Mon-a-Therm™ General Purpose Temperature Probe 400TM (90050, 90044)
K240077 · Covidien · FLL · Aug 20, 2024 · General Hospital
Device Facts
Record ID
K240077
Device Name
Mon-a-Therm™ General Purpose Temperature Probe 400TM (90050, 90044)
Applicant
Covidien
Product Code
FLL · General Hospital
Decision Date
Aug 20, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Mon-a-Therm™ General Purpose Temperature Probe 400TM is indicated for use in the routine monitoring of temperature in an anesthetized patient. The device is intended primarily for insertion into the esophagus or rectum, although medical judgment may dictate the selection of other anatomical sites such as the nasopharynx for some patients.
Device Story
The Mon-a-Therm™ General Purpose Temperature Probe 400TM is a sterile, single-use clinical electronic thermometer. It features a thermistor assembly at the distal tip, which senses body temperature when inserted into the esophagus, rectum, or nasopharynx. The probe is connected via a reusable interface cable to a compatible patient monitor (YSI-400 series compatible). The monitor processes the electrical signal from the thermistor to display the patient's temperature. Used in clinical settings (e.g., OR) by healthcare professionals, the device provides continuous temperature data to assist in monitoring anesthetized patients. The probe features a 'Satin Slip' surface for ease of insertion and is constructed from blue, opaque, semi-rigid PVC (phthalate/DEHP-free). Benefits include fully enclosed sensor for patient safety and compatibility with standard multifunction monitors.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench performance testing, including safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), usability (IEC 62366-1), biocompatibility (ISO 10993), and sterilization (ISO 11135).
Technological Characteristics
Thermistor-based temperature probe; YSI-400 series compatible. Materials: Blue, opaque, semi-rigid PVC (phthalate/DEHP-free, RoHS3 compliant). Dimensions: 9 Fr and 12 Fr. Sterilization: Ethylene Oxide. Connectivity: Wired connection to patient monitor via reusable interface cable. Accuracy: ±0.2°C (34°C-42°C range).
Indications for Use
Indicated for routine temperature monitoring in anesthetized patients (adults, pediatric, and neonates). Intended for esophageal, rectal, or nasopharyngeal placement.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
Predicate Devices
Hi-Lo Temp® General Purpose Temperature Probe (K874514)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 20, 2024
Covidien Angelica Medina Regulatory Affairs Specialist 6135 Gunbarrel Avenue Boulder, Colorado 80301
Re: K240077
Trade/Device Name: Mon-a-Therm™ General Purpose Temperature Probe 400TM (90050, 90044) Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical electronic thermometer Regulatory Class: Class II Product Code: FLL Dated: July 16, 2024 Received: July 19, 2024
Dear Angelica Medina:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
David Walloschek
David Wolloscheck, Ph.D. Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors
{2}------------------------------------------------
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K240077
#### Device Name
Mon-a-Therm™ General Purpose Temperature Probe 400TM (90050, 90044)
### Indications for Use (Describe)
The Mon-a-Therm™ General Purpose Temperature Probe 400TM is indicated for use in the routine monitoring of temperature in an anesthetized patient. The device is intended primarily for insertion into the esophagus or rectum, although medical judgment may dictate the selection of other anatomical sites such as the nasopharynx for some patients.
Type of Use (Select one or both, as applicable)
| <span style="font-size:10pt">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="font-size:10pt">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "Medtronic" in blue font. The font is sans-serif and appears to be bolded. The background is white.
## K240077 - Mon-a-Therm™ General Purpose Temperature Probe 400TM 510(k) Summary
This summary of 510(k) safety and effectiveness information for the Mon-a-Therm™ General Purpose Temperature Probe 400TM is submitted in accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with the requirements of 21 CFR §807.92.
## SUBMITTER INFORMATION
### Submitted By:
Covidien, Ilc 6135 Gunbarrel Avenue Boulder, CO 80301
Date Prepared: August 20th, 2024
### Contact Person:
Official Correspondent: Anqelica Medina, Requlatory Affairs Specialist Phone: 303-319-5125 Email: angelica.medina@medtronic.com
## Secondary Correspondent: Greeshma Kayala, Regulatory Affairs Director
Phone: (303) 579-1718 Email: greeshma.kayala@medtronic.com
## DEVICE
| Trade Name: | Mon-a-Therm™ General Purpose Temperature Probe 400TM |
|----------------------------|------------------------------------------------------|
| Common Name: | Clinical electronic thermometer |
| Classification Regulation: | 880.2910 |
| Classification Name: | Clinical Electronic Thermometer |
| Regulatory Class: | Class II |
| Product Code: | FLL |
| Review Panel: | General Hospital |
### PREDICATE DEVICE
| Predicate Manufacturer: | Mallinckrodt Critical Care |
|-------------------------|-----------------------------------------------|
| Predicate Trade Name: | Hi-Lo Temp® General Purpose Temperature Probe |
| Predicate 510(k): | K874514 |
{5}------------------------------------------------
# Medtronic
## DEVICE DESCRIPTION
The subject device of this premarket 510(k) notification is referred to as the Mon-a-Therm™ General Purpose Temperature Probe 400TM throughout this submission.
