study was performed using a pooled negative female urine sample spiked with hCG (traceable to the 5th WHO IS) to obtain samples with hCG concentration of 0, 12.5, 18.75, 25, 50, and 100 mIU/mL. Each sample was tested using 3 lots of HCG Home Use Pregnancy Test Cassette (Colloidal Gold), 3 lots of HCG Home Use Pregnancy Test Strip (Colloidal Gold) and 3 lots of HCG Home Use Pregnancy Test Midstream (Colloidal Gold). Each sample was tested in replicates of 20 per lot of device format at 1 site on 1 day by 3 different operators. A total of 60 replicates were performed per test format per hCG concentration. Summary of results are presented in the following tables.
HCG Home Use Pregnancy Test Cassette (Colloidal Gold)
| hCG concentration(mIU/mL) | Lot 1 | | Lot 2 | | Lot 3 | | Total result | | % Negative | % Positive |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | - | + | - | + | - | + | - | + | | |
| 0 | 20 | 0 | 20 | 0 | 20 | 0 | 60 | 0 | 100 | 0 |
| 12.5 | 20 | 0 | 20 | 0 | 20 | 0 | 60 | 0 | 100 | 0 |
| 18.75 | 12 | 8 | 11 | 9 | 12 | 8 | 35 | 25 | 58.33 | 41.67 |
| 25 | 0 | 20 | 0 | 20 | 0 | 20 | 0 | 60 | 0 | 100 |
| 50 | 0 | 20 | 0 | 20 | 0 | 20 | 0 | 60 | 0 | 100 |
| 100 | 0 | 20 | 0 | 20 | 0 | 20 | 0 | 60 | 0 | 100 |
HCG Home Use Pregnancy Test Strip (Colloidal Gold)
| hCG concentration(mIU/mL) | Lot 4 | | Lot 5 | | Lot 6 | | Total result | | % Negative | % Positive |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | - | + | - | + | - | + | - | + | | |
| 0 | 20 | 0 | 20 | 0 | 20 | 0 | 60 | 0 | 100 | 0 |
| 12.5 | 20 | 0 | 20 | 0 | 20 | 0 | 60 | 0 | 100 | 0 |
| 18.75 | 11 | 9 | 12 | 8 | 12 | 8 | 35 | 25 | 58.33 | 41.67 |
| 25 | 0 | 20 | 0 | 20 | 0 | 20 | 0 | 60 | 0 | 100 |
| 50 | 0 | 20 | 0 | 20 | 0 | 20 | 0 | 60 | 0 | 100 |
| 100 | 0 | 20 | 0 | 20 | 0 | 20 | 0 | 60 | 0 | 100 |
K240242 - Page 5 of 13
{5}
HCG Home Use Pregnancy Test Midstream (Colloidal Gold)
| hCG concentration (mIU/mL) | Lot 7 | | Lot 8 | | Lot 9 | | Total result | | % Negative | % Positive |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | - | + | - | + | - | + | - | + | | |
| 0 | 20 | 0 | 20 | 0 | 20 | 0 | 60 | 0 | 100 | 0 |
| 12.5 | 20 | 0 | 20 | 0 | 20 | 0 | 60 | 0 | 100 | 0 |
| 18.75 | 11 | 9 | 12 | 8 | 12 | 8 | 35 | 25 | 58.33 | 41.67 |
| 25 | 0 | 20 | 0 | 20 | 0 | 20 | 0 | 60 | 0 | 100 |
| 50 | 0 | 20 | 0 | 20 | 0 | 20 | 0 | 60 | 0 | 100 |
| 100 | 0 | 20 | 0 | 20 | 0 | 20 | 0 | 60 | 0 | 100 |
# Single-Site/Multi-Day Precision Study
A multi-day precision study was performed using a pooled negative female urine sample spiked with hCG (traceable to the $5^{\text{th}}$ WHO IS) to obtain samples with hCG concentration of 0, 12.5, 18.75, 25, 50, and $100\mathrm{mIU / mL}$ . Each sample was tested using 3 lots of HCG Home Use Pregnancy Test Cassette (Colloidal Gold), 3 lots of HCG Home Use Pregnancy Test Strip (Colloidal Gold) and 3 lots of HCG Home Use Pregnancy Test Midstream (Colloidal Gold). Each sample was tested in 20 replicates per run, 2 runs per day over the course of 5 days per lot per device test format. A total of 600 replicates per sampling method per hCG concentration. The table below summarized the precision data for each test format.
