K240226 · Oystershell NV · ORE · Oct 18, 2024 · SU
Device Facts
Record ID
K240226
Device Name
Pixie Skin Tag
Applicant
Oystershell NV
Product Code
ORE · SU
Decision Date
Oct 18, 2024
Decision
SESE
Submission Type
Traditional
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K240226 · Oct 18, 2024
Pixie Skin Tag
Oystershell NV
Post-market complaint records; International sales/incident reports (CE-marked devices)
The sponsor utilized post-market complaint data from international sales of the device and US sales of a reference device to demonstrate the safety profile and low incidence of adverse events in real-world use.
Post-market review of CE-marked PST devices; Post-market surveillance / Complaint analysis; Follow-up/Duration: Through end of 2022; Study Period: Up to end of 2022
Users of CE-marked Pixie Skin Tag devices outside the US; Sample Size: Not specified (based on sales through end of 2022); Number of Sites: Multiple (International)
Not applicable for this study
Incidence of serious and non-serious adverse events
Post-market review of reference Pixie devices; Post-market surveillance / Complaint analysis; Follow-up/Duration: 2021-2022; Study Period: 2021-2022
Users of reference Pixie devices (wart treatment) in the US; Sample Size: Over 3 million devices sold (over 2 million in US); Number of Sites: United States
Not applicable for this study
Incidence of serious and non-serious adverse events
Indications for Use
Pixie® Skin Tag is a cryogenic device for home use intended to treat skin tags in persons over the age of 21.
Device Story
Pen-like cryogenic device for home use; utilizes pressurized nitrous oxide (N2O) to destroy skin tag tissue. User adheres disposable thermally insulating adhesive bandage to skin surrounding tag; dispenses N2O into single-use applicator tip; rapid evaporation cools applicator to approximately -83°C. Applicator contacted with skin tag for 20 seconds; freezing causes cell death and blood flow stasis. Device intended for OTC use by patients; treats tags on neck, breast, armpits, or underarms. Benefits include removal of skin tags via cryosurgery. Post-treatment, tag falls away and new epidermis grows. Maximum 3 treatments per tag, 2-week intervals.
Clinical Evidence
No clinical trials performed. Evidence includes bench testing (temperature comparison, misuse/safety testing, bandage thermal protection), animal study (rabbit skin healing within 7 days), and usability studies (novice laypersons). Post-market data from CE-marked devices (outside US) showed no serious incidents and low rates of non-serious events (pain, discoloration, burns). Reference device post-market data (over 3 million sold) showed one serious event and very low non-serious event rates (0.012-0.019%).
Technological Characteristics
Pen-like cryogenic device; pressurized N2O propellant. Applicator tip cools via rapid evaporation to -83°C. Includes disposable thermally insulating adhesive bandages. Non-sterile. Materials same as reference device (K172373). Standards: ISO 10993-1, -5, -10, -12; ISO 14155; ISO 14971; ISO 20417; IEC 62366.
Indications for Use
Indicated for the treatment of skin tags in persons over the age of 21.
{0}------------------------------------------------
October 18, 2024
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Oystershell NV % Phil Triolo Founder and Principal Consultant Phil Triolo and Associates LC 86 Skycrest Lane Salt Lake City, Utah 84108
Re: K240226
Trade/Device Name: Pixie Skin Tag Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical Unit And Accessories Regulatory Class: Class II Product Code: ORE, GEH Dated: September 19, 2024 Received: September 20, 2024
Dear Phil Triolo:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
James H. Digitally signed by James H. Jang -S Date: 2024.10.18 Jang -S 17:30:13 -04'00'
For
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
K240266
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Submission Number (if known)
K240266
Device Name
Pixie Skin Tag
Indications for Use (Describe)
Pixie® Skin Tag is a cryogenic device for home use intended to treat skin tags in persons over the age of 21.
Type of Use (Select one or both, as applicable)
[X] Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Prescription Use (Part 21 CFR 801 Subpart D)
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows the logo for Oystershell. The logo consists of a stylized image of a person with their arms raised above their head, with a red dot above the head. To the right of the image is the word "Oystershell" in blue font. Below the word "Oystershell" is the phrase "Innovating Consumer Health" in a smaller font.
