ERI Portable X-ray System (CVX-air); ERI Portable X-ray System (CVX-lite); ERI Portable X-ray System (CVX-E)
K234108 · Energy Resources International Co., Ltd. · IZL · Sep 20, 2024 · Radiology
Device Facts
Record ID
K234108
Device Name
ERI Portable X-ray System (CVX-air); ERI Portable X-ray System (CVX-lite); ERI Portable X-ray System (CVX-E)
Applicant
Energy Resources International Co., Ltd.
Product Code
IZL · Radiology
Decision Date
Sep 20, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1720
Device Class
Class 2
Indications for Use
ERI Portable X-ray System is a portable x-ray source with a fixed tube current and voltage for producing x-ray images of extremities using computed radiography (CR) detection plate or digital radiography (DR) detector, and the applicable objects are adults. This product is for use by qualified personnel. It is not intended to replace a radiographic system with variable tube current and voltage (kVp) which may be required for full optimization of image quality and radiation exposure for different exam types.
Device Story
Portable X-ray system (models CVX-air, CVX-lite, CVX-E) generates X-rays for extremity imaging. Device comprises X-ray module, oscillating circuit board, control board, beam limiting device, LCD screen, extension wire control, and 14.4VDC Li-ion polymer battery. Operated by qualified personnel in clinical settings. User selects exposure time via push buttons; system uses fixed tube current and voltage settings (specific to model). Output is X-ray image captured on external CR/DR detector. Provides diagnostic imaging for adult patients; assists clinicians in evaluating extremity anatomy. Benefits include portability for point-of-care imaging.
Clinical Evidence
Clinical trials compared imaging results of the ERI Portable X-ray System against commercially available X-ray devices. Results indicated that image quality produced by the ERI system is on par with comparison devices, supporting substantial equivalence.
Indicated for producing diagnostic X-ray images of extremities in adult patients using CR or DR detectors. For use by qualified personnel. Not intended to replace variable kVp/mA radiographic systems.
Regulatory Classification
Identification
A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 20, 2024
Energy Resources International Co., Ltd. % Pearl Tsai QA & RA Specialist 2F .- 5, No.6-1, Sec.2, Shengyi Rd., Zhubei City Hsinchu County, 302058 TAIWAN
Re: K234108
Trade/Device Name: ERI Portable X-ray System (CVX-air); ERI Portable X-ray System (CVX-lite); ERI Portable X-ray System (CVX-E) Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: Class II Product Code: IZL Dated: December 22, 2023 Received: December 27, 2023
Dear Pearl Tsai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang
Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
K234108
Device Name
ERI Portable X-ray System (CVX-air); ERI Portable X-ray System (CVX-lite);
ERI Portable X-ray System (CVX-E)
Indications for Use (Describe)
ERI Portable X-ray System is a portable x-ray source with a fixed tube current and voltage for producing x-ray images of extremities using computed radiography (CR) detection plate or digital radiography (DR) detector, and the applicable objects are adults. This product is for use by qualified personnel.
It is not intended to replace a radiographic system with variable tube current and voltage (kVp) which may be required for full optimization of image quality and radiation exposure for different exam types.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
{4}------------------------------------------------
The assigned 510(K) number: K234108 Date Prepared: 07/22/2024
#### Submitter Information I.
Submitter: ENERGY RESOURCES INTERNATIONAL CO., LTD. Address: 2F .- 5, No.6-1, Sec.2, Shengvi Rd., Zhubei City, Hsinchu County 302058, Taiwan (R.O.C) Phone Number: +886-3-6585156 Fax Number: +886-3-6585802
Primary Contact Person: Pearl Tsai Email: pearl.tsai@eriray.com Phone Number: +886-3-6585156 ext.112 Fax Number: +886-3-6585802
