XT Series Power Wheelchair (XT2), XT Series Power Wheelchair (XT4)
K233877 · Magic Mobility · ITI · Mar 4, 2024 · Physical Medicine
Device Facts
Record ID
K233877
Device Name
XT Series Power Wheelchair (XT2), XT Series Power Wheelchair (XT4)
Applicant
Magic Mobility
Product Code
ITI · Physical Medicine
Decision Date
Mar 4, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3860
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
Magic Mobility power chairs are designed for the exclusive use of people (adults and children) who are unable to walk or have limited mobility and have the cognitive, physical and visual ability to control the vehicle safely. The XT Series Power Wheelchairs are intended to be selfpropelled on a range of surfaces and can be used on both indoor surfaces (i.e., concrete, asphalt, indoor flooring such as carpet, gravel, grass, and bark/woodchips).
Device Story
XT Series Power Wheelchair (XT2/XT4) is a battery-powered, electric motor-driven mobility device for indoor/outdoor use. Operated by user via joystick and hand control buttons. Provides mobility to individuals restricted to a sitting position. Features include adjustable seating (MPS or Rehab options), electromagnetic/regenerative braking, and suspension. Used in care facilities, residences, and on varied terrain (concrete, asphalt, grass, gravel). Output is physical movement of the wheelchair. Benefits include restored mobility for users with physical limitations.
Clinical Evidence
Bench testing only. No clinical data. Performance verified via ISO 7176 series standards for stability, braking, energy consumption, dimensions, speed, obstacle climbing, fatigue, and EMC/electrical safety. Biocompatibility assessed per ISO 10993-1.
Technological Characteristics
Steel frame; battery-powered (2x 12V AGM/Gel); 24V 4-pole direct drive gear motors; electromagnetic/regenerative braking; joystick controller; front/rear suspension. Software: R-Net (PGDT). Standards: ISO 7176 series (stability, braking, fatigue, EMC), ISO 10993-1 (biocompatibility), IEC 62304 (software lifecycle).
Indications for Use
Indicated for adults and children with limited mobility or inability to walk who possess the cognitive, physical, and visual capability to safely operate a battery-operated power wheelchair. Intended for indoor and outdoor use on various surfaces.
Regulatory Classification
Identification
A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
Quickie® Q700-UP M / Sunrise Medical (US) LLC (K172384)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 4, 2024
Magic Mobility % Pierre Bounaud Principal Consultant Ram+ 2251 San Diego Ave, Suite B-257 San Diego, California 92110
Re: K233877
Trade/Device Name: XT Series Power Wheelchair (XT2), XT Series Power Wheelchair (XT4) Regulation Number: 21 CFR 890.3860 Regulation Name: Powered Wheelchair Regulatory Class: Class II Product Code: ITI Dated: November 26, 2023 Received: December 7, 2023
Dear Pierre Bounaud:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
# Tushar Bansal -S
for Heather Dean, PhD
Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Rehabilitation Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K233877
#### Device Name
XT Series Power Wheelchair (XT2); XT Series Power Wheelchair (XT4)
#### Indications for Use (Describe)
The XT2 power wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position who have the capability of operating a power wheelchair. The XT4 power wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position who have the capability of operating a power wheelchair.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | |
|----------------------------------------------|---|
| Over-The-Counter Use (21 CFR 801 Subpart C) | X |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
510(k) Summary K233877
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Magic Mobility. The word "magic" is written in a stylized font, with the first half of the word in red and the second half in black. Below the word "magic" is the word "Mobility" written in a cursive font. The logo is simple and modern, and it conveys a sense of movement and innovation.
