K233881 · Ivwatch, LLC · PMS · Mar 15, 2024 · General Hospital
Device Facts
Record ID
K233881
Device Name
ivWatch® Model 400
Applicant
Ivwatch, LLC
Product Code
PMS · General Hospital
Decision Date
Mar 15, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The ivWatch Model 400 is intended to aid in the detection of infiltrations and extravasations during peripheral IV infusion therapy in pediatric and adult patients. The user profile is healthcare practitioners who are experienced in IV administration and management and located at hospitals and similar medical care facilities.
Device Story
The ivWatch Model 400 provides continuous, non-invasive monitoring of tissue near PIV insertion sites to detect infiltration/extravasation. The system includes a patient monitor, fiber optic sensor cable, sensor receptacle, extension module, patient cable, and SmartTouch sensor. The monitor emits visible and near-infrared light through the sensor to the skin and measures reflected light. Signal processing algorithms analyze changes in optical properties; if consistent with subcutaneous infusate pooling, the device triggers audible and visual alerts. Used in hospitals/medical facilities by trained healthcare practitioners. Output assists clinicians in identifying potential infiltration events, enabling earlier intervention and reducing patient risk. The subject device adds iron sucrose (1-20 mg/ml) to the previously cleared optically clear infusates.
Clinical Evidence
Bench testing only. Testing evaluated detection of iron sucrose concentrations from 1mg/ml to 20mg/ml. Results showed mean signal drops of 47.1% (std: 7.97%) for iron compared to 23.7% (std: 6.79%) for saline. Iron injection signals were not weaker than saline signals (p≈1.000).
Technological Characteristics
Uses visible and near-infrared light sensing. Components include patient monitor, fiber optic sensor cable, sensor receptacle, extension module, patient cable, and SmartTouch sensor. Biocompatibility per ISO 10993-1, 10993-5, 10993-10. Electrical safety/EMC per IEC 60601-1 and IEC 60601-1-2. Sterilization of sensors/receptacles via EO half-cycle overkill (ISO 11135).
Indications for Use
Indicated for detection of subcutaneous infiltrations and extravasations of 10 cc or less of optically clear and iron sucrose infusates, as an adjunctive device to clinical evaluation in healthcare settings for adults and pediatrics with peripherally-inserted catheters (PIVs). Not a substitute for regular clinician assessment. For use by trained healthcare practitioners.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the agency's name, "U.S. Food & Drug Administration."
March 15, 2024
ivWatch, LLC Holly Novak Vice President, Regulatory Affairs & Quality Assurance 700 Tech Center Parkway Suite 300 Newport News, Virginia 23606
Re: K233881
Trade/Device Name: ivWatch® Model 400 Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion pump Regulatory Class: Class II Product Code: PMS Dated: February 16, 2024 Received: February 16, 2024
Dear Holly Novak:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
David Walloschek
David Wolloscheck, Ph.D. Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
{2}------------------------------------------------
Office of Product Evaluation and Quality Center for Devices and Radiological Health
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K233881
Device Name ivWatch® Model 400
#### Indications for Use (Describe)
The ivWatch Model 400 is indicated for the detection of subcutaneous infiltrations and extravasations of 10 cc or less of optically clear and iron sucrose infusates, as an adjunctive device to the clinical evaluation in the healthcare setting of adults and pediatrics with peripherally-inserted catheters (PIVs). The device is indicated to assess patients for the subcutaneous infiltrations and extravasations but should not serve as a substitute for regular clinician assessment of the PIV site. The iv Watch Model 400 is intended for use by healthcare practitioners who have been tramed in the use of the device.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|-----------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 20px;">☑</span> |
| Over-The-Counter Use (21 CFR 801 Subpart C) | ☐ |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# K233881 - 510(K) SUMMARY
| Device Submitter | ivWatch, LLC | |
|------------------|----------------------------------------------------------------------------|--|
| | 700 Tech Center Parkway, Suite 300 | |
| | Newport News, VA 23606 | |
