K233982 · Cerenovus, Inc. · QJP · May 9, 2024 · Cardiovascular
Device Facts
Record ID
K233982
Device Name
CEREGLIDE 92 Catheter System
Applicant
Cerenovus, Inc.
Product Code
QJP · Cardiovascular
Decision Date
May 9, 2024
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The CEREGLIDE 92 Catheter System is indicated for use in facilitating the insertion and guidance of appropriately sized interventional devices into a selected blood vessel in the neurovascular system.
Device Story
The CEREGLIDE 92 Catheter System is a neurovascular intervention tool used by physicians to navigate blood vessels. It consists of an intermediate catheter, an INNERGLIDE 9 delivery aid, a rotating hemostasis valve (RHV), and slit introducers. The system is introduced over a steerable guidewire. The intermediate catheter features variable stiffness, a PTFE-lined lumen, and a hydrophilic coating to reduce friction. The INNERGLIDE 9 delivery aid facilitates the delivery of the intermediate catheter to the target vessel. The device is used in a clinical setting under fluoroscopic guidance, enabled by radiopaque markers on the catheter and delivery aid. The system allows for the delivery of other interventional devices, aiding in neurovascular procedures. The device is single-use and sterile.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including design verification (visual inspection, dimensional analysis, leak resistance, tensile strength, particulate count, coating durability, kink resistance, torque performance, trackability, burst pressure) and in vitro usability studies in a neurovascular model.
Technological Characteristics
Variable stiffness single-lumen catheter; materials include PTFE (liner), stainless steel/tungsten (braid/marker), Pebax, Urethane, Nylon, Makrolon, Vestamid (jackets/hubs). Hydrophilic coating. Sterilized via 100% Ethylene Oxide (ISO 11135). Complies with ISO 10555-1, ISO 80369-7, and ISO 10993-1. No software or electronic components.
Indications for Use
Indicated for facilitating insertion and guidance of interventional devices into neurovascular blood vessels. No specific patient population or contraindications defined.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
May 9, 2024
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and reads "FDA U.S. FOOD & DRUG ADMINISTRATION".
Cerenovus, Inc. Niall Fox Director of Regulatory Affairs 6303 Waterford District Drive, Suites 215 & 315 Miami. Florida 33126
### Re: K233982
Trade/Device Name: CEREGLIDE 92 Catheter System Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: QJP Dated: April 9, 2024 Received: April 9, 2024
### Dear Niall Fox:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
## Naira Muradyan -S
Naira Muradyan, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K233982
Device Name CEREGLIDE 92 Catheter System
Indications for Use (Describe)
The CEREGLIDE 92 Catheter System is indicated for use in facilitating the insertion and guidance of appropriately sized interventional devices into a selected blood vessel in the neurovascular system.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary K233982
| III. Device<br>Information | Table 1. Device Information | | | |
|----------------------------------------|-----------------------------|------------------------------------------|---------------|-----------------|
| | Device Proprietary Name | CEREGLIDE 92 Catheter System | | |
| | Common or Usual name | Catheter, Percutaneous, Neurovasculature | | |
| | Classification Name | 21 CFR 870.1250 - Percutaneous Catheter | | |
| | Regulatory Classification | II | | |
| | FDA Product Code | QJP | | |
| IV. Predicate<br>Device<br>Information | Table 2. Predicate Device | | | |
| | 510(k) Number | Date Cleared | Name | Manufacturer |
| | K230726 | November 29, | CEREGLIDE™ 92 | Cerenovus, Inc. |
2023
### V. Device Description
The CEREGLIDE™ 92 Catheter System consists of the CEREGLIDE™ 92 Intermediate Catheter and associated accessories, including the INNERGLIDE™ 9 Delivery Aid, Rotating Hemostasis Valve (RHV), and Slit Introducers.
Intermediate Catheter
The CEREGLIDE™ 92 Intermediate Catheter is a single lumen, variable stiffness catheter designed to be introduced over a steerable guidewire along with a microcatheter and/or compatible support device into the neuro vasculature. The catheter consists of a lubricious PTFE lined inner lumen to facilitate movement of the guidewires and other devices, variable pitch stainless steel and tungsten braid, and various durometer polymer jackets. These jackets provide distal flexibility and gradually transition to a stiffer proximal shaft to facilitate the advancement of the catheter in the anatomy. The outer surface of the catheter is hydrophilic coated in order to reduce friction during manipulation in the vessel. A radiopaque marker at the distal end of the catheter provides fluoroscopic visualization of the catheter tip. The proximal end of the catheter incorporates a standard Luer adapter to facilitate the attachment of accessories, a hub, and an ID band.
