← Product Code [BWF](/productcode/BWF) · K233855

# AllPEP (K233855)

_Enchant Tek Co. , Ltd. · BWF · Feb 2, 2024 · Anesthesiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K233855

## Device Facts

- **Applicant:** Enchant Tek Co. , Ltd.
- **Product Code:** [BWF](/productcode/BWF.md)
- **Decision Date:** Feb 2, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5690
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic

## Indications for Use

The Enchant AllPEP is intended for use as a Positive Expiratory Pressure (PEP) by patients aged 18 years and older. Patients must be capable of generating an exhalation flow of 10 lpm for 3-4 seconds to promote secretion clearance. This device may be used to prevent or reverse atelectasis and as an inspiratory, deep breathing positive exerciser. It may be used in hospital and clinical settings.

## Device Story

AllPEP is a Positive Expiratory Pressure (PEP) device used in hospital and clinical settings to promote secretion clearance and prevent/reverse atelectasis. Device consists of housing, top cover, and internal swinger mechanism. Patient inhales room air through one-way valve; during exhalation, air is directed through upper portion of device where swinger mechanism oscillates air, creating vibration. Device provides PEEP and oscillations during exhalation. Healthcare providers use device to assist patients in respiratory therapy; output is physical vibration and pressure during exhalation. Benefits include improved airway clearance and lung expansion. Device is single-patient, multi-use, intended for replacement every 30 days.

## Clinical Evidence

Bench testing only. Testing included aging/post-aging performance, cleaning validation, post-cleaning performance, frequency (Hz), amplitude (cmH2O), and pressure at min/max settings. Biocompatibility testing performed per ISO 10993-1 (cytotoxicity, sensitization, intracutaneous reactivity, material-mediated pyrogenicity, chemical characterization) and ISO 18562 (particulate matter, VOCs). Results met pre-defined acceptance criteria.

## Technological Characteristics

PEP device featuring swinger mechanism for oscillation and one-way valve for inhalation. Materials meet ISO 10993-1 for external communicating, mucosal membrane contact, prolonged duration (>24h, <30 days). Performance metrics: frequency range 9.82-16.32 Hz and amplitude 3.24-9.17 cmH2O across 5-30 lpm flow rates. Mechanical operation; no energy source required. Single-patient, multi-use; cleaned with neutral dishwashing agent.

## Regulatory Identification

An incentive spirometer is a device that indicates a patient's breathing volume or flow and that provides an incentive to the patient to improve his or her ventilation.

## Predicate Devices

- DHD Acapella ([K991561](/device/K991561.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.

February 2, 2024

Enchant Tek Co. Ltd. Leonard Ling QA/RA Supervisor No. 210 Xiangzhong Rd. Dongshan Township Yilan County, 26950 Taiwan

Re: K233855

Trade/Device Name: AllPEP Regulation Number: 21 CFR 868.5690 Regulation Name: Incentive Spirometer Regulatory Class: Class II Product Code: BWF Dated: December 4, 2023 Received: December 5, 2023

Dear Leonard Ling:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Rachana Visaria -S

Rachana Visaria, Ph.D. Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and

{2}------------------------------------------------

Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known) K233855

Device Name AllPEP

Indications for Use (Describe)

The Enchant AllPEP is intended for use as a Positive Expiratory Pressure (PEP) by patients aged 18 years and older. Patients must be capable of generating an exhalation flow of 10 lpm for 3-4 seconds to promote secretion clearance. This device may be used to prevent or reverse atelectasis and as an inspiratory, deep breathing positive exerciser. It may be used in hospital and clinical settings.

| Type of Use (Select one or both, as applicable)                                    |                                                                                   |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| <span style="font-size:12px">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="font-size:12px">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

| Date Prepared:             | February 2, 2024                                                                                                               |
|----------------------------|--------------------------------------------------------------------------------------------------------------------------------|
| Sponsor:                   | Enchant Tek Co. Ltd.<br>No. 210 Xiangzhong Rd.<br>Dongshan Township, Yilan County<br>Taipei, TW 26950<br>Tel +886- 03-959-3500 |
| Sponsor Contact:           | Ling Leonard - QA/RA Supervisor                                                                                                |
| Submission Correspondent:  | Paul Dryden<br>ProMedic, LLC                                                                                                   |
| Proprietary or Trade Name: | AllPEP                                                                                                                         |
| Common/Usual Name:         | PEP device                                                                                                                     |
| Classification Name:       | Spirometer, therapeutic (incentive)                                                                                            |
| Product Code:              | 21CFR 868.5690<br>BWF                                                                                                          |
| Predicate Device:          | DHD Acapella K991561                                                                                                           |
| Common/Usual Name:         | Spirometer, therapeutic (incentive)                                                                                            |
| Classification Name:       | 21CFR 868.5690                                                                                                                 |
| Product Code:              | BWF                                                                                                                            |

Device Description: AllPEP provides PEEP during user exhalation and features a mechanism which provides oscillations during exhalation as well.

