K234027 · Embecta Medical I, LLC · QFG · Aug 30, 2024 · Clinical Chemistry
Device Facts
Record ID
K234027
Device Name
embecta Insulin Delivery System
Applicant
Embecta Medical I, LLC
Product Code
QFG · Clinical Chemistry
Decision Date
Aug 30, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5730
Device Class
Class 2
Attributes
Therapeutic, PCCP
Indications for Use
The embecta Insulin Delivery System with interoperable technology (the Patch) is intended for subcutaneous delivery of insulin at set and variable rates for the management of diabetes mellitus in persons requiring insulin, for individuals 18 years of age and older. The Patch is able to reliably and securely communicate with compatible, digitally connected devices, including automated insulin dosing software, to receive, execute and confirm commands from these devices. The patch is intended for single patient, home use and requires a prescription.
Device Story
Disposable, wearable patch pump for subcutaneous insulin delivery; 300 U reservoir capacity; 72-hour wear time. Operated by a locked-down smartphone controller app via Bluetooth Low Energy (BLE) and Wi-Fi. System delivers basal and bolus insulin doses; provides status notifications (battery, reservoir, occlusion, faults). Used in home settings by patients; programmed by patient via smartphone app. Interoperable with automated insulin dosing software to receive/execute commands. Benefits include precise insulin delivery, remote programming, and integration with digital diabetes management ecosystems.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including basal and bolus delivery accuracy (n=90 pumps), occlusion detection timing, biocompatibility (ISO 10993), sterility (SAL 10^-6), electromagnetic compatibility, and human factors validation (n=16 lay users, n=16 HCPs).
Indicated for subcutaneous insulin delivery for diabetes mellitus management in patients 18 years and older. Contraindicated for patients unable to monitor glucose levels, maintain contact with healthcare providers, or follow system instructions.
Regulatory Classification
Identification
An alternate controller enabled infusion pump (ACE pump) is a device intended for the infusion of drugs into a patient. The ACE pump may include basal and bolus drug delivery at set or variable rates. ACE pumps are designed to reliably and securely communicate with external devices, such as automated drug dosing systems, to allow drug delivery commands to be received, executed, and confirmed. ACE pumps are intended to be used both alone and in conjunction with digitally connected medical devices for the purpose of drug delivery.
Special Controls
Alternate controller enabled infusion pumps must comply with the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include the following:
(i) Evidence demonstrating that device infusion delivery accuracy conforms to defined user needs and intended uses and is validated to support safe use under actual use conditions.
(A) Design input requirements must include delivery accuracy specifications under reasonably foreseeable use conditions, including ambient temperature changes, pressure changes (
*e.g.,* head-height, backpressure, atmospheric), and, as appropriate, different drug fluidic properties.(B) Test results must demonstrate that the device meets the design input requirements for delivery accuracy under use conditions for the programmable range of delivery rates and volumes. Testing shall be conducted with a statistically valid number of devices to account for variation between devices.
(ii) Validation testing results demonstrating the ability of the pump to detect relevant hazards associated with drug delivery and the route of administration (
*e.g.,* occlusions, air in line, etc.) within a clinically relevant timeframe across the range of programmable drug delivery rates and volumes. Hazard detection must be appropriate for the intended use of the device and testing must validate appropriate performance under the conditions of use for the device.(iii) Validation testing results demonstrating compatibility with drugs that may be used with the pump based on its labeling. Testing must include assessment of drug stability under reasonably foreseeable use conditions that may affect drug stability (
*e.g.,* temperature, light exposure, or other factors as needed).(iv) The device parts that directly or indirectly contact the patient must be demonstrated to be biocompatible. This shall include chemical and particulate characterization on the final, finished, fluid contacting device components demonstrating that risk of harm from device-related residues is reasonably low.
(v) Evidence verifying and validating that the device is reliable over the ACE pump use life, as specified in the design file, in terms of all device functions and in terms of pump performance.
(vi) The device must be designed and tested for electrical safety, electromagnetic compatibility, and radio frequency wireless safety and availability consistent with patient safety requirements in the intended use environment.
