The Medline ReNewal Reprocessed Siemens ACUSON AcuNav Diagnostic Ultrasound Catheter is inteacardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart of adult patients. The catheter is intended for imaging guidance only, not treatment delivery, during cardiac interventional percutaneous procedures.
Device Story
Reprocessed 8F ultrasound catheter; 64-element phased array transducer; provides intracardiac/intraluminal visualization of cardiac/great vessel anatomy and physiology; used for imaging guidance during percutaneous cardiac interventions. Operated by physicians in clinical settings; connects to compatible ultrasound consoles (e.g., ACUSON, Vivid I, Vivid q) via module interface. Physician maneuvers catheter tip (four-way deflection: anterior/posterior, left/right) to optimize tissue visualization. Output displayed on console for real-time guidance; aids clinical decision-making by providing anatomical visualization. Reprocessed once; tracked via cycle count; subject to cleaning, sterilization, and functional testing.
8F ultrasound catheter; 64-element phased array transducer; four-way tip deflection (anterior/posterior, left/right); 90 cm length. Connects to compatible ultrasound consoles via module interface. Reprocessed device; single-use per cycle; maximum cycle limit enforced. Sterilization via ethylene oxide.
Indications for Use
Indicated for intracardiac and intraluminal visualization of cardiac and great vessel anatomy, physiology, and other devices in the heart of adult patients. Intended for imaging guidance only during cardiac interventional percutaneous procedures.
Regulatory Classification
Identification
An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services-USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The FDA logo is a symbol of the agency's mission to protect and promote public health through the regulation and supervision of food safety, tobacco, pharmaceuticals, medical devices, and other related products.
## September 10, 2024
Surgical Instrument Service and Savings Inc. (dba Medline Renewal) Stephanie Boyle Mays Senior Specialist, Regulatory Affairs Quality Assurance 1500 NE Hemlock Ave. Redmond, Oregon 97756
## Re: K234064
Trade/Device Name: Medline ReNewal Reprocessed Siemens ACUSON AcuNav 8F Ultrasound Catheter (for Siemens Systems) (10135936); Medline ReNewal Reprocessed Siemens ACUSON AcuNav 8F Ultrasound Catheter (for GE Systems) (10135910) Regulation Number: 21 CFR 870.1200 Regulation Name: Diagnostic Intravascular Catheter Regulatory Class: Class II Product Code: OWQ Dated: August 12, 2024 Received: August 12, 2024
Dear Stephanie Boyle Mays:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Ouality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Aneesh S. Deoras -S
Aneesh Deoras Assistant Director
{2}------------------------------------------------
Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
The item numbers included in the scope of this submission are as follows:
| OEM Model<br>No. | Transducer | Curve | Size |
|------------------|----------------------------------------------|------------------------------------------------------------|-------------|
| 10135910 | GE Systems - 64 element<br>phased array | Four-way tip deflection:<br>anterior/posterior; left/right | 8 F x 90 cm |
| 10135936 | Siemens Systems - 64<br>element phased array | Four-way tip deflection:<br>anterior/posterior; left/right | 8 F x 90 cm |
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known) K234064
#### Device Name
Medline ReNewal Reprocessed Siemens ACUSON AcuNav Diagnostic Ultrasound Catheter, 8F (for Siems) (10135936); Medline ReNewal Reprocessed Siemens ACUSON AcuNav Diagnostic Ultrasound Catheter, 8F (for GE Systems) (10135910)
#### Indications for Use (Describe)
The Medline ReNewal Reprocessed Siemens ACUSON AcuNav Diagnostic Ultrasound Catheter is inteacardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart of adult patients. The catheter is intended for imaging guidance only, not treatment delivery, during cardiac interventional percutaneous procedures.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| | |
| X Prescription Use (Part 21 CFR 801 Subnart D) | Over-The-Counter Use (21 CFR 801 Subnart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image contains the Medline and Renewal logos. The Medline logo is on the left side of the image and consists of a blue square with a white star inside. The Renewal logo is on the right side of the image and consists of the word "Renewal" in green and blue letters. Below the word "Renewal" is the phrase "Full Circle Reprocessing" in smaller, black letters.