The CEREGLIDE™ 92 Catheter System is packaged with an INNERGLIDE™ 9 Delivery Aid, an RHV with a side port, and two Slit Introducer accessories. The INNERGLIDE™ 9 Delivery Aid is a compatible support device comprising of a single lumen catheter with a tapered tip, hydrophilic coating for increased lubricity, radiopaque tip for fluoroscopic visualization, and a proximal luer hub that is designed to be used as part of the CEREGLIDE™ 92 Catheter System to facilitate delivery of the CEREGLIDE™ 92 Intermediate Catheter to select blood vessels in the neurovasculature. The RHV with side port is used for flushing and
{4}------------------------------------------------
# 510(k) Summary (continued)
| V. Device<br>Description<br>(continued) | insertion of catheters. The Slit Introducers are designed to introduce the<br>CEREGLIDE™ 92 Intermediate Catheter into the base catheter and protect the distal<br>tip of the CEREGLIDE™ 92 Intermediate Catheter during insertion into the<br>hemostasis valve of the base catheter. |
|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| VI. Indications<br>for Use | The CEREGLIDE 92 Catheter System is indicated for use in facilitating the insertion<br>and guidance of appropriately sized interventional devices into a selected blood vessel<br>in the neurovascular system. |
| VII. Predicate<br>Comparison | A comparison of indications for use and technological characteristics between the<br>CEREGLIDE™ 92 Catheter System and the predicate device is presented in Table 3. |
| | | Table 3. Predicate and Subject Device Comparison Summary | |
|-------------------------------|------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Description | Predicate Device:<br>CEREGLIDE™ 92 Intermediate<br>Catheter<br>K230726 | Subject Device:<br>CEREGLIDE™ 92 Catheter System<br>K233982 |
| | Product Code | QJP | Same |
| | Regulatory Name | Catheter, Percutaneous | Same |
| | Classification | Class II - 21 CFR 870.1250 | Same |
| | Basic Design | Variable stiffness single lumen catheter | Same |
| | Indications For Use | The CEREGLIDE 92 Intermediate<br>Catheter is indicated for use in<br>facilitating the insertion and guidance<br>of appropriately sized interventional<br>devices into a selected blood vessel in<br>the neurovascular system. | The CEREGLIDE 92 Catheter System is<br>indicated for use in facilitating the insertion and<br>guidance of appropriately sized interventional<br>devices into a selected blood vessel in the<br>neurovascular system. |
| | CEREGLIDE 92 Intermediate Catheter Dimensions: | | |
| | Length | 114 - 135 cm | Same |
| | ID | 0.092" (2.34 mm) | Same |
| | Distal OD | 0.102" (2.59 mm) | Same |
| | Proximal OD | 0.108" (2.74 mm) | Same |
| | Catheter Coating | Hydrophilic | Same |
| | Coating Length | 50 cm | Same |
| | Materials: | | |
| | Marker Band | Stainless Steel and Tungsten | Same |
| | Braid | Stainless Steel and Tungsten | Same |
| | Liner | PTFE Liner | Same |
| | Hub | Nylon | Same |
| | Strain Relief | | Same |
| | Outer Jacket | Pebax, Urethane, Nylon | Same |
| | Accessories Included: | | |
| | Hemostasis valve | Tuohy Borst Hemostasis Valve with<br>Side Port Extension Tubing | RHV with Side Port |
| | Introducer Sheath | Slit Introducer (2) | Same |
| INNERGLIDE™ 9<br>Delivery Aid | Dimensions: | | |
| | Length | None | 165 cm |
| | ID | | 0.030" (0.76 mm) |
| | Distal OD | | 0.042" (1.07 mm) |
| | Proximal OD | | 0.078" (1.98 mm) |
| | Catheter Coating | | Hydrophilic |
| | Coating Length | | 50 cm |
Continued the next page
{5}------------------------------------------------
# 510(k) Summary (continued)
| | Materials: | | |
|-------------------------------|---------------------------------|----------------------------------------------------------|-----------------|
| INNERGLIDE™<br>9 Delivery Aid | Marker Band | | Tungsten |
| | Liner | | PTFE Liner |
| | Hub | None | Makrolon |
| | Strain Relief | | Pebax |
| | Outer Jacket | | Pebax, Vestamid |
| | Sterilization Method | Ethylene Oxide | Same |
| | Sterility Assurance Level (SAL) | 10-6 | Same |
| | Packaging | Polyethylene Hoop and Mounting Card, Tyvek Pouch, Carton | Same |
| | Shelf Life | 6 months | Same |
### VIII. Non-Clinical Performance Testing
#### Performance Testing - Bench
There are no changes to the CEREGLIDE 92 Intermediate Catheter (K230726). Appropriate testing was identified based on design, risk analyses, and the intended use of the INNERGLIDE 9 Delivery Aid to demonstrate that the CEREGLIDE 92 Catheter System is substantially equivalent to the legally marketed predicate device. The following performance data are being provided in support of the substantial equivalence determination. All testing was conducted using sampling methods as required by internal design control procedures.
