K240220 · Taiwan Aulisa Medical Devices Technologies, Inc. · FLL · Sep 27, 2024 · General Hospital
Device Facts
Record ID
K240220
Device Name
Aulisa Temperature Module (TM0002)
Applicant
Taiwan Aulisa Medical Devices Technologies, Inc.
Product Code
FLL · General Hospital
Decision Date
Sep 27, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Software as a Medical Device, Pediatric
Indications for Use
Aulisa Temperature Module (TM0002) is a battery-operated electronic device with indication for use in continuously measuring and monitoring armpit body temperature of adults, pediatrics, and infants and transmission of the measuring result via wireless signal. Aulisa Temperature Module (TM0002) is a non-invasive and reusable device for single patient use with intended environments of use are hospitals, medical facilities, home care, and subacute environments. The parameters derived by the Aulisa Temperature Module (TM0002) are transmitted to a commercially available mobile device that runs an Aulisa-developed application for display and review.
Device Story
Battery-operated electronic sensor module; measures axillary body temperature continuously. Inputs: thermal signals from armpit. Transformation: raw temperature data transmitted via Bluetooth 5.2 to third-party mobile device (iOS/Android). Output: temperature display and review via Aulisa-developed application. Used in hospitals, clinics, home care, and subacute settings by patients or caregivers. Software application provides visual and auditory alarms if temperature exceeds pre-set limits or technical errors occur. Benefits: enables continuous remote monitoring of body temperature, facilitating timely clinical intervention.
Clinical Evidence
Bench testing only. Accuracy validated per ISO 80601-2-56 and ASTM E1112. No clinical validation required as temperature output is derived directly from input signal without adjustment.
Technological Characteristics
Reusable sensor module (24mm x 32.8mm x 5.5mm) powered by CR1216 coin battery. Sensing principle: electronic thermometer. Connectivity: Bluetooth 5.2 to mobile device. Biocompatibility: ISO 10993-5, -10. Standards: IEC 60601-1, IEC 60601-1-11, IEC 60601-1-2, ANSI C63.27.
Indications for Use
Indicated for continuous armpit body temperature monitoring of adults, pediatrics, and infants. Non-invasive, reusable, single-patient use device for hospital, medical facility, home care, and subacute environments.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
Guardian Angel Rx GA2000 Series Digital Vital Sign Monitoring System (K203208)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
September 27, 2024
Taiwan Aulisa Medical Devices Technologies, Inc. Jou An Hsu Regulatory Affairs Specialist 6F-2, No. 3-1, YuanQu St., Nangang Dist., Taipei City, NA 115 Taiwan
Re: K240220
Trade/Device Name: Aulisa Temperature Module (TM0002) Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical electronic thermometer Regulatory Class: Class II Product Code: FLL Dated: August 28, 2024 Received: August 28, 2024
Dear Jou An Hsu:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
David Walloschek
David Wolloscheck, Ph.D. Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K240220
Device Name Aulisa Temperature Module (TM0002)
Indications for Use (Describe)
Aulisa Temperature Module (TM002) is a battery-operated electronic device with indication for use in continuously measuring and monitoring armpit body temperature of adults, pediatrics, and transmission of the measuring result via wireless signal.
Aulisa Temperature Module (TM0002) is a non-invasive and reusable device for single patient use with intended environments of use are hospitals, medical facilities, home care, and subacute environments.
The parameters derived by the Aulisa Temperature Module (TM0002) are transmitted to a commercially available mobile device that runs an Aulisa-developed application for display and review.
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### K240220 - 510(k) Summary
This 510(k) Summary is being submitted in accordance with the requirements of 21 CFR 807.92.
#### 5.1 General Information
| Date of Preparation | September 27, 2024 |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company Identification | Taiwan Aulisa Medical Devices Technologies,<br>Inc.<br>6F-2, No. 3-1, YuanQu St., Nangang Dist.,<br>Taipei City 115 TW<br>TEL.: +886-2-2655-7297<br>FAX: +886-2-2655-7260 |
| Contact Person | Ann Hsu<br>Regulatory Affairs Specialist<br>Taiwan Aulisa Medical Devices Technologies,<br>Inc.<br>Email: ann.hsu@aulisa.com |
#### 5.2 Trade/Device Name
Aulisa Temperature Module (TM0002)
#### 5.3 Regulatory Information
Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Common Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL
## 5.4 Predicate Device(s)
#### Predicate Device
- K181013, Fever Scout™ Continuous Monitoring Thermometer, VivaLnk Inc. ●
#### Reference Device
- K203208, Guardian Angel Rx GA2000 Series Digital Vital Sign Monitoring System (specifically model GA2002), Taiwan Aulisa Medical Devices Technologies, Inc.
