K234002 · Boston Scientific · GEH · Mar 13, 2024 · General, Plastic Surgery
Device Facts
Record ID
K234002
Device Name
ICEfx Cryoablation System (FPRCH8000-02)
Applicant
Boston Scientific
Product Code
GEH · General, Plastic Surgery
Decision Date
Mar 13, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4350
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ICEfy™ Cryoablation System is indicated for use as a cryosurgical tool in the fields of general surgery, dermatology, neurology (including cryoanalgesia), thoracic surgery (with the exception of cardiac tissue), ENT, gynecology, proctology, and urology. This system is destroy tissue (including prostate and kidney tissue, liver metastases, tumors, and skin lesions) by the application of extremely cold temperatures. The ICEfx Cryoablation System has the following specific indications: · Urology - Ablation of prostate tissue in cases of prostate cancer and Benign Prostate Hyperplasia (BPH) · Oncology - Ablation of cancerous or malignant tissue and benign tumors, and palliative intervention. Palliation of pain associated with metastatic lesions involving bone in patients who have failed or are not candidates for standard radiation therapy. · Dermatology Ablation or freezing of skin cancers and other cutaneous disorders Destruction of warts or lesions, angiomas, sebaceous hyperplasia, basal cell tumors of the eyelid or canthus area, ulcerated basal cell tumors, dermatoffbromas, small hemangiomas, multiple warts, plantar warts, actinic and sebortheic keratosis, cavernous hemangiomas, peri-anal condylomata, and palliation of tumors of the skin · Gynecology - Ablation of malignant neoplasia or benign dysplasia of the female genitalia · General surgery - Palliation of tumors of the rectum, anal fissures, pilonidal cysts, and recurrent cancerous lesions, ablation of breast fibroadenomas • ENT - Palliation of tumors of the oral cavity and ablation of leukoplakia of the mouth · Thoracic surgery - (with the exception of cardiac tissue) · Proctology - Ablation of benign or malignant growths of the anus or rectum
Device Story
Mobile cryoablation system; uses Joule-Thomson effect with high-pressure argon gas circulating through closed-tip needles to induce tissue freezing. Features CX technology; internal heating element in needle enables active tissue thawing. System computer-controlled via touchscreen interface; allows user to monitor and control procedure. Used in clinical settings by physicians for minimally invasive tissue destruction. Output consists of controlled thermal energy delivery to target tissue; enables ablation of tumors, lesions, and diseased tissue. Smaller, lighter weight, and fewer needle ports than predicate; does not support helium thaw option.
Clinical Evidence
Bench testing only. Verification and validation included system, electrical, mechanical, packaging, labeling, software, and usability testing. No new clinical data generated for this submission; relies on established safety/effectiveness of predicate device for indicated uses, including pain palliation of metastatic bone lesions.
Technological Characteristics
Mobile cryosurgical unit; Joule-Thomson gas expansion principle; high-pressure argon gas source; electrical heating (CX technology) for thawing. Touchscreen user interface. Smaller form factor than predicate. Software-controlled. Sterilization method not specified.
Indications for Use
Indicated for patients requiring cryosurgical tissue destruction in urology (prostate cancer, BPH), oncology (malignant/benign tumors, bone metastasis pain palliation), dermatology (skin cancers, warts, lesions), gynecology (genital neoplasia/dysplasia), general surgery (rectal tumors, breast fibroadenomas), ENT (oral tumors, leukoplakia), thoracic surgery (non-cardiac), and proctology (anal/rectal growths).
Regulatory Classification
Identification
(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.
{0}------------------------------------------------
March 13, 2024
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left side of the logo is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Boston Scientific Benjamin Van Santen Regulatory Affairs Specialist 1 SciMed Place Maple Grove, Minnesota 55311
Re: K234002
Trade/Device Name: ICEfx™ Cryoablation System Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical Unit And Accessories Regulatory Class: Class II Product Code: GEH Dated: December 18, 2023 Received: December 19, 2023
Dear Benjamin Van Santen:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jesse Muir, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K234002
Device Name ICEfx™ Cryoablation System
#### Indications for Use (Describe)
The ICEfy™ Cryoablation System is indicated for use as a cryosurgical tool in the fields of general surgery, dermatology, neurology (including cryoanalgesia), thoracic surgery (with the exception of cardiac tissue), ENT, gynecology, proctology, and urology. This system is destroy tissue (including prostate and kidney tissue, liver metastases, tumors, and skin lesions) by the application of extremely cold temperatures.
The ICEfx Cryoablation System has the following specific indications:
· Urology - Ablation of prostate tissue in cases of prostate cancer and Benign Prostate Hyperplasia (BPH)
· Oncology - Ablation of cancerous or malignant tissue and benign tumors, and palliative intervention. Palliation of pain associated with metastatic lesions involving bone in patients who have failed or are not candidates for standard radiation therapy.
