K234030 · Clear Guide Medical · JAK · Jan 17, 2024 · Radiology
Device Facts
Record ID
K234030
Device Name
Clear Guide SCENERGY
Applicant
Clear Guide Medical
Product Code
JAK · Radiology
Decision Date
Jan 17, 2024
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1750
Device Class
Class 2
Indications for Use
The device is a stereotaxic accessory intended to provide fusion of images from certain imaging modalities. The device is intended to be used in any interventional or diagnostic procedure where the combination of these modalities is used for visualization, except for procedures on the brain. The device is intended for use in a clinical setting.
Device Story
Software system for image-guided procedures; inputs include real-time ultrasound stream, static CT/MRI volumes, and video feed from Clear Guide Optical Head. Uses optical detection of rigid instruments and fiducial markers to track position. Proprietary algorithms perform automated segmentation and registration to fuse static volumetric data with real-time imaging. Displays fused images and guidance overlays on a clinical monitor. Used by physicians in clinical settings for needle-based interventions (e.g., biopsy, ablation). Provides real-time visual feedback of target regions and instrument path, accounting for patient movement/deformation. Benefits include improved accuracy and navigation during interventional procedures.
Clinical Evidence
Clinical validation study (NCT05573048) with 10 patients. Evaluated workflow, intervention accuracy, and safety of modifications (transperineal grid support). Results confirmed device accurately achieved intended use; no adverse effects or new risks observed compared to cadaveric prostate ablation experiments.
Technological Characteristics
Optical detection via stereo cameras; software-based image fusion of static (CT/MRI) and real-time (US) data. Hardware: Clear Guide CORE and Optical Head (compliant with ES60601-1, IEC 60601-1-2). Connectivity: Integrated with ultrasound machines. Accessories: SteriMASK, VisiMARKER, TP Access tool, SteriGRID. Software: Automated segmentation and registration algorithms. Non-sterile.
Indications for Use
Indicated for fusion of CT, MR, and US images in patients ≥ 25 kg. US images can be fused with CT or MR. Used in interventional/diagnostic procedures (excluding brain) to display target regions and projected path of interventional instruments (e.g., biopsy/ablation needles) using Clear Guide CORE and Optical Head platform, accounting for patient movement/deformation.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
January 17, 2024
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA acronym and name on the right. The FDA part of the logo is in blue, with the acronym "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION" next to it.
Clear Guide Medical % Alican Demir Principal Engineer / Director of Quality Systems 3600 Clipper Mill Road, Suite 400 BALTIMORE, MD 21211
Re: K234030
Trade/Device Name: Clear Guide SCENERGY Regulation Number: 21 CFR 892.1750 Regulation Name: Computed Tomography X-Ray System Regulatory Class: Class II Product Code: JAK, IYO, LLZ, ITX Dated: November 12, 2023 Received: December 20, 2023
Dear Alican Demir:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerelv.
Lu Jiang
Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K234030
Device Name Clear Guide SCENERGY
#### Indications for Use (Describe)
The Clear Guide SCENERGY is software that provides fusion of images from Computed Tomography (CT). Magnetic Resonance (MR), and Ultrasound (US) modalities. US images can be fused with either CT or MR.
The Clear Guide SCENERGY utilizes the Clear Guide CORE and Clear Guide Optical Head platform to display images of the target regions and the projected path of the intervent, while taking into account patient movement and deformation. Instrumentation used with the Clear Guide SCENERGY might include an interventional needle or needle-like rigid device, such as a biopsy needle, or an ablation needle. The device is intended to be used in any interventional or diagnostic procedure where the combination of these modalities is used for visualization, except for procedures on the brain. The device is intended for use in a clinical setting, in patients of sufficient size (at least 25 kg).
Type of Use (Select one or both, as applicable)
| <span style="font-size: 10pt;"> <span style="font-family: Arial;"> <span style="font-weight: bold;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </span> |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size: 10pt;"> <span style="font-family: Arial;"> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K234030
Image /page/3/Picture/1 description: The image is a logo for Clear Guide Medical. The logo consists of a stylized image of a needle and the words "CLEAR GUIDE MEDICAL" in bold, sans-serif font. The words "CLEAR GUIDE" are in black, and the word "MEDICAL" is in gray.