Traditional 510(k) Premarket Notification
Pixie Skin Taq
510(k) Summary Page 1
# 510(k) Summary
Date Prepared: 18 October 2024
## 1. SUBMITTER
- 510(k) Sponsor: Oystershell NV
Contact Person: Phil Triolo, PhD, RAC President and Senior Consultant Phil Triolo and Associates LC Phone: 801-699-9846 / Fax 801-328-2399 Email: philt@philt.com
- Phil Triolo and Associates, LC Prepared By: 86 Skycrest Lane SLC UT 84108 www.philt.com
### 2. DEVICE
| Trade/Device Name: | Pixie Skin Tag |
|----------------------|---------------------------------------------|
| Common Name: | Over-the-counter cryogenic skin tag remover |
| Classification Name: | Cryosurgical unit and accessories |
| Regulation number: | 21 CFR 878.4350 |
| Product Code: | ORE (Cryogenic Skin Tag Removal, OTC) |
| Device Class: | Class II |
#### 3. PREDICATE DEVICE
K211099 CryoConcepts Freeze'n Clear Skin Clinic Warts & Tags, Regulation 21 CFR 878.4350, Product Code GEH
## REFERENCE DEVICE
K172373 Compound W Nitro Freeze, Regulation 21 CFR 878.4350, Product Code GEH
### DEVICE DESCRIPTION 4.
Pixie Skin Taq is a pen-like cryogenic device that utilizes pressurized nitrous oxide (NoO, liquefied) to destroy skin tag tissue. The N2O is released into a single-use, disposable applicator, cooling the applicator as the result of rapid evaporation. The cooled applicator is briefly contacted with the skin tag while the surrounding tissue is protected by a disposable thermally insulating adhesive bandage. Each Pixie Skin Tag device is supplied in a carton with instructions for use and replacement applicator tips and insulating adhesive bandages.
### 5. INDICATIONS FOR USE
{5}------------------------------------------------
Pixie® Skin Tag is a cryogenic device for home use intended to treat skin tags in persons over the age of 21.
#### COMPARISON OF CHARACTERISTICS WITH THE PREDICATE DEVICE 6.
The main difference between the new Pixie Skin Tag and the predicate device is that the new Pixie Skin Tag uses compressed nitrous oxide as the cryogen instead of a combination of dimethyl ether and propane. The evaporation of nitrous oxide results in cooling of the applicator that is contacted with the skin tag during treatments to a lower temperature (-83°C) than the evaporation of the dimethyl ether/propane mixture (-55°C). Contact of the skin tag with the chilled applicator of the new and predicate devices results in cell death due to freezing and blood flow stasis that occurs during thawing of the affected tissue. To protect non-target skin surrounding the skin tag during treatment using the Pixie Skin Tag applicator, thermally insulating bandages are provided for the user to adhere to the skin surrounding the skin tag before treatments. In contrast, the predicate device is provided with tweezers that are used to isolate the skin tag from the surrounding skin to make it easier to contact the skin tag with the applicator without contacting non-target skin.
The Pixie Skin Tag device design and materials are the same as those of the reference Compound W Nitro Freeze Product cleared after FDA review of K172373. The devices differ in the indication for use of the new device to treat skin tags and the inclusion of the single-use, disposable thermally insulating bandages with the new devices.
The following table compares Pixie Skin Tag to the predicate devices with respect to intended use and technological characteristics. This comparison of the devices provides the detailed information that is the basis for the determination of substantial equivalence.
| Substantial Equivalence Table: Comparison of New Device with Legally Marketed Predicate<br>Device<br>None of the identified differences raise different questions of safety and effectiveness | | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ATTRIBUTE | NEW DEVICE<br>Oystershell Pixie Skin Tag (PST)<br>K240226 | PREDICATE DEVICE<br>CryoConcepts Freeze'n Clear (FnC)<br>Skin Clinic Warts & Tags<br>K211099 |
| Description | Pixie Skin Tag is a pen-like cryogenic<br>device that utilizes pressurized nitrous<br>oxide (N₂O, liquefied) to destroy skin tag<br>tissue. The N₂O is released into a single-<br>use, disposable applicator, cooling the<br>applicator to approximately -83°C as the<br>result of rapid evaporation. The cooled<br>applicator tip is contacted with the skin tag<br>for 20s while the surrounding tissue is<br>protected by a disposable thermally<br>insulating adhesive bandage. The frozen<br>skin tag falls away over time and new<br>epidermis grows in its place. One to three<br>treatments delivered on a 2-week schedule<br>may be delivered (maximum of 3<br>applications for any one skin tag). | The Freeze 'n Clear Skin Clinic for Warts<br>and Skin Tags is for OTC use and utilizes<br>a combination of dimethyl ether, propane<br>and isobutane (DMEP) delivered from the<br>canister into a single-use, disposable<br>applicator which acts as a reservoir for the<br>cryogenic gas. The gas rapidly evaporates<br>and cools the applicator to approximately<br>-55°C. After a 15-second wait, the<br>applicator is placed against the wart or skin<br>tag for 40 seconds which freezes the<br>targeted tissue. The frozen skin tag or wart<br>falls away over time and new epidermis<br>grows in its place. One to four treatments<br>with intervals of two weeks may be<br>required. |