· Dermatology Ablation or freezing of skin cancers and other cutaneous disorders
Destruction of warts or lesions, angiomas, sebaceous hyperplasia, basal cell tumors of the eyelid or canthus area, ulcerated basal cell tumors, dermatoffbromas, small hemangiomas, multiple warts, plantar warts, actinic and sebortheic keratosis, cavernous hemangiomas, peri-anal condylomata, and palliation of tumors of the skin
· Gynecology - Ablation of malignant neoplasia or benign dysplasia of the female genitalia
· General surgery - Palliation of tumors of the rectum, anal fissures, pilonidal cysts, and recurrent cancerous lesions, ablation of breast fibroadenomas
• ENT - Palliation of tumors of the oral cavity and ablation of leukoplakia of the mouth
· Thoracic surgery - (with the exception of cardiac tissue)
· Proctology - Ablation of benign or malignant growths of the anus or rectum
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| Device Trade Name: | ICEfx™™ Cryoablation System |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer: | Boston Scientific Corporation<br>300 Boston Scientific Way<br>Marlborough, MA 01752 |
| Manufacturing Site: | Boston Scientific Corporation<br>4100 Hamline Ave North<br>St. Paul, MN 55112 |
| Primary Contact: | Ben Van Santen<br>Regulatory Affairs Specialist<br>Boston Scientific Corporation<br>Cell: 763-923-5435<br>Email: benjamin.vansanten@bsci.com |
| Alternate Contact: | Rachel Owens<br>Senior Regulatory Affairs Manager<br>Boston Scientific Corporation<br>Cell: 763-273-6865<br>Email: Rachel.Owens@bsci.com |
| Date Prepared: | December 18, 2023 |
| Classifications: | 21 CFR 878.4350 Cryosurgical unit and accessories |
| Class: | II |
| Product Codes: | GEH |
{4}------------------------------------------------
Primary Predicate: Visual-ICE™ Cryoablation System (K230551)
Additional Predicate: N/A
### Device Description:
The ICEfx Cryoablation System is a mobile system intended for cryoablative tissue destruction using a minimally invasive procedure. The system is computer-controlled with a touch screen user interface that allows the user to control and monitor the procedure. The therapy delivered by the system is based on the Joule-Thomson effect displayed by compressed gases. The ICEfx System uses high-pressure argon gas that circulates through closed-tip cryoablation needles to induce tissue freezing. Active tissue thawing is achieved by the use of CX technology in which a heating element inside the cryoablation needle can be energized to cause thawing.
| Description of Submitted<br>Device: | Comments related to Predicate: |
|------------------------------------------------|----------------------------------------------------------------------------|
| ICEfx Cryoablation System | Visual-ICE Cryoablation System (K230551) |
| Design and Construction | |
| Console | Similar to predicate; ICEfx is smaller and lighter weight |
| Needle ports | ICEfx contains fewer needle ports |
| Channel Lock | Same as predicate |
| User interface | Similar to predicate; touchscreen with equivalent<br>function and features |
| Needle Interface | Same as predicate |
| Performance and Function | |
| Principles of Operation:<br>Freezing / Thawing | Same as predicate; ICEfx does not support the option<br>for helium thaw |
| CX Technology for Electrical<br>Thawing | Same as predicate |
| Needle Compatibility | Same as predicate; ICEfx supports fewer needles |
The table below provides a summary comparison of the submitted device compared to the predicate device.
In summary, the submitted ICEfx Cryoablation System has the same technology and principle of operation as the predicate device.
{5}------------------------------------------------
## Performance Testing Summary:
A full battery of verification and validation testing was conducted on the ICEfx Cryoablation System to ensure that the design, functionality, and performance met the specified requirements. Testing was conducted according to protocols based on international standards and in-house requirements. Testing included system testing, electrical testing, mechanical testing, packaging and labeling testing, software testing, design and usability testing.
System testing assessed whether the functional requirements of the system as a whole were satisfied by the design. Mechanical testing evaluated the mechanical robustness of the system and sub-systems, functional testing of the gas system, and tests of mechanical safety requirements. Electrical testing assessed the functional aspects of the circuit assemblies, tests of electromagnetic compatibility and immunity (EMC/EMI), and tests of electrical safety requirements. Labeling verification evaluated user manual and labeling accuracy with respect to design requirements and risk mitigations. Software testing exercised individual units of software as well as tests of the functionality of the entire software package. Validation testing included design and usability testing. Test results demonstrated that the ICEfx Cryoablation System meets defined specifications, is substantially equivalent to the predicate device, and does not raise any new issues of safety and effectiveness for its intended use.
Additionally, the safety and effectiveness of use of the predicate Visual-ICE Cryoablation System for pain palliation of metastatic lesions in bone has been evaluated in a clinical study. Results of the clinical study demonstrated that the specific use did not introduce new risks not normally associated with the general use of the device and demonstrated the safety and effectiveness of the system when used for the specific indication of pain palliation of metastatic lesions in bone.
### Conclusion (Substantial Equivalence):
The subject device is identical in technology and functionality to the primary predicate Visual-ICE Cryoablation System. The subject Traditional 510(k) proposes a specific indication which represents a subset of the general indications cleared for the Visual-ICE Cryoablation System. The data provided in this Traditional 510(k) Premarket Notification supports a determination that the ICEfx Cryoablation System is substantially equivalent to the legally marketed predicate device, with regard to performance, safety, and effectiveness for its intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.