K240127 · Conmed Corporation · FYD · Aug 5, 2024 · General, Plastic Surgery
Device Facts
Record ID
K240127
Device Name
PlumeSafe X5 Smoke Management System
Applicant
Conmed Corporation
Product Code
FYD · General, Plastic Surgery
Decision Date
Aug 5, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5070
Device Class
Class 2
Indications for Use
The CONMED PlumeSafe® X5™ Smoke Management System is designed to remove and filter smoke and aerosols from a surgical site produced during electrosurgical and laser procedures.
Device Story
Portable, stand-alone smoke management system; removes surgical smoke/aerosols from surgical site via vacuum motor. Inputs: surgical smoke drawn through accessories (pencils, tubing, fluid traps). Processing: smoke passes through multi-stage filter (pre-filter, ULPA, coconut-based carbon, impregnated alumina, carbon dust capture). Output: filtered clean air returned to environment. Used in OR/clinical settings by physicians/hospital staff. Features multiple modes (open, laparoscopic, pencil) and flow settings. Provides occlusion detection and motor ramp-up. Benefits: protects healthcare personnel from surgical smoke contamination. Integrates with CONMED MEP-1 Electrosurgical unit.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including filter efficiency (ISO 29463-5), sound pressure (ISO 3744, ISO 16571), electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and software validation (IEC 62304).
Technological Characteristics
Materials: aluminum, steel, plastic, sound-reducing insulation. Sensing: RFID for filter/accessory identification, occlusion detection. Energy: 120V AC. Filtration: multi-stage (pre-filter, ULPA, carbon, alumina). Connectivity: direct docking to CONMED MEP-1. Software: IEC 62304 compliant. Form factor: stackable rectangular console.
Indications for Use
Indicated for physicians and trained hospital staff to remove and filter surgical smoke and aerosols during electrosurgical and laser procedures.
Regulatory Classification
Identification
Air-handling apparatus for a surgical operating room is a device intended to produce a directed, nonturbulent flow of air that has been filtered to remove particulate matter and microorganisms to provide an area free of contaminants to reduce the possibility of infection in the patient.
Special Controls
*Classification.* Class II (special controls). The device, when it is an air handling bench apparatus, an air handling room apparatus, or an air handling enclosure apparatus, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 05, 2024
CONMED Corporation Erin Hoffman Principal Admin of Regulatory Affairs 6455 South Yosemite Street Greenwood Village. CO 80111
Re: K240127
Trade/Device Name: PlumeSafe® X5™ Smoke Management System Regulation Number: 21 CFR 878.5070 Regulation Name: Air-Handling Apparatus For A Surgical Operating Room Regulatory Class: Class II Product Code: FYD Dated: January 15, 2024 Received: July 09, 2024
Dear Erin Hoffman:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Katharine Segars -S
Katharine Segars, Ph.D. Assistant Director DHT4C: Division of Infection Control Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
Submission Number (if known)
K240127
Device Name
PlumeSafe® X5™ Smoke Management System
Indications for Use (Describe)
The CONMED PlumeSafe® X5™ Smoke Management System is designed to remove and filter smoke and aerosols from a surgical site produced during electrosurgical and laser procedures.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary – K240127 PlumeSafe® X5™ Smoke Management System
In accordance with the requirements of the Safe Medical Device Act of 1990 and 21 CFR 807.92, ConMed Corporation is hereby submitting the 510(k) Summary of Safety and Effectiveness for this 510(k).
#### l. SUBMITTER
ConMed Corporation 525 French Road Utica, NY 13502
### Company Contact:
Erin Hoffman Specialist Regulatory Affairs 720-299-6422 erinhoffman@conmed.com
### Date Prepared:
Aug 02, 2024
#### II. DEVICE NAME
Proprietary Name: PlumeSafe® X5™ Smoke Management System
### Model Numbers:
- PX5-120 PlumeSafe® X5™ Smoke Management System ●
- PX5-120NF PlumeSafe® X5™ Smoke Management System, 120V, No Filter ●
- PX5-135 PlumeSafe® X5™ Filter, 1" (25 mm), 3/8″ (9.5 mm), 1″ (25 mm), single pack ●
- PX5-1354 PlumeSafe® X5™ Filter, 1" (25 mm), 3/8″ (9.5 mm), 1″ (25 mm), 4pack .
