Anterior restorations (Class III, IV) Class V restorations (cervical caries, root erosion, wedge-shaped defects) Restorations in the posterior region (Class I and II) Veneering of discolored anterior teeth Splinting of mobile teeth Repair of composite and ceramic veneers
Device Story
Light Cure Composite is a radiopaque, nano-hybrid restorative material; supplied as a paste in a PP syringe. Used by dental professionals in clinics or authorized laboratories for restorative therapy. Operates via light-curing mechanism using blue light (400–500 nm wavelength); requires external light curing unit (power ≥ 400mW/cm2). Material is applied to tooth structure or existing restorations and cured for 20 seconds to achieve hardening. Provides structural restoration of tooth anatomy and aesthetics; benefits patient by repairing defects, restoring function, and improving appearance of teeth.
Clinical Evidence
Bench testing only. Performance evaluated against ISO 4049 standards including depth of cure, flexural strength, water sorption, solubility, color stability, and container delivery performance. Mechanical properties (compressive strength, elastic modulus, surface hardness) met specified acceptance criteria. Biocompatibility confirmed per ISO 10993-1.
Indicated for patients requiring dental restorations including anterior (Class III, IV) and posterior (Class I, II) cavities, Class V cervical/root/wedge defects, veneering of discolored teeth, splinting of mobile teeth, and repair of composite or ceramic veneers.
Regulatory Classification
Identification
Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 15, 2024
Sincera Technology (Changchun) Co., Ltd. % Eva Li Consultant Shanghai SUNGO Management Consulting Co., Ltd. Room 1401, Dongfang Building, 1500# Century Ave. Shanghai, China
Re: K240112
Trade/Device Name: Light Cure Composite Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth Shade Resin Material Regulatory Class: Class II Product Code: EBF Dated: January 16, 2024 Received: January 16, 2024
Dear Eva Li:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Bobak Shirmohammadi -S
For Michael E. Adjodha, M.ChE., RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices
{2}------------------------------------------------
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K240112
Device Name Light Cure Composite
Indications for Use (Describe) Anterior restorations (Class III, IV) Class V restorations (cervical caries, root erosion, wedge-shaped defects) Restorations in the posterior region (Class I and II) Veneering of discolored anterior teeth Splinting of mobile teeth Repair of composite and ceramic veneers
Type of Use (Select one or both, as applicable)|X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(K) Summary
## 1. Information of the submitter:
Submitter's Name: Sincera Technology (Changchun) Co., Ltd. Address: No. 177 Changda Rd, Hi-tech District, Changchun City, Jilin Province, P. R. China Contact person: Tony Su
Tel: +8615143153582
Email: tony.su@sinceradental.com
Preparation Date: Jan 2nd, 2024
## 2. Basic information of the device
Classification Name: Material, Tooth Shade, Resin Common / Usual Name: Tooth shade resin material. Proprietary Name: Light Cure Composite Product Code: EBF Class: Class II Regulation: 21 CFR 872.3690
## 3. Predicate Device
K042819 TETRIC EVOCERAM VOCLAR VIVADENT, INC.
## 4. Indication for Use
Anterior restorations (Class III, IV) Class V restorations (cervical caries, root erosion, wedge-shaped defects) Restorations in the posterior region (Class I and II) Veneering of discolored anterior teeth Splinting of mobile teeth Repair of composite and ceramic veneers
## 5. Device Description
Light Cure Composite is a state-of-the-art light-curing, radiopaque, nano-hybrid composite for the restorative therapy. The device cures with light at the wavelength range of 400–500 nm (blue light). The device is paste contained in PP syringe.
