K234012 · bioMerieux, Inc. · LON · Mar 14, 2025 · Microbiology
Device Facts
Record ID
K234012
Device Name
VITEK COMPACT PRO
Applicant
bioMerieux, Inc.
Product Code
LON · Microbiology
Decision Date
Mar 14, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.1645
Device Class
Class 2
Indications for Use
The VITEK® COMPACT PRO is intended for the automated quantitative and/or qualitative antimicrobial susceptibility testing of isolated colonies for most clinically significant aerobic Gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus spp., and yeast. The VITEK® COMPACT PRO is also intended for the automated identification of most clinically significant anaerobic organisms and Corynebacterium species, fermenting and nonfermenting Gram-negative bacilli, Gram-positive organisms, fastidious organisms, and yeasts and yeast-like organisms.
Device Story
VITEK® COMPACT PRO is an automated instrument for antimicrobial susceptibility testing (AST) and microbial identification (ID) in clinical/industry laboratories. Input: isolated colonies from growth media, prepared as McFarland suspensions in VITEK® 2 reagent cards. Operation: instrument performs vacuum-filling of cards, heat-sealing, incubation at 35.5°C, and continuous optical monitoring. Optical readings (660 nm for AST; 428/568/680 nm for ID) are processed by VITEK® 2 System software to determine growth and susceptibility. Output: quantitative/qualitative susceptibility results and organism identification. The Advanced Expert System™ (AES) provides automated validation and phenotypic interpretation. Results are communicated via bidirectional interface to laboratory information systems (LIS). Benefits: standardized, automated testing workflow; rapid identification and susceptibility profiling to guide clinical antibiotic therapy.
Clinical Evidence
Clinical performance evaluation compared VITEK® COMPACT PRO to VITEK® 2 using clinically relevant strains across GN, GP, ST, and YS card classes. Results showed high agreement: AST GN (98.8% EA, 97.3% CA), AST GP (99.5% EA, 97.4% CA), AST ST (98.5% EA, 98.7% CA), and AST YS (100% EA, 100% CA). Quality control testing across 39 antimicrobial agents and 13 organisms showed >99% pass rate. Reproducibility testing demonstrated >95% to >97% reproducibility across all card classes.
Technological Characteristics
Microprocessor-controlled electromechanical system. Uses VITEK® 2 64-well ID/AST cards. Vacuum-fill process, automated heat-sealing. Incubation at 35.5°C ± 1°C. Optical reader: 660 nm (AST), 428/568/680 nm (ID). Self-calibrating optics. Connectivity: Bidirectional Computer Interface (BCI) to LIS. Software: VITEK® 2 System software with Advanced Expert System™ (AES) for phenotypic analysis.
Indications for Use
Indicated for automated identification and antimicrobial susceptibility testing of isolated colonies of clinically significant aerobic Gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus spp., yeast, anaerobic organisms, Corynebacterium species, fermenting/nonfermenting Gram-negative bacilli, Gram-positive organisms, and fastidious organisms. For use by trained professional laboratory personnel.
Regulatory Classification
Identification
A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”
Predicate Devices
VITEK® 2 SYSTEM (N50510 Supplement S082)
Submission Summary (Full Text)
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
INSTRUMENT ONLY
## I Background Information:
A 510(k) Number
K234012
B Applicant
bioMerieux Inc.
C Proprietary and Established Names
Vitek Compact Pro
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| LON | Class II | 21 CFR 866.1645 - Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility System | MI - Microbiology |
## II Submission/Device Overview:
A Purpose for Submission:
To obtain a substantial equivalence determination for the VITEK COMPACT PRO.
B Type of Test:
In vitro diagnostic test software used for identification and antimicrobial susceptibility testing of microorganisms.
## III Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
B Indication(s) for Use:
The VITEK COMPACT PRO is intended for the automated quantitative and/or qualitative antimicrobial susceptibility testing of isolated colonies for most clinically significant aerobic Gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus spp., and yeast.
The VITEK COMPACT PRO is also intended for the automated identification of most clinically significant anaerobic organisms and Corynebacterium species, fermenting and nonfermenting Gram-negative bacilli, Gram-positive organisms, fastidious organisms, and yeasts and yeast-like organisms.
