# Automatic Registration (K234047)

_Brainlab AG · OLO · Mar 20, 2024 · Neurology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K234047

## Device Facts

- **Applicant:** Brainlab AG
- **Product Code:** [OLO](/productcode/OLO.md)
- **Decision Date:** Mar 20, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.4560
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** AI/ML

## Indications for Use

Automatic Registration is a software device for image guided surgery intended to be used in combination with compatible Brainlab navigation systems such as the Brainlab Spine & Trauma Navigation System. Automatic Registration provides an image registration for intraoperatively acquired 3D CT/CBCT or fluoroscopic images.

## Device Story

Software accessory for Brainlab Spine & Trauma Navigation System; correlates intraoperative 3D CT/CBCT or fluoroscopic images to surgical environment. Uses registration matrices, adhesive markers, and calibration phantom. New feature: 'Registration Update' method uses static CNN AI/ML algorithm to detect vertebrae in 2D X-ray images; enables rough pre-positioning of 2D X-rays relative to 3D datasets; reduces manual outlining and pre-positioning. Used in OR by surgeons/clinical staff. Output provides patient registration for navigation; improves surgical accuracy and workflow efficiency by automating landmark detection and image alignment.

## Clinical Evidence

No clinical testing performed. Bench testing included system registration accuracy verification in human cadavers (P95 ≤ 2.5 mm; mean 3D deviation ≤ 1.5 mm). AI/ML algorithm performance quantified via object detection, vertebra level assignment, and landmark prediction quality. Summative usability testing conducted with 15 representative users to validate new UI and workflow.

## Technological Characteristics

Software-based image registration system; utilizes optical tracking. Hardware includes reusable registration matrices, adhesive markers, and calibration phantom. Software modules: Automatic Registration 2.6, Universal Atlas Performer 6.0, Universal Atlas Transfer Performer 6.0. Connectivity: DICOM-compatible with 3rd party scanners (Airo, LoopX). Algorithm: Static (locked) Convolutional Neural Network (CNN) for vertebra detection. UI: HTML5-based.

## Regulatory Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

## Predicate Devices

- Automatic Registration 2.5 ([K203679](/device/K203679.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

March 20, 2024

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Brainlab AG Esther Moreno Garcia QM Consultant Olof-Palme-Str.9 Munich, 81829 Germany

Re: K234047

Trade/Device Name: Automatic Registration Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: December 21, 2023 Received: December 21, 2023

# Dear Esther Moreno Garcia:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Tejen D. Soni -S

For

Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K234047

Device Name Automatic Registration

Indications for Use (Describe)

Automatic Registration is a software device for image guided surgery intended to be used in combination with compatible Brainlab navigation systems such as the Brainlab Spine & Trauma Navigation System. Automatic Registration provides an image registration for intraoperatively acquired 3D CT/CBCT or fluoroscopic images.

Type of Use (Select one or both, as applicable)

| <div> <span style="font-size:10pt"> <span style="font-family:Wingdings">☑</span> </span>     Prescription Use (Part 21 CFR 801 Subpart D)    </div> |
|-----------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span style="font-size:10pt"> <span style="font-family:Wingdings">☐</span> </span>     Over-The-Counter Use (21 CFR 801 Subpart C)    </div>  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the Brainlab logo. The logo consists of a stylized pink brain icon on the left, followed by the word "BRAINLAB" in pink, uppercase letters. The brain icon appears to be made up of several curved lines, giving it a modern and abstract look.

# 510(k) Summary K234047

## March 20, 2024

| General Information                    |                                                       |
|----------------------------------------|-------------------------------------------------------|
| Manufacturer                           | Brainlab AG; Olof-Palme-Str.9, 81829, Munich, Germany |
| Establishment Registration             | 8043933                                               |
| Trade Name                             | Automatic Registration                                |
| Classification Name                    | Orthopedic Stereotaxic Instrument                     |
| Product Code                           | OLO                                                   |
| Regulation Number                      | 882.4560                                              |
| Regulatory Class                       | II                                                    |
| Panel                                  | Orthopedic                                            |
| Predicate Device(s)                    | K203679 - Automatic Registration                      |
| Contact Information                    |                                                       |
| Primary Contact                        | Alternate Contact                                     |
| Esther Moreno Garcia                   | Chiara Cunico                                         |
| QM Consultant - Regulatory Affairs     | Phone: +49 89 99 15 68 0                              |
| Phone: +49 89 99 15 68 0               | Fax: +49 89 99 15 68 5033                             |
| Email: regulatory.affairs@brainlab.com | Email: chiara.cunico@brainlab.com                     |

#### 1. Indications for Use

Automatic Registration is a software device for image guided surgery intended to be used in combination with compatible Brainlab navigation systems such as the Brainlab Spine & Trauma Navigation System. Automatic Registration provides an image registration for intraoperatively acquired 3D CT/CBCT or fluoroscopic images.