HCG Home Use Pregnancy Test Cassette (Colloidal Gold)
| hCG concentration (mIU/mL) | Lot 1 | | Lot 2 | | Lot 3 | | Total result | | % Negative | % Positive |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | - | + | - | + | - | + | - | + | | |
| 0 | 200 | 0 | 200 | 0 | 200 | 0 | 600 | 0 | 100 | 0 |
| 12.5 | 200 | 0 | 200 | 0 | 200 | 0 | 600 | 0 | 100 | 0 |
| 18.75 | 115 | 85 | 111 | 89 | 114 | 86 | 340 | 260 | 56.67 | 43.33 |
| 25 | 0 | 200 | 0 | 200 | 0 | 200 | 0 | 600 | 0 | 100 |
| 50 | 0 | 200 | 0 | 200 | 0 | 200 | 0 | 600 | 0 | 100 |
| 100 | 0 | 200 | 0 | 200 | 0 | 200 | 0 | 600 | 0 | 100 |
HCG Home Use Pregnancy Test Strip (Colloidal Gold)
| hCG concentration (mIU/mL) | Lot 4 | | Lot 5 | | Lot 6 | | Total result | | % Negative | % Positive |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | - | + | - | + | - | + | - | + | | |
| 0 | 200 | 0 | 200 | 0 | 200 | 0 | 600 | 0 | 100 | 0 |
| 12.5 | 200 | 0 | 200 | 0 | 200 | 0 | 600 | 0 | 100 | 0 |
| 18.75 | 113 | 87 | 113 | 87 | 112 | 88 | 338 | 262 | 56.33 | 43.67 |
| 25 | 0 | 200 | 0 | 200 | 0 | 200 | 0 | 600 | 0 | 100 |
| 50 | 0 | 200 | 0 | 200 | 0 | 200 | 0 | 600 | 0 | 100 |
| 100 | 0 | 200 | 0 | 200 | 0 | 200 | 0 | 600 | 0 | 100 |
K240242 - Page 6 of 13
{6}
HCG Home Use Pregnancy Test Midstream (Colloidal Gold)
| hCG concentration (mIU/mL) | Lot 4 | | Lot 5 | | Lot 6 | | Total result | | % Negative | % Positive |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | - | + | - | + | - | + | - | + | | |
| 0 | 200 | 0 | 200 | 0 | 200 | 0 | 600 | 0 | 100 | 0 |
| 12.5 | 200 | 0 | 200 | 0 | 200 | 0 | 600 | 0 | 100 | 0 |
| 18.75 | 118 | 82 | 113 | 87 | 114 | 86 | 345 | 255 | 57.5 | 42.5 |
| 25 | 0 | 200 | 0 | 200 | 0 | 200 | 0 | 600 | 0 | 100 |
| 50 | 0 | 200 | 0 | 200 | 0 | 200 | 0 | 600 | 0 | 100 |
| 100 | 0 | 200 | 0 | 200 | 0 | 200 | 0 | 600 | 0 | 100 |
## Multi-Site Precision Study
A multi-site precision study was performed using pooled negative female urine sample spiked with hCG (traceable to the 5th WHO IS) to obtain samples with hCG concentration of 0, 12.5, 18.75, 25, 50, and 100 mIU/mL. Each sample was tested using 3 lots of HCG Home Use Pregnancy Test Cassette (Colloidal Gold), 3 lots of HCG Home Use Pregnancy Test Strip (Colloidal Gold) and 3 lots of HCG Home Use Pregnancy Test Midstream (Colloidal Gold) at 3 sites. At each site, each sample was tested in 10 replicates per run, 2 runs per day over the course of 5 days per lot per device test format. All lots were tested at each site for each test format and a total of 900 replicates per test format per hCG concentration were obtained at 3 sites together. The table below summarized the precision data for each test format.
