K233999 · Tepha, Inc. · OOD · Apr 9, 2024 · General, Plastic Surgery
Device Facts
Record ID
K233999
Device Name
GalaFLEX LITE Scaffold
Applicant
Tepha, Inc.
Product Code
OOD · General, Plastic Surgery
Decision Date
Apr 9, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
GalaFLEX LITE Scaffold is intended to reinforce soft tissue, where weakness exists, in patients undergoing plastic and reconstructive surgery, or for use in procedures involving soft tissue repair, such as the repair of fascial defects that require the addition of a reinforcing or bridging material to obtain the desired surgical result.
Device Story
GalaFLEX LITE Scaffold is a sterile, single-use, bioabsorbable surgical mesh; manufactured from poly-4-hydroxybutyrate (P4HB) polymer extruded into monofilament fibers and knitted into a scaffold. Device is intended for use by surgeons in clinical settings to reinforce soft tissue where weakness exists or to repair fascial defects. Upon implantation, the scaffold provides mechanical reinforcement; it is absorbed by the body via hydrolysis and hydrolytic enzymatic digestion over 18 to 24 months. The device is provided as a nondyed (clear) scaffold. It functions as a permanent implant with tissue/bone contact (>30 days).
Clinical Evidence
No clinical data. Evidence consists of bench testing and an in vivo rabbit study evaluating performance specifications and absorption time. Testing demonstrated the device met all performance requirements and is substantially equivalent to the predicate.
Technological Characteristics
Material: Poly-4-hydroxybutyrate (P4HB) polymer. Form: Knitted monofilament fiber scaffold. Sterilization: Ethylene oxide (SAL 10^-6). Classification: Class II, 21 CFR 878.3300. Packaging: Tyvek envelope in foil pouch. Absorption: 18-24 months via hydrolysis/enzymatic digestion.
Indications for Use
Indicated for patients undergoing plastic and reconstructive surgery or soft tissue repair requiring reinforcement or bridging of fascial defects.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 09, 2024
Tepha, Inc Rajagopalan Prithi Regulatory Affairs Specialist 99 Hayden Avenue Suite 360 Lexington, Massachusetts 02421
Re: K233999
Trade/Device Name: GalaFLEX LITE Scaffold Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: Class II Product Code: OOD Dated: December 18, 2023 Received: December 18, 2023
#### Dear Rajagopalan Prithi:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{1}------------------------------------------------
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product: and 21 CFR 820.100. Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.
{2}------------------------------------------------
See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Tek N. Digitally signed by Tek N. Lamichhane -S Lamichhane -Date: 2024.04.09 S 08:35:21 -04'00' Tek N. Lamichhane, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
Submission Number (if known)
K233999
Device Name
GalaFLEX LITE Scaffold
#### Indications for Use (Describe)
GalaFLEX LITE Scaffold is intended to reinforce soft tissue, where weakness exists, in patients undergoing plastic and reconstructive surgery, or for use in procedures involving soft tissue repair, such as the repair of fascial defects that require the addition of a reinforcing or bridging material to obtain the desired surgical result.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
Prepared on: 2023-12-18
### Contact Details
| Applicant Name | Tepha, Inc. | |
|--------------------------------------------------|-------------------------------------------------------------|---------------------|
| Applicant Address | 99 Hayden Avenue Suite 360 Lexington MA 02421 United States | |
| Applicant Contact Telephone | 4259001327 | |
| Applicant Contact | Mrs. Prithi Rajagopalan | |
| Applicant Contact Email | Prithi.Rajagopalan@bd.com | |
| Device Name | | 21 CFR 807.92(a)(2) |
| Device Trade Name | GalaFLEX LITE Scaffold | |
| Common Name | Surgical mesh | |
| Classification Name | Surgical Film | |
| Regulation Number | 878.3300 | |
| Product Code | OOD | |
| Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| K113721 | TephaFLEX Light Mesh | OOD |
| Device Descrintion Summary<br>1 CED 207 00/2\/1\ | | |
GalaFLEX LITE Scaffold is a sterile, single-use device for prescription use only. It is a bioabsorbable surgical mesh manufactured from poly-4-hydroxybutyrate (P4HB). P4HB is a biologically derived polymer which is extruded into monofilament fibers and knitted into a surgical scaffold. The P4HB material is the predicate device, TephaFLEX Light Mesh (K113721) and the reference device, Phasix Mesh (K142818) in terms of formulation and the manufacturing processes. P4HB is absorbed by the body through a process of hydrolysis and hydrolytic enzymatic digestion typically within 18 to 24 months. The material components and manufacturing processes used to produce the subject device is substantially equivalent to the subject device will only be offered as a nondyed (clear) surgical scaffold. The subject device has the predicate device: soft tissue repair/reinforcement.
