← Product Code [PBF](/productcode/PBF) · K240113

# CORIOGRAPH Knee Pre-Op Plan (K240113)

_Smith & Nephew, Inc. · PBF · Mar 18, 2024 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K240113

## Device Facts

- **Applicant:** Smith & Nephew, Inc.
- **Product Code:** [PBF](/productcode/PBF.md)
- **Decision Date:** Mar 18, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Software as a Medical Device

## Indications for Use

CORIOGRAPH Pre-Op Planning and Modeling Service is intended to provide pre-operative planning for surgical procedures based on patient imaging, provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans.

## Device Story

CORIOGRAPH Knee Pre-Op Plan V1.0 is a software module within the CORIOGRAPH Pre-Op Planning and Modeling Service. It processes patient imaging data to generate 3D bone models and patient-specific surgical plans. Smith & Nephew personnel operate the system to create these plans, which are then reviewed by surgeons. The output consists of a PDF plan and a JSON file containing planning parameters, which serve as inputs for the REAL INTELLIGENCE CORI Surgical System. The device is used in professional healthcare facilities. By providing precise, patient-specific pre-operative planning, the system assists surgeons in implant alignment and positioning, potentially improving surgical outcomes for patients undergoing TKA or UKR.

## Clinical Evidence

Bench testing only. Verification and validation testing confirmed the software applications and generated plans meet design inputs. Summative usability testing in a simulated use environment demonstrated that surgeons could use the device safely and effectively. No clinical data was required.

## Technological Characteristics

Software as a Medical Device (SaMD) for pre-operative surgical planning. Operates by processing patient imaging to generate 3D bone models and surgical plans. Outputs include PDF reports and JSON files for integration with the REAL INTELLIGENCE CORI Surgical System. Compatible with various Smith & Nephew knee implant systems. The system utilizes a software-based planning workflow and implant database.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- VISIONAIRE UK Patient Matched Cutting Guides ([K211512](/device/K211512.md))

## Reference Devices

- REAL INTELLIGENCE CORI ([K221224](/device/K221224.md))
- REAL INTELLIGENCE CORI ([K240139](/device/K240139.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

March 18, 2024

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Smith & Nephew, Inc Wilcox Miriam Sr. Regulatory Affairs Specialist 1450 Brooks Road Memphis, Tennessee 38116

Re: K240113

Trade/Device Name: CORIOGRAPH Knee Pre-Op Plan Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: January 16, 2024 Received: January 16, 2024

Dear Wilcox Miriam:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shumaya Ali-S

Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

### Indications for Use

Submission Number (if known)

K240113

Device Name

CORIOGRAPH Knee Pre-Op Plan

Indications for Use (Describe)

Intended Use

CORIOGRAPH Pre-Op Planning and Modeling Service is intended to provide pre-operative planning for surgical procedures based on patient imaging, provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans.

Indications for Use

The CORIOGRAPH Pre-op Planning and Modeling Service is indicated for use for the following procedures:

- · unicondylar knee replacement (UKR)
- · total knee arthroplasty (TKA)

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

ver-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Smith & Nephew, Inc.

1450 Brooks Road Memphis, TN 38116 USA

T: + 1-901-396-2121 T: + 1-800-821-5700 (USA toll free) www.smith-nephew.com

# Smith Nephew

## 510(k) Summary

| 510(k) Owner             | Smith & Nephew, Inc.<br>1450 Brooks Road<br>Memphis, Tennessee 38116                                                                                                                |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person           | Miriam Wilcox<br>Senior Regulatory Affairs Specialist<br>Email: miriam.wilcox@smith-nephew.com                                                                                      |
| Date Prepared            | 18-March-2023                                                                                                                                                                       |
| Classification Reference | 21 CFR 888.3030                                                                                                                                                                     |
| Product Code             | PBF                                                                                                                                                                                 |
| Supported Codes          | JWH, MBH, OOG, HSX                                                                                                                                                                  |
| Common/Usual Name        | Orthopaedic Surgical Planning and Instrument Guides                                                                                                                                 |
| Trade/Proprietary Name   | CORIOGRAPH Knee Pre-Op Plan                                                                                                                                                         |
| Predicate Device(s)      | VISIONAIRE UK Patient Matched Cutting Guides (K211512)                                                                                                                              |
| Reason for Submission    | This Traditional 510(k) submission is seeking clearance for a new<br>offering, CORIOGRAPH Knee Pre-Op Plan V1.0, a module of the<br>CORIOGRAPH Pre-Op Planning and Modeling Service |

