This Blood Pressure Monitor is a digital monitor intended for use in measuring blood pressure and pulse rate with arm circumference ranging from 22cm to 32cm (about 8¾"-12½"), 22cm to 42cm (about 8¾"-16½"), 22cm to 45cm (about 8¾"-17¾") or 40cm to 52cm (about 15¾"-20½"). It is intended for adult indoor use only.
Device Story
Digital blood pressure monitor (TMB-2287-B) for home use; measures systolic/diastolic pressure and pulse rate via oscillometric method. Device consists of main unit (HIPS housing) and inflatable arm cuff; uses electronic pressure sensor to convert cuff pressure fluctuations into electrical signals. Features irregular pulse detection based on pulse interval variability analysis. Bluetooth module transfers data to external APP for storage/display; software/APP functions as non-device MDDS. Operated by adult patients; results displayed on LCD. Provides clinical data for blood pressure management.
Clinical Evidence
Clinical validation performed per ISO 81060-2:2018/Amd.1:2020. Two studies conducted with 87 and 91 adult subjects (total 530 datasets). Mean differences for SBP/DBP were -0.77±2.09/-0.08±1.69 mmHg (Criterion 1) and 0.2±2.96/0.76±3.42 mmHg (Criterion 1) across studies. Device met both validation criteria 1 and 2.
Technological Characteristics
Oscillometric measurement; HIPS housing; electronic pressure sensor; Bluetooth connectivity. Power: 4x1.5V AAA batteries or 5V DC adapter. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 80601-2-30, ISO 10993-1/5/10/23. Biocompatibility: surface-contacting, <30 days.
Indications for Use
Indicated for adult patients requiring non-invasive measurement of systolic/diastolic blood pressure and pulse rate in a home environment. Compatible with arm circumferences 22-52cm.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath. The logo is simple and professional, and it is easily recognizable.
April 29, 2024
Guangdong Transtek Medical Electronics Co., Ltd. Jerry Fan RA Manager Zone A, No. 105, Dongli Road, Torch Development District Zhongshan, Guangdong, China 528437
Re: K240102
Trade/Device Name: Blood Pressure Monitor (TMB-2287-B) Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: April 3, 2024 Received: April 3, 2024
Dear Jerry Fan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Stephen C. Browning -S
LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
K240102
Device Name
Blood Pressure Monitor (TMB-2287-B)
Indications for Use (Describe)
This Blood Pressure Monitor is a digital monitor intended for use in measuring blood pressure and pulse rate with arm circumference ranging from 22cm to 32cm (about 8% - 12½"), 22cm to 42cm (about 8%"-16%"), 22cm to 45cm (about 8%"-17%") or 40cm to 52cm (about 15%"-20%"). It is intended for adult indoor use only.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### 510(k) Summary
Prepared in accordance with the requirement of 21 CFR Part 807.92
Prepared Date: 03/11/2024
# 1. Submitter
| Name: | Guangdong Transtek Medical Electronics Co., Ltd. |
|----------|-------------------------------------------------------------------------------------------------|
| Address: | Zone A, No. 105, Dongli Road, Torch Development District, 528437 Zhongshan,<br>Guangdong, China |
| Contact: | Jerry Fan |
| Title: | RA Manager |
| E-mail: | gt-rateam@transtekcorp.com |
| TEL: | +86 15728668528 |
#### Subject Device 2.
| Trade/Device Name | Blood Pressure Monitor |
|-------------------|------------------------|
| Model | (TMB-2287-B) |
| Common Name | Blood Pressure Monitor |
| Classification | Class II |
| Product Code | DXN |
| Submission Type | Traditional 510(k) |
#### Predicate Device 3.
| Manufacturer: | Guangdong Transtek Medical Electronics Co., Ltd. |
|--------------------|--------------------------------------------------|
| Device Name/Model: | Blood Pressure Monitor / TMB-2266 |
| 510(k) Number: | K232713 |
| Classification | Class II |
| Product Code | DXN |
#### 4. Device Description
Blood Pressure Monitor TMB-2287-B is designed to measure systolic pressure, diastolic pressure and pulse rate of adult by a non-invasive technique, with an inflatable cuff wrapped around the upper arm. The method to define systolic and diastolic pressure is similar to auscultatory method, though it uses an electronic pressure sensor rather than a stethoscope and mercury manometer. The sensor converts tiny alternations of cuff pressure into electrical signals.
Based on analysis of these signals, the systolic and diastolic blood pressure is defined, and the pulse rate is calculated. This is an extensively used technique applied in blood pressure monitors, also known as "oscillometric method".
The main components of the Blood Pressure Monitor include main unit and cuff. For the outer housing of the main unit, it's made of HIPS material. As for accompanying cuffs, five types of cuffs have been clinically validated to be matched with the device, suitable for population aged at or over three and with arm circumference from 22cm~52cm. The cuff is consisted of fabric and an inflatable bladder inside. For critical electronic components, there is medical switch power supply, PCB, thermistor, pressure pump, motor, release valve and pressure sensor.
