K233911 · Carl Zeiss Meditec, AG · HQF · Sep 6, 2024 · Ophthalmic
Device Facts
Record ID
K233911
Device Name
VISULAS combi
Applicant
Carl Zeiss Meditec, AG
Product Code
HQF · Ophthalmic
Decision Date
Sep 6, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.4390
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
VISULAS combi is intended for use in photocoagulating and photodisrupting ocular tissues in the treatment of diseases of the eye, including: • Photocoagulation of the retina • Trabeculoplasty • Iridotomy • Posterior capsulotomy • Posterior membranectomy
Device Story
VISULAS combi is an ophthalmic laser system for photocoagulation (532 nm) and photodisruption (1064 nm) of ocular tissues. System components include laser console, touch control panel, laser slit lamp or indirect ophthalmoscope, foot switch, and table. Operates in single-spot, multi-spot, and YAG disruption modes. Used in clinical settings by physicians. Photocoagulation uses continuous wave thermal energy absorption; photodisruption uses laser-induced plasma formation to create mechanical shock waves for tissue incision. Output is controlled laser energy delivered to ocular structures. Benefits include precise treatment of retinal and anterior segment pathologies. Physician visualizes target via slit lamp and controls parameters via console to affect clinical decision-making and therapeutic outcomes.
Clinical Evidence
No clinical data. Substantial equivalence established via bench testing, biocompatibility testing (ISO 10993-1), and software verification/validation. System testing confirmed compliance with electrical/laser safety standards (ANSI/AAMI ES60601-1, IEC 60601-1-2, IEC 60825-1, etc.).
Technological Characteristics
Solid-state frequency-doubled laser (532 nm) and Q-switched Nd:YAG laser (1064 nm). Features include LED-illuminated slit lamp, micromanipulator, and foot switch. Connectivity via laser console. Software-controlled treatment modes (VERTE, VITE, YAG licenses). Complies with IEC 60601-1, IEC 60825-1, and ISO 15004-2.
Indications for Use
Indicated for patients requiring photocoagulation or photodisruption of ocular tissues for eye diseases, including retinal photocoagulation, trabeculoplasty, iridotomy, posterior capsulotomy, and posterior membranectomy.
Regulatory Classification
Identification
An ophthalmic laser is an AC-powered device intended to coagulate or cut tissue of the eye, orbit, or surrounding skin by a laser beam.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 6, 2024
Carl Zeiss Meditec AG Tanesha Bland Sr. Regulatory Affairs Specialist Carl Zeiss Meditec, Inc. 5300 Central Parkway Dublin, California 94568
Re: K233911
Trade/Device Name: VISULAS combi Regulation Number: 21 CFR 886.4390 Regulation Name: Ophthalmic Laser Regulatory Class: Class II Product Code: HQF Dated: August 2, 2024 Received: August 5, 2024
Dear Tanesha Bland:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Claudine H. Krawczyk -S
Claudine Krawczyk Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K233911
Device Name VISULAS combi
Indications for Use (Describe)
VISULAS combi is intended for use in photocoagulating and photodisrupting ocular tissues in the treatment of diseases of the eye, including:
• Photocoagulation of the retina
- Trabeculoplasty
- · Iridotomy
- · Posterior capsulotomy
· Posterior membranectomy
This device is Prescription Use (Rx) only.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "ZEISS" in white text on a blue background. The text is in a bold, sans-serif font. The blue background is a solid color, and the word "ZEISS" is centered on the background. The bottom of the blue background is curved.
In accordance with 21 CFR 807.87(h) and (21 CFR 807.92), the 510(k) Summary for VISULAS combi is provided below.
