Aneroid Sphygmomanometer with Stethoscope; Aneroid Sphygmomanometer
K240165 · Wenzhou Longwan Medical Device Factory · DXQ · Mar 20, 2024 · Cardiovascular
Device Facts
Record ID
K240165
Device Name
Aneroid Sphygmomanometer with Stethoscope; Aneroid Sphygmomanometer
Applicant
Wenzhou Longwan Medical Device Factory
Product Code
DXQ · Cardiovascular
Decision Date
Mar 20, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1120
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The device is intended to be used by medical professionals or in the home for the measurement of systolic and diastolic pressure on new borns, infants, children, young adults. The device is intended to be manually attached to a patient and manually inflated along with a manual method for detecting Korotkoff sounds.
Device Story
Manual blood pressure monitor; consists of inflatable cuff, inflation bulb, pressure gauge, air valve, and optional stethoscope. Operator wraps cuff around patient's upper arm or leg; manually inflates using bulb; listens for Korotkoff sounds via stethoscope while manually controlling deflation via valve. Pressure gauge displays systolic and diastolic readings in mmHg. Used in hospitals, clinics, ambulances, or home settings by medical professionals or patients. Provides visual pressure data to assist healthcare providers in clinical decision-making regarding patient cardiovascular status.
Clinical Evidence
No clinical data. Bench testing only. Compliance with ISO 81060-1:2007 for non-automated sphygmomanometers and biocompatibility testing per ISO 10993-5:2009, ISO 10993-10:2021, and ISO 10993-23:2021.
Indicated for measurement of systolic and diastolic blood pressure in newborns, infants, children, and young adults. Intended for use by medical professionals or in home settings.
Regulatory Classification
Identification
A blood pressure cuff is a device that has an inflatable bladder in an inelastic sleeve (cuff) with a mechanism for inflating and deflating the bladder. The cuff is used in conjunction with another device to determine a subject's blood pressure.
Predicate Devices
Aneroid sphygmomanometer, Aneroid sphygmomanometer with stethoscope (K211084)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 20, 2024
Wenzhou Longwan Medical Device Factory Bernard Zeng, Manager Unit 4, Building 27, No. 706 Yanyun Road, Oujiangkou Industrial Cluster District Wenzhou. Zhejiang 325026 China
Re: K240165
Trade/Device Name: Aneroid Sphygmomanometer with Stethoscope; Aneroid Sphygmomanometer Regulation Number: 21 CFR 870.1120 Regulation Name: Blood Pressure Cuff Regulatory Class: Class II Product Code: DXO, LDE Dated: January 15, 2024 Received: January 22, 2024
Dear Bernard Zeng:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Stephen C. Browning -S
LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known)
K240165
Device Name
## ANEROID SPHYGMOMANOMETER WITH STETHOSCOPE; ANEROID SPHYGMOMANOMETER
#### Indications for Use (Describe)
The device is intended to be used by medical professionals or in the home for the measurement of systolic and diastolic pressure on new borns, infants, children, young adults. The device is intended to be manually attached to a patient and manually inflated along with a manual method for detecting Korotkoff sounds.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(K) Summary
Prepared in accordance with the requirements of 21 CFR Part 807.92
#### Prepared Date: Dec. 11,2023
## 1. Submitter's Information
The submitter of this pre-market notification is:
| Name: | WENZHOU LONGWAN MEDICAL DEVICE FACTORY |
|-----------------|----------------------------------------------------------------------------------------------------------------------|
| Address: | Unit 4, Building 27, No. 706 Yanyun Road, Oujiangkou Industrial<br>Cluster District, Wenzhou, Zhejiang, 325026 China |
| Contact person: | Bernard Zeng |
| Title: | Manager |
| E-mail: | bernard@taiyumed.com |
| Tel: | 86-13957715667 |
## 2. Device Identification
| 510(K) number: | TBD |
|--------------------|-------------------------------------------|
| Trade/Device Name: | ANEROID SPHYGMOMANOMETER, |
| | ANEROID SPHYGMOMANOMETER WITH STETHOSCOPE |
| Common name: | Blood Pressure Cuff |
| Regulation Number: | 870.1120 |
| Regulation Name: | Blood Pressure Cuff |
| Regulation Class: | Class 2 |
| Panel: | Cardiovascular |
| Product Code: | DXQ,LDE |
## 3. Predicate Device
| 510(K) number: | K211084 |
|--------------------|-------------------------------------------|
| Trade/Device Name: | Aneroid sphygmomanometer, |
| | Aneroid sphygmomanometer with stethoscope |
| Common name: | Blood Pressure Cuff |
| Regulation Number: | 870.1120 |
| Regulation Name: | Blood Pressure Cuff |
| Regulation Class: | Class 2 |
| Panel: | Cardiovascular |
| Product Code: | DXQ,LDE |
{5}------------------------------------------------
Traditional 510(k) Submission of ANEROID SPHYGMOMANOMETER
#### 4. Indication for Use
The device is intended to be used by medical professionals or in the home for the measurement of systolic and diastolic pressure on new borns, infants, children, young adults. The device is intended to be manually attached to a patient and manually inflated along with a manual method for detecting Korotkoff sounds.