{4}------------------------------------------------

Smith & Nephew, Inc. 1450 Brooks Road Memphis, TN 38116 USA

-821-5700 (USA toll free) -nephew.com

## Intended Use

CORIOGRAPH Pre-Op Planning and Modeling Service is intended to provide pre-operative planning for surgical procedures based on patient imaging, provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans.

## Indications for Use

The CORIOGRAPH Pre-op Planning and Modeling Service is indicated for use for the following procedures:

- unicondylar knee replacement (UKR)
- . total knee arthroplasty (TKA)

### Device Description

The subject CORIOGRAPH Knee Pre-Op Plan is a medical function module within the CORIOGRAPH Pre-Op Planning and Modeling Service. This service is an additional offering being introduced by Smith & Nephew to allow for preoperative planning for surgical procedures based on patient imaging for total knee arthroplasty (TKA) and unicondylar knee replacement (UKR). The CORIOGRAPH Knee Pre-Op Plans will be generated for specific patients and anatomy by Smith and Nephew personnel using the CORIOGRAPH Knee Pre-Op Plan Software System V1.0, which is the subject of this submission. The CORIOGRAPH Knee Pre-Op Plans will output bone models of the patient anatomy and will generate patient-specific pre-operative plans which may be used on the REAL INTELLIGENCE CORI Surgical System.

{5}------------------------------------------------

Smith & Nephew, Inc. 1450 Brooks Road Memphis, TN 38116 USA

: + 1-901-396-2121 T: + 1-800-821-5700 (USA toll free) www.smith-nephew.com

## Currently Supported Knee Implants

The CORIOGRAPH Knee Pre-Op Plan is compatible with implant systems aligned to those supported with the REAL INTELLIGENCE CORI Surgical System. Compatible implant systems and clearance numbers are provided in Table 1.

TABLE 1. CURRENTLY SUPPORTED SMITH+ NEPHEW KNEE IMPLANTS FOR THE CORIOGRAPH KNEE PRE-OP PLAN V1.0

| Implant Model Name                         | 510(k) Number             | Classification<br>Product Code |
|--------------------------------------------|---------------------------|--------------------------------|
| JOURNEY II Unicompartmental Knee System    | K191211                   | HSX                            |
| JOURNEY UNI                                | K102069                   | HSX                            |
| JOURNEY II CR                              | K121443                   | JWH                            |
| JOURNEY II BCS                             | K111711                   | JWH                            |
| JOURNEY II XR                              | K141471, K152726          | JWH                            |
| LEGION CR/PS                               | K951987, K962557, K093746 | JWH                            |
| LEGION Porous CR Femoral Components        | K073325, K091543          | MBH                            |
| LEGION Porous CR Narrow Femoral Components | K210566                   | MBH                            |
| LEGION Porous Tibia                        | K100897                   | MBH                            |
| Porous Tibia Baseplate                     | K211221                   | MBH                            |
| GENESIS II CR/PS                           | K951987, K962557          | JWH                            |
| ANTHEM                                     | K142807                   | JWH                            |
| LEGION Knee System                         | K180334                   | JWH, MBH                       |

## REAL INTELLIGENCE CORI Surgical System Compatibility

The CORIOGRAPH Knee Pre-Op Plan is compatible with the following versions of the REAL INTELLIGENCE CORI surgical system.