{5}------------------------------------------------
#### 5.5 Device Description
The Aulisa Temperature Module (TM0002) is indicated for continuous armpit body temperature monitoring of adults, pediatrics, and infants which measures body temperature and sends the physiological data to an Aulisa-developed software application as cleared through previous Premarket Notification 510(k) submissions under Guardian Angel Rx GA2000 Series Digital Vital Sign Monitoring System (K203208) using Bluetooth technology. If the physiological data sent from the subject device to the software application falls outside of pre-set limits or when a technical error is detected, both auditory and visual alarm signals are generated through the software application to alert the caregiver.
The sensor module uses Bluetooth technology to interact with a commercial, thirdparty mobile device whether under iOS or Android operating system, such as iPad, iPhone, Google Pixel ... etc. which is not part of this submission. The vital-signs are transmitted to mentioned mobile device that runs Aulisa-developed application for display and review. The software application can be downloaded from an official APP store, such as iOS APP store or Google Play.
The Subject Device is reusable, and its intended environments of use are hospitals, medical facilities, home care, and subacute environments.
The device model, device name, population and parameters of the Subject Device are described below:
| Device Model | Aulisa Sensor Module(s) | Population | Parameters |
|--------------|---------------------------|------------------------------------|------------------|
| TM0002 | Aulisa Temperature Module | • Adult<br>• Pediatric<br>• Infant | Body Temperature |
#### 5.6 Indications for use
Aulisa Temperature Module (TM0002) is a battery-operated electronic device with indication for use in continuously measuring and monitoring armpit body temperature of adults, pediatrics, and infants and transmission of the measuring result via wireless signal. Aulisa Temperature Module (TM0002) is a non-invasive and reusable device for single patient use with intended environments of use are hospitals, medical facilities, home care, and subacute environments.
The parameters derived by the Aulisa Temperature Module (TM0002) are transmitted to a commercially available mobile device that runs an Aulisa-developed application for display and review.
#### 5.7 Comparison with Predicate Device
{6}------------------------------------------------
The Subject Device has similar intended use and technological characteristics to the Predicate device, Fever Scout™ Continuous Monitoring Thermometer (K181013), except that the Predicate device is powered by Rechargeable Lithium Battery, but the Subject Device is powered by non-Rechargeable Coin Battery.
The Subject Device is used together with pre-installed Aulisa-developed software application which has been cleared through previous Premarket Notification 510(k) submission under Guardian Angel Rx GA2000 Series Digital Vital Sign Monitoring System (K203208), thereafter referred to as the Reference device in this submission.
These differences do not affect the safety and effectiveness of the device when used as labeled. Table 5.7.1 and 5.7.2 list the similarities and differences between the Subject Device and Predicate device (K181013), along with Reference device (K203208, specifically model GA2002).
