← Product Code [HQC](/productcode/HQC) · K233902

# Centurion™ Vision System (Active Sentry™) (8065753057) (K233902)

_Alcon Laboratories, Inc. · HQC · Jan 10, 2024 · Ophthalmic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K233902

## Device Facts

- **Applicant:** Alcon Laboratories, Inc.
- **Product Code:** [HQC](/productcode/HQC.md)
- **Decision Date:** Jan 10, 2024
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 886.4670
- **Device Class:** Class 2
- **Review Panel:** Ophthalmic
- **Attributes:** Therapeutic

## Indications for Use

The Centurion™ Vision System (Active Sentry™) is indicated for emulsification, separation, irrigation, and aspiration of cataracts, residual cortical material and lens epithelial cells, vitreous aspiration and cutting associated with anterior vitrectomy, bipolar coagulation, and intraocular lens injection. The AutoSert™ IOL Injector Handpiece is intended to deliver qualified AcrySof™ intraocular lenses into the eye following cataract removal. The AutoSert™ IOL Injector Handpiece achieves the functionality of injection of intraocular lenses. The AutoSert™ IOL Injector Handpiece is indicated for use with AcrySof ™ lenses SN60WF. SN6AD1, SN6AT3 through SN6AT9, as well as approved AcrySof ™ lenses that are specifically indicated for use with this inserter, as indicated in the approved labeling of those lenses.

## Device Story

Ophthalmic surgical system for small incision cataract extraction and IOL injection; utilizes CENTURION™ OZil™, Active Sentry™, and INFINITI™ OZil™ handpieces. System emulsifies and aspirates lens material while replacing with balanced salt solution (BSS™) to maintain stable eye chamber volume. Surgeon controls phaco power, aspiration vacuum, flow rate, vitrectomy cut rate, IOL injection rate, and coagulation power via footswitch. Used in clinical surgical settings by ophthalmologists. Provides real-time fluidics management to assist in cataract removal and lens implantation; benefits patient through stable intraocular pressure and controlled surgical environment.

## Clinical Evidence

No clinical data provided. Substantial equivalence is supported by the fact that the only change is a labeling modification (new warning statement for accessories), with no changes to the device design, materials, or technology.

## Technological Characteristics

Phacofragmentation system; includes ultrasonic handpieces (OZil™), fluidics management, and IOL injection capability. Operates via controlled power, aspiration, and irrigation. System components, materials, and sterilization processes remain identical to the predicate device (K161794).

## Regulatory Identification

A phacofragmentation system is an AC-powered device with a fragmenting needle intended for use in cataract surgery to disrupt a cataract with ultrasound and extract the cataract.

## Predicate Devices

- CENTURION™ Vision System (Active Sentry™) ([K161794](/device/K161794.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

January 10, 2024

Alcon Laboratories, Inc. Rita De Rama Manager, Regulatory Affairs 6201 South Freeway Fort Worth, Texas 76134

Re: K233902

Trade/Device Name: Centurion™ Vision System (Active Sentry™) (8065753057) Regulation Number: 21 CFR 886.4670 Regulation Name: Phacofragmentation System Regulatory Class: Class II Product Code: HQC Dated: December 6, 2023 Received: December 11, 2023

Dear Rita De Rama:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

{1}------------------------------------------------

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product: and 21 CFR 820.100. Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

{2}------------------------------------------------

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Claudine H. Krawczyk -S

Claudine Krawczyk Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure

{3}------------------------------------------------

### Indications for Use

Submission Number (if known)

K233902

Device Name

Centurion TM Vision System (Active Sentry TM) (8065753057)

Indications for Use (Describe)

The Centurion™ Vision System (Active Sentry™) is indicated for emulsification, separation, irrigation, and aspiration of cataracts, residual cortical material and lens epithelial cells, vitreous aspiration and cutting associated with anterior vitrectomy, bipolar coagulation, and intraocular lens injection.

The AutoSert™ IOL Injector Handpiece is intended to deliver qualified AcrySof™ intraocular lenses into the eye following cataract removal.

The AutoSert™ IOL Injector Handpiece achieves the functionality of injection of intraocular lenses. The AutoSert™ IOL Iniector Handbiece is indicated for use with AcrySof ™ lenses SN60WF. SN6AD1, SN6AT3 through SN6AT9, as well as approved AcrySof ™ lenses that are specifically indicated for use with this inserter, as indicated in the approved labeling of those lenses.