HCG Home Use Pregnancy Test Cassette (Colloidal Gold)
| hCG concentration (mIU/mL) | Site 1 (Lot 1, Lot 2, Lot 3) | | Site 2 (Lot 1, Lot2, Lot 3) | | Site 3 (Lot 1, Lot 2, Lot 3 | | Total result | | % Negative | % Positive |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | - | + | - | + | - | + | - | + | | |
| 0 | 300 | 0 | 300 | 0 | 300 | 0 | 900 | 0 | 100 | 0 |
| 12.5 | 300 | 0 | 300 | 0 | 300 | 0 | 900 | 0 | 100 | 0 |
| 18.75 | 150 | 150 | 150 | 150 | 150 | 150 | 450 | 450 | 50 | 50 |
| 25 | 0 | 300 | 0 | 300 | 0 | 300 | 0 | 900 | 0 | 100 |
| 50 | 0 | 300 | 0 | 300 | 0 | 300 | 0 | 900 | 0 | 100 |
| 100 | 0 | 300 | 0 | 300 | 0 | 300 | 0 | 900 | 0 | 100 |
K240242 - Page 7 of 13
{7}
HCG Home Use Pregnancy Test Strip (Colloidal Gold)
| hCG concentration (mIU/mL) | Site 1 (Lot4, Lot5, Lot6) | | Site 2 (Lot4, Lot5, Lot6) | | Site 3 (Lot4, Lot5, Lot6) | | Total result | | % Negative | % Positive |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | - | + | - | + | - | + | - | + | | |
| 0 | 300 | 0 | 300 | 0 | 300 | 0 | 900 | 0 | 100 | 0 |
| 12.5 | 300 | 0 | 300 | 0 | 300 | 0 | 900 | 0 | 100 | 0 |
| 18.75 | 150 | 150 | 150 | 150 | 140 | 160 | 440 | 460 | 51.11 | 48.89 |
| 25 | 0 | 300 | 0 | 300 | 0 | 300 | 0 | 900 | 0 | 100 |
| 50 | 0 | 300 | 0 | 300 | 0 | 300 | 0 | 900 | 0 | 100 |
| 100 | 0 | 300 | 0 | 300 | 0 | 300 | 0 | 900 | 0 | 100 |
HCG Home Use Pregnancy Test Midstream (Colloidal Gold)
| hCG concentration (mIU/mL) | Site 1 (Lot7, Lot8, Lot9) | | Site 2 (Lot7, Lot8, Lot9) | | Site 3 (Lot7, Lot8, Lot9) | | Total result | | % Negative | % Positive |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | - | + | - | + | - | + | - | + | | |
| 0 | 300 | 0 | 300 | 0 | 300 | 0 | 900 | 0 | 100 | 0 |
| 12.5 | 300 | 0 | 300 | 0 | 300 | 0 | 900 | 0 | 100 | 0 |
| 18.75 | 170 | 130 | 150 | 150 | 140 | 160 | 480 | 420 | 53.53 | 46.67 |
| 25 | 0 | 300 | 0 | 300 | 0 | 300 | 0 | 900 | 0 | 100 |
| 50 | 0 | 300 | 0 | 300 | 0 | 300 | 0 | 900 | 0 | 100 |
| 100 | 0 | 300 | 0 | 300 | 0 | 300 | 0 | 900 | 0 | 100 |
2. Linearity:
Linearity is not applicable since this is a qualitative test.