(vii) For any device that is capable of delivering more than one drug, the risk of cross-channeling drugs must be adequately mitigated.
(viii) For any devices intended for multiple patient use, testing must demonstrate validation of reprocessing procedures and include verification that the device meets all functional and performance requirements after reprocessing.
(2) Design verification and validation activities must include appropriate design inputs and design outputs that are essential for the proper functioning of the device that have been documented and include the following:
(i) Risk control measures shall be implemented to address device system hazards and the design decisions related to how the risk control measures impact essential performance shall be documented.
(ii) A traceability analysis demonstrating that all hazards are adequately controlled and that all controls have been validated in the final device design.
(3) The device shall include validated interface specifications for digitally connected devices. These interface specifications shall, at a minimum, provide for the following:
(i) Secure authentication (pairing) to external devices.
(ii) Secure, accurate, and reliable means of data transmission between the pump and connected devices.
(iii) Sharing of necessary state information between the pump and any digitally connected alternate controllers (
*e.g.,* battery level, reservoir level, pump status, error conditions).(iv) Ensuring that the pump continues to operate safely when data is received in a manner outside the bounds of the parameters specified.
(v) A detailed process and procedure for sharing the pump interface specification with digitally connected devices and for validating the correct implementation of that protocol.
(4) The device must include appropriate measures to ensure that safe therapy is maintained when communications with digitally connected alternate controller devices is interrupted, lost, or re-established after an interruption (
*e.g.,* reverting to a pre-programmed, safe drug delivery rate). Validation testing results must demonstrate that critical events that occur during a loss of communications (*e.g.,* commands, device malfunctions, occlusions, etc.) are handled appropriately during and after the interruption.(5) The device design must ensure that a record of critical events is stored and accessible for an adequate period to allow for auditing of communications between digitally connected devices and to facilitate the sharing of pertinent information with the responsible parties for those connected devices. Critical events to be stored by the system must, at a minimum, include:
(i) A record of all drug delivery
(ii) Commands issued to the pump and pump confirmations
(iii) Device malfunctions
(iv) Alarms and alerts and associated acknowledgements
(v) Connectivity events (
*e.g.,* establishment or loss of communications)(6) Design verification and validation must include results obtained through a human factors study that demonstrates that an intended user can safely use the device for its intended use.
(7) Device labeling must include the following:
(i) A prominent statement identifying the drugs that are compatible with the device, including the identity and concentration of those drugs as appropriate.
(ii) A description of the minimum and maximum basal rates, minimum and maximum bolus volumes, and the increment size for basal and bolus delivery, or other similarly applicable information about drug delivery parameters.
(iii) A description of the pump accuracy at minimum, intermediate, and maximum bolus delivery volumes and the method(s) used to establish bolus delivery accuracy. For each bolus volume, pump accuracy shall be described in terms of the number of bolus doses measured to be within a given range as compared to the commanded volume. An acceptable accuracy description (depending on the drug delivered and bolus volume) may be provided as follows for each bolus volume tested, as applicable: Number of bolus doses with volume that is <25 percent, 25 percent to <75 percent, 75 percent to <95 percent, 95 percent to <105 percent, 105 percent to <125 percent, 125 percent to <175 percent, 175 to 250 percent, and >250 percent of the commanded amount.
(iv) A description of the pump accuracy at minimum, intermediate, and maximum basal delivery rates and the method(s) used to establish basal delivery accuracy. For each basal rate, pump accuracy shall be described in terms of the amount of drug delivered after the basal delivery was first commanded, without a warmup period, up to various time points. The information provided must include typical pump performance, as well as worst-case pump performance observed during testing in terms of both over-delivery and under-delivery. An acceptable accuracy description (depending on the drug delivered) may be provided as follows, as applicable: The total volume delivered 1 hour, 6 hours, and 12 hours after starting delivery for a typical pump tested, as well as for the pump that delivered the least and the pump that delivered the most at each time point.
(v) A description of delivery hazard alarm performance, as applicable. For occlusion alarms, performance shall be reported at minimum, intermediate, and maximum delivery rates and volumes. This description must include the specification for the longest time period that may elapse before an occlusion alarm is triggered under each delivery condition, as well as the typical results observed during performance testing of the pumps.
