K234070 · Dexcom, Inc. · SAF · Mar 5, 2024 · Clinical Chemistry
Device Facts
Record ID
K234070
Device Name
Stelo Glucose Biosensor System
Applicant
Dexcom, Inc.
Product Code
SAF · Clinical Chemistry
Decision Date
Mar 5, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1355
Device Class
Class 2
Indications for Use
The Stelo Glucose Biosensor System is an over-the-counter (OTC) integrated Continuous Glucose Monitor (iCGM) intended to continuously measure, record, and display glucose values in people 18 years and older not on insulin. The Stelo Glucose Biosensor System helps to detect normal (euglycemic) and low or high (dysglycemic) glucose levels. The Stelo Glucose Biosensor System may also help the user better understand how lifestyle and behavior modification, including diet and exercise, impact glucose excursion. The user is not intended to take medical action based on the device output without consultation with a qualified healthcare professional.
Device Story
Stelo Glucose Biosensor System is an OTC iCGM for non-insulin-using adults; measures interstitial fluid glucose via subcutaneous enzymatic electrochemical sensor (glucose oxidase). Wearable subsystem (sensor, transmitter, patch) converts glucose to electrical current; transmitter processes signal via onboard algorithm to estimated glucose values (EGV). Data transmitted via Bluetooth to smartphone app; app displays glucose readings and trend graphs; provides insights on time-in-range. No alerts/alarms; no insulin device integration. Used in home setting by patient. Benefits include understanding lifestyle/diet/exercise impact on glucose. Healthcare providers use output for patient counseling; patients use for behavioral modification.
Clinical Evidence
Pivotal clinical study (N=130) evaluated safety and effectiveness over 15.5 days. Accuracy compared to YSI 2300 STAT Plus analyzer. Overall MARD was 8.5% (70-180 mg/dL range). Percent within 20% of comparator across 40-400 mg/dL range was 93.1%. Precision study (N=18) showed 8.8% mean paired absolute relative difference and 6.2% CV. Sensor survival rate was 73.5% at 15 days in initial study; subsequent study with improved patch showed 77.9% survival. Data availability rate remained >97% throughout wear.
Technological Characteristics
Integrated CGM; wire-type subcutaneous sensor; BLE transmitter; factory-calibrated. 15-day sensor life. Materials/hardware identical to predicate G7. Software includes firmware for signal processing and mobile app for display. Cybersecurity controls implemented per FDA guidance. Sterilization validation performed.
Indications for Use
Indicated for people 18 years and older not on insulin to continuously measure, record, and display glucose values, detect euglycemic/dysglycemic levels, and monitor impact of lifestyle/behavior modifications on glucose excursions.
Regulatory Classification
Identification
An integrated continuous glucose monitoring system (iCGM) is intended to automatically measure glucose in bodily fluids continuously or frequently for a specified period of time. iCGM systems are designed to reliably and securely transmit glucose measurement data to digitally connected devices, including automated insulin dosing systems, and are intended to be used alone or in conjunction with these digitally connected medical devices for the purpose of managing a disease or condition related to glycemic control.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include the following:
(i) Robust clinical data demonstrating the accuracy of the device in the intended use population.
(ii) The clinical data must include a comparison between iCGM values and blood glucose values in specimens collected in parallel that are measured on an FDA-accepted laboratory-based glucose measurement method that is precise and accurate, and that is traceable to a higher order (
*e.g.,* an internationally recognized reference material and/or method).(iii) The clinical data must be obtained from a clinical study designed to fully represent the performance of the device throughout the intended use population and throughout the measuring range of the device.
(iv) Clinical study results must demonstrate consistent analytical and clinical performance throughout the sensor wear period.
(v) Clinical study results in the adult population must meet the following performance requirements:
(A) For all iCGM measurements less than 70 milligrams/deciliter (mg/dL), the percentage of iCGM measurements within ±15 mg/dL of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 85 percent.
(B) For all iCGM measurements from 70 mg/dL to 180 mg/dL, the percentage of iCGM measurements within ±15 percent of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 70 percent.
(C) For all iCGM measurements greater than 180 mg/dL, the percentage of iCGM measurements within ±15 percent of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 80 percent.
(D) For all iCGM measurements less than 70 mg/dL, the percentage of iCGM measurements within ±40 mg/dL of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 98 percent.
(E) For all iCGM measurements from 70 mg/dL to 180 mg/dL, the percentage of iCGM measurements within ±40 percent of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 99 percent.
(F) For all iCGM measurements greater than180 mg/dL, the percentage of iCGM measurements within ±40 percent of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 99 percent.
(G) Throughout the device measuring range, the percentage of iCGM measurements within ±20 percent of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 87 percent.
(H) When iCGM values are less than 70 mg/dL, no corresponding blood glucose value shall read above 180 mg/dL.
(I) When iCGM values are greater than 180 mg/dL, no corresponding blood glucose value shall read less than 70 mg/dL.
(J) There shall be no more than 1 percent of iCGM measurements that indicate a positive glucose rate of change greater than 1 mg/dL per minute (/min) when the corresponding true negative glucose rate of change is less than −2 mg/dL/min as determined by the corresponding blood glucose measurements.
(K) There shall be no more than 1 percent of iCGM measurements that indicate a negative glucose rate of change less than −1 mg/dL/min when the corresponding true positive glucose rate of change is greater than 2 mg/dL/min as determined by the corresponding blood glucose measurements.
(vi) Data demonstrating similar accuracy and rate of change performance of the iCGM in the pediatric population as compared to that in the adult population, or alternatively a clinical and/or technical justification for why pediatric data are not needed, must be provided and determined by FDA to be acceptable and appropriate.
(vii) Data must demonstrate that throughout the claimed sensor life, the device does not allow clinically significant gaps in sensor data availability that would prevent any digitally connected devices from achieving their intended use.
(2) Design verification and validation must include a detailed strategy to ensure secure and reliable means of iCGM data transmission to provide real-time glucose readings at clinically meaningful time intervals to devices intended to receive the iCGM glucose data.