C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
Following VITEK 2 ID & AST Cards can only be used with the VITEK COMPACT PRO:
GN, GP, ST, YS AST cards
ANC, BCL, CBC, GN, GP, NH, YST ID cards
IV Device/System Characteristics:
A Device Description:
The VITEK COMPACT PRO is intended for laboratory use by professional users who are trained in microbiology and good laboratory practices. The VITEK COMPACT PRO instrument is an automated instrument designed for use in both clinical and industry laboratories. The instrument performs sample well filling, incubation, and optical readings. The VITEK COMPACT PRO instrument is a two-step automated instrument for:
- Hydrating reagents with sample inoculum prepared from isolated colonies
- Pre-processing cards, incubating cards, and continuous reading for growth
The VITEK 2 Systems Software receives the instrument optical readings and performs analysis to report the ID/AST results. The instrument then ejects the used reagent card into the waste area for disposal. The system includes a VITEK COMPACT PRO instrument with an internal computer, monitor, keyboard, mouse, handheld barcode scanner, and USB hub. The software provided with the internal computer includes analysis and limited data management programs. A bidirectional computer interface (BCI) may transfer results automatically to the user's laboratory information system (LIS). A Quality Control System is available to track the quality control results of the test cards. The Advanced Expert System (AES, Clinical Use) is available to provide online, systematic validation of results and interpretation of resistant phenotypes found during susceptibility testing.
The VITEK COMPACT PRO is intended for organism identification and AST results (quantitative MIC values). Then, VITEK2 Systems softwares (e.g., LIS and AES) use the organism ID and antimicrobial agent MIC values to determine the qualitative categorical interpretation based on the best phenotype match.
K234012 - Page 2 of 10
{2}
K234012 - Page 3 of 10
# B Instrument Description Information:
1. Instrument Name:
VITEK COMPACT PRO
2. Specimen Identification:
Patient information may be entered manually or downloaded from a data management system. The instrument reads the bar code information from the cards and cassettes and automatically sends the information to the system software.
3. Specimen Sampling and Handling:
Isolated colonies are inoculated into saline and then the density was manually adjusted to a 0.5 McFarland standard using a VITEK 2 DENSICHECK. Specimens and a cassette are loaded into the instrument for the autoinoculation into test cards. Then, the cassette is manually transferred into the load/unload station for incubation.
4. Calibration:
Self-calibrating.
5. Quality Control:
Quality Control (QC) testing of the VITEK COMPACT PRO should be performed as dictated by the protocol in the package insert of each AST card and antimicrobial agent. The purpose of QC testing is to monitor performance of the VITEK COMPACT PRO as well as the proficiency of the laboratory personnel who use the system. The QC testing is performed with the manufacturer recommended QC strains. The QC workflow follows the same steps as the sample workflow. The QC results are generated and displayed on the user interface indicating the expected QC MIC range, the VITEK COMPACT PRO AST interpretive criteria call, and whether the relevant antimicrobial agents for the organism have passed or failed.
During validation testing of each AST card, indicated antimicrobial agents and CLSI-recommended reference strains are summarized in Table 1. Purity of all the isoaltes was confirmed by purity plates, with 100% of purity plates showing monomicrobial growth. Results are summarized in Table 2 and the quality control testing is acceptable.
