K240132 · Emoled, Srl · GEX · Mar 12, 2024 · General, Plastic Surgery
Device Facts
Record ID
K240132
Device Name
EmoLED (v. 2-USA)
Applicant
Emoled, Srl
Product Code
GEX · General, Plastic Surgery
Decision Date
Mar 12, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The EmoLED device is intended to provide photo therapeutic light to the body. The EmoLED device is generally indicated to treat dermatological conditions. The EmoLED device is specifically indicated to emit visible blue/violet light to treat moderate inflammatory acne vulgaris.
Device Story
Portable, battery-powered phototherapy device; emits non-coherent blue light (415 ± 5 nm) via six LEDs. Features polycarbonate body, touchscreen interface, and rotatable optic head with distance sensor. Operator-controlled; projects light at 120 mW/cm2 intensity over 50.3 cm2 area. Distance sensor ensures optimal 4 ± 1 cm skin-to-device spacing; pauses emission if outside 3-5 cm range. Used in clinical settings for targeted treatment of acne vulgaris on face, back, or shoulders. Provides 36 J/cm2 dose over 5-minute treatment time. Benefits patients by delivering localized phototherapeutic energy to affected skin areas.
Clinical Evidence
No clinical testing performed. Substantial equivalence supported by non-clinical bench testing, including photobiological safety (IEC 62471), medical electrical equipment safety (IEC 60601-2-57), and usability engineering (IEC 62366-1, CEI EN 60601-1-6).
Technological Characteristics
Polycarbonate housing; 6 LED light sources (415 ± 5 nm); rechargeable lithium-ion battery; 120 mW/cm2 optical power intensity. Includes distance sensor (3-5 cm range), touchscreen, and optic head rotatable 180°. Standards: IEC 62471, IEC 60601-2-57, IEC 62366-1, CEI EN 60601-1-6, ISTA 3E.
Indications for Use
Indicated for patients with moderate inflammatory acne vulgaris and general dermatological conditions requiring blue/violet light phototherapy.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 12, 2024
Emoled. Srl % Robert Swain Principal Consultant RESound Clinical Consulting LLC 19 Chipping Norton Lane Bedford, New Hampshire 03110
Re: K240132
Trade/Device Name: EmoLED (v. 2-USA) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: January 17, 2024 Received: January 17, 2024
Dear Robert Swain:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/ofdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).
{2}------------------------------------------------
Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Tanisha L. Tanisha L. Hithe
Tanisha L. -s
Michael - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -2024.03.12 Hithe -S 09:56:46 -04'00'
Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K240132
Device Name EmoLED (v. 2-USA)
Indications for Use (Describe)
The EmoLED device is intended to provide photo therapeutic light to the body. The EmoLED device is generally indicated to treat dermatological conditions. The EmoLED device is specifically indicated to emit visible blue/violet light to treat moderate inflammatory acne vulgaris.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
| <span style="font-family: sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: sans-serif;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
Prepared on: 2024-03-07
| Contact Details | 21 CFR 807.92(a)(1) |
|-----------------|---------------------|
|-----------------|---------------------|
| Applicant Name | Emoled, srl |
|---------------------------------|----------------------------------------------------------------------|
| Applicant Address | Via di Rimaggio 141/Interno 25 Sesto Fiorentino Florence 50019 Italy |
| Applicant Contact Telephone | +390550751960 |
| Applicant Contact | Mr. Lorenzo Targetti |
| Applicant Contact Email | I.targetti@emoled.com |
| Correspondent Name | RESound Clinical Consulting LLC |
| Correspondent Address | 19 Chipping Norton Lane Bedford NH 03110 United States |
| Correspondent Contact Telephone | +16032335166 |
| Correspondent Contact | Dr. Robert Swain |
| Correspondent Contact Email | rswain@resoundclinical.com |
| Device Name | 21 CFR 807.92(a)(2) |
|-------------|---------------------|
|-------------|---------------------|
| Device Trade Name | EmoLED (v. 2-USA) |
|---------------------|-------------------------------------------------------------------------------------|
| Common Name | Laser surgical instrument for use in general and plastic surgery and in dermatology |
| Classification Name | Powered Laser Surgical Instrument |
| Regulation Number | 878.4810 |
| Product Code | GEX |
| Legally Marketed Predicate Devices | 21 CFR 807.92(a)(3) |
|------------------------------------|---------------------|
|------------------------------------|---------------------|
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
|-------------|----------------------------------------------------------|--------------|
| K200659 | Dermalux Tri-Wave MD | GEX |
| K232656 | BLU TOTALE | OLP |
| Device Description Summary | 21 CFR 807.92(a)(4) |
|----------------------------|---------------------|
|----------------------------|---------------------|
Emol.ED is a portable, reusable medical device powered by rechargeable lithium-ion batteries that illuminate six (6) LED light sources which emit non-coherent blue light between 415 ± 5 nanometers (nm). The EmoLED device emits a uniform and constant optical power intensity of 120 mW/cm2 over an area of approx. 50.3 cm in diameter) at a distance of 4 +/- 1 cm. The EmoLED device consists of:
a polycarbonate body containing a software-controlled processor that controls the emission of light; .