#### TABLE 2. CURRENTLY SUPPORTED SYSTEM COMPATIBILITY FOR THE CORIOGRAPH KNEE PRE-OP PLAN V1.0

| Device Name            | Version | 510(k) Number | Classification<br>Product Code |
|------------------------|---------|---------------|--------------------------------|
| REAL INTELLIGENCE CORI | 3.0     | K240139       | HSX                            |

## Discussion of Similarities and Differences

The purpose of this Traditional 510(k) submission is to seek clearance for the CORIOGRAPH Knee Pre-Op Plan V1.0, which is a medical function module of the CORIOGRAPH Pre-Op Planning and Modeling Service. This Software as a Medical Device (SaMD) is based on the VISIONAIRE Patient-Matched Technology Software Process also used to generate preoperative plans and physical instruments by the predicate device VISIONAIRE Patient-Matched Cutting Guides (K211512).

The modifications made to the VISIONAIRE Patient-Matched Technology Software Process include an update to a software application to allow for:

- A new pre-operative planning workflow within the software application. This workflow will use the same implant database used on the CORI Surgical System.
- Added ability to detect if the case is intended for Visionaire Patient Matched Cutting Guide or for CORIOGRAPH Pre-operative planning from the surgeon inputs.

{6}------------------------------------------------

1-800-821-5700 (USA toll free) www.smith-nephew.com

- . Added ability to create a .json file and pre-operative plan images as outputs for pre-operative planning workflow.
- Implementation of minor bug fixes.

The software planning tools and processes used for the CORIOGRAPH Knee Pre-Op Plan have similar intended use, indications for use and technological characteristics as the software planning tools used for the preoperative plans generated for surgeon concurrence in the predicate VISIONAIRE UK Patient Matched Cutting Guides (K211512).

Smith & Nephew believes that CORIOGRAPH Knee Pre-Op Plan V1.0 is subject to premarket notification requirements under Section 510(k) of the Federal Food, Drug and Cosmetic Act ("FFDCA" or "the Act") and is substantially equivalent to the previously cleared REAL INTELLIGENCE CORI (K221224).

#### TABLE 3. PREDICATE DEVICE

| Manufacturer        | Description                                     | Submission Number | Clearance Date |
|---------------------|-------------------------------------------------|-------------------|----------------|
| Smith & Nephew, Inc | VISIONAIRE UK Patient<br>Matched Cutting Guides | K211512           | 09/16/2021     |

Substantial equivalence analysis for the CORIOGRAPH Knee Pre-Op Plan is provided in Table 4.