#### 2. Device Description

The Subject Device Automatic Registration is an accessory to the Brainlab Spine & Trauma Navigation System. It correlates intraoperatively acquired patient data (3D CT/CBCT or fluoroscopic images) to the surgical environment in order to provide a patient registration for subsequent use by the Brainlab Spine & Trauma Navigation. The device includes the following software modules:

- Automatic Registration 2.6 ●
- Universal Atlas Performer 6.0 ●
- Universal Atlas Transfer Performer 6.0 .

And uses as well several hardware devices, mainly registration matrices, adhesive flat markers and a calibration phantom, for performing the registration. The software is installed on an Image

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the Brainlab logo. The logo consists of a stylized brain icon on the left and the word "BRAINLAB" in capital letters on the right. The color of the logo is pink.

Guided Surgery (IGS) platform. The registration matrices are reusable devices, delivered nonsterile and having patient contact.

The image registration can be provided using different methods, depending on the compatible CT or Cone-Beam CT scanner used. There is a specific workflow integration with the 3ª party scanners Airo and LoopX, as well as a universal workflow that can be used with any CT scanner using Dicom data. As a new feature, Automatic Registration 2.6 implements the possibility to update an automatic registration generated with LoopX by using also a static (locked) Al/ ML algorithm provided by the Universal Atlas for detection of vertebrae in 2D X-Ray images. This vertebra detection is then used for a rough pre-positioning of 2D X-Rays in relation to the 3D dataset and reduces user interaction for outlining 2D images and manual pre-positioning of images.

# 3. Substantial Equivalence

For the Substantial Equivalence determination, comparison of the Subject Device features with the following predicate device(s) was carried out:

K203679 - Automatic Registration 2.5 (product code: OLO)

The predicate was chosen since it's the predecessor device and therefore very similar to the subject device w.r.t the indications for use, technological characteristics and use cases.

The main difference compared to the predicate device is the new Registration Update method using AI/ ML. The two already existing registration methods have been reimplemented in a new GUI, without significant changes.

