# Human Lubricant (K240081)

_Jiangxi Wulinxiang Biological Technology Co., Ltd. · NUC · Aug 16, 2024 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K240081

## Device Facts

- **Applicant:** Jiangxi Wulinxiang Biological Technology Co., Ltd.
- **Product Code:** [NUC](/productcode/NUC.md)
- **Decision Date:** Aug 16, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.5300
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology

## Indications for Use

WLX & Bio Human Lubricant is a water based personal lubricant, for penile and/or vaginal application, intended to lubricate and moisturize, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is not compatible with natural rubber latex, polyurethane, and polyisoprene condoms.

## Device Story

WLX & Bio Human Lubricant is a non-sterile, water-based, clear gel personal lubricant. It is intended for penile and/or vaginal application to provide lubrication and moisture during intimate sexual activity. The device is packaged in 100 mL polyethylene (PE) bottles. It is an over-the-counter product used by consumers to enhance comfort during sexual activity. The lubricant is not compatible with natural rubber latex, polyurethane, or polyisoprene condoms. It functions by supplementing the body's natural lubrication.

## Clinical Evidence

Bench testing only. Biocompatibility testing performed per ISO 10993-1, including cytotoxicity (ISO 10993-5), sensitization (ISO 10993-10), vaginal irritation (ISO 10993-23), and acute systemic toxicity (ISO 10993-11). Results confirmed the device is non-cytotoxic, non-irritating, non-sensitizing, and not systemically toxic. Shelf-life validated via real-time and accelerated aging (ASTM F1980-16) over 6 months. Microbial limits and antimicrobial effectiveness tested per USP <51>, <61>, and <62>.

## Technological Characteristics

Water-based gel; ingredients include Water, Glycerin, Propylene Glycol, Carbomer, Hydroxyethylcellulose, Triethanolamine, and Phenoxyethanol. pH 5.5-7.5; viscosity 100-300 cps at 25°C; osmolality 750-950 mOsm/kg. Packaged in 100 mL polyethylene (PE) bottles. Non-sterile. Shelf-life 6 months validated per ASTM F1980-16.

## Regulatory Identification

A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.

## Special Controls

*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.

## Predicate Devices

- pjur® AQUA Baseline ([K200731](/device/K200731.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and "ADMINISTRATION" in a smaller font size below.

August 16, 2024

Jiangxi Wulinxiang Biological Technology Co., Ltd. Yang Qi GM 1F,Building 1,Nanchang LED Industry Innovation Demonstration Park, No. 1111 Changdong Avenue, Qingshan, Nanchang, Jiangxi 330012 CHN

Re: K240081

Trade/Device Name: WLX & Bio Human Lubricant Regulation Number: 21 CFR\$ 884.5300 Regulation Name: Condom Regulatory Class: II Product Code: NUC Dated: July 11, 2024 Received: July 15, 2024

Dear Yang Qi:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

{2}------------------------------------------------

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Monica D. Garcia -S

Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known) K240081

Device Name WLX & Bio Human Lubricant

### Indications for Use (Describe)

WLX & Bio Human Lubricant is a water based personal lubricant, for penile and/or vaginal application, intended to lubricate and moisturize, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is not compatible with natural rubber latex, polyurethane, and polyisoprene condoms.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 201 Subpart D) | <input type="checkbox"/>            |
|----------------------------------------------|-------------------------------------|
| Over-The-Counter Use (21 CFR 201 Subpart G)  | <input checked="" type="checkbox"/> |

|_ | Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

### 510(k) Summary — K240081 WLX & Bio Human Lubricant

#### l. General Information on Submitter

| Applicant:      | Jiangxi Wulinxiang Biological |
|-----------------|-------------------------------|
| Address:        | 1F, Building 1, Nanchang LED  |
|                 | Industry Innovation           |
|                 | Demonstration Park, No. 1111  |
|                 | Changdong Avenue, Qingshan    |
|                 | Nanchang, Jiangxi 330012      |
|                 | CHN                           |
| Telephone:      | 86 (791) 881-10717            |
| Contact Person: | Yang Qi,                      |
| Contact Title:  | GM                            |
| Email:          | 13755676363@163.com           |
| Date Prepared:  | August 15, 2024               |

#### II. General Information on Device

| Trade Name:       | WLX & Bio Human Lubricant |
|-------------------|---------------------------|
| Common Name:      | Personal Lubricant        |
| Regulation Name:  | Condom                    |
| Regulation Number | 21 CFR 884.5300           |
| Device Class:     | II                        |
| Product Code:     | NUC (Lubricant, Personal) |

#### lll. Predicate Device

| Predicate Device    | 510(k) Number |
|---------------------|---------------|
| pjur@ AQUA Baseline | K200731       |

The predicate device has not been subject to a design-related recall.