(3) Design verification and validation must include adequate controls established during manufacturing and at product release to ensure the released product meets the performance specifications as defined in paragraphs (b)(1) and (b)(2) of this section.
(4) The device must demonstrate clinically acceptable performance in the presence of clinically relevant levels of potential interfering substances that are reasonably present in the intended use population, including but not limited to endogenous substances and metabolites, foods, dietary supplements, and medications.
(5) The device must include appropriate measures to ensure that disposable sensors cannot be used beyond its claimed sensor wear period.
(6) Design verification and validation must include results obtained through a usability study that demonstrates that the intended user can use the device safely and obtain the expected glucose measurement accuracy.
(7) The labeling required under § 809.10(b) of this chapter must include a separate description of the following sensor performance data observed in the clinical study performed in conformance with paragraph (b)(1) of this section for each intended use population, in addition to separate sensor performance data for each different iCGM insertion or use sites (
*e.g.,* abdomen, arm, buttock):(i) A description of the accuracy in the following blood glucose concentration ranges: less than 54 mg/dL, 54 mg/dL to less than 70 mg/dL, 70 to 180 mg/dL, greater than 180 to 250 mg/dL, and greater than 250 mg/dL.
(ii) A description of the accuracy of positive and negative rate of change data.
(iii) A description of the frequency and duration of gaps in sensor data.
(iv) A description of the true, false, missed, and correct alert rates and a description of the available glucose concentration alert settings, if applicable.
(v) A description of the observed duration of iCGM life for the device.
Predicate Devices
Dexcom G7 Continuous Glucose Monitoring System (K231081)
Submission Summary (Full Text)
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY AND INSTRUMENT
## I Background Information:
A 510(k) Number
K234070
B Applicant
Dexcom, Inc.
C Proprietary and Established Names
Stelo Glucose Biosensor System
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| SAF | Class II | 21 CFR 862.1355 - Integrated Continuous Glucose Monitoring System | CH - Clinical Chemistry |
## II Submission/Device Overview:
A Purpose for Submission:
New device
B Measurand:
Glucose in Interstitial Fluid
C Type of Test:
Quantitative, amperometric assay (Glucose Oxidase)
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
K234070 - Page 2 of 21
## III Intended Use/Indications for Use:
### A Intended Use(s):
See Indications for Use below.
### B Indication(s) for Use:
The Stelo Glucose Biosensor System is an over-the-counter (OTC) integrated Continuous Glucose Monitor (iCGM) intended to continuously measure, record, analyze, and display glucose values in people 18 years and older not on insulin. The Stelo Glucose Biosensor System helps to detect normal (euglycemic) and low or high (dysglycemic) glucose levels. The Stelo Glucose Biosensor System may also help the user better understand how lifestyle and behavior modification, including diet and exercise, impact glucose excursion.
The user is not intended to take medical action based on the device output without consultation with a qualified healthcare professional.
### C Special Conditions for Use Statement(s):
OTC - Over The Counter
No MRI/CT/diathermy — MR unsafe: Don't wear any Stelo Glucose Biosensor system component during magnetic resonance imaging (MRI) or high-frequency electrical heat (diathermy) treatment. However, it's safe to have a CT scan if you keep the sensor out of the scanned area and cover the sensor with a lead apron during the scan. The Stelo Glucose Biosensor system hasn't been tested in those situations when used during an MRI scan, diathermy, or in the scanned area of a CT scan. The magnetic fields and heat could damage Stelo Glucose Biosensor system components, which may cause inaccurate sensor readings.
With the Stelo Glucose Biosensor system, you can take a standard or maximum acetaminophen dose of 1 gram (1,000 mg) every 6 hours and still use the Stelo Glucose Biosensor system. Taking higher than the maximum dose of acetaminophen (e.g. >1 gram every 6 hours in adults) may affect the sensor readings and make them look higher than they really are.
If you are taking hydroxyurea, your sensor readings will be higher than your actual glucose. The level of inaccuracy depends on the amount of hydroxyurea in your body. Talk to your physician about alternative glucose monitoring approaches.
Don't use if you have problematic hypoglycemia. The Stelo Glucose Biosensor system hasn't been designed for these populations. Consult with your healthcare provider to discuss which Dexcom product is right for you.
Don't use if you are on dialysis. The Stelo Glucose Biosensor system performance hasn't been evaluated in these populations and sensor readings may be inaccurate.
Only insert your sensor on the back of your upper arm. Don't wear your sensor on other sites (such as your abdomen), as it may not work as expected.
### D Special Instrument Requirements:
Not applicable.
{2}
K234070 - Page 3 of 21
## IV Device/System Characteristics:
### A Device Description:
The Stelo Glucose Biosensor System System is a home use device that is intended to continuously measure the glucose in the interstitial fluid, calculate the glucose reading and make this value available to the user. The Stelo System can reliably and securely transmit glucose measurement data to authorized digitally connected devices. The Stelo System is not intended to be used in conjunction with insulin devices such as insulin pens and Automated Insulin Dosing (AID) systems.
The Stelo System is based on the same principle of operation, fundamental design, and physical characteristics as the G7 CGM System, Dexcom’s most recent prescription integrated continuous glucose monitoring (iCGM) system (predicate device: K231081). The Stelo System comprises the following two subsystems:
1. **Glucose Sensing Subsystem (GSS): Wearable (Sensor, Transmitter, and Patch) and Applicator**
The sensor is a small and flexible wire inserted by the applicator into subcutaneous tissue where it converts glucose into electric current. The transmitter is pre-connected to the sensor and is worn on the body by the adhesive patch. The transmitter measures the electric current produced by the sensor and converts these measurements into estimated glucose values (EGV) using an onboard algorithm. The transmitter sends glucose data to a mobile application.
Each GSS box also includes an overpatch, which is a general adhesive tape that helps with the adhesion of the wearable to the user’s body.
2. **Mobile Applications Subsystem (MAS): iOS and Android Mobile Application (App)**
The App provides in-app guidance for the user on how to apply and set up the wearable and how to create a user account. Once the set-up has been completed, the App receives information from the transmitter and acts as a user interface by indicating the system state (e.g., warm-up period, signal loss, etc.) and displaying glucose readings and trend graphs.
