K240194 · Riverpoint Medical · MBI · Feb 21, 2024 · Orthopedic
Device Facts
Record ID
K240194
Device Name
ProCinch, QuadCinch
Applicant
Riverpoint Medical
Product Code
MBI · Orthopedic
Decision Date
Feb 21, 2024
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ProCinch, QuadCinch Adjustable Loop Device is intended for use in the fixation of bone and soft tissue in orthopedic procedures requiring ligament or tendon reconstruction.
Device Story
ProCinch/QuadCinch is an adjustable button loop system for orthopedic ligament or tendon reconstruction. Device consists of an ultra-high molecular weight polyethylene (UHMWPE) loop and a titanium (Ti-6Al-4V ELI) plate. Used in hospital/clinic/surgical settings by surgeons. The system facilitates fixation by passing the titanium plate through a femoral channel and securing the loop. Some configurations come pre-assembled; others require intraoperative assembly. Additional sutures (UHMWPE or polyester) assist in assembly and passage. Device is sterile, single-use, and ethylene oxide sterilized. Provides mechanical fixation to stabilize tissue during healing.
Clinical Evidence
No clinical data. Bench testing included cyclic mechanical testing to verify fixation strength, sterilization adoption validation, biocompatibility per ISO 10993-1:2018, packaging validation per ISO 11607-1:2006, and usability engineering validation in cadaveric models per EN62366:2015. LAL and rabbit pyrogenicity testing performed.
Technological Characteristics
Materials: UHMWPE loop, titanium plate (Ti-6Al-4V ELI per ASTM F136). Principle: Manual bone/soft tissue fixation. Form factor: Adjustable loop (15-60mm). Sterilization: Ethylene oxide (EtO). Connectivity: None. Software: None.
Indications for Use
Indicated for fixation of bone and soft tissue in orthopedic procedures requiring ligament or tendon reconstruction.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Riverpoint Medical OrthoButton AL Adjustable Button Loop (K171060, K230212)
Submission Summary (Full Text)
{0}------------------------------------------------
February 21, 2024
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Riverpoint Medical Paulie Vagts Senior Regulatory Affairs Associate III 825 NE 25th Ave Portland, Oregon 97232
Re: K240194
Trade/Device Name: ProCinch, QuadCinch Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: MBI Dated: January 24, 2024 Received: January 24, 2024
Dear Paulie Vagts:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Sara S. Thompson -S
For
Jesse Muir, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
K240194
Device Name
ProCinch, QuadCinch
Indications for Use (Describe)
The ProCinch, QuadCinch Adjustable Loop Device is intended for use in the fixation of bone and soft tissue in orthopedic procedures requiring ligament or tendon reconstruction.
Type of Use (Select one or both, as applicable)
> | Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for Riverpoint. The logo consists of the letters "RP" in a green square, followed by the word "RIVERPOINT" in green, sans-serif font. The logo is simple and modern, and the green color suggests that the company is environmentally friendly.
# 510(k) SUMMARY K240194 ProCinch, QuadCinch Adjustable Button Loop
### Submitter Information
| Submitter's Name: | Riverpoint Medical |
|-------------------------|-----------------------------------------------------------------|
| Address: | 825 NE 25th Ave.<br>Portland, OR 97232 |
| Phone Number: | (503) 517-8001 or (971) 288-1083 |
| Fax Number: | (503) 517-8002 |
| Registration Number: | 3006981798 |
| Contact Person: | Paul Vagts<br>(503) 517-8001 |
| Date of Preparation: | February 14th, 2024 |
| Trade Name: | ProCinch, QuadCinch |
| Common or Usual Names: | Suture Retention Device, Button Loop |
| Classification Name: | Fastener, Fixation, Non-Degradable, Soft Tissue |
| FDA Class: | II |
| Product Classification: | 888.3040: Smooth or threaded metallic bone<br>fixation fastener |
| Classification Code: | MBI |
#### Predicate Device
K202399 - Riverpoint Medical OrthoButton FL
#### Device Description
The Riverpoint Medical ProCinch, QuadCinch adjustable button loop is comprised of an ultra-high molecular weight polyethylene (UHMWPE) loop provided with or without a titanium (Ti-6Al-4V ELI per ASTM F136) plate. Additional non-absorbable sutures consisting of UHMWPE, or polyester are looped through the titanium plate to aide in assembly of the device and passing the plate through the intended void. The UHMWPE is available undyed (white) or with trace filaments of blue or green color suture.