## K233877 510(k) SUMMARY
#### DATE PREPARED
November 26, 2023
#### MANUFACTURER AND 510(k) OWNER
Magic Mobility 3 International Court, Scoresby, VIC, 3179 Australia Telephone: Official Contact: Terence Walsh, Quality and Regulatory Affairs Manager
#### REPRESENTATIVE/CONSULTANT
Pierre Bounaud, Ph.D. Allison C. Komiyama, Ph.D., RAC RQM+ Telephone: +1 (412) 816-8139 Email: pbounaud@rqmplus.com; akomiyama@rqmplus.com
## DEVICE INFORMATION
| Proprietary Name/Trade Name: | XT Series Power Wheelchair (XT2 and XT4 models) |
|------------------------------|---------------------------------------------------------------|
| Common Name: | Powered wheelchair |
| Regulation Number: | 21 CFR 890.3860 |
| Class: | II |
| Product Code: | ITI |
| Premarket Review: | OHT5/Neuromodulation and Physical Medicine Devices<br>(DHT5B) |
| Review Panel: | Physical Medicine |
## PREDICATE DEVICE IDENTIFICATION
The XT Series Power Wheelchair is substantially equivalent to the following predicates:
| 510(k) Number | Predicate Device Name / Manufacturer | Primary<br>Predicate | Reference<br>Device |
|---------------|-----------------------------------------------|----------------------|---------------------|
| K211574 | Extreme X8 / Magic Mobility | ✓ | |
| K172384 | Quickie® Q700-UP M / Sunrise Medical (US) LLC | | ✓ |
The predicate devices have not been subject to a design related recall.
#### DEVICE DESCRIPTION
The XT Series Power Wheelchair has two configurations: XT2 and XT4. They are designed for everyday use for both indoor and outdoor environments including care facilities and private residences. The subject device is intended to provide mobility to persons who are restricted or limited to a sitting position.
The XT Series Power Wheelchairs are battery powered, electric motor driven devices that can be
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Magic Mobility. The word "magic" is written in a stylized font, with the top half of the letters in red and the bottom half in black. Below the word "magic" is the word "Mobility" written in a cursive font. The letters in "Mobility" are black.
used on both indoor and outdoor surfaces (i.e., concrete, asphalt, indoor flooring such as carpet, gravel, grass, and bark/woodchips). The XT Series Power Wheelchair offers two basic seating options: MPS and Rehab. The MPS option is more static and does not allow for additional aftermarket cushions. The Rehab option is more adjustable to adhere to recommendations from the user's physician or physical therapist.
The XT Series Power Wheelchairs include the following accessories:
- Extra spreader bar ●
- Slide in table ●
- Lights ●
- Luggage rack ●
- Accessory charger
- Posture belt ●
- Roho cushion ●
- Jay cushion ●
- MPS push rail ●
- MPS peg push handle ●
- Scooter stopper
- Fold forward kit ●
## INTENDED USE
Magic Mobility power chairs are designed for the exclusive use of people (adults and children) who are unable to walk or have limited mobility and have the cognitive, physical and visual ability to control the vehicle safely. The XT Series Power Wheelchairs are intended to be selfpropelled on a range of surfaces and can be used on both indoor surfaces (i.e., concrete, asphalt, indoor flooring such as carpet, gravel, grass, and bark/woodchips).
## INDICATIONS FOR USE
The XT2 power wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position who have the capability of operating a power wheelchair.
The XT4 power wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position who have the capability of operating a power wheelchair.
## COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
Magic Mobility believes that the XT Series Power Wheelchair is substantially equivalent to the predicate device based on the information summarized here:
The subject device has a similar design and dimensions and uses similar or identical materials as the devices cleared in K211574. The subject device has the same intended use and similar technological characteristics (i.e.. base technology and OEM jovstick control) to the devices cleared in K211574 and K172384. The subject device has the same intended use environment. including off road capabilities, as the device cleared in K211574. The subject device uses the same software as the device cleared in K211574 and K172384. The Magic XT Series Power Wheelchairs have undergone testing to ensure that any differences in technological characteristics (i.e., battery size, frame design) do not affect safety and effectiveness when compared to the predicate devices.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Magic Mobility. The word "magic" is written in a stylized font, with the "ma" in red and the "gic" in black. Below the word "magic" is the word "Mobility" in a cursive font. The logo is simple and modern.
