K233979 · Greiner Bio-One North America, Inc. · JKA · Jul 19, 2024 · Clinical Chemistry
Device Facts
Record ID
K233979
Device Name
VACUETTE® QUICKSHIELD Complete (Plus)
Applicant
Greiner Bio-One North America, Inc.
Product Code
JKA · Clinical Chemistry
Decision Date
Jul 19, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1675
Device Class
Class 2
Indications for Use
The VACUETTE® QUICKSHIELD Complete is intended to be used only with VACUETTE® Blood Collection Tubes as a system in routine venipuncture procedures. This device is designed with a safety shield, which can be activated to cover the needle immediately following venipuncture to aid in the protection against accidental needlestick injury. They are for single-use only and should be used by properly trained healthcare professionals only in accordance with these instructions. The VACUETTE® QUICKSHIELD Complete PLUS is intended to be used only with VACUETTE® Blood Collection Tubes as a system in routine venipuncture procedures. This device is designed with a safety shield, which can be activated to cover the needle immediately following venipuncture to aid in the protection against accidental needlestick injury. VACUETTE® VISIO PLUS Needles are designed for use in the daily blood collection routine when delegated by a qualified practitioner. The flashback window is situated in the transparent part of the cannula hub, which assists the user to recognize successful vein penetration. They are for single- use only and should only be used by adequately trained healthcare personnel in accordance with these instructions.
Device Story
VACUETTE® QUICKSHIELD Complete (Plus) is a sterile, single-use blood collection system consisting of a pre-assembled safety tube holder and a multiple-use drawing needle. The device is used by healthcare professionals during routine venipuncture. The 'Plus' variant includes a flashback window in the cannula hub to confirm successful vein penetration. Following blood collection, the user manually activates a safety shield to cover the needle, locking it in place to prevent accidental needlestick injuries. The device is designed to be used exclusively with VACUETTE® Blood Collection Tubes. By providing a secure, integrated safety mechanism, the device reduces the risk of sharps-related injuries to the clinician, thereby improving workplace safety during blood sampling procedures.
Clinical Evidence
No clinical data. Bench testing only. Performance testing included simulated use for venipuncture, safety mechanism functionality, and mechanical integrity (force to activate, torque, leakage, puncture force). Biocompatibility testing performed per ISO 10993 series (cytotoxicity, sensitization, irritation, hemocompatibility, pyrogenicity, acute systemic toxicity). Sterility validated per ISO 11137-1.
Technological Characteristics
Materials: Polypropylene (PP) for needle cap, holder, and safety shield; AISI 304 stainless steel for needle; rubber sleeve. Dimensions: 21G or 22G needles; 1", 1.25", or 1.5" lengths. Sterilization: E-Beam irradiation (SAL 10^-6). Connectivity: None. Mechanism: Manual, one-hand activated safety shield. Standards: ISO 23908, ISO 7864, ISO 9626, ISO 10993, ISO 11137-1, ISO 11737-1.
Indications for Use
Indicated for routine venous blood collection in patients requiring venipuncture. For use by trained healthcare professionals only. Contraindications: None stated.
Regulatory Classification
Identification
A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
July 19, 2024
Greiner Bio-One North America Inc Manfred Abel Regulatory Affairs & Quality Systems Manager 4238 Capital Drive Monroe, North Carolina 28110
Re: K233979
Trade/Device Name: VACUETTE® QUICKSHIELD Complete (Plus) Regulation Number: 21 CFR 862.1675 Regulation Name: Blood specimen collection device Regulatory Class: Class II Product Code: JKA, FMI Dated: June 20, 2024 Received: June 20, 2024
Dear Manfred Abel:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
David Walloschek
David Wolloscheck, Ph.D. Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors
{2}------------------------------------------------
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K233979
Device Name VACUETTE® QUICKSHIELD Complete (Plus)
#### Indications for Use (Describe)
The VACUETTE® QUICKSHIELD Complete is intended to be used only with VACUETTE® Blood Collection Tubes as a system in routine venipuncture procedures. This device is designed with a safety shield, which can be activated to cover the needle immediately following venipuncture to aid in the protection against accidental needlestick injury. They are for single-use only and should be used by properly trained healthcare professionals only in accordance with these instructions.