The Mon-a-Therm™ General Purpose Temperature Probe 400TM is a finished medical device that monitors temperature and are versatile probes that may be used for esophageal, nasopharyngeal or rectal placement.
Features and benefits:
- Fully enclosed sensor helps ensure patient safety
- Satin Slip™ finish for easy insertion ●
- Compatible with most multifunction patient monitors. Refer to Table 1: Summary of Mon-a-. Therm™ 400 Series Thermistor Interface Cables.
Image /page/5/Picture/9 description: The image shows a collection of radioactive seeds used in brachytherapy, a type of radiation therapy. The seeds are small, cylindrical objects attached to thin, flexible wires. These seeds are implanted directly into or near a tumor to deliver a high dose of radiation while minimizing exposure to surrounding healthy tissue. The seeds appear to be grouped together, possibly for storage or preparation before implantation.
Figure 1: Illustration of subject device Mon-a-Therm™ General Purpose Temperature Probe 400TM
The Mon-a-Therm™ General Purpose Temperature Probe 400TM is packaged individually as a sterile, single-use device and is available in the following sizes: 9 Fr 100/case CFN 90050 and 12 Fr 100/case, CFN 90050.The device and its packaging are not made with natural rubber latex or phthalates. The type of probe and device size are designated on the unit package.
The Mon-a-Therm™ General Purpose Temperature Probe 400TM components are illustrated in Figure 2.
Image /page/5/Picture/13 description: The image shows a diagram of a catheter with four different labels. The catheter is labeled with the word "catheter" and has an arrow pointing to the length of the catheter. The numbers 1, 2, 3, and 4 are used to label different parts of the catheter.
Figure 2: Illustration of Subject Device Components
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "Medtronic" in blue font. The font is a sans-serif font and the letters are closely spaced together. The background of the image is white. The word is the logo of the medical technology company Medtronic.
- 1. Tube Blue
- 2. Sleeve GP
- 3. Thermistor Assembly
- 4. Slurry Mixture
The structure of Mon-a-Therm™ General Purpose Temperature Probe 400TM is illustrated in Figure 3 below. The tip of the probe is sealed with PVC and silicone at the end of the thermistor which is where the sealed tube directly contacts the mucosa as shown in "Detail A".
A temperature probe is located near the distal tip. Refer to Figure 3: Structure of Subject Device. The catheter of the subject device features a frosted external surface, referred to as Satin Slip surface. This specialized surface coating extends across the entire external area of the subject device.
Image /page/6/Figure/8 description: This image shows a technical drawing of a thermistor. The overall length of the thermistor is 18 3/4. Detail A shows a close up of the end of the thermistor, which is sealed in a tube with a maximum length of 3/8.
Figure 3: Structure of Subject Device
Each probe is electrically connected to a compatible reusable cable which is specified in Table 1: Summary of Mon-a-Therm™ 400 Series Thermistor Interface Cables and compatible monitors by the connector in the tail of the probe. The reusable connects the probe to a patient monitor, which is compatible with YSI-400 Series thermistor, so that the temperature measurement value is displayed on the screen of monitor. All patient monitors that meet the specifications for YSI-400 thermistor, temperature accuracy, and compatible interface cables are compatible. Refer to Figure 4: Illustration of Patient Monitor Compatibility.
{7}------------------------------------------------
## Table 1: Summary of Mon-a-Therm™ 400 Series Thermistor Interface Cables and compatible monitors
| Product Code | Cable Length | Monitor Compatibility | Monitor<br>Connector |
|--------------|---------------|------------------------------------------------------------------------------------------------------|-------------------------------------|
| 502-0400A | 16 feet/4.87m | | |
| 502-0410A | 10 feet/3.05m | Most patient monitors using series<br>400 thermistor sensors, such as<br>Mindray Passport2™* Monitor | 1/4 inch, right<br>angle phono plug |
| 502-0405A | 16 feet/4.87m | | |
| 502-0415A | 10 feet/3.05m | Siemens Monitor series SC | 7 pin<br>connector |
| 502-0401A | 16 feet/4.87m | | |
| 502-0411A | 10 feet/3.05m | HP or Philips Monitors such as<br>Philips Intellivue™* Monitor | 2 pin<br>connector |
Image /page/7/Figure/4 description: The image shows a patient monitor connected to a temperature probe. The patient monitor displays the temperature as XX.X degC. A reusable cable connects the temperature module to the Mon-a-Therm General Purpose Temperature Probe 400TM.
Figure 4: Illustration of Patient Monitor Compatibility
## INDICATIONS FOR USE
The Mon-a-Therm™ General Purpose Temperature Probe 400TM is indicated for use in the routine monitoring of temperature in an anesthetized patient. The device is intended primarily for insertion into the esophagus or rectum, although medical judgment may dictate the selection of other anatomical sites such as the nasopharynx for some patients.