### Stelo Glucose Sensing Subsystem (GSS)
The Stelo GSS has the same hardware design as the predicate device (G7 GSS); the devices use the same sensor, patch, applicator, and transmitter hardware. In addition, the Stelo transmitter firmware has the same primary responsibilities as the predicate G7 transmitter firmware.
While the firmware modes, firmware modules, wireless communication requirements, and algorithm inputs/outputs of the Stelo transmitter firmware remain the same as the predicate device, the Stelo transmitter firmware includes the following differences:
- No optional calibration
- The Stelo transmitter firmware is factory calibrated only and does not use any Blood Glucose (BG) entered by the lay user for calibration.
- Device connectivity restrictions
- The Stelo transmitter firmware does not pair/connect to an additional dedicated primary display device (e.g., a receiver), insulin pens, AID systems, or other
{3}
unauthorized devices. The Stelo transmitter can only pair/connect to the user's (personal) smartphone.
- Extended (15.5 days) device wear duration
- The Stelo transmitter firmware supports a sensor wear duration of 15 days, compared to 10 days for the predicate.
- Algorithm modifications
- The algorithm was modified to support the extended wear duration while maintaining the accuracy of the device. Additional algorithm modifications and parameter optimizations were also introduced to help improve the user experience (e.g., reducing jitteriness of glucose readings displayed in the trend graph and better handling of edge cases or specific scenarios users may encounter in the field).
# Stelo Mobile Application System (MAS)
The Stelo MAS consists of the Stelo app that is available on both iOS and Android platforms. The app can be downloaded to a compatible, Bluetooth Low Energy (BLE)-enabled smart device. The Stelo app was developed using the same software development kit (SDK) as the predicate device (G7 app) and inherited the main functionality and UI elements: in-app onboarding, current glucose value, trend graph, in-app Clarity card, event logging, and connectivity to Dexcom's cloud servers. The following main UI elements were tailored to the Stelo user population:
- Stelo app onboarding
- Similar to the predicate device, the Stelo app includes an onboarding experience to guide users through inserting the wearable and setting up the app. Instructions and safety information are tailored to the Stelo user population.
- App update interval
- The Stelo app wirelessly receives glucose information from the transmitter every five minutes. All the sensor readings are then displayed on the mobile app every 15 minutes
- Narrowed displayed glucose range
- Similar to the predicate device, the sensor measures glucose values within the 40-400 mg/dL range and sends the glucose data to the Stelo app. The Stelo app then narrows the displayed glucose values to 70-250 mg/dL. The displayed glucose range is intended to provide the glucose range pertinent to the Stelo user population who may not often have glucose values less than 70 mg/dL and more than 250 mg/dL. Glucose values below this range are reported as “Below 70 mg/dL” and glucose values above this range are reported as “Above 250 mg/dL.”
- Insights
- The user can review their latest time in range (day by day and at the end of a session) through "Insights," a feature that calculates how much time the user's glucose was in range during the preceding day(s). The Insights also present suggestions for how the user can increase the time in range. Insights are delivered to the user as an out-of-app notification and can also be accessed from within the Stelo app.
- Alerts removal
- Glucose alerts and alarms (e.g., Urgent Low Alarm, Urgent Low Soon, Low, High, Rising Fast, and Falling Fast Alerts) are not included in the Stelo app,
K234070 - Page 4 of 21
{4}
unlike the predicate G7 app. The Stelo app provides lock screen and in-app visual notifications to the user such as Bluetooth off, Signal Loss, or Sensor Expired.
## B Principle of Operation:
The Stelo Glucose Biosensor System detects glucose levels from the fluid just beneath the skin (interstitial fluid). The sensor probe continuously measures glucose concentration in the interstitial fluid via an enzymatic electrochemical reaction using glucose oxidase. The enzyme, glucose oxidase, catalyzes the oxidation of glucose and produces hydrogen peroxide. The production of hydrogen peroxide generates an electrical current that is proportionate to the interstitial glucose concentration. The transmitter converts the signal using an algorithm to a glucose value read in $\mathrm{mg/dL}$, which is then transmitted to the mobile application for the user to see and use accordingly.
## C Instrument Description Information:
1. Instrument Name:
Stelo Glucose Biosensor System
2. Specimen Identification:
Not applicable.
3. Specimen Sampling and Handling:
Not applicable.
4. Calibration:
The Stelo Glucose Biosensor System is factory calibrated. Users may not enter optional calibrations based on fingerstick blood glucose values.
5. Quality Control:
Not applicable.
This medical device product has functions subject to FDA premarket review as well as functions that are not subject to FDA premarket review. For this application, if the product has functions that are not subject to FDA premarket review, FDA assessed those functions only to the extent that they either could adversely impact the safety and effectiveness of the functions subject to FDA premarket review or they are included as a labeled positive impact that was considered in the assessment of the functions subject to FDA premarket review.