a touch screen (display), placed in the front part of the device that allows the operator to control the device;
a multi-function (on/off) button, placed in the posterior part of the body of the device;
{5}------------------------------------------------
- a power jack and a micro-USB plug, placed on the left side of the body of the device;
an optic part containing the LED light source, the LED's control board, an optical lens (optic head) and a distance sensor connected to the body which can be rotated by the operator up to 180° to project the light beam on the desired skin surface;
- accessories, which include a visual comfort screen and 24V power supply (with an integrated connection cable).
### Intended Use/Indications for Use
The EmoLED device is intended to provide photo therapeutic light to the body. The EmoLED device is generally indicated to treat dermatological conditions. The EmoLED device is specifically indicated to emit visible blue/violet light to treat moderate inflammatory acne vulgaris.
## Indications for Use Comparison
The subject device is substantially equivalent to the primary predicate device and primary predicate device have similar indications for use and intended users, and use environments, with minor differences that do not introduce new questions for safety and effectiveness.
The Dermalux Tri-Wave MD (the primary provides phototherapeutic light energy from three distinct spectra; red light, blue light, and near infrared. The primary predicate device's indications for use states that each individual spectra is intended to treat specific dermatological conditions:
· The blue light (415nm wavelength) is generally indicated to treat dermatological conditions and specifically indicated to treat moderate inflammatory acne vulgaris;
· The red light (633nm wavelength) is generally indicated for treatment of superficial, benign, vascular and pigmented lesions;
- The near-infrared light (830mm wavelength) is generally indicated for the temporary relieve of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation where applied.
The indications for use for the EmoLED (the subject device) is limited to the blue light region of the visible light spectrum (415nm wavelength), and has the identical indications for use as the primary predicate device's blue light indication:
· The blue light (415nm wavelength) is generally indicated to treat dermatological conditions and specifically indicated to treat moderate inflammatory acne vulgaris.
The primary predicate and subject device have the same indications for use limited to the blue light region and the intended use to generally treat dermatological conditions and specifically treat moderate inflammatory acne vulgaris. The primary predicate device has additional indications for use involving red light (633mm wavelength) and near infrared light (830m wavelength) for the treatment of conditions other than those identified for the blue light (415nm wavelength) indication, and which are independent of the indications for use involving the 415nm wavelength light. The use of the EmoLED device at 415nm wavelength for the same indications as the primary predicate device at the same wavelength does not raise new questions on safety and effectiveness, as the devices have substantially equivalent indications. This rationale should meet the requirements of 21 CFR 807.92(a)(5).
## Technological Comparison
#### 21 CFR 807.92(a)(6)
The Dermalux Tri-Wave MD primary predicate device (K200659) employs a total of 480 LEDs that emit blue light at a wavelength of 415 nm and produces a maximum power density of 40 mW/cm2. In K200659, the Dermalux Tri-Wave MD device was compared to the Phototherapy Systems device, cleared under K190938; the latter device has the same indications for use as the EmoLED and Dermalux Tri-Wave MD device (with relation to the use of blue light at 415nm wavelength). The Phototherapy Systems device blue light LEDs produce a maximum power density of 120 mW/cm2. The EmoLED employs 6 blue light LEDs that produce a maximum power density of 120 mW/cm2, identical to the Phototherapy Systems device. This indicates that based on the identical indications for use and the differential in the maximum power density between the Dermalux Tri-Wave MD and its predicate, the Phototherapy Systems device, there are no new questions of safety and effectiveness raised by this difference in technological characteristic. It is therefore reasonable to conclude that the difference in maximum power density between the EmoLED and the Dermalux Tri-Wave MD raises no new questions of safety and effectiveness, since this same difference failed to raise these questions when the Dermalux Tri-Wave MD was compared to its predicate (Phototherapy Systems).