3. Analytical Specificity/Interference:
Interference from exogenous and endogenous substances:
A study was performed to evaluate potential interference from certain exogenous and endogenous substances for the HCG Home Use Pregnancy Test Cassette (Colloidal Gold), HCG Home Use Pregnancy Test Strip (Colloidal Gold) and HCG Home Use Pregnancy Test Midstream (Colloidal Gold). A urine pool from non-pregnant females was used to prepare samples with hCG levels of $0\mathrm{mIU / mL}$, $10\mathrm{mIU / mL}$ and $25\mathrm{mIU / mL}$, which were then spiked with potentially interfering substances at the concentrations listed below. Samples were tested in replicates of 5 per lot using 3 lots of each device format. The results demonstrated no interferences at the concentrations shown in the table below.
K240242 - Page 8 of 13
{8}
| Potential Interferent | Highest Concentration Tested that demonstrated no interference |
| --- | --- |
| **Endogenous Interferents** | |
| Albumin | 2000 mg/dL |
| Bilirubin | 2mg/dL |
| Ethanol | 1% |
| Glucose | 2000 mg/dL |
| Hemoglobin | 1000 mg/dL |
| Ketone | >80 mg/dL |
| Protein | 5.65 mmol/L |
| Erythrocytes | >250/μL |
| Leukocyte | >500/μL |
| Uric acid | 450 mmol/L |
| **Exogenous Interferents** | |
| Ascorbic acid | 20 mg/dL |
| Acetaminophen | 20 mg/dL |
| Aspirin | 20 mg/dL |
| Tetracycline | 20 mg/dL |
| Ampicillin | 20 mg/dL |
| β hydroxybutyrate | 2000 mg/dL |
| Benzoylecgonine | 10 mg/dL |
| Caffeine | 20 mg/dL |
| Cannabinol | 10 mg/dL |
| EDTA | 80 mg/dL |
| Ephedrine | 20 mg/dL |
| Folic acid | 800 μg/mL |
| Phenylpropanolamine | 20 mg/dL |
| Pheothiazine | 20 mg/dL |
| Thiophene | 20 mg/dL |
| Vitamin B1 | 800 μg/mL |
**Cross-Reactivity of Structurally Related Compounds:**
A study was performed to evaluate the potential cross-reactivity from luteinizing hormone (LH), follicle-stimulating hormone (FSH), and thyroid-stimulating hormone (TSH) on the HCG Home Use Pregnancy Test Cassette (Colloidal Gold), HCG Home Use Pregnancy Test Strip (Colloidal Gold) and HCG Home Use Pregnancy Test Midstream (Colloidal Gold) devices. Urine specimens from non-pregnant females were pooled and used to prepare samples with hCG levels of 0 mIU/mL, 12.5 mIU/mL and 25 mIU/mL. Then hCG samples were spiked with each cross-reactants at the following concentrations: 500 mIU/mL LH, 1000 mIU/mL FSH and 1 mIU/mL TSH. The samples were tested in replicates of 5 per lot using 3 lots of each device format. The results demonstrated no cross reactivity with LH at 500 mIU/mL, FSH at 1000 mIU/mL and TSH at 1 mIU/mL in either negative or positive urine samples.
K240242 - Page 9 of 13
{9}
K240242 - Page 10 of 13
## Effects of hCG β-core fragment:
A study was performed to evaluate the effect of hCG β-core fragment on the HCG Home Use Pregnancy Test Cassette (Colloidal Gold), HCG Home Use Pregnancy Test Strip (Colloidal Gold) and HCG Home Use Pregnancy Test Midstream (Colloidal Gold) devices. Negative female urine samples (0 mIU/mL) and negative samples spiked with hCG to levels of 10 mIU/mL and 25 mIU/mL were spiked with hCG β-core fragment at concentrations of 125,000, 250,000, 500,000 and 1,000,000 pmol/mL. The samples were tested in replicates of 10 per lot using three lots of each device format. The results demonstrated that the candidate devices are not affected by concentrations of hCG β-core fragment up to 1,000,000 pmol/mL.
