K233876 · Alcon Laboratories, Inc. · HQC · Jun 21, 2024 · Ophthalmic
Device Facts
Record ID
K233876
Device Name
UNITY VCS (8065000296); UNITY CS (8065000297)
Applicant
Alcon Laboratories, Inc.
Product Code
HQC · Ophthalmic
Decision Date
Jun 21, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.4670
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The UNITY VCS console, when used with compatible devices, is indicated for use during anterior segment (i.e. phacoemulsification and removal of cataracts) and posterior segment (i.e. vitreoretinal) ophthalmic surgery. In addition, with the optional laser this system is indicated for photocoagulation (i.e. vitreoretinal and macular pathologies), iridotomy, and trabeculoplasty procedures. The UNITY CS console, when used with compatible devices, is indicated for use during anterior segment (i.e. phacoemulsification and removal of cataracts) ophthalmic surgery.
Device Story
Multifunctional surgical consoles for anterior and posterior segment ophthalmic surgery; UNITY VCS supports both, UNITY CS supports anterior only. Systems drive handpieces for tissue cutting, lens emulsification, illumination, and diathermy coagulation. Inputs include touchscreen, remote control, and foot controllers; flow/vacuum-controlled aspiration and pressure-controlled irrigation/infusion manage ocular fluid. Optional integrated 532-nm green laser module for UNITY VCS performs photocoagulation, iridotomy, and trabeculoplasty. Used in clinical/hospital settings by ophthalmic surgeons. Output provides real-time surgical control and visualization integration (e.g., NGENUITY 3D, LuxOR, Verion). Benefits include precise fluidics management and integrated surgical capabilities, reducing need for multiple consoles. System uses RFID for accessory recognition and e-Connectivity for data management.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including biocompatibility (ISO 10993), sterilization validation, electromagnetic compatibility (IEC 60601-1-2), electrical/mechanical safety (IEC 80601-2-58), and software verification/validation (IEC 62304).
Indicated for patients undergoing anterior segment (phacoemulsification/cataract removal) and posterior segment (vitreoretinal) ophthalmic surgery. UNITY VCS additionally indicated for photocoagulation, iridotomy, and trabeculoplasty. Prescription use only.
Regulatory Classification
Identification
A phacofragmentation system is an AC-powered device with a fragmenting needle intended for use in cataract surgery to disrupt a cataract with ultrasound and extract the cataract.
CENTURION Vision System with Active Sentry (K161794)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
June 21, 2024
Alcon Laboratories, Inc. Ganesh Balachandar Regulatory Affairs Manager 6201 South Freeway Fort Worth, Texas 76134
Re: K233876
Trade/Device Name: UNITY VCS (8065000296); UNITY CS (8065000297) Regulation Number: 21 CFR 886.4670 Regulation Name: Phacofragmentation system Regulatory Class: Class II Product Code: HQC, HQE, HQF Dated: May 17, 2024 Received: May 20, 2024
Dear Ganesh Balachandar:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
{1}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Claudine H. Krawczyk -S
Claudine Krawczyk Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K233876
Device Name UNITY VCS (8065000296) ; UNITY CS (8065000297)
#### Indications for Use (Describe) UNITY VCS
The UNITY VCS console, when used with compatible devices, is indicated for use during anterior segment (i.e. phacoemulsification and removal of cataracts) and posterior segment (i.e. vitreoretinal) ophthalmic surgery.
In addition, with the optional laser this system is indicated for photocoagulation (i.e. vitreoretinal and macular pathologies), iridotomy, and trabeculoplasty procedures.
#### UNITY CS
The UNITY CS console, when used with compatible devices, is indicated for use during anterior segment (i.e. phacoemulsification and removal of cataracts) ophthalmic surgery.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
In accordance with 21 CFR 807.92, Alcon hereby provides the 510(k) summary for the UNITY VCS and UNITY CS.
