BD Prevue™ II Peripheral Vascular Access System

K240146 · Bard Access Systems, Inc. · IYO · Feb 18, 2024 · Radiology

Device Facts

Record IDK240146
Device NameBD Prevue™ II Peripheral Vascular Access System
ApplicantBard Access Systems, Inc.
Product CodeIYO · Radiology
Decision DateFeb 18, 2024
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 892.1560
Device ClassClass 2
AttributesPediatric

Indications for Use

BD Prevue™ II Peripheral Vascular Access System is indicated for Vascular Access Imaging Applications performed by appropriately trained healthcare professionals in a medical setting. Typical examinations performed using the BD Prevue™ II Peripheral Vascular Access System include: Vascular [Exam Type (Adult and Pediatric)]: Assessment of vessels in the extremities and neck leading to or coming from the heart, superficial veins in the arms and legs, and vessel mapping. Assessment of superficial thoracic vessels. Vascular Access [Exam Type (Adult and Pediatric)]: Guidance for PICC, CVC, dialysis catheter, port, PIV, midline, arterial line placement, access to fistula and general vein and artery access in adult and pediatric patients, including those patients with difficult intravascular access who may have small, fragile, and/or non-palpable vessels.

Device Story

Portable 2D ultrasound system with integrated Cue™ Needle Tracking System; used by trained healthcare professionals in medical settings for vascular access guidance. System inputs include ultrasound signals and passive magnetic fields emitted from magnetized needles detected by sensors in specialized ultrasound probes. Software processes sensor data to generate a virtual needle image overlaid on the real-time ultrasound display, showing needle position, trajectory, and intersection window. Provides clinicians with visual guidance during needle insertion; facilitates accurate vessel access, particularly in patients with difficult intravascular access (small, fragile, or non-palpable vessels). Includes procedure documentation, vessel measurement tools, and electronic connectivity.

Clinical Evidence

No clinical data. Bench testing conducted for the predicate device (K211193) covers the proposed patient population; no additional testing required for this modification.

Technological Characteristics

Portable 2D ultrasound system; includes Cue™ Needle Tracking System using passive magnetic field sensing. Components: Cue™ Traditional/Vascular Access Probes with integrated sensors, Cue™ Magnetizer, and Cue™ enabled needles. Connectivity: electronic data transfer. Software-based needle tracking and visualization. No changes to materials or design from predicate.

Indications for Use

Indicated for vascular and vascular access imaging in adult and pediatric patients, including those with difficult intravascular access (small, fragile, or non-palpable vessels). Used by trained healthcare professionals in medical settings for vessel assessment and guidance for PICC, CVC, dialysis catheter, port, PIV, midline, arterial line, and fistula/graft access.