The device also enjoys a function of detecting irregular pulse rate. When measurements were
{5}------------------------------------------------
performed, the monitor will record all pulse intervals and calculate the average. If two or more pulse intervals were recorded, and the difference between each interval and the average is larger than ±25% of the average; or if four or more pulse intervals were recorded, and the difference between each interval and the average is larger than ± 15% of the average, the irregular pulse symbol will be displayed along with measurement results.
An embedded Bluetooth wireless connection module in the device allows it to connect with matching receiving ends. When a measurement is done, the results will be displayed on LCD, and the measured data will be transferred to the APP via Bluetooth.
With the use of software (including APP) and Bluetooth communication module, the wireless software function and hardware function are solely intended to transfer, store, convert formats, or display medical device data and results (blood pressure and pulse rate readings), without controlling or altering the functions or parameters of any connected medical devices, which is not be intended for active patient monitoring, therefore, based on the FDA guidance titled "Medical Device Data Systems, Medical Image Storage Devices, and Medical Image Communications Devices" (issued on September 28, 2022.), this software function is belong to Non-device-MDDS, and the hardware function is belong to Device-MDDS, they are not subject to FDA laws and regulations applicable to devices.
### 5. Indications for use
This Blood Pressure Monitor is a digital monitor intended for use in measuring blood pressure and pulse rate with arm circumference ranging from 22cm to 32cm (about 8¾˜-12½´´), 22cm to 42cm (about 8% -16½"), 22cm to 45cm (about 8¾ -17¾ ") or 40cm to 52cm (about 15¾ -20½ "). It is intended for adult indoor use only.
| Item | Subject Device | Predicate Device | Note |
|----------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|
| 510(k) | Applying | K232713 | / |
| Model | TMB-2287-B | TMB-2266 | / |
| Applicant | Guangdong Transtek Medical<br>Electronics Co., Ltd. | Guangdong Transtek Medical<br>Electronics Co., Ltd. | Same |
| Device Name | Blood Pressure Monitor | Blood Pressure Monitor | Same |
| Product Code | DXN | DXN | Same |
| Classification | Class II | Class II | Same |
| Regulation # | 21 CFR 870.1130 | 21 CFR 870.1130 | Same |
| Type of Use | OTC | OTC | Same |
| Intended Use /<br>Indication for<br>Use | This Blood Pressure Monitor is a<br>digital monitor intended for use in<br>measuring blood pressure and pulse<br>rate with arm circumference ranging<br>from 22cm to 32cm (about<br>8¾″-12½″), 22cm to 42cm (about | This Blood Pressure Monitor is a<br>digital monitor intended for use in<br>measuring blood pressure and pulse<br>rate with arm circumference ranging<br>from 16cm to 36cm (about<br>6⅓″-14⅕″), 22cm to 32cm (about | Similar, refer<br>to Note 1. |
| | 8¾"-16½"), 22cm to 45cm (about<br>8¾"-17¾") or 40cm to 52cm (about<br>15¾"-20½").<br>It is intended for adult indoor use<br>only. | 8¾"-12½"), 22cm to 42cm (about<br>8¾"-16½") or 22cm to 45cm (about<br>8¾"-17¾").<br>The cuff with arm circumference<br>range of 16~36cm is intended for<br>children older than 6 years old or<br>adults.<br>The cuffs with arm circumference<br>range of 22~32cm or 22~42cm or<br>22~45cm, which are intended for<br>adult population.<br>It is intended indoor use only. | |
| Patient<br>Population | Adult | Population at or over 3 years old | Similar, refer<br>to Note 2. |
| Principle | Oscillometric method | Oscillometric method | Same |
| Anatomical<br>Site | Upper Arm | Upper Arm | Same |
| Where used<br>(hospital,<br>home,<br>ambulance,<br>etc.) | Home | Home | Same |
| Power Supply | 4*1.5V AAA batteries;<br>Or by DC 5V adapter | 4*1.5V AAA batteries;<br>Or by DC 5V adapter | Same |
| Human Factors | Blood pressure | Blood pressure | Same |
| Measurement<br>Items | Measuring systolic and diastolic<br>blood pressure and pulse rate of<br>intended population, including<br>irregular pulse rhythm detection. | Measuring systolic and diastolic<br>blood pressure and pulse rate of<br>intended population, including<br>irregular pulse rhythm detection. | Same |
| Cuff Deflation | Automatic deflation | Automatic deflation | Same |
| Blood Pressure<br>Measurement | 0mmHg ~ 299mmHg,<br>5°C - 40°C within ±3mmHg (0.4kPa) | 0mmHg ~ 299mmHg,<br>5°C - 40°C within ±3mmHg (0.4kPa) | Same |
| Pulse Rate<br>Measurement | 40-199 beat/minute, ±5% | 40-199 beat/minute, ±5% | Same |
| Display | LCD | LCD | Same |
| Memory | 2×199 | 2×199 | Same |
| Operation<br>Environment | Temperature: 5°C~40°C;<br>Relative Humidity: 15%~90% RH; | Temperature: 5°C~40°C;<br>Relative Humidity: 15%~90% RH; | Same |
| | Atmospheric: 700hPa~1060hPa | Atmospheric: 700hPa~1060hPa | |
| | Temperature: -20°C~60°C; | Temperature: -20°C~60°C; | |
| Storage and | Relative humidity≤93%RH, | Relative humidity≤93%RH, | |
| transportation | non-condensing; | non-condensing; | Same |
| Environment | Atmospheric: 500hPa~1060hPa | Atmospheric: 500hPa~1060hPa | |
| Performance | Compliance with IEC 80601-2-30 | Compliance with IEC 80601-2-30 | Same |
| Clinical | Compliance with ISO 81060-2 | Compliance with ISO 81060-2 | Same |
| Biocompatibilit | All patient contact parts meet the | All patient contact parts meet the | |
| y | requirements of ISO 10993-1/5/10/23 | requirements of ISO 10993-1/5/10/23 | Same |
| Electrical | Compliance with IEC 60601-1 and | Compliance with IEC 60601-1 and | |