| Item | Subject device | Predicated device | Comparison |
|-----------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Device Name | Light Cure Composite | TETRIC VOCERAM | N/A |
| 510(K) No. | | K042819 | N/A |
| Classification | Class II | Class II | Same |
| Regulation | 21 CFR 872.3690 | 21 CFR 872.3690 | Same |
| Indication for use | Anterior restorations | Anterior restorations (Class) | Same |
| No. 177 Changda Rd, Hi-tech District, Changchun City, Jilin Province, P. R. China | | | |
| | (Class III, IV)<br>Class V restorations<br>(cervical caries, root<br>erosion, wedge-shaped<br>defects)<br>Restorations in the<br>posterior region (Class I<br>and II)<br>Veneering of discolored<br>anterior teeth<br>Splinting of mobile teeth<br>Repair of composite and<br>ceramic veneers | (Class III, IV)<br>Class V restorations<br>(cervical caries, root<br>erosion, wedge-shaped<br>defects)<br>Restorations in the<br>posterior region (Class I<br>and II)<br>Veneering of discolored<br>anterior teeth<br>Splinting of mobile teeth<br>Repair of composite and<br>ceramic veneers | |
| Product Code | EBF | EBF | Same |
| Patient Population | All the groups | All the groups | Same |
| Prescription Use | Rx only | Rx only | Same |
| Environment | Dental prosthetics and<br>authorized laboratories<br>and clinics | Dental prosthetics and<br>authorized laboratories<br>and clinics | Same |
| Applicable Standards | ISO 4049; ISO 10993-1 | ISO 4049; ISO 10993-1 | Same |
| Device Sterilization | No | No | Same |
| Shelf life | 2 years | 3 years | Different*1 |
| Light curing unit | Power: ≥ 400mW/cm2<br>Wavelength: 400-500nm<br>Curing time:20s | Power: ≥500mW/cm2<br>Wavelength: 400-500nm<br>Curing time:20s | Different*2 |
| particle size | 0.4-2μm | 0.4-3μm | Different*3 |
| Elastic Modulus(GPa) | Mean:8.5 | Mean:7.4 | Different*3 |
| Compressive<br>Strength(MPa) | Mean:286.5 | Mean:248.4 | Different*3 |
| Surface | Mean:63.1 | Mean:53.8 | Different*3 |
| Hardness(HV) | | | |
| Radio-opacity | Mean:2.27mm | Not provided | N/A |
#### 6. Technological Characteristics Comparison:
{5}------------------------------------------------
## 7. Non-clinical performance Data:
| Test Item | Acceptance criterial | Result |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| Outward appearance-ISO 4049 | According to visual 1nspect1on, the product should be uniform in texture and should be a paste without visible impurities and foreign matter. | Meet the requirement |
| Sensitivity to ambient light-ISO 4049 | The material of sample remains physically homogeneous. | Meet the requirement |
| Depth of Cure-ISO 4049 | ≥1.5mm | ≥1.5mm |
| Flexural strength-ISO 4049 | ≥80Mpa | ≥80Mpa |
| Water sorption-ISO 4049 | ≤40µg/mm³ | ≤40µg/mm³ |
| solubility-ISO 4049 | ≤7.5µg/mm³ | ≤7.5µg/mm³ |
{6}------------------------------------------------
Sincera Technology (Changchun) Co., Ltd. No. 177 Changda Rd, Hi-tech District, Changchun City, Jilin Province, P. R. China
| Shade-ISO 4049 | Without the need for a magnifying glass to observe with the naked eye, the cured material should be evently coloredk | Meet the requirement |
|-----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| Color comparison for color stability-ISO 4049 | Only slight changes are allowed | Meet the requirement |
| Appearance of container-ISO 4049 | Visually observe that the surface of the container should be smooth, without scratches, cracks, burrs, small notches or macroscopic defects caused by mechanical processing. | Meet the requirement |
| Delivery performance of container-ISO 4049 | Remove the syringe cap and turn the container push rod clockwise allowing resin to be smoothly squeezed out of the syringe without obstruction | Meet the requirement |
| Compressive Strength | $\ge$ 200 Mpa | Meet the requirement |
| Elastic Modulus | $\ge$ 6 Gpa | Meet the requirement |
| Surface Hardness | $\ge$ 60(HV) | Meet the requirement |
| particle size | 0.4-2μm | |
| Radio-opacity | Mean: 2.27mm | |
## 8. Biocompatibility
Meet ISO 10993-1 - Biological evaluation of medical devices — Part 1: Evaluation and testing.
## 9. Conclusion
The results of the testing described above provide reasonable assurance that the subject device is as safe and effective as the predicate device(s) which cleared under K042819 and supports a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.