## Effects of Urine pH:
A study was performed to evaluate the effects of pH on the HCG Home Use Pregnancy Test Cassette (Colloidal Gold), HCG Home Use Pregnancy Test Strip (Colloidal Gold) and HCG Home Use Pregnancy Test Midstream (Colloidal Gold) devices. Negative female urine samples (0 mIU/mL) and negative samples spiked with hCG to levels of 10 mIU/mL and 25 mIU/mL were adjusted to have pH values of 3.0, 3.5, 4.0, 4.5, 5.0, 5.5, 6.0, 6.5, 7.0, 7.5, 8.0, 8.5 and 9.0. The negative and positive hCG samples with the different pH levels were tested in replicates of 10 per lot using 3 lots of each device. The results demonstrated that changes in pH ranging from 3.0 to 9.0 do not interfere with either negative or positive results from the devices.
## Effects of Urine Specific Gravity:
A study was performed to evaluate the effects of urine specific gravity HCG Home Use Pregnancy Test Cassette (Colloidal Gold), HCG Home Use Pregnancy Test Strip (Colloidal Gold) and HCG Home Use Pregnancy Test Midstream (Colloidal Gold) devices. Negative female urine samples (0 mIU/mL) and negative samples spiked with hCG to levels of 10 mIU/mL and 25 mIU/mL were adjusted to have urine specific gravities of 1.000, 1.005, 1.010, 1.015, 1.020, 1.025, 1.030 and 1.035. The negative and positive hCG samples with the different urine gravity levels were tested in replicates of 10 per lot using 3 lots of each device format. The results demonstrated that changes in urine specific gravity from 1.000 to 1.035 do not interfere with either negative or positive results from the devices.
## Hook Effect:
A study was performed to evaluate the high dose hook effect on HCG Home Use Pregnancy Test Cassette (Colloidal Gold), HCG Home Use Pregnancy Test Strip (Colloidal Gold) and HCG Home Use Pregnancy Test Midstream (Colloidal Gold) devices. Negative female urine samples were spiked with hCG at concentrations of 5000, 10,000, 100,000, 500,000, 650,000, 850,000, 950,000 and 1,000,000 mIU/mL. The samples were tested in replicates of 30 using three device lots per test format. The results demonstrated that no hook effect was observed at hCG concentration up to 1,000,000 mIU/mL.
## 4. Assay Reportable Range:
Not applicable
{10}
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
HCG Home Use Pregnancy Test Cassette (Colloidal Gold), HCG Home Use Pregnancy Test Strip (Colloidal Gold) and HCG Home Use Pregnancy Test Midstream (Colloidal Gold) devices are calibrated against reference material traceable to WHO International Standard 5th edition (NIBSC code 07/364)
6. Detection Limit:
The detection limit was determined in the precision study (see Section VII.A.1 above)
7. Assay Cut-Off:
The device’s cut-off is 25 mIU/mL., See Precision/Reproducibility (Section VII.A.1) section above.
B Comparison Studies:
1. Method Comparison with Predicate Device:
A total of 300 urine samples were collected from women whose ages ranged from 18 to 48 years, who presented to the clinics for pregnancy testing. Approximately half of the 300 women were suspected to be pregnant and most of them were in the early stage of pregnancy (less than 5 weeks). The samples were masked and randomized prior to testing by professionals using 1 lot of the candidate devices and a single lot of the predicate device. One hundred (100) samples were tested using the HCG Home Use Pregnancy Test Cassette (Colloidal Gold), 100 samples were tested using the HCG Home Use Pregnancy Test Strip (Colloidal Gold, and 100 samples were tested using HCG Home Use Pregnancy Test Midstream (Colloidal Gold). Summary results are presented in the tables below.