Table 1. Quality Control Antimicrobial Agents and Strains Tested
| AST Card | Antimicrobial Agent | Strain |
| --- | --- | --- |
| GN AST card | Amoxicillin/clavulanic acid, Aztreonam, Cefazolin, Cefepime, Ceftazidime, Ceftriaxone, Ciprofloxacin, Ertapenem, ESBL, Meropenem/vaborbactam, Minocycline, Tigecycline, Tobramycin, Trimethoprim/sulfamethoxazole | Escherichia coli, ATCC 25922
Escherichia coli, ATCC 35218
Klebsiella pneumoniae, ATCC 700603
Klebsiella pneumoniae, ATCC BAA-2814
Pseudomonas aeruginosa ATCC 27853 |
{3}
| AST Card | Antimicrobial Agent | Strain |
| --- | --- | --- |
| GP AST card | Ampicillin, Cefoxitin Screen, Clindamycin, Daptomycin, Erythromycin, Gentamicin High Level, Linezolid, Moxifloxacin, Nitrofurantoin, Oxacillin, Streptomycin High Level, Tetracycline, Vancomycin | Enterococcus faecalis, ATCC 29212 Enterococcus faecalis, ATCC 51299 Staphylococcus aureus, ATCC 29213 Staphylococcus aureus, ATCC BAA-1026 |
| ST AST card | Benzylpenicillin, Ceftriaxone, Clindamycin, Erythromycin, Inducible Clindamycin Resistance, Levofloxacin, Linezolid, Tetracycline, Trimethoprim/sulfamethoxazole, Vancomycin | Staphylococcus aureus, ATCC 29213 Staphylococcus aureus, ATCC BAA-977 Streptococcus pneumoniae, ATCC 49619 |
| YS AST card | Caspofungin, Micafungin | Candida krusei, ATCC 6258 Candida parapsilosis, ATCC 22019 |
All gram-negative (GN), gram-positive (GP) and Streptococcus (ST) reproducibility isolates were sub-cultured twice onto trypticase soy agar and incubated appropriately at $35 \pm 2^{\circ}\mathrm{C}$ , for $18 - 24$ hours (GN and GP) or $18 - 20$ hours (ST) before testing. All yeast (YS) isolates were sub-cultured twice onto Sabouraud dextrose agar and incubated at $35 \pm 2^{\circ}\mathrm{C}$ for 24 hours. MIC values for each drug/organism combination were compared to the expected MIC range. Purity of all the suspensions was confirmed by purity plates. A minimum of 20 replicates of each QC strain were tested at each clinical trial site on the VITEK COMPACT PRO and the VITEK 260 instruments. Each quality control organism had no more than five MIC results per drug and instrument after each day's testing. Results are summarized in Table 2 stratified by each AST card. The cumulative reporting and stratification by each card is presented for simplicity and was based on the analysis for each drug/organism combination which revealed acceptable quality control test results.
Table 2. Quality Control Results Stratified by AST Card
| AST card | Number of MIC values within QC Range | |
| --- | --- | --- |
| | VITEK COMPACT PRO | VITEK 2 60 |
| GN AST card | 3469/3482 (99.6%) | 3480/3482 (99.9%) |
| GP AST card | 3386/3393 (99.8%) | 3384/3393 (99.7%) |
| ST AST card | 1124/1134 (99.1%) | 1123/1134 (99.0%) |
| YS AST card | 298/300 (99.3%) | 298/300 (99.3%) |
V Substantial Equivalence Information:
A Predicate Device Name(s):
VITEK 2 System
K234012 - Page 4 of 10
{4}
B Predicate 510(k) Number(s):
N50510/S82
C Comparison with Predicate(s):
Table 3: Predicate Comparison
| Device & Predicate Device(s): | K234012 VITEK COMPACT PRO | N50510/S0082 VITEK 2 System |
| --- | --- | --- |
| Device Trade Name | VITEK COMPACT PRO | VITEK 2 System |
| General Device Characteristic Similarities | | |
| Intended Use/Indications for Use | The VITEK COMPACT PRO is intended for the automated quantitative and/or qualitative antimicrobial susceptibility testing of isolated colonies for most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus spp., and yeast.