Traditional 510(k) Notification Medline ReNewal Reprocessed ACUSON AcuNav 8F Ultrasound Catheter
# 510(k) 234064 Summary
This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR § 807.92.
| Submitter/<br>Owner | Surgical Instrument Service and Savings Inc. (dba Medline ReNewal)<br>1500 NE Hemlock Ave., Redmond, OR 97756 | | |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Contact/<br>Prepared by | Stephanie Boyle Mays<br>Senior Specialist, Regulatory Affairs Quality Assurance<br>P: 541-516-4205 • F: 541-923-3375 • smays@medline.com | | |
| Date Prepared | December 21, 2023 | | |
| Device Name<br>and<br>Classification | Proprietary/Trade<br>Name: | Medline ReNewal Reprocessed Siemens ACUSON AcuNav 8F Ultrasound Catheter<br>(for Siemens Systems) (10135936); Medline ReNewal Reprocessed Siemens<br>ACUSON AcuNav 8F Ultrasound Catheter (for GE Systems) (10135910) | |
| Device Name<br>and<br>Classification | Common or Usual<br>Name | Ultrasound catheter, reprocessed | |
| Device Name<br>and<br>Classification | Regulatory<br>Name/Reference | Reprocessed intravascular, ultrasound catheter,<br>21 CFR § 870.1200 | |
| Device Name<br>and<br>Classification | Regulatory Class | 2 | |
| Device Name<br>and<br>Classification | Product Code | OWQ | |
| Device Name<br>and<br>Classification | Panel | Cardiovascular | |
| Device Name<br>and<br>Classification | Predicate<br>selection rationale | The predicate models in K170263 include the subject<br>device models of this submission. The predicate<br>devices and subject devices are the same except the<br>subject devices have been reprocessed. | |
| Device Name<br>and<br>Classification | 510(k) Number | K170263 | |
| Predicate<br>Device | Proprietary or<br>Trade Name | AcuNav Diagnostic Ultrasound Catheter 8F, 10F | |
| Predicate<br>Device | Common or Usual<br>Name | Ultrasound catheter | |
| Predicate<br>Device | Regulatory<br>Name/Reference | Catheter, ultrasound, intravascular<br>21 CFR § 870.1200 | |
| Predicate<br>Device | Regulatory Class | 2 | |
| Predicate<br>Device | Product Code | OBJ | |
| Predicate<br>Device | Panel | Cardiovascular | |
| Device<br>Description | 510(k) applicant | Siemens Medical Solutions, 685 E. Middlefield Rd.,<br>Mountain View, CA 94043 | |
| Device<br>Description | The Medline ReNewal Reprocessed AcuNav Diagnostic Ultrasound<br>Catheters 8F are licensed for single use only. The catheter is optimized for<br>intracardiac scanning. With the catheter, the physician can maneuver the<br>imaging plane located inside the catheter tip to see the region of interest.<br>The physician can steer the catheter to optimize tissue visualization. | | |
| | The catheters are to be used only on systems with which they have been<br>tested and found compatible. | | |
| Technological<br>Characteristics | The technological characteristics, materials, and the fundamental scientific<br>technology of the subject device is equivalent to the predicate device. The<br>proposed device is a reprocessed version of the predicate device. Each<br>device is marked, tracked, and taken out of service once the maximum<br>number of cycles has been reached. K170263 AcuNav Diagnostic<br>Ultrasound Catheter was used as the primary predicate to support intended<br>use, technological characteristics, and functional performance<br>specifications. | | |
| Non-clinical<br>Testing<br>Summary | The functional characteristics of the subject device have been evaluated<br>and found to be substantially equivalent to the predicate device based on<br>the following tests:<br>Functional performance:<br>●<br>simulated use and artificial soiling;<br>●<br>visual inspection;<br>●<br>mechanical characteristics<br>●<br>ultrasound transducer testing<br>●<br>dimensional analysis<br>●<br>device functionality:<br>●<br>Electrical Safety<br>●<br>dielectric and current leakage<br>●<br>Cleaning Validation<br>●<br>Biocompatibility<br>●<br>Sterilization Validation<br>●<br>bioburden testing; and<br>●<br>ethylene oxide and ethylene chlorohydrin residuals testing<br>●<br>bacteriostasis/fungistasis<br>●<br>Packaging and shelf life validation<br>●<br>Product stability | | |