| Contents | Each Pixie Skin Tag (PST) device is<br>supplied in an outer carton with an<br>instructions for use leaflet and a blister<br>pack containing 4 replacement single-use<br>applicator tips for Pixie 10mL and 9<br>replacement single-use tips for Pixie 21mL,<br>and a sheet of 10 disposable, single-use<br>insulating adhesive bandages. | Freeze 'n Clear Skin Clinic for Warts and<br>Skin Tags is supplied in an outer carton<br>with an instructions for use leaflet, multiple-<br>use actuator, 8 single-use applicator tips,<br>and a multiple-use tweezer. |
| Substantial Equivalence Table: Comparison of New Device with Legally Marketed Predicate<br>Device<br>None of the identified differences raise different questions of safety and effectiveness | | |
| ATTRIBUTE | NEW DEVICE<br>Oystershell Pixie Skin Tag (PST)<br>K240226 | PREDICATE DEVICE<br>CryoConcepts Freeze'n Clear (FnC)<br>Skin Clinic Warts & Tags<br>K211099 |
| Classification | Class II | Class II |
| Regulation | 21 CFR 878.4350 | 21 CFR 878.4350 |
| Name | Unit, Cryosurgical, and Accessories | Unit, Cryosurgical, and Accessories |
| Panel | General and Plastic Surgery | General and Plastic Surgery |
| Product Code | ORE<br>Cryogenic Skin Tag Removal, OTC | GEH<br>Cryosurgical unit and accessories |
| Intended and<br>Indicated Use | Pixie® Skin Tag is a cryogenic device for<br>home use intended to treat skin tags in<br>persons over the age of 21. | The Freeze 'n Clear Skin Clinic for Warts<br>and Skin Tags product is intended for the<br>OTC treatment of common warts, plantar<br>warts, and skin tags. |
| Target<br>Population | Persons over the age of 21 | No age restriction |
| Anatomical<br>Site(s) | Only treat skin tags if they are located on<br>the neck, breast, armpits, or underarms via<br>a narrow stalk | Instructions do not restrict the use of the<br>device to treat skin tags to any specific<br>anatomical locations. |
| Where used | Home (over-the-counter) Use | Home (over-the-counter) Use |
| Principles of<br>Operation | The application of extreme cold removes<br>heat from the targeted skin tag, resulting in<br>freezing and destruction of the tissue. | The application of extreme cold removes<br>heat from the targeted skin tag, resulting in<br>freezing and destruction of the tissue. |
| Technology | Nitrous Oxide is used to cool an applicator<br>which is applied to the skin tag.<br>The same technology is used by Pixie<br>devices cleared for use to treat warts after<br>FDA review of K172373. | DMEP (mixture of dimethyl ether and<br>propane) is used to cool a foam applicator<br>which is applied to the lesion. |
| Mode of<br>Application | User adheres a disposable, thermally<br>insulating bandage to the skin surrounding<br>the skin tag.<br>The user dispenses Nitrous Oxide for 2s<br>into the disposable applicator tip attached<br>to the pen. The applicator is immediately<br>applied to the skin tag for a recommended<br>duration of 20s.<br>The bandage is removed and discarded.<br>Up to 3 treatments may be applied to each<br>skin tag, separated by 2 weeks. | The applicator stick is attached to the<br>actuator. The actuator is pressed into the<br>canister of DMEP (mixture of dimethyl<br>ether and propane) for 3s to 5s to dispense<br>the DMEP and saturate the applicator. The<br>applicator and actuator are removed from<br>the cannister assembly, and the user waits<br>15s before attempting treatment.<br>To treat the skin tag, the provided tweezers<br>are used to grasp the skin tag and pull it<br>away from its base, exposing the stalk (or<br>stem) of the skin tag. The applicator tip is<br>then applied to the stalk of the skin tag for<br>up to a recommended duration of 40s.<br>Up to 4 treatments may be applied to each<br>skin tag, separated by 2 weeks. |
| Number of<br>treatments | Available in 10mL and 21mL sizes,<br>capable of delivering, respectively, 5 and<br>10 treatments. | Available in one size, capable of delivering<br>up to 8 treatments. |
| Flammability | Non-flammable, but will support fire | Highly Flammable |
{6}------------------------------------------------
## K240226
Oystershell
## Traditional 510(k) Premarket Notification
# Pixie Skin Tag
{7}------------------------------------------------
| Image: logo | Oystershell<br>INNOVATING CONSUMER HEALTH |
|-------------|-------------------------------------------|
|-------------|-------------------------------------------|
## Traditional 510(k) Premarket Notification
Pixie Skin Tag
| Substantial Equivalence Table: Comparison of New Device with Legally Marketed Predicate<br>Device<br>None of the identified differences raise different questions of safety and effectiveness | | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ATTRIBUTE | NEW DEVICE<br>Oystershell Pixie Skin Tag (PST)<br>K240226 | PREDICATE DEVICE<br>CryoConcepts Freeze'n Clear (FnC)<br>Skin Clinic Warts & Tags<br>K211099 |
