Nephron Nitrile™, Nitrile Powder-Free Examination Blue Gloves (Tested For Use With Chemotherapy Drugs and Fentanyl)

K233970 · Nephron Nitrile, LLC · LZA · Jan 16, 2024 · General Hospital

Device Facts

Record IDK233970
Device NameNephron Nitrile™, Nitrile Powder-Free Examination Blue Gloves (Tested For Use With Chemotherapy Drugs and Fentanyl)
ApplicantNephron Nitrile, LLC
Product CodeLZA · General Hospital
Decision DateJan 16, 2024
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 880.6250
Device ClassClass 1

Indications for Use

Nephron Nitrile™, Nitrile Powder-Free Examination Blue Gloves (Tested For Use With Chemotherapy Drugs and Fentanyl) is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs and fentanyl in accordance with ASTM D6978-05 (2019) Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.

Device Story

Disposable nitrile examination glove; worn by healthcare examiners to prevent cross-contamination between patient and examiner. Device is blue, powder-free, ambidextrous, fingertip-textured, and features a beaded cuff. Manufactured from nitrile (NBR) latex. Tested for permeation resistance against various chemotherapy drugs and fentanyl per ASTM D6978-05 (2019). Used in clinical settings; single-use only. Provides a physical barrier to protect the wearer from patient contaminants and hazardous drug exposure.

Clinical Evidence

No clinical data; bench testing only. Performance verified via ASTM standards for physical properties (tensile strength, elongation), dimensions, freedom from holes, residual powder, and chemotherapy/fentanyl permeation resistance. Biocompatibility confirmed via ISO 10993 series (irritation, sensitization, cytotoxicity, systemic toxicity).

Technological Characteristics

Material: Nitrile (NBR) latex. Form factor: Ambidextrous, fingertip-textured, beaded cuff. Sizes: XS, S, M, L, XL, XXL. Standards: ASTM D6319-19 (physical properties/dimensions), ASTM D6978-05 (2019) (chemotherapy permeation), ASTM D5151-19 (holes), ASTM D6124-06 (powder). Biocompatibility: ISO 10993-5, -10, -11, -23. Non-sterile, single-use.

