The VELYS Robotic-Assisted Solution is intended for stereotaxic surgery to aid the surgeon in identifying the relative position and orientation of anatomical structures, planning the position of the femoral and tibial implant components intraoperatively, and preparing the bones during unicompartmental knee arthroplasty. When indicated for unicompartmental knee arthroplasty, the VELYS Robotic-Assisted Solution is indicated for use with the SIGMA HP Partial Knee Implants and INTUITION Instruments for SIGMA HP Partial Knee and their cleared indications for use.
Device Story
Image-free robotic-assisted surgery system for unicompartmental knee arthroplasty (UKA). Uses optical tracking to acquire intraoperative anatomical landmarks; surgeon plans femoral/tibial implant positions based on preferred technique. System includes robotic arm mounted on OR bed rail; provides real-time compensation for patient leg movement during bone resection. Surgeon manually manipulates saw handpiece within planned resection planes; system constrains handpiece orientation. Operated by surgeon in OR via touchscreen and footswitch. Stores case reports for surgeon review. Benefits include precise bone preparation and alignment during knee arthroplasty.
Clinical Evidence
Bench testing only. No clinical data required. Testing included sagittal border guard evaluation, dynamic compensation performance, cadaveric resection accuracy and soft tissue study, summative usability evaluation, and design validation.
Technological Characteristics
Image-free robotic-assisted system; optical tracking; robotic arm mounted on OR bed rail; touchscreen/footswitch interface. Software-controlled bone resection guidance. Complies with 21 CFR Part 820 design controls.
Indications for Use
Indicated for patients undergoing unicompartmental knee arthroplasty (UKA) requiring stereotaxic surgical assistance for anatomical landmark identification, implant component planning, and bone preparation. Used in conjunction with SIGMA HP Partial Knee Implants and INTUITION Instruments.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 5, 2024
DePuy Ireland UC Camille Chua Regulatory Affairs Manager Loughbeg Ringaskiddy Ireland
Re: K240211
Trade/Device Name: VELYS Robotic-Assisted Solution Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic instrument Regulatory Class: Class II Product Code: OLO Dated: May 6, 2024 Received: May 6, 2024
Dear Camille Chua:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely. Shumaya Ali -S Shumaya Ali, M.P.H. Assistant Director
DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
Submission Number (if known)
K240211
Device Name
VELYS Robotic-Assisted Solution
#### Indications for Use (Describe)
The VELYS Robotic-Assisted Solution is intended for stereotaxic surgery to aid the surgeon in identifying the relative position and orientation of anatomical structures, planning the position of the femoral and tibial implant components intraoperatively, and preparing the bones during unicompartmental knee arthroplasty.
When indicated for unicompartmental knee arthroplasty, the VELYS Robotic-Assisted Solution is indicated for use with the SIGMA HP Partial Knee Implants and INTUITION Instruments for SIGMA HP Partial Knee and their cleared indications for use.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) #: K240211
510(k) Summary
Prepared on: 2024-06-05
### Contact Details
### 21 CFR 807.92(a)(1)
| Applicant Name | | DePuy Ireland UC | | | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------|-----------------------------------|------------------------------|---------------------|--------------|
| Applicant Address | | Loughbeg Ringaskiddy Ireland | | | |
| Applicant Contact Telephone | | +1 949-795-6191 | | | |
| Applicant Contact | | Ms. Camille Chua | | | |
| Applicant Contact Email | | vchua1@its.jnj.com | | | |
| Correspondent Name | | | DePuy Ireland UC | | |
| Correspondent Address | | | Loughbeg Ringaskiddy Ireland | | |
| Correspondent Contact Telephone | | | +1 949-795-6191 | | |
| Correspondent Contact | | | Ms. Camille Chua | | |
| Correspondent Contact Email | | | vchua1@its.jnj.com | | |
| Device Name | | | | 21 CFR 807.92(a)(2) | |
| Device Trade Name | | VELYS Robotic-Assisted Solution | | | |
| Common Name | | Stereotaxic instrument | | | |
| Classification Name | | Orthopedic Stereotaxic Instrument | | | |
| Regulation Number | | 882.4560 | | | |
| Product Code(s) | | OLO | | | |
| Legally Marketed Predicate Devices<br>21 CFR 807.92(a)(3) | | | | | |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | | | | Product Code |
| K172301 | Mako Partial Knee Application | | | | OLO |
| K233227 | VELYS Robotic-Assisted Solution | | | | OLO |
| Device Description Summary | | | | 21 CFR 807.92(a)(4) | |
| The VELYS™ Robotic-Assisted Solution is an image-free robotic-assisted surgery system. It is intended for stereotaxic surgery to aid the<br>surgeon in identifying the relative position and orientation of anatomical structures and landmarks, planning the position of the femoral<br>and tibial implant components intraoperatively, and preparing the bones during total knee arthroplasty (TKA) and unicompartmental<br>knee arthroplasty (UKA). | | | | | |
The image-free system uses a dedicated optical tracking device to acquire anatomical landmarks intra-operatively. These landmarks are then used to plan the femoral and tibial implant locations based on the surgeon's preferred surgical technique and placement preferences. Following the planning step, the VELYS™ Robotic-Assisted Solution helps the surgeon to execute the bone preparation
{4}------------------------------------------------
#### K240211 510(k) #:
according to the plan.
