K240192 · Scivita Medical Technology Co., Ltd. · LQR · Oct 15, 2024 · Gastroenterology, Urology
Device Facts
Record ID
K240192
Device Name
Single-use Extraction Baskets
Applicant
Scivita Medical Technology Co., Ltd.
Product Code
LQR · Gastroenterology, Urology
Decision Date
Oct 15, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5010
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Single-use Extraction Baskets is intended for the endoscopic removal of stones and stone fragments in the biliary system.
Device Story
Single-use, sterile, disposable biliary extraction basket; consists of basket, sheath, and handle. Basket constructed of stainless steel/nitinol wires. Operated endoscopically by physician; handle activation opens/closes basket to capture stones/fragments. Used in clinical setting for biliary stone removal. Not compatible with mechanical lithotripters. Benefits patient by facilitating minimally invasive stone extraction.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including dimensions, tensile pull, simulated-use functionality/durability, stone capture, and deflection testing. Biocompatibility per ISO 10993-1; sterilization per ISO 11135:2014; shelf life per ASTM F1980-21.
Technological Characteristics
Materials: stainless steel/nitinol wires. Form factor: catheter-based basket with handle. Energy source: manual mechanical actuation. Sterilization: ethylene oxide (ISO 11135:2014). Shelf life: accelerated aging (ASTM F1980-21).
Indications for Use
Indicated for endoscopic removal of stones and stone fragments in the biliary system in patients requiring such intervention.
Regulatory Classification
Identification
A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.
Special Controls
*Classification.* Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services - USA seal. To the right of the seal is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on top of the word "ADMINISTRATION".
October 15, 2024
Scivita Medical Technology Co., Ltd. Dan Jiang Senior Regulatory Affairs Specialist No. 2, Qingqiu Street, Suzhou Industrial Park Suzhou, Jiangsu 215000 China
Re: K240192/A001
Trade/Device Name: Single-use Extraction Baskets (EBF-1529S, EBF-1529C, EBF-1511C, EBF-1511S, EBF-1506S, EBF-1506C, EBF-1507S, EBF-1507C, EBF-1508S, EBF-1508C, EBF-2029S, EBF-2029C, EBF-2006S, EBF-2006C, EBF-2007S, EBF-2007C, EBF-2008S, EBF-2008C) Regulation Number: 21 CFR 876.5010 Regulation Name: Biliary catheter and accessories Regulatory Class: Class II Product Code: LOR Dated: October 11, 2024 Received: October 11, 2024
Dear Dan Jiang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
U.S Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov
{1}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product: and 21 CFR 820.100. Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical
{2}------------------------------------------------
devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Anthony Lee -S
Anthony Lee, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity, and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
Submission Number (if known)
#### K240192
#### Device Name
Single-use Extraction Baskets (EBF-1529S, EBF-1529C, EBF-1511C, EBF-1511S, EBF-1506S, EBF-1506C, EBF-1507S, EBF-1507C, EBF-1508S, EBF-1508C, EBF-2029S, EBF-2029C, EBF-2006S, EBF-2006C, EBF-2007S, EBF-2007C, EBF-2008S, EBF-2008C)
Indications for Use (Describe)
The Single-use Extraction Baskets is intended for the endoscopic removal of stones and stone fragments in the biliary system.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## Classification
Add a primary product code and any associated product codes below. You may type in the primary product code directly (only the product code field is required) or you may filter down by choosing first a medical specialty, regulation, then product code. If a device specific quidance is available for the ? product code, the guidance name and web link will be displayed. Use the Product Classification Website resource in the help text to obtain information about your product code and check the regulation text for any special controls that need to be considered (e.g, PAE and 21 CFR 890.3450).
| Medical Specialty | Gastroenterology & Urology |
|-------------------|---------------------------------------------|
| Regulation | 876.5010 - Biliary catheter and accessories |
| Product Code | LQR (Class 2) - Dislodger, Stone, Biliary |
The primary product code of your device indicates a device specific guidance document is available to aid you in preparing a comprehensive submission. The document entitled "Guidance for the Content of Premarket Notifications for Metal Expandable Biliary Stents: Final" is available at the link below. If you have any questions about applicability of this guidance, please contact the CDRH review Office.
https://www.fda.gov/regulatory-information/search-fda-guidance-documents/metal-expandable-biliarystents-premarket-notification-510k-submissions
Associated Product Code(s)
?