{7}------------------------------------------------
# Table 5.7.1 Comparison Table to Predicate
| | Subject Device | Predicate | Comparison |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------|
| Device name | Aulisa Temperature Module<br>(TM0002) | Fever Scout TM Continuous<br>Monitoring Thermometer | -- |
| K number | -- | K181013 | -- |
| Product code | FLL | FLL | Identical |
| Regulation<br>Number | 21 CFR 880.2910 | 21 CFR 880.2910 | Identical |
| Regulation<br>description | Clinical Electronic Thermometer | Clinical Electronic Thermometer | Identical |
| Classification | II | II | Identical |
| Indications<br>for use | Aulisa Temperature Module<br>(TM0002) is a battery-operated<br>electronic device with indication<br>for use in continuously measuring<br>and monitoring armpit body<br>temperature of adults, pediatrics,<br>and infants and transmission of<br>the measuring result via wireless<br>signal.<br>Aulisa Temperature Module<br>(TM0002) is a non-invasive and<br>reusable device for single patient<br>use with intended environments of<br>use are hospitals, medical<br>facilities, home care, and subacute<br>environments.<br>The parameters derived by the<br>Aulisa Temperature Module<br>(TM0002) are transmitted to a | The wireless Fever ScoutTM<br>Continuous Monitoring<br>thermometer is a non-invasive and<br>re-usable electronic device for<br>home use and a non-invasive and<br>single patient use in the hospital.<br>This product is intended for non-<br>urgent ambulatory continuous<br>armpit body temperature<br>monitoring from ages 29 days and<br>older. | Different<br>Note 1 |
{8}------------------------------------------------
| Section 5 | 510(k) Summary |
|------------------------------------------------------------------------------------------------------------------|----------------|
| commercially available mobile<br>device that runs an Aulisa-<br>developed application for<br>display and review. | |
{9}------------------------------------------------
| | Subject Device | Predicate | Comparison |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Patient<br>population | adult, pediatric, infant | ages 29 days and older | Different<br>Note 2 |
| Use<br>environment | hospitals, medical facilities, home<br>care, subacute environments | home, hospital | Different<br>Note 3 |
| Dimension | 24mm x 32.8mm x 5.5mm | 61mm x 41mm x 5.5mm | Different<br>Note 4 |
| Measurement | body temperature | body temperature | Identical |
| Disposable<br>or reusable | Reusable (Temperature Module)<br>disposable (Patch) | reusable/ disposable | Identical |
| Wireless<br>communication | Bluetooth 5.2 | Bluetooth BLE<br>Wireless (Wi-Fi)<br>2.4G | Different<br>Note 5 |
| Operating<br>platform | Mobile device with pre-<br>installed Aulisa-developed<br>application<br>• iOS 13.0 or later<br>• Android 6.0 or later | • iPhone 5S+ or later & iOS 8.0<br>or later<br>• Android 4.3 or later | Identical<br>Both devices operate on a platform with compatible<br>application and indicate a software operating system<br>has been used instead of a hardware. |
| | Subject Device | Predicate | Comparison |
| Alarm | Visual and auditory alarms | NA | Different<br>Note 6 |
| Measurement<br>accuracy | Under 37°C:<br>$\pm$ 0.2°C<br>37.1~39.0°C:<br>$\pm$ 0.1°C<br>39°C and above: $\pm$ 0.2°C | 98.6°F-102.2°F:<br>$\pm$ 0.2°F (37°C-39°C:<br>$\pm$ 0.1°C)<br>95°F-98.6°F:<br>$\pm$ 0.4°F (35°C-<br>37°C: $\pm$ 0.2°C)<br>102.2°F-107.6°F:<br>$\pm$ 0.4°F (39°C-42°C:<br>$\pm$ 0.2°C) | Different<br>Note 7 |
| Display range | 89.6°F-107.6°F (32°C-42°C)<br><89.6°F (32.0°C) displays "Lo"<br>>107.6°F (42.0°C) displays "Hi" | 95°F-<br>107.6°F<br>(35°C-<br>42°C) | Different<br>Note 8 |
| Operation<br>mode | Continuous | Continuous | Identical |
| Measuring site | Axillary (armpit) | Axillary (armpit) | Identical |
{10}------------------------------------------------
{11}------------------------------------------------
| | Subject Device | Predicate | Comparison |
|--------------|----------------|-----------------------------------------|---------------------|
| Power supply | CR1216 | MS Lithium rechargeable battery<br>3.0V | Different<br>Note 9 |
| | | | |
#### Note 1
Both devices function as a thermometer which have the same indication for use for body temperature monitoring, except that the patient population of Subject Device is "adults, pediatrics, and infants" while Predicate Device is "from ages 29 days and older". Clinical accuracy testing was conducted to ISO 80601-2-56. Therefore, the difference does not raise different questions of safety and effectiveness.
#### Note 2
The patient population of subject device is "adults, pediatrics, and infants" while the predicate device is "from ages 29 days and older". Clinical accuracy testing was conducted to ISO 80601-2-56. Therefore, the difference does not raise different questions of safety and effectiveness.
#### Note 3
The use environment of subject device is "hospitals, medical facilities, home care, subacute environments" whereas the predicate device is limited to "home, and hospital" only. The Subject Device has undergone testing per IEC 60601-1:2005 + A1:2012 + A2:2020, FCC Part 15 Subpart C§ 15.249. Additionally, performance data comprising software verification, cybersecurity analysis, wireless coexistence testing, and usability testing demonstrate that this difference does not raise different questions of safety and effectiveness.
#### Note 4
The dimension of Subject Device is smaller than the Predicate Device. However, the difference does not raise different questions of safety and effectiveness.