#### SUBMITTER 1.
| Applicant: | Carl Zeiss Meditec AG<br>Goeschwitzer Strasse 51-52<br>07745 Jena<br>Germany |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact | Dr. Ling Ren<br>Regulatory Affairs Manager<br>Carl Zeiss Meditec AG<br>Goeschwitzer Strasse 51-52 07745 Jena, Germany<br>Phone: +49 7364 2062845<br>E-mail: tanesha.bland@zeiss.com (preferred) |
| Primary Correspondent | Tanesha Bland<br>Sr. Regulatory Affairs Specialist<br>Carl Zeiss Meditec, Inc.<br>5300 Central Parkway Dublin, CA 94568<br>Phone: (925) 216-7963 Fax: (925) 557-4259<br>E-mail: tanesha.bland@zeiss.com (preferred) |
| Date Prepared: | August 30, 2024 |
Page 1 of 8
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "ZEISS" in white text on a dark blue background. The text is in a bold, sans-serif font. The logo is simple and modern. The image is likely a company logo.
#### SUBJECT DEVICE 2.
| Device Trade Name: | VISULAS combi |
|--------------------|---------------------------------|
| 510(k) number: | K233911 |
| Classification: | 21CFR886.4390 laser, Ophthalmic |
| Regulatory Class: | II |
| Product Code: | HQF |
#### PREDICATE DEVICE 3.
#### Table 1. Primary Predicate Device- VISULAS green
| Predicate Device: | VISULAS green |
|-------------------|---------------------------------|
| 510(k) Number: | K232051 |
| Classification: | 21CFR886.4390 Laser, Ophthalmic |
| Regulatory Class: | II |
| Product Code: | HQF |
| Table 2. Secondary Predicate Device- VISULAS yag | | |
|--------------------------------------------------|--|--|
| | | |
| Predicate Device: | VISULAS yag |
|-------------------|---------------------------------|
| 510(k) Number: | K230350 |
| Classification: | 21CFR886.4390 Laser, Ophthalmic |
| Regulatory Class: | II |
| Product Code: | HQF |
#### DEVICE DESCRIPTION 4.
VISULAS combi is an ophthalmic laser used for standard photocoagulation treatments of ocular tissue with 532 nm wavelength and standard photodisruption treatments of ocular tissues at a wavelength of 1064 mm. VISULAS combi is operated in the following treatment modes:
- single-spot mode (software license VERTE) -
- multi-spot mode (software license VITE)
- YAG disruption mode (software license YAG). -
VISULAS combi consists of a laser console, touch control panel, optional laser light applicators such as laser slit lamp or laser indirect ophthalmoscope, foot switch and instrument table. The device can also be used with various accessories, such as SL Imaging Solution, contact lenses or Applanation Tonometer.
#### ನ. INDICATIONS FOR USE
VISULAS combi is intended for use in photocoagulating and photodisrupting ocular tissues in the treatment of diseases of the eye, including
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "ZEISS" in white letters against a blue background. The word is in a bold, sans-serif font. The blue background is a solid color, and the white letters stand out clearly against it. The logo is simple and modern.
- . Photocoagulation of the retina
- Trabeculoplasty
- Iridotomy •
- Posterior capsulotomy •
- Posterior membranectomy .
This device is Prescription Use (Rx) only.