# 1. SUBMITTER
| Applicant | Alcon Laboratories, Inc.<br>6201 South Freeway<br>Fort Worth, TX 76134-2099 |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Primary Correspondent | Ganesh Balachandar, M.S.<br>Regulatory Affairs Manager<br>Alcon Research LLC,<br>on behalf of Alcon Laboratories, Inc.<br>20511 Lake Forest Drive<br>Lake Forest, CA 92630-7741<br>+1 (949) 322-5101<br>ganesh.balachandar@alcon.com |
| Secondary Correspondent | Ophelia Biggs<br>Director, Regulatory Affairs<br>Alcon Research LLC,<br>on behalf of Alcon Laboratories, Inc.<br>20511 Lake Forest Drive<br>Lake Forest, CA 92630-7741<br>+1 (949) 217-3854<br>ophelia.biggs@alcon.com |
| Date Prepared | June 21, 2024 |
{5}------------------------------------------------
# 2. SUBJECT DEVICES
| Device Trade Name | UNITY VCS (8065000296);<br>UNITY CS (8065000297) |
|-----------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification<br>Number and Name<br>(Product Code in<br>Parentheses) | For UNITY VCS:<br>• 21 CFR 886.4150, Vitreous aspiration and cutting instrument<br>(HQE)<br>• 21 CFR 886.4670, Phacofragmentation system (HQC)<br>• 21 CFR 886.4690, Ophthalmic laser (HQF)<br>For UNITY CS:<br>• 21 CFR 886.4670, Phacofragmentation system (HQC) |
| Regulatory Class | II |
# 3. PREDICATE DEVICES
| FOR UNITY VCS: | |
|-----------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|
| Predicate Device | CONSTELLATION Vision System |
| 510(k) Number | K141065 |
| Classification<br>Number and Name<br>(Product Code in<br>Parentheses) | 21 CFR 886.4150, Vitreous aspiration and cutting instrument (HQE) 21 CFR 886.4670, Phacofragmentation system (HQC) |
| Regulatory Class | II |
| FOR UNITY VCS (Laser Functionality): | |
|-----------------------------------------------------------------------|--------------------------------------------|
| Predicate Device | Next-Generation Laser |
| 510(k) Number | K062624 |
| Classification<br>Number and Name<br>(Product Code in<br>Parentheses) | • 21 CFR 886.4390, Laser, Ophthalmic (HQF) |
| Regulatory Class | II |
| FOR UNITY CS: | |
|------------------|--------------------------------------------------|
| Predicate Device | CENTURION Vision System with Active Sentry |
| 510(k) Number | K161794 |
| Classification | 21 CFR 886.4670, Phacofragmentation system (HQC) |
{6}------------------------------------------------
| Number and Name<br>(Product Code in<br>Parentheses) | |
|-----------------------------------------------------|----|
| Regulatory Class | II |
{7}------------------------------------------------
#### DEVICE DESCRIPTION 4.
The UNITY VCS is a multifunctional surgical instrument for use in anterior and posterior segment ophthalmic surgeries. The capabilities of the UNITY VCS include driving a variety of handpieces that provide the ability to cut vitreous and tissues, emulsify the crystalline lens, illuminate the posterior segment of the eye, and apply diathermy to stop bleeding. Flow-controlled or vacuumcontrolled aspiration is used to remove ocular matter from the eye. Pressure-controlled irrigation/infusion capability is provided to replace fluid in the eye and enters the eye directly through either an infusion cannula or a handpiece. The graphical user interface is menu-driven. The operator provides inputs using the touchscreen panel, the remote control, and the foot controllers.
An optional, fully integrated laser module is available that can be installed in the base of the UNITY VCS configuration for vitreoretinal surgery. The laser delivers the same visible 532-nm green treatment beam designed for ophthalmic use as in the predicate device.
When used with compatible components and accessories, the system will perform anterior and posterior segment ophthalmic procedures, including irrigation, phacoemulsification, extraction, vitrectomy, diathermy coagulation, intraocular illumination, photocoagulation, viscous fluid control, tissue grabbing and cutting.