(vi) For wireless connection enabled devices, a description of the wireless quality of service required for proper use of the device.
(vii) For any infusion pumps intended for multiple patient reuse, instructions for safely reprocessing the device between uses.
Predicate Devices
Omnipod DASH Insulin Management System with interoperable technology (K191679)
Submission Summary (Full Text)
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
## I Background Information:
A 510(k) Number
K234027
B Applicant
embecta Medical I, LLC
C Proprietary and Established Names
embecta Insulin Delivery System
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| QFG | Class II | 21 CFR 880.5730 - Alternate Controller Enabled Pump | CH - Clinical Chemistry |
| NDC | Class II | 21 CFR 868.1890 - Predictive pulmonary-function value calculator | CH - Clinical Chemistry |
## E Purpose for Submission:
New device
## II Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
B Indication(s) for Use:
The embecca Insulin Delivery System (the Patch) with interoperable technology is intended for subcutaneous delivery of insulin at set and variable rates for the management of diabetes mellitus in persons requiring insulin, for individuals 18 years of age and older. The Patch is able to reliably and securely communicate with compatible, digitally connected devices, including automated
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
insulin dosing software, to receive, execute and confirm commands from these devices. The Patch is intended for single patient, home use and requires a prescription.
## C Special Conditions for Use Statement(s):
This device is for prescription use only.
This device is intended for one-person use ONLY. Do not share the system with anyone – including other family members.
Insulin pump therapy is NOT recommended for people who are:
- Unable to monitor glucose levels as recommended by their health care provider.
- Unable to maintain contact with their health care provider.
- Unable to use the embecta Insulin Delivery System according to instructions.
Do Not adjust your Bolus Calculator or your Bolus Presets without the presence of your healthcare provider. Only your healthcare provider can help you change your Target and Minimum Glucose Level(s), Insulin-to-Carb Ratio, Insulin Sensitivity Factor, Duration of Insulin Action, and Bolus Presets. Incorrect settings can result in over delivery or under delivery of insulin. This can cause hypoglycemia (low BG) or hyperglycemia (high BG) events.
Before having an X-ray, magnetic resonance imaging (MRI), diathermy, electrocautery, or computed tomography (CT) scan (or any similar test or procedure), you'll need to disconnect and remove your patch. These types of tests and procedures can make your patch stop working. After your procedure is done, you can start over with a new patch.
The patch and controller are designed to use rapid-acting U-100 insulin. The following U-100 rapid acting insulin analogs have been tested and found to be safe for use in the pump: NovoLog or Humalog U-100 insulin for ages 18 years and older. NovoLog and Humalog are compatible with the system for use up to 72 hours (3 days). If you have questions about using other insulins, contact your health care provider. Always consult your health care provider and refer to the insulin labeling prior to use.
## III Device Description
The embecca Insulin Delivery System is a prescription home use device intended to support insulin therapy for diabetes mellitus management. The pump is a disposable Insulin Delivery Device that is controlled by a Bluetooth Low Energy enabled Wireless Controller App housed on a locked down Android smartphone. In addition, the pump and controller will be packed with a syringe and needle for filling. The pump is a single use disposable sterile tubeless patch pump device intended to be worn by the patient for a nominal period of up to 72 hours. The pump is adhered to the patient using a medical grade adhesive patch and delivers insulin using a pump-gear mechanism. The body-worn pump stores and administers up to 300U of user-filled U-100 insulin with a selected basal and bolus dosage setting that is agreed upon between the user and their healthcare practitioner or provider. The U-100 insulin is delivered to the patient subcutaneously and both the pump and the controller app will send feedback on interrupted or
K234027 - Page 2 of 11
{2}
fault status to the patient. This will include successful activation and status changes such as calculation of a low or empty insulin reservoir volume, and pump failure (expiry/near expiry, hardware, software, communication failures, and occlusion).