#### IV. Description of Device

WLX & Bio Human Lubricant is a non-sterile, water-based, clear gel personal lubricant. It is a Class II medical device classified under 21 CFR 884.5300 (condom) and product code NUC (lubricant, personal). It contains Water, Glycerin, Propylene Glycol, Carbomer, Hydroxyethylcellulose, Triethanolamine, and Phenoxyethanol. The lubricant is not compatible with natural rubber latex, polyurethane, and polyisoprene condoms.

{5}------------------------------------------------

The lubricant is packaged in polyethylene (PE) bottles of 100 mL capacity (Model -WLX100).

The specifications for WLX & Bio Human Lubricant are described in Table 1.

| Parameter                                                                                                     | Test<br>Method | Specification                                                                                                                                                                                                                                                                                                           |
|---------------------------------------------------------------------------------------------------------------|----------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Appearance/Color                                                                                              | Visual         | Clear gel                                                                                                                                                                                                                                                                                                               |
| Odor                                                                                                          | Olfactory      | Odorless                                                                                                                                                                                                                                                                                                                |
| pH                                                                                                            | USP<791>       | 5.5-7.5                                                                                                                                                                                                                                                                                                                 |
| Viscosity                                                                                                     | USP<912>       | 100-300 (cps at 25°C)                                                                                                                                                                                                                                                                                                   |
| Osmolality                                                                                                    | USP<785>       | 750-950 mOsm/kg                                                                                                                                                                                                                                                                                                         |
| Antimicrobial Effectiveness                                                                                   | USP <51>       | Meets USP <51> acceptance<br>criteria for Category 2<br>products. Category 2, bacteria should<br>show not be less than 2.0 log reduction<br>at 14 days and no increase from 14-day<br>Count at the 28 day count. Yeasts and<br>molds should show no increase from<br>the initial calculated count at 14 and 28<br>days. |
| Total Aerobic Microbial Count<br>(TAMC)                                                                       | USP<61>        | Less than 100 cfu/mL                                                                                                                                                                                                                                                                                                    |
| Total Yeast and Mold Count<br>(TYMC)                                                                          | USP<61>        | Less than 10 cfu/mL                                                                                                                                                                                                                                                                                                     |
| Absence of Pathogenic<br>Organisms (Candida<br>albicans, Pseudomonas<br>aeruginosa, Staphylococcus<br>aureus) | USP<62>        | Absent                                                                                                                                                                                                                                                                                                                  |

# Table 1. Device Specifications

#### V. Indications for Use

WLX & Bio Human Lubricant is a water based personal lubricant, for penile and/or vaginal application, intended to lubricate and moisturize, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is not compatible with natural rubber latex, polyurethane, and polyisoprene condoms.

#### VI. Substantial Equivalence Discussion

The following table compares the intended use and technological characteristics of the subject and predicate device:

{6}------------------------------------------------

| Characteristic /<br>Feature             | WLX & Bio Human Lubricant<br>(subject device)                                                                                                                                                                                                                                                                                                                                                                                                     | pjur@ AQUA Baseline<br>(predicate device) - K200731                                                                                                                                                                                                                                                                                                                                                                                               | Comparison                                                                                                                                |
|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for use                     | WLX & Bio Human Lubricant is a<br>water based personal lubricant, for<br>penile and/or vaginal application,<br>intended to lubricate and moisturize,<br>to enhance the ease and comfort of<br>intimate sexual activity and<br>supplement the body's natural<br>lubrication. This product is not<br>compatible with natural rubber latex,<br>polyurethane, and polyisoprene<br>condoms.                                                            | Pjur@ AQUA Baseline for penile,<br>vaginal and/or anal application,<br>intended to moisturize and lubricate,<br>enhance the ease and comfort of<br>intimate sexual activity and<br>supplement the body's natural<br>lubrication. This product is not<br>compatible with natural rubber latex,<br>polyisoprene, and polyurethane<br>condoms.                                                                                                       | Similar: The<br>subject and<br>predicate<br>devices have<br>similar<br>indications for<br>use, and they<br>have the same<br>intended use. |
| Water-Based<br>Lubricant                | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                               | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                               | Same                                                                                                                                      |
| Over the Counter                        | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                               | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                               | Same                                                                                                                                      |
| Not a<br>contraceptive or<br>Spermicide | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                               | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                               | Same                                                                                                                                      |
| Non-sterile                             | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                               | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                               | Same                                                                                                                                      |
| Primary<br>Ingredients                  | Water, Glycerin, Propylene Glycol,<br>Carbomer, Hydroxyethylcellulose,<br>Triethanolamine, and Phenoxyethanol                                                                                                                                                                                                                                                                                                                                     | water, glycerin, ethoxydiglycol,<br>phenoxyethanol, hydroxypropyl<br>guar hydroxypropyltrimonium<br>chloride, hydroxyethylcellulose,<br>and citric acid                                                                                                                                                                                                                                                                                           | Different                                                                                                                                 |
| Microbial Limits                        | Total mold/yeast count <10 cfu/mL<br>Total aerobic microbial count <100<br>cfu/mL<br>Absence of pathogens (Candida<br>albicans, Pseudomonas aeruginosa,<br>Staphylococcus aureus)<br>Antimicrobial effectiveness for<br>Category 2, bacteria: not less than<br>2.0 log reduction at 14 days and no<br>increase from 14-day count at the<br>28 day count. Yeasts and molds: no<br>increase from the initial calculated<br>count at 14 and 28 days. | Total mold/yeast count <10 cfu/mL<br>Total aerobic microbial count <100<br>cfu/mL<br>Absence of pathogens (Candida<br>albicans, Pseudomonas aeruginosa,<br>Staphylococcus aureus)<br>Antimicrobial effectiveness for<br>Category 2, bacteria: not less than<br>2.0 log reduction at 14 days and no<br>increase from 14-day count at the<br>28 day count. Yeasts and molds: no<br>increase from the initial calculated<br>count at 14 and 28 days. | Same                                                                                                                                      |
| Viscosity                               | 100-300 cps at 25°C                                                                                                                                                                                                                                                                                                                                                                                                                               | 750-1250                                                                                                                                                                                                                                                                                                                                                                                                                                          | Different                                                                                                                                 |
| Osmolality                              | 750-950 mOsm/kg                                                                                                                                                                                                                                                                                                                                                                                                                                   | 500-625 mOsm/kg (diluted 1:3.9)                                                                                                                                                                                                                                                                                                                                                                                                                   | Different                                                                                                                                 |
| pH                                      | 5.5-7.5                                                                                                                                                                                                                                                                                                                                                                                                                                           | 4.1-4.7                                                                                                                                                                                                                                                                                                                                                                                                                                           | Different                                                                                                                                 |
| Condom<br>Compatibility                 | Not compatible with natural rubber<br>latex, polyurethane, and<br>polyisoprene condoms.                                                                                                                                                                                                                                                                                                                                                           | Not compatible with natural rubber<br>latex, polyurethane, and polyisoprene<br>condoms.                                                                                                                                                                                                                                                                                                                                                           | Same                                                                                                                                      |

{7}------------------------------------------------

K240081 Page 4 of 5

The subject and predicate devices have similar indications for use and have the same intended use - to provide lubrication during intimate sexual activity. The subject and predicate devices have different technological characteristics, including different formulations and device specifications. The different technological characteristics do not raise different questions of safety and effectiveness.

#### VII. Summary of Non-Clinical Performance Testing

### Biocompatibility

Biocompatibility testing on the subject lubricant was performed in accordance with the 2023 FDA quidance document Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices – Part 1: Evaluation and testing within a risk management process" and ISO 10993-1:2009 as follows:

- Cytotoxicity (ISO 10993-5:2009/(R)2014) ●
- Guinea Pig Maximization Sensitization Test (ISO 10993-10: 2021)
- Vaginal Irritation (ISO 10993-23: 2021)
- Acute Systemic Toxicity (ISO 10993-11:2017)

The results of this testing demonstrate that the subject lubricant is non-cytotoxic, nonirritating, non-sensitizing, and not systemically toxic.

# Shelf Life

The subject device is a non-sterile personal lubricant packaged in PE bottles of 100 mL capacity with a 6-month shelf-life in accordance with the results of real time and an accelerated aging study per ASTM F1980-16, Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices. The device specifications listed in Table 1 were tested across the device shelf-life and the subject device met the specifications at all time points.

# Condom Compatibility

Condom compatibility testing was not conducted for the subject device. The subject

{8}------------------------------------------------

device is not compatible with natural rubber latex, polyisoprene, and polyurethane condoms.

#### VIII. Conclusion

The results of the testing described above demonstrate that WLX & Bio Human Lubricant is as safe and effective as the predicate device and supports a determination of substantial equivalence.

---

**Source:** [https://fda-staging.innolitics.com/device/K240081](https://fda-staging.innolitics.com/device/K240081)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).