| Indications for<br>use                 | Automatic Registration is a<br>surgical device for image guided<br>surgery intended to be used in<br>combination with compatible<br>Brainlab navigation systems.<br>Automatic Registration provides<br>an image registration for<br>intraoperatively acquired 3D CT/<br>CBCT or fluoroscopic images. It<br>consists of the software module<br>Automatic Registration and<br>hardware accessories. | Automatic Registration is a<br>software device for image guided<br>surgery intended to be used in<br>combination with compatible<br>Brainlab navigation systems such<br>as the Brainlab Spine & Trauma<br>Navigation System. Automatic<br>Registration provides an image<br>registration for intraoperatively<br>acquired 3D CT/CBCT or<br>fluoroscopic images. |
|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                        |                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                 |
|                                        |                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                 |
| Intended use<br>environment            | Automatic Registration is<br>intended to be used in an OR<br>environment. The device is not<br>intended to be used in the MR<br>environment..                                                                                                                                                                                                                                                     | Automatic Registration is<br>intended to be used in an OR<br>environment. The device is not<br>intended to be used in the MR<br>environment.                                                                                                                                                                                                                    |
| Software version                       | Automatic Registration 2.5                                                                                                                                                                                                                                                                                                                                                                        | Automatic Registration 2.6<br>Universal Atlas Performer 6.0<br>Universal Atlas Transfer<br>Performer 6.0                                                                                                                                                                                                                                                        |
| Hardware<br>devices                    | Registration Matrix CT –<br>Open Surgery (Spine) Registration Matrix CT –<br>Small Incision (Spine) Registration Matrix CT –<br>Minimally Invasive (Cranial<br>and Spine) Adhesive flat markers Calibration Phantom iCT                                                                                                                                                                           | Registration Matrix CT -<br>Open Surgery (Spine) Registration Matrix CT -<br>Small Incision (Spine) Registration Matrix CT -<br>Minimally Invasive (Cranial<br>and Spine) Adhesive flat markers Calibration Phantom iCT                                                                                                                                         |
| Localization<br>technique              | Optical tracking                                                                                                                                                                                                                                                                                                                                                                                  | Optical tracking                                                                                                                                                                                                                                                                                                                                                |
| Registration<br>methods                | Universal AIR Pre-Calibrated Automatic<br>Registration                                                                                                                                                                                                                                                                                                                                            | Universal AIR Pre-Calibrated Automatic<br>Registration Registration Update                                                                                                                                                                                                                                                                                      |
| Image<br>registration<br>accuracy      | Registration accuracy of ≤ 2.5<br>mm (P95, i.e. within the 95th<br>percentile) with a mean<br>navigation accuracy with 3D<br>deviation ≤ 1.5 mm.                                                                                                                                                                                                                                                  | Registration accuracy of ≤ 2.5<br>mm (P95, i.e. within the 95th<br>percentile) with a mean<br>navigation accuracy with 3D<br>deviation ≤ 1.5 mm.                                                                                                                                                                                                                |
| UI technology                          | WPF                                                                                                                                                                                                                                                                                                                                                                                               | HTML5                                                                                                                                                                                                                                                                                                                                                           |
| Compatible 3rd<br>party CT<br>scanners | CT and Cone Beam CT Imaging<br>devices with DICOM<br>Conformance Statement are<br>compatible (Universal AIR).                                                                                                                                                                                                                                                                                     | CT and Cone Beam CT Imaging<br>devices with DICOM<br>Conformance Statement are<br>compatible (Universal AIR).                                                                                                                                                                                                                                                   |
|                                        | Airo and LoopX                                                                                                                                                                                                                                                                                                                                                                                    | Airo and LoopX                                                                                                                                                                                                                                                                                                                                                  |

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "BRAINLAB" in pink font. To the left of the word is a pink symbol that looks like three stylized brains stacked on top of each other. The symbol is slightly smaller than the word "BRAINLAB".

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the word "BRAINLAB" in pink, with a stylized pink symbol to the left of the word. The symbol appears to be three curved lines stacked on top of each other. The word "BRAINLAB" is written in a sans-serif font, with each letter capitalized. The overall image has a clean and modern look.

## 4. Performance Data

The following testing was conducted on the Subject Device to establish substantial equivalence with the predicate device:

#### Software Verification and Validation Testing

Software verification and validation testing has been conducted and documentation is provided as recommended by FDA's Guidance for Industry and FDA Staff, "Content of Premarket Submissions for Device Software Functions". These include successful implementation of product specifications, incremental testing for different release candidates, testing of risk control measures, compatibility testing or cybersecurity tests. The documentation submitted is for Enhanced Documentation level.

#### Bench Tests

- . Accuracy: For the new reqistration method "Registration Update" the overall system registration accuracy was verified in human cadavers under a representative worst case scenario. The following acceptance criteria has been fulfilled: Registration accuracy of ≤ 2.5 mm (P95, i.e. within the 95th percentile) with a mean navigation accuracy with 3D deviation ≤ 1.5 mm. Therefore the subject device achieves the same accuracy values as the registration methods available in the predicate device.
- Machine Learning: The AI/ML algorithm used in the new registration method is a Convolutional Neural network (CNN) developed using a Supervised Learning approach. The algorithm was developed using a controlled internal process that defines activities from the inspection of input data to the training and verification of the algorithm. This is a static algorithm (locked). The assessment of the AI/ML detected landmarks on X-rays has been done by quantifying the object detection, the quality of vertebra level assignment, the quality of landmark predictions, and the performance of the observer view direction for 2D X-rays from the Universal Atlas Transfer Performer 6.0.
- . Usability: Summative usability testing was performed in order to validate the new user interface and workflow guidance for the Registration Update method. The evaluation was conducted with 15 representative users which had to use the Registration Update under different circumstances. The summative usability testing identified no critical use related problems.

No clinical testing was required for the subject device.

#### 5. Conclusion

The comparison of the Subject Device with the predicate device shows that Automatic Registration (2.6) has similar functionality, intended use and technological characteristics as the predicate device. Based on the comparison to the predicate and the performance testing conducted, the Subject Device is considered substantially equivalent to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K234047](https://fda-staging.innolitics.com/device/K234047)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).