| Table 4. Performance Testing Summary | | |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------|
| Test | Test Summary | Result |
| Design Verification | | |
| Visual<br>Inspection | Confirm that the INNERGLIDE 9 Delivery Aid meets the<br>visual requirements described in ISO 10555-1 Section 4.4. | PASS:<br>Samples met the established<br>acceptance criteria |
| Catheter ID | Verify that the INNERGLIDE 9 Delivery Aid internal<br>diameters meet the design requirements. | PASS:<br>Samples met the established<br>acceptance criteria |
| Catheter OD | Verify that the INNERGLIDE 9 Delivery Aid outer diameters<br>meet the design requirements. | PASS:<br>Samples met the established<br>acceptance criteria |
| Catheter<br>Working<br>Length | Confirm the working length of the INNERGLIDE 9 Delivery<br>Aid as defined in ISO 10555-1 Section 3.6. | PASS:<br>Samples met the established<br>acceptance criteria |
| Catheter Tip<br>Length | Verify that the tip length and navigation length of the<br>INNERGLIDE 9 Delivery Aid meet the design requirements. | PASS:<br>Samples met the established<br>acceptance criteria |
| Hub Luer | Verify that the hub luer on the INNERGLIDE 9 Delivery Aid<br>meets dimensional, interconnectability, and performance<br>requirements defined in ISO 80369-7. | PASS:<br>Samples met the established<br>acceptance criteria |
| System Air<br>Leak Resistance | Verify that the INNERGLIDE 9 Delivery Aid has no air leak. | PASS:<br>Samples met the established<br>acceptance criteria |
| System Liquid<br>Leak Resistance | Verify that the INNERGLIDE 9 Delivery Aid joint strength<br>meets the freedom from leakage (liquid during pressurization)<br>requirements of ISO 10555-1:2013, section 4.7. | PASS:<br>Samples met the established<br>acceptance criteria |
| Hub Pull<br>Testing | Verify that the INNERGLIDE 9 Delivery Aid and<br>CEREGLIDE 92 Intermediate Catheter hub to joint strength<br>meets acceptance criteria. | PASS:<br>Samples met the established<br>acceptance criteria |
| Shaft Tensile<br>Strength | Verify that the INNERGLIDE 9 Delivery Aid and<br>CEREGLIDE 92 Intermediate Catheter joint strength meets the<br>acceptance criteria. | PASS:<br>Samples met the established<br>acceptance criteria |
| Particulate<br>Count | Verify the particulate size and counts of the CEREGLIDE 92<br>Catheter System under simulated use conditions with<br>comparison to the predicate devices. | PASS:<br>Samples met the established<br>acceptance criteria |
{6}------------------------------------------------
#### VIII. Non-Clinical Performance Testing (continued)
| Table 4. Performance Testing Summary (continued) | | |
|-------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------|
| Test | Test Summary | Result |
| | Design Verification (continued) | |
| Coating<br>Durability and<br>Friction | Verify that the friction force and durability of the hydrophilic<br>coating on the INNERGLIDE 9 Delivery Aid meet the design<br>requirements. | PASS:<br>Samples met the established<br>acceptance criteria |
| Coating Length | Verify that the INNERGLIDE 9 Delivery Aid hydrophilic<br>coating length meets the design requirements. | PASS:<br>Samples met the established<br>acceptance criteria |
| Coating<br>Thickness | Verify that the INNERGLIDE 9 Delivery Aid hydrophilic<br>coating thickness meets the design requirements. | PASS:<br>Samples met the established<br>acceptance criteria |
| Coating<br>Integrity | Verify that the coating on the INNERGLIDE 9 Delivery Aid is<br>free of uncoated individual voids when evaluated with Congo<br>Red Dye Coverage Test. | PASS:<br>Samples met the established<br>acceptance criteria |
| Kink Resistance | Confirm that the INNERGLIDE 9 Delivery Aid meets the<br>requirement to remain stable and not kink during use. | PASS:<br>Samples met the established<br>acceptance criteria |
| Tip Flexibility | Test the tip flexibility of the INNERGLIDE 9 Delivery Aid,<br>relative to cleared devices of similar design. | PASS:<br>Samples met the established<br>acceptance criteria |
| Torque<br>Performance | Confirm that the INNERGLIDE 9 Delivery Aid meets the<br>torque strength requirement. | PASS:<br>Samples met the established<br>acceptance criteria |
| Trackability<br>(Delivery and<br>Withdrawal<br>Force) | Confirm that the INNERGLIDE 9 Delivery Aid trackability<br>meets the design requirements. | PASS:<br>Samples met the established<br>acceptance criteria |
| Burst Pressure | Verify the CEREGLIDE 92 Intermediate Catheter continues to<br>meet minimum static burst pressure requirements. | PASS:<br>Samples met the established<br>acceptance criteria |
| Design Validation | | |
| In Vitro<br>Usability<br>Studies | In vitro studies were conducted in a neurovascular model to<br>evaluate compatibility, trackability, and durability of the<br>INNERGLIDE 9 Delivery Aid and use for delivery of the<br>CEREGLIDE 92 Intermediate Catheter. | PASS:<br>Samples met the established<br>acceptance criteria |
### Performance Testing - Animal
No animal studies were required as appropriate verification and validation of the subject device design were achieved based on the similarities of the proposed device and the predicate device, and from results of bench testing.