The VITEK COMPACT PRO is also intended for the automated identification of most clinically significant anaerobic organisms and Corynebacterium species, fermenting and nonfermenting gram-negative bacilli, gram-positive organisms, fastidious organisms, and yeasts and yeast-like organisms. | The VITEK 2 is intended to be used for the automated quantitative or qualitative susceptibility testing of isolated colonies for most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus agalactiae and S. pneumoniae. |
| Test Cards | VITEK 2 AST 64-well cards | Same |
| AST Card Inoculum | Isolated colonies prepared with a DENSICHECK densitometer and autoinoculation after loading specimens and AST card into the instrument. | Same |
| Filling and Sealer process | Vacuum filled and | Same |
K234012 - Page 5 of 10
{5}
| Device & Predicate Device(s): | K234012
VITEK COMPACT PRO | N50510/S0082
VITEK 2 System |
| --- | --- | --- |
| | automated heat seal | |
| Incubation and Reading | Automated, 35.5°C +/- 1°
with automated readings
every 15 minutes with 660
nm optics | Same |
| AST analysis | Analysis of changes in
growth based turbidity | Same |
| Advanced Expert
System | Yes | Same |
| Waste | Automated removal of
ejected test cards to waste
container at end of cycle | Same |
| LIS communication | Bidirectional Computer
Interface (BCI) | Same |
| Results Reported | Identification (ID),
quantitative (MIC) and/or
qualitative AST results. | Same |
| General Device
Characteristic Differences | VITEK COMPACT PRO | VITEK 2 System |
| Test Setup | VITEK FLEXPREP | Smart Carrier Station or
VITEK FLEXPREP |
| Card Capacity | 15 cards with licenses to
expand to 30 and 60 cards | 60 for VITEK 2 and
120 for VITEK 2 XL |
| Cards per Cassette | 10 | 15 |
| AST Dilution Preparation | Manual | Automated or manual |
## VI Standards/Guidance Documents Referenced:
- Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA
- Guidance for Content of Premarket Submissions for Device Software Functions
- Guidance for Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions
- Guidance for Electromagnetic Compatibility (EMC) of Medical Devices
The following FDA-recognized Consensus Standards were referenced and pertain to device and study design for the VITEK COMPACT PRO:
- IEC 60601-1-2 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests (Edition 4.1 2020-09)
K234012 - Page 6 of 10
{6}
- IEC 61010-1 Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 1: General requirements (Edition 3.1 2017-01 CONSOLIDATED VERSION)
- IEC 62471 Photobiological safety of lamps and lamp systems (First edition 2006-07)
## VII Performance Characteristics (if/when applicable):
## A Analytical Performance:
### 1. Precision/Reproducibility:
The reproducibility of MIC results was evaluated with VITEK COMPACT PRO. For this study, three instruments were used at three testing sites. The following VITEK AST test card panels were used during testing: Gram-Negative (GN), Gram-Positive (GP), Streptococcus (ST) and Yeast (YS). All GN, GP and ST reproducibility isolates were sub-cultured twice onto trypticase soy agar and incubated appropriately at $35 \pm 2^{\circ}\mathrm{C}$, for 18 – 24 hours (GN and GP) or 18 – 20 hours (ST) before testing. All yeast isolates were sub-cultured twice onto Sabouraud dextrose agar and incubated at $35 \pm 2^{\circ}\mathrm{C}$ for 24 hours. Each reproducibility isolate was tested on the VITEK COMPACT PRO using the manual dilution method. The reproducibility set was tested in triplicate for three days at three clinical trial sites per instrument producing at least 54 results per antimicrobial (2 organisms $\times$ 3 replicates $\times$ 3 sites $\times$ 3 days = 54). For all antimicrobials evaluated, best-case and worst-case reproducibility was $\geq 95\%$. Additionally, purity plates were analyzed for monomicrobial growth. All purity plates displayed monomicrobial growth. The list of organism groups and antimicrobial agents tested in the reproducibility study is shown in Table 4. Results are summarized in Table 5 stratified by each AST card. The cumulative reporting and stratification by each card is presented for simplicity and was based on the analysis for each drug/organism combination which revealed acceptable reproducibility results.