# 510(k) Summary
(per 21 CFR § 807.92)
# Submitter's Information
| Name | Clear Guide Medical, Inc. |
|----------------|-----------------------------------------------------------------------|
| Address | 3600 Clipper Mill Road, Suite 400<br>Baltimore, MD 21211 USA |
| Phone Number | +1 (443) 602-8950 |
| Contact Person | Alican Demir, PhD<br>Principal Engineer / Director of Quality Systems |
| Date Prepared | January 15, 2024 |
# Device Information
| Trade Name | | Clear Guide SCENERGY | | |
|-------------------|-----|----------------------------------------------------------------------------------|--|--|
| Common Name | | System, Imaging Software | | |
| Classification(s) | JAK | Computed tomography x-ray system<br>Class II under 21 CFR 892.1750 | | |
| | | Subsequent/Secondary Classification(s) are: | | |
| | IYO | Ultrasonic pulsed echo imaging system<br>Class II under 21 CFR 892.1560 | | |
| | LLZ | Medical image management and processing system<br>Class II under 21 CFR 892.2050 | | |
| | ITX | Diagnostic ultrasound transducer<br>Class II under 21 CFR 892.1570 | | |
| | | | | |
# Predicate Device Information
| Device Name | Clear Guide SCENERGY (Clear Guide Medical, Inc.) |
|-------------------|--------------------------------------------------------------------------------------|
| 510(k) Number | K201188 |
| Common Name | System, Imaging Software |
| Classification(s) | JAK Computed tomography x-ray system<br>Class II under 21 CFR 892.1750 |
| | Subsequent/Secondary Classification(s) for all predicate devices are. |
| | IYO Ultrasonic pulsed echo imaging system<br>Class II under 21 CFR 892.1560 |
| | LLZ Medical image management and processing system<br>Class II under 21 CFR 892.2050 |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Clear Guide Medical. The logo consists of a stylized letter "L" formed by a thick black line intersecting with two thinner gray lines. To the right of the "L" is the text "CLEAR GUIDE" in bold, black letters, with the word "MEDICAL" underneath in a smaller, gray font.
# Device Description
The CLEAR GUIDE SCENERGY is a medical guidance and navigation software system that provides image fusion, instrument tracking, marker tracking, and target planning functionalities to operators during imageguided diagnostic or interventional procedures. The system uses data from multiple compatible imaging modalities that includes computed-tomography (CT), magnetic resonance imaging (MRI), or diagnostic ultrasound (US).
The CLEAR GUIDE SCENERGY uses optical detection technology to identify and track objects in the field of view. By pairing this information with the aforementioned imaging data. the CLEAR GUIDE SCENERGY executes proprietary software algorithms to display fused images in real-time to the clinician. These segmentation and registration algorithms are automated. Segmentation results are ning that new inputs (e.g., a new CT or MRI) would be required to change the segmentation output. Registration can be reset by the user at any time during use.
### Intended Use
The device is a stereotaxic accessory intended to provide fusion of images from certain imaging modalities. The device is intended to be used in any interventional or diagnostic procedure where the combination of these modalities is used for visualization, except for procedures on the brain. The device is intended for use in a clinical setting.
## Indications For Use
The Clear Guide SCENERGY is software that provides fusion of images from Computed Tomography (CT), Magnetic Resonance (MR), and Ultrasound (US) modalities. US images can be fused with either CT or MR.
The Clear Guide SCENERGY utilizes the Clear Guide CORE and Clear Guide Optical Head platform to display images of the target regions and the projected path of the interventional instrument. while taking into account patient movement and deformation. Instrumentation used with the Clear Guide SCENERGY might include an interventional needle or needle-like rigid device, such as a biopsy needle, or an ablation needle. The device is intended to be used in any interventional or diagnostic procedure where the combination of these modalities is used for visualization, except for procedures on the brain. The device is intended for use in a clinical setting, in patients of sufficient size (at least 25 kg).
## Technological Characteristics
Real-time marker tracking and instrument guidance is achieved by optical detection of needle-like rigid instrumentation by means of stereo cameras, and overlay of graphical indicators on the input imaging modality (Ultrasound, CT, or MR). Optionally, fiducial markers are attached to the patient and/or instrument to facilitate tracking.
Image fusion is the joining of static, volumetric, high-resolution imaging data (e.g., CT or MRI scans) with real-time, streamed image data (e.g., ultrasound). CLEAR GUIDE SCENERGY uses proprietary algorithms capable of combining three input sources (image stream from an ultrasound machine, a static CT/MRI volume, and video feed from the CLEAR GUIDE Optical Head) to provide an accurate visual feedback for the underlying image-guided procedure. Fusion of static imaging data (CT/MRI) is facilitated by the application of multi-modality fiducial markers prior the imaging. When Ultrasound and CT/MR fusion is enabled, the real-time ultrasound image stream is shown on the display together with dynamically computed CT/MB slices from the volumetric data, and guidance overlays.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image is a logo for Clear Guide Medical. The logo consists of two parts: a stylized graphic on the left and the company name on the right. The graphic on the left features a thick, black vertical line with two thinner, gray lines curving away from it. To the right of the graphic, the words "CLEAR GUIDE" are stacked on top of each other in bold, black letters, and below them, the word "MEDICAL" is written in gray.