Phasix Mesh (K142818) is used as a reference device specification values of the subject device are being harmonized with the BD P4HB product line, particularly Phasix Mesh (K142818), following the acquisition of Tepha, Inc. by BD in July 2021.
# Intended Use/Indications for Use
GalaFLEX LITE Scaffold is intended to reinforce soft tissue, where weakness exists, in patients undergoing plastic and reconstructive surgery, or for use in procedures involving soft tissue repair of fascial defects that require the addition of a reinforcing or bridging material to obtain the desired surgical result.
# Indications for Use Comparison
The subject device and the cited predicate/reference devices have identical intended uses of soft tissue repair/reinforcement. The subject device, and the predicate have similar in use. The subject device and the reference device have similar indications of
### 21 CFR 807.92(a)(5)
21 CFR 807.92(a)(5)
{5}------------------------------------------------
use wherein the reference device is intended for use specifically in hernia repairs and the subject device is used in general soft tissue repair/reinforcement and this does not constitute a new intended use.
# Technological Comparison
The intended use, materials, technological characteristics (i.e., design, material, chemical composition), packaging materials/ components, sterilization (ethylene oxide with SAL 10^-6), principal of operation, and manufacturing process for the subject device and the predicate device, TephaFLEX Light Mesh (K113721) are substantially equivalent. Since the and predicate device are similar, no new safety and effectiveness concerns are raised.
21 CFR 807.92(a)(6)
Refer to the 'Technological Comparison Table' attached below for a comparison of the technological characteristics for the subject device, GalaFLEX LITE Scaffold, and the predicate device, TephaFLEX Light Mesh (K113721).
# Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
FDA Guidance "Guidance for the Preparation of a Premarket Notification for a Surgical Mesh" issued March 02, 1999 was used to determine appropriate bench and in vivo testing for the subject device includes the physical and functional characteristics as recommended in the Surgical Mesh quidance. In vivo rabbit study for the subject device includes the performance specifications and absorption time.
Clinical Tests Summary: Not Applicable.
The subject device met all of the performance requirements as described. All samples utilized in the performance testing were representative of the subject device in terms of device design, manufacturing, packaqing and sterilization. The testing results demonstrate that GalaFLEX LITE Scaffold is substantially equivalent to the predicate device, TephaFLEX Light Mesh (K113721).
{6}------------------------------------------------
| Device<br>Characteristics | GalaFLEX LITE Scaffold<br>(Subject Device) | TephaFLEX Light Mesh (K113721)<br>(Predicate Device) |
|----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device<br>Classification | Identical to predicate device. | Class II<br>Surgical Mesh 21 CFR 878.3300<br>Product Code: OOD |
| Material | Identical to predicate device. | Poly-4-hydroxybutyrate (P4HB) |
| Device Sizes | Identical to predicate device. | Offered in single sheets of varying shapes<br>and sizes ranging from 1x2 inches to 10x14<br>inches and may be cut to the shape or sizes<br>desired for a specific application. |
| Intended Use | Identical to predicate device. | Soft tissue repair/reinforcement |
| Indications for<br>Use | GalaFLEX LITE™ Scaffold is intended to<br>reinforce soft tissue where weakness exists<br>in patients undergoing plastic and<br>reconstructive surgery, or for use in<br>procedures involving soft tissue repair, such<br>as the repair of fascial defects that require<br>the addition of a reinforcing or bridging<br>material to obtain the desired surgical result. | TephaFLEX Light Mesh is intended to<br>reinforce soft tissue where weakness exists<br>in patients undergoing plastic and<br>reconstructive surgery, or for use in<br>procedures involving soft tissue repair, such<br>as the repair of hernia or other fascial<br>defects that require the addition of a<br>reinforcing or bridging material to obtain<br>the desired surgical result. |
| Patient<br>Contact<br>Classification | Identical to predicate device. | Permanent implant device with tissue/bone<br>contact (>30 days) |
| Packaging<br>materials and<br>configurations | Identical to predicate device. | The device is held within a medical grade<br>Tyvek envelope which is placed in a foil<br>pouch. The pouch is placed into an SBS<br>carton. |
| Sterilization | Identical to predicate device. | Ethylene Oxide SAL < 10-6 |
### K233999 510k Summary: Technological Comparison Table
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.