Table 4. List of Organism group and Antimicrobial Agents for Reproducibility Testing
| AST card | Organism group | Antimicrobial Agents |
| --- | --- | --- |
| GN AST card | Enterobacterales, Acinetobacter spp., and/or Pseudomonas aeruginosa | Amoxicillin/clavulanic acid, Aztreonam, Ceftazidime, Ciprofloxacin, Ceftriaxone, Ertapenem, Cefepime, Minocycline, Trimethoprim/sulfamethoxazole, Tigecycline, and Tobramycin |
| GP AST card | Enterococcus spp. and/or Staphylococcus spp. | Ampicillin, Clindamycin, Daptomycin, Erythromycin, Nitrofurantoin, Linezolid, Moxifloxacin, Oxacillin, Quinupristin/dalfopristin, Tetracycline, and Vancomycin |
| ST AST card | Streptococcus spp. | Ceftriaxone, Erythromycin, Clindamycin, Levofloxacin, Benzylpenicillin, Linezolid, Trimethoprim/sulfamethoxazole, Tetracycline, and Vancomycin |
| YS AST card | Candida spp. | Fluconazole and Micafungin |
K234012 - Page 7 of 10
{7}
Table 5. Reproducibility Results Stratified by AST card
| AST card | Antimicrobial Agents | Best case (%) | Worst case (%) |
| --- | --- | --- | --- |
| GN AST card | Site 1 | 564/567 (99.5%) | 561/567 (98.9%) |
| | Site 2 | 564/567 (99.5%) | 560/567 (98.8%) |
| | Site 3 | 566/567 (99.8%) | 560/567 (98.8%) |
| GN AST card total | | 1694/1701(99.6%) | 1681/1701 (98.8%) |
| GP AST card | Site 1 | 732/738 (99.2%) | 724/738 (98.1%) |
| | Site 2 | 731/738 (99.1%) | 728/738 (98.6%) |
| | Site 3 | 734/738 (99.5%) | 713/738 (96.6%) |
| GP AST card total | | 2197/2214 (99.2%) | 2165/2214 (97.8%) |
| ST AST card | Site 1 | 627/630 (99.5%) | 626/630 (99.4%) |
| | Site 2 | 622/630 (98.7%) | 617/630 (97.9%) |
| | Site 3 | 630/630 (100%) | 630/630 (100%) |
| ST AST card total | | 1879/1890 (99.4%) | 1873/1890 (99.1%) |
| YS AST card | Site 1 | 112/117 (95.7%) | 112/117 (95.7%) |
| | Site 2 | 116/117 (99.1%) | 116/117 (99.1%) |
| | Site 3 | 116/117 (99.1%) | 116/117 (99.1%) |
| YS AST card total | | 345/351 (98.3%) | 345/351 (98.3%) |
2. Linearity:
Not applicable.
3. Analytical Specificity/Interference:
Not applicable.
4. Accuracy (Instrument):
The accuracy of MICs obtained by the VITEK COMPACT Pro was evaluated testing representative strains of clinically relevant microorganisms with the VITEK AST cards. The test cards were categorized as each product class of gram-negative (GN), gram-positive (GP), Streptococcus (ST) and yeast (YS) with the list of indicated organisms per antimicrobial. All GN, GP and ST challenge isolates were sub-cultured twice onto trypticase soy agar and incubated appropriately at $35 \pm 2^{\circ}\mathrm{C}$ , for $18 - 24$ hours (GN and GP) or $18 - 20$ hours (ST) before testing. All yeast isolates were sub-cultured twice onto Sabouraud dextrose agar and incubated at $35 \pm 2^{\circ}\mathrm{C}$ for $24$ hours. Three different VITEK COMPACT PRO and VITEK 260 instruments operated by three different operators were used to test isolated colonies of representative isolates of different species of Enterobacterales, Acinetobacter spp., Pseudomonas aeruginosa, Staphylococcus spp., Streptococcus spp., Enterococcus spp., and Candida spp.. Table 6 shows the test antimicrobial agents and organism groups used for the AST challenge study.
Table 6. List of Organism group and Antimicrobial Agents for Challenge Study
| AST card | Organism group | Antimicrobial Agents |
| --- | --- | --- |
| GN AST card | Enterobacterales, Acinetobacter spp., | Amoxicillin/clavulanic acid, Aztreonam, Ceftazidime, Ciprofloxacin, Ceftriaxone, Cefazolin, Cefepime, ESBL Screen, Ertapenem, |
K234012 - Page 8 of 10
{8}
| AST card | Organism group | Antimicrobial Agents |
| --- | --- | --- |
| | and Pseudomonas aeruginosa | Meropenem/vaborbactam, Minocycline, Trimethoprim/sulfamethoxazole, Tigecycline, Tobramycin |
| GP AST card | Enterococcus spp., Staphylococcus spp, and Streptococcus spp. | Ampicillin, Cefoxitin Screen, Clindamycin, Daptomycin, Erythromycin, Gentamicin High Level Screen, Linezolid, Moxifloxacin, Nitrofurantoin, Oxacillin, Streptomycin High Level Screen, Tetracycline, Vancomycin |
| ST AST card | Streptococcus spp. | Benzylpenicillin, Ceftriaxone, Clindamycin, Erythromycin, Inducible Clindamycin Resistance Screen, Levofloxacin, Linezolid, Tetracycline, Trimethoprim/sulfamethoxazole, Vancomycin |
| YS AST card | Candida spp. | Caspofungin, Micafungin |
The MIC results obtained using the VITEK COMPACT PRO were compared to the MIC results obtained using the VITEK 2. Essential Agreement (EA) was defined as MIC results obtained from the VITEK COMPACT PRO that were within one doubling dilution of the MIC results obtained from the VITEK 2. Category Agreement (CA) was defined as MIC interpretations (S/SDD/I/R) that were the same between the VITEK COMPACT PRO and VITEK 2. Very major errors were defined as false susceptible results from the VITEK COMPACT PRO, major errors were defined as false resistance results from the VITEK COMPACT PRO, and minor errors were defined as results with minor discrepancies (i.e., an intermediate result reported as either resistant or susceptible, or vice versa). Since this is a method-to-method comparison, results were considered acceptable if the EA was $\geq 95\%$.