{5}------------------------------------------------
..
| 510(k) #: | K240192 | | K240192 - Page 1 of 2 |
|---------------------------------|----------------------------------------------------------------------------------------------------------------|----------------|-------------------------|
| | | 510(k) Summary | Prepared on: 2024-10-11 |
| Contact Details | | | 21 CFR 807.92(a)(1) |
| Applicant Name | Scivita Medical Technology Co., Ltd. | | |
| Applicant Address | No. 2, Qingqiu Street, Suzhou Industrial Park, 215000 Suzhou, Jiangsu<br>Prov., P.R. CHINA Suzhou 215000 China | | |
| Applicant Contact Telephone | +86-512-8187778 | | |
| Applicant Contact | Mrs. Wu Ruqin | | |
| Applicant Contact Email | wuruqin@scivitamedical.com | | |
| Correspondent Name | Scivita Medical Technology Co., Ltd. | | |
| Correspondent Address | No. 2, Qingqiu Street, Suzhou Industrial Park, 215000 Suzhou,<br>Jiangsu Prov., P.R. CHINA Suzhou 215000 China | | |
| Correspondent Contact Telephone | 86-512-81877788 | | |
| Correspondent Contact | Dan Jiang | | |
| Correspondent Contact Email | ra@scivitamedical.com | | |
Device Name
21 CFR 807.92(a)(2)
| Device Trade Name | Single-use Extraction Baskets (EBF-1529S, EBF-1529C, EBF-1511C,<br>EBF-1511S, EBF-1506S, EBF-1506C, EBF-1507S, EBF-1507C,<br>EBF-1508S, EBF-1508C, EBF-2029S, EBF-2029C, EBF-2006S,<br>EBF-2006C, EBF-2007S, EBF-2007C, EBF-2008S, EBF-2008C) |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name | Biliary catheter and accessories |
| Classification Name | Dislodger, Stone, Biliary |
| Regulation Number | 876.5010 |
| Product Code(s) | LQR |
Legally Marketed Predicate Devices
21 CFR 807.92(a)(3)
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
|-------------|----------------------------------------------------------|--------------|
| K203322 | SpyGlass™ Discover Retrieval Basket | LQR |
Device Description Summary
21 CFR 807.92(a)(4)
The Single-use Extraction Baskets is a single-use, sterile, disposable device. It is intended for the endoscopic removal of stones and stone
fragments in the biliary system. The device consists of basket, sheath, and handle components. The basket component is constructed of
stainless steel/nitinol wires. And the basket is connected to the handle of the operating portion through the insertion portion. Activating
the handle allows the basket to open or close. This device is not compatible with any mechanical lithotripter.## Intended Use/Indications for Use
21 CFR 807.92(a)(5)
{6}------------------------------------------------
The Single-use Extraction Baskets is intended for the endoscopic removal of stones and stone fragments in the billary system.
## Indications for Use Comparison
The Single-use Extraction Baskets has the same intended use as the predicate device. Both the Single-use Extraction Baskets and predicate device are intended for the endoscopic removal of stone fragments in the biliary system.
## Technological Comparison
The subject Single-use Extraction Baskets and predicate device (K203322, SpyGlass™ Discover Retrieval Basket) share the similar technological characteristics. The differences in technological characteristics do not raise different questions of safety and/or effectiveness when compared to the predicate device.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
The following performance data were provided in support of the substantial equivalence determination. The test results demonstrated that the subject device complies with the standard requirements.
Biocompatibility testing
Biocompatibility testing is performed in accordance, "Use of International Standard ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" including cytotoxicity, sensitization, and irritation testing.
Performance testing including sterilization validation, shelf life, and mechanical performance/bench to demonstrate the basic performance of the subject device and confirmed that the subject device performs as intended.
Sterilization validation
Sterilization validation was carried out with Methods Half-cycle approach in accordance with ISO 11135:2014.
Shelf life testing
Shelf life testing was conducted based on an accelerated aging test in accordance with ASTM F1980-21, the standard guide for accelerated aging of sterile barrier systems for medical devices. Three years real-time aging test will be performed to demonstrate longer stability and support the results of the accelerated aging test.
Mechanical performance testing
Mechanical performance testing was conducted on the following items to demonstrate the basic performance of the subject device and confirmed that the subject device performs as intended.
- · Dimensions Testing
- · Tensile Pull Testing
- · Simulated-Use Functionality & Durability Testing
- · Stone Capture Testing
- · Deflection Testing
Comparative testing was conducted on the Single-use Extraction Baskets and the predicate SpyGlass Discover Retrieval Basket (K203322).
The clinical data is not applicable.
The nonclinical test were conducted to demonstrate that the subject is as safe and effective as the predicate.
## 21 CFR 807.92(a)(6)
21 CFR 807.92(a)(6)
21 CFR 807.92(a)(5)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.