| Storage<br>Conditions<br>and Safety<br>Precautions | Contents under pressure<br>Never pierce or burn product, even after<br>use.<br>Keep away from fire or flame.<br>Do not smoke while using this product.<br>Do not expose to heat or store at<br>temperatures above 122°F (50°C).<br>Store unused product at room temperature<br>away from heat.<br>Keep out of reach of children. | Do not smoke while using the product.<br>Do not puncture or incinerate canister.<br>Do not expose to heat or store at<br>temperatures above 120°F.<br>Store at room temperature, away from the<br>reach of children, and out of the way of<br>direct sunlight or other heat sources. |
| Disposal | Device is disposable. | Device is disposable after emptied of<br>cryogen. |
| Shelf Life | 3 years (unopened) | 3 years (unopened) |
| Sterility | Product is provided non-sterile | Product is provided non-sterile |
## 7. PERFORMANCE EVALUATIONS
### 7.1 Bench Studies
A comparison of the temperatures reached by the new and predicate applicators was conducted and demonstrating that, after dispensing the cryogenic gases into the applicators, the new Pixie Skin Tag applicator reaches and maintains a temperature of approximately -83°C, whereas the predicate applicator reaches a temperature of approximately -33°C, over recommended application durations.
Results of numerous studies demonstrate the ability of PST, after exposure to elevated temperatures or anticipated device misuse, to be activated and dispense N2O to the applicators, and, independently, to safely evacuate the pressurized nitrous oxide; and to protect the N2O cartridge when the device is dropped.
An analysis of the ability of the thermally insulating bandages to protect the skin from thermal damage during Pixie Skin Tag treatments demonstrated adequate thermal protection.
## 7.2 Animal Studies
A study conducted in rabbits demonstrated that application of PST to healthy rabbit skin resulted in epithelial damage that did not extend deeper than the superficial fascia and evidence of wound healing occurring within 7 days after application.
## 7.3 Clinical Evidence
Results of three usability studies demonstrated that novice laypersons could set up and apply simulated treatments after reading the provided instructions for use.
CE-marked PST devices have been sold in the EU and other countries for several years. Post-market review of complaints reported for CE-marked PST devices sold through the end of 2022 in countries outside the US revealed no reports of serious incidents, and a total of only 5 reported incidents for non-serious events identified as pain, discoloration, blister formation, skin burns, and hypersensitivity reactions. These are known side effects/risks associated with use of
{8}------------------------------------------------
### Traditional 510(k) Premarket Notification 510(k) Summary Oystershel Pixie Skin Tag Page 5
the device. Further, the low incidence of reported skin burns strongly implies that the insulating bandages are being used properly, and are effective.
Post-market review of complaints reported for the over 3 million devices that have been sold, including over 2 million sold in the US, for reference Pixie devices intended to treat warts, revealed only one serious event. The very low and acceptable rates of reported non-serious events, all of which are communicated to end users in the instructions for use, were 0.019% in 2021 and 0.012% in 2022 (based on an assumption that 50% of devices sold were used).
### 8. SOFTWARE
No software is employed in the Pixie Skin Tag device.
### 9. BIOCOMPATIBILITY
Pixie Skin Tag pen and applicator components and their processing are the same as those of the reference predicate device which is intended for the same user (patient) contact. Therefore, the biological and toxicological risks associated with the Pixie Skin Tag applicator assembly are the same as the reference predicate device and are low and acceptable.
The thermally insulating bandages make direct contact with intact skin. Results of in-vitro cytotoxicity (MEM Elution), sensitization, and intracutaneous injection studies, conducted in accordance with ISO 10993-5, ISO 10993-10, and ISO 10993-10, respectively, met all acceptance criteria, demonstrating low and acceptable biological risks.
## 10. STERILIZATION
Neither the new Pixie Skin Tag nor the predicate device is sterilized.
## 11. STANDARDS
●
The following standards were referenced in this 510k to support substantial equivalence:
- ISO 10993-1
- ISO 10993-12
- ISO 14155 ●
- ISO 10993-5
-
-
- ISO 20417 ●
- ISO 10993-10 ●
- IEC 62366 ●
- ISO 14971
## 12. CONCLUSION
The results of bench, pre-clinical, and usability studies and comparisons with the predicate device provide the evidence needed to find that Oystershell's new Pixie Skin Tag devices are substantially equivalent to the predicate CryoConcepts Freeze'n Clear Skin Clinic Warts & Tags.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.