Overall performance of AST card results was $>95\%$ for GN AST card (n=625 including screening results), GP AST card (n=554 including screening results), ST AST card (n=437 including screening results), and YS AST card (n=89 results), which is acceptable.
For all species and antimicrobial agents in GN AST card, 576/583 (98.8%) of MIC results were within one doubling dilution of the comparator result and 567/576 (97.3%) of SIR categorizations were in agreement. For each microorganism and antimicrobial agent, essential agreement was $>95\%$ except ceftriaxone (91.3% EA). One very major error (false susceptible) occurred with Cefepime. One major error (false resistance) occurred with Meropenem/vaborbactam and Trimethoprim/sulfamethoxazole each. All purity plates (100%) exhibited monomicrobial growth.
For all species and antimicrobial agents in GP AST card, 427/429 (99.5%) of MIC results were within one doubling dilution of the comparator result and 418/429 (97.4%) of SIR categorizations were in agreement. For each microorganism and antimicrobial agent, essential agreement was $>95\%$. Two very major errors (false susceptible) occurred with Daptomycin. Two major errors (false resistance) occurred with Ampicillin. All purity plates (100%) exhibited monomicrobial growth.
For all species and antimicrobial agents in ST AST card, 387/393 (98.5%) of MIC results were within one doubling dilution of the comparator result and 388/393 (98.7%) of SIR
K234012 - Page 9 of 10
{9}
categorizations were in agreement. For each microorganism and antimicrobial agent, essential agreement was >95%. One very major error (false susceptible) occurred with Cetriaxone, Clindamycin, and Tetracycline. One major error (false resistance) occurred with Tetracycline. All purity plates (100%) exhibited monomicrobial growth.
For all species and antimicrobial agents in YS AST card, 89/89 (100%) of MIC results were within one doubling dilution of the comparator result and 89/89 (100%) of SIR categorizations were in agreement. For each microorganism and antimicrobial agent, essential agreement was >95%. No very major error (false susceptible) and major error (false resistance) occurred. All purity plates (100%) exhibited monomicrobial growth.
Given a large volume of data, Table 7 summarizes the results stratified by AST card only. The cumulative reporting and stratification by each AST card is presented for simplicity and was based on the analysis for each drug/organism combination which revealed acceptable high degree of agreement between VITEK COMPACT PRO and VITEK 2 instruments.
Table 7. Summary of AST Results, Stratified by AST Card
| AST card | Organism group | Total tested | # EA | % EA | Total Eval. | # EA Eval. | % EA Eval. | # CA | % CA | # S | # R | # vmj | # maj | # min |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| GN AST | Enterobacterales
Acinetobacter spp.
Pseudomonas aeruginosa | 583 | 576 | 98.8% | 232 | 225 | 97.0% | 567 | 97.3% | 302 | 229 | 1 | 2 | 13 |
| GP AST | Staphylococcus spp.
Enterococcus spp.
Streptococcus spp. | 429 | 427 | 99.5% | 192 | 190 | 99.0% | 418 | 97.4% | 272 | 130 | 2 | 2 | 7 |
| ST AST | Streptococcus spp. | 393 | 387 | 98.5% | 244 | 238 | 97.5% | 388 | 98.7% | 122 | 260 | 3 | 1 | 1 |
| YS AST | Candida spp. | 89 | 89 | 100% | 72 | 72 | 100% | 89 | 100% | 81 | 8 | 0 | 0 | 0 |
5. Carry-Over:
Not applicable.
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K234012 - Page 10 of 10
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.