- PX5-135-22 - PlumeSafe® X5™ Filter, 7/8″ (22 mm), 3/8″ (22 mm), single pack
- . PX5-1354-22 - PlumeSafe® X5™ Filter, 7/8″ (22 mm), 3/8″ (9.5 mm), 7/8″ (22 mm), 4pack
| Common Name: | PlumeSafe® X5™ |
|--------------------|-----------------------------|
| Panel: | General and Plastic Surgery |
| Product Code: | FYD |
| Device Class: | II |
| Regulation Number: | 21 CFR 878.5070 |
{4}------------------------------------------------
#### lll. PREDICATE
| Trade Name: | VisiClear Smoke Evacuation Unit |
|---------------|---------------------------------|
| Company Name: | CONMED |
| 510(k): | K131402 |
#### IV. DEVICE DESCRIPTION
The PlumeSafe® X5™ Smoke Management System is a portable, stand-alone, capital unit, that when paired with a reusable filter and disposable accessories makes up a Smoke Management System. The smoke unit is accompanied by a reusable filter and is available in two different configurations, a 22mm or 25mm size. Each filter is designed to capture components of surgical smoke and return clean air to the environment and can connect to different accessories which include pencils, adapters, fluid traps, and tube sets. Each filter is a seff-contained device that is completely enclosed to protect health care personnels from potential contamination during filter changes. Each filter is comprised of filtration (gross particulate pre-filter, Ultra-Low Penetration Air (ULPA), coconutbased carbon, impregnated alumina, and carbon dust capture media) that produce capture efficiency ratings equal to or better than 99.999% for particles ranging in size from 0.01μm to 0.2μm.
PlumeSafe® X5™ is comprised of a vacuum motor, aluminum, steel, and plastic components combined with sound reducing insulation. The vacuum motor is used to draw the surgical site, through an accessory, and into the PlumeSafe® X5™ filter. The PlumeSafe® X5™ Smoke Management System provides multiple modes (open, laparoscopic, and pencil) and settings to adjust the flow to the clinical need.
### V. INTENDED USE / INDICATIONS FOR USE
The CONMED PlumeSafe® X5™ Smoke Management System is designed to remove and filter smoke and aerosols from a surgical site produced during electrosurgical and laser procedures.
### VI. COMPARISON OF THE TECHNOLOGICAL CHARACTERISTICS
The PlumeSafe® X5™ Smoke Management System is similar to the predicate device in that the design of both includes an identical Intended Use, RFID functionality, automatic activation through the EZ Link clip accessory, material, packaging, accessories, unit housing shape, operational features, and design features. The main difference in the subject device from the predicate device is number of filtration stages, black box collection, and direct docking to the CONMED MEP-1 Electrosurgical device (cleared through K213354). The PlumeSafe® X5™ Smoke Management System is and substantially equivalent to the predicate as demonstrated by non-clinical performance testing listed in section VII.
| | Subject Device (K240127) | Predicate Device (K131402) | Comparison |
|--------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | PlumeSafe® X5™ Smoke<br>Management System | VisiClear Smoke Evacuation System | N/A |
| Manufacturer | CONMED | CONMED | Identical. |
| Regulation<br>Number | 878.5070 | 878.5070 | Identical. |
| Product Code | FYD | FYD | Identical. |
| Device<br>Class/Name | Class II/ Air-handling apparatus for a<br>surgical operating room | Class II/ Air-handling apparatus for a<br>surgical operating room | Identical. |
| Fundamental<br>Scientific<br>Technology | Smoke Management System | Smoke Management System | Identical. |
| Intended<br>Use/Indications for Use | The CONMED PlumeSafe® X5TM<br>Smoke Management System is<br>designed to remove and filter<br>smoke and aerosols from a surgical<br>site produced during electrosurgical<br>and laser procedures. | The VISICLEAR Smoke Evacuation<br>System is designed to remove and<br>filter smoke and aerosols from a<br>surgical site produced during<br>electrosurgical and laser<br>procedures. | Identical. |
| Target<br>Population | For physicians and trained hospital<br>staff during the use of lasers or<br>electrosurgery. | For physicians and trained hospital<br>staff during the use of lasers or<br>electrosurgery. | Identical. |
| Modes | Laparoscopic, Open Tubing,<br>Electrosurgical, Pencil | Laparoscopic, Open Tubing,<br>Electrosurgical Pencil | Identical. |
| Display | Capacitive Touch Screen | LCD | Similar.<br>Display does not impact<br>functionality of the device. |
| Power Source | Standard 100/120V alternating<br>current | Standard 120 Volt | Similar.<br>Power Source does not<br>impact functionality of the<br>device. |
| Unit Housing<br>Shape | Rectangle | Trapezoid | Similar.<br>PlumeSafe® X5TM shape<br>designed to be stacked with<br>the CONMED MEP-1<br>Electrosurgical Unit. The<br>differences in the unit<br>housing shape does not affect<br>the functionality of the<br>device. |