The embecta-provided smartphone will be non-sterile and is locked to run only the embecta controller app that enables controlling the pump discreetly. The controller will enable the user to pair, prime, and program a basal and bolus dose via wireless transmission to the pump as well as provide users with system alerts, including status information and notifications. This includes, but is not limited to, battery life, total insulin delivered, calculation of a low insulin reservoir volume, occlusion, and other possible device faults. The controller's app is designed to program and display the patient's basal insulin delivery rate, delivered bolus doses, historical data, and insulin usage on a color display touch screen. The controller can be recharged with the provided charger.
## IV Substantial Equivalence Information:
A Predicate Device Name(s):
Omnipod DASH Insulin Management System with Interoperable Technology
B Predicate 510(k) Number(s):
K191679
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K234027 | K191679 |
| --- | --- | --- |
| Device Trade Name | embecta Insulin Delivery System | Omnipod DASH Insulin Management System with Interoperable Technology |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | Intended for the subcutaneous delivery of insulin at variable rates for the management of diabetes mellitus in people requiring insulin. Intended to be interoperable with connected devices including CGMs and automated insulin | Same |
K234027 - Page 3 of 11
{3}
| | dosing algorithms. | |
| --- | --- | --- |
| Operating Environment | Home Use | Same |
| Insulin Delivery Modes | Both basal and bolus | Same |
| Bolus Dose Range | 0.05 – 30 U | 0.05 – 30 U |
| General Device Characteristic Differences | | |
| Specific Drug/Biologic Use | U-100 Insulin.
System has been tested with Novolog and Humalog. | U-100 Insulin.
System has been tested with Novolog, Humalog, Admelog and Apidra. |
| Reservoir Capacity | 300 U | 200 U |
V Standards/Guidance Documents Referenced:
- FDA Guidance “Infusion Pumps Total Product Life Cycle” dated December, 2014
- FDA Guidance “Applying Human Factors and Usability Engineering to Medical Devices” dated February, 2016
- FDA Guidance “Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions”, September, 2023
- FDA Guidance “Design Considerations and Pre-market Submission Recommendations for Interoperable Medical Devices” dated September 6, 2017
- FDA Guidance “Content of Human Factors Information in Medical Device Marketing Submissions”, dated December, 2022
- ISO 10993-1:2018 Biological Evaluation of Medical Devices: Evaluation and Testing
- ISO 10993-3:2014 Biological Evaluation of Medical Devices – Part 3: Tests for Genotoxicity, Carcinogenicity and Reproductive Toxicity
- ISO 10993-4:2017 Biological Evaluation of Medical Devices – Part 4: Selection of Tests for Interactions with Blood.
- ISO 10993-5:2009 Biological Evaluation of Medical Devices – Part 5: Tests for in Vitro Cytotoxicity.
- ISO 10993-6: 2016 Biological Evaluation of Medical Devices – Part 6: Tests for Local Effects after Implantation.
- ISO 10993-10:2021 Biological Evaluation of Medical Devices – Part 10: Tests for Irritation and Skin Sensitization.
- ISO 10993-11:2017 Biological Evaluation of Medical Devices – Part 11: Tests for Systemic Toxicity.
K234027 - Page 4 of 11
{4}
- ISO 10993-23:2021 Biological Evaluation of Medical Devices – Part 23: Tests for Irritation.
- ISO 14971 Third Edition 2019 Medical Devices – Application of Risk Management to Medical Devices
- IEC 60601-1-2 Medical electrical equipment – Part 1-2: General Requirements for Basic Safety and Essential Performance – Collateral Standard: Electromagnetic Disturbances- Requirements and Tests
- IEC 60601-1-6 Medical electrical equipment – Part 1-6: General requirements for basic safety and essential performance – Collateral standard: Usability
- IEC 60601-1-8 Medical Electrical Equipment – Part 1-8: General Requirements for Basic Safety and Essential Performance – Collateral Standard: General Requirements, Tests and Guidance for Alarm Systems in Medical Electrical Equipment and Medical Electrical Systems
- IEC 60601-1-11 Medical Electrical Equipment – Part 1-11: General Requirements for Basic Safety and Essential Performance – Collateral Standard- Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment
- IEC 62304:2015 Medical Device Software – Software Life Cycle Processes
- IEC 62366-1:2020 Medical Devices – Part 1: Application Of Usability Engineering To Medical Devices
- ISO 11607-1:2019 Packaging for Terminally Sterilized Medical Devices – Part 1: Requirements for Materials, Sterile Barrier Systems and Packaging Systems
- IEC 62366-1:2020 Medical Devices – Part 1: Application of Usability Engineering to Medical Devices, including Amendment 1.
