K234098 · Yongkang Yile Vehicle Co., Ltd. · INI · Apr 18, 2024 · Physical Medicine
Device Facts
Record ID
K234098
Device Name
Mobility Scooter (C4+)
Applicant
Yongkang Yile Vehicle Co., Ltd.
Product Code
INI · Physical Medicine
Decision Date
Apr 18, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3800
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The mobility scooter is intended for providing assistance to people restricted to a sitting position.
Device Story
Mobility Scooter (C4+) is a motorized three-wheeled vehicle designed to assist individuals restricted to a sitting position. The device consists of a steel frame, four wheels (two front, two rear), a seat, a tiller console, an electric motor, an electromagnetic brake, and lead-acid batteries. The user operates the scooter via the tiller console to navigate. The device is intended for use on flat surfaces; it provides mobility by converting electrical energy from batteries into mechanical motion via the motor. The electromagnetic brake provides stopping capability. The device is intended for over-the-counter use by the patient. It benefits the patient by providing independent mobility and assistance for those with restricted sitting capabilities.
Clinical Evidence
No clinical data. Bench testing only. Performance testing conducted per ISO 7176 series (stability, braking, energy consumption, dimensions, speed, strength, climatic, obstacle climbing, power/control systems, etc.). Biocompatibility testing (cytotoxicity, intracutaneous reactivity, sensitization) per ISO 10993-1. EMC testing per ISO 7176-21, IEC 60601-1-2, and IEC TR 60601-4-2. Package validation per ASTM D4169-16. Usability and human factors testing performed per FDA guidance.
Technological Characteristics
Motorized 3-wheeled vehicle. Materials: Cold rolled steel/seamless steel pipe frame; PVC seat/armrest/handlebar/foot pedal. Power: 24V DC 950W motor, 12V lead-acid batteries. Electromagnetic braking. Macpherson suspension. Dimensions: 1366mm x 665mm x 1323mm. Connectivity: None. Standards: ISO 7176 series (performance), ISO 10993 (biocompatibility), IEC 60601-1-2 (EMC).
Indications for Use
Indicated for individuals restricted to a sitting position requiring mobility assistance. No specific age or gender contraindications provided.
Regulatory Classification
Identification
A motorized three-wheeled vehicle is a gasoline-fueled or battery-powered device intended for medical purposes that is used for outside transportation by disabled persons.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name, "U.S. Food & Drug Administration," written out next to it.
April 18, 2024
Yongkang Yile Vehicle Co., Ltd. % Evan Hu Technical and Regulatory Director Shanghai Mind-link Consulting Co., Ltd. 377 Tianzhu Road, Jiading Shanghai, 201801 China
Re: K234098
Trade/Device Name: Mobility Scooter (C4+) Regulation Number: 21 CFR 890.3800 Regulation Name: Motorized Three-Wheeled Vehicle Regulatory Class: Class II Product Code: INI Dated: December 22, 2023 Received: December 26, 2023
Dear Evan Hu:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Tushar Bansal -S
for Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Rehabilitation Devices
{2}------------------------------------------------
OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K234098
Device Name Mobility Scooter (C4+)
Indications for Use (Describe)
The mobility scooter is intended for providing assistance to people restricted to a sitting position.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### 510(k) Summary
#### K234098
#### 1. Preparation date: Dec. 12th, 2023
#### 2. Submitter
Manufacturer: YONGKANG YILE VEHICLE CO., LTD. Address: No.46 Zerong Road, Dahou Industrial Zone, Tangxian Town, Yongkang, Zhejiang, 321314, PRC Contact person: David SHI, Chief Manager, +86 15158900066, 41724003@gq.com Submission correspondent: Evan Hu, Technical and Regulatory Director, +86-18616124827, evan.ww.hu@outlook.com
### 3. Device
Trading name: Mobility Scooter (C4+) Common name: Scooter Regulation No.: 21 CFR 890.3800 Classification name: Vehicle, Motorized 3-Wheeled Classification: Class II Product code: INI
### 4. Predicate device
Predicate device: K132855 Trading name: HEARTWAY Power Mobility Scooter, S15 Common name: Scooter Regulation No.: 21 CFR 890.3800 Classification name: Vehicle, Motorized 3-Wheeled Classification: Class II Product code: INI
### 5. Device description
The mobility scooter mainly consists of steal frame, two front wheels, two rear wheels, a seat, a tiller console, an electric motor, an electromagnetic brake, front lights, two rechargeable Lead-acid Batteries, and a battery charger. The product is intended to be used for assisting people restricted to a sitting position. It is driven by the end users for movement. It is compact, maneuverable but not necessarily able to overcome
{5}------------------------------------------------
obstacles. The maximum load is 150 kg. It is powered by rechargeable batteries with 25.3 km theoretical distance range which maximum speed up to 10.5 km/h (tested value).
