ACS® LD FB Knee System

K234044 · Implantcast GmbH · JWH · Sep 6, 2024 · Orthopedic

Device Facts

Record IDK234044
Device NameACS® LD FB Knee System
ApplicantImplantcast GmbH
Product CodeJWH · Orthopedic
Decision DateSep 6, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3560
Device ClassClass 2
AttributesTherapeutic

Indications for Use

General total knee arthroplasty indications include: • Painful, disabling joint disease of the knee resulting from: degenerative arthritis, rheumatoid arthritis or post-traumatic arthritis - · Post-traumatic loss of knee ioint configuration and function - Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function · Revision of previous unsuccessful knee replacement or other soft tissue stability is adequate · Fracture of the distal femur and/or proximal tibia that cannot be standard fracture management techniques The posterior stabilized variant is also indicated for PCL instability requiring implant bearing surface geometries with increased anterior-posterior constraint and absent or non-functioning posterior cruciate ligament. The ACS® LD FB Knee System is intended for cemented use, single use only. The ACS® LD FB Knee System is intended for use in total knee arthroplasty in skeletally mature patients, to provide increased mobility and reduce pain by replacing the damaged knee joint articulation where there is evidence of sufficient sound bone to seat and support the components.

Device Story

Modular knee replacement system; replaces damaged knee joint articulation. Components include femoral components, tibial components, PE-inserts, stems, spacers, and patella replacements. Used in total knee arthroplasty to reduce pain and increase mobility. Implanted by orthopedic surgeons in clinical settings. System provides mechanical articulation; fixed bearing design. Cemented, single-use device. Benefits patient through restoration of joint function and stability.

Clinical Evidence

Bench testing only. No clinical data provided. Performance characterized via mechanical testing including femoral/tibial fatigue (ASTM F2083, F1800), taper/augment fatigue (ASTM F1875), interlocking strength (ASTM F1814), patellofemoral/tibiofemoral constraint and contact pressure (ASTM F1223, F2083), cone connection testing (ASTM F2722), and range of motion evaluation (ISO 21536).

Technological Characteristics

Modular, cemented, fixed-bearing knee prosthesis. Materials include metal and polymer components. Design features include femoral components, tibial components, PE-inserts, stems, and spacers. Testing conducted per ASTM F2083, F1800, F1875, F1814, F1223, F2722, and ISO 21536. No software or electronic components.

Indications for Use

Indicated for skeletally mature patients requiring total knee arthroplasty due to painful, disabling joint disease (degenerative, rheumatoid, or post-traumatic arthritis), post-traumatic loss of joint configuration, moderate varus/valgus/flexion deformity, revision of previous unsuccessful knee replacement, or distal femur/proximal tibia fractures. Posterior stabilized variant indicated for PCL instability. Contraindicated if insufficient bone stock exists to support components.