#### 6. SUBSTANTIAL EQUIVALENCE TO PRIMARY PREDICATE
#### Table 1. Subject to Primary Predicate Device Comparison Table – Indications for Use
| | VISULAS combi | VISULAS green |
|-------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K number | Subject Device | K232051 |
| Legal manufacturer | Carl Zeiss Meditec AG | Carl Zeiss Meditec AG |
| Indications for Use Statement | VISULAS combi is intended for use<br>in photocoagulating and<br>photodisrupting ocular tissues in the<br>treatment of diseases of the eye,<br>including<br>• Photocoagulation of the retina<br>• Trabeculoplasty<br>• Iridotomy<br>• Posterior capsulotomy<br>• Posterior membranectomy | VISULAS green is intended for use<br>in photocoagulating ocular tissues in<br>the treatment of diseases of the eye,<br>including<br>• Photocoagulation of the retina<br>• Trabeculoplasty<br>• Iridotomy |
| | General Laser Specifications | |
| Laser type | solid state laser, frequency-doubled | solid state laser, frequency-doubled |
| Wavelength | 532 nm | 532 nm |
| Power (cw laser) | 50 to 1500 mW | 50 to 1500 mW |
| Length of pulse | 10 - 2500 ms and cw | 10 - 2500 ms and cw |
| Repeat mode | yes | yes |
| Laser spot size settings | 50 μm to 1000 μm | 50 μm to 1000 μm |
| Aiming beam | 620 - 650 nm | 620 - 650 nm |
| Types of laser applicators | LSL, laser endoprobes, LIO | LSL, laser endoprobes, LIO |
| Multi-spot treatment mode | yes | yes |
| Increments of power | 50-200 mW (10 mW steps), 200-<br>500 mW (20 mW steps), 500-1500<br>mW (50 mW steps) | 50-200 mW (10 mW steps), 200-<br>500 mW (20 mW steps), 500-1500<br>mW (50 mW steps) |
| Laser spot size setting | Continuously adjustable from<br>50 μm to 1000 μm | Continuously adjustable from<br>50 μm to 1000 μm |
| | Conditions of Use | |
| Site in the body | eye | eye |
| Principle of operation | Photocoagulation of ocular tissue | Photocoagulation of ocular tissue |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "ZEISS" in a bold, sans-serif font. The word is white and is set against a dark blue background. The bottom of the image shows a curved white shape, which appears to be part of a larger design element.
# Table 2. Subject to Primary Predicate Device Comparison Table - Laser Slit Lamp
The Laser Slit Lamp (LSL) is an optional component of the VISULAS laser.
| | LSL green combi | LSL green comfort |
|---------------------|---------------------------------------------------------|----------------------------------------------------------|
| K number | Component of subject device<br>VISULAS combi | Component of predicate device<br>VISULAS green (K232051) |
| Legal manufacturer | Carl Zeiss Meditec AG | Carl Zeiss Meditec AG |
| Design | | |
| Slit lamp type | ZEISS type | ZEISS type |
| Tube system | Parallel or convergent | Parallel or convergent |
| Power supply | Via laser console | Via laser console |
| Specifications | | |
| Slit width | 0 mm to 14 mm (continuously) | 0 mm to 14 mm (continuously) |
| Slit length | 1/3/5/9/14 mm | 1/3/5/9/14 mm |
| Slit image rotation | 0°, ±45°, 90° | 0°, ±45°, 90° |
| Magnification | 5 magnifications, in steps of 5x, 8x,<br>12x, 20x, 32x | 5 magnifications, in steps of 5x, 8x,<br>12x, 20x, 32x |
| Light source | LED illumination, continuously<br>adjustable brightness | LED illumination, continuously<br>adjustable brightness |
| Micromanipulator | yes | yes |
The subject device VISULAS combi and the primary predicate device VISULAS green are both intended as ophthalmic lasers for photocoagulation treatments, the indications for use are within the same intended use as the predicate device and do not raise different questions of safety and effectiveness.
The subject device VISULAS combi and the primary predicate laser system share the same fundamental parameters and the same fundamental principle of operation:
Photocoagulation: Continuous wave operation in which thermal energy created by the absorption of the laser energy by the ocular tissue is causing coagulation.