The UNITY CS is a multifunctional surgical instrument for use in anterior segment ophthalmic surgeries. The capabilities of the UNITY CS include driving a variety of handpieces that provide the ability to cut anterior vitreous and tissues, emulsify the crystalline lens, and apply diathermy to stop bleeding. Flow-controlled or pressure-controlled aspiration is used to remove ocular matter from the eye. Pressure-controlled irrigation/infusion capability is provided to replace fluid in the eye and enters the eye directly through either an infusion cannula or a handpiece. The graphical user interface is menu-driven, and the operator provides inputs using the touchscreen panel, the remote control, and the foot controllers.
When used with compatible components and accessories, the system will perform anterior segment ophthalmic procedures, including irrigation, phacoemulsification, extraction, vitrectomy, and diathermy coagulation.
The UNITY VCS and UNITY CS are intended to be used within a clinic(s)/hospital(s)/surgical practice network.
Both devices are for prescription use only.
#### INDICATIONS FOR USE ડાં
### UNITY VCS:
The UNITY VCS console, when used with compatible devices, is indicated for use during anterior segment (i.e. phacoemulsification and removal of cataracts) and posterior segment (i.e.
{8}------------------------------------------------
vitreoretinal) ophthalmic surgery.
In addition, with the optional laser this system is indicated for photocoagulation (i.e. vitreoretinal and macular pathologies), iridotomy and trabeculoplasty procedures.
### UNITY CS:
The UNITY CS console, when used with compatible devices, is indicated for use during anterior segment (i.e. phacoemulsification and removal of cataracts) ophthalmic surgery.
{9}------------------------------------------------
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS TO PREDICATE 6.
| Characteristic | CONSTELLATION<br>Vision System<br>(Predicate Device for<br>Subject Device 1) | UNITY VCS<br>(Subject Device 1) | CENTURION Vision<br>System with Active<br>Sentry (Predicate Device<br>for Subject Device 2) | UNITY CS<br>(Subject Device 2) |
|-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Administrative | | | | |
| 510(k) | K141065 | K233876 | K161794 | K233876 |
| Intended Use | Anterior and posterior<br>segment ophthalmic<br>surgery | Same<br>(The subject device,<br>consisting of the console<br>and compatible devices, is<br>intended to facilitate<br>management of fluid and<br>gases, as well as removal,<br>grasping, cutting,<br>illumination, and<br>coagulation of ocular<br>materials)1 | Anterior segment<br>ophthalmic surgery | Same<br>(The subject device,<br>consisting of the console<br>and compatible devices, is<br>intended to facilitate<br>management of fluid as<br>well as removal, cutting,<br>and coagulation of ocular<br>materials)1 |
| Indications for Use | The CONSTELLATION®<br>Vision System is indicated<br>for the following:<br>The CONSTELLATION®<br>Vision System is an<br>ophthalmic microsurgical<br>system that is indicated for<br>both anterior segment (i.e.<br>phacoemulsification and | The UNITY VCS console,<br>when used with<br>compatible devices, is<br>indicated for use during<br>anterior segment (i.e.<br>phacoemulsification and<br>removal of cataracts) and<br>posterior segment (i.e.<br>vitreoretinal) ophthalmic | The Centurion Vision<br>System is indicated for<br>emulsification, separation,<br>irrigation, and aspiration<br>of cataracts, residual<br>cortical material and lens<br>epithelial cells, vitreous<br>aspiration and cutting<br>associated with anterior | The UNITY CS console,<br>when used with<br>compatible devices, is<br>indicated for use during<br>anterior segment (i.e.<br>phacoemulsification and<br>removal of cataracts)<br>ophthalmic surgery. |