Principle of Operation: The oscillating positive exhalation pressure device, AllPEP, includes a housing, a top cover, and an oscillating mechanism. The patient inhales room air through a one-way valve and when exhaling the exhaled gas is directed through the upper portion of the device which contains a swinger mechanism to oscillate the air, causing the air to vibrate.

### Indications for Use:

The Enchant AllPEP is intended for use as a Positive Expiratory Pressure (PEP) by patients aged 18 years and older. Patients must be capable of generating an exhalation flow of 10 lpm for 3-4 seconds to promote secretion clearance. This device may be used to prevent or reverse atelectasis and as an inspiratory, deep breathing positive exerciser. It may be used in hospital and clinical settings.

### Patient Population:

Patients 18 years and older who are capable of generating an exhalation flow of 10 1pm for 3-4 seconds.

{5}------------------------------------------------

### Environments of use:

For hospital and clinical settings.

### Substantial Equivalence Discussion

The table below compares the key features of the proposed with the identified predicate - DHD Acapella K991561. The comparison demonstrates that the subject device is substantially equivalent to the predicate device.

#### . Indications for Use -

The indications for use are similar for the subject device when compared to the predicate device.

Discussion - Since the predicate, newer cleared PEP devices have more details within the indications for use. We have added this information, it does not however raise different questions of safety or effectiveness.

#### . Technology and construction -

The technology and principle of operation is similar for the subject device when compared to the predicate device.

Discussion - While the predicate may have a different means of creating oscillation, there are a number of cleared PEP devices which also have slightly different ways of creating oscillations during the exhalation cycle. This difference does not raise different concerns of safety or effectiveness.

#### . Environment of Use -

The environments of use are similar to predicate which are hospital and clinical settings. Discussion - The environments of use are similar.

#### Patient Population -.

The patient population for the subject device is patients requiring secretion clearance and those who may have atelectasis. They should be capable of generating an exhalation flow of 10 lpm for 3-4 seconds.

Discussion - The subject and predicate device patient populations are similar.

### Non-Clinical Testing Summary -

### Bench testing -

The following tests were performed to demonstrate substantial equivalence:

- Aging and post-aging performance ●
- Cleaning validation ●
- Post-cleaning performance
- Frequency (Hz)
- Amplitude (cmH20) ●
- Pressure at min / max settings (cmH20) 0

Discussion - The test results are similar to the predicate and within pre-defined acceptance criteria.

{6}------------------------------------------------

## Biocompatibility -

According to ISO 10993-1 and the FDA Guidance related to biocompatibility from 2020, the subject device is considered:

- External Communicating
- Tissue Contacting
- Prolonged duration of use (>24 hours, <30 days) o

For those materials in direct contact, they would be considered:

- Surface Device
- Mucosal Membrane ●
- o Prolonged duration of use (>24 hours, <30 days)

Based upon ISO 10993-1, the sponsor has performed the following biocompatibility tests for the subject device:

- ISO 10993-5:2009 - Cytotoxicity
- ISO 10993-10:2010 Sensitization and Intracutaneous Reactivity ●
- ISO 10993-11:2017 Material-Mediated Pyrogenicity ●
- ISO 10993-18:2020 Chemical Characterization and Risk Assessment ●
- ISO 18562-2:2018 Particulate Matter testing ●
- ISO 18562-3:2018 Volatile Organic Compounds (VOC) with risk assessment ●

Discussion - The subject device was found to meet the applicable requirements for biocompatibility safety for the intended population.

### Discussion of Differences

There are no significant differences in indications for use, population, use environments, technological characteristics or performance which raise different concerns of safety compared to the predicate and thus should be found to be substantially equivalent.

### Substantial Equivalence Conclusion

Through performance testing, design, and non-clinical testing the proposed device and predicate have been found to be substantially equivalent.