The VACUETTE® QUICKSHIELD Complete PLUS is intended to be used only with VACUETTE® Blood Collection Tubes as a system in routine venipuncture procedures. This device is designed with a safety shield, which can be activated to cover the needle immediately following venipuncture to aid in the protection against accidental needlestick injury. VACUETTE® VISIO PLUS Needles are designed for use in the daily blood collection routine when delegated by a qualified practitioner. The flashback window is situated in the transparent part of the cannula hub, which assists the user to recognize successful vein penetration. They are for single- use only and should only be used by adequately trained healthcare personnel in accordance with these instructions.
| Type of Use (Select one or both, as applicable) | | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| <table style="border:none"><tr><td><span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)</td></tr><tr><td><span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)</td></tr></table> | <span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
| <span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | | |
| <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) | | |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Greiner Bio-One. The logo features a stylized blue wave-like shape above the word "greiner" in a bold, sans-serif font. Below "greiner" is the text "BIO-ONE" in a smaller, sans-serif font.
#### K233979 - Premarket Notification 510(K) Summary
#### 1. SUBMITTER
| Applicant name: | Greiner Bio-One GmbH.<br>Bad Haller Strasse 32<br>4550 Kremsmünster,<br>Austria |
|-----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact person: | Manfred Abel<br>Greiner Bio-One NA Inc.<br>4238 Capital Drive<br>Monroe, NC 28110<br>Phone +1 704 261 7800<br>manfred.abel@gbo.com |
| Establishment registration number: | 8020040 |
| Date prepared: | July 19, 2024 |
| <b>2. DEVICE</b> | |
| Trade name: | VACUETTE® QUICKSHIELD Complete (Plus) |
| Common name: | Safety tube holder pre-assembled with the needle |
| Regulation: | 862.1675, Blood Specimen Collection Device |
| Primary classification product code: | JKA – Blood specimen collection device |
| Subsequent classification product code: | FMI – Needle, hypodermic, single lumen |
| Medical device class: | II |
| Review panel: | General Hospital |
| <b>3. PREDICATE DEVICES</b> | |
| Primary predicate device: | Blood Collection Needle with/without Holder,<br>Safety Blood Collection Needle with/without Holder,<br>Luer Access Device-holder with Pre-attached Multiple Sample<br>Adapter (K200027) |
## 4. DEVICE DESCRIPTION
Secondary predicate device:
The sterile VACUETTE®QUICKSHIELD Complete is a single packed VACUETTE®QUICKSHIELD Safety Tube Holder with pre-threaded VACUETTE® Multiple Use Drawing Needle in a blister pack. It is a sterile, single-use plastic tube holder, designed with a safety shield, which can be activated to cover the needle immediately following venipuncture.
VACUETTE® QUICKSHIELD Complete (K072320)
The sterile VACUETTE® QUICKSHIELD Complete PLUS is a single packed VACUETTE® QUICKSHIELD Safety Tube Holder with pre-threaded VACUETTE® VISIO PLUS Blood Collection Needle in a blister pack. It is a sterile, single-use plastic tube holder, designed with a safety shield, which can be activated to cover the
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Greiner Bio-One. The logo features a stylized blue swoosh above the word "greiner" in a bold, sans-serif font. Below "greiner" is the text "BIO-ONE" in a smaller, sans-serif font.
needle immediately following venipuncture.
#### 5. INDICATION FOR USE
The VACUETTE® QUICKSHIELD Complete is intended to be used only with VACUETTE® Blood Collection Tubes as a system in routine venipuncture procedures. This device is designed with a safety shield, which can be activated to cover the needle immediately following venipuncture to aid in the protection against accidental needlestick injury. They are for single-use only and should be used by properly trained healthcare professionals only in accordance with these instructions.