| | Subject Device | Predicate Device | Reference Device | Statement of Equivalence |
|----------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|
| | Magic Mobility<br>XT Series Power Wheelchair<br>K233877 | Magic Mobility<br>Extreme X8 Power Wheelchair<br>K211574 | Sunrise Medical (US) LLC<br>Quickie® Q700-UP M<br>K172384 | |
| Indications for<br>Use | The XT2 power wheelchair is a battery-<br>operated device with wheels that is<br>intended for medical purposes to provide<br>mobility to persons restricted to a sitting<br>position who have the capability of<br>operating a power wheelchair.<br>The XT4 power wheelchair is a battery-<br>operated device with wheels that is<br>intended for medical purposes to provide<br>mobility to persons restricted to a sitting<br>position who have the capability of<br>operating a power wheelchair. | The Extreme X8 Power Wheelchair is a<br>battery-operated device with wheels that<br>is intended for medical purposes to<br>provide mobility to persons restricted to<br>a sitting position who have the capability<br>of operating a power wheelchair. | The Sunrise Medical Quickie® Q700-<br>UP M power wheelchairs are battery<br>operated devices, that are indicated for<br>medical purposes to provide mobility<br>and repositioning of the user,<br>including a stand-up feature. | Identical to predicate device. No<br>impact on safety or effectiveness. |
| Product Code /<br>Regulation<br>Number | ITI / 21 CFR 890.3860 | ITI / 21 CFR 890.3860 | IPL / 21 CFR 890.3900 | Identical to predicate device. No<br>impact on safety and effectiveness. |
| Regulation<br>Description | Powered Wheelchair | Powered Wheelchair | Standup Wheelchair | Identical to predicate device. No<br>impact on safety and effectiveness. |
| Maximum User<br>Weight (lbs) | XT2: 300<br>XT4: 400 | 400 | 265 | Similar to predicate device. No<br>impact on safety and effectiveness. |
| Storage<br>Temperature<br>(°C) | -40 to 70°C | -40 to 70°C | -40 to 70°C | Identical to predicate and reference<br>devices. No impact on safety and<br>effectiveness. |
| Location for Use | Indoors and outdoors including care<br>facilities, residences, and soft/rough<br>terrain | Indoors and outdoors including care<br>facilities, residences, and soft/rough<br>terrain | Indoors and outdoors including care<br>facilities, and residences | Identical to predicate and reference<br>devices. No impact on safety and<br>effectiveness. |
| Frame Material | Steel | Steel | Steel and aluminum | Identical to predicate and reference<br>devices. No impact on safety and<br>effectiveness. |
| Biocompatibility | Uses materials common to many<br>wheelchairs | Uses materials common to many<br>wheelchairs | Uses materials common to many<br>wheelchairs | Identical to predicate and reference<br>devices. No impact on safety and<br>effectiveness. |
| Ground<br>Clearance | 5" | 4" | Unknown | Similar to predicate device. No<br>impact on safety and effectiveness. |
| | Subject Device | Predicate Device | Reference Device | |
| | Magic Mobility<br>XT Series Power Wheelchair<br>K233877 | Magic Mobility<br>Extreme X8 Power Wheelchair<br>K211574 | Sunrise Medical (US) LLC<br>Quickie® Q700-UP M<br>K172384 | Statement of Equivalence |
| Overall Dimensions<br>(length by width; inches) | XT2: 43.5" x 27"<br>XT4: 39.5"x 27.5" | 45.25" x 27" | 25" x 26" | Similar to predicate device. No impact on safety and effectiveness. |
| Rolling Base Weight (lbs) | XT2: 167<br>XT4: 183 | 145 | 152 | Similar to predicate device. No impact on safety and effectiveness. |
| Power Source | Batteries | Batteries | Batteries | Identical to predicate and reference devices. No impact on safety and effectiveness. |
| Battery Details | 2- Group 31 12-volt AGM or Gel<br>20h capacity in 95-115 AMP/hour range. | 2- 73 AMP/hour AGM Group 24 12-volt<br>Weight 55 lbs | 24V (2x12V) / 73 Ah/20h | Similar to predicate device. No impact on safety and effectiveness. |
| Battery Charger Input Voltage<br>(volts) | 110 | 110 | Unknown | Identical to predicate device. No impact on safety and effectiveness. |
| Motors | XT2: 2 motors<br>XT4: 4 motors<br>24V, 4 pole direct drive gear in line motors/gearbox. Freewheeling lever/motor lock releases & engages motor brakes | 4-24V, 4 pole direct drive gear in line motors/gearbox. Freewheeling lever/motor lock releases & engages rear motor brakes | Unknown | Similar to predicate device. No impact on safety and effectiveness. |
