K240164 · Bain Medical Equipment (Guangzhou) Co., Ltd. · FIB · Jun 28, 2024 · Gastroenterology, Urology
Device Facts
Record ID
K240164
Device Name
NovaLine SP-C35 Transducer Protector (956007)
Applicant
Bain Medical Equipment (Guangzhou) Co., Ltd.
Product Code
FIB · Gastroenterology, Urology
Decision Date
Jun 28, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5820
Device Class
Class 2
Indications for Use
The NovaLine SP-C35 Transducer Protector is a single-use, disposable, prescription device intended for use as a protective device for pressure monitors and to help prevent contamination of the fluid pathway in hemodialysis delivery systems.
Device Story
Single-use, disposable transducer protector; plastic housing with hydrophobic 0.2 µm PTFE composite membrane filter; male/female luer lock connectors. Installed between blood tubing set and hemodialysis machine pressure monitor. Principle: hydrophobic filter allows air passage for pressure monitoring while blocking liquid, bacteria, viruses, and particulates from entering the machine. Used in clinical hemodialysis settings; operated by trained healthcare professionals. Protects machine from fluid contamination; maintains sterility of blood tubing pathway. Benefits: prevents cross-contamination and protects sensitive pressure monitoring equipment.
Clinical Evidence
No clinical data. Bench testing only. Performance verified via structural integrity, dimensional testing (ISO 80369-7), and microbial challenge testing (ASTM F1671, ASTM F838). Biocompatibility confirmed via ISO 10993 series (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogen, complement activity, PTT, platelet count, hematology).
Technological Characteristics
Housing: Polyethylene/Polyvinylchloride; Filter: PTFE composite membrane (0.2 µm pore size). Dimensions conform to ISO 80369-7:2021. Pressure limits: ±500 mmHg. Sterilization: E-beam (SAL 10^-6). No software or electronics.
Indications for Use
Indicated for use as a protective device for pressure monitors and to prevent fluid pathway contamination in hemodialysis delivery systems. Prescription use only.
Regulatory Classification
Identification
A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo for the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
June 28, 2024
Bain Medical Equipment (Guangzhou) Co., Ltd. Zoe Zeng Regulatory Supervisor No.10 Juncheng Road, No.10, Juncheng Road, Eastern Area, Economic and Technological Development Guangzhou, Guangdong 510760 China
Re: K240164
Trade/Device Name: NovaLine SP-C35 Transducer Protector Regulation Number: 21 CFR§ 876.5820 Regulation Name: Hemodialysis system and accessories Regulatory Class: II Product Code: FIB Dated: May 31, 2024 Received: May 31, 2024
Dear Zoe Zeng:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
3
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Maura Rooney -S
Maura Rooney Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known) K240164
Device Name NovaLine SP-C35 Transducer Protector
7.
#### Indications for Use (Describe)
The NovaLine SP-C35 Transducer Protector is a single-use, disposable, prescription device intended for use as a protective device for pressure monitors and to help prevent contamination of the fluid pathway in hemodialysis delivery systems.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
ﻧﻬﺎﻳﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
న్న
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
{4}------------------------------------------------
## Tab #29 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.
The assigned 510(k) Number: K240164
- 1. Date of Preparation: 05/31/2024
- 2. Sponsor Identification
### Bain Medical Equipment (Guangzhou) Co., Ltd.
No.10, Juncheng Road, Eastern Area, Economic and Technological Development District, 510760, Guangdong, P.R.China
Establishment Registration Number: 3013217742
Contact Person: Sophia Shao Position: Management Representative Tel: +86-20-66856868 ext. 218 Fax: +86-20-32067500 Email: sophia@baingz.com
#### 3. Designated Submission Correspondent
### Bain Medical Equipment (Guangzhou) Co., Ltd
No.10, Juncheng Road, Eastern Area, Economic and Technological Development District, 510760, Guangdong, P.R.China
Ms. Zoe Zeng (Primary Contact Person) Tel: +86-20-66856868 ext.290 Fax: +86 20 3206 7500 Email: zoe@baingz.com
Ms. Sophia Shao (Alternative Contact Person) Tel: +86-20-66856868 ext.218 Fax: +86 20 3206 7500 Email: sophia@baingz.com
{5}------------------------------------------------
Trade Name: NovaLine SP-C35 Transducer Protector Common Name: Transducer Protector Models: 956007
### Regulatory Information
Regulation Name: Hemodialysis system and accessories Classification: II Product Code: FIB Regulation Number: CFR 876.5820 Review Panel: Gastroenterology/Urology
### Indications for Use:
The NovaLine SP-C35 Transducer Protector is a single-use, disposable, prescription device intended for use as a protective device for pressure monitors and to help prevent contamination of the fluid pathway in hemodialysis delivery systems.
### Device Description
The NovaLine SP-C35 Transducer Protector consists of a plastic housing that contains a hydrophobic filter with a reference pore size of 0.2 µ m. The housing has male and female luer lock connectors that allow its attachment between a blood tubing set and the pressure monitor on a hemodialysis machine.
The NovaLine SP-C35 Transducer Protector is to be used as protective device for pressure monitors as well as to help maintain the sterility of the blood tubing fluid pathway. The 0.2 ↓ m hydrophobic filter helps prevent cross-contamination by viruses, bacterial and particulate matter while preventing the flow of fluids to the hemodialysis machine pressure monitor.
The devices are packaged sterile and labeled for single use only.
The proposed devices are provided in sterile condition, it is subject to e-beam radiation sterilization prior to release to achieve a Sterility Assurance Level (SAL) of 10-6.
| Components | Material | Type of Contact | Contact Duration | Applicable standard |
|----------------------------|-------------------|------------------------------------------------------|------------------|---------------------|
| Cap | Polyethylene | Surface device (skin) | <24 hours | ISO 10993 |
| Female luer lock connector | Polyvinylchloride | External communication device (blood path, indirect) | <24 hours | ISO 10993 |
| Male luer lock | Polyvinylchloride | External communication device (blood path, indirect) | <24 hours | ISO 10993 |
{6}------------------------------------------------
| Filter paper | PTFE composite<br>membrane | External communication device<br>(blood path, indirect) | <24 hours | ISO 10993 |
|--------------|----------------------------|---------------------------------------------------------|-----------|-----------|
|--------------|----------------------------|---------------------------------------------------------|-----------|-----------|
#### ട. Identification of Predicate Device
### Predicate Device
510(k) Number: K072988
Product Name: Nipro Transducer Protector TP-SURE Manufacturer: NIPRO MEDICAL CORPORATION
#### 6. Non-Clinical Test Conclusion
Non-clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device.
### Finished Product Performance Test
The proposed devices were tested with appearance, structural integrity, Dimension(ISO 80369-7:2021) and sterility and the test results demonstrated that the proposed device complies with the standards requirements. A side-by-side parallel test with the predicate device demonstrate that the proposed device performs equivalent to the predicate device and is safe and effective when used as intended. A human factors assessment was conducted to demonstrate its safe and effective use.
### Product Packaging
Tests were performed on sterilized product packages according to ASTM F88/F88-15 and ASTM F1929-15 to support compliance with pre-established acceptance criteria.
### Simulated Distribution Test
Tests were conducted according to ASTM D4169-22 to support compliance with pre-established acceptance criteria.
### Shelf life
Packaging performance(Seal Strength and Dye Penetration) and Performance Tests were conducted, including tests per ISO 80369-7. The test resulted demonstrated that the accelerated aged samples complied with the pre-established acceptance criteria.
### Bacterial Endotoxin Testing
Endotoxin limit is established as 20 USP Endotoxin Unit (EU) for the proposed device. Test has been conducted by Limulus Amebocyte Lysate (LAL) method according to USP <85> Bacterial Endotoxin Limit.
