K233971 · Arthrex, Inc. · MAI · Jun 4, 2024 · Orthopedic
Device Facts
Record ID
K233971
Device Name
Arthrex AlloSync PushLock Suture Anchor
Applicant
Arthrex, Inc.
Product Code
MAI · Orthopedic
Decision Date
Jun 4, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Arthrex AlloSync™ PushLock® suture anchors are intended to be used for suture (soft tissue) fixation to bone in the shoulder in the following procedures: Shoulder: Rotator Cuff Repair, Bankart Repair, Superior Labrum from Anterior (SLAP) Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction
Device Story
Arthrex AlloSync PushLock Suture Anchor is a two-component orthopedic fixation device; consists of human cortical bone anchor body and PEEK eyelet; includes disposable driver inserter. Used by orthopedic surgeons in clinical settings for soft tissue-to-bone fixation in shoulder procedures. Device functions as mechanical anchor to secure sutures to bone; provides stable fixation during healing. Delivered sterile for single-use. Benefits include biocompatible allograft material for bone fixation, reducing need for synthetic materials in specific clinical scenarios.
Clinical Evidence
Bench testing only. Mechanical testing (pull-out, insertion, degradation) demonstrated statistically equivalent strength to predicate. Biocompatibility testing conducted per ISO 10993-1:2018 (cytotoxicity, sensitization, irritation, genotoxicity, acute systemic toxicity, implantation, material-mediated pyrogenicity, chemical characterization). Bacterial Endotoxin Testing (BET) performed per ANSI/AAMI ST72:2011/(R)2016, USP <161>, USP <85>, and EP 2.6.14 confirmed compliance with pyrogen limits. Allograft donor screening and testing performed in accordance with 21 CFR 1270/1271 and AATB standards.
Technological Characteristics
Anchor body: human cortical bone; Eyelet: PEEK. Dimensions: 2.9mm diameter, 12.5mm total length. Sterile, single-use. Mechanical fixation principle. Complies with 21 CFR 1271 for HCT/Ps. Biocompatible per ISO 10993-1:2018.
Indications for Use
Indicated for soft tissue-to-bone fixation in the shoulder for patients undergoing rotator cuff repair, Bankart repair, SLAP lesion repair, biceps tenodesis, AC separation repair, deltoid repair, or capsular reconstruction.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
June 4, 2024
Image /page/0/Picture/1 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right is the FDA logo, with the letters "FDA" in a blue square and the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Arthrex, Inc. Emmarie Halteman Senior Regulatory Affairs Specialist 1370 Creekside Boulevard Naples, Florida 34108-1945
Re: K233971
Trade/Device Name: Arthrex AlloSync PushLock Suture Anchor Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: MAI, HWC Dated: May 8, 2024 Received: May 9, 2024
Dear Emmarie Halteman:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by
Jesse Muir -S Jesse Muir -S
Date: 2024.06.04
15:32:02 -04'00'
Jesse Muir, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K233971
Device Name Arthrex AlloSync™ PushLock® Suture Anchor
Indications for Use (Describe)
The Arthrex AlloSync™ PushLock® suture anchors are intended to be used for suture (soft tissue) fixation to bone in the shoulder in the following procedures:
Shoulder: Rotator Cuff Repair, Bankart Repair, Superior Labrum from Anterior (SLAP) Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction
Type of Use (Select one or both, as applicable)
| <span></span> | <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|---------------|--------------------------------------------------------------------------------------|
| <span></span> | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Arthrex. The logo consists of the word "Arthrex" in a sans-serif font, with a stylized image of surgical scissors replacing the "x". The scissors are depicted in a simple, line-art style, and the overall logo has a clean and modern look.