The VACUETTE® QUICKSHIELD Complete PLUS is intended to be used only with VACUETTE® Blood Collection Tubes as a system in routine venipuncture procedures. This device is designed with a safety shield, which can be activated to cover the needle immediately following venipuncture to aid in the protection against accidental needlestick injury. VACUETTE® VISIO PLUS Needles are designed for use in the daily blood collection routine when delegated by a qualified practitioner. The flashback window is situated in the transparent part of the cannula hub, which assists the user to recognize successful vein penetration. They are for singleuse only and should only be used by adequately trained healthcare personnel in accordance with these instructions.
#### 6. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
Greiner Bio-One NA Inc. believes that the VACUETTE® QUICKSHIELD Complete (Plus) is substantially equivalent in intended use, indications for use, principles of operation, and technological characteristics to the predicate devices. Any differences between VACUETTE® QUICKSHIELD Complete (Plus) and the predicate devices do not raise new types of safety or effectiveness issues, as further discussed below.
| | Candidate Device | Primary<br>Predicate<br>Device | Secondary<br>Predicate<br>Device | Remark |
|----------------|-----------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------|--------|
| Name | VACUETTE®<br>QUICKSHIELD<br>Complete (Plus) | Blood Collection<br>Needle with/without<br>Holder, Safety<br>Blood Collection<br>Needle with/without<br>Holder, Luer<br>Access Device- | VACUETTE®<br>QUICKSHIELD<br>Complete | n/a |
| 510(k) Number | K233979 | K200027 | K072320 | n/a |
| Classification | Class II | Class II | Class II | n/a |
| Regulation | 862.1675 | 862.1675 | 880.5570 | n/a |
| Product Code | Primary product code:<br>JKA<br>Subsequent product code:<br>FMI | Primary product code:<br>JKA<br>Subsequent product<br>code: FMI | FMI | n/a |
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image contains the logo for Greiner Bio-One. The logo features a stylized blue arc above the word "greiner" in a bold, sans-serif font. Below "greiner" is the text "BIO-ONE" in a smaller, sans-serif font.
| Indications for Use | The VACUETTE® QUICKSHIELD Complete | The Blood Collection Needle with/without holder | The QUICKSHIELD Complete | Different. S&E evaluated by external and internal performance testing, see further information below. |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|
| | The VACUETTE® QUICKSHIELD Complete is intended to be used only with VACUETTE® Blood Collection Tubes as a system in routine venipuncture procedures. This device is designed with a safety shield, which can be activated to cover the needle immediately following venipuncture to aid in the protection against accidental needlestick injury. They are for single-use only and should be used by properly trained healthcare professionals only in accordance with these instructions. The VACUETTE® QUICKSHIELD Complete PLUS is intended to be used only with VACUETTE® Blood Collection Tubes as a system in routine venipuncture procedures. This device is designed with a safety shield, which can be activated to cover the needle immediately following venipuncture to aid in the protection against accidental needlestick injury VACUETTE® VISIO PLUS Needles are designed for use in the daily blood collection routine when delegated by a qualified practitioner. The flashback window is situated in the transparent part of the cannula hub, which assists the user to recognize successful vein penetration. They are for single-use only and should only be used by adequately trained healthcare personnel in accordance with these instructions. | The Blood Collection Needle with/without holder is intended to be used with vacuum blood collection tube for the collection of venous blood. The Safety Blood Collection Needle with/without holder is intended to be used with vacuum blood collection tube for the collection of venous blood. The safety shield is intended to aid in the protection against accidental needle stick injury. The Luer access device-holder with preattached multiple sample adapter is a sterile, non-invasive device used to connect devices with male or female luer connectors to blood collection tubes for the collection of blood. | The QUICKSHIELD Complete is intended to be used only with VACUETTE® Blood Collection Tubes as a system in routine venipuncture procedures. This device is to be used by properly trained healthcare professionals only in accordance with these instructions. This device is designed with a safety shield, which can be activated to cover the needle immediately following venipuncture to aid in the protection against accidental needlestick injury. VACUETTE® VISIO PLUS Blood Collection Needles are designed for use in the daily blood collection routine when delegated by a qualified practitioner. The flashback window is situated in the transparent part of the cannula hub, which assists the user to recognize successful vein penetration. They are for single-use only and should only be used by adequately trained healthcare personnel in accordance with these instructions. | Different. S&E evaluated by external and internal performance testing, see further information below. |
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for Greiner BIO-ONE. The logo features a blue, curved shape above the word "greiner" in a bold, black sans-serif font. Below "greiner" is the text "BIO-ONE" in a smaller, black sans-serif font.