| Contact Person | Holly Novak, Vice President of Regulatory Affairs<br>and Quality Assurance | |
| | holly.novak@ivwatch.com 855-489-<br>2824 x7046 | |
| | Fax: 757-224-5009 | |
| Date Prepared | December 07, 2023 | |
## Subject Device
| Trade Name | ivWatch® Model 400 |
|---------------------|---------------------------------------------------|
| Device Class | II |
| Regulation Number | 21 CFR 880.5725 |
| Product Code | PMS |
| Common Name | Peripheral Intravenous (PIV) Infiltration Monitor |
| Classification Name | Infusion pump |
# Predicate Device
| Trade Name | ivWatch® Model 400 |
|---------------------|---------------------------------------------------|
| 510(k) Number | K192385 |
| Device Class | II |
| Regulation Number | 21 CFR 880.5725 |
| Product Code | PMS |
| Common Name | Peripheral Intravenous (PIV) Infiltration Monitor |
| Classification Name | Infusion pump |
{5}------------------------------------------------
#### Device Description
The ivWatch Model 400 is a medical device that provides continuous, non-invasive monitoring of human tissue adjacent to peripheral intravenous (PIV) insertion sites to aid in the early detection of infiltration and extravasation events. The predicate device includes a Patient Monitor, a Fiber Optic Sensor Cable, and a disposable Sensor Receptacle. Under K192385 the system was expanded to add three Device Accessories including an Extension Module, a Patient Cable and SmartTouch Sensor. The Device Accessories expand system's functionality to support two sensor technologies (fiber optic and electronic).
The ivWatch Model 400 is a medical device that provides continuous, non-invasive monitoring of human tissue adjacent to peripheral intravenous (PIV) insertion sites to aid in the early detection of infiltration and extravasation events.
The ivWatch Model 400 uses visible and near-infrared light to measure changes in the optical properties of the tissue near a PIV insertion site. The ivWatch Patient Monitor (IPM) contains an optical system that generates visible and near-infrared light signals that are sent through the sensor cable to the patient's skin. Simultaneously, the IPM measures the light reflected back through the sensor cable from the patient's skin. Measured changes between the emitted and reflected signal are processed by ivWatch signal processing algorithms to determine if an infiltration event may have occurred. If changes in the optical properties of the tissue near the peripheral IV insertion site are consistent with an infusate pooling in the subcutaneous tissue, the IPM emits audible and visual notifications intended to prompt the clinician to inspect the peripheral IV site for a possible infiltration event.
### Intended Use
The ivWatch Model 400 is intended to aid in the detection of infiltrations and extravasations during peripheral IV infusion therapy in pediatric and adult patients. The user profile is healthcare practitioners who are experienced in IV administration and management and located at hospitals and similar medical care facilities.
The intended use of the subject device is the same as the predicate (K192385).
#### Indications for Use
The ivWatch Model 400 is indicated for the detection of subcutaneous infiltrations and extravasations of 10 cc or less of optically clear and iron sucrose infusates, as an adjunctive device to the clinical evaluation in the healthcare setting of adults and pediatrics with peripherally-inserted catheters (PIVs). The device is indicated to assess patients for the subcutaneous infiltrations and extravasations but should not serve as a substitute for regular clinician assessment of the PIV site. The ivWatch Model 400 is intended for use by healthcare practitioners who have been trained in the use of the device.
{6}------------------------------------------------
The indications for use of the subject device are the predicate (K192385), with the addition of iron sucrose as an indicated infusate.
## Comparison of the Subject Device to the Predicate Devices
The subject device includes all the same device accessories as the predicate. Under this 510K submission, the only change is to the Indications for use to add iron sucrose as an indicated infusate. There are no changes to the devices.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for IVWatch. The logo consists of a stylized teardrop shape on the left, with the letters "iv" inside the teardrop. To the right of the teardrop is the word "Watch" in a sans-serif font. The logo is simple and modern, and the use of the teardrop shape likely represents intravenous fluids.
