K233827 · Oxiwear, Inc. · DQA · Aug 28, 2024 · Cardiovascular
Device Facts
Record ID
K233827
Device Name
OxiWear
Applicant
Oxiwear, Inc.
Product Code
DQA · Cardiovascular
Decision Date
Aug 28, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Indications for Use
The OxiWear system is indicated for non-invasive, spot-checking and/or continuous data collection of adult (22+) and adolescent (12-21) patients who are well or poorly perfused, during motion conditions. It is intended for use in hospitals, medical facilities, home healthcare environments, and mobile environments. The recommended application site is the helix of the ear.
Device Story
OxiWear is a wireless, reusable ear pulse oximeter measuring SpO2 and pulse rate via red and infrared light transmission/reflection through the ear helix. The system comprises a wearable device and a mobile application. The wearable attaches to the ear, powered by an internal rechargeable lithium-ion battery. It transmits data via Bluetooth to a smartphone running the OxiWear Mobile Application. The app provides historical trending, live monitoring, and text alerts, serving as a data repository for physician/caregiver review. Used in clinical, home, and mobile environments by patients or caregivers. Output allows healthcare providers to monitor patient oxygenation and pulse rate, supporting clinical decision-making for well or poorly perfused patients.
Clinical Evidence
Clinical testing included controlled laboratory SpO2 validation against a co-oximeter (no motion, normal perfusion) and treadmill-based motion testing for SpO2 and pulse rate against a reference pulse oximeter. Results demonstrate accuracy comparable to industry-standard pulse oximeters. Bench testing confirmed performance under low perfusion and no-motion conditions.
Technological Characteristics
Materials: Silicone (SH516U, SH0130U, MasterSil 151MED), Polyamide (AKULON CARE K1U). Sensing: Red and Infrared LEDs. Power: Internal rechargeable lithium-ion battery. Form factor: Ear-worn wearable (~50g). Connectivity: Bluetooth (wearable to smartphone), Wi-Fi (via smartphone). Standards: ISO 80601-2-61, IEC 60601-1, IEC 60601-1-2, IEC 62304, ISO 10993-1/5/10/21.
Indications for Use
Indicated for non-invasive, spot-checking and/or continuous data collection of adult (22+) and adolescent (12-21) patients, well or poorly perfused, during motion and non-motion conditions. Recommended site: helix of the ear. Intended for hospital, medical facility, home, and mobile environments.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
The Rad-G Pulse Oximeter and Accessories (K201770)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. Next to that is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 28, 2024
OxiWear, Inc. % Rita King Chief Executive Officer MethodSense. Inc. 1 Copley Parkway, Suite 130 Morrisville, North Carolina 27560
Re: K233827
Trade/Device Name: OxiWear Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA, DPZ Dated: July 26, 2024 Received: July 29, 2024
Dear Rita King:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products: and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.
{2}------------------------------------------------
See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Bradley Q. Quinn -S
Bradley Quinn Assistant Director DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K233827
Device Name OxiWear
Indications for Use (Describe)
The OxiWear system is indicated for non-invasive, spot-checking and/or continuous data collection of adult (22+) and adolescent (12-21) patients who are well or poorly perfused, during motion conditions. It is intended for use in hospitals, medical facilities, home healthcare environments, and mobile environments. The recommended application site is the helix of the ear.
Type of Use (Select one or both, as applicable)
| <span>☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-------------------------------------------------------------|
| <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# Summary of 510(k)
# OxiWear, Inc. K233827
This 510(k) Summary is in conformance with 21CFR 807.92
| Submitter: | OxiWear Inc.<br>1111 Arlington Blvd<br>#305<br>Arlington, VA 22209<br>Phone: 571-212-7526 |
|--------------------------------|--------------------------------------------------------------------------------------------------------------------|
| Primary Contact: | Rita King, CEO<br>MethodSense, Inc.<br>Email: ritaking@methodsense.com<br>Phone: 919-313-3961<br>Fax: 919-313-3979 |
| Company Contact: | Shavini Fernando<br>Chief Executive Officer |
| Date Prepared: | July 26, 2024 |
| Device Name and Classification | |
| Trade Name: | OxiWear |
| Common Name: | Oximeter |
| Classification: | Class II |
| Regulation Number: | 21 CFR Part 870.2700 – Oximeter |
| Classification Panel: | Cardiovascular |
| Product Code: | DQA |
| Subsequent Product Code: | DPZ |
Predicate Device:
{5}------------------------------------------------
| | Primary Predicate | Reference Device |
|---------------------------|----------------------------------------------------------------------------|------------------------------------------|
| Trade Name | Models 8100Q2, 8101Q2, and 8102Q2 Reusable, Ear Clip Pulse Oximeter Sensor | The Rad-G Pulse Oximeter and Accessories |
| Common Name | Pulse Oximeter Sensor | Oximeter |
| 510(k) Submitter / Holder | Nonin Medical, Inc. | Masimo Corporation |
| 510(k) Number | K160865 | K201770 |
| Regulation Number | 21 CFR 870.2700 | 21 CFR 870.2700 |
| Classification Panel | Cardiovascular | Cardiovascular |
| Product Code | DQA, DPZ | DQA, DPZ, BZQ |
{6}------------------------------------------------
### Device Description and Intended Use
The OxilVear System is an ear pulse oximeter that non-invasively measures and displays patient blood oxygen saturation (SpO2) and Pulse Rate for spot-checking and/or continuous data collection. The OxiWear System is intended to be used with well or poorly perfused patients of all skin types in both motion conditions. The system is intended for use in clinical and home environments for adult (22+) and adolescent (12-21) patients of all skin types who are well or poorly perfused, and can safely wear the oximeter device on the helix of the OxilVear System can be used with patients that are in both motion and stationary conditions.