Regulatory Classification

Identification

A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. September 6, 2024 implantcast GmbH % Dave McGurl Vice President, Regulatory Affairs - Orthopedics MCRA, LLC 803 7th St NW, Floor 3 Washington, District of Columbia 20001 Re: K234044 Trade/Device Name: ACS® LD FB Knee System Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JWH Dated: August 6, 2024 Received: August 6, 2024 Dear Dave McGurl: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory {2}------------------------------------------------ assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Lixin Liu -S Lixin Liu, Ph.D Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use Submission Number (if known) K234044 Device Name ACS® LD FB Knee System Indications for Use (Describe) General total knee arthroplasty indications include: • Painful, disabling joint disease of the knee resulting from: degenerative arthritis, rheumatoid arthritis or post-traumatic arthritis - · Post-traumatic loss of knee ioint configuration and function - Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function · Revision of previous unsuccessful knee replacement or other soft tissue stability is adequate · Fracture of the distal femur and/or proximal tibia that cannot be standard fracture management techniques The posterior stabilized variant is also indicated for PCL instability requiring implant bearing surface geometries with increased anterior-posterior constraint and absent or non-functioning posterior cruciate ligament. The ACS® LD FB Knee System is intended for cemented use, single use only. The ACS® LD FB Knee System is intended for use in total knee arthroplasty in skeletally mature patients, to provide increased mobility and reduce pain by replacing the damaged knee joint articulation where there is evidence of sufficient sound bone to seat and support the components. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW! * The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ ## 510(k) Summary | Device Trade Name: | ACS® LD FB Knee System | |----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Manufacturer: | implantcast, GmbH<br>Lüneburger Schanze 26<br>2164 Buxtehude, Germany | | Contact: | Ms. Juliane Höppner<br>Head of Regulatory & Clinical & Biological Affairs<br>Phone: +49 4161 744 135<br>Email: j.hoeppner@implantcast.de | | Prepared by: | Mr. Dave McGurl<br>Vice President, Regulatory Affairs - Orthopedics<br>MCRA, LLC<br>803 7th St NW, Floor 3<br>Washington, DC 20001<br>Phone: 202.552.5800<br>dmcgurl@mcra.com | | Date Prepared: | December 20, 2023 | | Classification: | 21 CFR §888.3560 - Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis | | Class: | II | | Product Code: | JWH | | Primary Predicate<br>Device: | Corin USA Limited Unity™ Total Knee System (K173884) | | Additional Predicate<br>Devices: | Stryker Corporation Scorpio® NRG and Scorpio® TS Knee Systems<br>(K994128, K011643, K041591, K042343)<br>Medacta International, SA GMK® Primary and GMK® Revision<br>Knee Systems (K090988, K102437) | #### Indications for Use: General total knee arthroplasty indications include: - . Painful, disabling joint disease of the knee resulting from: degenerative arthritis, rheumatoid arthritis or post-traumatic arthritis - . Post-traumatic loss of knee joint configuration and function {5}------------------------------------------------ - . Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function - . Revision of previous unsuccessful knee replacement or other procedure, where soft tissue stability is adequate - Fracture of the distal femur and/or proximal tibia that cannot be stabilized by standard ● fracture management techniques The posterior stabilized variant is also indicated for PCL instability requiring implant bearing surface geometries with increased anterior-posterior constraint and absent or non-functioning posterior cruciate ligament. The ACS® LD FB Knee System is intended for cemented use, single use only. The ACS® LD FB Knee System is intended for use in total knee arthroplasty in skeletally mature patients, to provide increased mobility and reduce pain by replacing the damaged knee joint articulation where there is evidence of sufficient sound bone to seat and support the components. ## Device Description: The ACS® LD FB Knee System is a modular knee replacement system offering various components that can be combined to replace the knee joint with various options depending upon the size and anatomy of each patient. The ACS® LD FB Knee System consists of: - . ACS® LD Femoral Components - o ACS® LD Femoral Component - o ACS® LD Femoral Component Slim - o ACS® LD PS Femoral Component Slim - o ACS® LD PS Femoral Component - o ACS® LD SC Femoral Component - ACS® LD FB+ Tibial Components ● - o ACS® LD FB+ Tibia - o Locking Plug for ACS® FB Tibial Component - o Cone Plug for ACS® FB Tibial Component - ACS® FB+ PE-Inserts ● - o ACS® FB+ PE-Insert - ACS® FB+ PE-Insert Hyperflex o - ACS® FB+ PE-Insert Ultra o - o ACS® FB+ PS PE-Insert Hyperflex - o ACS® FB+ SC PE-Insert - ACS® Double Taper - ACS® Stem ● ● - o ACS® LD Extension Stem Male Taper - o ACS® Stem - ACS® Spacers . - o ACS® FB Tibial Spacer - o ACS® FB screw for spacer - o MK Femoral Spacer {6}------------------------------------------------ - o MK Screw for Spacer - ACS® Patella Replacements - o ACS® PE-Patella The ACS® LD FB Knee System is available as a non-coated (LD) fixed bearing (FB) version. The tibial and femoral components are available in cemented version. ### Substantial Equivalence: ● The ACS® LD FB Knee System is substantially equivalent to the predicate devices cited on the previous page with respect to intended use, design, and materials. #### Performance Testing: All necessary testing has been performed for the worst-case configuration of the ACS® LD FB Knee System to assure substantial equivalence to its predicates and to demonstrate the subject devices perform as intended. All testing was performed on test units representative of finished devices. The performance of the ACS® LD FB Knee System was characterized through the following tests: - Femoral Fatigue Testing (ASTM F2083) ● - Tibial Fatigue Testing (ASTM F1800) - Fatigue Testing of Tapers and Augments (ASTM F1875) - Interlocking Strength (ASTM F1814) - Patellofemoral Constraint Testing (ASTM F1223) - Patellofemoral Contact Area and Pressure Testing (ASTM F2083) - Posterior Stabilized Shear Fatigue Testing (ASTM F2083) - Cone Connection Testing (ASTM F2722) - Tibiofemoral Constraint Testing (ASTM F1223) ● - Tibiofemoral Contact Area and Pressure Testing (ASTM F2083) - Range of Motion Evaluation (ISO 21536) #### Conclusion: The ACS® LD FB Knee System possesses the same intended use and similar technological characteristics as the predicate devices. Therefore, the ACS® LD FB Knee System is substantially equivalent to the cited predicate devices.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...