Type of Use (Select one or both, as applicable)

escription Use (Part 21 CFR 801 Subpart D)

he-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

## 510(k) Summarv

Prepared on: 2024-01-08

| Contact Details                    | 21 CFR 807.92(a)(1)                                      |              |
|------------------------------------|----------------------------------------------------------|--------------|
| Applicant Name                     | Alcon Laboratories, Inc.                                 |              |
| Applicant Address                  | 6201 South Freeway Fort Worth TX 76134 United States     |              |
| Applicant Contact Telephone        | 949 886 5787                                             |              |
| Applicant Contact                  | Mrs. Rita De Rama                                        |              |
| Applicant Contact Email            | rita.derama@alcon.com                                    |              |
| Device Name                        | 21 CFR 807.92(a)(2)                                      |              |
| Device Trade Name                  | Centurion™ Vision System (Active Sentry™) (8065753057)   |              |
| Common Name                        | Phacofragmentation system                                |              |
| Classification Name                | Unit, Phacofragmentation                                 |              |
| Regulation Number                  | 886.4670                                                 |              |
| Product Code                       | HQC                                                      |              |
| Legally Marketed Predicate Devices | 21 CFR 807.92(a)(3)                                      |              |
| Predicate #                        | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| K161794                            | CENTURION™ Vision System (Active Sentry™)                | HQC          |
| Device Description Summary         | 21 CFR 807.92(a)(4)                                      |              |

Alcon's CENTURION™ Vision System is an ophthalmic surgical instrument designed to provide cataract extraction using the CENTURION™ OZil™handpiece, the CENTURION™ Active Sentry™ handpiece, and the INFIN门!" OZZI" handpiece.

The CENTURION™ Vision System is intended for use in small incision cataraction and IOL injection surgical procedures. This system allows the surgeon to emulsify and aspirate the eye, while replacing aspirated fluid and lens material with balanced salt solution (BSS™). This process maintains a stable (inflated) eye chamber volume. Using system controls, the surgeon regulates the amount of power applied to the handpiece tip, the rate of aspiration vacuum, and the flow of BSS™ irrigating solution. The system includes a footswitch to enable the surgeon to control flow of fluidics, aspiration rate, phaco power, vitrectory cut rate, IOL injection rate, and coagulation power.

## Intended Use/Indications for Use

The Centurion™ Vision System (Active Sentry") is indication, separation, irrigation, and aspiration of cataracts, residual cortical material and lens epithelial cells, vitreous and cutting associated with anterior vitrectomy, bipolar coagulation, and intraocular lens injection.

The AutoSert™ IOL Injector Handpiece is intended to deliver qualified AcrySof™

intraocular lenses into the eye following cataract removal.

The AutoSert™ IOL Injector Handpiece achieves the function of intraocular lenses. The AutoSert™ IOL Injector Handpiece is indicated for use with AcrySof ™ lenses SN60WF, SN6AT3 through SN6AT9, as well as approved AcrySof™ lenses that are specifically indicated for use with this inserter, as indicated in the approved labeling of those lenses.

### Indications for Use Comparison

21 CFR 807.92(a)(5)

21 CFR 807.92(a)(5)

{5}------------------------------------------------

The indications for use are the same between the subject device and the predicate device.

## Technological Comparison

21 CFR 807.92(a)(6)

A modification to add a new Warning statement in the labeling for the TurboSonics™ ultrasonic tips, which are accessories for the CENTURION® Vision System, is the only change being made to the predicate device.

The modified CENTURION™ Vision System (Active Sentry") is substantially equivally submitted CENTURION™ Vision System (Active Sentry™) in that both devices:

· Have the same indication for use.

· Use the same operating principle.

· Use the same technology, including components used.

· Are of the same design, incorporate the same materials, have the same shelf life, and are packaged and sterilized using the same materials and processes.

#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)

Non-clinical testing was previously performed in the predicate device and can be referenced in K161794. A modification to add a new Warning statement in the labeling for the TurboSonics™ ultrasonic tips, which are accessories for the CENTURION® Vision System, is the only change being made to the predicate device. Thus, the subject device described in this submission is substantially equivalent to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K233902](https://fda-staging.innolitics.com/device/K233902)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).