## V Substantial Equivalence Information:
## A Predicate Device Name(s):
Dexcom G7 Continuous Glucose Monitoring (CGM) System
K234070 - Page 5 of 21
{5}
B Predicate 510(k) Number(s):
K231081
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K234070 | K231081 |
| --- | --- | --- |
| Device Trade Name | Stelo Glucose Biosensor System | Dexcom G7 Continuous Glucose Monitoring (CGM) System |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | Automatically measure glucose in bodily fluids continuously for a specified period of time | Same |
| Use Setting | Home use | Same |
| General Device Characteristic Differences | | |
| Intended Use Population | Persons who are not on insulin therapy age 18 years and older | Persons with diabetes age 2 years and older |
| Type of Use | Over-the-counter use | Prescription use |
| Sensor Calibration | Factory calibrated only | Factory calibrated with optional manual calibration |
| Sensor Useful Life | Up to 15 days with 12 hour grace period (automatic sensor shut off) | Up to 10 days with 12 hours grace period (automatic sensor shut off) |
| Primary Display Device | Mobile app | Mobile app or receiver |
| Displayed Range | 70-250 mg/dL | 40-400 mg/dL |
| Display Device Update Interval | Every 15 minutes | Every 5 minutes |
| Glucose Alerts and Alarms | None | Mandatory Alarms: Urgent Low
Optional Alerts: Urgent Low Soon, Low Glucose, High Glucose, Falling Fast, Rising Fast |
| Compatibility with Connected Devices | Compatible with digitally connected devices excluding | Compatible with digitally connected devices including |
K234070 - Page 6 of 21
{6}
| | insulin devices such as insulin pens and AID systems | insulin devices. |
| --- | --- | --- |
| Connectivity with cloud-based applications | The App can communicate wirelessly to Dexcom Clarity. | The App can communicate wirelessly to Dexcom Clarity and Dexcom Follow App. |
VI Standards/Guidance Documents Referenced:
1. ISO 14971 Third Edition 2019-12: Medical devices - Application of risk management to medical devices
2. IEC 62304 62304 Edition 1.1 2015-06 CONSOLIDATED VERSION: Medical device software - Software life cycle processes
3. ANSI AAMI ES60601-1:2005/(R)2012 & A1:2012 C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021]: Medical electrical equipment – Part 1: General requirements for basic safety and essential performance
4. IEC 60601-1-2 Edition 4.1 2020-09 CONSOLIDATED VERSION: Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance
5. IEC 60601-1-6 Edition 3.2 2020-07 CONSOLIDATED VERSION: Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance
6. IEC 60601-1-11 Edition 2.1 2020-07 CONSOLIDATED VERSION: Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance
7. ISO 14155 Third Edition 2020-07: Clinical investigation of medical devices for human subjects - Good clinical practice
8. IEC 1162366-1 Edition 1.1 2020-06 CONSOLIDATED VERSION: Medical devices - Part 1: Application of usability engineering to medical devices
9. ISO 10993-1 Fifth edition 2018-08: Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
10. ISO 10993-2 Second edition 2006-07-15: Biological Evaluation of medical devices - Part 2: Animal welfare requirements
11. ISO 10993-3 Third edition 2014-10-1: Biological evaluation of medical devices - Part 3: Tests for genotoxicity carcinogenicity and reproductive toxicity
12. ISO 10993-5 Third edition 2009-06-01 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
13. ISO 10993-6 Third edition 2016-12-01: Biological evaluation of medical devices -- Part 6: Tests for local effects after implantation
14. ISO 10993-7 Second edition 2008-10-15 Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals
15. ISO 10993-10 Third Edition 2010-08-01 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization
16. ISO 10993-11 Third edition 2017-09 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity
17. ISO 10993-12 Fourth edition 2012-07-01 Biological evaluation of medical devices - Part 12: Sample preparation and reference materials
18. ISO 10993-17 First edition 2002-12-01 Biological evaluation of medical devices - Part 17: Establishment of allowable limits for leachable substances
K234070 - Page 7 of 21
{7}
19. ISO 10993-18 Second edition 2020-01 Biological evaluation of medical devices - Part 18: Chemical characterization of medical device materials within a risk management process
20. ISO 11607-1 Second edition 2019-02 Packaging for terminally sterilized medical devices - Part 1: Requirements for materials sterile barrier systems and packaging systems
21. ISO 11607-2 Second edition 2019-02 Packaging for terminally sterilized medical devices - Part 2: Validation requirements for forming sealing and assembly processes
22. ISO 11135 Second edition 2014-07-15 Sterilization of health-care products - Ethylene oxide - Requirements for the development, validation, and routine control of a sterilization process for medical devices
23. ISO 11737-1 Third edition 2018-01 [including AMD:21] Sterilization of health care products - Microbiological methods - Part 1: Determination of a population of microorganisms on products
24. ISO 11737-2 Third edition 2019-12 Sterilization of medical devices - Microbiological methods - Part 2: Tests of sterility performed in the definition, validation and maintenance of a sterilization process
25. ISO 15223-1 Fourth edition 2021-07 Medical devices - Symbols to be used with information to be supplied by the manufacturer - Part 1: General requirements
26. ASTM D4169-22 Standard Practice for Performance Testing of Shipping Containers and Systems
27. ASTM F2503-23 Standard Practice for Marking Medical Devices and Other Items for Safety in the Magnetic Resonance Environment
28. ISO 23908 First edition 2011-06-11 Sharps injury protection - Requirements and test methods - Sharps protection features for single-use hypodermic needles, introducers for catheters and needles used for blood sampling
29. IEC 62133-2 Edition 1.0 2017-02: Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems
30. IEC TR 60601-4-2 Edition 1.0 2016-05 Medical electrical equipment - Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems
31. AIM Standard 7351731 Rev. 3.00 2021-06-04 Medical Electrical Equipment and System Electromagnetic Immunity Test for Exposure to Radio Frequency Identification Readers
32. IEEE ANSI USEMCSC C63.27-2021 American National Standard for Evaluation of Wireless Coexistence
33. AAMI TIR69:2017/(R2020) Technical Information Report Risk management of radio-frequency wireless coexistence for medical devices and systems
34. IEC 60417:2002 Graphical symbols for use on equipment
35. ANSI AAMI HE75:2009/(R)2018 Human factors engineering - Design of medical devices
36. ISO 11138-1 Third edition 2017-03 Sterilization of health care products - Biological indicators - Part 1: General requirements
K234070 - Page 8 of 21
{8}
VII Performance Characteristics (if/when applicable):
A Analytical Performance:
1. Precision/Reproducibility:
Stelo Glucose Biosensor performance was evaluated in clinical studies described below in section C(3). A subset of subjects wore two Stelo Glucose Biosensor at the same time (N=18). Precision was evaluated by comparing the glucose reading from the two Systems worn by the same subject on the backs of both upper arms.
The mean paired absolute relative difference (between the 2 concurrently worn devices) was 8.8% and the mean coefficient of variation (mean %CV) was 6.2%.
Precision Analysis
| Matched Pairs (n) | 63,024 |
| --- | --- |
| Number of subjects (N) | 18 |
| Paired absolute difference (mg/dL) | 13.3 |
| Paired absolute relative difference (%) | 8.8 |
| Coefficient of variation (%) | 6.2 |
2. Linearity:
The measurement range of the System is 40-400 mg/dL and the reportable range is 70-250 mg/dL. Data supporting this claimed measurement range was generated in the clinical study described in Section C(3) below.