For configurations that do not have a titanium plate preattached, a titanium plate component is affixed to the loop during the procedure. For models that come with a preattached titanium plate, the procedure is the same except the titanium plate is passed through the channel. In both configurations, additional sutures are used to pass the loop and titanium plate (if pre-attached) parallel through the femoral channel and
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for Riverpoint. The logo consists of the letters "RP" in a green square on the left, followed by the word "RIVERPOINT" in green, sans-serif font. The logo is simple and modern, and the green color suggests a connection to nature or sustainability.
secure into place. The UHMWPE is available undyed (white), dyed blue, or with trace filaments of blue or green color suture. The device is sterilized by ethylene oxide gas and is provided sterile for single use. The device is intended for use in a hospital/clinic/surgical setting.
The classification for the ProCinch, QuadCinch is FDA Class II device with product classification 21 CFR §888.3040: Smooth or threaded metallic bone fixation fastener, Product Code MBI.
#### Intended Use / Indications for Use
The ProCinch, QuadCinch is intended for use in the fixation of bone and soft tissue in orthopedic procedures requiring ligament or tendon reconstruction.
#### Performance Data
The sutures used to construct the ProCinch. QuadCinch meet requirements established by the United States Pharmacopeia (USP), except for diameter. The UHMWPE, polyester, or polyblend sutures are tested per USP performance requirements for tensile strength. Non-clinical performance testing for the ProCinch, QuadCinch included a sterilization adoption validation, biocompatibility testing per ISO10993-1:2018 -Biological Evaluation of Medical Devices, product packaging validation per ISO 11607-1:2006 - Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems, usability engineering validation with simulated use in a cadaveric models performed per EN62366: 2015-Medical devices - Application of usability engineering to medical devices. LAL and rabbit pyrogencity testing have demonstrated that the ProCinch, QuadCinch does not raise any additional concerns regarding pyrogenicity. Non-clinical mechanical testing was performed to verify the fixation strength of the ProCinch, QuadCinch using cyclic testing as compared to the predicate device. Results of performance testing for the ProCinch, QuadCinch device concluded that the device performed comparably to the predicate device and the validations performed demonstrated that the ProCinch, QuadCinch met all requirements for its intended use.
## Substantial Equivalence and Comparison of Technical Characteristics
As reflected in the following substantial equivalence table, the ProCinch, QuadCinch has the same intended use and indications for use, the same principles of operation, and similar technical characteristics as the predicate device. Both the ProCinch, QuadCinch and the predicate device are sterilized using the same processes, are composed of the same materials, and are tested per the same performance requirements.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Riverpoint. The logo consists of the letters RP in a green rounded square. To the right of the square is the word RIVERPOINT in green, bold, sans-serif font.
| Device | Riverpoint Medical<br>ProCinch, QuadCinch<br>Adjustable Button Loop | Riverpoint Medical<br>OrthoButton AL Adjustable<br>Button Loop<br>(K171060, K230212) | Comparison |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Intended Use /<br>Indications for Use | The Riverpoint Medical<br>ProCinch, QuadCinch is intended<br>for use in the fixation of bone and<br>soft tissue in orthopedic<br>procedures requiring ligament or<br>tendon reconstruction. | The Riverpoint Medical<br>OrthoButton AL is intended for<br>use in the fixation of bone and<br>soft tissue in orthopedic<br>procedures requiring ligament or<br>tendon reconstruction. | Identical |
| Loop Material | UHMWPE/Polyester | UHMWPE/Polyester | Identical |
| Additional Sutures | UHMWPE, polyester | UHMWPE, nylon, polyester,<br>polypropylene | Included |
| Color Additive | 21CFR74 | 21CFR74 and 21CFR73 | Included |
| Button Material | Titanium (Ti-6AI-4V ELI per<br>ASTM F136) | Titanium (Ti-6AI-4V ELI per<br>ASTM F136) | Identical |
| Finished Loop Length | 15-60mm | 15-60mm | Identical |
| Regulation Number | 888.3040 | 888.3040 | Identical |
| Regulatory Class | II | II | Identical |
| Prescription or OTC | Prescription | Prescription | Identical |
| Product Code | MBI | MBI | Identical |
| Principle of Operation | Manual fixation | Manual fixation | Identical |
| Packaging | Tyvek/Poly Pouch | Tyvek/Poly Pouch | Identical |
| Biocompatibility | Biocompatible per ISO 10993 | Biocompatible per ISO 10993 | Identical |
| Sterilization | EtO Sterilization | EtO Sterilization | Identical |
| Shelf Life | 5 Years | 5 Years | Identical |
The minor difference in technical characteristics is limited to 1) the subject device is provided with the adjustable loop fully assembled whereas the predicate device is assembled intraoperatively and 2) new packaging insert. These minor differences do not raise new questions of safety or effectiveness; therefore, the ProCinch, QuadCinch is substantially equivalent to the currently marketed OrthoButton FL predicate device.
#### Conclusion
The information provided in this Special 510(k) demonstrates that the Riverpoint Medical ProCinch, QuadCinch is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.