## 510(k) Summary
| Date Prepared | May 22, 2024 |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter | Arthrex Inc. |
| | 1370 Creekside Boulevard |
| | Naples, FL 34108-1945 |
| Contact Person | Name: Emmarie Halteman |
| | Title: Senior Regulatory Affairs Specialist |
| | Phone: 1-239-643-5553 ext. 74232 |
| | Email: emmarie.halteman@arthrex.com |
| Trade Name | Arthrex AlloSync™ PushLock® Suture Anchor |
| Common Name | Single/multiple component metallic bone fixation appliances and accessories |
| Product Code | MAI - Fastener, Fixation, Biodegradable, Soft Tissue<br>HWC - Screw, Fixation, Bone |
| Classification Name | 21 CFR 888.3030: Single/multiple component metallic bone fixation appliances and accessories; Fastener, Fixation, Biodegradable, Soft Tissue<br>21 CFR 888.3040: Smooth or threaded metallic bone fixation fastener |
| Regulatory Class | II |
| Predicate Device | K063479: Arthrex 2.5mm PushLock |
| Reference Device | K070347: Allofix Push-In Anchor |
| Purpose of Submission | This Traditional 510(k) premarket notification is submitted to obtain clearance for the Arthrex AlloSync™ PushLock® Suture Anchor |
| Device Description | The Arthrex AlloSync™ PushLock® Suture Anchor is a two-component suture anchor comprised of a hollow anchor body manufactured from human cortical bone, a Polyetheretherketone (PEEK) eyelet, and a disposable driver inserter. The anchor will be offered in a 2.9mm diameter and 12.5mm in total length (anchor + eyelet). The anchor is sold sterile, single-use. |
| Indications for Use | The Arthrex AlloSync™ PushLock® suture anchors are intended to be used for suture (soft tissue) fixation to bone in the shoulder in the following procedures:<br><br>Shoulder: Rotator Cuff Repair, Bankart Repair, Superior Labrum from Anterior to Posterior (SLAP) Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction |
| Performance Data | The submitted mechanical test data, pull-out, insertion, and degradation, demonstrates that the Arthrex AlloSync™ PushLock® Suture Anchor results in statistically equivalent strength to the predicate device Arthrex 2.5mm PushLock (K063479). |
| | Bacterial Endotoxin Testing (BET) was performed on the Arthrex<br>AlloSync™ PushLock® Suture Anchor utilizing the Kinetic<br>Turbidimetric limulus amebocyte lysate (LAL) method in<br>accordance with ANSI/AAMI ST72:2011/(R)2016, USP <161>, USP<br><85>, EP 2.6.14. Testing was performed in compliance with US FDA<br>good manufacturing practice (GMP) regulations 21 CFR Parts 210,<br>211 and 820. The testing conducted demonstrates that the Arthrex<br>AlloSync™ PushLock® Suture Anchor meets pyrogen limit<br>specifications. |
| | All donors have been screened and tissues recovered, processed,<br>stored, tested, and distributed in accordance with current U.S.<br>federal regulations as promulgated in 21 CFR 1270 and 1271,<br>current Standards for Tissue Banking set forth by the American<br>Association of Tissue Banks (AATB) and international laws and<br>regulations as required. This allograft was deemed suitable for<br>implantation by LifeNet Health. A physician medical director<br>evaluated the following donor variables to determine donor<br>suitability: infectious disease test results, current donor medical<br>history, behavioral risk assessment interview, physical assessment,<br>relevant medical records, including previous medical history,<br>laboratory test results, and autopsy or coroner reports (if<br>performed). All donors are tested for relevant infectious diseases.<br>Testing is performed by laboratories that are registered with the<br>U.S. Food and Drug Administration (FDA) and certified under the<br>Clinical Laboratory Improvement Amendments of 1988 (CLIA) and<br>42 CFR 493. Test methods that are FDA-licensed, approved, or<br>cleared for donor screening are used as available. The following<br>test criteria were met for the donor of this allograft bio-implant:<br>Hepatitis B virus (HBV) surface antigen, HBV core antibody,<br>Hepatitis C virus (HCV) antibody, HIV-1/12 antibody, Syphilis, HIV-1<br>NAT, HCT NAT, and HBV NAT. |
| | Cytotoxicity, Sensitization, Irritation, Genotoxicity, Acute Systemic<br>Toxicity, Implantation, Material Mediated Pyrogenicity, and<br>Material Chemical Characterization was conducted on the Arthrex<br>AlloSync™ PushLock® Suture Anchor in accordance with ISO 10993-<br>1:2018 and deemed biocompatible when used as intended. |
| Technological Comparison | Compared to the predicate device Arthrex 2.5mm PushLock<br>(K063479), the proposed Arthrex AlloSync™ PushLock® Suture<br>Anchor has the same intended use/indications for the shoulder,<br>fundamental scientific technology, sterility, and shelf-life. The<br>proposed Arthrex AlloSync™ PushLock® Suture Anchor's anchor<br>body is made of human cortical bone; whereas the primary<br>predicate device Arthrex 2.5mm PushLock is made of Poly (I-<br>lactide) (PLLA). However, the anchor body cleared under reference |
| | device Allofix Push-In Anchor (K070347) is made of human allograft<br>bone and they both comply with all requirements in 21 CFR 1271<br>to reduce the risk of the introduction, transmission, and spread of<br>communicable diseases by human cells, tissues, and cellular and<br>tissue-based products (HCT/Ps). |
| | Any differences between the proposed device and the predicate<br>device are considered minor and do not raise new or different<br>questions concerning safety or effectiveness. |
| Conclusion | The Arthrex AlloSync™ PushLock® Suture Anchor is substantially<br>equivalent to the predicate device in which the basic design<br>features and intended uses are the same. Any differences between<br>the proposed device and the predicate device are considered<br>minor and do not raise different questions concerning safety or<br>effectiveness. Based on the indications for use, technological<br>characteristics, and the summary of data submitted, Arthrex Inc.<br>has determined that the proposed device is substantially<br>equivalent to the currently marketed predicate device. |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the word "Arthrex" in a stylized font. The "x" in "Arthrex" is stylized to look like a medical instrument. The logo is simple and modern.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "Arthrex" in a bold, sans-serif font. To the right of the word is an image of a surgical tool, possibly scissors or forceps. The tool is depicted in a simple, stylized manner, with its handles open and blades visible. The logo is likely for a medical device company specializing in arthroscopic surgery.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.