| Characteristic | CONSTELLATION<br>Vision System<br>(Predicate Device for<br>Subject Device 1) | UNITY VCS<br>(Subject Device 1) | CENTURION Vision<br>System with Active<br>Sentry (Predicate Device<br>for Subject Device 2) | UNITY CS<br>(Subject Device 2) |
| | removal of cataracts) and<br>posterior segment (i.e.,<br>vitreoretinal) ophthalmic<br>surgery in addition to the<br>indications included with<br>the optional Next-<br>Generation laser. The<br>AutoSert® IOL Injector<br>Handpiece is intended to<br>deliver qualified<br>ACRYSOF® intraocular<br>lenses into the eye<br>following cataract<br>removal.<br>The AutoSertⓇ<br>IOL Injector Handpiece<br>achieves the functionality<br>of injection of intraocular<br>lenses. The AutoSert is<br>indicated for use with<br>ACRYSOF® lenses<br>SN60WF, SN6AD1,<br>SN6AT3 through<br>SN6AT9, as well as<br>approved ACRYSOF®<br>lenses that are specifically<br>indicated for use with this<br>inserter as indicated in the | surgery.<br><br>In addition, with the<br>optional laser this system<br>is indicated for<br>photocoagulation (i.e.<br>vitreoretinal and macular<br>pathologies), iridotomy<br>and trabeculoplasty<br>procedures.2 | vitrectomy, bipolar<br>coagulation, and<br>intraocular lens injection.<br>The AutoSert IOL Injector<br>Handpiece is intended to<br>deliver qualified AcrySof<br>intraocular lenses into the<br>eye following cataract<br>removal.<br><br>The AutoSert IOL Injector<br>Handpiece achieves the<br>functionality of injection<br>of intraocular lenses. The<br>AutoSert IOL Injector<br>Handpiece is indicated for<br>use with the AcrySof<br>lenses SN60WF,<br>SNWAD1, SN6AT3<br>through SN6AT9,<br>as well as approved<br>AcrySof lenses that are<br>specifically indicated for<br>use with this inserter, as<br>indicated in the approved<br>labeling of those lenses. | |
| Characteristic | CONSTELLATION<br>Vision System<br>(Predicate Device for<br>Subject Device 1) | UNITY VCS<br>(Subject Device 1) | CENTURION Vision<br>System with Active<br>Sentry (Predicate Device<br>for Subject Device 2) | UNITY CS<br>(Subject Device 2) |
| | approved labeling of those<br>lenses. | | | |
| Console Features | | | | |
| Monolith | Yes | Yes | Yes | Yes |
| Tray | Yes | Yes | Yes | Yes |
| Touch Panel Display | Yes | Yes | Yes | Yes |
| RFID for accessory<br>connections | Yes | Yes | No | Yes |
| e-Connectivity | Yes | Yes | Yes | Yes |
| User Interface | Touch panel display, foot<br>controllers, remote control | Same | Same | Same |
| Electrical Power<br>Specifications | 100-240 V AC, 50/60 Hz | Same | Same | Same |
| Operating<br>Temperature Range | 10-35°C | Same | Same | Same |
| Console Height,<br>Width and Depth | Height: 62.4"<br>Width: 30.3"<br>Depth: 36" | Height: 64.6"<br>Width: 25.4"<br>Depth:28.3" | Height:64.6"<br>Width:19.6'<br>Depth:31.6" | Height: 64.6"<br>Width: 25.4"<br>Depth:28.3" |
| Maximum Humidity<br>(RH = Relative<br>Humidity) | 95% RH non-condensing | Same | Same | Same |
| Fluidics Features | | | | |
| Push Prime | Yes | Yes | No | Yes |
| Infusion/Irrigation<br>Source Change | Yes | Yes | Yes | Yes |
| Aspiration Pressure<br>Control | Yes | Yes | Yes | Yes |
| Aspiration Flow<br>Control | Yes | Yes | Yes | Yes |
| Characteristic | CONSTELLATION<br>Vision System<br>(Predicate Device for<br>Subject Device 1) | UNITY VCS<br>(Subject Device 1) | CENTURION Vision<br>System with Active<br>Sentry (Predicate Device<br>for Subject Device 2) | UNITY CS<br>(Subject Device 2) |
| Suction Flow<br>Control | Yes | Yes | No | No |
| Flow Control –<br>Irrigation | No | Yes | Yes | Yes |
| Flow Control –<br>Infusion | No | Yes | No | No |
| IOP-Controlled<br>Infusion/Irrigation | Yes | Yes | Yes | Yes |
| Pressurized Infusion<br>Irrigation | Yes | Yes | No | No |
| | Yes | Yes | Yes | Yes |
| Proportional Reflux | Yes | Yes | No…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.