Regulatory Classification

Identification

An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Special Controls

*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ March 1, 2024 Image /page/0/Picture/1 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. Bard Access Systems, Inc. % Samira Saeed Regulatory Affairs Specialist I 605 N 5600 W SALT LAKE CITY UT 84116 Re: K240146 Trade/Device Name: BD Prevue™ II Peripheral Vascular Access System Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic Pulsed Echo Imaging System Regulatory Class: Class II Product Code: IYO, IYX, LLZ Dear Samira Saeed: The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated February 18, 2014. Specifically, FDA is updating this SE Letter to include the correct IFU statement and 510(k) Summary, as an administrative correction. Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Yanna Kang, OHT8: Office of Radiological Health, (301) 796-6704, Yanna.Kang@fda.hhs.gov. Sincerely, ## Yanna S. Kang -S Yanna Kang, Ph.D. Assistant Director Mammography and Ultrasound Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health {1}------------------------------------------------ February 18, 2024 Image /page/1/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services. To the right of the symbol, there is a blue square with the letters 'FDA' in white. Next to the blue square, the text 'U.S. FOOD & DRUG ADMINISTRATION' is written in blue. Bard Access Systems, Inc. % Samira Saeed Regulatory Affairs Specialist I 605 N 5600 W SALT LAKE CITY UT 84116 Re: K240146 Trade/Device Name: BD Prevue™ II Peripheral Vascular Access System Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic Pulsed Echo Imaging System Regulatory Class: Class II Product Code: IYO, IYX, LLZ Dated: January 18, 2024 Received: January 19, 2024 Dear Samira Saeed: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). {2}------------------------------------------------ Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerelv. ### Yanna S. Kang -S Yanna Kang, Ph.D. Assistant Director Mammograph and Ultrasound Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health {3}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K240146 Device Name BD Prevue™ II Peripheral Vascular Access System Indications for Use (Describe) BD Prevue™ II Peripheral Vascular Access System is indicated for Vascular Access Imaging Applications performed by appropriately trained healthcare professionals in a medical setting. Typical examinations performed using the BD Prevue™ II Peripheral Vascular Access System include: [Table] Imaging Applications: Vascular [Exam Type (Adult and Pediatric)]: Assessment of vessels in the extremities and neck leading to or coming from the heart, superficial veins in the arms and legs, and vessel mapping. Assessment of superficial thoracic vessels. Vascular Access [Exam Type (Adult and Pediatric)]: Guidance for PICC, CVC, dialysis catheter, port, PIV, midline, arterial line placement, access to fistula and general vein and artery access in adult and pediatric patients, including those patients with difficult intravascular access who may have small, fragile, and/or non-palpable vessels. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logos of two companies, BD and Bard. The BD logo is in blue and consists of the letters "BD" next to an orange sun-like symbol. The Bard logo is in green and consists of the word "BARD" in a stylized font, with the text "has joined BD" underneath it in a smaller font. BD Prevue™ II Peripheral Vascular Access System Special 510(k) Premarket Notification # 510(k) Summary 21 CFR 807.92(a) | General<br>Provisions | Submitter Name: | Bard Access Systems, Inc. (Bard has joined BD) | |----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | Submitter Address: | 605 North 5600 West<br>Salt Lake City, Utah 84116 | | | Contact Person: | Samira Saeed<br>Regulatory Affairs Specialist | | | Telephone Number: | 801.522.5000 | | | Date of Preparation: | January 16 2023 | | Subject<br>Device: | Trade Name: | BD Prevue II Peripheral Vascular Access System | | | Common Name: | Ultrasound System with Needle Tracking | | | Regulation Name: | Ultrasonic Pulsed Echo Imaging System<br>Diagnostic Ultrasonic Transducer<br>Picture Archiving and Communication System | | | Regulation Number: | 21 CFR 892.1560<br>21 CFR 892.1570<br>21 CFR 892.2050 | | | Product Code: | IYO<br>ITX<br>LLZ | | | Regulatory Class: | Class II | | | Classification Panel: | Radiology | | Predicate<br>Device | Trade Name: | BD Prevue II Peripheral Vascular Access System | | | Common Name: | Ultrasound System with Needle Tracking | | | Regulation Name: | Ultrasonic Pulsed Echo Imaging System<br>Diagnostic Ultrasonic Transducer<br>Picture Archiving and Communication System | | | Regulation Number: | 21 CFR 892.1560<br>21 CFR 892.1570<br>21 CFR 892.2050 | | | Product Code: | IYO<br>ITX<br>LLZ | | | Regulatory Class: | Class II | | | Classification Panel: | Radiology | | | Premarket Notification Number: | K211193 | | Device<br>Description | BD Prevue™ II Peripheral Vascular Access System Device Description: | | | | The