For photocoagulation treatments, the subject device VISULAS combi is substantially equivalent to the primary predicate laser system presented in the 510(k) premarket notification in terms of indications for use and technological characteristics. Differences between subject devices do not raise any new issues of safety or effectiveness.
| | | | Table 3. Subject to Secondary Predicate Device Comparison Table - Indications for Use | |
|--|--|--|---------------------------------------------------------------------------------------|--|
| | | | | |
| | VISULAS combi | VISULAS yag |
|-------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K number | Subject Device | K230350 |
| Legal manufacturer | Carl Zeiss Meditec AG | Carl Zeiss Meditec AG |
| Indications for Use Statement | VISULAS combi is intended for use<br>in photocoagulating and<br>photodisrupting ocular tissues in the<br>treatment of diseases of the eye,<br>including<br>Photocoagulation of the retina Trabeculoplasty | VISULAS yag is intended for use in<br>photodisrupting ocular tissues in the<br>treatment of diseases of the eye,<br>including Posterior capsulotomy Iridotomy Posterior membranectomy |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the ZEISS logo. The logo consists of the word "ZEISS" in a bold, sans-serif font. The text is white and is set against a dark blue background.
Page 5 of 8
| | Iridotomy Posterior capsulotomy Posterior membranectomy | |
|---------------------------------------------------------|----------------------------------------------------------------------------|----------------------------------------------------------------------------|
| General Laser Specifications | | |
| Optic principle | Q-switched Nd:YAG laser | Q-switched Nd:YAG laser |
| Wavelength of treatment beam | 1064 nm | 1064 nm |
| Pulse duration | < 4 ns | < 4 ns |
| Laser class in accordance with<br>IEC 60825-1 | 4 | 4 |
| Light hazard classification according<br>to ISO 15004-2 | Group 2 | Group 2 |
| Pulse Mode 1 (Single Pulse) | 9.0 mJ to 13.0 mJ, at max. 2.5 Hz | 9.0 mJ to 13.0 mJ, at max. 2.5 Hz |
| Pulse Mode 2 (Double Pulse) | 18.0 mJ to 28.0 mJ, at max. 1 Hz<br>(burst frequency 33 kHz) | 18.0 mJ to 28.0 mJ, at max. 1 Hz<br>(burst frequency 33 kHz) |
| Pulse Mode 3 (Triple Pulse) | 29.0 mJ to 45.0 mJ, at max. 1 Hz (2<br>shots/ 2s) (burst frequency 33 kHz) | 29.0 mJ to 45.0 mJ, at max. 1 Hz (2<br>shots/ 2s) (burst frequency 33 kHz) |
| Conditions of Use | | |
| Site in the body | eye | eye |
| Principle of operation | Photodisruption of ocular tissue | Photodisruption of ocular tissue |
#### Table 4. Subject to Reference Device Comparison Table - Laser Slit Lamp
The Laser Slit Lamp (LSL) is an optional component of the VISULAS laser.
| | LSL green combi | LSL yag |
|-------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| K number | Component of subject device<br>VISULAS combi | Component of reference device<br>VISULAS yag (K230350) |
| Legal manufacturer | Carl Zeiss Meditec AG | Carl Zeiss Meditec AG |
| Design | | |
| Laser beam guidance | Free beam arrangement in the head of<br>the laser slit lamp | Free beam arrangement in the head of<br>the laser slit lamp |
| Energy attenuation | 22 levels:<br>2/4/6/8/10/12/14/16/20/24/<br>28/32/36/40/42/48/56/60/6 4/70/80/100<br>% | 22 levels:<br>2/4/6/8/10/12/14/16/20/24/<br>28/32/36/40/42/48/56/60/6 4/70/80/100<br>% |
| YAG focus diameter | 6.5 µm ± 20% in air | 6.5 µm ± 20% in air |
| YAG laser beam divergence | 14° ± 20% (round angle) | 14° ± 20% (round angle) |
| YAG aiming beam | Diode, 2-point / 4-point (switchable)<br>660 nm to 680 nm, max. 150 µm at the<br>cornea (class 1 in accordance with<br>IEC 60825-1) | Diode, 2-point / 4-point (switchable)<br>660 nm to 680 nm, max. 150 µm at the<br>cornea (class 1 in accordance with<br>IEC 60825-1) |
| Focus shift between aiming and<br>therapy beam | 0/150/225/300 µm, toggle between<br>anterior and posterior | 0/150/225/300 µm, toggle between<br>anterior and posterior |
| Nominal Ocular Hazard Distance<br>(NOHD) of YAG laser | 4 m | 4 m |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the word ZEISS in a bold, sans-serif font. The word is white and is set against a dark blue background. The logo is simple and modern, and the contrast between the white text and the blue background makes it easy to read.