ivWatch, LLC 700 Tech Center Parkway, Suite 300 Newport News, VA 23606
1.855.IVWATCH (489.2824) Fax: 757.224.5009 www.ivwatch.com
# Summary Table
| Item | Predicate Device<br>K192385 | Subject Device | Substantial<br>Equivalence<br>Discussion |
|-----------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | ivWatch Model 400 | Same | Equivalent |
| Manufacturer | ivWatch, LLC | Same | Equivalent |
| 510(k) Number | K192385 | K233881 | N/A |
| Product Code | PMS | Same | Equivalent |
| Device Class | II | Same | Equivalent |
| Common Name | Peripheral Intravenous (PIV) Infiltration<br>Monitor | Same | Equivalent |
| Regulation Number | 21 CFR 880.5725 | Same | Equivalent |
| Intended Use | The ivWatch Model 400 is intended to aid in<br>the detection of infiltrations and<br>extravasations during peripheral IV<br>infusion therapy in pediatric and adult<br>patients. The user profile is healthcare<br>practitioners who are experienced in IV<br>administration and management and<br>located at hospitals and similar medical<br>care facilities. | Same | Equivalent |
| Indications for Use | The ivWatch Model 400 is indicated for the<br>detection of subcutaneous infiltrations and<br>extravasations of 10 cc or less of optically<br>clear infusates, as an adjunctive device to<br>the clinical evaluation in the healthcare | The ivWatch Model 400 is indicated for<br>the detection of subcutaneous<br>infiltrations and extravasations of 10 cc<br>or less of optically clear and iron sucrose<br>infusates, as an adjunctive device to the | Equivalent<br>Bench testing results<br>support device<br>detection of iron |
| Item | Predicate Device<br>K192385 | Subject Device | Substantial<br>Equivalence<br>Discussion |
| | setting of adults and pediatrics with<br>peripherally-inserted catheters (PIVs). The<br>device is indicated to assess patients for<br>subcutaneous infiltrations and<br>extravasations but should not serve as a<br>substitute for regular clinician assessment<br>of the PIV site. The ivWatch Model 400 is<br>intended for use by healthcare<br>practitioners who have been trained in the<br>use of the device. | clinical evaluation in the healthcare<br>setting of adults and pediatrics with<br>peripherally-inserted catheters (PIVs).<br>The device is indicated to assess patients<br>for subcutaneous infiltrations and<br>extravasations but should not serve as a<br>substitute for regular clinician<br>assessment of the PIV site. The ivWatch<br>Model 400 is intended for use by<br>healthcare practitioners who have been<br>trained in the use of the device. | sucrose; iron infusate<br>do not impact device<br>safety or effectiveness |
| Sterilization | The SmartTouch Sensor and Sensor<br>Receptacles are sterilized using the EO<br>half-cycle overkill approach. A minimum<br>SAL of<br>$10^{-6}$ per ISO 11135, EO residuals of < 4mg<br>and ECH residuals of < 9mg were achieved. | Same | Equivalent |
| Shelf-Life | Packaging passed Burst Testing per ASTM<br>F1140, and Bubble Emission Leak Test per<br>ASTM F2096.<br>Sensor Receptacle 2 years<br>SmartTouch Sensor 1 year | Packaging passed Burst Testing per<br>ASTM F1140, and Bubble Emission Leak<br>Test per ASTM F2096.<br>Sensor Receptacle 3 years<br>SmartTouch Sensor 3 years | Equivalent<br>Shelf life was<br>determined by<br>meeting acceptance<br>criteria in compliance<br>with requirements<br>outlined in ASTM<br>F1886, ASTM F2096<br>and ASTM F1140 |
| Item | Predicate Device<br>K192385 | Subject Device | Substantial<br>Equivalence<br>Discussion |
| Reprocessing | The Patient Monitor, Extension Module,<br>Fiber Optic Sensor Cable and Patient<br>Cables are to be cleaned/disinfected in<br>between uses. Cleaning and low-level<br>disinfection validation testing passed all<br>pre-defined acceptance criteria in<br>compliance with AAMI TIR12 and AAMI<br>TIR30. | Same | Equivalent |
| Biocompatibility | The Sensor Receptacle, Fiber Optic Sensor<br>Cable, SmartTouch Sensor and Patient<br>Cable are classified as having prolonged<br>duration (greater than 24 hours but less<br>than 30 days) patient-contacting<br>components.<br>Biocompatibility testing was conducted in<br>compliance with ISO 10993-1, 10993-5,<br>10993-10. Results show that the patient-<br>contacting components pass all pre-<br>defined acceptance criteria defined for<br>cytotoxicity, sensitization, and irritation<br>testing. | Same | Equivalent |