Indications for Use

Indicated for use as a disposable medical glove worn on the examiner's hand to prevent cross-contamination between patient and examiner. Tested for resistance to permeation by specific chemotherapy drugs and fentanyl.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. January 16, 2024 Nephron Nitrile, LLC. Lou Kennedy CEO/President/Manager 4777 12th Street Extension West Columbia, South Carolina 29172 Re: K233970 Trade/Device Name: Nephron Nitrile™, Nitrile Powder-Free Examination Blue Gloves (Tested For Use With Chemotherapy Drugs and Fentanyl) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I. reserved Product Code: LZA, LZC, OPJ, QDO Dated: December 15, 2023 Received: December 15, 2023 Dear Lou Kennedy: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products: and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {2}------------------------------------------------ Sincerely, # Bifeng Qian -S Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K233970 #### Device Name Nephron Nitrile™, Nitrile Powder-Free Examination Blue Gloves (Tested For Use With Chemotherapy Drugs and Fentany() Indications for Use (Describe) Nephron Nitrile™. Nitrile Powder-Free Examination Blue Gloves (Tested For Use With Chemotherapy Drugs and Fentanyl) is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs and fentanyl in accordance with ASTM D6978-05 (2019) Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs. The following chemotherapy drugs and concentration had NO breakthrough detected up to 240 minutes: Bleomycin Sulfate (15.0 mg/ml) Busulfan (6.0 mg/ml) Carboplatin (10.0 mg/ml) Cisplatin (1.0 mg/ml) Cyclophosphamide (20.0 mg/ml) Cytarabine (100.0 mg/ml) Dacarbazine (10.0 mg/ml) Daunorubicin HCl (5.0 mg/ml) Docetaxel (10.0 mg/ml) Doxorubicin HCl (2.0 mg/ml) Epirubicin HCl (2.0 mg/ml) Etoposide (20.0 mg/ml) Fludarabine (25.0 mg/ml) Fluorouracil (50.0 mg/ml) Gemcitabine (38.0 mg/ml) Idarubicin HCl (1.0 mg/ml) Ifosfamide (50.0 mg/ml) Irinotecan (20.0 mg/ml) Mechlorethamine HCl (1.0 mg/ml) Melphalan (5.0 mg/ml) Methotrexate (25.0 mg/ml) Mitomycin C (0.5 mg/ml) Mitoxantrone HCl (2.0 mg/ml) Paclitaxel (6.0 mg/ml) Rituximab (10.0 mg/ml) Trisenox (1.0 mg/ml) Vincristine Sulfate (1.0 mg/ml) The following chemotherapy drugs have low permeation times: Carmustine (3.3 mg/ml) : 33.8 minutes Thiotepa (10.0 mg/ml) : 128.1 minutes Warning: Not for Use with: Carmustine, Thiotepa The tested Opioid is: Fentanyl Citrate Injection (100 mcg/2 mL) {4}------------------------------------------------ Prescription Use (Part 21 CFR 801 Subpart D) - 区 Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {5}------------------------------------------------ ### A. APPLICANT INFORMATION | 510(K) Owner's Name | Nephron Nitrile, LLC. | |---------------------|-------------------------------------------------------------------| | Address | 4777 12th Street Extension, West Columbia, SC<br>29172. | | Phone | 844-937-3888 | | Fax | 1-803-926-9853 | | E-mail | lkennedy@nephronpharm.com<br>NitrileRegulatory@nephronnitrile.com | | Contact Person | Lou Kennedy | | Designation | Chief Executive Officer | | Contact Number | 1-803-569-3110 | | Contact Email | lkennedy@nephronpharm.com<br>NitrileRegulatory@nephronnitrile.com | | Preparation date | 28 September 2023 | | Date Submitted | 15 December 2023 | ### B. DEVICE IDENTIFICATION | Name of the device | Nephron Nitrile™, Nitrile Powder-Free<br>Examination Blue Gloves (Tested For Use With<br>Chemotherapy Drugs and