The OxilVear System contains the OxilVear Wearable, and the OxiMear Mobile Application. The OxiMear Wearable is a wireless, reusable, non-invasive ear pulse oximeter that is intended to measure SpO2 and Pulse Rate based on the amount of transmitted, reflected light through the ear. The OxiWear Wearable is powered by an internal rechargeable lithium ion battery, and attaches directly to the ear. The OxilVear Wearable is intended to be recharged by the OxiWear Charger, which is a rechargeable portable charger,
The OxiMear Wearable is designed to be used with the OxiMeation which receives data from the OxiMear Wearable via Bluetooth connection to the user's smart phone. The OxiMear Mobile Application acts as a historical trending, live data monitoring, and text alert application is intended to be used alongside the OxiWear Wearable during daily use, as well as a data repository to refer back to previous reports containing information about the patient, pulse rate, patient contacts, and more, to present to a physician or caregiver for review.
#### Indications for Use
The OxilWear system is indicated for non-invasive, spot-checking and/or continuous data collection of adult (12-21) patients who are well or poorly perfused, during motion conditions. It is intended for use in hospitals, medical facilities, home healthcare environments. The recommended application site is the helix of the ear.
# Risk Analysis Method
The OxilVear system was assessed to determine risks to health associated with the use of the device. Risks related to safety and performance were considered. A risk analysis was conducted in accordance with ISO 14971:2019. Medical devices – Application of risk management to medical devices.
#### Substantial Equivalence
The table below provides a detailed comparison of OxiWear to the predicate device.
{7}------------------------------------------------
| Characteristic | Subject Device<br>OxiWear | Predicate Device<br>Models 8100Q2, 8101Q2, and<br>8102Q2 Reusable, Ear Clip Pulse<br>Oximeter Sensor<br>(K160865) | Reference Device<br>Rad-G Pulse Oximeter and<br>Accessories<br>(K201770) | Comparison |
|----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | | | | |
| Intended Use | OxiWear is intended for non-<br>invasive spot-checking and/or<br>continuous data collection of<br>patients. | Models 8100Q2, 8101Q2, and<br>8102Q2 Reusable, Ear Clip Pulse<br>Oximeter Sensor is intended for<br>non-invasive, spot-checking and/or<br>continuous monitoring of patients. | The Rad-G Pulse Oximeter and<br>Accessories are intended for the<br>non-invasive spot-checking or<br>continuous monitoring of patients. | Identical as it relates to<br>spot-checking and<br>continuous data collection. |
| Indications for Use | | | | |
| Indications for Use | OxiWear is indicated for non-<br>invasive spot-checking and/or<br>continuous data collection of adult<br>(22+) and adolescent (12-21)<br>patients who are well or poorly<br>perfused, during motion and non-<br>motion conditions. It is intended<br>for use in hospitals, medical<br>facilities, home healthcare<br>environments, and mobile<br>environments. The recommended<br>application site is the helix of the<br>ear. | Models 8100Q2, 8101Q2, and<br>8102Q2 Reusable, Ear Clip Pulse<br>Oximeter Sensor is indicated for<br>non-invasive, spot-checking and/or<br>continuous monitoring of adult<br>and pediatric patients (> 40 kg / 88<br>lb) who are well or poorly<br>perfused, during non-motion<br>conditions. It is intended for use in<br>hospitals, medical facilities,<br>Emergency Medical Service (EMS)<br>environments, home healthcare<br>environments, and mobile<br>environments. The recommended<br>application site is the earlobe. | The Rad-G Pulse Oximeter and<br>Accessories are intended for the<br>non-invasive spot-checking or<br>continuous monitoring of<br>functional oxygen saturation of<br>arterial hemoglobin (SpO2), Pulse<br>Rate (PR), and Pleth Respiration<br>Rate (RRp).<br>The Rad-G Pulse Oximeter and<br>Accessories are indicated for<br>noninvasive spot-checking or<br>continuous monitoring of<br>functional oxygen saturation of<br>arterial hemoglobin (SpO₂) and | Different – This difference<br>does not change the<br>intended use of the device<br>as it relates to the spot-<br>checking and continuous<br>data collection. |