The effective treatment area of the Dermalux Tri-Wave MD is 792 cm2, where the effective treatment area of the EmoLED device is approximately 50.3cm2, which represents the area of an 8 cm diameter circle. Given the intended use of treating dermatological conditions in general and moderate acne vulgaris specifically, focusing light energy within an 8 cm diameter circle on the face and/or
21 CFR 807 92(a)(5)
{6}------------------------------------------------
back/shoulders is appropriate since illuminating blue light on healthy skin provides no therapeutic benefit, and the subject device is to treat the affected area. In the case of acne vulgaris, which most frequently affects the face and back/shoulders, a treatment area of 50.3 cm2 is appropriate with the intended use even if multiple, overlapping treatments are reguired on the back/ shoulders. The difference in effective treatment area between the subject and primary predicate device does not raise any new questions of safety and effectiveness as the products both achieve therapeutic benefit regardless of the exposure area.
The standard dose of the subject device is 36.0 J/cm2. This is calculated by multiplying the output power (120 mW/cm2) by 300 seconds (5 minutes) of treatment time. This is compared to the primary predicate, which is 48 /cm2 and is calculated by multiplying its output power (40 mW/cm2) by 1200 seconds (20 minutes) of treatment time. For both devices, the adjusted by increasing or decreasing the exposure time of the difference in the standard dosage (36 J/cm2 for the subject device, 48 J/ cm2 for the primary predicate device) does not raise new questions of safety and effectiveness.
The recommended treatment time for the EmoLED device is 5 minutes for the primary predicate device. A secondary predicate, the BLU TOTALE (K232656), is a hand-held OTC blue-light phototherapy device indicated for use in dermatology for the treatment of mild to moderate acne. The recommended treatment time for the BLU TOTALE is 4 minutes per treated area. This treatment time is comparable to the 5-minute treatment time for the 1 minute difference between the recommended treatment times between the subject device and the BLUE TOTALE secondary predicate device fails to raise new questions of safety and effectiveness.
The subject device is designed to deliver uniform, constant light energy to the source is 4 ± 1 cm from the surface of the skin. The distance sensor integrated into the operator establish this optimal distance. If the optical head is placed too close (<3 cm) or too far (>5 cm) from the skin's surface, the light energy will be paused until the optimal distance is re-established. The primary predicate device's minimum distance that the LED light panel should be to the skin surface is 2.5 cm, with no "optimal" distance identified. The minimum distance for the subject device is 3 cm, where the primary predicate's suggested minimum distance is 2.5 cm, both of which are very similar, with the minimum difference raising no issues with safety and effectiveness of the subject device.
The subject device has a much smaller footprint than the primary predicate device with the subject device whereas the predicate device is a console that is wheeled into and out of the subject device allows the user to achieve the indications for use to treat dermatological conditions, in general, and acne vulgaris, specifically, just as the primary predicate device's design does. Therefore, the differences in the devices does not raise any questions of safety and effectiveness.
The subject device is powered using rechargeable lithium-ion batteries; a 24V charger (with integrated cable) is provided. When fully charged the subject device can provide approximately 150 minutes of treatment. All relevant electromagnetic compatibility testing and electrical safety testing has been performed and passed. The primary predicate device is powered by mains power. As both devices use electrical energy to power the devices for their intended use and the subject device has been validated through the applicable testing, there are no new questions of safety and effectiveness introduced through the different types of electrical energy sources for the devices.
While the intended use of the subject and primary predicate device is identical (with respect to blue light), there are notable differences when comparing technological characteristics between the two (e.q., output power, treatment time, footprint, etc.) None of the differences in technological characteristics highlighted, however, raise any new questions of safety and the performance of the subject device, as verified and validated through testing in accordance with relevant standards, has been demonstrated to be substantially equivalent to the primary predicate device.
# Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
Performance testing of the EmoLED device (v. 2) was conducted to verify that the device met all design specifications. The test results demonstrate that the EmoLED device complies with the following testing standards:
IEC 62471 First Edition 2006-07 (FDA #12-249) Photobiological safety of lamps and lamp systems
IEC 60601-2-57 Edition 1.0 2011-01 (FDA #12-242) Medical electrical equipment -- Part 2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic, aesthetic use
IEC 62366-1 Edition 1.1 2020-06 Consolidated Version (FDA #5-129) Medical devices -- Part 1: Application of usability engineering to medical devices
CEI EN 60601-1-6/A 1:2016 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance --Collateral Standard: Usability
ISTA 3E 2017 (FDA #5-128) Similar Packaged-Products in Unitized Loads of Truckload Shipment
No clinical testing has been performed to support this submission.
{7}------------------------------------------------
The EmoLED device was subjected to an identical series of nonclinical tests as its primary predicate device and the test results demonstrated that the EmoLED device performed as safe and as effective as the legally marketed primary predicate device.
ﻳﻮﻡ ﺍﻟﻤﻮﺿﻮﻉ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﺘﻲ ﺗﻘﺮﻳﺒﺎ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮ
1
:
।
.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.