Fentanyl) | |-----------------------------------|---------------------------------------------------------------------------------------------------------------------------| | 510(k) Number | K233970 | | Product proprietary or trade name | Nephron Nitrile™ | | Common or usual name | Nitrile Examination Glove (Tested For Use With<br>Chemotherapy Drugs and Fentanyl) | | Classification name | Non-Powdered Patient Examination Glove,<br>Specialty | | Device Classification | Class-1 | | Product Code | LZA, LZC, OPJ, QDO | | Regulation Number | 21 CFR 880.6250 | | Review Panel | General Hospital | ### C. PREDICATE DEVICE | Predicate Device | Nephron Nitrile Powder-Free Nitrile Examination<br>Gloves (Tested For Use With Chemotherapy<br>Drugs and Fentanyl) | |------------------|--------------------------------------------------------------------------------------------------------------------| | 510(k) Number | K231349 | | Regulatory Class | Class-1 | | Product code | LZA, LZC, OPJ, QDO | | Manufacturer | Nephron Nitrile, LLC. | {6}------------------------------------------------ ### D. DESCRIPTION OF THE DEVICE: Nephron Nitrile™. Nitrile Powder-Free Examination Blue Gloves (Tested For Use With Chemotherapy Drugs and Fentanyl) are Class I patient examination gloves, bearing the product codes LZA, LZC, OPJ, QDO (21 CFR 880.6250). They meet all the current specifications listed under the ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application and also complies with requirements for Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs as per ASTM D6978-05 (2019). The powder-free gloves are made from Nitrile (NBR) latex and are blue in color. The product is non-sterile, fingertip textured, ambidextrous with beaded cuff, and single use only. Nephron Nitrile™, Nitrile Powder-Free Examination Blue Gloves (Tested For Use With Chemotherapy Drugs and Fentanyl) include glove sizes: X-Small, Small, Medium, Large, X-Large and XX-Large. ### E. INDICATION FOR USE OF THE DEVICE: Nephron Nitrile™, Nitrile Powder-Free Examination Blue Gloves (Tested For Use With Chemotherapy Drugs and Fentanyl) is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs and fentanyl in accordance with ASTM D6978-05 (2019) Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs. | The following chemotherapy drugs and concentration had NO breakthrough detected up to | | | The tested | |----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------| | 240 minutes: | | | Opioid is: | | Bleomycin Sulfate (15.0 mg/ml)<br>Busulfan (6.0 mg/ml)<br>Carboplatin (10.0 mg/ml)<br>Cisplatin (1.0 mg/ml)<br>Cyclophosphamide (20.0 mg/ml)<br>Cytarabine (100.0 mg/ml)<br>Dacarbazine (10.0 mg/ml)<br>Daunorubicin HCl (5.0 mg/ml)<br>Docetaxel (10.0 mg/ml) | Doxorubicin HCl (2.0 mg/ml)<br>Epirubicin HCl (2.0 mg/ml)<br>Etoposide (20.0 mg/ml)<br>Fludarabine (25.0 mg/ml)<br>Fluorouracil (50.0 mg/ml)<br>Gemcitabine (38.0 mg/ml)<br>Idarubicin HCl (1.0 mg/ml)<br>Ifosfamide (50.0 mg/ml)<br>Irinotecan (20.0 mg/ml) | Mechlorethamine HCl (1.0 mg/ml)<br>Melphalan (5.0 mg/ml)<br>Methotrexate (25.0 mg/ml)<br>Mitomycin C (0.5 mg/ml)<br>Mitoxantrone HCl (2.0 mg/ml)<br>Paclitaxel (6.0 mg/ml)<br>Rituximab (10.0 mg/ml)<br>Trisenox (1.0 mg/ml)<br>Vincristine Sulfate (1.0 mg/ml) | Fentanyl<br>Citrate<br>Injection<br>(100 mcg/2 mL)<br><br>Permeation: no<br>breakthrough<br>up to 240<br>minutes | | The following chemotherapy drugs have low permeation times:<br>Carmustine (3.3 mg/ml) : 33.8 minutes Thiotepa (10.0 mg/ml) : 128.1 minutes<br>Warning: Not for Use with: Carmustine, Thiotepa | | | | {7}------------------------------------------------ ### F. SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS OF THE DEVICE COMPARED TO THE PREDICATE DEVICE | CHARACTERISTICS | STANDARDS | DEVICE PERFORMANCE | | Comparison | | |-----------------------------------------------------------------------|-----------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------|------------| | | | PREDICATE | SUBJECT | | | | 510(K) Number | --- | K231349 | K23970 | | | | Name of device | --- | Nephron Nitrile Powder-Free<br>Nitrile Examination Gloves<br>(Tested For Use With<br>Chemotherapy Drugs and<br>Fentanyl) | Nephron Nitrile™, Nitrile<br>Powder-Free Examination<br>Blue Gloves (Tested For Use<br>With Chemotherapy Drugs<br>and Fentanyl) | Similar | | | Product Code | --- | LZA, LZC, OPJ, QDO | LZA, LZC, OPJ, QDO | Identical | | | Indication for use | --- | Nephron Nitrile Powder-Free<br>Nitrile Examination Gloves<br>(Tested For Use With<br>Chemotherapy Drugs and<br>Fentanyl) is a disposable<br>device intended for medical<br>purpose that is worn on the<br>examiner's hand to prevent<br>contamination between<br>patient and examiner. In<br>addition, these gloves were<br>tested for use with<br>chemotherapy drugs and<br>fentanyl in accordance with<br>ASTM D6978-05 (2019)<br>Standard Practice for<br>Assessment of Resistance of<br>Medical Gloves to<br>Permeation by<br>Chemotherapy Drugs. | Nephron Nitrile™, Nitrile<br>Powder-Free Examination<br>Blue Gloves (Tested For Use<br>With Chemotherapy Drugs<br>and Fentanyl) is a disposable<br>device intended for medical<br>purpose that is worn on the<br>examiner's hand to prevent<br>contamination between<br>patient and examiner. In<br>addition, these gloves were<br>tested for use with<br>chemotherapy drugs and<br>fentanyl in accordance with<br>ASTM D6978-05 (2019)<br>Standard Practice for<br>Assessment of Resistance of<br>Medical Gloves to<br>Permeation by<br>Chemotherapy Drugs. | Identical | | | Regulation Number | --- | 21 CFR 880.6250 | 21 CFR 880.6250 | Identical | | | Material | --- | Nitrile | Nitrile | Identical | | | Color | --- | Blue Gloves | Blue | Identical | | | Size | --- | M, L, XL, XXL | XS, S, M, L, XL, XXL | Similar | | | | | | | Adding size XS and S | | | Single Use | --- | Single-use | Single-use | Identical | | | Sterile/Non-sterile | --- | Non-Sterile | Non-Sterile | Identical | | | Rx Only or OTC | --- | OTC | OTC | Identical | | | CHARACTERISTICS | | DEVICE PERFORMANCE | | Comparison | | | 510(K) Number | STANDARDS | PREDICATE | SUBJECT | | | | K231349 | --- | K231349 | K23970 | | | | Dimensions - Length | ASTM D6319-19 | Minimum 230 mm<br>(sizes M – XXL) | Minimum 220 mm<br>(Sizes XS-S)<br>Minimum 230 mm<br>(sizes M – XXL) | Similar<br>Adding size XS and S | | | Dimensions - Width | ASTM D6319-19 | M: 95±10 mm<br>L: 110±10 mm<br>XL: 120±10 mm<br>XXL: 130±10 mm | XS: 70±10 mm<br>S: 80±10 mm<br>M: 95±10 mm<br>L: 110±10 mm<br>XL: 120±10 mm<br>XXL: 130±10 mm | Similar<br>Adding size XS and S | | | Physical Properties-Tensile Strength | ASTM D6319-19 | Before aging<br>14MPa, min<br><br>After aging<br>14MPa, min | Before aging<br>14MPa, min<br><br>After aging<br>14MPa, min | Identical | | | Physical Properties-Ultimate Elongation | ASTM D6319-19 | Before aging<br>500%, min<br><br>After aging<br>400%, min | Before aging<br>500%, min<br><br>After aging<br>400%, min | Identical | | | Thickness | ASTM D6319-19 | Palm: Minimum<br>0.05 mm<br>Finger: Minimum<br>0.05 mm | Palm: Minimum<br>0.05 mm<br>Finger: Minimum<br>0.05 mm | Identical | | | Powder Free Residue | ASTM D6319-19 | ≤2 mg per glove | ≤ 2 mg per glove | Identical | | | Freedom from holes | ASTM D5151-2019 | In accordance with<br>ASTM D 5151-19,<br>following ASTM D6319-19,<br>G-I, AQL 2.5 | In accordance with<br>ASTM D 5151-19,<br>following ASTM D6319-19,<br>G-I, AQL 2.5 | Identical | | | Chemotherapy Drugs Tested with Minimum Breakthrough Detection Time | ASTM D6978-05 (2019) | Bleomycin Sulfate<br>15 mg/ml (15,000 ppm)<br>>240 Minutes<br>Busulfan<br>6 mg/ml (6,000 ppm)<br>>240 Minutes<br>Carboplatin<br>10 mg/ml (10,000 ppm)<br>>240 Minutes<br>Carmustine<br>3.3 mg/ml (3,300 ppm)<br>33.8 Minutes<br>Cisplatin<br>1 mg/ml (1,000 ppm)<br>>240 Minutes<br>Cyclophosphamide<br>20 mg/ml (20,000 ppm)<br>>240 Minutes | Bleomycin Sulfate<br>15 mg/ml (15,000 ppm)<br>>240 Minutes<br>Busulfan<br>6 mg/ml (6,000 ppm)<br>>240 Minutes<br>Carboplatin<br>10 mg/ml (10,000 ppm)<br>>240 Minutes<br>Carmustine<br>3.3 mg/ml (3,300 ppm)<br>33.8 Minutes<br>Cisplatin<br>1 mg/ml (1,000 ppm)<br>>240 Minutes<br>Cyclophosphamide<br>20 mg/ml (20,000 ppm)<br>>240 Minutes | Identical | | | CHARACTERISTICS | | STANDARDS | DEVICE PERFORMANCE | | Comparison | | 510(K) Number | | --- | PREDICATE | SUBJECT | | | | | | K231349 | K23970 | | | | | | Cytarabine<br>100 mg/ml (100,000 ppm)<br>>240 Minutes | Cytarabine<br>100 mg/ml (100,000 ppm)<br>>240 Minutes | Identical | | | | | Dacarbazine<br>10 mg/ml (10,000 ppm)<br>>240 Minutes | Dacarbazine<br>10 mg/ml (10,000 ppm)<br>>240 Minutes | Identical | | | | | Daunorubicin HCl<br>5 mg/ml (5,000 ppm)<br>>240 Minutes | Daunorubicin HCl<br>5 mg/ml (5,000 ppm)<br>>240 Minutes | Identical | | | | | Docetaxel<br>10 mg/ml (10,000 ppm)<br>>240 Minutes | Docetaxel<br>10 mg/ml (10,000 ppm)<br>>240 Minutes | Identical | | | | | Doxorubicin HCl<br>2 mg/ml (2,000 ppm)<br>>240 Minutes | Doxorubicin HCl<br>2 mg/ml (2,000 ppm)<br>>240 Minutes | Identical | | | | | Epirubicin HCl<br>2 mg/ml (2,000 ppm)<br>>240 Minutes | Epirubicin HCl<br>2 mg/ml (2,000 ppm)<br>>240 Minutes | Identical | | | | | Etoposide<br>20 mg/ml (20,000 ppm)<br>>240 Minutes | Etoposide<br>20 mg/ml (20,000 ppm)<br>>240 Minutes | Identical | | | | | Fludarabine<br>25 mg/ml (25,000 ppm)<br>>240 Minutes | Fludarabine<br>25 mg/ml (25,000 ppm)<br>>240 Minutes | Identical | | | | | Fluorouracil<br>50 mg/ml (50,000 ppm)<br>>240 Minutes | Fluorouracil<br>50 mg/ml (50,000 ppm)<br>>240 Minutes | Identical | | | | | Gemcitabine<br>38 mg/ml (38,000 ppm)<br>>240 Minutes | Gemcitabine<br>38 mg/ml (38,000 ppm)<br>>240 Minutes | Identical | | | | | Idarubicin HCl<br>1 mg/ml (1,000 ppm)<br>>240 Minutes | Idarubicin HCl<br>1 mg/ml (1,000 ppm)<br>>240 Minutes | Identical | | | | | Ifosfamide<br>50 mg/ml (50,000 ppm)<br>>240 Minutes | Ifosfamide<br>50 mg/ml (50,000 ppm)<br>>240 Minutes | Identical | | | | | Irinotecan<br>20 mg/ml (20,000 ppm)<br>>240 Minutes | Irinotecan<br>20 mg/ml (20,000 ppm)<br>>240 Minutes | Identical | | | | | Mechlorethamine HCl<br>1 mg/ml (1,000 ppm)<br>>240 Minutes | Mechlorethamine HCl<br>1 mg/ml (1,000 ppm)<br>>240 Minutes | Identical | | | | | Melphalan<br>5 mg/ml (5,000 ppm)<br>>240 Minutes | Melphalan<br>5 mg/ml (5,000 ppm)<br>>240 Minutes | Identical | | | | | Methotrexate<br>25 mg/ml (25,000 ppm)<br>>240 Minutes | Methotrexate<br>25 mg/ml (25,000 ppm)<br>>240 Minutes | Identical | | | | | Mitomycin C<br>0.5 mg/ml (500 ppm)<br>>240 Minutes | Mitomycin C<br>0.5 mg/ml (500 ppm)<br>>240 Minutes | Identical | | CHARACTERISTICS | STANDARDS | DEVICE PERFORMANCE |…
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...