3. Analytical Specificity/Interference:
Users of the Stelo Glucose Biosensor System can take a standard or maximum dose of acetaminophen (up to 1 gram every 6 hours in adults) and still use the Stelo System readings. Sensor glucose readings will be falsely higher if the user is taking more than a standard acetaminophen dose.
Sensor glucose readings will also be falsely higher if the user is taking hydroxyurea. Users should talk to their physician about alternative glucose monitoring approaches if they are taking hydroxyurea.
4. Assay Reportable Range:
See Linearity section above.
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
The Stelo GSS (sensor and transmitter) has a storage shelf-life of up to 10 months. Shelf life was evaluated at 86°F. Sensors should be stored at 36°- 86°F and 10-90% relative humidity.
K234070 - Page 9 of 21
{9}
6. Detection Limit:
If a glucose measurement is less than 70 mg/dL, the result is displayed by the system as “Below 70 mg/dL.” If a glucose measurement exceeds 250 mg/dL, result is displayed as “Above 250 mg/dL.” Users can view historical results across the full device measurement range (40-400 mg/dL) via Dexcom Clarity, a cloud-based software available on the web and on apps.
7. Assay Cut-Off:
Not applicable
8. Accuracy (Instrument):
Not applicable.
9. Carry-Over:
Not applicable.
B Comparison Studies:
1. Method Comparison with Predicate Device:
Not applicable. Accuracy is determined by comparing device values to an FDA cleared laboratory grade glucose analyzer (Yellow Springs Instrument 2300 STAT Plus™ Glucose Analyzer) and referred to as the “comparator method” in Section C(3) below.
2. Matrix Comparison:
Not applicable. Interstitial fluid is the only indicated matrix.
C Clinical Studies:
1. Clinical Sensitivity:
Not applicable.
2. Clinical Specificity:
Not applicable.
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
The sponsor conducted a pivotal study to evaluate the safety and effectiveness of the Stelo Glucose Biosensor system for up to 15.5 days. The study was conducted in the United States at six clinic sites. Although the real-time reporting range of the system is 70-250 mg/dL, users can view the full device measurement range of 40-400 mg/dL retrospectively via
K234070 - Page 10 of 21
{10}
Dexcom Clarity. Therefore, the device performance across the full measuring range was included in the analyses below and considered in the substantial equivalence determination.
## iCGM Reading Accuracy to Comparator
Below are the percent of values 15%/15 mg/dL, and 40%/40 mg/dL stratified by glucose ranges of <70, 70-180, and >180 mg/dL for the iCGM and comparator. The 95% CI in the below tables is the 95% confidence interval (two-sided, alpha = 0.05) and the 95% LB is the lower bound of 95% confidence limit (one-sided, alpha = 0.05).
### Percent and Point Accuracy Between iCGM and Comparator by iCGM Glucose Range (N=130)
| iCGM Glucose Range^{1} (mg/dL) | Matched Pairs (N) | Percent Within 15 mg/dL (95% LB) | Percent within 40 mg/dL (95% LB) | Percent within 15% (95% LB) | Percent within 40% (95% LB) | Mean Bias (mg/dL) (95% CI) |
| --- | --- | --- | --- | --- | --- | --- |
| <70 | 1,807 | 92.0 (88.2) | 99.3 (98.1) | --- | --- | -6.9 (-9.3, -4.5) |
| 70-180 | 10,902 | --- | --- | 84.3 (82.0) | 99.6 (99.4) | -5.2 (-7.4, -2.9) |
| >180 | 7,621 | --- | --- | 88.1 (84.4) | 99.9 (99.8) | -3.3 (-6.0, -0.6) |
¹iCGM readings are within 40-400 mg/dL, inclusive
### Percent and Point Accuracy Between iCGM and Comparator by Comparator Glucose Ranges (N=130)
| Comparator Glucose Range (mg/dL) | Matched Pairs^{1} (N) | Percent Within 15 mg/dL (95% LB) | Percent within 40 mg/dL (95% LB) | Percent within 15% (95% LB) | Percent within 40% (95% LB) | Mean Bias (mg/dL) (95% CI) |
| --- | --- | --- | --- | --- | --- | --- |
| <70 | 2,008 | 93.9 (91.6) | 99.9 (99.0) | --- | --- | -1.3 (-3.4, 0.7) |
| 70-180 | 10,463 | --- | --- | 86.0 (83.8) | 99.7 (99.5) | -3.2 (-5.2, -1.1) |
| >180 | 7,859 | --- | --- | 86.7 (82.8) | 99.9 (99.6) | -7.5 (-10.3, -4.7) |
¹iCGM readings are within 40-400 mg/dL, inclusive
The Stelo percent and point accuracy between iCGM and the comparator is comparable to that of the predicate G7 CGM System (K213919).
The percent of values within 20% of comparator method were calculated across the measuring range overall.
K234070 - Page 11 of 21
{11}
Percent of iCGM Values within 20% of Comparator Glucose (N=130)
| iCGM Glucose Range | Matched Pairs (n) | Percent within 20% (95% LB) |
| --- | --- | --- |
| 40-400 mg/dL | 20,330 | 93.1 (91.2) |
Percent of values within 15%/15 mg/dL, 20%/20 mg/dL, and 40%/40 mg/dL stratified by glucose ranges of <54, 54-69, 70-180, 181-250, and >250 mg/dL for iCGM and comparator were also provided.
System Accuracy to Comparator Within iCGM Glucose Ranges (N=130)
| iCGM Glucose Range^{1} (mg/dL) | Matched Pairs (N) | Percent Within 15 mg/dL | Percent Within 20 mg/dL | Percent within 40 mg/dL | Percent within 15% | Percent within 20% | Percent within 40% | Mean Bias (mg/dL) | MARD (%) |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| <54 | 288 | 70.5 | 84.7 | 99.7 | --- | --- | --- | -10.3 | 17.4 |
| 54-69 | 1,519 | 92.6 | 96.4 | 99.5 | --- | --- | --- | -2.2 | 9.2 |
| 70-180 | 10,902 | --- | --- | --- | 84.8 | 93.3 | 99.6 | -1.1 | 8.5 |
| 181-250 | 3,301 | --- | --- | --- | 84.3 | 91.5 | 100.0 | -11.3 | 8.3 |
| >250 | 4,320 | --- | --- | --- | 91.6 | 96.8 | 100.0 | -9.2 | 6.8 |
¹CGM readings are within 40-400 mg/dL, inclusive.