| Manufacturer | Greiner Bio-One (GBO) | Jiangsu Caina<br>Medical Co., Ltd | Greiner Bio-One (GBO) | n/a |
|----------------------------------------------|--------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|----------------------------------------------------------------------------|-----------------------------------------------------------------------------|
| Safety<br>mechanism | Active safety<br>mechanism. One-hand<br>activation. | The safety shield is<br>intended to prevent<br>needle sticks. Only on<br>the Safety | Active safety<br>mechanism. One-<br>hand activation. | Same |
| | Manual operation for<br>activation. | Manual | Manual operation<br>for activation. | |
| Cannula<br>variants | • Multiple<br>Sampling Needle<br>(K973620)<br>• Visio Plus Needle<br>(K061483) | Various gages and<br>length | • Multiple Samling<br>Needle (K973620)<br>• Visio Plus Needle<br>(K061483) | SE |
| Cannula<br>Gauges | 21G or 22G | 27G, 25G, 23G, 22G,<br>21G, 20G, 18G | 21G or 22G | SE |
| Cannula<br>Length | 1" or 1.25" or 1.5" | 1" or 1.25" or 1.5" | 1" or 1.25" or 1.5" | SE |
| Sterile | Yes | Yes | Yes | SE |
| SAL | 10-6 | 10-6 | 10-6 | SE |
| Sterilization | E-Beam Irradiation | ETO | ETO | Different,<br>evaluation<br>according to ISO<br>11137-1 and ISO<br>11137-2. |
| Non-pyrogenic | Yes | Yes | Yes | SE |
| Materials -<br>comply<br>with ISO<br>10993-1 | Yes | Yes | Yes | SE |
| Use | Single use only | Single use only | Single use only | SE |
| | Materials: | | | |
| Needle Cap | PP | Same | Same | SE |
| Needle | Stainless Steel AISI 304 | Same | Same | SE |
| Rubber sleeve | Rubber Sleeve | Rubber Sleeve | Rubber Sleeve | SE |
| Holder | PP | Same | Same | SE |
| Safety shield | PP | Same | Same | SE |
| | Connection of: | | | |
| Needle and<br>Holder | Pre-attached | Pre-attached | Pre-attached | SE |
| Holder and<br>shield | Rotational shield | n/a | Non-rotational shield | Different, S&E<br>evaluated by<br>internal test<br>method |
## Comparison of Technological Characteristics
As can be seen in the above table a comparison of the VACUETTE® QUICKSHIELD Complete (Plus) and its features show that both the candidate device and the predicate devices have the same fundamental scientific technological characteristics. The VACUETTE® QUICKSHIELD Complete (Plus) and the predicate devices include the same safety mechanism covering the needle after the blood collection. The used safety mechanisms need to be activated manual activation of the safety mechanism of candidate device as well as the predicate devices the safe and locked position of the safety
{8}------------------------------------------------
Premarket Notification VACUETTE®QUICKSHIELD Complete (Plus) Greiner Bio-One
Image /page/8/Picture/1 description: The image shows the logo for Greiner Bio-One. The logo features a blue, curved shape above the word "greiner" in bold, black letters. Below "greiner" is the text "BIO-ONE" in smaller, black letters. The blue curved shape is reminiscent of a stylized wave or swoosh.
mechanism is reached by the Quickshield-needle-cover. The used materials for all products are the same or at least comparable and the differences do not have an influence on the safety and effectiveness of the candidate device. The candidate device and the predicate devices are provided sterile and single use only.