| Wheel Size<br>(inches) | 4" x 14" | 4" x 14" | Unknown | Identical to predicate device. No impact on safety and effectiveness. |
| Turning Radius<br>(inches) | XT2: 69"<br>XT4: 55.5" | 52" | Unknown | Similar to predicate device. No impact on safety and effectiveness. |
| Range (miles) | XT2: 33<br>XT4: 24 | 15.5 | Unknown | Similar to predicate device. No impact on safety and effectiveness. |
| Maximum Speed (mph) | 6.2 | 6.2 | 6 (with an option of 8) | Identical to predicate device. No impact on safety and effectiveness. |
| Anti-pitch Mechanism for Climbing | None | None | Additional anti-pitch lock out | Identical to predicate device. No impact on safety and effectiveness. |
| Lift Range<br>(inches) | 0-12" | 0-12" | 0-12" | Identical to predicate device. No impact on safety and effectiveness. |
| Tilt Range<br>(degrees) | 0-50 | 0-50 | 0-50 | Identical to predicate device. No impact on safety and effectiveness. |
| | Subject Device | Predicate Device | Reference Device | |
| | Magic Mobility<br>XT Series Power Wheelchair<br>K233877 | Magic Mobility<br>Extreme X8 Power Wheelchair<br>K211574 | Sunrise Medical (US) LLC<br>Quickie® Q700-UP M<br>K172384 | Statement of Equivalence |
| Recline Range<br>(degrees) | 0-170 | 0-170 | 0-172 | Identical to predicate device. No<br>impact on safety and effectiveness. |
| Suspension | Front and Rear | None | Unknown | Similar to predicate device. No<br>impact on safety and effectiveness. |
| Braking System | Electromagnetic, regenerative brakes with<br>a free-wheeling mode | Electromagnetic, regenerative brakes<br>with a free-wheeling mode | Unknown | Identical to predicate device. No<br>impact on safety and effectiveness. |
| Minimum<br>Braking<br>Distance at<br>Maximum<br>Speed (meters) | 1.7 | 1.7 | Unknown | Identical to predicate device. No<br>impact on safety and effectiveness. |
| User Controller | Joystick and hand control buttons | Joystick and hand control buttons | Joystick and hand control buttons | Identical to predicate device. No<br>impact on safety and effectiveness. |
| Joystick Mount | Fixed mount, height adjustable, swing-<br>away | Fixed mount, height adjustable, swing-<br>away | Fixed mount, height adjustable, swing-<br>away | Identical to predicate device. No<br>impact on safety and effectiveness. |
| Software | R-Net from PGDT | R-Net from PGDT | R-Net from PGDT | Identical to predicate device. No<br>impact on safety and effectiveness. |
| Seat Height<br>(minimum,<br>inches) | 19" | 19" | 16.2" | Identical to predicate device. No<br>impact on safety and effectiveness. |
| Seat Width<br>(inches) | 12"-24" | 16"-24" | 16"-22" | Similar to predicate device. No<br>impact on safety and effectiveness. |
| Armrest | Height adjustable, removable, flip up<br>option | Height adjustable, removable, flip up<br>option | Unknown | Identical to predicate device. No<br>impact on safety and effectiveness. |
| Footrest | 1 or 2 pieces, fixed or flip up, angle and<br>height adjustable rigid footplates, and<br>others | 1 or 2 pieces, fixed or flip up, angle and<br>height adjustable rigid footplates, and<br>others | Unknown | Identical to predicate device. No<br>impact on safety and effectiveness. |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Magic Mobility. The word "magic" is written in a stylized font, with the first two letters in red and the last three letters in black. The word "Mobility" is written in a cursive font below the word "magic". The logo is simple and modern, and it is likely used to represent a company that sells mobility products or services.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Magic Mobility. The word "magic" is written in a stylized font, with the "ma" in red and the "gic" in black. Below the word "magic" is the word "Mobility" in a cursive font. The logo is simple and modern.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for Magic Mobility. The word "magic" is written in a stylized font, with the first half in red and the second half in black. Below the word "magic" is the word "Mobility" written in a cursive font, also in black. The TM symbol is in the bottom right corner of the word "Mobility".
