Hair Boom Air / Ulike Hair UpUp / Hair Boom 69 / Hair Boom
K240064 · Wontech Co., Ltd. · OAP · Feb 7, 2024 · Physical Medicine
Device Facts
Record ID
K240064
Device Name
Hair Boom Air / Ulike Hair UpUp / Hair Boom 69 / Hair Boom
Applicant
Wontech Co., Ltd.
Product Code
OAP · Physical Medicine
Decision Date
Feb 7, 2024
Decision
SESE
Submission Type
Special
Regulation
21 CFR 890.5500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Hair Boom Air is indicated to promote hair growth in females with androgenetic alopecia who have Ludwig-Savin Classifications of - II, males who have Norwood-Hamilton Classifications of Ila - V and for both. Fitzpatrick Classification of Skin Phototypes I to IV.
Device Story
Hair Boom Air is a helmet-based diode laser device for home use. It utilizes diode lasers to provide full coverage of the upper 1/3 of the head (scalp). The device operates on a 30-minute therapy cycle. Upon completion of the cycle, the system emits an audible beep pattern to signal the user that therapy is finished and the device is ready to be powered down. The device is intended for OTC use by patients to promote hair growth. It is designed with an outer helmet and a protective inner liner containing foam-lined spacers (polyester/spandex) that contact the user's scalp.
Clinical Evidence
No clinical studies were performed. Substantial equivalence is supported by non-clinical bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and laser safety (IEC 60825-1) standards, alongside software validation and biocompatibility assessments.
Indicated for females with androgenetic alopecia (Ludwig-Savin I-II) and males (Norwood-Hamilton IIa-V) with Fitzpatrick Skin Phototypes I-IV to promote hair growth.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 7, 2024
Won Tech Co., Ltd. Yoon Hyunsik Team Leader 64 Techno8-ro, Yuseong-gu Daejeon, 34028 Korea, South
### Re: K240064
Trade/Device Name: Hair Boom Air / Ulike Hair UpUp / Hair Boom 69 / Hair Boom Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared Lamp Regulatory Class: Class II Product Code: OAP Dated: January 5, 2024 Received: January 9, 2024
Dear Yoon Hyunsik:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).
{2}------------------------------------------------
Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Tanisha L. Hithe Tanisha L. Hithe -S 2024.02.07 20:11:33 -05'00'
Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K240064
Device Name Hair Boom Air / Ulike Hair UpUp / Hair Boom 69 / Hair Boom
Indications for Use (Describe)
The Hair Boom Air is indicated to promote hair growth in females with androgenetic alopecia who have Ludwig-Savin Classifications of - II, males who have Norwood-Hamilton Classifications of Ila - V and for both. Fitzpatrick Classification of Skin Phototypes I to IV.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | |
|----------------------------------------------|--------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | <div> <span> ☒ </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for WONTECH. The logo consists of an orange circle with a stylized "W" inside, followed by the text "WONTECH" in a sans-serif font. The "WON" part of the text is in orange, while the "TECH" part is in gray.
# 510(k) Summary
[As required by 21 CFR 807.92]
#### Date Prepared [21 CFR 807.92(a)(a)] 1.
January 05, 2024
#### Submitter's Information & Contact Person [21 CFR 807.92(a)(1)] 2.
- Name of Manufacturer: WON TECH Co., Ltd.
| - Address: | 64 Techno 8-ro, Yuseong-gu, Daejeon, 34028 Republic of Korea |
|------------------|--------------------------------------------------------------|
| - Contact Name: | Hyun Sik Yoon |
| - Telephone No.: | +82-10-6750-5346 |
| - Fax No.: | +82-70-7836-0110 |
| - Email Address: | yoonhs21@wtlaser.com |
#### 3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]
Common name: Lamp, non-heating, for promotion of hair growth Trade name: Hair Boom 69, Hair Boom Air, Hair Boom, Ulike Hair UpUp
| Classification Description | 21 CFR Section | Product Code |
|----------------------------|----------------|--------------|
| Infrared lamp | 890.5500 | OAP |
As stated in 21 CFR, parts 878.4810, this generic type of the device has been classified as Class II.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for WONTECH. The logo consists of an orange circle with a stylized white "W" inside of it. To the right of the circle is the text "WONTECH" in a sans-serif font. The "WON" portion of the text is in orange, while the "TECH" portion is in gray.
#### 4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)[
The identified predicate devices within this submission are shown as follow:
Predicate device
- . K223862 510(k) Number: ● Applicant: WON Tech Co., Ltd Classification Name: Infrared lamp per 21 CFR 890.5500 ●
- Trade Name: Hair Boom Air, Hair Boom 69, Hair Boom, Ulike Hair UpUp
#### Description of the Device [21 CFR 807.92(a)(4)] ട്.
The Hair Boom Air diode laser is configured within an outer helmet and protective inner liner. The use of diode lasers provides for a full coverage of the upper 1/3 of the head; i.e., the area commonly covered with stylized hair. At the end of the therapy cycle, the system signals that therapy is complete and ready to be powered down, by emitting an audible beep pattern.
#### 6. Indications for Use [21 CFR 807.92(a)(5)]
The Hair Boom 69 is indicated to promote hair growth in females with androgenetic alopecia who have Ludwig-Savin Classifications ofl - II, males who have Norwood-Hamilton Classifications of Ila - V and for both, Fitzpatrick Classification of Skin Phototypes I to IV.