Secondary Contact Person: Robert Yang
Email: robert@eriray.com Phone Number: +886-3-6585156 ext.100
Fax Number: +886-3-6585802
#### Subject Device II.
Trade Name: ERI Portable X-ray System (CVX-air),
ERI Portable X-ray System (CVX-lite),
ERI Portable X-ray System (CVX-E)
Model: CVX-air, CVX-lite, and CVX-E
Common Name and Classification:
| No. | Product<br>Code | Common Name | Regulator<br>y Class | Regulation Number | Panel |
|-----|-----------------|---------------------|----------------------|-------------------|-----------|
| 1 | IZL | Mobile x-ray system | II | 21 CFR 892.1720 | Radiology |
# III. Predicate Device
| | Predicate Device | | |
|----------------|----------------------------|----------------------------|---------------|
| Subject Device | Predicate Device | Submitter/510(k)<br>Holder | 510(k) Number |
| | EZER Portable X-ray System | Livermoretech | K193535 |
### IV. Device Description
ERI Portable X-ray System generates and controls X-ray with a fixed tube current and voltage (kV). It uses X light source to produce X-ray images of human anatomical structures for diagnosis, and the applicable objects are adults. It operates on 14.4VDC supplied by a rechargeable Lithium-Ion Polymer battery pack.
The ERI Portable X-ray System, is a portable x-ray device that comes in three models: CVX-air, CVX-lite and CVX-E. The differences is as follows:
{5}------------------------------------------------
| Diversity Description | Main Model | Series Model | |
|-------------------------------------------------------------------------------------------|------------|--------------|-------|
| | CVX-air | CVX-lite | CVX-E |
| X-ray tube voltage | 70kV | 67 kV | 60 kV |
| X-ray tube current | 1mA | 0.8mA | 0.6mA |
| Power Consumption | 70 W | 53.6 W | 36W |
| Others are the same as CVX-air | | | |
| The main model and series models have the same internal and external appearance, only the | | | |
| X-ray irradiation parameter settings are different
The X-ray tube head, X-ray controls and power source are assembled in a portable case. The ERI Portable X-ray System is comprised of the following components:
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
- (1) X-ray module
- (2) Oscillating circuit board
- (3) Control board
- (4) Beam limiting device
- (5) LCD screen
- (6) Extension wire control
- (7) rechargeable battery.
#### V. Indications for Use
ERI Portable X-ray System is a portable x-ray source with a fixed tube current and voltage for producing x-ray images of extremities using computed radiography (CR) detection plate or digital radiography (DR) detector, and the applicable objects are adults. This product is for use by qualified personnel.
It is not intended to replace a radiographic system with variable tube current and voltage (kVp) which may be required for full optimization of image quality and radiation exposure for different exam types.
# VI. Comparison of Technology Characteristics with the Predicate Device
Energy Resources International claims that the ERI Portable X-ray System is substantially equivalent to the EZER Portable X-ray System cleared by the FDA in K193535. Energy Resources International claims substantial equivalence because the CVX-air, CVX-lite and CVX-E has a similar intended use, operating principles, and operational specifications as compared to the predicate device.
The specific details regarding the similarities and differences between the ERI Portable X-ray System and EZER Portable X-ray System have been identified and explained in the Comparison Table, see table below. A summary of these similarities and differences is included below. These differences do not present any new issues related to safety and effectiveness.