## VI Performance Characteristics:
### A. Analytical Performance
#### 1. Basal delivery accuracy
To assess basal delivery accuracy at 0.05, 1.0 and 30.0 U/h, 44 aged pumps and 45 unaged pumps were tested. Delivery accuracy was assessed by pumping insulin into a container on a scale and measuring the weight of the liquid at 1h, 6h, and 12h intervals for minimum and intermediate rates (0.1 and 1.0 U/hr) and 1h and 6h intervals for maximum rate (25 U/hr).
The following tables report the typical basal performance (median) observed, along with the lowest and highest results observed for minimum, intermediate, and maximum basal rate settings for all pumps tested.
K234027 - Page 5 of 11
{5}
Table 1: Amount of fluid delivered after 1, 6, and 12 hours with 0.05 U/hr (minimum) basal rate setting
| 0.1 U/hr Basal Duration | 1 hour | 6 hours | 12 hours |
| --- | --- | --- | --- |
| Total expected delivery volume | 0.05 U | 0.30 U | 0.60 U |
| Median amount delivered [min, max] | 0.048 U
[0.00, 0.18] | 0.29 U
[0.07, 0.45] | 0.58 U
[0.27, 0.72] |
Table 2: Amount of fluid delivered after 1, 6, and 12 hours with 1 U/hr (intermediate) basal rate setting
| 1 U/hr Basal Duration | 1 hour | 6 hours | 12 hours |
| --- | --- | --- | --- |
| Total expected delivery volume | 1.0 U | 6.0 U | 12.0 U |
| Median amount delivered [min, max] | 0.99 U
[0.74, 1.20] | 5.98 U
[5.21, 6.14] | 11.97 U
[10.94, 12.26] |
Table 3: Amount of fluid delivered after 1, 6, and 12 hours with 30 U/hr (high) basal rate setting
| 25 U/hr Basal Duration | 1 hour | 6 hours |
| --- | --- | --- |
| Total expected delivery volume | 30 U | 180 U |
| Median amount delivered [min, max] | 30.13 U
[29.77, 30.58] | 180.49 U
[180.10, 180.93] |
2. Bolus delivery accuracy
To assess bolus delivery accuracy, a total of 90 pumps were used, half of which were aged. 30 pumps each were tested by delivering at minimum, intermediate, and maximum bolus volumes (0.05 U, 6.0 U, and 30 U).
Tables 4-6 below show the number (and %) of boluses within the specified range of the minimum bolus, one intermediate bolus and the maximum bolus volume.