#### 5. Device Description
The ANEROID SPHYGMOMANOMETER and ANEROID SPHYGMOMANOMETER WITH STETHOSCOPE are manual blood pressure monitor used to measure blood presure.It consists of a cuff,a inflation bulb,a pressure gauge, an air valve and a stethoscope(optional).
Cuff: The cuff is made of a durable fabric and is wrapped around the upper arm of the patient. It inflates and deflates during the measurement process. The material of outside layer is nylon cloth.
Inflation Bulb: The inflation bulb is a rubber bulb that connects to the cuff and is used to manually inflate the cuff with air. The material is Neoprene or Silicon rubber or Nature latex or PVC.
Pressure Gauge: The pressure gauge is a circular dial with a needle that indicates the pressure within the cuff. It is calibrated in millimeters of mercury (mmHg) and shows the systolic and diastolic blood pressure readings. The material is Aluminum or ABS.
Air Valve: The valve is a control mechanism that allows for the release of air from the cuff once the measurement is complete.
Stethoscope: Although not a part of the device itself, a stethoscope is typically used in conjunction with an Aneroid Sphygmomanometer to listen for the blood flow sounds, known as Korotkoff sounds, in the brachial artery during the measurement. Three model are single head Stethoscope(Model: TY-S01),Dual Head Stethoscope(Model: TY-S05) and Spraque Rappaport Stethoscope(Model: TY-S04) individually.These stethoscopes are listed in FDA database.
#### 6. Compared to Predicate Device
Compared to the predicate devices, the subject device has the same intended use, similar product design, similar performance, same safety as the predicate device, the summarized comparison information is listed in the following table
| SE<br>Comparisons | Subject Device | Predicate Device | Similarities/<br>Differences |
|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|
| 510(K) number | TBD | K211084 | / |
| Name | ANEROID<br>SPHYGMOMANOMETER,<br>ANEROID<br>SPHYGMOMANOMETER<br>WITH STETHOSCOPE | Aneroid sphygmomanometer,<br>Aneroid sphygmomanometer<br>with stethoscope | Same |
| Indication for<br>Use | The device is intended to be<br>used by medical professionals | The device is intended to be<br>used by medical professionals | Same |
| | or in the home for the<br>measurement of systolic and<br>diastolic pressure on new<br>borns, infants, children, young<br>adults and adults. The device<br>is intended to be manually<br>attached to a patient and<br>manually inflated along with a | or in the home for the<br>measurement of systolic and<br>diastolic pressure on new borns,<br>infants, children, young adults<br>and adults. The device is<br>intended to be manually<br>attached to a patient and<br>manually inflated along with a | |
| | manual method for detecting<br>Korotkoff sounds. | manual method for detecting<br>Korotkoff sounds. | |
| Product code | DXQ,LDE | DXQ,LDE | Same |
| Class | II | II | Same |
| Regulation<br>Number | 21 CFR 870.1120 | 21 CFR 870.1120 | Same |
| Common name | Blood Pressure Cuff | Blood Pressure Cuff | Same |
| Type of Use | Over the Counter | Over the Counter | Same |
| Environment of<br>Use | Home, Hospital, healthcare<br>facility,ambulance etc. | Home, Hospital, healthcare<br>facility, ambulance etc. | Same |
| Target<br>population | New borns, infants, children,<br>young adults and adults | New borns, infants, children,<br>young adults and adults | Same |
| Site | Upper Arm (leg for child) | Upper Arm (leg for child) | Same |
| Measurement<br>Method | Auscultatory Korotkoff sounds<br>Method | Auscultatory Korotkoff sounds<br>Method | Same |
| Inflation | Manual by inflation bulb | Manual by inflation bulb | Same |
| Deflation | Manual deflation via valve | Manual deflation via valve | Same |
| Display | Aneroid Manometer | Aneroid Manometer | Same |
| Monitor scale | From 0 to 300 mmHg with a<br>minimum interval of 2 mmHg. | From 0 to 300 mmHg with a<br>minimum interval of 2 mmHg. | Same |
| Design of the<br>device | The device comprises tubing<br>attached to a soft inelastic cuff<br>with an integrated inflatable<br>bladder that is wrapped around<br>the patient's limb and secured<br>by hook and loop closure. | The device comprises tubing<br>attached to a soft inelastic cuff<br>with an integrated inflatable<br>bladder that is wrapped around<br>the patient's limb and secured<br>by hook and loop closure. | Same |