A comparison of CLEAR GUIDE SCENERGY's technological characteristics to its predicate device is provided below.
| Category | Modified Device:<br>Clear Guide SCENERGY | Predicate Device:<br>Clear Guide SCENERGY |
|-------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K234030 | K201188 |
| Product Code | JAK | JAK |
| Classification Regulation | 21 CFR 892.1750 | 21 CFR 892.1750 |
| Regulatory Class | Class II | Class II |
| Classification Name | Computed tomography x-ray system | Computed tomography x-ray system |
| Subsequent<br>Product Code(s) | IYO, LLZ, ITX | IYO, LLZ |
| Intended Use(s)<br>(i.e., List of Main Functions) | - Image Fusion<br>- Instrument Tracking and Guidance | - Image Fusion<br>- Instrument Tracking and Guidance |
| Indications for Use | The Clear Guide SCENERGY is software that provides<br>fusion of images from Computed Tomography (CT), Mag-<br>netic Resonance (MR), and Ultrasound (US) modalities.<br>US images can be fused with either CT or MR.<br>The Clear Guide SCENERGY utilizes the Clear Guide<br>CORE and Clear Guide Optical Head platform to display<br>images of the target regions and the projected path of<br>the interventional instrument, while taking into account<br>patient movement and deformation. Instrumentation used<br>with the Clear Guide SCENERGY might include an in-<br>terventional needle or needle-like rigid device, such as<br>a biopsy needle, an aspiration needle, or an ablation<br>needle. The device is intended to be used in any inter-<br>ventional or diagnostic procedure where the combination<br>of these modalities is used for visualization, except for<br>procedures on the brain. The device is intended for use<br>in a clinical setting, in patients of sufficient size (at least<br>25 kg). | The Clear Guide SCENERGY is software that provides<br>fusion of images from Computed Tomography (CT), Mag-<br>netic Resonance (MR), and Ultrasound (US) modalities.<br>US images can be fused with either CT or MR.<br>The Clear Guide SCENERGY utilizes the Clear Guide<br>CORE and Clear Guide SuperPROBE platform to display<br>images of the target regions and the projected path of<br>the interventional instrument, while taking into account<br>patient movement and deformation. Instrumentation used<br>with the Clear Guide SCENERGY might include an in-<br>terventional needle or needle-like rigid device, such as<br>a biopsy needle, an aspiration needle, or an ablation<br>needle. The device is intended to be used in any inter-<br>ventional or diagnostic procedure where the combination<br>of these modalities is used for visualization, except for<br>procedures on the brain. The device is intended for use<br>in a clinical setting, in patients of sufficient size (at least<br>25 kg). |
| Principle of Operation<br>(i.e., Mechanism of Action) | Optical Detection & Software | Optical Detection & Software |
| Hardware Components | Clear Guide CORE and Optical Head | Clear Guide CORE and Optical Head |
| Conditions of Use | Image-guided Procedures | Image-guided Procedures |
| Fusion Capabilities | US-CT & US-MR | US-CT & US-MR |
| Input Imaging Data | CT, MR, or Ultrasound | CT, MR, or Ultrasound |
| Device Interactions | Connected/Anchored to Transducer | Connected/Anchored to Transducer |
| Segmentation Process | Automatic | Automatic |
| Registration Process | Automatic | Automatic |
| Fusion Algorithms | Automatic | Automatic |
| Tracking/Guidance<br>Algorithms | Automatic | Automatic |
| User Interface Modes | SETUP (3D), PLAN, NAVIGATION | SETUP (3D), NAVIGATION<br>continued |
| Category | Modified Device:<br>Clear Guide SCENERGY | Predicate Device:<br>Clear Guide SCENERGY |
| Supported Accessories<br>(Marketed separately) | - SteriMASK transducer cover kit<br>- VisiMARKER fiducial markers<br>- TP Access tool with SteriGRID needle guide grid | - SteriMASK transducer cover kit<br>- VisiMARKER fiducial markers |
| Intended User | Physician | Physician |
| Environment of Use | Clinical Setting | Clinical Setting |
| Duration of Use | ≤ 24 Hours | ≤ 24 Hours |
| Number of Uses | Reusable | Reusable |
| Sterility | Non-Sterile | Non-Sterile |
| Bench Testing | N/A (no modifications) | N/A (no modifications) |
| Animal Testing | N/A | N/A |
| Clinical Testing | Yes | Yes |
| Standards | N/A (for software) | N/A (for software)…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.