K234027 - Page 6 of 11
{6}
Table 4: Amount of fluid delivered after a 0.05 U bolus request (n=15000 boluses)
| Units delivered after a 0.2 U bolus request (% of commanded units) | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | <0.0125
(<25%) | 0.0125 - 0.0375
(25-75%) | 0.0375- 0.045
(75-90%) | 0.045- 0.0475
(90-95%) | 0.0475- 0.0525
(95-105%) | 0.0525- 0.055
(105-110%) | 0.055- 0.0625
(110-125%) | 0.0625- 0.0875
(125-175%) | 0.0875- 0.125
(175-250%) | >0.125
(>250%) |
| Number and percent of boluses | 5/15000
(0.03 %) | 66/15000
(0.44 %) | 1678/150000
(11.19 %) | 4259/150000
(28.39 %) | 7077/15000
(47.18 %) | 1105/150000
(7.37 %) | 792/15000
(5.28 %) | 18/15000
(0.12 %) | 0/150000
(0 %) | 0/15000
(0 %) |
Table 5: Amount of fluid delivered after a 6.0 U bolus request (n=750 boluses)
| Units delivered after a 6.0 U bolus request (% of commanded units) | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | <1.50
(<25%) | 1.50-4.50
(25-75%) | 4.5-5.40
(75-90%) | 5.40-5.70
(90-95%) | 5.70-6.30
(95-105%) | 6.30-6.60
(105-110%) | 6.60-7.5
(110-125%) | 7.5-10.5
(125-175%) | 8.75-15.0
(175-250%) | >15.0
(>250%) |
| Number and percent of boluses | 0/750
(0.0%) | 0/750
(0.0%) | 0/750
(0.0%) | 1/750
(0.13%) | 749/750
(99.87%) | 0/750
(0.0%) | 0/750
(0.0%) | 0/750
(0.0%) | 0/750
(0.0%) | 0/750
(0.0%) |
Table 6: Amount of fluid delivered after a 30 U bolus request (n=300 boluses)
| Units delivered after a 30.0 U bolus request (% of commanded units) | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | <7.5
(<25%) | 7.5-22.5
(25-75%) | 22.5-27.0
(75-90%) | 27.0-28.5
(90-95%) | 28.5-31.5
(95-105%) | 31.5-33.0
(105-110%) | 33.0-37.5
(110-125%) | 37.5-52.5
(125-175%) | 52.5-75.0
(175-250%) | >75.0
(>250%) |
| Number and percent of boluses | 0/300
(0.0%) | 0/300
(0.0%) | 0/300
(0.0%) | 0/300
(0.0%) | 300/300
(100%) | 0/300
(0.0%) | 0/300
(0.0%) | 0/300
(0.0%) | 0/300
(0.0%) | 0/300
(0.0%) |
K234027 - Page 7 of 11
{7}
K234027 - Page 8 of 11
3. Occlusion detection
To assess occlusion detection performance and unintended bolus upon occlusion release performance, 160 pumps were tested, half of which were aged.
The following table depicts the typical (average) time to occlusion detection for three different situations when using U-100 insulin.
Table 7: Timing of occlusion detection alarms
| | Typical time to occlusion detection | Maximum time to occlusion detection |
| --- | --- | --- |
| 5.35 U Bolus* | 2 minutes 15 seconds | 3 minutes 34 seconds |
| 1.0 U/hr Basal | 4 hour 4 minutes | 5 hours 27 minutes |
| 0.05 U/hr Basal | 79 hours 59 minutes | 80 hours (Pump expiration) |
*A 30 U bolus was commanded, and the reported value represents maximum amount of fluid at occlusion detection.
B. Other Supportive Instrument Performance Characteristics Data
1. Hazard Analysis
A comprehensive hazard analysis was reviewed, in which design inputs and outputs, risks, and risk mitigations for hardware and software associated with proper functioning of the insulin pump were reviewed. The sponsor performed a hazard analysis to account for the unique intended use, design elements, and risks of their ACE pump. This analysis identified hazards which could reasonably be anticipated to impact the proper use of the device, traced all identified risks to adequate design controls, and demonstrated that design features were appropriately implemented and validated.
2. Human Factors
The results of Human Factors (HF) validation study for the embecca Insulin Delivery System, a patch style insulin pump, support substantial equivalence to the predicate device. The HF validation study focused on two user groups: people with diabetes mellitus who are 18+ years or older and treated as adults and healthcare providers (HCP). The people with diabetes mellitus consisted of 16 "lay users," where there were 6 pump-inexperienced, 6 pump-experienced, and 4 patients with caregivers represented as a dyad. The HCP group consisted of 15 certified diabetes care and education specialists and 1 clinical pharmacist. Users were trained according to a "train the trainer" model where trainers from embecca provided training to HCPs (registered nurses, diabetes educators, etc.) and the trained HCPs provided training to the people with diabetes. Lay-user training consisted of two one-on-one training sessions spread across two days. The first training focused on initial device set-up and basic
{8}
pump functionality and the second training focused on advanced pump functionality. There was a 1-hour decay period after training prior to validation testing. The tasks evaluated during simulated use testing consisted of administering a single rate basal program, manual boluses, adjusting time zones, temporary basal programs, multi-rate basal programs, bolus calculation based on carbohydrate and blood glucose values, pausing insulin delivery, alert customization, disposal of the device, and responding to alerts and alarms. Lay-users were not validated to set-up a new insulin pump independently and initial pump set-up, which consists of inputting pump settings such as setting up target glucose levels, preset boluses, insulin to carb ratio, insulin sensitivity factor, duration of insulin action, and minimum glucose level should occur under the supervision of a healthcare provider.