BD Prevue™ II Peripheral Vascular Access System is a portable device that features real-<br>time 2D ultrasound imaging, customized vascular access applications, procedure<br>documentation, vessel measurement tools, and electronic connectivity (if enabled). | | | | Cue™ Needle Tracking System Description: | | | | Imaging Applications | Exam Type (Adult and Pediatric) | | Indications for Use | Vascular | Assessment of vessels in the extremities and neck leading to or<br>coming from the heart, superficial veins in the arms and legs, and<br>vessel mapping. Assessment of superficial thoracic vessels. | | | Vascular Access | Guidance for PICC, CVC, dialysis catheter, port, PIV, midline, arterial<br>line placement, access to fistula and grafts, and general vein and<br>artery access in adult and pediatric patients, including those<br>patients with difficult intravascular access who may have small,<br>fragile, and/or non-palpable vessels. | | Technological<br>Characteristics | Technological characteristics of the subject BD Prevue™ II Peripheral Vascular Access System<br>are substantially equivalent with respect to design and function to those of the cited predicate<br>device. The key modification made to the subject device when compared to the predicate<br>device is the specification of difficult intravascular access (DIVA) patients in the indications for<br>use. As a result, the indications for use in the product instructions for use was updated to<br>reflect the modification. There are no other differences between the subject and predicate<br>device. | | | Performance<br>Tests | As part of Bard Access Systems, Inc.'s design controls, risk analysis was<br>conducted to assess the impact of the proposed subject device indications for use<br>modifications. The results of the risk analysis determined that no verification or validation<br>activities were required because the subject device modifications to the Indications for use and<br>resulting modifications to the instructions for use and labeling do not include any changes to<br>the design, materials, performance, or risk profile of the cited predicate device and the<br>performance testing conducted in the predicate device clearance (K211193) covers the<br>proposed specified difficult intravascular access patient population. Therefore, it is not<br>necessary to conduct additional verification and validation testing. | | {5}------------------------------------------------ Image /page/5/Picture/1 description: The image shows two company logos. The first logo consists of an orange starburst-like symbol to the left of the letters "BD" in blue. The second logo shows the word "BAIRD" in green. Below the word "BAIRD" is the phrase "has joined BD" in a smaller font. BD Prevue™ II Peripheral Vascular Access System Special 510(k) Premarket Notification The BD Prevue™ II Peripheral Vascular Access System is equipped with Cue™ Needle Tracking System which is designed to track and display the location and trajectory of a needle under ultrasound guidance. The technology consists of software installed on an ultrasound system and sensors incorporated into the ultrasound probes. The sensors detect a passive magnetic field emitted from the needle. The software interprets the data from the sensors and creates a virtual image of the needle on the ultrasound display, providing clinicians with a visual representation of the needle during the insertion process. The Cue™ Needle Tracking System requires the use of either the Cue™ Traditional Probe or Cue™ Vascular Access Probe, the Cue™ Magnetizer and a Cue™ enabled needle. The Cue™ Traditional Probe and Vascular Access Probe contain sensors for tracking Cue™ compatible needles (following magnetization by the Cue™ Magnetizer). The tracked needle's current position, trajectory, and intersection window are displayed over the ultrasound image. BD Prevue™ II Peripheral Vascular Access System is indicated for Vascular and Vascular Access Imaging Applications performed by appropriately trained healthcare professionals in a medical setting. Typical examinations performed using the BD Prevue™ II Peripheral Vascular Access System include: {6}------------------------------------------------ . Image /page/6/Picture/1 description: The image shows the logos of two companies, BD and BARD. The BD logo is in blue, and the BARD logo is in green. Underneath the BARD logo, it says "has joined BD" in a smaller font. The BD logo is on the left, and the BARD logo is on the right. BD Prevue™ II Peripheral Vascular Access System Special 510(k) Premarket Notification | Summary of | The modification to the indications for use and resulting modifications to the product | |-------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Substantial | instructions for use has no impact on the intended use, technological characteristics, or risk | | Equivalence | predicate device. Therefore, the subject BD Prevue™ II Peripheral Vascular Access System is<br>substantially equivalent to the predicate BD Prevue™ II Peripheral Vascular Access System. |
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...