HCG Home Use Pregnancy Test Cassette (Colloidal Gold)
| Candidate Device | Predicate Device | | Total | |
| --- | --- | --- | --- | --- |
| | | Positive | Negative | |
| Cassette | Positive | 50 | 0 | 50 |
| | Negative | 0 | 50 | 50 |
| | Total | 50 | 50 | 100 |
HCG Home Use Pregnancy Test Strip (Colloidal Gold)
| Candidate Device | Predicate Device | | Total | |
| --- | --- | --- | --- | --- |
| | | Positive | | Negative |
| Test Strip | Positive | 50 | 0 | 50 |
| | Negative | 0 | 50 | 50 |
| | Total | 50 | 50 | 100 |
HCG Home Use Pregnancy Test Midstream (Colloidal Gold)
K240242 - Page 11 of 13
{11}
| Candidate Device | Predicate Device | | Total | |
| --- | --- | --- | --- | --- |
| | | Positive | Negative | |
| Midstream | Positive | 50 | 0 | 50 |
| | Negative | 0 | 50 | 50 |
| | Total | 50 | 50 | 100 |
An additional method comparison study was performed in which 100 people tested their own sample for each test format. One false positive result was observed using the cassette format of the device when results were compared to the predicate. Limiting statements have been added to the labeling to indicate that false positives are possible when using this device.
2. Matrix Comparison:
Not Applicable. The device is intended for urine samples only.
C Clinical Studies:
1. Clinical Sensitivity:
Not applicable.
2. Clinical Specificity:
Not applicable.
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
A lay user study was conducted at 3 sites with 300 volunteers with diverse educational and occupational backgrounds who were between the age of 18 and 48. This included 100 lay users using the HCG Home Use Pregnancy Test Cassette (Colloidal Gold) at site 1, 100 lay-users using the HCG Home Use Pregnancy Test Strip (Colloidal Gold) at site 2, and 100 lay-users using the HCG Home Use Pregnancy Test Midstream (Colloidal Gold) at site 3. Lay users collected their own sample, and these samples were split into two groups, one group was for the lay-user testing, and the other was for the professional testing. All samples were masked and randomized prior to professional testing. Each lay user tested her own urine sample with only one test method on the candidate device following the instructions on the package insert. The same sample was tested by the professional using the predicate device. Then the results of lay-user tests were compared to results reported by a laboratory professional. The data shows that the agreement between lay-user results and professional results was 100% (below):
HCG Home Use Pregnancy Test Cassette (Colloidal Gold)
| Cassette | Professional User Predicate Device | | Total | |
| --- | --- | --- | --- | --- |
| | | Positive | | Negative |
| Lay-User | Positive | 50 | 0 | 50 |
| | Negative | 0 | 50 | 50 |
| | Total | 50 | 50 | 100 |
K240242 - Page 12 of 13
{12}
HCG Home Use Pregnancy Test Strip (Colloidal Gold)
| Test Strip | Professional User Predicate Device | | Total | |
| --- | --- | --- | --- | --- |
| | | Positive | | Negative |
| Lay-User | Positive | 50 | 0 | 50 |
| | Negative | 0 | 50 | 50 |
| | Total | 50 | 50 | 100 |
HCG Home Use Pregnancy Test Midstream (Colloidal Gold)
| Midstream | Professional User Predicate Device | | Total | |
| --- | --- | --- | --- | --- |
| | | Positive | | Negative |
| Lay-User | Positive | 50 | 0 | 50 |
| | Negative | 0 | 50 | 50 |
| | Total | 50 | 50 | 100 |
Each lay person was given a questionnaire that was completed at the end of the study. The questionnaire results indicated that lay-users found the tests easy to use, the results clear and easy to read, and the instructions for use easy to understand.
D Clinical Cut-Off:
Not applicable
E Expected Values/Reference Range:
Not applicable
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K240242 - Page 13 of 13
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.