{7}------------------------------------------------

|                                           | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                      | Predicate Device                                                                                                                                                                                                                                                                                                                           | Subject and Predicate Comparison                                                                                                                                                                |
|-------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company                                   | Enchant Tek Co., Ltd.                                                                                                                                                                                                                                                                                                                                                                                                               | DHD Acapella                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                 |
| 510(k)                                    | K230622                                                                                                                                                                                                                                                                                                                                                                                                                             | K991561                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                 |
| Classification No. &<br>Product code      | 868.5690<br>BWF - spirometer, therapeutic (incentive)                                                                                                                                                                                                                                                                                                                                                                               | 868.5690<br>BWF - spirometer, therapeutic (incentive)                                                                                                                                                                                                                                                                                      | Identical                                                                                                                                                                                       |
| Indications for Use                       | The Enchant AllPEP is intended for use as a<br>Positive Expiratory Pressure (PEP) by<br>patients 18 years and older. Patients must be<br>capable of generating an exhalation flow of<br>10 lpm for 3-4 seconds to promote secretion<br>clearance. This device may be used to<br>prevent or reverse atelectasis and as an<br>inspiratory, deep breathing positive<br>exerciser. It may be used in hospital and<br>clinical settings. | Indicated for use as a Positive Expiratory<br>Pressure (PEP) Device<br>improves clearance of secretions<br>may reduce the need for postural drainage<br>facilitates opening of airways in patients<br>with Cystic Fibrosis, COPD, asthma, and<br>lung diseases with secretory problems<br>may be used to prevent or reverse<br>atelectasis | Similar<br>More recent PEP clearances have<br>included the environment of use and<br>target population in the indications<br>statement. This does not change the<br>intended use.               |
| Environments of Use                       | Hospital and clinical settings                                                                                                                                                                                                                                                                                                                                                                                                      | Hospital, clinical and home care settings                                                                                                                                                                                                                                                                                                  | Similar, but not for home use                                                                                                                                                                   |
| Prescriptive                              | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                 | Yes                                                                                                                                                                                                                                                                                                                                        | Similar                                                                                                                                                                                         |
| Patient Population                        | Patients requiring PEP therapy aged 18<br>years and older.<br>Patient capable of generating an exhalation<br>flow of 10 lpm for 3-4 seconds                                                                                                                                                                                                                                                                                         | Patients requiring PEP therapy                                                                                                                                                                                                                                                                                                             | Similar                                                                                                                                                                                         |
| Single Patient, multi-use                 | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                 | Yes                                                                                                                                                                                                                                                                                                                                        | Similar                                                                                                                                                                                         |
| Patient Interface                         | Mouthpiece                                                                                                                                                                                                                                                                                                                                                                                                                          | Mouthpiece                                                                                                                                                                                                                                                                                                                                 | Similar                                                                                                                                                                                         |
| Basic Components                          | Swinging Flap valve which generates<br>oscillation during exhalation<br>One-way valve for inhalation<br>Mouthpiece                                                                                                                                                                                                                                                                                                                  | Rocker which generates oscillation during<br>exhalation<br>One-way valve for inhalation<br>Mouthpiece                                                                                                                                                                                                                                      | Similar<br>The principle of operation is the<br>oscillation during airflow. This is similar<br>to the predicate device and does not raise<br>different questions of safety and<br>effectiveness |
| Mean frequency range<br>(Hz)              | 5lpm - 9.82<br>10lpm - 9.96<br>15lpm - 11.85<br>20lpm - 13.01<br>25lpm - 14.89<br>30lpm - 16.32                                                                                                                                                                                                                                                                                                                                     | Not published                                                                                                                                                                                                                                                                                                                              | Similar based on bench testing of subject<br>device and predicate                                                                                                                               |
|                                           | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                      | Predicate Device                                                                                                                                                                                                                                                                                                                           | Subject and Predicate Comparison                                                                                                                                                                |
| Mean amplitude<br>Pressure (cmH20)        | 5lpm - 3.24<br>10lpm - 6.31<br>15lpm - 8.31<br>20lpm - 9.42<br>25lpm - 9.12<br>30lpm - 9.17                                                                                                                                                                                                                                                                                                                                         | Not published                                                                                                                                                                                                                                                                                                                              | Similar based on bench testing of subject<br>device and predicate                                                                                                                               |
| Lowest Pressure (cmH20)                   | 5lpm - 0.50<br>10lpm - 0.78<br>15lpm - 2.09<br>20lpm - 4.00<br>25lpm - 7.26<br>30lpm - 10.75                                                                                                                                                                                                                                                                                                                                        | Not published                                                                                                                                                                                                                                                                                                                              | Similar based on bench testing of subject<br>device and predicate                                                                                                                               |
| Highest Pressure<br>(cmH20)               | 5lpm - 3.74<br>10lpm - 7.10<br>15lpm - 10.40<br>20lpm - 13.42<br>25lpm - 16.38<br>30lpm - 19.92                                                                                                                                                                                                                                                                                                                                     | Not published                                                                                                                                                                                                                                                                                                                              | Similar based on bench testing of subject<br>device and predicate                                                                                                                               |
| Cleaning method                           | Neutral dishwashing agent<br>Single patient, multi-use                                                                                                                                                                                                                                                                                                                                                                              | Cleaning instructions not provided<br>Single patient, multi-use                                                                                                                                                                                                                                                                            | Similar based on bench testing of subject<br>device and predicate                                                                                                                               |
| Performance testing                       | Aging and post-aging performance<br>Cleaning and post-cleaning<br>performance<br>Cleaning validation<br>Transportation - shock and vibration<br>Frequency (Hz)<br>Amplitude (cmH2O)<br>Pressure at min / max settings (cmH2O)                                                                                                                                                                                                       | Frequency (Hz)<br>Amplitude (cmH2O)<br>Pressure at min I max settings (cmH2O)                                                                                                                                                                                                                                                              | Similar                                                                                                                                                                                         |
| Type of patient contact<br>with materials | Externally communicating/Tissue Surface<br>contact /mucosa<br>Prolonged duration (>24 hours, < 30 days)                                                                                                                                                                                                                                                                                                                             | Not published                                                                                                                                                                                                                                                                                                                              | Patients are instructed to replace device<br>every 30 days.                                                                                                                                     |

{8}------------------------------------------------

---

**Source:** [https://fda-staging.innolitics.com/device/K233855](https://fda-staging.innolitics.com/device/K233855)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).