The system includes a Robotic-Assisted Device that constrion and orientation of the saw handpiece and blade inside each resection plane on the patient's femur and tibia. The surgeon actuates and manipulates the saw handpiece, within the planned resection plane, to execute the bone resection. This is analogous to using manual instruments in TKA or UKA. If the patient's leg moves during the resection, the Robotic-Assisted Device compensates for such movement in real-time.
The Robotic-Assisted Device is assembled with a Robotic-Assisted Device arm, mounted on the Operating Room (OR) bed rail, for a minimal footprint
The VELYS® Robotic-Assisted Solution incorporates several software subsystems, including applications responsible for general operation of the system and clinical applications dedicated to the surgery workflow.
The users interact with the clinical applications via a touchscreen and footswitch to navigate through the surgery steps. Case Reports including key surgical procedure information are stored on the system and can be retrieved by the surgeon for future use. Cases Reports including PHI are only available to the surgeon who performed the procedure.
# Intended Use/Indications for Use
The VELYS Robotic-Assisted Solution is intended for stereotaxic surgery to aid the surgeon in identifying the relative position and orientation of anatomical structures, planning the femoral and tibial implant components intraoperatively, and preparing the bones during unicompartmental knee arthroplasty.
When indicated for unicompartmental knee arthroplasty, the VELYS Robotic-Assisted Solution is indicated for use with the SIGMA HP Partial Knee Implants and INTUITION Instruments for SIGMA HP Partial Knee and their cleared indications for use.
# Indications for Use Comparison
The subject device has the same indications for use as the predicate device.
# Technological Comparison
The subject device and the predicate device have similar technological characteristics with respect to principles of operation, system design, and performance. The subject and predict device achieve the same outcomes, however the main design differences of the subject device include: image-free planning and patient anatomy assessment, robotic-assisted bone resection, and surgeon control. Performance testing on the subject device and comparison to the reference device, VELYS™ Robotic-Assisted Solution with TKA, support that there are no new questions of safety or effectiveness.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
Bench testing was performed on the subject device, VELYS™ Robotic-Assisted Solution for Unicompartmental Knee Arthroplasty (UKA), in accordance with the product risk analysis and product requirements, in line with FDA 21 CFR Part 820, and to demonstrate substantial equivalence with the predicate device. Where available, FDA-consensus standards and/or guidance documents were used. New bench testing specific to the expanded indication to include UKA and leveraged testing from the VELYS™ Robetic-Assisted Solution Total Knee Arthroplasty (TKA) reference device were performed for assessing device performance.
Performance testing activity included
- · Sagittal Border Guard Evaluation
- · Dynamic Compensation Performance Test
- · Cadaveric Resection Accuracy and Soft Tissue Study
- · Summative Usability Evaluation
- · Design Validation
No clinical data was necessary to support the determination of substantial equivalence
Both the subject device and predicate device have intended use, same indications for use, and similar technological characteristics. Where there are technological differences, performance testing was conducted on the subject device and VELYS™ Robotic-Assisted Solution with TKA was utilized as a reference device to show that no new questions of safety or effectiveness were raised; therefore DePuy Synthes concludes that the subject and predicate devices are substantially equivalent.
# 21 CFR 807.92(a)(5)
### 21 CFR 807.92(a)(6)
## 21 CFR 807.92(a)(5)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.