#### SUMMARY OF NON-CLINICAL TESTING
The following tests were performed to demonstrate substantial equivalence based on current industry and FDA recognized standards. The results of these tests indicate that the XT Series Power Wheelchair is substantially equivalent to the predicate device.
| BENCH TESTING | |
|---------------------------|-------------------------------------------------|
| ISO 7176-1 | Static stability |
| ISO 7176-2 | Dynamic stability |
| ISO 7176-3 | Effectiveness of brakes |
| ISO 7176-4 | Energy consumption |
| ISO 7176-5 | Dimensions, mass, and maneuvering space |
| ISO 7176-6 | Maximum speed, acceleration, and deceleration |
| ISO 7176-7 | Measurement of seat and wheel dimensions |
| ISO 7176-8 | Static, impact, and fatigue |
| ISO 7176-9 | Climatic test |
| ISO 7176-10 | Obstacle climbing |
| ISO 7176-11 | Test dummies |
| ISO 7176-14 | Power and control systems for power wheelchairs |
| ISO 7176-15 | Documentation and labeling |
| ISO 7176-16 | Resistance to ignition |
| ISO 7176-19 | Dynamic Test |
| ISO 7176-26 | Vocabulary |
| ISO 7176-30 | Wheelchairs for changing occupant posture |
| EMC AND ELECTRICAL SAFETY | |
| ISO 7176-21 | EMC testing |
| ISO 7176-25 | Batteries and chargers |
| BIOCOMPATIBILITY | |
|------------------|---------------------------------------------------------|
| ISO 10993-1 | Evaluation and testing within a risk management process |
The materials used in the subject device are identical composition, formulation processing, and geometry as the materials found in the Extreme X8 Power Wheelchair by Magic Mobility LLC (K211574) and the Quickie Q700-UP M by Sunrise Medical (US) LLC (K172384) device. Furthermore, the T Series Power Wheelchair has the same type and duration of patient contact as the predicate and reference devices. Therefore, based on previous use, the subject device is considered to have met the requirements of ISO 10993-1 and FDA's Guidance for Industry and Food and Drug Administration Staff - Use of International ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process".
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for Magic Mobility. The word "magic" is written in a stylized font, with the top half of the letters in red and the bottom half in black. The word "Mobility" is written in a cursive font below the word "magic". The logo is simple and modern, and it is likely used to represent a company that provides mobility solutions.
| SOFTWARE | |
|-----------|-----------------------------|
| IEC 62304 | Software life cycle process |
The results of these tests indicate that the XT Series Power Wheelchair is substantially equivalent to the predicate devices.
## CONCLUSION
Based on the testing performed (including wheelchair dynamic testing and flammability) it can be concluded that the subject device does not raise new issues of safety or effectiveness compared to the predicate devices. The similar indications for use, technological characteristics and performance characteristics for the proposed XT Series Power Wheelchair are assessed to be substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.