System Accuracy to Comparator Within Comparator Glucose Ranges (N=130)
| iCGM Glucose Range (mg/dL) | Matched Pairs^{1} (N) | Percent Within 15 mg/dL | Percent Within 20 mg/dL | Percent within 40 mg/dL | Percent within 15% | Percent within 20% | Percent within 40% | Mean Bias (mg/dL) | MARD (%) |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| <54 | 206 | 88.8 | 93.2 | 98.5 | --- | --- | --- | 6.0 | 15.4 |
| 54-69 | 1,802 | 94.1 | 97.8 | 100.0 | --- | --- | --- | 1.8 | 10.0 |
| 70-180 | 10,463 | --- | --- | --- | 86.2 | 93.8 | 99.7 | -0.3 | 8.2 |
| 181-250 | 2,988 | --- | --- | --- | 88.0 | 94.9 | 99.9 | -6.5 | 7.4 |
| >250 | 4,871 | --- | --- | --- | 86.1 | 92.5 | 99.9 | -15.8 | 8.0 |
¹CGM readings are within 40-400 mg/dL, inclusive.
K234070 - Page 12 of 21
{12}
Concurrence
Concurrence of iCGM values compared to the comparator method across the entire measuring range was also evaluated. iCGM glucose ranges of <40, 40-60, 61-80, 81-120, 121-160, 161-200, 201-250, 251-300, 301-350, 351-400, and >400 mg/dL were evaluated against the comparator glucose ranges and the percentages of iCGM values within those ranges were reported in the following tables.
Concurrence of iCGM and Comparator by iCGM Glucose Range (N=130)
| | Comparator Glucose Values (mg/dL) | | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| iCGM (mg/dL) | <40 | 40-60 | 61-80 | 81-120 | 121-160 | 161-200 | 201-250 | 251-300 | 301-350 | 351-400 | >400 | Total |
| <40 | --- | 23
63.9% | 12
33.3% | 1
2.8% | | | | | | | | 36 |
| 40-60 | 1
0.1% | 407
55.4% | 308
41.9% | 19
2.6% | | | | | | | | 735 |
| 61-80 | | 276
10.4% | 1,995
75.1% | 381
14.3% | 4
0.2% | | | | | | | 2,656 |
| 81-120 | 1
0.0% | 13
0.3% | 540
12.0% | 3,426
76.1% | 503
11.2% | 13
0.3% | 1
0.0% | 3
0.1% | | | | 4,500 |
| 121-160 | | | 1
0.0% | 468
13.2% | 2,584
72.9% | 468
13.2% | 23
0.6% | 2
0.1% | 1
0.0% | | | 3,547 |
| 161-200 | | | | 3
0.1% | 353
15.2% | 1,487
64.1% | 444
19.1% | 32
1.4% | 1
0.0% | | | 2,320 |
| 201-250 | | | | | 1
0.0% | 204
9.1% | 1,350
59.9% | 571
25.4% | 119
5.3% | 7
0.3% | | 2,252 |
| 251-300 | | | | | | | 185
7.9% | 1,299
55.6% | 772
33.0% | 80
3.4% | | 2,336 |
| 301-350 | | | | | | | | 205
12.9% | 1,114
70.2% | 261
16.4% | 8
0.5% | 1,588 |
| 351-400 | | | | | | | | | 129
32.6% | 249
62.9% | 18
4.5% | 396 |
| >400 | | | | | | | | | 1
1.6% | 36
59.0% | 24
39.3% | 61 |
K234070 - Page 13 of 21
{13}
Concurrence of iCGM and Comparator by Comparator Glucose Range (N=130)
| | Comparator Glucose Values (mg/dL) | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| iCGM (mg/dL) | <40 | 40-60 | 61-80 | 81-120 | 121-160 | 161-200 | 201-250 | 251-300 | 301-350 | 351-400 | >400 |
| <40 | --- | 23
3.2% | 12
0.4% | 1
0.0% | | | | | | | |
| 40-60 | 1
50.0% | 407
56.6% | 308
10.8% | 19
0.4% | | | | | | | |
| 61-80 | --- | 276
38.4% | 1,995
69.9% | 381
8.9% | 4
0.1% | | | | | | |
| 81-120 | 1
50.0% | 13
1.8% | 540
18.9% | 3,426
79.7% | 503
14.6% | 13
0.6% | 1
0.0% | 3
0.1% | | | |
| 121-160 | | | 1
0.0% | 468
10.9% | 2,584
75.0% | 468
21.5% | 23
1.1% | 2
0.1% | 1
0.0% | | |
| 161-200 | | | | 3
0.1% | 353
10.2% | 1,487
68.5% | 444
22.2% | 32
1.5% | 1
0.0% | | |
| 201-250 | | | | | 1
0.0% | 204
9.4% | 1,350
67.4% | 571
27.0% | 119
5.6% | 7
1.1% | |
| 251-300 | | | | | | | 185
9.2% | 1,299
61.5% | 772
36.1% | 80
12.6% | |
| 301-350 | | | | | | | | 205
9.7% | 1,114
52.1% | 261
41.2% | 8
16.0% |
| 351-400 | | | | | | | | | 129
6.0% | 249
39.3% | 18
36.0% |
| >400 | | | | | | | | | 1
0.0% | 36
5.7% | 24
48.0% |
| Total | 2 | 719 | 2,856 | 4,298 | 3,445 | 2,172 | 2,003 | 2,112 | 2,137 | 633 | 50 |
K234070 - Page 14 of 21
{14}
# Trend Accuracy
Trend accuracy describes the accuracy of the sensor during times of rapidly changing glucose and is characterized by slopes, such as from $>2\mathrm{mg/dL/min}$ to $<-2\mathrm{mg/dL/min}$ . Trend accuracy was assessed by the concurrence rate of the glucose rate of change (changes in $\mathrm{mg/dL}$ of glucose per minute) determined by the iCGM values and the corresponding comparator values for each iCGM-comparator measurement pair.