3. Biocompatibility
Biocompatibility testing was performed per FDA Guidance Document: Use of International Standard ISO 10993-1 “Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process,” and FDA special controls for alternate controller enabled infusion pump. Results support a determination of substantial equivalence.
4. Sterility
The patch pump and filling syringe are provided sterile in a sealed sterile barrier tray. These components are terminally sterilized in its final package using Ethylene Oxide and the process has been validated to ensure a Sterility Assurance Level (SAL) of 10⁻⁶ in accordance with ISO 11135: 2014.
5. Insulin Compatibility and Stability
The embecta system was found to be compatible with U-100 insulin Novolog (insulin aspart), and Humalog (insulin lispro) for up to 3 days.
6. Additional Bench Testing
In addition to the performance testing described above, mechanical testing, simulated use testing, and other device verification testing was conducted to demonstrate that the system meets its intended use and is safe, reliable, and all safety and reliability critical requirements have been adequately verified. Summaries for reliability, safety, and verification testing follow:
| Testing to Support System Reliability |
| --- |
| Pump lifetime testing |
| Transient Temperature |
| Transient Pressure |
| Ingress Protection |
| Cleaning |
| Label legibility |
K234027 - Page 9 of 11
{9}
| Against mechanical stressors (e.g., vibrations, shock) |
| --- |
| Against chemical stressors (e.g., sunscreens, insect repellents, chlorine and saltwater, cleaning agents) |
| Delivery Accuracy and Occlusion Detection Testing |
| Testing to Support System Safety |
| --- |
| Environmental Safety Testing to 60601-1-11 |
| Safety and Essential Performance Testing to 60601-1 |
| Worst Case Accuracy |
| Incidental Delivery |
| Occlusion Detection Testing |
| Leakage |
| User Guide |
| Testing to Support System Verification |
| --- |
| Environmental Safety Testing to 60601-1-11 |
| Safety and Essential Performance Testing to 60601-1 |
| Occlusion Detection |
| Insulin Delivery Verification |
| Environmental Conditions |
7. Electromagnetic Compatibility and Wireless Coexistence
Electromagnetic compatibility, electromagnetic immunity and wireless coexistence testing was performed for the pump. All tests demonstrated that the device would perform as expected in the home healthcare environment.
8. Basic Safety and Essential Performance (Electrical Safety)
The sponsor provided verification evidence for compliance with the IEC 60601-1 and applicable collateral standards. Verification results support the finding of substantial equivalent for this device.
9. Data Logging
The sponsor provided a summary of pump and controller logging capability which enable the device to record critical events including insulin delivery, pump commands and confirmations, connectivity states, malfunctions, and alarms. These were reviewed and found to be adequate.
10. Interoperability
A plan and approach for interoperability were provided according to the FDA Guidance "Design Considerations and Pre-market Submission Recommendations for Interoperable
K234027 - Page 10 of 11
{10}
"Medical Devices" and determined to be adequate to support and clearly specify expectations, requirements, and interface specifications to potential interoperable devices. In addition, their plan covered their approach to working with connected device companies regarding contractual approaches, interfaces for data communication and exchange, and post-market reporting procedures and responsibilities (e.g., who is responsible for investigating and reporting complaints, malfunctions, and adverse events).
The sponsor additionally provided validated software protocols intended to ensure secure, accurate, and reliable communication with digital interfacing devices, as well as failsafe design features to mitigate the risks associated with interruption of communication with digitally connected devices. These protocols were reviewed and found to be adequate.
## 11. Software and Cybersecurity
Detailed Information on software and cybersecurity of the device was reviewed and found acceptable.
## VII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
## VIII Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K234027- Page 11 of 11
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.