### 7. Determination of Substantial Equivalence [21 CFR 807.92(a)(6) and 21 CFR 807.92(b)]
There are no significant differences between the previous and renewed Hair Boom 69 that would adversely affect the use of the product. It is substantially equivalent to this device in design, function, and technical characteristics.
| | Proposed Device | Predicate Device #1 | SE Decision |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K Number | - | K223862 | N/A |
| Manufacturer | WON TECH Co., Ltd. | WON TECH Co., Ltd. | Same |
| Model Name | Hair Boom Air,<br>Hair Boom 69,<br>Ulike Hair Up Up | Hair Boom Air,<br>Hair Boom 69,<br>Ulike Hair Up Up | Same |
| Indications<br>for Use | The Hair Boom 69 is<br>indicated to promote hair<br>growth in females with<br>androgenetic alopecia who | The Hair Boom 69 is<br>indicated to promote hair<br>growth in females with<br>androgenetic<br>alopecia who | Same |
| | | | |
| | have Ludwig-Savin Classifications ofl - II, males who have Norwood-Hamilton Classifications of Ila - V and for both, Fitzpatrick Classification of Skin Phototypes I to IV. | who have Ludwig-Savin Classifications ofl - II, males who have Norwood-Hamilton Classifications of Ila - V and for both, Fitzpatrick Classification of Skin Phototypes I to IV. | |
| Anatomical<br>site | Head Skin | Head Skin | Same |
| Design | Helmet Design | Helmet Design | Same |
| Clearance | Females & Males OTC | Females & Males OTC | Same |
| Technical<br>characteristics | The Hair Boom 69 diode laser is configured within an outer helmet and protective inner liner. The use of diode lasers provides for a full coverage of the upper 1/3 of the head; i.e., the area commonly covered with stylized hair. At the end of the therapy cycle, the system signals that therapy is complete and ready to be powered down, by emitting an audible beep pattern. | The Hair Boom 69 diode laser is configured within an outer helmet and protective inner liner. The use of diode lasers provides for a full coverage of the upper 1/3 of the head; i.e., the area commonly covered with stylized hair. At the end of the therapy cycle, the system signals that therapy is complete and ready to be powered down, by emitting an audible beep pattern. | Same |
| Wavelength | 650 +/- 5nms | 650 +/- 5nms | Same |
| Laser Mode | 30 minutes | 30 minutes | Same |
| Helmet Design | Image: Black helmet | Image: White and blue helmet | Different<br>Color changed<br>from<br>"White&Blue" to<br>"Black" |
| Controller<br>Design | Image: Black controller | Image: White and blue controller | Different<br>Color changed<br>from<br>"White&Blue" to<br>"Black" |
| Adapter<br>Connector | | | Different<br>Previously, there<br>was separate<br>connector part for<br>connecting the<br>adapter to the<br>controller.<br>However, after the<br>change, the<br>connector is<br>integrated into the<br>adapter, so that the<br>adapter and<br>controller are<br>directly connected. |
| Classification | OAP, Infrared Lamp | OAP, Infrared Lamp | Same |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for WONTECH. The logo consists of an orange stylized letter "W" on the left, followed by the word "WONTECH" in orange and gray. The "WON" part of the word is in orange, while the "TECH" part is in gray.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for WONTECH. The logo consists of an orange circular icon with a stylized "W" inside, followed by the text "WONTECH". The "WON" part of the text is in orange, while the "TECH" part is in gray.
### Non-Clinical Test Summary [21 CFR 807.92(b)(1)]
The WON TECH Co., Ltd. claims conformity with the following Electrical & Mechanical Safety, and Electromagnetic Compatibility standards:
- 1) Electrical Safety, Electromagnetic Compatibility Testing
As aim of this submission is limited to the controller design change and to add a new timer mode, the result of electrical safety test reports and standards were demonstrated in the previous cleared submission under K223862
| Standard<br>(Edition) | Description |
|------------------------------|------------------------------------------------------------------|
| IEC 60601-1:2005/AMD2012 | General requirements for basic safety and essential performance. |
| IEC 60601-1-2:2014 | Electromagnetic disturbances - Requirements and test. |
| IEC 60825-1 Edition 2.0 2007 | Equipment Classifications and Requirements. |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for WONTECH. The logo consists of an orange circle with a stylized "W" inside it, followed by the text "WONTECH" in a sans-serif font. The "WON" part of the text is in orange, while the "TECH" part is in gray.
### 2) Software Validation
The Hair Boom Air contains MINOR level of concern software was designed and developed according to a software development process and was verified and validated.
The software information is provided in accordance with FDA guidance: The content of premarket submissions for software contained in medical devices, on May 11, 2005.
### 3) Biocompatibility
As aim of this submission is limited to the controller design change and to add a new timer mode, the part that contacts the user is same as the previous cleared submission under K223862
| Part | Material | Patient<br>Contact | Duration of Contact<br>by ISO 10993-1 | Bio-<br>compatibility |
|-----------------------------------------|---------------------|--------------------|---------------------------------------|-----------------------|
| Foam lined spacers<br>inside the helmet | Polyester / Spandex | Intact Skin | Limited<br>(< 24 hours) | Yes |
- The material is same as the predicate device (K223862) as they use same helmet design ●
### Clinical Test Summary [21 CFR 807.92(b)(2)]
No clinical studies were considered necessary and performed.
### Conclusion [21 CFR 807.92(b)(3)]
In according with the Federal Food & drug and cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification WON TECH Co., Ltd. concludes that the Hair Boom 69 is substantially equivalent to previous device (Hair Boom 69, K223862) as described herein.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.