#### Performance Testing - Clinical
Clinical studies were not required as appropriate verification and validation of the subject device design were achieved based on the similarities of the proposed device and the predicate device, and from results of bench testing.
#### Sterilization
The CEREGLIDE™ 92 Catheter System, as packaged with included accessories, is sterilized using a validated 100% Ethylene Oxide sterilization process to ensure sterility assurance level (SAL) of 10° in accordance with ISO 11135. The CEREGLIDE™ 92 Catheter System and all accessories meet EO residuals per EN ISO 10993-7 for a limited contact delivery system - externally communicating. The CEREGLIDE™ 92 Catheter System and all accessories are for single use only.
#### Shelf-Life
The CEREGLIDE™ 92 Catheter System will have a shelf life of 6 months based on the successful completion of stability testing. Shelf-life testing was performed using standard test methods and acceptance criteria. Prior to aging, all samples were exposed to standard transportation conditioning. Results of testing on the subject device all met established acceptance criteria.
{7}------------------------------------------------
# 510(k) Summary (continued)
| VIII. Non- |
|-------------|
| Clinical |
| Performance |
| Testing |
| (continued) |
#### Biocompatibility Testing
The INNERGLIDE 9 Delivery Aid was assessed for biocompatibility in accordance with International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process," and FDA Guidance for Industry and FDA Staff, "Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process"," (issued September 4, 2020). The subject device is considered an externally communicating medical device with circulating blood contact for less than 24 hours. The following biocompatibility testing was completed as part of this evaluation:
| Table 5. Biocompatibility Test Summary for the INNERGLIDE 9 Delivery Aid | | | |
|--------------------------------------------------------------------------|---------------------------------------------------------------|--------|--|
| Test Method | Acceptance Criteria | Result | |
| Cytotoxicity - MEM Elution | Grade ≤ 2 | PASS | |
| Sensitization - Guinea Pig<br>Maximization | Grade < 1 | PASS | |
| Irritation - Rabbit<br>Intracutaneous Reactivity | Difference of ≤ 1 from control mean score | PASS | |
| Acute Systemic Toxicity<br>in Mice | No biologically significant signs of systemic<br>toxicity | PASS | |
| Pyrogenicity - Materials<br>Mediated Rabbit Pyrogen | Temperature change (△T) < 0.5°C | PASS | |
| ASTM Hemolysis Study -<br>Direct and Extraction Methods | Hemolytic Index < 2% | PASS | |
| SC5b-9 Complement<br>Activation Assay | Results evaluated based on comparison to<br>negative controls | PASS | |
| In Vivo Thromboresistance | Results evaluated based on comparison to<br>negative controls | PASS | |
#### Based upon the intended use, design, materials, function, performance testing and IX. Conclusion biocompatibility testing, it is concluded that the subject device, CEREGLIDE™ 92 Catheter System (K233982) is substantially equivalent to the predicate CEREGLIDE™ 92 Intermediate Catheter (K230726). The addition of the INNERGLIDE™ 9 Delivery Aid did not raise any new questions regarding the safety and effectiveness of the device. The CEREGLIDE™ 92 Catheter System, as designed, manufactured, packaged, and sterilized, is substantially equivalent to the predicate device currently marketed under the Federal Food, Drug and Cosmetic Act.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.