| Safety | IEC 60601-1-11 | IEC 60601-1-11 | Same |
| EMC | Compliance with IEC 60601-1-2 | Compliance with IEC 60601-1-2 | Same |
| Wireless | Bluetooth | Bluetooth | Same |
#### Comparison to Predicate Device 6.
{6}------------------------------------------------
{7}------------------------------------------------
# Note 1:
The subject device shares a similar intended use with the predicate device, and the substantial difference between the subject device and the predicate device regarding the intended use lies in the arm circumference range and patient population. However, the range of the intented patient population of the subject device is within that of the predicate device, and used on its intended patient population with the intended arm circumference, the subject device with matched cuffs has also been validated according to IEC 80601-2-30 and ISO 81060-2. As demonstrated in relevant test reports, the difference here does not raise any issues concerning safety and effectiveness.
# Note 2:
Although there are some differences between the intended patient population of the subject device and the predicate device, the range of the intended patient population of the subject device is within that of the predicate device, and used on its intended patient population with the intended arm circumference, the subject device with matched cuffs has also been validated according to IEC 80601-2-30 and ISO 81060-2. As demonstrated in relevant test reports, the difference here does not raise any issues concerning safety and effectiveness.
## Conclusion: _
Based on the comparison and analysis in this submission, it can be concluded that: the subject device is substantially equivalent to the predicate device and reference device regarding safety and effectiveness.
## 7. Performance Data
The following performance data were provided in support of the substantial equivalence determination:
{8}------------------------------------------------
# Biocompatibility testing:
The biocompatibility evaluation for the device was conducted in accordance with the FDA Guidance for Industry and Food and Drug Administration Staff: Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". The biocompatibility testing includes the following tests:
- Cytotoxicity
- Sensitization
- Irritation
The subject device is considered as surface contacting for a duration of exceed 24 hours but not exceed 30 days.
# Non-clinical data
The device has been tested according to following standards:
- IEC 60601-1: Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2: Medical electrical equipment - Part 1-2: General requirement for basic safety and essential performance - Collateral standard: Electromagnetic compatibility -Requirements and tests
- IEC 80601-2-30: Medical electrical equipment -- Particular requirements for basic safety and essential performance of automated non-invasive sphygmomanometers.
- IEC 60601-1-11: Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance – Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- FDA Guidance for Non-Automated Sphygmomanometer
# Wireless testing
- 47 CFR Part 15, Subpart C
- Radio Frequency Wireless Technology in Medical Devices: Guidance for Industry and Food ● and Drug Administration Staff (August 14, 2013)
# Clinical data
The device was tested according to ISO 81060-2:2018/Amd.1:2020. Non-invasive sphygmomanometers - Part 2: Clinical validation of intermittent automated measurement type. Two clinical studies were performed on the device matching with different models of cuff. The study population consisted of 87 and 91 qualified subjects respectively. 258 datasets were collected from 87 subjects with aged 58.8+8.4. The mean differences between reference BPs and device readings were -0.77±2.09/-0.08±1.69mmHg for systolic BP (SBP)/diastolic BP (DBP) of criterion 1, and -0.77± 1.58/-0.08±1.17mmHg for SBP/DBP of criterion 2. 272 datasets were collected from 91 subjects with aged 50.6±12.04. The mean differences between reference BPs and device readings were 0.2+2.96/0.76±3.42mmHg for systolic BP (SBP)/diastolic BP (DBP) of criterion 1, and 0.2± 2.46/0.76±2.83mHg for SBP/DBP of criterion 2. TMB-2287-B fulfilled both validation criteria 1 and 2 of the ISO 81060-2:2018/Amd.1:2020.
{9}------------------------------------------------
### 8. Conclusion
Based on the information presented in this 510(k) premarket notification submission, a conclusion can be drawn that the proposed subject device is considered substantially equivalent to the predicate device. The differences between the subject device and the predicate devices were successfully tested with relevant standards and FDA guidance, and do not affect equivalent safety and effectiveness or raise new issues concerning safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.