| Electromagnetic<br>Compatibility (EMC) and<br>Electrical Safety | EMC and electrical safety testing were<br>conducted on the and results show that the<br>ivWatch® Model 400 with Device<br>Accessories complies with the<br>requirements of EN / IEC 60601-1-2 4th<br>Edition and IEC 60601-1 3rd Edition. | Same | Equivalent |
| Item | Predicate Device<br>K192385 | Subject Device | Substantial<br>Equivalence<br>Discussion |
| Software Verification and<br>Validation Testing | Software verification and validation testing for the ivWatch® Model 400 with Device Accessories were conducted and documentation was provided in accordance FDA's Guidance for Industry and Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and IEC 62304. | Same | Equivalent |
| Performance Testing | Optical safety testing was performed and indicated that the optical radiation emitted by the sensors significantly less than the limits defined in ANSI Z136.1. | Same | Equivalent |
| | A series of IRB-approved clinical studies were performed for the verification and validation of the ivWatch Model 400 with Device Accessories. All submitted clinical studies were conducted in the United States and in compliance with 21 CRF 50, 21 CFR 54, 21 CFR 56 and 21 CRF 812, ICH E6 and ISO 14155. The results demonstrated that the ivWatch Model 400 with Device Accessories meets the devices intended use and raises no concerns regarding safety and efficacy. | Same | Equivalent |
| Item | Predicate Device<br>K192385 | Subject Device | Substantial<br>Equivalence<br>Discussion |
| | N/A, Iron data and claims were not<br>included in predicate submission. | Data supports the update to the ivWatch<br>Patient Monitoring System's Indications<br>for Use to detect iron sucrose infusions in<br>addition to optically clear infusions. Iron<br>concentrations from 1mg/ml to 20mg/ml<br>were tested.<br>Results showed:<br>Mean signal drops<br>Saline: 23.7% (std: 6.79%)<br>Iron: 47.1% (std: 7.97%) Iron injection signals not weaker than<br>saline signals (p≈1.000) Diluted iron signal (1mg/ml) is<br>stronger signal than saline. Signal strength increases with higher<br>concentration of iron. | Equivalent<br>Bench testing results<br>support device<br>detection of iron<br>sucrose; iron infusates<br>do not impact device<br>safety or effectiveness. |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for IVWatch. The logo consists of a stylized black teardrop shape with a grid pattern inside, followed by the text "ivWatch" in a sleek, modern font. The logo is simple, clean, and professional.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for "ivWatch". The logo consists of a stylized teardrop shape on the left, with the letters "iv" inside the teardrop. To the right of the teardrop is the word "Watch" in a simple, sans-serif font. The logo is black and white.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for IVWatch. The logo consists of a stylized water droplet shape with a grid pattern in the upper left portion. The text "ivWatch" is written in a modern, sans-serif font to the right of the droplet. A small circle is present to the right of the text.
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for "ivWatch". The logo consists of a stylized black teardrop shape with the letters "iv" inside, formed by white squares. To the right of the teardrop is the word "Watch" in a simple, sans-serif font, with a trademark symbol next to it.
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the logo for "ivWatch". The logo consists of a stylized black teardrop shape on the left, with the letters "iv" inside the teardrop. To the right of the teardrop, the word "Watch" is written in a simple, sans-serif font. The logo is clean and modern, with a focus on the company name.
ivWatch, LLC 700 Tech Center Parkway, Suite 300 Newport News, VA 23606
1.855.IVWATCH (489.2824) Fax: 757.224.5009 www.ivwatch.com
### Conclusions
The technological characteristics, principles of operation and intended use of the ivWatch® Model 400 subject device and the predicate device are similar. Test results show that the ivWatch® Model 400 subject device meets all pre-defined acceptance criteria. The nonclinical and clinical tests demonstrate that the ivWatch® Model 400 is substantially equivalent to the predicate device, K192385.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.