| Device name | EZER Portable X-ray System<br>(Predicate Device: K193535) | ERI Portable X-ray System<br>(Proposed Device) |
|----------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification,<br>regulation<br>number, product<br>code | Mobile x-ray system,<br>21 CFR 892.1720,<br>IZL | Mobile x-ray system,<br>21 CFR 892.1720,<br>IZL |
| Indications for<br>Use/Intended<br>Use | EZER Portable X-Ray system is a<br>portable x-ray source with a fixed<br>tube current and voltage for<br>producing diagnostic x-ray images<br>of extremities using digital or film<br>image receptors. Its use is<br>intended to be used by trained<br>clinician or technicians for both<br>adult and pediatric subjects age 2<br>and above.<br>It is not intended to replace a<br>radiographic system with variable<br>tube current and voltage (kVp)<br>which may be required for full<br>optimization of image quality and<br>radiation exposure for different<br>exam types. | ERI Portable X-ray System is a<br>portable x-ray source with a fixed<br>tube current and voltage for<br>producing x-ray images of<br>extremities using computed<br>radiography (CR) detection plate or<br>digital radiography (DR) detector,<br>and the applicable objects are<br>adults. This product is for use by<br>qualified personnel.<br>It is not intended to replace a<br>radiographic system with variable<br>tube current and voltage (kVp)<br>which may be required for full<br>optimization of image quality and<br>radiation exposure for different<br>exam types. |
| TECHNOLOGICAL | | |
| Size: Body | 9.2" L x 6.4" W x 4.6" H | 147(H)×211.1(W)×161.7(D) mm |
| Weight | 2.6 kg | 3.2 kg |
| Source to skin<br>distance | 30 cm | 45 cm |
| Focal Spot | 1.2 mm | 0.45 mm |
| Collimator | Four manually and steplessly<br>adjustable shutters with light beam<br>type central x-ray indicator<br>(Advantech R72) | Stepless manual button control,<br>using a motor to drive the light gate<br>to adjust the X-ray irradiation range. |
| Triggering<br>Mechanism | Two stage triggering | Two stage triggering |
| User Interface | Up-down push buttons for kVp<br>selections and exposure time<br>selections with LED indicators<br>and mAs indicators. | Up-down push buttons for exposure<br>time selection to adjust mAs, LED<br>on/off buttons, and collimator<br>horizontal/vertical control buttons. |
| Energy Source | Rechargeable 22.2 V DC<br>Lithium Ion Polymer battery pack | Rechargeable 14.4 V DC<br>Lithium Ion Polymer battery pack |
| Exposure Time | 0.03~1.30 seconds in 0.01<br>increments | 0.01~1.00 seconds in 0.01<br>increments |
| mA | 2.0 mA fixed | 1.0 mA fixed / 0.8 mA fixed / 0.6<br>mA fixed |
| kVp | 60 kVp fixed | 70 kVp fixed / 67 kVp fixed / 60<br>kVp fixed |
{6}------------------------------------------------
The subject device is similar to the predicate device in terms of the indications for use and technological application. Both the subject and predicate devices are portable X-ray system for taking diagnostic X-rays of human anatomy using a fixed tube current and voltage (kVp). Other differences in device design such as exposure time, size, and user interface.
{7}------------------------------------------------
### VII. Non-clinical Test Data
Testing was performed in accordance with the following international standards:
- IEC 62304 Edition 1.1 [2015-06] Medical device software Software life cycle processes ●
- IEC 60601-1 Edition 3.2 [2020-08 CONSOLIDATED VERSION] Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2 Edition 4.1 [2020-09 CONSOLIDATED VERSION] Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance -Collateral Standard: Electromagnetic Compatibility - Requirements And Tests
- IEC 60601-1-3 Edition 2.2 [2021-01 CONSOLIDATED VERSION] Medical electrical equipment - Part 1-3: General requirements for basic safety and essential performance -Collateral Standard: Radiation protection in diagnostic X-ray equipment
- IEC 60601-2-54 Edition 2.0 [2022-09] Medical electrical equipment Part 2-54: Particular ● requirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy
- ISO14971 Third Edition [2019-12] Medical devices - Applications of risk management to medical devices
- 21 CFR 1020.30: Diagnostic x-ray system and their major components ●
- 21 CFR 1020.31: Radiographic Equipment
# VIII. Clinical Performance Data
The ERI Portable X-ray System has undergone comprehensive non-clinical design verification and validation testing, including assessments of Electrical Safety and performance. The results confirm that the device's design is appropriate and effective for its intended use. Additionally, clinical trials were conducted to compare the imaging results of the ERI Portable X-ray System with those obtained from commercially available X-ray devices. These trials demonstrated that the image quality produced by the ERI system is on part with that of the comparison devices, supporting its substantial equivalence as a portable X-ray system.
# IX. Conclusions
The Energy Resources International ERI Portable X-ray System has the same intended use and same basic technology as the predicate device, thus is able to achieve the same effectiveness and safety as the predicate device. The ERI Portable X-ray System contains in some combination similar features and design as the predicates. Other differences include device design such as exposure time, size and user interface. The subject device is substantially equivalent to the predicate device with its intended use, mechanical and electrical performance as described.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.