| Motor | 120V motor: 904098 Domel 4.75in<br>(1150-1250 Watts) | 120V motor: 904169 Domel 4.32in<br>(800 Watts) | Similar.<br>The larger motor in the<br>PlumeSafe® X5TM does not<br>affect the functionality of the<br>device. |
| Accessories | Footswitch, electrosurgical smoke<br>pencils, cables, remote activation,<br>fluid trap, cart | Footswitch, electrosurgical smoke<br>pencils, cables, remote activation,<br>fluid trap, cart | Identical. |
| Packaging | Filter is supplied with biohazard bag<br>for proper disposal after useful life<br>is expired. Some filters can be<br>placed inside the unit<br>and then bagged as part of the<br>packaging. | Filter is supplied with biohazard bag<br>for proper disposal after useful life<br>has expired. | Similar.<br>Packaging, including<br>biohazard bag, does not<br>impact functionality of the<br>device. |
| Performance | The ULPA-grade filter is 99.9997%<br>efficiency at 0.01 to 0.20 micron<br>particle size with a filter life of up to<br>35 hours.<br><br>Flow Rate Testing<br><br>In addition, compared to the<br>predicate, CONMED performed the<br>following non-clinical tests:<br>Sound per ISO 16571 Continuous Use Intermittent Use Motor Ramp-up Detection Transportation Reliability | The ULPA-grade filter is 99.999%<br>efficiency at 0.10 to 0.20 micron<br>particle size with a filter life of up<br>to 35 hours.<br><br>Flow Rate Testing | Similar.<br><br>Both include ULPA-grade filter<br>and can accommodate the<br>same filter life. The<br>PlumeSafe® X5TM can filter a<br>smaller particle size.<br>Flow rate testing<br>demonstrated similar flow<br>rates and does not impact<br>functionality.<br><br>Discussion of nonclinical tests<br>performed are provided in<br>table 2 below. |
| Electrical<br>Safety and<br>EMC Testing | Fulfills Electrical Safety<br>requirements per IEC 60601-1.<br>Fulfills Electromagnetic<br>compatibility (EMC) per IEC 60601-<br>1-2. | Fulfills Electrical Safety<br>requirements per IEC 60601-1.<br>Fulfills Electromagnetic<br>compatibility (EMC) per IEC 60601-<br>1-2. | Identical. |
| Software | Software Verification and Validation<br>per IEC 62304. | Software Verification and Validation<br>per IEC 62304. | Identical. |
| Test | Brief Description | Applicable Standards | Results<br>(Pass/Fail) |
| Sound | Verify sound pressure and noise level characteristics per ISO Standards. | ISO 3744<br>ISO16571 | Pass |
| Initial Filter<br>Efficiency | Verify filter configurations maintained the ULPA efficiency rating to remove 99.999% of the most penetrating particle sizes (MPPS) at set flow rate and concentration factor per ISO Standards. | ISO 29463-5:2018 | Pass |
| Final Filter<br>Efficiency | Verify filter configurations maintained the ULPA efficiency rating to remove 99.999% of the most penetrating particle sizes (MPPS) at set flow rate and concentration factor over a period per ISO Standards. | ISO 29463-5:2018 | Pass |
| Continuous Use | Verify continuous operation and continuous surgical plume evacuation. | N/A | Pass |
| Intermittent<br>Use | Verify the device runs intermittently over a specified amount of time without the motor stopping or overheating at a set airflow. | N/A | Pass |
| Motor Ramp-up | Verify the device can reach 90% of maximum suction within 2 seconds of activation when used with a surgical pencil. | N/A | Pass |
| Detection | Verify the device detects occlusion. | N/A | Pass |
| Reliability | Verify lifetime of motor at 350 hours. | N/A | Pass |
| Transportation | Verify the Filter and Console packaging configurations per ASTM D4169. | ASTM D4169 | Pass |
| Flow | Verify airflow is maintained in Open, Pencil, and Laparoscopic modes. | N/A | Pass |
| Software<br>verification and<br>validation | Software testing completed per IEC 62304. | IEC 62304 | Pass |
| Electrical safety | Electrical Safety testing completed per IEC 60601-1 | IEC 60601-1 | Pass |
| Electromagnetic<br>Compatibility | Electromagnetic Capability testing completed per IEC 60601-1-2 (EMC and Immunity) | IEC 60601-1-2 | Pass |
### Table 1 - Comparison of Subject PlumeSafe® X5™ and Predicate VisiClear (K131402) Devices
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
#### VII. SUMMARY OF NONCLINICAL TESTING
Benchtop testing was completed to support substantial equivalence to the predicate as it relates to safety and effectiveness. The following tests were conducted to support substantial equivalence for this submission:
### Table 2: Non-Clinical Performance tests
#### VIII. SUMMARY OF CLINICAL TESTING
N/A
{8}------------------------------------------------
#### IX. CONCLUSION
The subject PlumeSafe® X5™ Smoke Management System is substantially equivalent in design, materials, indications for use, principles of operation and technological characteristics to the predicate VisiClear. The conclusions drawn from the nonclinical performance testing (discussed above) demonstrate that the device is as safe, as effective, and performs as well as or better than the predicate device, VisiClear (K131402).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.