Trend Accuracy Between iCGM and Comparator (N=130)
| | Comparator Rate Range (mg/dL/min) | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| iCGM Rate Range (mg/dL/min) N (Row %) | <-2 | [-2,-1) | [-1,-0) | [0,1] | (1,2] | >2 | Number of Paired iCGM-Comparator (Row N) |
| <-2 | 80 (31.5%) | 99 (39.0%) | 57 (22.4%) | 14 (5.5%) | 2 (0.8%) | 2 (0.8%) | 254 |
| [-2,-1) | 96 (7.0%) | 618 (45.0%) | 574 (41.8%) | 77 (5.6%) | 7 (0.5%) | 1 (0.1%) | 1,373 |
| [-1,0) | 40 (0.4%) | 590 (6.6%) | 6,563 (73.2%) | 1,666 (18.6%) | 100 (1.1%) | 11 (0.1%) | 8,970 |
| [0,1] | 12 (0.2%) | 73 (1.1%) | 1,661 (24.2%) | 4,392 (64.0%) | 652 (9.5%) | 69 (1.0%) | 6,859 |
| (1,2] | 1 (0.1%) | 8 (0.5%) | 84 (4.9%) | 542 (31.8%) | 822 (48.3%) | 246 (14.4%) | 1,703 |
| >2 | 0 (0.0%) | 3 (0.4%) | 18 (2.7%) | 78 (11.6%) | 209 (31.1%) | 365 (54.2%) | 673 |
| Number of Paired iCGM-Comparator (Column N) | 229 | 1,391 | 8,957 | 6,769 | 1,792 | 694 | 19,832 |
Note: RoC was calculated using comparator data with consecutive measurements greater than 5 minutes.
K234070 - Page 15 of 21
{15}
Agreement when iCGM Reads “Below 70 mg/dL” or “Above 250 mg/dL”
The Stelo Glucose Biosensor System reports glucose readings between 70 and 250 mg/dL. When the system determines the sensor reading is below 70 mg/dL, it displays “Below 70 mg/dL” on the mobile app. When the system determines the sensor reading is above 250 mg/dL, it displays “Above 250 mg/dL” on the mobile app. Because the System does not display glucose values below 70 mg/dL or above 250 mg/dL, the comparisons to the actual blood glucose levels (as determined by the comparator method) when the iCGM value is classified as “Below 70 mg/dL” or “Above 250 mg/dL” is evaluated separately, and the cumulative percentages of when the comparator values were less than certain glucose values (for “Below 70 mg/dL”) and when comparator values were greater than certain glucose values (for “Above 250 mg/dL”) are presented in the table below.
Distribution of Comparator for iCGM Readings “Below 70 mg/dL” or “Above 250 mg/dL” (N=130)
| | | Comparator (mg/dL) | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| iCGM Readings | iCGM-Comparator Pairs | <80 | <90 | <100 | <110 | ≥110 | Total |
| Below 70 mg/dL" | n | 1740 | 1814 | 1833 | 1841 | 2 | 1843 |
| | Cumulative Percent | 94 | 98 | 99 | 100 | 0 | |
| | | | | | | | |
| iCGM Readings | iCGM-Comparator Pairs | >200 | >180 | >160 | >130 | ≤130 | Total |
| "Above 250 mg/dL" | n | 4381 | 4381 | 4381 | 4381 | 0 | 4381 |
| | Cumulative Percent | 100 | 100 | 100 | 100 | 0 | |
Sensor Stability
Sensor stability describes the performance over the sensor’s intended lifetime (up to 15 days, including the optional 12-hour grace period). Performance was estimated by calculating the percentage of iCGM readings within 15 mg/dL or 15% (15/15%), 20 mg/dL or 20% (20/20%), and 40 mg/dL or 40% (40/40%) of the laboratory comparator values during the beginning (Day 1 to 3), early middle (Day 4 to 7), late middle (Day 9 to 12), and end (Day 13 to the first half of Day 15) of the System lifecycle. The mean of the absolute relative difference was evaluated over the sensor life within the measuring range.
Accuracy of iCGM vs Comparator by Wear Period (N=130)
| Wear period | Matched pairs (n) | MARD (%) | %15/15 (%) | %20/20 (%) | %40/40 (%) |
| --- | --- | --- | --- | --- | --- |
| Beginning | 5,718 | 9.1 | 83.6 | 92.4 | 99.7 |
| Early middle | 5,509 | 7.8 | 89.2 | 95.2 | 99.9 |
| Late middle | 5,197 | 7.3 | 90.4 | 96.0 | 99.9 |
| End | 3,906 | 9.1 | 85.1 | 92.1 | 99.7 |
K234070 - Page 16 of 21
{16}
K234070 - Page 17 of 21
# Sensor Life
A total of 131 sensors worn on the arm by adult subjects were evaluated to determine the percentage of sensors that lasted through the 15-day sensor life. Of the 131 Sensors, 73.5% of evaluable sensors lasted until the final day of use. Some sensors were excluded from this analysis if they were intentionally removed prior to day 15. For example, on or before day 15, 60 sensors were intentionally removed by the sponsor as part of the clinical study. 26 Sensors (19.8%) had “early sensor shut-off” (ESS), which is when a sensor automatically ends a sensor session as a result of self-diagnostics. Survival rates were calculated using the Kaplan Meier method.