{8}------------------------------------------------
# Medtroi
## Technological Characteristics
The Mon-a-Therm™ General Purpose Temperature Probe 400TM and its predicate device, utilize the same indications for use, intended purpose, target population, user base and environmental conditions. Compared to the predicate device, the subject device maintains consistency in terms of performance specifications such as measurement range, accuracy, and shelf life.
The subject and predicate device originally adhered to the EN ISO 80601-2-56:2012 standard, requiring a temperature output range of 35°C to 42°C. However, a subsequent standard revision (EN ISO 80601-2-56:2017) expanded the required range to 34°C to 42°C. The subject device meets the stringent accuracy requirements of the newer standard, maintaining a deviation of no more than ±0.2°C within the primary range and ±0.3°C in the extended range. While the predicate device complies with the older standard, the subject device has been updated to align with the latest regulations.
The subject and predicate device utilize PVC material for their tube and sleeve components. though the materials differ in composition. The predicate device employs a clear. 73 Shore A PVC material. In contrast, the subject device features a blue, opaque, semi-rigid PVC material. Notably, the subject device's PVC formulation is not made with DEHP or Phthalates and adheres to RoHS3 compliance standards.
The subject and predicate device utilize PVC material for their tube and sleeve components, though the materials differ in composition. The predicate device employs a clear, 73 Shore A PVC material. In contrast, the subject device features a blue, opaque, semi-rigid PVC material. Notably, the subject device's PVC formulation is not made with DEHP or Phthalates and adheres to RoHS3 compliance standards.
To assess the impact of PVC material used in the tube and sleeve components, temperature accuracy testing was performed on both the subject and predicate device. The results from "Verification Test Report for Temperature Accuracy and Storage Temperature" confirmed that the PVC variations do not compromise temperature accuracy.
These differences do not raise concerns regarding the safety or effectiveness of the subject device when compared to the predicate device as concluded in the Mon-a-Therm General Purpose Temperature Probe Verification and Validation Summary Test Plan and Summary report which demonstrates that these modifications do not have any adverse impact on the performance of the device. Refer to Bench Performance Testing in the below table.
Consequently, it can be affirmed that the Mon-a-Therm™ General Purpose Temperature Probe 400TM is substantially equivalent to the predicate. Hi-Lo Temp® General Purpose Temperature Probe.
## PERFORMANCE DATA
The following performance data were provided to support the substantial equivalence determination with the predicate device.
## Bench Performance Testing
The following benchtop performance testing was provided to support the substantial equivalence determination with the predicate device.
Covidien IIc
{9}------------------------------------------------
| Bench Test<br>Category | Guidance/Standard |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Safety | IEC 60601-1:2005 + AMD1:2012 + AMD2:2020 & EN 60601-1:2006/A2<br>2021<br>General requirements for basic safety and essential performance |
| Safety and<br>performance | ISO 80601-2-56:2017 + A1: 2018 & EN ISO 80601-2-56:2017 + A1:2020<br>Particular requirements for basic safety and essential performance of<br>clinical thermometers for body temperature measurement. |
| Electromagnetic<br>Compatibility | IEC 60601-1-2:2014 + AMD1:2020 & EN 60601-1-2:2015 + A1:2021<br>General requirements for basic safety and essential performance –<br>Collateral Standard: Electromagnetic disturbances – Requirements and<br>tests. |
| Usability | IEC 62366-1:2015 + AMD1:2020 &EN 62366-1:2015 + A1:2020<br>Application of usability engineering to medical devices<br>IEC 60601-1-6:2010 + A1:2013 + A2:2020 & EN 60601-1-6:2010 +<br>A1:2013 + A2:2021<br>General requirements for basic safety and essential performance –<br>Collateral standard: Usability. |
| Biocompatibility | ISO 10993-1:2018 & EN ISO 10993-1:2020<br>Biological evaluation of medical devices — Evaluation and testing within<br>a risk management process.<br>ISO 10993-23:2021 & EN ISO 10993-23:2021<br>Biological evaluation of medical devices — Part 23: Tests for irritation.<br>Device Classification:<br>Contact Type - Surface Medical device, Mucosal Membrane<br>Contact Duration - Prolonged contact, 24 hrs to 30 days<br>Target Population - Adults, Pediatric and Neonates |
| Sterilization | ISO 11135-1:2014 + A1:2018 & EN ISO 11135:2014 + A1:2019<br>Sterilization of health care products – Ethylene Oxide.<br>EN 556-1:2001<br>Sterilization of medical devices. Requirements for medical devices to be<br>designated "STERILE". Requirements for terminally sterilized medical<br>devices. |
# Animal Performance Testing
Not applicable. No animal performance testing was required to demonstrate device safety and effectiveness.
Covidien Ilc
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image contains the word "Medtronic" in a bold, blue font. The word is centered and takes up most of the image. The background is white.
## Clinical Performance Testing
Not applicable. No clinical performance testing was required to demonstrate device safety and effectiveness.
## CONCLUSIONS
Results from comprehensive verification and validating testing demonstrate that the Mon-a-Therm™ General Purpose Temperature Probe 400TM and Hi-Lo Temp® General Purpose Temperature Probe are substantially equivalent with respect to material, technological and performance characteristics.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.