| | Summary Table: Predicate and Medline ReNewal Reprocessed AcuNav Diagnostic<br>Ultrasound Catheter comparison chart. | | |
| | Predicate | Proposed | Comparison |
| View | Siemens Medical Solutions<br>AcuNav Diagnostic<br>Ultrasound Catheter | Medline ReNewal<br>Reprocessed Siemens<br>ACUSON AcuNav<br>Diagnostic Ultrasound<br>Catheter 8F | As Stated |
| 510(k) | K170263 | 234064 | N/A |
| Model Numbers | 10135936 and 10135910, 8F<br>08255790 and 10043372, 10F | 10135936 and 10135910 8F | As stated |
| Common Name | Ultrasound catheter | Ultrasound catheter<br>Reprocessed | As stated |
| ' Regulation No. | 21 CFR § 870.1200 | 21 CFR § 870.1200 | Same |
| Regulatory<br>Class | 2 | 2 | Same |
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for Medline Renewal Full Circle Reprocessing. The Medline logo is on the left, with the word "MEDLINE" in white letters on a blue square. To the right of the Medline logo is the word "Renewal" in large green and blue letters. Below the word "Renewal" is the phrase "Full Circle Reprocessing" in smaller blue letters.
Traditional 510(k) Notification
Medline ReNewal Reprocessed ACUSON AcuNav 8F Ultrasound Catheter
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the Medline Renewal logo. The Medline logo is on the left side of the image and is a blue square with a white star and the word "MEDLINE" in white letters. To the right of the Medline logo is the word "Renewal" in green and blue letters. Below the word "Renewal" is the phrase "Full Circle Reprocessing" in smaller letters.
# K234064
Traditional 510(k) Notification Medline ReNewal Reprocessed ACUSON AcuNav 8F Ultrasound Catheter
| Product Code | OBJ | OWQ | As stated |
|---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Indications for<br>Use | The catheter is intended for<br>intra-cardiac and intra-luminal<br>visualization of cardiac and<br>great vessel anatomy and<br>physiology as well as<br>visualization of other devices<br>in the heart of adult and<br>pediatric patients. The catheter<br>is intended for imaging<br>guidance only, not treatment<br>delivery, during cardiac<br>interventional percutaneous<br>procedures. | The Medline ReNewal<br>Reprocessed Siemens<br>ACUSON AcuNav Diagnostic<br>Ultrasound Catheter 8F is<br>intended for intra-cardiac and<br>intra-luminal visualization of<br>cardiac and great vessel<br>anatomy and physiology as<br>well as visualization of other<br>devices in the heart of adult<br>patients. The catheter is<br>intended for imaging guidance<br>only, not treatment delivery,<br>during cardiac interventional<br>percutaneous procedures. | Same |
| Technological<br>Characteristics<br>a | To operate, the device is<br>connected to a compatible<br>console (ACUSON, Vivid I,<br>Vivid q) via a compatible<br>module interface. | To operate, the device is<br>connected to a compatible<br>console (ACUSON, Vivid I,<br>Vivid q) via a compatible<br>module interface. | Same |
| Reprocessing | Each catheter is reprocessed no more than one time. Medline ReNewal does<br>not reprocess the catheters of other reprocessors. | | |
| Conclusion | The predicate and proposed devices in this application have the same<br>indications for use and technological characteristics. Based on this and the<br>non-clinical testing data presented in this 510(k) submission, the Medline<br>ReNewal Reprocessed AcuNav Ultrasound Catheter 8F models 10135936<br>and 10135910 are substantially equivalent to the predicate device. | | |
| a | Neither K170263 nor the current submission included the consoles or any other system<br>components as part of their respective submissions. Only devices 10135936 and 10135910 are<br>included in the project scope. | | |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.