#### Note 5
Both devices transmit the measuring result via wireless signal, except that Subject Device sends out data through Bluetooth only while Predicate sends out data through both Bluetooth and Wi-fi. The Subject Device has undergone testing per IEC 60601-1:2005 + A 1:2012 + A2:2020, FCC Part 15 Subpart C \$ 15.249. Additionally, performance data comprising software verification, cybersecurity analysis, wireless coexistence testing, and usability testing demonstrate that this difference does not raise different questions of safety and effectiveness.
{12}------------------------------------------------
#### Note 6
The predicate device does not have an alarm system, whereas the subject device includes "Visual and auditory alarms" integrated with a pre-installed application that is identical to that of the reference device. This difference does not raise new questions of safety or effectiveness, as the alarm system has been previously validated in the reference device.
## Note 7
Both devices have the same "Measurement accuracy between 37.1°C-39°C with +0.1°C". The subject device has Measurement accuracy under 37°C with ±0.2°C, while the predicate device has Measurement accuracy between 35°C-37°C and 39°C-42°C with ±0.2°C. Clinical accuracy testing was conducted per ISO 80601-2-56. Therefore, the difference does not raise different questions of safety and effectiveness.
### Note 8
The subject device has the Display range between 32°C-42°C and have "Hi. Lo" display while out of range. The predicate device has the same Display range between 35°C-42°C but does not have "Hi, Lo" display while out of range. Performance testing was conducted to ISO 80601-2-56. Therefore, the difference does not raise different questions of safety and effectiveness.
### Note 9
Both devices are battery powered, while the only difference being that the subject device uses a Coin Battery, while the predicate uses a Rechargeable Lithium Battery. The subject device underwent testing in accordance with IEC 60601-1:2005 + A1:2012 + A2:2020, and the results demonstrated compliance with safety requirements. Therefore, this difference does not impact the safety and effectiveness of the device.
#### 5.8 Summarv of Non-clinical Testing
The results of the testing demonstrate equivalence with the predicate device(s) and compliance with recognized standards. The table below summarizes test results for the Subject Device, which met the relevant requirements of the applicable recognized standards or claimed specifications.
| Item | Reference | Result |
|-------------------------------------------|-------------------------------|--------|
| Electrical Safety | IEC 60601-1<br>IEC 60601-1-11 | Pass |
| Temperature and Humidity | IEC 60601-1<br>IEC 60601-1-11 | Pass |
| Atmospheric Pressure<br>(Altitude) | IEC 60601-1 | Pass |
| Electromagnetic Immunity<br>and Emissions | IEC 60601-1-2 | Pass |
{13}------------------------------------------------
| Performance | ISO 80601-2-56<br>ASTM E1112 | Pass |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|
| Mechanical Durability | IEC 60601-1<br>IEC 60601-1-11 | Pass |
| Biocompatibility | ISO 10993-5 | Pass |
| • In Vitro Cytotoxicity | ISO 10993-10 | |
| • Skin Sensitization | | |
| • Skin Irritation | | |
| Wireless Coexistence Testing | ANSI C63.27 | Pass |
| FDA Guidance | | |
| Software V&V | Guidance for the Content of Premarket<br>Submissions for Device Software Functions | Pass |
| Security Testing | Cybersecurity in Medical Devices, Quality<br>System Considerations and Content of<br>Premarket Submissions | Pass |
| Usability testing | Applying Human Factors and Usability<br>Engineering to Medical Devices | Pass |
| Clinical electronic<br>thermometer | Enforcement Policy for Clinical Electronic<br>Thermometers<br>Guidance on the Content of Premarket<br>Notification [510(K)] Submissions for<br>Clinical Electronic Thermometers | Pass |
| Manufacturer's specifications | | |
| Bench testing | Durability and Performance testing<br>Battery Life Testing | Pass |
{14}------------------------------------------------
#### 5.9 Summary of Clinical Testing
The accuracy of the Subject Device was validated in accordance with ISO 80601-2-56 and ASTM E1112. The output temperature of the Subject Device is derived directly from input signal without any adjustment, and therefore no clinical validation is needed.
#### 5.10 Conclusion
Based on the testing summarized in the 510(k) submission, the results demonstrate that the subject device is substantially equivalent to the predicate devices. The differences do not raise any new questions of safety or effectiveness when compared to its predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.