Summary of System Survival by Day with 15-day Wear Period (N Subject=130)
| Day of wear | Number of sensors (N=131) | Survival rate (%) |
| --- | --- | --- |
| 1 | 130 | 99.2 |
| 2 | 130 | 99.2 |
| 3 | 128 | 97.7 |
| 4 | 124 | 94.7 |
| 5 | 124 | 94.7 |
| 6 | 123 | 93.9 |
| 7 | 121 | 93.1 |
| 8 | 120 | 92.4 |
| 9 | 119 | 91.6 |
| 10 | 115 | 89.3 |
| 11 | 112 | 87.7 |
| 12 | 107 | 83.8 |
| 13 | 104 | 81.4 |
| 14 | 75 | 78.9 |
| 15 | 38 | 73.5 |
{17}
A subsequent study was conducted in 108 participants in the intended user population to assess the impact of a new sensor patch intended to improve the sensor survival rate. Subjects attended clinic sessions for device insertion and removal. Sensor insertions were performed at the clinic by the subjects. All subjects wore an overlay with the device (also called an overpatch). No blood draws or home self-monitored blood glucose testing were required for this study. No display devices were provided to subjects, so all system data were blinded to the subjects for the full wear period and could not be used for subjects' diabetes management. A total of 80 of 110 systems lasted until Day 15, 6 of which were intentionally removed on or prior to day 15 and were censored from the survival rate analysis. 17 sensors had early sensor shutoff (ESS) and 5 had an adhesive failure. Survival rates were calculated using the Kaplan Meier method.
Summary of System Survival by Day in Subsequent Study (N Subject=108)
| Day of wear | Number of sensors (N=110) | Survival rate (%) |
| --- | --- | --- |
| 1 | 107 | 97.3 |
| 2 | 107 | 97.3 |
| 3 | 107 | 97.3 |
| 4 | 107 | 97.3 |
| 5 | 106 | 96.4 |
| 6 | 106 | 96.4 |
| 7 | 106 | 96.4 |
| 8 | 104 | 94.5 |
| 9 | 102 | 92.7 |
| 10 | 101 | 91.8 |
| 11 | 97 | 90.0 |
| 12 | 93 | 86.3 |
| 13 | 91 | 84.4 |
| 14 | 89 | 82.6 |
| 15 | 80 | 77.9 |
K234070 - Page 18 of 21
{18}
# Data Capture Rate
The data capture rate characterizes the reliability of the communication between components of the system. The next table describes the data availability rate as the percentage of readings expected to be calculated throughout the 15-day life span of the sensor life based on information collected in the pivotal trial.
Data Availability Rate by Wear Day (N=130)
| Wear day | Number of sensors | Data availability rate (%) |
| --- | --- | --- |
| 1 | 130 | 99.4 |
| 2 | 130 | 99.7 |
| 3 | 130 | 99.7 |
| 4 | 128 | 99.6 |
| 5 | 124 | 99.8 |
| 6 | 124 | 99.7 |
| 7 | 123 | 99.8 |
| 8 | 121 | 99.6 |
| 9 | 120 | 99.5 |
| 10 | 119 | 99.2 |
| 11 | 115 | 98.9 |
| 12 | 112 | 98.6 |
| 13 | 107 | 98.9 |
| 14 | 104 | 97.8 |
| 15 | 75 | 97.1 |
# D Clinical Cut-Off:
Not applicable.
# E Expected Values/Reference Range:
Not applicable.
# F Other Supportive Instrument Performance Characteristics Data:
The following supportive performance characteristics were established through nonclinical testing of the predicate device and are applicable to the Stelo Glucose Biosensor System in this 510(k):
- Biocompatibility
- Chemical/Material Characterization (not Biocompatibility-related)
- Sterilization Validation
- Electrical and Mechanical Performance
- Operating Environmental Conditions Testing
- Wireless Coexistence
- Electrical Safety and Electromagnetic Compatibility
- Packaging Validation
- Interoperability
K234070 - Page 19 of 21
{19}
The following performance characteristics were verified or validated through studies conducted on the subject device, Stelo Glucose Biosensor System:
## Human Factors
Sixty participants across different regional geographies evaluated four self-selection e-commerce platforms in the HF validation test through both the Dexcom website and e-commerce website via mobile and desktops. Fifteen of the 60 participants who correctly self-identified themselves as intended users of the Stelo System (based off of the above e-commerce websites) completed a simulated Human Factors validation testing using the Stelo System mobile app, Sensor/Applicator and associated instructional materials.
## Software Verification and Validation
Software verification and validation testing was conducted to confirm that the software used in the Stelo Glucose Biosensor System performed in accordance with established specifications, IEC 62304 and FDA Guidance document “Guidance for the Content of Premarket Submissions for Device Software Functions,” June 14, 2023. Evaluation activities included code review, unit, system integration, and system level testing which verified functionality of the device against established software requirements. Results of the software executed protocols for the Stelo Glucose Biosensor System are acceptable for their intended use.
## Cybersecurity
Dexcom has provided cybersecurity risk management documentation for the Stelo Glucose Biosensor System that includes analysis of confidentiality, integrity, and availability for data, information and software related to the Stelo Glucose Biosensor System in accordance with the FDA Guidance Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions” (September 27, 2023). For each identified threat and vulnerability risk event scenario, risk assessment of impact to confidentiality, integrity, and availability was performed and documented within the cybersecurity risk management documentation. Appropriate risk mitigation controls have been implemented and tested. In addition, Dexcom has controls and processes in place to ensure continued support for keeping the device secure and to ensure that the device firmware, software and components are malware free. Additional controls are also in place in manufacturing through distribution to ensure that the medical device firmware and software are malware free from point of origin to the hands of the end user.
## Electrical and Mechanical Performance (session dependent testing):
Performance testing was performed to ensure the device specifications for sensor sensitivity, linearity, and fatigue, as well as wearable battery life and current leakage were met.
## Operating Environmental Conditions Testing:
Environmental testing was performed on the Stelo Glucose Biosensor System to ensure the device specifications for operational robustness, operating conditions (evaluation of performance under various operational temperatures and humidity conditions), and chemical robustness (evaluation of performance when subject to soap water) were met.
## Shelf-Life:
Shelf-life testing was performed to evaluate the stability of Stelo Glucose Sensing Subsystem under real time anticipated storage conditions and supported its useful life to be up to 10 months. The test results for the Stelo Glucose Sensing Subsystem met specifications.
K234070 - Page 20 of 21
{20}
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K234070 - Page 21 of 21
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.