| Characteristic | Subject Device<br>OxiWear | Predicate Device<br>Models 8100Q2, 8101Q2, and<br>8102Q2 Reusable, Ear Clip Pulse<br>Oximeter Sensor<br>(K160865) | Reference Device<br>Rad-G Pulse Oximeter and<br>Accessories<br>(K201770) | Comparison |
| | | | Pulse Rate (PR) of adult, pediatric,<br>infant, and neonate patients<br>during both no motion and motion<br>conditions, and for patients who<br>are well or poorly perfused in<br>hospitals, hospital-type facilities,<br>transport, and home<br>environments. | |
| | | | The Rad-G Pulse Oximeter and<br>Accessories are indicated for the<br>spot-checking or continuous<br>monitoring of Respiration Rate<br>from the photoplethysmogram<br>(RRp) of adult and pediatric<br>patients during no motion<br>conditions in hospitals, hospital-<br>type facilities, transport, and home<br>environments. | |
| Regulation and<br>Product Code | 21 CFR 870.2700<br>DQA, DPZ | 21 CFR 870.2700<br>DQA, DPZ | 21 CFR 870.2700<br>DQA, DPZ, BZQ | Identical to the predicate<br>device. |
| Characteristic | Subject Device<br>OxiWear | Predicate Device<br>Models 8100Q2, 8101Q2, and<br>8102Q2 Reusable, Ear Clip Pulse<br>Oximeter Sensor<br>(K160865) | Reference Device<br>Rad-G Pulse Oximeter and<br>Accessories<br>(K201770) | Comparison |
| Environment of<br>Use | No motion and motion conditions<br>in hospitals, hospital-type facilities,<br>home healthcare environments,<br>and mobile environments. | Non-motion conditions in<br>hospitals, hospital-type facilities,<br>Emergency Medical Service (EMS)<br>environments, home healthcare<br>environments, and mobile<br>environments. | No motion and motion conditions<br>in hospitals, hospital-type facilities,<br>transport, and home<br>environments. | Equivalent to the<br>combination of the<br>predicate and reference<br>devices. The only<br>difference being that the<br>predicate also includes<br>Emergency Medical Service<br>environments whereas the<br>subject device does not. |
| Patient Population | Adult (22+) and adolescent (12-21)<br>patients. | Adult and pediatric patients (> 40<br>kg / 88 lb) | Adult, pediatric, infant, and<br>neonate patients. | Different – This difference<br>in patient population does<br>not change the intended<br>use of the device. |
| Application Site | Helix of the ear | Earlobe | Fingertip | Different – This difference<br>in application site does not<br>change the intended use of<br>the device. |
| Display | | | | |
| Display Type | LCD Touchscreen | LED | LCD Touchscreen | Identical to the reference<br>device. |
| Notifications | | | | |
| Characteristic | Subject Device<br>OxiWear | Predicate Device<br>Models 8100Q2, 8101Q2, and<br>8102Q2 Reusable, Ear Clip Pulse<br>Oximeter Sensor<br>(K160865) | Reference Device<br>Rad-G Pulse Oximeter and<br>Accessories<br>(K201770) | Comparison |
| Type of<br>Notification | Haptic<br>Notifications | Visual, Audible Alarms | Visual, Audible Alarms | Different – This difference<br>does not change the<br>intended use of the device<br>as it relates to spot-<br>checking and continuous<br>data collection. Continuous<br>monitoring is a more<br>focused use with a higher<br>risk profile. |
| Technological Characteristics | | | | |
| Measured<br>Parameters | SpO2, PR | SpO2, PR | SpO2, PR, RRp | Identical to the predicate<br>device. |