The scooter is highly recommended to be driven on flat roads. Frequent usage of the scooter on slopes, rough terrain, or when climbing curbs will result in a reduction of driving distance.
# 6. Indications for use/Intended use
The mobility scooter is intended for providing assistance to people restricted to a sitting position.
# 7. Comparison of technological characters between proposed and predicate devices
| Items | Proposed device<br>(K-234098-Mobility Scooter, C4+) | Predicate device<br>(K132855-HEARTWAY Mobility Scooter, S15) | Comments |
|------------------------------------------------|-----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------|
| Indications for use/Intended use | The mobility scooter is intended for providing assistance to people restricted to a sitting position. | The device is intended for medical purposes to provide mobility to persons restricted to a sitting position. | #1 |
| Product code | INI | INI | Same |
| Frame and material | Folded, Cold rolled steel and seamless steel pipe | Folded, Carbon Steel Pipe | #2 |
| Overall dimension | Length * width * height:<br>1366mm * 665mm * 1323mm | Length * width * height:<br>1400mm * 700mm * 1360mm | #2 |
| Seat width | 460mm | 620mm | #2 |
| Seat height | 660mm | 820mm | #2 |
| Max occupant mass/Weight limit | 150kg | 160kg | #2 |
| Overall weight | w/ batteries 130Kg (w/o package)<br>w/o batteries 105Kg (w/o package) | w/ batteries 136Kg<br>w/o batteries 98Kg | #2 |
| Max speed | 10.5km/h (tested value)<br>12km/h (set value) | 9.6km/h | #2 |
| Batteries | Lead acid battery<br>Four batteries in two packs, 12V | Lead acid battery<br>Two batteries packs, 12V DC, 50Ah. | #2 |
| | DC, 20Ah. | | |
| Range per charge/<br>Energy consumption | 25.3km | 40km | #2 |
| Motor | 24V DC, 950W | 24V DC, 700 W | #2 |
| Controller | PG S-DRIVE | RHINO DS120 Drive | #6 |
| Brake | Electromagnetic brake | Electromagnetic brake | Same |
| Barking time and<br>distance | <1s, 2.8m (min value at max forward<br>speed on flat floor) | Not mentioned. | #2 |
| Charger output | 24V DC, 6A<br>(IEC 60335-2-29 and ISO 7176-25<br>certified) | 24V DC 4F24050<br>(UL E241359 certified) | Same |
| Rear wheel | 320mm * 100mm, pneumatic* 2 | 13" * 3.5" (330mm*89mm) solid * 2 | #2 |
| Front wheel | 320mm * 100mm, pneumatic* 2 | 13" * 5.0" (330mm*127mm) solid * 2 | #2 |
| Suspension | Macpherson suspension | Cross brace | #2 |
| Incline | 9 degrees | 10 degrees | #2 |
| Turning radius | 1800mm | 1280mm | #2 |
| Ground clearance | 1000mm | 750mm | #2 |
| Kerb climbing ability | 100mm | 80mm | #2 |
| Armrest type | Flip backward | Flip backward | Same |
| Wheel lock | Push to lock | Push to lock | Same |
| Warranty | 1 year: Main frame, controller,<br>motor, battery. | 3 years: Main frame<br>1 year: Controller / gear motor /<br>batteries w/o exhaustive and wear<br>parts | #3 |
| User contacting<br>components and<br>materials | Foot pedal: PVC sheet<br>Handlebar: PVC<br>Seat, Seat belt: PVC fabrics<br>Backrest, Armrest: PVC fabrics | Foot pedal: ABS<br>Handlebar: PVC materials<br>Seat, Seat belt: PVC materials<br>Backrest, Armrest: PVC materials | #4 |
| Biocompatibility | Complied with ISO 10993 serials | Complied with ISO 10993 serials | Same |
| Performance | Meet requirements of ISO 7176<br>serials | Meet requirements of ISO 7176<br>serials | Same |
| EMC | Meet requirements of<br>ISO 7176-21:2009,<br>IEC 60601-1-2: 2014+A1:2020,<br>IEC TR 60601-4-2:2016. | Meet requirements of<br>ISO 7176-21:2009,<br>IEC 60601-1-2,<br>ANSI / RESNA WC/Vol.2: 2009,<br>CISPR 11: 2004+A2:2006,<br>EN 61000-4-2: 2008,<br>IEC 61000-4-3:2006,<br>IEC 61000-4-8: 2001 | #5 |
Table 1. Characters comparison
{6}------------------------------------------------
Comments:
#1: The proposed and predicate devices have different wording in their Indications for Use,
{7}------------------------------------------------
but the meanings are the same and this does not affect the device's safety and effectiveness, demonstrating the substantial equivalence.