| Design of<br>Stethoscope | Three types option:<br>Single head<br>Dual head<br>Sprague Rappaport | Three types option:<br>Single head<br>Dual head<br>Sprague Rappaport | Same |
| Materials | The manometer: aluminum or<br>ABS<br>The tubing, inflation bulb and<br>cuff bladder: Neoprene or<br>Silicon rubber or Nature latex<br>or PVC<br>Cuff: Nylon cloth for outside<br>layer | The manometer: aluminum or<br>ABS<br>The tubing, inflation bulb and<br>cuff bladder: Neoprene or<br>Silicon rubber or Nature latex or<br>PVC<br>Cuff: Nylon cloth for outside<br>layer | Same |
| Accuracy | ± 3 mmHg | ± 3 mmHg | Same |
| Cuff sizes | 500 ×140mm<br>350 ×100mm<br>600 ×170mm<br>780 × 220mm<br>260 × 80mm<br>200 × 60mm<br>255 × 75mm<br>380 × 110mm<br>560 × 150mm<br>660 × 170mm<br>760 × 210mm | 800mm × 220mm<br>620mm × 175mm<br>520mm × 140mm<br>340mm × 110mm<br>250mm × 80mm<br>210mm × 60mm | Different,<br>See Note 1 |
| Cuff<br>circumference | Fits arm circumferences<br>60mm-660mm, the standard<br>cuff should be available for use<br>in measuring a New born,<br>infant, child's leg blood<br>pressure and for child, young<br>adult, and adult with larger<br>arms | Fits arm circumferences<br>100mm-620mm, the standard<br>cuff should be available for use<br>in measuring a New born, infant,<br>child's leg blood pressure and<br>for child, younq adult, and adult<br>with larger arms | Different,<br>See Note 1 |
| Cuff bladder<br>size | 220 × 120mm<br>175 × 85mm<br>315 × 145mm<br>375 × 185mm<br>110 × 60mm<br>80 × 40mm<br>115 × 55mm<br>175 × 80mm<br>215 × 120mm<br>310 × 140mm<br>340 × 170mm | 340mm × 170mm<br>310mm × 140mm<br>220mm × 120mm<br>170mm × 80mm<br>110mm × 60mm<br>80mm × 40mm | Different.<br>See Note 1 |
| Biocompatibility | ISO 10993-5:2009,<br>ISO 10993-10:2021,<br>ISO 10993-23:2021 | ISO 10993-5:2009,<br>ISO 10993-10:2010 | Different,<br>See Note 2 |
| Performance | ISO 81060-1:2007 | ISO 81060-1:2007 | Same |
{6}------------------------------------------------
#### Traditional 510(k) Submission of ANEROID SPHYGMOMANOMETER
{7}------------------------------------------------
#### Traditional 510(k) Submission of ANEROID SPHYGMOMANOMETER
Note 1: All of the cuff were tested alongside aneroid sphygmomanometer according to ISO 81060-1:2007, these differences do not affect safety and effectiveness.
Note 2: The skin irritation testing and skin sensitization testing were conducted in accordance with ISO 10993-23:2021 and ISO 10993-10:2021 due to the version update of the standard, this difference does not affect safety and effectiveness.
The subject device and predicate device are substantially equivalent in the areas of technological characteristics such as basic design, features, method of operation, general function,
{8}------------------------------------------------
Traditional 510(k) Submission of ANEROID SPHYGMOMANOMETER
application, and intended use. The subject device does not raise any new potential safety risks and is equivalent in performance to the existing legally marketed devices.
## 8. Performance Testing Summary
#### The subject device comply with:
#### Clinical test:
Clinical testing is not required.
#### Non-clinical data
#### Safety and performance
ISO 81060-1:2007 Non-invasive sphygmomanometers - Part 1: Requirements and test methods for non-automated measurement type
#### Biocompatibility
ISO 10993-5:2009 Biological Evaluation of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity ISO 10993-10:2021 Biological Evaluation of Medical Devices - Part 10: Tests for Skin Sensitization ISO 10993-23:2021 Biological Evaluation of Medical Devices - Part 23: Tests for Irritation
#### 9. Conclusion
The conclusions drawn from the nonclinical tests demonstrate that the subject devices ANEROID SPHYGMOMANOMETER,ANEROID SPHYGMOMANOMETER WITH STETHOSCOPE are as safe, as effective, and performs as well as the legally marketed predicated devices Aneroid sphygmomanometer,Aneroid sphygmomanometer with stethoscope(K211084).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.