| Materials | SH516U (SHORE 60A) [Silicone]<br>MasterSil 151MED QUART KIT<br>[Silicone]<br>SH0130U SILICONE (SHORE 30A)<br>AKULON CARE K1U (PA6)<br>[Polyamide] | Specific materials unknown;<br>Biocompatible per ISO 10993-1 | Specific materials unknown;<br>Biocompatible per ISO 10993-1 | Different – This difference<br>does not change the<br>intended use of the device. |
| Sensor Type | Red and Infrared LEDs | Red and Infrared LEDs | Red and Infrared LEDs | Identical to the predicate<br>device. |
| | | Predicate Device | Reference Device | |
| Characteristic | Subject Device<br>OxiWear | Models 8100Q2, 8101Q2, and<br>8102Q2 Reusable, Ear Clip Pulse<br>Oximeter Sensor<br>(K160865) | Rad-G Pulse Oximeter and<br>Accessories<br>(K201770) | Comparison |
| Principle of<br>Operation | The pulse oximeter determines<br>functional oxygen saturation of<br>arterial hemoglobin (SpO2) by<br>measuring the absorption of red<br>and infrared light passing through<br>perfused tissue. Changes in the<br>absorption caused by the pulsation<br>of blood in the vascular bed are<br>used to determine oxygen<br>saturation and pulse rate. | The pulse oximeter determines<br>functional oxygen saturation of<br>arterial hemoglobin (SpO2) by<br>measuring the absorption of red<br>and infrared light passing through<br>perfused tissue. Changes in the<br>absorption caused by the pulsation<br>of blood in the vascular bed are<br>used to determine oxygen<br>saturation and pulse rate. | The pulse oximeter determines<br>functional oxygen saturation of<br>arterial hemoglobin (SpO2) by<br>measuring the absorption of red<br>and infrared light passing through<br>perfused tissue. Changes in the<br>absorption caused by the pulsation<br>of blood in the vascular bed are<br>used to determine oxygen<br>saturation and pulse rate. | Identical to the predicate<br>device. |
| Performance Specification | | | | |
| SpO2<br>Measurement<br>Range | 70-100% | 70-100% | 70-100% | Identical to the predicate<br>device. |
| SpO2, no motion | ± 3.5% (ARMS) | ± 3% (ARMS) | Adults/Pediatrics/Infants: ± 2%<br>(ARMS)<br>Neonates: ± 3% (ARMS) | Different - This difference<br>does not change the<br>intended use of the device. |
| SpO2, motion | ± 3.5% (ARMS) | N/A | ± 3% (ARMS) | Different - This difference<br>does not change the<br>intended use of the device. |
| Characteristic | Subject Device<br>OxiWear | Predicate Device<br>Models 8100Q2, 8101Q2, and<br>8102Q2 Reusable, Ear Clip Pulse<br>Oximeter Sensor<br>(K160865) | Reference Device<br>Rad-G Pulse Oximeter and<br>Accessories<br>(K201770) | Comparison |
| SpO2, low<br>perfusion | $\pm 3.5\% (A_{RMS})$ | $\pm 2\% (A_{RMS})$ | Adults/Pediatrics/Infants: $\pm 2\% (A_{RMS})$<br>Neonates: $\pm 3\% (A_{RMS})$ | Different - This difference<br>does not change the<br>intended use of the device. |
| Pulse Rate<br>Measurement<br>Range | 20-240 BPM | 25-240 BPM | 25-240 BPM | Different - This difference<br>does not change the<br>intended use of the<br>device. |
| Pulse Rate, no<br>motion | $\pm 3 \text{ BPM } (A_{RMS})$ | $\pm 3 \text{ BPM } (A_{RMS})$ | $\pm 3 \text{ BPM } (A_{RMS})$ | Identical to the predicate<br>device. |
| Pulse Rate, motion | $\pm 5 \text{ BPM } (A_{RMS})$ | N/A | $\pm 5 \text{ BPM } (A_{RMS})$ | Identical to the reference<br>device. |
| Pulse Rate, low<br>perfusion | $\pm 3 \text{ BPM } (A_{RMS})$ | $\pm 3 \text{ BPM } (A_{RMS})$ | $\pm 3 \text{ BPM } (A_{RMS})$ | Identical to the predicate<br>device. |
| Environmental | | | | |
| Operating<br>Temperature | 0 to 40 °C (32 to 104 °F) | -20 to 50 °C (-4 to 122 °F) | 0 to 40 °C (32 to 104 °F) | Identical to the reference<br>device. |
| Storage<br>Temperature | -20 to 60 °C (-4 to 140 °F) | -40 to 70 °C (-40 to 158 °F) | -20 to 60 °C (-4 to 140 °F) | Equivalent to the reference<br>device. |
| Characteristic | Subject Device<br>OxiWear | Predicate Device<br>Models 8100Q2, 8101Q2, and<br>8102Q2 Reusable, Ear Clip Pulse<br>Oximeter Sensor<br>(K160865) | Reference Device<br>Rad-G Pulse Oximeter and<br>Accessories<br>(K20177…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.