#2: The proposed device has different specifications for scooter components and performance compared to the predicate device. It underwent performance testing according to ISO 7176 series, and the results demonstrated that it met all the standard requirements. Therefore, both the proposed and predicate devices meet the performance testing standards, demonstrating the substantial equivalence.
#3: The main frame of the proposed device has a shorter warranty period compared to the predicate device. However, the warranty period for other components remains unchanged. This discrepancy in warranty period does not raise any safety or effectiveness concerns, demonstrating the substantial equivalence.
#4: The user-contact components only differ in the material used for the foot pedal, while the direct contact components remain unchanged. Biocompatibility testing has confirmed the long-term safety of use, demonstrating the substantial equivalence.
#5: The proposed and predicate devices for EMC testing both used the ISO 7176-21 standard. In addition, the proposed device underwent supplementary testing in accordance with the IEC 60601-1-2 and IEC 60601-4-2 standards, and both of these two standards encompass the testing methods outlined in the IEC 61000 series. The ANSI / RESNA WC/Vol.2 is the consensus standard of IEC 60601-1-2, Consequently, the use of different standards does not affect the EMC testing process, demonstrating the substantial equivalence.
#6: The proposed and predicate devices utilized distinct controllers with differing firmware configurations. Specifically, the proposed device employed the PG S-Drive controller, as cleared in K150987 and K191256. Performance testing of the controller was carried out in accordance with ISO 7176-14, yielding results that confirm adherence to the relevant standards. As a result, the utilization of different controllers has no bearing on the safety and effectiveness of the device, thereby demonstrating the substantial equivalence.
# 8. Non-clinical testing
# PERFORMANCE TESTING
- (1) EMC report was conducted according to ISO 7176-21:2009, IEC 60601-1-2: 2014+A1:2020, and IEC TR 60601-4-2:2016.
- (2) Physical properties were conducted according to the following standards serials.
- ISO 7176-1:2014, Static stability
- ISO 7176-2:2017, Dynamic stability ।
- ISO 7176-3: 2012, Effectiveness of brakes -
- -ISO 7176-4:2008, Energy consumption and theoretical distance range
- -ISO 7176-5:2008, Dimensions, mass and turning space
{8}------------------------------------------------
- ISO 7176-6:2018, Maximum speed, acceleration and deceleration -
- -ISO 7176-7:1998, Seating and wheel dimensions
- ISO 7176-8:2014, Static, Impact and fatigue strengths -
- ISO 7176-9:2009, Climatic tests -
- -ISO 7176-10:2008, Obstacle-climbing ability
- -ISO 7176-11:2012, Test dummies
- ISO 7176-13:1989, Coefficient of friction -
- ISO 7176-14:2022, Power and control systems -
- -ISO 7176-15:1996, Information disclosure, documentation and labelling
- -ISO 16840-10:2021, Resistance to ignition
- ISO 7176-22: 2014, Set up procedure -
- ISO 7176-25: 2013, Battery and charger -
## BIOCOMPATIBILITY TESTING:
The proposed device was tested in compliance with the FDA Guidance document Use of International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process", as the Surface device- Intact skin-Long term contact (>30 days).
The items listed below underwent testing in accordance with ISO 10993-1.
- Cytotoxicity .
- Intracutaneous Reactivity -
- -Sensitization
# PACKAGE VALIDATION
Package integrity testing under simulated shipping conditions was conducted to satisfy the requirements in ASTM D4169-16 Standard Practice for Performance Testing of Shipping Containers and Systems. All packaging was deemed acceptable for protection of product.
### USABILITY AND HUMAN FACTOR
A usability and human factor testing was conducted to assess the usability of the scooter from the users' perspective by following the FDA guidance-Applying Human Factors and Usability Engineering to Medical Devices. The results of observation and questionnaire indicated that users were generally satisfied with the scooter's usage and new risks were identified to be monitored.
### 9. Clinical testing
Not applicable for this submission.
{9}------------------------------------------------
## 10. Conclusion
The differences between the predicate and the proposed device do not raise any new or different questions of safety or effectiveness. The proposed device is substantially equivalent to the predicate device with respect to indications for use and technological characteristics.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.