### Biocompatibility Tests
The following tests were conducted to support the biological safety of the proposed device: cytotoxicity(ISO10993-5), sensitization(ISO 10993-10), irritation(ISO 10993-23), systemic
{7}------------------------------------------------
toxicity(ISO10993-11), pyrogen(ISO10993-11), complement activity (ISO10993-4), partial thromboplastin time(ISO 10993-4), platelet count(ISO 10993-4) and hematology(ISO 10993-4).
Microbial Challenge Testing
Tests were conducted to support the sterile barrier according to ASTM F1671/F1671M-2022, YYT1497-2016 and ASTM F838-2020.
Electrical Safety and EMC Not applicable.
Software Verification and Validation Testing Not applicable.
Animal Studies No animal studies were performed.
#### 7. Clinical Test Conclusion
No clinical study is included in this submission.
#### Substantially Equivalent (SE) Comparison 8.
| Item | Proposed Device | Predicate Device<br>K072988 |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Model | 956007 | / |
| Device Class | II | II |
| Product Code | FIB | FIB |
| Regulation Number | 21CFR part 876.5820 | 21CFR part 876.5820 |
| Indications for Use | The NovaLine SP-C35 Transducer Protector<br>is a single-use, disposable, prescription device<br>intended for use as a protective device for<br>pressure monitors and to help prevent<br>contamination of the fluid pathway in<br>hemodialysis delivery systems. | The Nipro Transducer Protector<br>TP-SURE is a single-use,<br>disposable, prescription device<br>intended for use as a protective<br>device for pressure monitors<br>and to help prevent<br>contamination of the fluid<br>pathway in hemodialysis<br>delivery systems. |
| Feature | Sterile<br>Non-pyrogenic<br>Single Use<br>Prescription Device | Sterile<br>Non-pyrogenic<br>Single Use<br>Prescription Device |
| Main Configuration | Filter Paper<br>Male Luer Lock Connector | Filter Paper<br>Male Luer Lock Connector |
{8}------------------------------------------------
| Item | | Proposed Device | Predicate Device<br>K072988 |
|-------------------------|----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | Cap | Cap |
| Physical<br>performance | Positive<br>pressure<br>Limitation<br>(mmHg) | 500 | 500 |
| | Negative<br>Pressure<br>Limitation<br>(mmHg) | -500 | -500 |
| Performance | | Conforms to<br>ISO 80369-7:2016 | Conforms to<br>ISO 80369-7:2016 |
| Materials | | PVC, PTFE, PET, PE | PP, PTFE, PET |
| Biocompatibility | | Cytotoxicity;<br>Skin Sensitization;<br>Intradermal Reactivity;<br>Acute Systemic Toxicity;<br>Material-mediated Pyrogens;<br>In Vitro Hemolysis;<br>Bacterial endotoxin testing;<br>Complement Activity;<br>Partial Thromboplastin Time;<br>Platelet Count;<br>Hematology | Cytotoxicity;<br>Pyrogenicity<br>Acute Toxicity;<br>Intracutaneous Reactivity;<br>Hemolysis Testing;<br>Implantation Testing;<br>Bacterial Endotoxin Testing |
| Sterilization | | Ebeam, SAL(10-6) | EO, SAL(10-6) |
| Labeling | | Direction for Use<br>Intended Use<br>Description<br>Warnings and Cautions | Direction for Use<br>Intended Use<br>Description<br>Warnings and Cautions |
The proposed device has the similar configuration, indication for use and safety feature as the predicated device. The non-clinical testing demonstrates the product performance of proposed device is similar as that of the predicate device or the product performance of proposed device is acceptable. The biocompatibility of the proposed device comply with ISO 10993 series standards.
#### 9. Substantially Equivalent (SE) Conclusion
Based on the comparison and analysis above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.