#### TABLE 4. SUBSTANTIAL EQUIVALENCE TABLE

| Design Aspect<br>Reviewed                         | CORIOGRAPH Knee Pre-Op<br>Plan<br>(Subject Device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | VISIONAIRE UK Patient<br>Matched Cutting Guides<br>(Predicate Device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Comparison                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|---------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number                                     | K240113                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | K211512<br>S.E. 09/16/2021                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Manufacturer                                      | Smith & Nephew, Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Smith & Nephew, Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Same as predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Design Aspect<br>Reviewed                         | CORIOGRAPH Knee Pre-Op<br>Plan<br>(Subject Device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | VISIONAIRE UK Patient<br>Matched Cutting Guides<br>(Predicate Device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Comparison                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Indications for<br>Use                            | Intended Use<br>CORIOGRAPH Pre-operative<br>Planning Service is intended<br>to provide pre-operative<br>planning for surgical<br>procedures based on patient<br>imaging, provided that<br>anatomic landmarks<br>necessary for alignment and<br>positioning of the implant<br>are identifiable on patient<br>imaging scans.<br><br>Indications for Use<br>The CORIOGRAPH Pre-<br>operative Planning Service is<br>indicated for use for the<br>following procedures:<br>unicondylar knee<br>replacement (UKR)total knee arthroplasty<br>(TKA) | Smith & Nephew's<br>VISIONAIRE UK Patient<br>Matched Cutting Guides are<br>intended to be used as<br>patient-specific surgical<br>instrumentation to assist in<br>the positioning of knee<br>replacement components<br>intra-operatively and in<br>guiding the marking of bone<br>before cutting provided that<br>anatomic landmarks<br>necessary for alignment and<br>positioning of the implant<br>are identifiable on patient<br>imaging scans.<br><br>The Smith & Nephew<br>VISIONAIRE UK Patient<br>Matched Cutting Guides are<br>intended for use with the<br>following existing Smith &<br>Nephew, Inc. Knee Systems<br>and their cleared indications<br>for use:<br>Genesis II Knee SystemLegion Knee SystemJourney BCS Knee<br>SystemJourney II Knee SystemJOURNEY II<br>Unicompartmental Knee<br>(JOURNEY II UK) System<br>(K190085)<br>The Smith & Nephew<br>VISIONAIRE UK Patient<br>Matched Cutting Guides are<br>intended for single use only. | Substantially equivalent -<br>Intended use is modified to<br>reflect that the output is a<br>software file instead of a<br>physical patient-matched<br>part. The indications for use<br>are updated to reflect the<br>surgical procedures instead<br>of the specific implant<br>systems. As the software<br>output is used within the<br>CORI surgical system, the<br>device is not limited by<br>manufacture of physical<br>parts. Compatible Implant<br>Systems will be disclosed in<br>the case management portal |
| Target<br>Population                              | Patients who are able to<br>undergo an MRI safely, fit<br>the indications, and do not<br>present any<br>contraindications for the<br>existing knee implant system<br>to which it is designed.                                                                                                                                                                                                                                                                                                                                                | Patients who are able to<br>undergo an MRI safely, fit<br>the indications, and do not<br>present any<br>contraindications for the<br>existing knee implant system<br>to which it is designed.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Same as predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Target<br>Anatomy                                 | Knee                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Knee                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Same as predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                                   | CORIOGRAPH Knee Pre-Op<br>Plan<br>(Subject Device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | VISIONAIRE UK Patient<br>Matched Cutting Guides<br>(Predicate Device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Comparison                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Design Aspect<br>Reviewed                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Where used                                        | Professional healthcare facility                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Professional healthcare facility                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Same as predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Compatible<br>Total Knee<br>Systems               | Genesis II CR<br>Genesis II PS<br>LEGION CR<br>LEGION CR Porous<br>LEGION CR Narrow Porous<br>LEGION PS<br>Journey II BCS<br>Journey II CR<br>Journey II XR<br>Journey II BCS ROX<br>Journey II CR ROX<br>Journey II UK<br>Journey Uni                                                                                                                                                                                                                                                                                                       | Genesis II CR<br>Genesis II PS<br>LEGION CR<br>LEGION PS<br>Journey II BCS<br>Journey II CR<br>Journey II XR<br>JOURNEY II<br>Unicompartmental Knee<br>(JOURNEY II UK)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Substantially equivalent - As<br>a software device,<br>compatible implant systems<br>for the subject device are<br>aligned with implants cleared<br>for use with the CORI<br>Surgical System, while<br>compatible implants for the<br>predicate device were<br>defined by cleared product<br>offerings for physical patient<br>matched guides.                                                                                                                                                                       |
| Design Type<br>Output                             | Patient Matched<br>Preoperative surgical plan<br>provided in PDF for surgeon<br>review. After surgeon<br>acceptance of the plan<br>parameters, PDF and/or<br>software file is provided to<br>be used as an input to<br>robotic surgical system.                                                                                                                                                                                                                                                                                              | Patient Matched<br>Preoperative surgical plan<br>provided in PDF for surgeon<br>review. After surgeon<br>acceptance of the plan<br>parameters, Physical Patient<br>Matched Cutting Block is<br>provided.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Same as predicate<br>Substantially equivalent -<br>Final output is a software file<br>instead of a physical patient-<br>matched cutting guide                                                                                                                                                                                                                                                                                                                                                                        |
| Common Name                                       | Orthopaedic Surgical<br>Planning and Instrument<br>Guides                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Knee Prosthesis                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Substantially Equivalent -<br>Output of Pre-Operative plan<br>is SaMD instead of a physical<br>guide, therefore product<br>code is updated to reflect<br>output.                                                                                                                                                                                                                                                                                                                                                     |
| Device Class                                      | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Same as predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Device<br>Classification<br>Name and<br>Reference | 21 CFR 888.3030 -<br>Single/multiple component<br>metallic bone fixation<br>appliances and accessories.                                                                                                                                                                                                                                                                                                                                                                                                                                      | 21 CFR 888.3520- Knee Joint<br>Femorotibial metal/polymer<br>nonconstrained cemented<br>prosthesis<br>21 CFR 888.3560- Knee Joint<br>Patellofemorotibial<br>polymer/metal/polymer<br>semi-constrained cemented<br>prosthesis                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Substantially Equivalent -<br>Output of Pre-Operative plan<br>is SaMD instead of a physical<br>guide, therefore product<br>code is updated to reflect<br>output.                                                                                                                                                                                                                                                                                                                                                     |
| Panel Code                                        | Orthopaedics/87                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Orthopaedics/87                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same as predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Product Code                                      | PBF                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | HSX<br>OOG                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Substantially Equivalent -<br>Output of Pre-Operative plan<br>is SaMD instead of a physical<br>guide, therefore product<br>code is updated to reflect<br>output.                                                                                                                                                                                                                                                                                                                                                     |
| Design Aspect<br>Reviewed                         | CORIOGRAPH Knee Pre-Op<br>Plan<br>(Subject Device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | VISIONAIRE UK Patient<br>Matched Cutting Guides<br>(Predicate Device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Comparison                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Supported<br>Product Code                         | JWH<br>MBH<br>OOG<br>HSX                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | JWH<br>MBH<br>OOG<br>HSX                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Same as predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |

{7}------------------------------------------------

## Smith Nephew

#### Smith & Nephew, Inc.

1450 Brooks Road Memphis, TN 38116
USA T: + 1-901-396-2121 T: + 1-800-821-5700 (USA toll free) www.smith-nephew.com

{8}------------------------------------------------

## Smith Nephew

Smith & Nephew, Inc.

1450 Brooks Road Memphis, TN 38116
USA T: + 1-901-396-2121 T: + 1-800-821-5700 (USA toll free) www.smith-nephew.com

{9}------------------------------------------------

Smith & Nephew, Inc. 1450 Brooks Road Memnhis TN 38116

USA

5700 (USA toll free) ephew.com

## Non-Clinical Testing (Bench)

Verification and validation testing demonstrated the safety and effectiveness of the software applications used in the CORIOGRAPH Knee Pre-Op Plan and demonstrated the safety and effectiveness of the generated CORIOGRAPH Knee Pre-Op Plan when used as an input to the CORI Surgical System. Summative usability testing demonstrated that participating surgeons were able use the subject device safely and effectively in a simulated use environment. Smith and Nephew, Inc. has concluded that all design inputs have been met and that the verification and validation testing performed did not raise any new questions of safety or effectiveness.

## Conclusion

The software tools and process used for CORIOGRAPH Knee Pre-Op Plan have similar intended use and indications for use and the same technological characteristics used in the software planning tools and processes used for the preoperative plans generated for the predicate VISIONAIRE UK Patient Matched Cutting Guides (K211512). The differences between the two systems include an update to support the implant database used by the CORI Surgical System as well as a modification of the output to include a JSON file and preoperative plan images for use on the CORI Surgical System.

Verification testing demonstrated that the system meets required design inputs. Summative usability testing results demonstrate that representative users can use the safely and effectively in a simulated use environment. The information presented in this 510(k) premarket notification demonstrates that the CORIOGRAPH Knee Pre-Op Plan System may be used to generative plans for surgical procedures based on patient imaging. This information also demonstrates that the CORIOGRAPH Knee Pre-Op Plan system is as safe and effective as the software and planning process used in the predicate VISIONAIRE UK Patient Matched Cutting Guides (K211512). Smith and Nephew, Inc. believes that FDA can find CORIOGRAPH Knee Pre-Op Plan V1.0 to be substantially equivalent to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K240113](https://fda-staging.innolitics.com/device/K240113)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).