The subject device VISULAS combi and the secondary predicate device VISULAS yag are both intended as ophthalmic lasers for photodisruption freatments, the indications for use are within the same intended use as the reference device and do not raise different questions of safety and effectiveness. The subject device VISULAS combi and the reference laser system share fundamental parameters and the same fundamental principle of operation:
Photodisruption: At the focus point of the laser beam pressure and temperature are very high to result in the formation of a laser-induced plasma allowing tissue photodisruption to take place. As the plasma expands, it cools down rapidly; hence, the thermal effect on adjacent tissue is very low. The mechanical stress to adjacent tissue due to the shock wave accompanying the photodisruption expansion of the plasma results in an incisional effect.
For photoisruption treatments, the subject device VISULAS combi is substantially equivalent to the secondary laser system presented in the 510(k) premarket notification in terms of indications for use and technological characteristics. Differences between subject device and reference devices do not raise any new issues of safety or effectiveness.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the ZEISS logo. The logo consists of the word "ZEISS" in a bold, sans-serif font, set against a solid blue background. Below the word, there is a curved shape, also in blue, that suggests a lens or optical element. The overall design is clean and modern, reflecting the company's focus on precision optics and technology.
#### 7. SUMMARY OF STUDIES
### Biocompatibility Testing
ZEISS conducted biocompatibility testing on patient-contacting accessories in accordance with ISO 10993-1. The evaluations addressed cytotoxicity, sensitization, aligning with the recommendations delineated in the FDA guidance document "Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices -Part 1: Evaluation and Testing within a Risk Management Process". The testing performed aligns with current recognized standards and meets or exceeds testing performed on the predicate device. Biocompatibility testing demonstrated equivalency between the subject device and the predicate device.
### Laser safety, electrical safety and electromagnetic compatibility (EMC)
VISULAS combi was evaluated against the following requirements and was found to comply with: · ANSI/AAMI ES60601-1:2005/(R) 2012
- ANSI Z80.36-2016
- · IEC 60601-1-2:2014
- · IEC 60825-1:2007
- IEC 60601-2-22:2012
- IEC 62133: 2012
- IEC 60601-4-2
# Software Verification and Validation Testing
The software of this device is considered to have an Enhanced Documentation Level. Software verification and validation testing was conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." Verification and validation of VISULAS combi demonstrated that the product works as designed.
### Bench Testing
Non-clinical system testing provided an evaluation of the system relevant to each of the system specifications. The functional and system level testing showed that the system met the defined specifications. ZEISS demonstrated non-clinical equivalency between the subject device and the predicate device.
### Animal Testing
Not applicable. Animal studies are not necessary to establish the substantial equivalence of this device.
# Clinical Data
Not applicable. Clinical studies are not necessary to establish the substantial equivalence of this device.
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image contains the word "ZEISS" in a bold, sans-serif font. The word is white and is set against a solid blue background. The logo is simple and modern, with a focus on the company name.
#### CONCLUSION 8.
The VISULAS combi is substantially equivalent to the primary predicate device VISULAS green (K232051) and the secondary predicate device VISULA yag (K230350).
The VISULAS combi is similar in technological characteristics, performance, principles of operation, and has identical indications for use as the primary predicate device and the secondary predicate device. Any differences between the proposed device and the predicate devices do not raise any new issues.
Results of the non-clinical performance testing support a determination of substantial equivalence between the subject device and the predicate devices for the proposed intended use.
Thus, the VISULAS combi meets the requirements for substantial equivalence as compared to the predicate devices, VISULAS green (K232051) and VISULAS yag (K230350).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.