#### Comparison of Intended Use
The above comparison table shows the indications for use of the candidate device as well as predicate devices. The indications are comparable in their content so that all products are intended to be used for venous blood collection.
### 7. PERFORMANCE DATA
The VACUETTE® QUICKSHIELD Complete (Plus) is manufactured and verified by testing in conformity with applicable recognized consensus standards like ISO 23908, ISO 7864, and ISO 9626 but not limited to those standards. Simulated use tests were performed for venous blood collection procedure as well as the functionality of the safety mechanism was tested.
The VACUETTE® QUICKSHIELD Complete (Plus) is manufactured and tested in conformity with internal specifications and recognized consensus standards. Performance testing was conducted to demonstrate and support our substantial equivalency claim that the device performs as intended and in an equivalent manner as the predicate devices.
| Requirement | Outcome<br>(Pass or fail) |
|----------------------------------------------------------|---------------------------|
| Force to remove shield | Pass |
| Torque to rotate shield | Pass |
| Torque to screw out needle | Pass |
| Spin out | Pass |
| Force to activate safety mechanism | Pass |
| Force necessary to bend safety shield with needle by 90° | Pass |
| Puncture force | Pass |
| Friction force | Pass |
| Leakage | Pass |
#### Sterility testing:
VACUETTE® QUICKSHIELD Complete (Plus) are sterilized by e-beam irradiation for a final Sterility Assurance Level (SAL) of 10° per ISO 11137-1 Sterilization of health care products – Radiation – Part 1: Requirements for development, validation, and routine control of a sterilization process for medical devices. Bioburden testing for determination of verification dose and subsequently the processing dose was performed in accordance with ISO 11737-1 - Sterilization of medical devices - microbiological methods - Part 1: Determination of a population of microorganisms on products.
#### Biocompatibility testing:
Biocompatibility for the VACUETTE® QUICKSHIELD Complete (Plus) was performed in accordance with the standards of the ISO 10993 series and under consideration of the Guidance for industry and Food and Drug Administration Stuff - Use of the International Standard ISO 10993-1. "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process" (Attachment A: Table A. 1 Endooints to be addressed in a biological risk assessment). The holder was classified as "surface device with short term contact to intact skin", and the cannula was classified as "external communicating device with short term contact to the blood path." Based on the classification the following biological endpoints were evaluated, and it was demonstrated that the candidate device meets the following standards:
{9}------------------------------------------------
Premarket Notification VACUETTE®QUICKSHIELD Complete (Plus) Greiner Bio-One
Image /page/9/Picture/1 description: The image shows the logo for Greiner Bio-One. The logo features the word "greiner" in a bold, sans-serif font, with the letters in black. Above the word "greiner" is a blue, curved shape that resembles a stylized "G". Below the word "greiner" is the text "BIO-ONE" in a smaller, sans-serif font, also in black.
- Cytotoxicity (ISO 10993-5) ●
- . Sensitization (ISO 10993-10)
- . Irritation (ISO 10993-23)
- Hemocompatibility (ISO 10993-4) .
- Material mediated pyrogenicity (ISO 10993-11) ●
- Acute Systemic Toxicity (ISO 10993-11) .
#### 8. CONCLUSION
The test results and the comparison table above show the VACUETTE® QUICKSHIELD Complete (Plus) is substantially equivalent to its predicate devices with respect to intended use, technological and performance characteristics and does not raise new safety or effectiveness concerns.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.